Orthopaedic Proceedings
Vol. 87-B, Issue SUPP_III | Pages 321 - 321
1 Sep 2005
MacDonald S Bourne R Rorabeck C Cleland D Chess D McCalden R
Full Access

Introduction and Aims: Polyethylene wear continues to be the most significant issue following total hip arthroplasty (THA), leading to the current increase in use of alternative bearing surfaces. We performed a prospective, randomised, blinded clinical trial comparing metal versus polyethylene bearing surfaces in patients receiving THA. Method: Forty-one patients were randomised to receive a metal (23) or a polyethylene (18) insert with identical femoral and acetabular components. Patients were evaluated pre-operatively at three, six, 12 months and annually thereafter, including an evaluation of erythrocyte and 24-hour urine cobalt, chromium and titanium metal ion levels, validated outcome measures (WOMAC, SF-12, Harris Hip Score) and radiographs. Results: No patients were lost to follow-up. One patient died of unrelated causes. At an average 4.4 (range 3.3–5.1) years follow-up there were no differences in any outcome measures or radiographic findings. Patients receiving metal liners had significantly elevated metal ion measurements. At the latest follow-up erythrocyte cobalt levels were seven times elevated (median 1.2μg/l (metal) Vs 0.18μg/l (poly), p< .001). Urine cobalt levels were 41 times elevated (median 11.9μg/day (metal) Vs 0.29μg/day (poly), p< .001) and urine chromium levels were 14 times elevated (median 4.9μg/day (metal) Vs 0.36μg/day (poly), p< .001). Erthrocyte chromium, titanium and urine titanium were not significantly different between groups. As well, contrary to previous reports, there was not a significant trend towards decreasing ion levels over time. Conclusion: In this prospective randomised blinded clinical trail comparing metal to polyethyle bearing surfaces, both cobalt and chromium ion levels were significantly elevated in the blood and urine of the patients randomised to receive the metal on metal THA


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_29 | Pages 57 - 57
1 Aug 2013
Schepers A Nwokeyi N van der Jagt D
Full Access

Purpose of Study:. To look at the difference in metal ion levels using Ceramic on Metal (COM) and Metal on Metal (MOM) bearings in Total Hip Replacements, comparing the results between well placed and poorly placed cups. Methods:. Metal ion levels using the ICP-MS method of assay have been studied as part of a prospective randomised trial between 4 different bearing combinations. The ion levels have been raised in the COM and MOM bearings. Metal ions are not raised in the Ceramic on Ceramic (COC) and Ceramic on X linked Poly(COP) bearings, showing that no other material in the study prosthesis aside from the bearing surfaces contribute to raised ion levels. Cup inclination and cup anteversion has been measured in all patients. Results:. At 3 year follow-up the Cobalt and Chrome levels are significantly lower with the COM bearing than the MOM bearing when both well and poorly placed cups are included, and seem to be returning to normal pre-operative levels. Taking cups put in in Lewinnik's safe zones, then the Chrome levels are similar between the two bearings, but the Cobalt levels remain raised in the MOM bearing. Conclusion:. Cobalt and Chrome levels are significantly lower in COM bearings than MOM bearings when cups are not placed in the ideal position. This represents an added advantage in the use of COM bearings as compared to MOM bearings


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_29 | Pages 59 - 59
1 Aug 2013
Myburgh J Monni T
Full Access

Purpose of the study:. Recent literature raised concern on the adverse effects, incidence of pseudo tumours and raised blood and urine Chromium and Cobalt levels in patients with a metal on metal articulation. The purpose of this study was to follow up a cohort of patients that received a metal on metal total hip replacement with minimum follow up of ten years and to compare the results with the literature. Methods:. All the available patients that underwent metal on metal total hip replacements during 1999 to 2000 with a minimum follow up of 10 years were clinically examined and evaluated with the Harris Hip Score; X-rays; Full Blood Count; CRP; LFT; U, K & E; blood and urine Chromium and Cobalt levels and Ultrasonic examination of the hip. The ultrasound was done by an independent experienced ultrasonographist in each case to exclude pseudo tumours or fluid collections. The patients were clinically examined to try and find any adverse effects of the metal on metal articulation. Results:. We found one patient that had a revision of the femoral component for a peri-prosthetic fracture retaining the metal on metal articulation that subsequently developed loosening and a pseudo tumour. The remaining patients had excellent Harris Hip Scores and their implants showed no osteolysis or radiological signs of loosening. Conclusion:. Despite the recent concern we obtained good results after ten year follow up with metal on metal prosthesis. However we do recommend careful selection of patients, detailed attention to surgical technique and close follow up when considering metal on metal articulation. Patients should be counselled about the possibility of tissue reaction to the bearing surface and possible future revision surgery


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 74 - 74
1 Jun 2012
Gill I Krishnan M Reed M Partington P
Full Access

Introduction. The aim of this study is to report the results of Revision hip arthroplasty using large diameter, metal on metal bearing implants- minimum 2 year follow up. Methods. A single centre retrospective study was performed of 22 consecutive patients who underwent acetabular revision surgery using metal on metal bearing implants between 2004 and 2007. Birmingham hip resurfacing (BHR) cup was used in all patients - monoblock, uncemented, without additional screws in 16 cases and cemented within reinforcement or reconstruction ring in 6 cases. Femoral revisions were carried out as necessary. Results. There were 16 men and 6 women with a mean age of 71 years (51-83). Revision surgery was performed for aseptic loosening in 10, infected primary hip arthroplasty in 8, infected Hemiarthroplasty in 1 and Peri-prosthetic fracture with loosening in 3 patients. A 2-stage revision was performed for all infected hips. One patient died and the remaining 21 patients had clinical and radiological assessment at a mean 35 months (24-60). The mean Harris hip score was 75 (23-98) with 50% good to excellent results. Only 1 patient had further revision to a proximal femoral replacement and constrained cup in 2 stages for recurrent infection at 24 months. There were 2 recurrent infections (both revised for septic loosening) and 1 non-union of trochanteric osteotomy. There were no dislocations in the group. No radiological loosening of implants or metal ion complications have been seen at last follow up. Conclusions. We believe this is the first reported series on the use of large diameter metal on metal bearing surfaces for revision hip arthroplasty. Our series shows satisfactory short to medium term results in this complex group of patients with no component loosenings, despite monoblock cups and no dislocations


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 1 - 1
1 Sep 2012
Ramaskandhan J Malviya A Bowman R Lingard E Holland J
Full Access

Introduction. Cemented stems have shown 90–100% survivorship when coupled with polyethylene acetabular component. This study aims to compare cemented stem behaviour in combination with large metal on metal (MOM) vs. metal on poly (MOP) bearings. Patients and Methods. 100 patients were recruited into a single centre RCT (we required 40 in each group for power .90 to confirm stem subsidence of >0.5mm at 2 years; p< 0.05). Recruits were randomized to MOP (28mm) or MOM femoral heads with CPCS cemented femoral stem. Assessments included X-rays (AP pelvis), Harris Hip Scores, blood metal ion levels and patient questionnaires (WOMAC, SF-36, satisfaction questionnaire). Evaluations were done pre-operatively and 3, 12 and 24 months post operatively; blood metal ion measures at 1 year. Results. There were 50 patients in each arm of study matched for age (64 ± 8.5) and BMI (29.04 ±5.5). There was no difference in femoral stem subsidence at 2 years 1.34 (±1.3) and 1.4 (±1.2) mm for MOM and MOP respectively (p=0.88). There was significant improvement in HHS from pre-op to 3 months: 41 to 87 for MOM and 44 to 86 for MOP (p=0.00). This was maintained with no difference between groups at 2 years (p=0.74). Similar pattern was seen for WOMAC and SF-36 scores for both groups at 2 years (p>0.05). Increased blood Cobalt and Chromium levels were observed in 17% and 0% for MOM and MOP group. MOM group reported better patient satisfaction for overall (91% vs. 79%), pain relief (82% vs 66%) and improvement in ADL activities (94% vs.70%) at 2 years. Conclusions. There were no significant differences between groups for stem behavior, clinical and patient reported outcomes. Despite higher patient satisfaction reported by MOM patients, increased metal ion levels had raised concerns regarding the use of MOM bearings with cemented stems for primary THR


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 148 - 148
1 Jul 2014
Smeekes C Ongkiehong B van der Wal B
Full Access

Summary. The M2a-38. tm. metal on metal total hip arthroplasty showed a high incidence of pseudotumors and an unexpected high revision rate in our thoroughly screened cross sectional cohort. Introduction. After the revival of the metal on metal (MoM) bearing in total hip arthroplasty (THA) at the beginning of this century, there are now serious questions about this type of bearing. The advantage of large head MoM bearing is the increase in range of motion and stability. In our institution the choice was made for 38 mm heads. During the last few years concerns have been raised about the relationship of MoM bearing and elevated serum cobalt and chromium ion levels, their local and systemic toxicological effects and the incidence of local tumorous masses (pseudotumors). Are these findings applicable for all MoM bearings or are there also product specific issues. We present the outcome of a cementless MoM THA using a 38mm head in a unique consecutive series of 377 THA who were performed in our institution. Patients and Methods. All 351 patients (377 THA) with a cementless MoM THA (M2a-38. tm. , Biomet Inc, Warsaw, IN, USA, and Taperloc® stem, Biomet UK, Bridgend) between 2008 and 2011 were evaluated. All patients were analyzed by a physical exam, serum levels of cobalt and chromium and an interview to determine if there were any complaints. An MRI of the hip was made if patients reported pain during physical activity, allergies to metals, serum cobalt or chromium ion levels ≥ 5 ppb or if the inclination of the acetabular component was more than 50 degrees. Nine patients deceased, three were lost to follow up and four already underwent a revision before the screening. We analyzed 361 hips with an average follow up of 30 (range 2–58) months. The average preoperative age was 63 years (41–88). Results. 219 patients with 235 THA (65%) reported no complaints. Median cobalt level in patients with complaints was 6.6 (0.2–173) ppb and in the group without complaints 3.7 (0.2–27.3) ppb. Median chromium level in patients with complaints was 5.0 (0.1–134) ppb and in the group without complaints 3.7 (0.2–27) ppb. On the 226 performed MRI scans, 56 pseudotumors were diagnosed and described using the Anderson classification (9 C1, 41 C2 and 5 C3). 71 hips had been revised after a mean follow of 30 months (range 0.2–50 months). Reasons for revision were because of pain, raised metal ions and a pseudotumor in 28; pain and raised metal ion levels in 15; aseptic loosening of the acetabular or femoral component in 11; raised metal ions and a pseudotumor in 7; combination of luxations, luxation feelings and fractures in 5; infections in 3 and for other reasons in 2. Conclusion. The short-term results of the THA with a 38mm head metal on metal articulation are higher compared with other MoM bearings. They show a revision rate of 7% and 10% in the Australian and England register. In other peer reviewed literature we find a revision rate between 0–13% after five years. We also observed a high incidence of elevated serum levels of metal ions, pseudotumors and an unexpected high early revision rate


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_I | Pages 74 - 74
1 Jan 2003
Ryu J Saito S Osaka S
Full Access

One hundred fifty one total hip arthroplasty (THA) with metal on metal hip system of without cement were followed at a mean of 5. 5 years (range 2. 6 years to 7 years). Original diseases were 134 hip in 123 cases with osteoarthritis and 17 hips in 13 with rheumatoid arthritis. The average Harris hip score of before the operation was 45. 3 points and at the follow-up time the average score was 88. 5 points. Radiographic findings in the incidence of partial radiolucent line around the component: in the acetabular component, no radiolucent line was observed, in the femoral component, marginal lucent lines were 8. 6% (13 hips). Cortical hypertrophy was observed in high ratio of 27. 2%(41 hips). There was no loosening and tilt of component over 5 degrees observed in any patient. Postoperative dislocation of the hip was observed 9 hips (6. 0%). Cobalt concentrations and chromium concentration in the serum were measured in cases in metal on metal total hip, metal on poly arthroplasty and in controls without implants. Serum was assayed using graphite furnaces atomic absorption spectrophotometer. No patients were seen of elevation in serum cobalt and chromium concentrations. The authors conclude that THA with metal on metal system has presented excellent results for mid term follow-up in our hospital


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 133 - 133
1 Mar 2009
Neumann D Berka J Dorn U
Full Access

Aims: A polyethylene free, metal on metal acetabular system (Hofer-Imhof cup; Lubrimet metal on metal articulation made of CoCr-forge alloy) was designed in an effort to improve total hip arthroplasty longevity. This study was undertaken to review the clinical performance of this implant and to determine if acetabular loosening or revision and wear and osteolysis were prevalent. The long term results (mean follow up period 110 months) are presented in this study. Methods: In this prospective, randomized study a minimum of 90 months follow up results involving the first 100 implanted metal liner total hip arthroplasties are presented. The mean follow up was 110 months. Between April 1995 and November 1996 ninety-eight patients (100 hips) had a total hip replacement consisting of a titanium cementless self reaming, parabolic cup, a cementless titanium stem and the Lubrimet® metal on metal articulation. Two acetabular and one femoral component had to be revised due to aseptic loosening without showing macroscopic evidence of metallosis and no histological evidence of excessive metal wear. One patient hat to be revised by a complete change of the total hip (acetabular component and stem) showing histologic evidence of a metal induced hyperergic immunologic reaction. One patient presented with a broken neck of the stem and had to be revised by stem exchange. As a none device related orthopedic complication one acetabular component required revision surgery due to a periprosthetic fracture. Conclusions: The long term results of the Hofer-Imhof Lubrimet®Metal-on-metal articulation are satisfying and so the system represented a viable alternative for total hip arthroplasty in younger higher demand patients in the past. Due to the possible development of a hyperergic immunologic reaction to the metal on metal articulation and the availibility of a ceramic on ceramic articulation we do not implant this liner anymore


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 5 - 5
1 Jun 2012
Asal MK
Full Access

Fracture of the acetabulum can lead to degenerative arthritis of the hip, avascular necrosis of the femoral head, or both. Total hip arthroplasty is a common form of surgical treatment when significant joint changes and pain are present. Ten patients with fracture acetabulum were treated in this study using metal on metal total hip arthoplasty. The initial fracture was posterior wall fracture in one patient, posterior column fracture in one patient, transverse fracture in 2 patients, fracture dislocation in 3 patients and fracture posterior wall and column in 3 patients. The indications of arthroplasty were secondry osteoarthritis after internal fixation or after conservative management or collapse of the femoral head. Arthroplasty was done after an average period of 1.8 years (range from 1 to 4 years). After a follow up period ranged from 3 to 7 years with a mean of 4.6 years, the Harris hip score was improved from a mean of 51 (range 20 to 65) to a mean of 92.5 (range 90 to 95). Infection occurred in one case and two stages revision was done. Another case developed loosening of the acetabular component and was revised using cementless cup fixed with screws and bone graft. Metal on metal THR after acetabular fracture are relatively uncomplicated and lead to a good outcome despite the difficulties faced during the procedure


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 94 - 95
1 Mar 2010
Randelli F Randelli P Visentin O Arrigoni P Randelli G
Full Access

Metal on Metal coupling in total hip replacement has been widely used since many years. After the rebirth of resurfacing a new trend to use very large diameter metal-metal coupling with standard stem prostheses has been started. New prostheses, old and new problems. We analyze first failures with new large diameter metal on metal coupling. The analysis focused on seven early failures of large diameter metal-metal prostheses (two resurfacing and five cementless prosthesis with XL head) over the first series of 350 cases (100 resurfacing and 250 cementless) in the first year. Synovial fluid aspiration have been performed in all failed patients searching for metal ions and bacterial proliferation. Moreover, prosthetic component positioning was also studied as a possible primum movens of these failures. Some failed patients underwent epicutaneous patch test for skin reaction to metal. One resurface prosthesis failed as a result of an vascular necrosis and conseguent fracture of the femoral neck and revealed a moderate increment in metal ion concentration in blood and synovial fluid obtained at time of revision. One resurface prosthesis failed as a result of an infected metallosis with a huge intraabdominal mass and revealed a huge increment in metal ion concentration in blood, addominal and synovial fluid. This patient underwent a lumbotomy to evacuate the abdominal retroperitoenal mass before prosthesis removal for a two step procedure. Out of the five failures of metal on metal cementless prosthesis with XL head four were the result of aseptic loosening and one was the result of a low grade infection discovered at coltures after revision surgery. Three showed clear metallosis caused by wrong positioning (more than 50° of cup inclination). All of these three presented an articular noise and elevated blood and sinovial fluid metal ion concentration. The fourth patient with aseptic loosening had a good component positioning but demonstrated an epicutanous allergic reaction to Cobalt. Also one of the three patients with metallosis resulted allergic to Cobalt. The blood and sinovial metal ion values were always elevated but particularly high in patients with cup inclination over 50°. Large head metal on metal prostheses demonstrated a higher percentage of early failure in our experience. They are very sensible to positioning. Blood and sinovial metal ion determination helps to promptly diagnose a bad metal on metal prosthetic functioning. A more accurate analysis about the different metals available on the market and their resistance to edge wear should not be delayed any further


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 228 - 228
1 Dec 2013
Ando W Yamamoto K Koyama T Fudo K Tsujimoto T Aihara M Ohzono K
Full Access

Metal on metal total hip arthroplasty provides the potential improvement in articular wear and the use of large-diameter femoral heads following the prospect for reduction in the risk of dislocation. The purpose of this study was to compare the clinical and radiographic outcomes as well as serum metal ion level between the two different component designs with small and large femoral heads in metal on metal total hip arthroplasty. We studied 39 patients with large head (Magnum®, Biomet; cup size minus 6 mm) and 37 patients with small head (M2a taper®, Biomet; 28 or 32 mm head) of metal on metal total hip arthroplasty between December 2009 to October 2011 with follow-up of 2.1 years (1.0–3.3 years) after surgery. Harris Hip Score, UCLA activity score, EQ-5D, radiographic assessment, and serum cobalt and chromium ion levels were evaluated. Harris Hip Score, UCLA activity score, and EQ-5D were improved after surgery in small and large head groups, however, no significant differences were observed between both groups. Cup inclination was below 50 degree in all prosthesis. No loosening and no osteolysis were observed. Cobalt and chromium ion was not detected before surgery; however, metal ion levels of both groups were increased after surgery in time dependent manner. There was no significant difference between two groups at one year after surgery. One patient in each group showed the increase of cobalt ion level over 7 ppb (15.4 ppb, 12.9 ppb) without any clinical symptom including pain. Cup inclination was 29 degree in both patients and cup anteversion was 38 and 41 degree, respectively. There was no significant difference of ion levels between both groups. No dislocation was observed in large head group while one dislocation occurred in small head group. No patients required the revision surgery. This metal on metal component, especially with large femoral heads, showed the good clinical results at the maximum follow up of 3.3 year after surgery. However, the ion level of two patients increased over 7 ppb and longer follow-up will be needed


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 75 - 75
1 Jun 2012
Gill I Krishnan M Reed M Partington P
Full Access

Introduction. To report the short to medium term results of acetabular reconstruction using reinforcement/reconstruction ring, morcellised femoral head allograft and cemented metal on metal cup. Methods. Single centre retrospective study of 6 consecutive patients who underwent acetabular reconstruction for revision hip surgery. The acetabulum was reconstructed using morcellised femoral head allograft and reinforcement or reconstruction ring fixed with screws. The Birmingham cup – designed for cementless fixation, was cemented into the ring in all cases. The uncemented Echelon stem with metal on metal modular head was used for reconstructing the femur. Data from our previous in-vitro study had shown good pull out strength of a cemented Birmingham cup. Results. There were 2 men and 4 women with a mean age of 75 years(57-83). Revision was performed for aseptic loosening in 2, septic loosening in 2 and peri-prosthetic fracture with loosening in 2 patients. All patients were reviewed clinically and radiographically at a mean of 36 months follow-up(range 24 - 42 months). Revision was not necessary in any patient for failure of acetabular or femoral fixation. However, 1 patient had revision to a proximal femoral replacement and constrained cup for recurrent infection and osteomyelitis at 24 months. This patient was excluded from the final analysis. The mean Harris hip score at last follow up was 79(range 70-89). Radiographic analysis revealed good graft incorporation and no signs of loosening or cup/ring migration. No dislocations or metal ion problems were recorded in this series. Conclusions. To the best our knowledge, this is the first series using cemented metal on metal cups within a reinforcement/reconstruction ring for revision hip arthroplasty. Excellent cemented fixation of the cup, manufactured for cementless fixation, was obtained at surgery with no evidence of loosening, and no dislocations at minimum 24 months follow up


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 274 - 274
1 Jul 2011
Kakwani R Wainwright C Tawari G Kashyap S Roysam A Nanu A
Full Access

Purpose: A single blind prospective randomised controlled trial comparing the Metal-on-polyethylene articulation with the metal-on-metal articulation in THA.

Method: The clinical and radiological findings of the consecutive patients who were enrolled in the RCT at the participating centres were recorded prospectively. The clinical evaluation was performed with the Harris scoring system as well as the Oxford Hip Scoring Sheet. The computer randomised option was revealed to the operative surgeon only after the patient was anaesthetised, during the recruitment period (June 1998 to July 2004). Of the total of 378 patients, 2 died prior to the final review and 63 were lost to follow-up. The final study group contained 315 patients, with 159 patients in the metal-on-polyethylene group and 156 patients in the metal-on-metal group.

Results: The indication for the hip arthroplasty for majority (309 patients) was primary osteoarthritis. The average age at the time of the surgery was 68.2 years and the average duration of follow-up was 85 months (42–115). There was an improvement of the Oxford hip scores from an average of 37 per-operatively to 16 postoperatively. The Harris hip scores also improved from an average of 47.0 pre-operatively to 87.3 post-operatively. The patient groups were statistically similar with respect to age, sex and duration of follow-up, and the final outcome scores revealed no statistical difference between the two groups.

Conclusion: The clinical results obtained with the use of the articulation are comparable to those obtained by the metal-on-polyethylene articulation encouraging the use of this alternative bearing surface.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 5 - 5
1 Sep 2012
Smith J Railton P Kinniburgh D Faris P David L Marshall D Burkart B Powell J
Full Access

In a cross sectional cohort study three different metal on metal total hip systems were assessed. Two monoblock acetabular designs; the Durom socket (Zimmer, Warsaw, In) and the Birmingham socket (Smith and Nephew, Memphis, TN), and one modular metal on metal total hip system (Pinnacle, Depuy Orthopaedics, Warsaw, In) in patients who have received these implants in the our region. 56 patients were recruited in total. All patients were assessed clinically, radiologically and biochemically. Whole blood Cobalt, Chromium and titanium levels were tested. The median head size used in the Birmingham group was 52mm (Range 44mm to 56mm), and in the Durom group, 48mm (Range 42 to 54mm). The median head size used in the modular Pinnacle group was 40mm (Range 36–44mm). The blood metal ions levels in the larger non modular acetabular sockets were significantly raised compared to the Pinnacle group. For Co 1.95 µg/l and 2.70 µg/l in the Durom and Birmingham groups respectively compared to only 0.52 µg/l in the Pinnacle group (P< 0.001). Mean Cr levels were the same for the two monoblock systems, 1.9 µg/l compared to the Pinnacle sockets 1.2 µg/l (P< 0.001). Our study clearly demonstrates that there is a significant difference in metal ion levels in patients following a monobloc large head arthroplasty system compared to a smaller modular metal on metal hip arthroplasty. The smaller head size appears to produce less metal ions whilst at the same time a 36mm–44mm head size is large enough to increase hip stability and range of movement as well as decreasing the risk of impingement. In our practice we are no longer using this design and the safest strategy, when considering metal on metal bearings, is to use a modular, smaller head system such as the Pinnacle


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 58 - 58
1 Mar 2008
Beaulé P Dorey F LeDuff M Amstutz H
Full Access

Ninety-four hips with a mean patient age 34.2 (range 15– 40) with a metal/metal surface arthroplasty (SA) were reviewed with 71% men and 14% with previous surgery. The Chandler risk index was calculated as well as the SA risk index (SARI). At a mean follow-up three years, three hips were converted at a mean of twenty-seven months (two to fifty), and ten patients had significant radiological changes. Mean SARI for the thirteen problematic hips versus remaining hips was significantly higher, 4.7 and 2.6, respectively (p=0.00). If SARI > 3 the relative risk of early problems is twelve times greater than if SARI ≤3. The purpose of this study was to evaluate the early outcome of a hybrid metal on metal surface arthroplasty of the hip in patients forty years and younger and identify potential risk factors. Surface Arthroplasty Risk Index can help identify patients who may be at increased risk of early failure following metal on metal surface arthroplasty. Proper patient selection and careful surgical technique may minimize early failures with the re-introduction of surface arthroplasty of the hip. Ninety-four hips mean age 34.2 (range 15– 40) with metal/metal surface arthroplasty (SA) were reviewed with 71% men and 14% with previous surgery. The Chandler risk index was calculated and SA risk index (SARI). Mean follow-up three years (range 2–5), three hips were converted at a mean of twenty-seven months (2–50), and ten patients had significant radiological changes. Mean SARI for these thirteen problematic hips versus remaining hips was significantly higher, 4.7 and 2.6, respectively (p=0.00). The mean angle between the prosthesis stem and femoral shaft in the problematic group was significantly smaller than the remaining hips (p=0.03): 133° and 139°, respectively. If SARI > 3 the relative risk of early problems is twelve times greater than if SARI ≤3


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 422 - 422
1 Apr 2004
Tuke MA
Full Access

Metal on Metal bearings have functioned satisfactorily for up to 30 years without a full understanding of what shape the components were at the start, or as they were worn over time. Modern metrology methods can use elegant computer driven coordinate measuring machines in both manufacturing and retrieval analysis as most manufacturers and laboratories do. This method however is fl awed in its ability to look at the much higher levels of resolution required for analysing these bearings. A method has been developed which will analyse to the sub micron deviances from roundness that the subleties of manufacturing can achieve, and allows an analysis of wear of explanted implants. Both new and explanted Ring, McKee, Stanmore, Metasul and Resurfacing implants have demonstrated initial form variances that range from 1 to over 100 microns out of round. Roundness trace analysis has allowed linear wear on explanted implants to be measured from 0.5 to 150 microns. This represents combined head and cup linear annual wear rates from 2 to 65 microns per year. Metal heads and cups are not manufactured as round as has been believed. The average wear rates over time for different Metal on Metal systems is not the same


Orthopaedic Proceedings
Vol. 87-B, Issue SUPP_I | Pages 5 - 5
1 Mar 2005
Dunstan E Sanghrajka A Tilley S Unwin P Cannon S Briggs T
Full Access

Retrospective analysis of 25 consecutive metal on metal proximal femoral replacements performed at our unit between 1965 and 1979. Methods. Patients were: clinically evaluated using the Modified Harris Hip and Enneking Scoring Systems and radiologically using the ISOLOS scoring system. The concentration of Cr, Co, Ti, Al, V, Mo & Ni in whole blood and urine was also measured by High-Resolution Inductively Coupled Mass Spectrometry and compared with controls and patients with other implants. Retrieved prostheses (in situ for in excess of 25 years) were analysed for roughness and wear using a Mitutoya form tracer and an electron microscope. Results. Thirteen patients have since died; nine from metastatic disease and four from other causes. Of the remainder, eleven (44%) are still alive, five still retaining metal on metal articulations and one has been lost to follow up. They have been in situ for an average of 32 years. The average modified Harris Hip score is 76 (53–93) and the average Enneking Score is 74 (63–90). In the retrieved prostheses the contact zones were found to be smoother (Ra 0.05?m), have fewer and smaller carbides together with evidence of ‘self-healing’ when compared to the original surface (Ra 0.32?m). Blood & urine levels of Co & Cr were significantly elevated. Co levels were exceptionally elevated in loose prostheses but levels quickly fell following revision. Conclusion. We have shown the potential longevity of metal on metal arthroplasty. The wear seen in retrieved specimens is low and we might expect to improve the fixation by reducing the torque with apical bearing and encouraging extra-cortical bone bridging with hydroxy-apatite coated collars. Elevated serum and urine Co levels may well predict a loose prosthesis and may be useful as a screening tool


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 79 - 79
1 Mar 2006
Pillai A Basapa P Hilmy O El-Hadidi M
Full Access

Background: Metal on metal resurfacing of the hip is an attractive conservative treatment option for the younger, active arthritic patient. It has the advantages of bone preservation and the longevity of metal bearing surfaces. Concerns remain over the performance of these implants due to historical failure of similar devices. The possible longer-term survival of these implants can only be justified if the early results are as successful as conventional hip replacement. Patients & Methods: The clinical and radiological follow-up of our first 25 consecutive resurfacing patients is presented. The Birmingham Hip Resurfacing prosthesis (Midland Medical Technologies, Birmingham, UK) was used, and all procedures performed by the senior author (EHM). Clinical assessment was done using the modified Harris Hip Score, and functional outcomes measured using the Western Ontario and McMaster Universities Osteoarthritis index (WOMAC). Radiographs were examined for radioleucent lines, migration of components, osteolysis and heterotropic ossification. Results: 28 primary hip resurfacings in 25 patients were reviewed. The mean follow up period was 18 mts (12– 36mts). The mean age of patients at time of operation was 51yrs (37–64 yrs). Main indication for surgery was Osteoarthrosis. The mean Harris Hip Score improved from 44 preoperative, to 94.5 postoperative. The mean preoperative WOMAC score was 30 and the mean postoperative score 88.The average duration of hospital stay was 5 days, and the average operative blood loss 3g %. All patients in paid employment returned to work at an average of 8 weeks postoperatively, and majority of patients are able to participate in sports and recreational activities. There were no cases of dislocations, femoral neck fractures, avascular necrosis or osteolysis. Conclusion: Hip resurfacing using metal on metal hybrid fixed components, offers a viable treatment alternative in younger active patients. Early outcomes are at least comparable to total hip replacement in terms of pain relief and restoration of function. Despite the lack of long-term data, the relative absence of complications and the potential benefits justify its continued use. We believe the future is very promising for this procedure


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 52 - 52
1 Mar 2009
Quinlan J Dillon J Walker E O’Sullivan T
Full Access

Patients with DDH are known to be at risk of early degenerative changes to their hips. To date, no consensus exists as to the most appropriate management of this group, with many surgical options being associated with specific complications such as dislocation and early wear. In addition, modern resurfacing methods are considered by many to be contra-indicated in patients with DDH due to the technical difficulty of the procedure. This prospective study analyses a single surgeon series of known DDH hips that underwent metal on metal resurfacing from November 1999 to July 2004 inclusive. There were 31 resurfacings carried out on 28 patients (11 males, 17 females). The mean age of the study group at the time of surgery was 43.9+/−9.1 years. No patient was lost to follow up. Pre-operatively, 23 hips were classified as Crowe I (n=9), II (n=5), III (n=5) and IV (n=4). Patients were followed up to a mean of 46.4+/−18.1 months. The mean Harris Hip scores were 54.9+/−9.3 pre-operatively and 98.1+/−4.9 post-operatively (p< 0.001, Student’s t-test). Using the UCLA activity profile, the mean scores were 3.2+/−1.0 pre-operatively and 6.4+/−1.8 post-operatively (p< 0.001, Student’s t-test). Although the management of young patients with early degenerative changes secondary to DDH remains controversial, the results of this study suggest that not only is resurfacing technically possible even in advanced cases, it also offers excellent functional outcomes and should be considered in appropriate cases


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 113 - 113
1 Mar 2013
Cho YJ Lee J Chun YS Rhyu KH Kwak S Ji H Won YY Yoo M
Full Access

Purpose. To evaluate the radiological changes after metal on metal resurfacing arthroplasty. Materials and Methods. Between December 1998 and August 2004, 166 hips in 150 patients who underwent metal resurfacing arthroplasty and followed up more than 4 years. Their mean age at the time of operation was 37.3 years(range, 15–68 years) and mean period of follow-up was 6.1 years(range, 48–95 months). The cause of arthroplasty included 115 avascular necrosis, 43 osteoarthritis, 7 ankylosing spondylitis, 1 haemophilic arthropathy. All patients had anteroposterior, translateral radiographs of the hip made preoperatively and each follow-up visit, and we analyzed radiographic findings such as radiolucencies or impingement signs around implant, neck narrowing and heterotopic ossification. Results. There was a no significant difference between preoperative and postoperative Harris hip score and range of motion. The mean stem-shaft angle was 137.4°, and 55.4% were ranged 130° to 140°. The mean inclination of acetabular component was 44.9°. There were no radiolucent lines or osteolytic lesion around the acetabular components, but 3 hips showed radiolucency around the head-neck junction(1.8%) and 4 hips showed radiolucent line around the stem (2.4%). 12 hips had impingement signs around the head-neck junction (7.2%), and 2 cases showed neck narrowing (1.2%). 3 cases had some heterotopic ossification (1.8%). In 12 cases with impingement signs, the stem-shaft angle and inclination of acetabular component were lower than control group. Pseudotumor was not found in this cohort. Conclusions. This study demonstrates no serious radiological problems till the midterm follow-up after resurfacing arthroplasty, but osteolytic lesion such as radiolucent line around head-neck junction, neck narrowing can be a potential cause of failure in future. Even though the radiolucent line around stem of femoral component revealed no subjective symptom yet, it suggests the micromotion of femoral component which can lead to femoral component loosening. The most common radiological findings, impingement signs, seem not to have clinical significance


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 93 - 93
1 Aug 2012
Clarke S Phillips A
Full Access

Metal on metal press-fit acetabular cups are the worst performing acetabular cup type with severe failure consequences compared to cups made from more inert materials such as polyethylene or ceramic. The cause of failure of these cup types is widely acknowledged to be multi-factorial, therefore creating a complex scenario for analysis through clinical studies. A factorial analysis has been carried out using an experimentally validated finite element analysis to investigate the relative influence of four input factors associated with acetabular cup implantation on output parameters indicating potential failure of the implantation. These input factors were: cup material stiffness; cup inclination; cup version; cup seating; and level of press-fit. The output parameter failure indicators were: wear; tensile strains in the underlying bone; bone remodelling; and cup-bone micromotions. The factorial analysis concluded that the most significant influence was that of cup inclination on wear, and the second most significant was the influence of the level of press-fit on bone remodelling at the acetabular rim. Significant influence was also observed between version angle and wear, and cup-seating and micro-motion. The results demonstrated the clear multi-factorial nature of implant failure and highlighted the importance of correct implant positioning and fit


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 531 - 531
1 Oct 2010
Thomas G Faisal M Young S
Full Access

Introduction: Disagreement exists in the literature as to whether hip resurfacing requires more acetabular reaming than uncemented hip replacement. Materials and Methods: Thirty-nine patients underwent total hip arthroplasty with a large metal on metal resurfacing acetabular cup and large metal head with a conventional uncemented femoral stem. Although there was no change in routine operative technique the femoral neck was measured (as if for resurfacing) before being cut. The size of the acetabular component actually inserted was compared with the smallest size which could have been used had a resurfacing been carried out for each patient. A two tailed paired t test was carried out. Results: We found that there was no difference in female patients but in men the actual cup used was 2.44mm smaller than the smallest possible resurfacing cup which equates to 6.39cm. 3. less bone (p < 0.0001). Discussion: By using paired samples we have reduced confounding factors and shown that a large metal on metal bearing on a conventional stem is more conservative of acetabular bone than hip resurfacing in male patients


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 139 - 139
1 Jan 2013
Bolland B Maul C Cook R Roques A Wood R Latham J
Full Access

The poor outcome of large head metal on metal total hip replacements (LHMOMTHR) in the absence of abnormal wear at the articulating surfaces has focussed attention on the trunnion/taper interface. The RedLux ultra-precision 3D form profiler provides a novel indirect optical method to detect small changes in form and surface finish of the head taper as well as a quantitative assessment of wear volume. This study aimed to assess and compare qualitatively the tapers from small diameter with LHMOMTHR's. 3 groups of retrieval tapers were analysed (Group 1: 28mm CoCr heads from MOMTHRs (n=5); Group 2: Large diameter CoCr heads from LHMOMTHRs (n=5); Group 3: 28mm heads from metal on polyethylene (MOP) THRs; n=3). Clinical data on the retrievals was collated. Both bearing surfaces and head tapers were measured for wear using the Redlux profiling non contact measurement system. Measurements included taper angle and 3D surface maps. Taper angles obtained with the Redlux method were compared to those obtained using CMM measurement on 3 parts. The Redlux profiling, including imprints, was also repeated 3 times to gauge potential errors. There was no difference in mean 12/14 taper angles between groups. There was no difference in volumetric and linear wear at the bearing surface between groups. Only the LHMOMs showed transfer of pattern from the stem to the internal head taper, with clear demarcation of the contact and damaged area between head taper and stem trunnion. 3D surface mapping demonstrated wear patterns compatible with motion or deformations between taper and trunnion in the LHMOM group alone. Discussion: Differences in appearance of the taper surface between LHMOMTHRs and MOP or MOM small diameter devices highlight an area of concern and potential contributor to the mode of early failure. Further work is required to fully qualify the Redlux method capabilities


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 18 - 18
1 Mar 2013
Bolland B Maul C Cook R Roques A Tuke M Wood R Latham J
Full Access

The poor outcome of large head metal on metal total hip replacements (LHMOMTHR) in the absence of abnormal wear at the articulating surfaces has focussed attention on the trunnion / taper interface. The RedLux ultra-precision 3D form profiler provides a novel indirect optical method to detect small changes in the form and surface finish of the head taper as well as a quantitative assessment of wear volume. This study aimed to assess and compare qualitatively the tapers from well functioning small diameter, with poorly functioning LHMOMTHR's using the above technique. Method. 3 groups of retrieval tapers were analysed (Group 1: 28 mm CoCr heads from well functioning MOMTHRs (n=5); Group 2: Large diameter CoCr heads from LHMOMTHRs revised for failure secondary to adverse reaction to metal debris (n=5); Gp 3 (control): 28 mm heads from well functioning metal on Polyethylene (MOP) THRs; n=3). Clinical data on the retrievals was collated. The Redlux profiling of modular head tapers involves a non direct method whereby an imprint of the inside surface of a modular head is taken, and this is subsequently scanned by an optical non contact sensor using dedicated equipment [1]. The wear was also measured on the bearing surface [1]. RedLux profiling of the tapers produced a taper angle and 3D surface maps. The taper angles obtained with the Redlux method were compared to those obtained using CMM measurement on 3 parts. The Redlux profiling, including imprints, was also repeated 3 times to gauge potential errors. Results. There was no difference in mean 12/14 taper angles between groups. There was no difference in volumetric and linear wear at the bearing surface between groups. Only the LHMOMs showed transfer of pattern from the stem to the internal head taper, with clear demarcation of the contact and damaged area between head taper and stem trunnion (see figure 1 – interpretation of head taper surface features demonstrated using Redlux optical imaging). 3D surface mapping demonstrated wear patterns compatible with motion or deformations between taper and trunnion in the LHMOM group. These appearances were not seen in tapers from small diameter MOM and MOP THRs (see Figure 2). Discussion. Differences in appearance of the taper surface between poorly functioning LHMOMTHRs and well functioning MOP or MOM small diameter devices highlight an area of concern and potential contributor to the mode of early failure. Further work is required to fully qualify the Redlux method capabilities, and to understand the origin of the damage seen on those tapers, and the possible partial contribution of damage caused during retrieval


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 378 - 378
1 Jul 2008
Khan M Kuiper J Richardson J
Full Access

In-vitro studies have shown that wear rates of the metal on metal (MOM) bearing hip prostheses decline once the bearing runs-in and the bearing subsequently enters a steady state wear phase. Baseline cobalt levels are thus expected to decline with time in patients. Several clinical studies have not found such a decline. Baseline cobalt levels are hence limited in their capacity to provide information on the wear performance of the bearing couple. We have demonstrated in a previous study that exercise causes a rise in plasma metal ion levels in patients with MOM bearing hip replacement. Would the exercise related cobalt rise be more sensitive to detect change in wear behaviour of the bearing couple? We tested the alternate hypothesis that exercise related rise in the plasma cobalt levels will correlate inversely with the duration of MOM implantation. Sixteen patients with three different well functioning MOM bearing hip replacement [two types of resurfacing (BHR, Cormet) and Metasul] were included into the study. Patients were divided in to two groups based on time since implantation, an early group of mean 18 months and a late group of mean 57 months. Plasma levels of cobalt were measured before (baseline) and after 1 hour of maximal exercise (peak). The difference between baseline and peak for each patient provided the exercise related cobalt rise. A significant increase in plasma cobalt levels of 13% was noticed after the exercise (p < 0.005). Baseline Cobalt levels in the late group (53nmol/l) were higher than early group (44nmol/l) but the difference was not significant (p=0.45). However, the mean exercise related Cobalt rise levels was lower in late group (3.5nmol/l) than the early group (6.5nmol/l). This lower rise in cobalt level in the late group precisely reflects on the steady state wear as seen in in-vitro tests. Baseline cobalt levels are limited in determining the in-vivo performance of the bearing couple. Exercise related rise in cobalt levels can differentiate the running in and steady state wear phases of metal on metal bearings and is thus a more accurate tool of assessing in-vivo wear performance of the bearing couple


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_18 | Pages 22 - 22
1 Apr 2013
Hosny H Srinivasan S Keenan J Fekry H
Full Access

Medical and Health care products Regulatory Agency (MHRA) released an alert in 2010 regarding metal on metal (MoM) bearings in hip arthroplasty owing to soft tissue reactions to Metal debris. Following this, we adopted a targeted screening protocol to review patients with this bearing couple. 218 Patients (252 hips), mean age 53.2 (25–71) years were assessed clinically using Oxford hip score (OHS) and X-ray examination. The mean follow up was 44.5 (12–71) months. Patients were considered at higher risk (118 patients/133 hips) if they had deterioration of OHS (50 hips), Small sized heads <50mm (114 hips), acetabular inclination >500 (37 hips), neck thinning (17 hips). These patients (107/118), (120/133 hips) were further investigated through measuring metal ion levels and magnetic resonance imaging (MRI). The mean blood levels of cobalt and chromium in this group were 6.7, 8.62 ug/L respectively. Metal ions increased significantly with high acetabular inclination angles (p=0.01, 0.004 respectively), but was not affected by the size of the head (p=0.13). MRI showed periprosthetic lesions around 28 hips (26 fluid collections, 2 pseudotumours). The screening protocol detected all patients who subsequently required elective revision. We believe that this protocol was beneficial in detecting problematic MoM hip replacements


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 402 - 403
1 Apr 2004
Band T Metcalf J Jones H
Full Access

The ‘first generation’ Metal on Metal bearing devices was typically produced from cast, high carbon CoCrMo alloy and was in the as-cast condition. They exhibited course, hard primary, and block carbides supported by a softer matrix material. This bi-phasic condition has been verified through reported literature and forensic scientific studies of ‘long-term survived’ retrieved ‘first generation’ devices. The as-cast microstructure of CoCrMo alloys possesses superior wear resistance to the microstructures formed following post cast thermal treatments. It has been well reported that the improvement of mechanical properties, such as tensile or fatigue strength, can be achieved through the thermal treatment of this alloy. Thermal treatments of this alloy have been found to alter its’ microstructure with a significant modification to the carbide phase morphology. The modifications vary with a tendency for a refinement of the carbide size through dissolution of the chromium and molybdenum through solid state solution. Through the examination of the wear patterns of retrieved devices and wear testing of this material in its’ various microstructural conditions, it has been shown that modifications to the carbide morphology, to achieve improved mechanical properties, reduces its’ bio-tribological properties/performance leading to a lower wear resistance. The as-cast carbide morphology is the most mechanically stable condition and with its’ volume fraction, reduces the potential for adhesive wear of the matrix through ‘matrix to matrix’ contact of the two opposing bearing surfaces. It has been reported that abrasive wear is the typical mechanism for metal on metal bearings due to the generation of ‘third body’ particles from carbide asperity tips fracturing during the initial ‘running-in’ period [typically 500k to 1M cycles]. After this stage the carbides become almost level with the surrounding softer matrix material with ‘third body’ scratches dominating the surface topography. Evidence of surface pitting on ‘first generation’ devices [McKee Farrar and Muller] and modern high carbon wrought devices [Metasul] has been attributed to adhesive/fatigue wear following surface-to-surface contact. Therefore, in microstructural conditions, where there is a reduced carbide volume fraction, or no carbides present, wear resistance is reduced. To test this hypothesis two wear tests have been carried out on CoCrMo samples produced from the same chemistry alloy, with varying microstructures, using Calowear [abrasive] and Pin on Dist [adhesive] tests. The as-cast microstructural condition was determined to have the lowest wear coefficient [k=mm3/Nm] in both tests, however statistical significance at 90% confidence interval was only confirmed in the Calowear Test. Examination of wear scars confirmed the mechanical stability of the as-cast carbide phase. It is noted, however that there are papers which have been published offering a divergence of opinion to this hypothesis and which have been considered by this author


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 317 - 317
1 Jul 2008
Khan M Kuiper J Takahashi T Richardson J
Full Access

Introduction: The wear particles produced from the metal-on-metal hip prosthesis causes measurable rise of metal ion levels in the patient’s body fluids. Wear of the bearing is directly related to its use. The goal of this study is to test two hypotheses. Firstly, that exercise causes increased wear particles in vivo which can cause immediate measurable rise in the serum metal ion levels. Secondly, that this rise in metal ion level is different for different types of bearings. Material and Methods: Eighteen participants were allocated to four different groups i.e. Birmingham Hip Resurfacing prosthesis group, Cormet 2000 resurfacing prosthesis group, Thrust plate prosthesis group (28mm metasul articulation) and group four with out any metal work. Blood samples were taken immediately before, immediately after and one hour after exercise. Plasma cobalt and chromium was determined using Inductively-Coupled-Plasma-Mass-Spectrometry and Dynamic-reaction-cell respectively with detection limit of 2nmol/l each. Results: The four patient groups were comparable. A significant increase (p< 0.005) in serum cobalt and chromium of 13% and 11% respectively, was noticed after the exercise. Rise of cobalt levels in patients with a resurfacing MOM was 8.5 times (BHR group) and 6.5 times (Cormet group) larger than in those with a Metasul (p=0.021 and p=0.047). Neither rise of metal levels nor baseline levels correlated with any other factor (p> 0.27). Discussion: Physiologic exercise causes immediate detectable rise in the serum metal ion levels. The increase is predominantly related to the size of the bearing surface. Exercise-related-cobalt-rise could be used to assess the tribology of the different metal on metal designs in vivo for future research


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 353 - 353
1 Mar 2004
Skipor A Campbell P Amstutz H Jacobs J
Full Access

Aims: Measure serum chromium (SrCr) and cobalt (SrCo) and urine chromium (UCr) levels in patients with metal on metal surface arthroplasty of the hip. Methods: Ion levels were measured prospectively in 22 patients implanted with the Conserve Plusª (Wright Medical, TN) CoCr hip resurfacing. There were 15 males and 7 females with an average age of 49 years (range 28 Ð 62 yr). Serum and urine samples were collected using strict anti-contamination techniques pre-operatively and at 3, 6 and 12 months using graphite furnace atomic absorption spectrophotometry. Results: All postoperative metal levels were increased compared to their pre-operative levels. SrCr and SrCo values are at their highest at 3 months post operative and then begin to decrease. UrCr although elevated at 3 and 6 months postoperatively compared to the preop values, the levels continued to increase after the 6-month interval. These values are approximately 4-fold, 7-fold and 3-fold higher in SrCr, UrCr and SrCo, respectively, compared to the values seen in a group of patients with well functioning conventional metal (CoCr) on polyethylene total hips at 84 months postoperative measured by our group. Conclusions: The present levels are 2-fold lower in both SrCr and UrCr and 3-fold lower in SrCo than a group of patients with older generation surface arthroplasties reported previously by our group, suggesting improved manufacturing techniques and material properties have resulted in reduced component wear and generation of wear particles. 12 and 24 month data are currently being collected and analyzed


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 3 - 3
1 Sep 2012
Langton D Ghandi J Sidaginamale R Mereddy P Joyce T Lord J Natu S Nargol A
Full Access

Introduction. We conducted independent wear analysis of retrieved metal on metal (MoM) hip components from around the world. All patients with resurfaced hips who developed adverse reactions to metal debris (ARMD) were found to have increased wear of the bearing surfaces. This was untrue in patients with large diameter (?36mm) MoM total hip replacements. This led us to search for other factors leading to ARMD. Methods. MoM THR explants retrieved from 78 patients suffering ARMD underwent full volumetric wear analysis of bearing surface and taper-junctions using coordinate measuring machine. Scanning electron microscopy (SEM) used to characterise material composition of specific areas. Results. 34 MoM THRs were found to have relatively low bearing surface wear (< 3mm. 3. /year). In each of these cases, material loss up to 60 microns wear depth was identified on the internal taper-junctions of femoral components. However, volumetric loss was rarely >5mm. 3. Similarly only 65% of metal ion levels of these patients were found to be greater than the MHRA guidance figure (7µg/L). Patterns of material loss at the tapers were consistent with antero-posterior force splaying open the taper-junction. This characteristic pattern was identified in number of commercially available devices (titanium and cobalt chromium stems). Soft tissue lesions were severe in patients found to have taper damage. Histology confirmed severe ALVAL with lymphoid neogenesis in majority cases, suggesting that wear debris from taper junctions may have greater potential to stimulate adverse immune response. Discussion. The results suggest that forces transmitted from large diameter hard-on-hard bearing surfaces are sufficient to cause mechanical damage to modular junctions with secondary localised corrosion. We urge caution in the use of these designs and recommend a re-evaluation of the stem head interface


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 12 - 12
1 Mar 2013
Bolland B Roques A Maul C Cook R Wood R Tuke M Latham J
Full Access

The poor outcome of large head metal on metal total hip replacements (LHMOMTHR) in the absence of abnormal articulating surface wear has focussed attention on the trunnion / taper interface. The RedLux ultra-precision 3D form profiler provides a novel indirect optical method to detect small changes in form and surface finish of the head taper as well as quantitative assessment of wear volume. This study aimed to assess and compare qualitatively tapers from small and large diameter MOMTHR's. Tapers from 3 retrieval groups were analysed. Group 1: 28mm CoCr heads from MOMTHRs (n=5); Group 2: Large diameter CoCr heads from LHMOMTHRs (n=5); Gp 3 (control): 28mm heads from metal on polyethylene (MOP) THRs; n=3). Clinical data on the retrievals was collated. RedLux profiling of tapers produced a taper angle and 3D surface maps. The taper angles were compared to those obtained using CMM measurements. There was no difference between groups in mean 12/14 taper angles or bearing surface volumetric and linear wear. Only LHMOMs showed transfer of pattern from stem trunnion to head taper, with clear demarcation of contact and damaged areas.3D surface mapping demonstrated wear patterns compatible with motion or deformations between taper and trunnion in the LHMOM group. These appearances were not seen in tapers from small diameter MOM and MOP THRs. Differences in appearance of the taper surface between poorly functioning LHMOMTHRs and well functioning MOP or MOM small diameter devices highlight an area of concern and potential contributor to the mode of early failure


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 130 - 130
1 May 2011
Nikolaou V Petit A Huk O Zukor D Bergeron S Antoniou J
Full Access

Introduction: Several studies have shown the presence of cobalt (Co) and chromium (Cr) ions in blood, urine, and organs of patients after THA using Co-Cr alloy-based implants. Even though it is well known that exposure to heavy metals may lead to significant alterations in human sperm morphology and motility, less is known on the effect of Co and Cr on semen parameters after metal on metal (MOM) hip replacement. Methods: Semen was collected form 10 patients between 41 and 49 years old (mean=45.9±3.0 years) by masturbation after 2–3 days of abstinence. The time of implantation varied from 1 to 9 years (mean=5.1±3.9 years). Samples were collected in a sterile container and examined within 1h after ejaculation for morphology, motility, and number of sperm cells following standard criteria. All patients were doing well at their follow-up visits (Harris Hip Score=94±4; UCLA activity a score=7±1) and no sign of osteolysis was observed on X-rays. Co and Cr concentrations were measured in both the seminal plasma and in the blood of patients by inductively coupled plasma-mass spectroscopy (ICP-MS). Results: results showed that the levels of Co in the seminal plasma and the blood of the patients were not statistically different. However, the level of Cr was significantly lower in the seminal plasma than in the blood of the patients. The ejaculate volume (2.1 ±0.6 ml), the sperm density (66±53 x 106), the total sperm count (151±75 x 106/ml), the pH (8), and the percentage of normal morphology (46±18%) were in the range of the WHO criteria for fertile population and also in the range of reference patients in the city of measurements. However, the viability was lower than that observed in a fertile population without prosthesis (41±19%). Conclusions: results of the present study strongly suggest that both Co and Cr ions crossover to the semen but that their concentrations were too low to significantly affect sperm parameters of young patients with MM prosthesis. Further longitudinal studies are however necessary to conclusively determine the effect of metal ions from MM prosthesis on sperm parameters


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 548 - 548
1 Nov 2011
Hart A Davda K Lali F Skinner J Sampson B
Full Access

Introduction: Local problems of metal on metal (MOM) hip arthroplasty such as pseudotumours, neck thinning and osteolysis maybe related to concentrations of cobalt and chromium ions in the synovial fluid. There is little reported on these values. Our aim was to determine the range of metal ion levels in synovial fluid, and to investigate the relationship between these samples and simultaneous blood samples. Methods: Synovial fluid and whole blood samples were taken from 30 consecutive patients at the time of revision surgery for a painful MOM hip. Aspirated fluid was not visibly contaminated with blood. Impants were in situ for a mean period of 31 months. All had normal renal function. Samples were analysed using ICP mass spectrometry and compared with 10 samples from patients without implants. Results: The mean (and range) of synovial fluid metal ion levels were 1965 ug/l (30 to 13618) and 6265 ug/l (11 to 81630) for Cobalt and Chromium respectively. There was a good correlation between synovial and blood levels for both cobalt (R=0.65, p = 0.0001) and chromium (R = 0.59, p = 0.006). Discussion and Conclusion: Metal ions in synovial fluid from MOM hips are generated from wear of the bearing surfaces, the correlation with blood metal ion levels, shown in this study, suggest that blood levels may be used as surrogate marker for hip wear rate. Our range of synovial fluid metal ion levels may be useful for those conducting in vitro studies on the biocompatibility of MOM hips


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 17 - 17
1 Dec 2013
Bolland B Gardner E Roques A Maul C Culliford D Zeineh N O'Hara L
Full Access

This study reports the mid-term results of a large bearing uncemented metal on metal total hip replacement (MOMHTHR) matched series using the Synergy stem and Birmingham modular head in 36 hips (mean follow up 61 months). All patients underwent clinical, metal ion and MRI assessment. Wear analysis was performed on retrieved heads using Redlux non-contact optical profilometry. Seven patients (19%) have undergone revision surgery. All revisions had two or more of either symptoms, high metal ions or an MRI suggestive of an adverse reaction to metal debris (ARMD). There was no evidence of component malposition or impingement. Frank staining of tissues together with high volume dark brown fluid collections were found in all cases. All stems and cups were well fixed. In 4 cases pubic and ischial lysis (adjacent to the inferior fins) was observed. All 7 cases had radiological, intraoperative and histological evidence of ARMD (Figure 1). The failure cohort had significantly higher whole blood cobalt ion levels and OHS (p = 0.001), but no significant difference in cup size (p = 0.77), gender predominance, stem offset or cup position (p = 0.12). Sleeves had been used in all revision cases. Wear analysis (n = 4) demonstrated increased wear at the trunnion/sleeve interface in a distribution compatible with micromotion (Figure 2). There was normal wear at the articulating surface. This series further demonstrates unacceptable failure rates in LHMOMTHR in a series where a compatible stem for the BHR modular head was used. Use of a CoCr sleeve within a CoCr head taper appears to contribute to abnormal wear and therefore potential ARMD and subsequent failure


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 392 - 392
1 Jul 2010
Theruvil Vasukutty N Higgs D Hancock N Dunlop D Latham J
Full Access

Introduction: The advantages of metal on metal bearings (MoM) include improved wear characteristics and lower dislocation rate. Metal sensitivity and Aseptic Lymphocyte Dominated Vasculitis Associated Lesion (ALVAL) reaction are becoming increasingly recognised. The BOA has recently issued a statement regarding MoM bearings. They suggested that any revisions for symptoms of pain and soft tissue reaction should be reported to the MHRA and histological specimens forwarded to the Stanmore Retrieval Centre. Methods and results: We report four patients (all females), who presented with late dislocation of a large diameter MoM bearing (three following total hip replacements and one following resurfacing). They all made good initial recovery with complete relief of pain. They developed pain around groin between one and two years following the surgery. Inflammatory markers were normal. MRI/Ultrasound scanning showed the presence of cystic lesions around the hip. They all presented with relatively late dislocation (26 months to six years). At exploration there was a large joint effusion with necrosis and detachment of the abductors. A characteristic finding was an avascular bare trochanter stripped free of any soft tissues. Histology showed a predominantly histiocytic response in keeping with the findings of Willert and Pandit. Discussion: In the absence of any obvious causes for dislocation, one should have a high index of suspicion for the possibility of metal hypersensitivity causing joint effusion, muscle necrosis and thereby leading to instability due to the progressive periosteal and soft tissue erosion. These patients should be considered for early revision of the bearing surface to prevent further damage to the muscle and bone. If the abductors are completely detached a more constrained type of bearing should be used to improve the stability. To the best of our knowledge this is the first report of this unusual complication following MoM bearings


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 4 - 4
1 Sep 2012
Bolland B Culliford D Langton D Millington J Arden N Latham J
Full Access

This study reports the mid-term results of a large bearing hybrid metal on metal total hip replacement (MOMHTHR) in 199 hips (185 patients) with mean follow up of 62 months. Clinical, radiological, metal ion and retrieval analysis were performed. Seventeen patients (8.6%) had undergone revision, and a further fourteen are awaiting surgery (defined in combination as failures). Twenty one (68%) failures were females. All revisions and ten (71%) awaiting revision were symptomatic. Twenty four failures (86%) showed progressive radiological changes. Fourteen revision cases showed evidence of adverse reactions to metal debris (ARMD). The failure cohort had significantly higher whole blood cobalt ion levels (p=0.001), but no significant difference in cup size (p=0.77), inclination (p=0.38) or cup version (p=0.12) compared to the non revised cohort. Female gender was associated with increased risk of failure (p=0.04). Multifactorial analysis demonstrated isolated raised Co levels in the absence of symptoms or XR changes were not predictive of failure (p=0.675). However the presence of pain (p<0.001) and XR changes (p<0.001) in isolation were significant predictors of failure. Wear analysis (n=5) demonstrated increased wear at the trunnion/head interface (mean out of roundness measurements 34.5 microns (normal range 8–10 microns) with normal wear levels at the articulating surfaces. Macroscopically corrosion was evident at the proximal and distal stem surfaces. Cumulative survival rate, with revision for any reason was 92.4% (95%CI: 87.4–95.4) at 5 years. Including those awaiting surgery, the revision rate would be 15.1% with 89.6% (95% CI: 83.9–93.4). Cumulative survival at 5 years. This MOMHTHR series has demonstrated unacceptable high failure rates with evidence of high wear at the head/trunnion interface and passive corrosion to the stem surface. Female gender was an independent risk factor of failure. Metal ion levels remain a useful aspect of the investigation work up but in isolation are not predictive of failure


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 3 - 3
1 Aug 2012
Bolland B Culliford D Langton D Millington J Arden N Latham J
Full Access

This study reports the mid-term results of a large bearing hybrid metal on metal total hip replacement (MOMHTHR) in 199 hips (185 patients) with mean follow up of 62 months. Clinical, radiological outcome, metal ion levels and retrieval analysis were performed. Seventeen patients (8.6%) had undergone revision, and a further fourteen are awaiting surgery (defined in combination as failures). Twenty one (68%) failures were females. All revisions and ten (71%) of those awaiting revision were symptomatic. Twenty four failures (86%) showed progressive radiological changes. Fourteen revision cases showed evidence of adverse reactions to metal debris (ARMD). The failure cohort had significantly higher whole blood cobalt ion levels (p=0.001), but no significant difference in cup size (p=0.77), inclination (p=0.38) or cup version (p=0.12) in comparison to the non revised cohort. Female gender was associated with an increased risk of failure (chi squared p=0.04). Multifactorial analysis demonstrated isolated raised Co levels in the absence of either symptoms or XR changes was not predictive of failure (p=0.675). However both the presence of pain (p<0.001) and XR changes (p<0.001) in isolation were both significant predictors of failure. Wear analysis (n=5) demonstrated increased wear at the trunnion/head interface (mean out of roundness measurements of 34.5 microns +/−13.3 (+/−2SD, normal range 8-10 microns) with normal levels of wear at the articulating surfaces. There was evidence of corrosion at the proximal and distal stem surfaces. The cumulative survival rate, with revision for any reason was 92.4% (95%CI: 87.4-95.4) at 5 years. Including those awaiting surgery, the revision rate would be 15.1% with cumulative survival at 5 years of 89.6% (95% CI: 83.9-93.4). This MOMHTHR series has demonstrated unacceptable high failure rates with evidence of high wear at the head/trunnion interface and passive corrosion to the stem surface. This raises concern with the use of large heads on conventional 12/14 tapers. Female gender was an independent risk factor of failure. Metal ion levels remain a useful aspect of the investigation work up but in isolation are not predictive of failure


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_I | Pages 57 - 57
1 Jan 2003
Williams S Stewart TD Ingham E Stone MH Fisher J
Full Access

In vivo and in vitro studies of ceramic on ceramic (COC) bearings have demonstrated that swing phase microseparation followed by the impact of the femoral head on the superior acetabular insert rim leads to accelerated wear. However, resultant wear remained low. The wear of ceramic on polyethylene (COP) and metal on metal (MOM) couples under swing phase microseparation is unknown, this study aimed to compare the wear of these total hip replacements under standard and microseparation conditions. A physiological hip simulator was used, loads and motions were applied to approximate in vivo conditions. Microseparation was achieved by displacing the femoral head inferiorly during swing phase, the head contacted the inferior cup rim and was laterally displaced. On heel strike the head contacted the superior cup rim prior to relocation. Components (as shown in table 1) were tested for 5 million cycles, at a frequency of 1 hertz in 25% (v/v) new born calf serum. Under standard conditions, wear of COC and MOM bearings was significantly lower than wear of COP couples. Under microseparation conditions the COC and MOM wear increased by 4 and 25 times respectively. Microseparation conditions reduced wear of COP couples by a factor of 4. Creep deformation and damage to the UHMWPE cup rim was observed, however, wear remained low. It is postulated that this reduction in polyethylene wear is due to the separation of the components in swing phase improving the entrapment of lubricant, hence wear is reduced via a squeeze film lubrication mechanism


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Joyce T Langton D Lord J Grigg H Jameson S Cooke N Tulloch C Logishetty R Meek D Nargol A
Full Access

Background: There is a paucity of published data with regard to the wear of failed metal on metal (MoM) resurfacing devices. Materials and Methods: MoM components retrieved from patients from two independent centres experiencing failure secondary to ARMD were analysed using a Mitutoyo Legex 322 coordinate measuring machine (CMM) which has an accuracy of 0.8 microns. Between 4000 – 6000 points were taken on each explant, dependent on the size of the bearing surface. Maximum wear depths and total volumetric wear were calculated. These values were compared to those from control samples retrieved following uncomplicated fractures/femoral collapse secondary to avascular necrosis (after calculating equivalent yearly wear rates). Results: 58 ARMD components were analysed. This included 22 36mm MoM THRs (DePuy Pinnacle), 28 DePuy ASRs and 8 Zimmer Duroms. There were 30 resurfacing fracture/avascular necrosis controls. Volumetric wear rates and maximum wear depths of ARMD resurfacing components were significantly greater than the resurfacing control group for both the ASR and Duroms (p< 0.05) however 2 ARMD components exhibited similar amounts of wear compared to controls. Wear rates of the ARMD THR group were significantly lower than the ARMD resurfacing group (p< 0.05). Conclusions: Increased articular wear is associated with an increased incidence of local adverse effects including tissue necrosis, joint effusions and fractures. However, there are a minority of patients who can develop tissue necrosis in the absence of accelerated wear, implying a spectrum of sensitivity This is reflected in the incidence of ARMD in the patient groups at the main study centre: > 5% in the ASR group and approx 1% in the THR group. We believe this indicates a failure of adequate lubrication and the resultant negative effects in larger bearing devices


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 313 - 313
1 May 2010
Tuke M Taylor A Maul C
Full Access

Retrieval analysis offers a direct insight into in vivo wear mechanisms. However, the 3D measurement of wear patch characteristics on spherical highly reflective bearings has been difficult. An instrument based on an optical technique has been developed over the past 3 years. It is capable of scanning metallic head and cup in a single measurement, within minutes, at a resolution of 20 nm. From the cloud of 3D points obtained during scanning (typically 35,000 To 1,000,000), a 3D image of the measured part can be obtained. The associated computer program allows for sphericity, roughness, radius and local radius to be calculated, and surface maps of the 3D model can easily be plotted. Both head and cup of two failed MoM resurfacing devices, a wear simulator test couple and intact components were analysed using the new technique. A successful McKee Farrar head (20 years in vivo) was also scanned. Results were compared with traces obtained on a Mitutoyo RA 300 roundness machine (resolution 0.01 microns). 3D maps of the bearing surfaces of MoM devices were obtained. The maximum linear wear values on heads were 2.5 microns, 99 microns 53.5 microns and 298 microns for the simulator sample, the McKee Farrar head and the two failed resurfacing devices respectively. The corresponding maximum linear wear values on cups were 11 microns, 529 microns and 645 microns for the simulator sample and the two failed resurfacing devices respectively. These results were in good agreement with results obtained on the Mitutoyo machine. Contrary to other worn samples, the two latter cups showed that the cup had worn on the edge of the bearing surface. This resulted in an oval shaped wear patch on the head. For the McKee Farrar device and the simulator device, the wear patch was away from the edge and the outline of the wear patch was circular in shape. This novel technique has allowed for high resolution 3D mapping of the full bearing surfaces on successful McKee Farrar device and on more recent resurfacing devices. Further studies are required. However, the results suggest that component positioning is paramount to wear performance of metal on metal devices


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 337 - 337
1 Jul 2008
Yousuf M Shankarappa YK
Full Access

The authors report their preliminary experience with a minimum of one year follow –up of hydroxyapatite coating as the means of fixation of the femoral head in hip resurfacing. Between Dec 2003 and Dec 2004, of the 23 cases performed by the senior author,22 were available for follow up,15 were women (68.2%) and 7 were men (31.8). The femoral and acetabular components of the uncemented version of the CORMET 2000(Corin,Cire ncester,UK) were used. The surgical approach was the Hardinge approach in all cases. Patients were assessed pre-operatively for pain and function,using the Harris Hip Score. Post operatively they were assessed in clinic with x rays at 6 weeks,6 months and annually thereafter. X rays were evaluated for pre and post op neck shaft angle,giving an indication of varus or valgus placement of the head prosthesis. The lateral view was assessed to reveal anterior or posterior tilting of the prosthesis. Neck thinning was evaluated by measuring the ratio of the metal cup and bony neck diameters at the cup neck junction, recorded post op and at one year. None of the femoral implants were placed in varus. Only one case had inferior notching, which had remodelled at one year. In the lateral view none of the cups showed a displacement of more than ten degrees in the AP direction. There were no stem lucencies or signs of femoral implant migration in any of the cases. None of our cases showed neck thinning (change in ratio greater than 10%). Uncemented femoral implant in a metal on metal resurfacing hip replacement appears to perform well and shows no catastrophic problems at the short term one year follow up. Longer follow up studies are necessary


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 94 - 94
1 Mar 2010
Kim Y Kim Y Park K Choi I
Full Access

The purpose of this study is to identify clinical and radiographic results of 78 uncemented total hip arthroplasties using Metasul. ®. metal on metal bearings with Wagner standard cup and proximal hydroxyapatite coated CLS stem. Mean age was 39 years and average follow-up period was 11.7 years. Mean Harris hip score had improved from 51.4 points preoperatively to 95.2 points finally. There were 2 hips with progressive osteolysis around the acetabular cup. Of them, one hip was revised due to loosening of the cup, and the other was observed because of patient’s refusal to revise. In histopathologic findings on osteolytic area, a lot of macrophage phagocytizing metal debris and perivascular lymphocyte infiltration were found. Immunohistochemical analysis suggested delayed metal hypersensitivity. Serum cobalt levels in hips with osteolysis were not higher than those in hips without osteolysis. Early osteolysis with sudden onset of groin pain in few hips remains a concern


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 154 - 154
1 Mar 2008
Harlan C Amstuts Le Duff MJ Campbell PA Dorey FJ Beaule’ PE Gruen TA
Full Access

The purpose of the present study was to analyze the-clinical and radiographic factors which determine the enduring fixation of metal on metal hybrid surface arthroplasty. The first 500 hips (in 436 patients) of a consecutive series of over 700 Conserve® plus were reviewed clinically and radiographically. The patients were 48.6 years-old in average, 74% were male and the dominant etiology was idiopathic osteoarthritis (64.6%). All femoral components were cemented but the femoral metaphyseal stem was cemented in 112 hips and press-fit in the remaining 388. Average follow-up was 4.8 years (range, 2.7 to 7.8 years). Eighteen hips were converted to a THR, 11 of them for aseptic loosening of the femoral component at an average follow-up of 40.5 months. Put in positives mention fx and loose incidence, survor ship details of loose and fx for other At last review, 14 hips presented metaphyseal stem radiolucencies and 16 hips (3.2%) showed a narrowing of the neck of 10% or more at the interface with the femoral component. None of the components with cemented stems showed femoral radiolucencies, neck narrowing, or were revised for aseptic loosening. Cementing the femoral stem significantly (p = 0.013) reduces the chances of early femoral failure or appearance of a metaphyseal stem radiolucency. Risk factors for loosening or radioluciences included cysts > 1 cm, small femoral head size, female gender, early cases (before improved acrylic fixation including femoral suctioning measures were instituted) and reduced stem shaft angle. Metal-on-metal surface arthroplasty provides a stable, bone conserving, low morbidity and wear resistant prosthetic solution well suited for young and highly active patients


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 28 - 28
1 Jan 2013
Langton D Sidaginamale R Lord J Joyce T Natu S Nargol A
Full Access

Background. Previous studies have suggested that the modular junction of metal on metal (MoM) total hip replacements (THR) is an important source of metallic debris. Methods. We carried out a prospective study using custom techniques to analyse one of the largest collections of failed contemporary MoM devices in the world. All explants from patients who had suffered adverse reactions to metal debris (ARMD) were included in this study. These explants included: 82 36mm THRs, and 147 resurfacing head THRs and 140 resurfacing arthroplasties from several manufactures. Volumetric wear analysis of the bearing surfaces and taper junctions was carried out using a coordinate measuring machine. The relationships between total metallic loss and metal ion concentrations and the macroscopic and histological tissue appearance of THR patients were compared to those in resurfacing patients. Mann Whitney test for non-parametric data was used to assess significant differences between groups. Results. Resurfacing explants retrieved from patients who had suffered ARMD were found to have significantly higher median rates of volumetric wear than the THRs (10.16 versus 2.25mm. 3. /yr (p < 0.001)). Total volumetric material loss from taper junctions ranged from 0.01 to 21.55mm. 3. When volumetric taper wear was combined with bearing surface wear in the THR patients this total rate of material loss was still significantly less than in the resurfacing patients 2.52 versus 10.16mm. 3. /yr (p < 0.001)). Despite this, macroscopic tissue destruction and extent of ALVAL infiltration was found to be significantly greater in the THR patients. Conclusion. Taper debris appears to more readily stimulate a destructive immune cascade than debris from primary bearing surfaces. This cascade can culminate in catastrophic tissue necrosis when blood metal ion concentrations appear normal. MHRA guidance should distinguish clearly between MOM THRs and resurfacings


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 6 - 6
1 Jun 2016
Ebreo D Felgate H Martinez-Estrada F Donell S Nolan J Clark I
Full Access

Introduction. The biological pathways responsible for adverse reactions to metal debris (ARMD) are unknown. Necrotic and inflammatory changes in response to Co-Cr nanoparticles in periprosthetic tissues may involve both a cytotoxic response and a type IV delayed hypersensitivity response. Our aim was to establish whether differences in biological cascade activation exists in tissues of patients with end-stage OA compared to those with aseptic loosening of a metal on polyethylene (MoP) THR and those with ARMD from metal-on-metal (MoM) THR. Patients & Methods. A microarray experiment (Illumina HT12-v4) was performed to identify the range of differential gene expression between 24 patients across 3 phenotypes: Primary OA (n=8), revision for aseptic loosening of MoP THR (n=8) and ARMD associated with MoM THR (n=8). Results were validated using Taqman Low Density Array (TLDA) selecting the top 36 genes in terms of fold-change (FC)>2 and a significant difference (p<0.05) on ANOVA. Pathways of cellular interaction were explored using Ingenuity IPA software. Results. There is a similar pattern of gene expression between MoP and MoM phenotypes versus primary OA across 33,777 genes. One hundred and thirty significantly differentially expressed genes across 3 phenotypes were identified. Fifteen pathways were associated with differentially expressed genes between MoP and MoM phenotypes. TLDA demonstrated qualitative mirroring of the expression pattern observed in the microarray and consistency in the direction of change for individual genes. Discussion. There were no signature pathways in which multiple genes are differentially expressed such that inferences between the contributions of innate macrophage and adaptive T-cell responses can be made. TIMP3 & MMP12 were consistently identified in 15 pathways that were associated with differential gene expression between MoP and MoM phenotypes. Conclusion. Analyses of the expression of individual genes such as PRG4 (lubricin) have demonstrated patterns that may provide avenues for further research into biomarkers for periprosthetic osteolysis and ARMD


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 8 - 8
1 Jun 2016
Cook R Bolland B Tilley S Latham J
Full Access

Using a femoral head from one manufacturer on the stem of another manufacturer poses the risk that the taper interface between the components do not contact correctly and the performance of the joint will be impaired. The cohorts in this study are a combination of modular Birmingham Hip Resurfacing (BHR) and Adept femoral heads on CPT stems. The study reviews the geometry of the taper interfaces to establish if the taper clearance angles was outside of the normal range for other taper interfaces. In addition the rates of material loss from the bearings and taper and a ranking of the stem damage were reviewed to determine if the levels of loss were above that seen for other similar joints.

The material loss analysis demonstrated that the rates or levels of loss from the bearings, taper and stem are no different to levels published for manufacturer matched joints and in many cases are lower. The results demonstrate that the taper clearance angles for the mixed manufacturer joints (BHR-CPT: 0.067 to −0.116, Adept-CPT: 0.101 to −0.056) were within the range of other studies and manufacturer matched clearances (0.134 to −0.149).

Using components from different manufacturers has not in this instance increased the level of material loss from the joints, when compared to other similar manufacturer matched joints.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 39 - 39
1 Jan 2011
El-Meligy M Armstrong C Miller D
Full Access

On hundred and ninety unselected patients, mean age 62(SD12.8) were operated upon by one surgeon, using same approach. HSS improved from 33 to 75 post-operatively, the low figures reflecting a high percentage of Charnley C patients (71%). There was a 5% mortality rate from unrelated causes. Early complications included a 2.97% femoral fracture rate within 4–8 weeks of surgery (analysis showed no relation between implant position, BMI, age or Charnley classification, but there was correlation between high level of satisfaction and early weight bearing). There was one infection and one dislocation. Late complications included 3 periprosthetic fractures, 3 dislocations and 2 aseptic cup failures. However, 30% of patients developed groin pain, increasing stiffness and clunking at 1–2 years following surgery. These patients were investigated with an ultrasound scan. 28 patients (14%) had an effusion and were aspirated. Chrome and cobalt levels in synovial fluid, blood and urine were very high. 13 patients have been revised. 1 showed infection, 10 showed metal allergy and 2 a mixed picture of allergy and foreign body reaction. In our MOM database (762 hips including the current study) a total of 35 patients were revised for similar problems. Two distinct histological finding were observed. Metal allergy occurred independently of cup angle. The implants were well fixed; there was tissue necrosis and histological evidence of ALVAL. There was no infection or metalosis. FB reaction occurred more frequently when the cup was open. The Implants were loose; there was naked eye evidence of metalosis and metallic particles seen on light microscopy. We concluded that Birmingham/Freeman Hybrid had high complication rate without significant functional superiority, and other MoM devises shared similar complication. Both Metal allergy and FB reaction can occur in MOM THR. Ultra sound is a good screening tool in symptomatic patients


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 433 - 433
1 Dec 2013
Morapudi S Zhou R Khan Y Barnes K
Full Access

Introduction:

The awareness of MoM hips in the general public, one would expect, to be good, given the media coverage. This study aims to look at the perceptions and knowledge of patients who have already got a MoM arthroplasty.

Methods:

All the patients who have had a MoM hip arthroplasty from Nov 2003 to the end of 2007 were identified from the database. Postal questionnaires were sent to all the patients, the responses received and analyzed. Those patients who had symptoms and those who fulfilled MHRA criteria were invited to a clinic for further surveillance.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 318 - 318
1 Jul 2008
Gajjar SM Porter ML
Full Access

Aim of the study: To evaluate the results of metal-on-metal resurfacing hip arthropalsty in young patients

Materials and Methods: Between February 2000 and December 2004, this operation was performed in 181 patients (205 hips) using the posterior approach. The main indications were osteoarthritis, osteonecrosis, hip dysplasia, post-traumatic arthritis, perthe’s disease and slipped capital femoral epiphysis. There were 142 males and 39 females aged 26 to 68 years (average 47.4 years). 136 patients had a minimum 3 year follow-up (range 1–5 years). Clinical and radiological evaluation was performed at 3 months, 1 year, 2 years and 5 years from time of surgery. All patients were scored using the Charnley’s modification of Merle d’Aubigne score preoperatively and at last follow-up. Radiological evaluation included grading acetabular and femoral zones for loosening, determining stem-shaft angle, hip ratio. The endpoint for outcome evaluation was revision which was two in this series. Survivourship analysis was performed by Kaplan-Meier analysis. The results in our patients were encouraging with most returning to sporting activities at an average of 12 months. 1 patient with bilateral hip resurfacing had evidence of osteolysis after surgery and had occasional hip pain until last follow-up. Complications like neck fracture, heterotrophic ossification were not seen in our series.

Conclusion: Early results of metal-on-metal resufacing arthroplasty in our series are encouraging. Continued follow-up and evaluation remains important to determine the long term results of this operation.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 132 - 132
1 Mar 2009
Lohmann C Laupichler D Junk-Jantsch S Fiedler M Pflüger G
Full Access

Introduction: Metal/metal total hip arthroplasties (THA) with improved qualities of the alloys and encouraging midterm clinical results are widely used. Hyperergic reactions have been observed in revision tissues in a series of failures. This study examined synovial fluids of patients with aseptic loosening of THA from metal/metal and ceramic/polyethylene endoprostheses or arthritis of the hip by analysis of various released cytokines. Materials and Methods: The aspirations of synovial fluids from 11 patients with arthritis of the hip, 6 THA revisions with ceramic polyethylene articulations, and 22 metal/metal articulations were retrieved. 15 of the 22 cases showed lymphocytic infiltration in the histologies. The aspirates were examined with a commercially available assay using a Multiplex Reader. The interleukins Il-1 beta, -2, -5, -6, -10, -12, -13, -15,-17 and IL-1 receptor antagonist (Il-1ra) were measured. Further G-CSF, GM-CSF, IFN gamma, MIP 1 beta, MIP alpha, MCP 1, and TNF alpha were assayed. Results: Samples from patients with aseptic loosenings showed increased Il-10 and MCP compared to osteoarthritis. TNF alpha, MIP alpha, and Il-1β were increased in metal/metal THA. Il-5, Il-12, Il-13 and Il-17 were only increased in patients with lymphocytic reactions, but not in ceramic/polyethylene articulations. GM-CSF, G-CSF, IFN gamma, Il-1ra Il-2, and Il-6 were only elevated in THA with lymphocytic reactions compared to metal/metal cases without. Diskussion: Aspirates from aseptic loosened THA are important diagnostic tools. The data showed a distinctly different cytokine profile joint fluids in aseptic loosenings of metal/metal THA vs. ceramic-UHMWPE articulations and fluids from osteoarthritis patients. The data may contribute to establish a cytokine profile to determine failures due to lymphocytic infiltrations before revision of metal/metal articulations


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 113 - 113
1 Jun 2012
Kindsfater K Terefenko C Bureau C Sherman C
Full Access

Introduction. Metal on metal hip bearings (MoM) are under scrutiny. Short and mid-term complications attributed to metal wear debris have been reported. Distinctions between MoM prostheses exist. Thus, generalizing findings from one design to another is questionable. This study reports minimum 5 yr. Pinnacle™ modular MoM results. Methods. Between September 2001 and October 2004, 95 consecutive MoM THAs were performed by one surgeon in a prospective cohort design. Mean age was 53 yrs (range 34-70); 57 were male. Mean BMI was 29 (range 20-46). OA was noted in 87 patients. Head size was 28mm in 3 and 36mm in 92. Surgical approach was postero-lateral in all patients. Harris Hip Score (HHS), WOMAC, ROM and radiographs were evaluated preoperatively, at 6-months, and yearly thereafter. Results. Mean follow-up was 6 yrs (SD 1.7). Mean radiographic follow-up was 5.2 yrs (SD 1.2). Radiographs were available for 83 hips. Mean HHS improved from 43 preoperative (range 15–68) to 98 (range 86-100, p < 0.0001). Mean WOMAC score improved from 56 (range 15-94) to 12 (range 0-63, p < 0.0001). Flexion improved from 95° (range 50-115) to 102° (range 90-120, p < 0.0001). Complications included trochanteric fx (6 mo) without component removal, traumatic femoral Fx (40 days) with stem replacement, and one dislocation. No other device related adverse events were reported. No peri-acetabular osteolysis was observed. Six year survival was 100% for the acetabular shell/modular metal liner combination. Conclusion. Excellent mid-term results were obtained in a young, active cohort treated with modular MoM THA. No AVAL, pseudotumor, unexplained pain or other complication attributable to wear debris or the MoM articulation was observed. These findings indicate that THA with the Pinnacle(tm) modular MoM bearing performs well in a high demand population at mid term follow-up


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 100 - 100
1 May 2012
Bills P Underwood RJ Cann PM Hart A
Full Access

INTRODUCTION

There is increasing worldwide interest in the assessment of wear in explanted hip components. This is due is part to high profile failures of orthopaedic components in the US, whilst in the UK hip resurfacings have been experiencing a higher than expected failure rate. The reasons for these failures are not well understood, with data from the NJR suggesting the 43% of MoM resurfacing failures are unexplained.

Wear analysis is a vital tool in determining failure mechanisms and ultimately improving the longevity of joint replacements through improved design and manufacturing control. There are currently no relevant measurement standards for the evaluation of retrieved orthopaedic components. This paper will assess two of the most commonly used techniques namely roundness measurement and co-ordinate measurement. The advantages and disadvantages of both techniques are considered in this paper.

ROUNDNESS MACHINE

The Talyrond 365 is a stylus based roundness machine. The component is located on a rotating table and the stylus measures the deviation from a perfect circle as the component is slowly rotated. The Talyrond measures a single profile to an accuracy of 30 nm and up to 72,000 data points per revolution. The air spindle has a radial accuracy of <0.02 μm and the Talymin gauge a minimum resolution of 12 nm. Individual roundness profiles can be stitched together to build up 3D cylinder maps, allowing 3D pictures of sections of explanted hip components to be generated.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 173 - 173
1 Mar 2008
Thomas PW Lucente DL Mantegna DN
Full Access

The pathologies of the hip (epiphisiolysis, Perthes, congenital dysplasia) cause invalidating outcomes in young patients. In the years many authors searched for the golden standard if total hip arthroplasty is necessary. From Smith-Petersen to McKee, to Amstutz the saving of the femoral neck has been the main objective, with the longevity of the system. Therefore the combination metal-metal has been used for articular components, because the friction of metallic components creates sub-micron metallic particles causing only minimal foreign body reactions andvery low wear rates.

Our study group (Arthro Surgery Group) has implanted, from April 2003, 28 cementless resurfacing hip prosthesis with metal-metal articular components.

The average follow-up has been of 6 months (from 3 months to 1 year). In all cases, the Harris Hip Score has been good or excellent (from 77 to 100), and radiographic exams didn’t show signs of radiolucence or resorption.

This type of prosthesis allows moreover an minimum-invasive approach to the hip and a second line of defense in case of revision.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 84 - 84
1 Sep 2012
Pamu J Hussain A Daniel J Kamali A Li C
Full Access

Introduction

Ion analysis has been used as one of the key indicators to assess the performance of MoM devices in patients. Modular devices, in particular having larger overall surface area (the stem and sleeve), and locking interfaces (head – bore, sleeve- taper and sleeve-bore, stem-taper surfaces) than other MoM devices are expected to release greater number of ions. Concerns have been expressed that the ion release at the taper junction might be a potential cause leading to the failure of the implant [Garbuz et al, 2010].

The aim of this study was to look into the wear and the associated ion release from the taper junction and the articulating surface of modular devices.

Method

For the first time a novel design has been used to isolate the taper junction on modular devices on the hip simulators in order to compare the wear at the taper junction and articulating surface. The taper junction has been isolated in a small gaiter, while the head and cup were contained in a large gaiter. CoCrMo sleeves having an offset of +8 mm have been used on 50 mm modular heads along with Ti6Al4V stems. The acetabular components were standard BHR cups. Three devices (Smith & Nephew, UK) have been tested with newborn calf serum as a lubricant (in the large gaiter) and also as the medium containing the taper junction (in the small gaiter). The serum samples from the articulating surface and taper junction were analysed using HR-ICPMS. The locking interfaces at the taper junction have been left intact throughout the duration of the test. Both the head and the cup have been tested under anatomical conditions using the standard implant development centre's (IDC) profile for 2 million cycles (Mc). The lubricant was newborn calf serum with 0.2% sodium azide diluted with de-ionised water to achieve protein concentration of 20 mg/ml. The flexion/extension was 30°/15° and the internal/external rotation was ±10°. The force was Paul-type stance phase loading with a maximum load of 3 kN and a standard ISO swing phase load of 0.3 kN. The frequency was 1 Hz, with an 8 hour stop after every 16 hours of testing.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVII | Pages 8 - 8
1 May 2012
Cronin M Lewis D John A Jones S
Full Access

INTRODUCTION. The risk of dislocation in large diameter metal on metal hip replacement is significantly lower than in standard THR. This is due to the increased primary arc, increased jump distance and possibly a suction effect. Our unit has performed over 1500 of these cases with an overall revision rate of <1%. We report a case series of dislocations in 5 large diameter metal on metal hips undertaken at our unit. METHOD. All cases were reduced closed and investigated for cause of dislocation. Radiological investigation included plain film radiographs and CT to exclude component mal-position and MRI to document soft tissue deficiency. Metal ion levels were measured and microbiological investigation was undertaken. RESULTS. In all cases component positioning was acceptable. Metal ion levels were significantly elevated with levels comparable to published work. MRI showed significant soft tissue defects in all patients. At revision all were found to have necrotic areas associated with a large turbid effusion. Histology confirmed metal related inflammatory change and microbiology specimens confirmed the absence of infection. DISCUSSION. The local detrimental effects of metal debris are well documented. We believe this is the first series with a proven link between these soft tissue problems and dislocation. Every case of large diameter metal on metal hip in our unit that has dislocated has been proven to be associated with metallosis and has required revision


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_III | Pages 547 - 547
1 Aug 2008
Chambers MJ Rana BD Kelly MP Gray AJR Roberts J
Full Access

Introduction: We hypothesised that a stemmed hybrid total hip replacement with a large metal head, (LMOM), with the same bearing technology, would give early functional results as good as hip resurfacing (HR) with fewer contraindications and reduced morbidity due to a less invasive approach. Methods: We performed a 12 month observational study of consecutive patients undergoing LMOM (n=40) and HR (n=60). Patients’ age, sex, blood loss, hospital stay and early complications were compared as well as pre and post-op Oxford scores and range of movement. Results: The HR group was younger with more normal hip morphology. Mean age 54years HR/ 59years LMOM, BMI 28 and 29 respectively. Diagnosis of primary osteoarthritis 77% HR compared to 47% in LMOM group. Operation time, blood loss and haemoglobin drop were similar in both groups. The LMOM group achieved an earlier discharge of 5.5 days compared to 6.3 (p< 0.12). Complications included 2 aseptic wound leakages in the LMOM group. In the HR group an aseptic wound leakage, superficial infection and a sciatic nerve palsy was observed. At 6 week and 3 month follow up, the range of moment of both groups was very similar. Mean Oxford hip scores were 44 preoperatively and 22 postoperatively in the HR group and 45 falling to 24 in the LMOM group. Discussion: LMOM compared to HR is preferable in respect to hospital stay and reduced perioperative complications despite an older cohort with a wider variety of arthritic pathologies. However longer follow up is required to conclude further


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 168 - 168
1 Mar 2008
Iavicoli I Falcone G Alessandrelli M Salvatori S Cresti R De Santis V Caroli S Carelli G
Full Access

During the past fifty years metal hip implants have been used in orthopaedics. While these implants are considered biologically inert, several studies indicated that prosthetic implants could release metal ions following wear. The aim of this study was to evaluate serum and urine concentrations of metal ions in patients treated with hybrid surface replacement of the hip. Co, Cr, Mn, Mo, and Ni in the serum and urine of 14 patients and 19 controls were quantified by Sector Field Inductively Coupled Plasma Mass Spectrometry (SF-ICP-MS). A Spectrometer ELEMENT (Finnigan MAT, Bremen, Germany) was used at the following resolution: 98Mo, 100Mo, (m/D;m = 300); 52Cr, 53Cr, 55Mn, 59Co, 60Ni, (m/D;m = 3000). The internal standard method (In) was applied to correct the random fluctuations of the signal. Indium was added to the sample solutions so as to reach a final concentration of 1 ng mL-1. Limits of Detection were calculated on the basis of the 3=555; criterion for 10 replicate measurements of solutions containing dilute pooled blood serum and dilute pooled urine, and recovery was obtained by quantifying the analytes of interest in spiked serum and urine. Recoveries of the elements ranged between 94.6% (serum Cr) and 118% (urine Ni). Data support the assumption that the concentrations of Co and Cr (the major components of the alloy) are substantially higher in serum and urine of patients than in those of controls while for the other elements no specific trend could be observed. Actually there isn’t any explanation on the possible biological consequences related to the increase of the metal ions levels although many hypothesis have been made. Further studies are thus mandatory to clarify this pattern


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 547 - 548
1 Nov 2011
Underwood R Cann P Ilo K Wagner C Skinner J Cobb J Porter M Muirhead-Allwood S Hart A
Full Access

Introduction: The London Implant Retrieval Centre (LIRC) was founded to investigate the high incidence of unexplained failures of Metal-on-Metal (MoM) hips. A multidisciplinary team analyse the failed hips, investigations include CT and MRI scans, blood and synovial fluid tests, wear measurements, X-rays and clinical data from the explanting surgeons.

Wear measurements of 100 explanted hips have been carried out on a Taylor Hobson 365 Roundness Machine using the LIRC Wear Protocol. It was found that 50% of explanted cups were wearing less than 5 μm/year and 60% of components were wearing less than 10 μm/year. Wear tests on hip joint simulators predict wear rates between 2 and 8 μm/year. However, 6% of cups are wearing faster than 100 μm/year, with 16% of cups have wear patches deeper than 100 μm and that 4% have a wear patch deeper than 300 μm.

Discussion: This paper considers the common characteristics of components in this very high wearing category. Engineering parameters such as head/cup clearance, surface finish, form errors and head cup contact conditions are investigated. This is correlated with clinical data and other results from the LIRC.

Cup position is an important factor, all of the high wearing components are outside the Lewinick’s Box, however it is shown that mal position is does not always lead to extreme wear. Further analysis is taking place to calculate the size of the contact patch between head and cup (based on patient data and biomechanics) and the proximity of the contact patch to the edge of the cup.

Conclusion: The study of explanted components shows that 6% exhibit extreme wear, and although several “risk” factors can be identified, it is not clear why only a proportion of these components show extremely high wear rates. This is the subject of current investigation.


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_II | Pages 158 - 158
1 Feb 2003
Nolan J Darrah C Fairman B Fleming J
Full Access

This prospective study evaluates the outcome of a new metal -on-metal total hip replacement in a younger group of patients. Fifty-five primary all-metal total hip replacements (THR) were evaluated prospectively at a follow-up of 2.8–5.5 years. Patients were selected according to age and activity levels. The mean age was 58 years (41–69). 33 males and 22 females were included in the study. Surgery was carried out for osteoarthritis in 52 patients and for non-union fractured femoral neck, ankylosing spondilitis and post slipped upper femoral epiphysis in the three remaining patients. A single surgeon (the senior author) through the posterior approach carried out the surgery. All patients received the porous coated titanium shell with a Morse taper cobalt chrome liner and double wedge tapered polished cobalt chrome stem and modular head. Blood metal ion analysis was performed on a cohort of 24 patients using High Resolution Inductively Coupled Plasma Mass Spectrometry, sampling taken preoperatively and then repeated post operatively at 6 months, 1 year and then annually. Clinical results have been excellent. X rays show Harris A cementation in all femurs, with no component migration or radiolucencies being identified on follow-up radiographs. No prosthesis to-date has required revision. One patient has died and one is lost to follow up. The following non-device related complications were reported in the group, 2 (4%) superficial wound infections, 1 (2%) dislocation, 1 (2%) thrombosis, 1 (2%) IT band defect and 2 (4%) impingement. The dislocation was treated with a closed reduction. The impingement has resolved by one year in both patients. The results of the pre and postoperative blood metal ion analysis demonstrate some elevated levels, these levels being similar to those previously reported in the literature. The hybrid all-metal THR may represent a valuable alternative in the younger, high demand patient


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 1 - 1
1 Jan 2011
Jeavons L Dixon S Reddy R Fern E Norton M
Full Access

We report a 10% failure rate for aseptic loosening and overall revision rate of 15% at 5 years mean follow up in 190 patients using the Cormet 2000 Dual coat ace-tabular component.

Between April 2001 and March 2004, this cup was used in our region by 4 surgeons. It was peripherally expanded and comprised a cobalt chrome shell, plasma sprayed with a further layer of cobalt chrome which was then coated with hydroxyapatite. This backing has since been abandoned in favour of a titanium plasma coat beneath the hydroxyapatite because of a higher than expected incidence of early loosening.

There were 190 cups implanted in 174 patients, 142 with resurfacing heads. The average age was 54 and 99 were male.

Revision for aseptic loosening was required in 20 cups (10%) at a mean interval of 23 months including five within 2 months, Of the early revisions three developed a deep infection. Undiagnosed groin pain in four further patients appeared at revision to arise from an inflamed bursa secondary to impingement of psoas tendon on the rim of the cup. There was one primary infection, one infection following revision of a trochanteric osteotomy and two neck fractures. Persistent unexplained groin pain was seen in three patients who have declined further surgery

Failure of this backing to integrate with bone led to an unacceptably high early loosening rate. Positioning of the larger resurfacing cups is critical to avoid painful irritation of psoas. Early revision carries a high risk of infection.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 391
1 Jul 2010
Langton D Jameson S Joyce T Natu S Logishetty R Tulloch C Nargol A
Full Access

In our independent centre, in the period from January 2003 to august 2008, over 1100 36mm MoM THRs have been implanted as well as 155 Birmingham Hip Resurfacing procedures, 402 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems. During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 11 failures of this nature in patients with ASR implants (10 females) and 2 in the 36 MoM THR group (one male one female). Tissue specimens from revision surgery showed varying degrees of ‘ALVAL’ as well as consistently high numbers of histiocytes. Metal debris was also a common finding. A fuller examination of our ASR cohort as a whole has shown that smaller components placed with inclinations > 45° and anteversions < 10 or > 20° are associated with increased metal ion levels. The 11 ASR failed joints were all sub optimally positioned (by the above definition), small components. Explant analysis using a coordinate measuring machine and out of roundness device confirmed greater than expected wear of each component. The lower number of failures in the 36mm MoM group, as well as the equal sex incidence, suggests that the majority of these failures are due to the instigation of an immune reaction by large amounts of wear debris rather than adverse reactions to well functioning joints. It is likely that small malpositioned ASRs function in mixed to boundary lubrication, and this, combined with the larger radius of these joints compared to the 36mm MoM joints, results in more rapid wear


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_6 | Pages 9 - 9
1 May 2015
Veettil M Ward A Smith E
Full Access

We retrospectively reviewed the mid term outcome of 88 MoM THA in 84 patients and 21Hip Resurfacing using Recap Magnum bearing surface performed during 2006 – 2009.

There were 41 males and 47 females in the THA group and 17 males and 4 females in the Hip Resurfacing group.

All procedures were performed through a posterior approach. The average head size for the THR group was 46mm and the cup size was 52mm and the average head size for the resurfacing was 50mm and cup size was 56mm respectively.

Median age for the THA group was 60 yrs. (28–73) and for the Resurfacing it was 51.5 yrs. (32–62). Average follow up was 76 months for the THA group and 78 months for the Resurfacing group.

Average serum cobalt for the THA and the Resurfacing groups were 53.2nmol/l (119) and 30.85 and the Chromium levels were 82.44nmol/l(134.5) and 67.49 respectively. Eight MRI scans showed abnormal fluid collections suspicious of ARMD in the THA group and 2 showed fluid collection in the Resurfacing group. There were five revisions in the THA group with the tissue diagnosis of ALVAL. In all except one case a well fixed uncemented stem (Taperloc) was retained.

In our series Recap Magnum on a Taperloc stem showed 94% survival at five years and therefore we continue to review the cases annually with serum cobalt chromium levels and MRI scans.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 25 - 25
1 Aug 2012
AL-Hajjar M Fisher J Williams S Tipper J Jennings L
Full Access

In vitro the introduction of microseparation and edge loading to hip simulator gait cycle has replicated clinically relevant wear rates and wear mechanisms in ceramic-on-ceramic bearings[1], and elevated the wear rates of MoM surface replacements (SR) to levels similar to those observed in retrievals[2]. The aim was to assess the wear of two different sized MoM total hip replacement bearings under steep cup inclination angles and adverse microseparation and edge loading conditions.

Two tests were performed on the Leeds II hip joint simulator using two different size bearings (28mm and 36mm). Cups were mounted to provide inclination angles of 45 degrees (n=3) and 65 degrees (n=3). The first three million cycles were under standard gait conditions. Microseparation and edge loading conditions as described by Nevelos et al[1] were introduced to the gait cycle for the subsequent three million cycles. The lubricant was 25% new born calf serum. The mean wear rates and 95% confidence limits were determined and statistical analysis was performed using One Way ANOVA.

Under standard gait conditions, when the cup inclination angle increased from 45 degrees to 65 degrees, the wear of size 28mm bearing significantly (p=0.004) increased by 2.7-fold, however, the larger bearings did not show any increase in wear (p=0.9). The introduction of microseparation conditions resulted in a significant (p=0.0001) increase in wear rates for both bearing sizes under both cup inclination angle conditions. Under microseparation conditions, the increase in cup inclination angle had no influence on the wear rate for both bearing sizes (Figure 1).

With larger bearings, head-rim contact occurs at a steeper cup inclination angle providing an advantage over smaller bearings. The introduction of edge loading and microseparation conditions resulted in a significant increase in wear rates for both bearing sizes. The wear rates obtained in this study under combined increased cup inclination angle and microseparation were half of those obtained when SR MoM bearings were tested under similar adverse conditions[2]. This study shows the importance of prosthesis design and accurate surgical positioning of the head and acetabular cup in MoM THRs.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_9 | Pages 23 - 23
1 Feb 2013
Parry M Bhabra G Sood A Figgitt M Case P Blom A
Full Access

Orthopaedic cobalt chromium particles and ions can induce indirect DNA damage and chromosome aberrations in human cells on the other side of a cellular barrier in tissue culture. This occurs by intercellular signalling across the barrier. We now show that the threshold for this effect depends on the metal form and the particle composition. Ionic cobalt and chromium induced single strand breaks at concentrations equivalent to those found in the blood of patients with well functioning metal on metal hip prostheses. However, they only caused double strand breaks if the chromium was present as chromium (VI), and did not induce chromosome aberrations. Nanoparticles of cobalt chromium alloy caused DNA double strand breaks and chromosome aberrations, of which the majority were tetraploidy. Ceramic nanoparticles induced only single strand breaks and/or alkaline labile sites when indirectly exposed to human fibroblasts. The assessment of reproductive risk from maternal exposure to biomaterials, especially those liberated by orthopaedic implants, is not yet possible with epidemiology. Whilst the barrier model used here differs from the in vivo situation in several respects, it may be useful as a framework to evaluate biomaterial induced damage across physiological barriers


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 413 - 413
1 Dec 2013
Garofolo G Snir N Park B Wolfson T Hamula M Marwin S
Full Access

Background. Revision surgery for failed metal-on-metal (MOM) total hip arthroplasty (THA) or hip resurfacing (HR) has been a challenge. Previous studies have reported high failure and complication rates, including dislocation, infection, aseptic loosening and lower patient satisfaction. Options for revision depend on the integrity and stability of the femoral and acetabular components. When both components fail, full revision is required; however, when the acetabular component remains well fixed and oriented, only the isolated femoral component revision can be performed. Dual mobility components can be utilized to match the size to the inner diameter of the metal cup. With the dual mobility implant, the morbidity and complications associated with cup revision are avoided while maintaining a natural femoral head size and potentially increasing range of motion and stability postoperatively compared to standard THA. Purpose. The aim of this study was to evaluate short- to mid-term results of revision THA after failed metal-on-metal THA or HR using the dual mobility device. Study Design. Retrospective case series with prospective follow-up. Methods. A cohort of consecutive patients who underwent revision THA for failed MOM THA or HR utilizing a dual mobility device with a minimum follow-up of a year was identified. Charts were retrospectively reviewed for surgical indications, comorbidities, concomitant procedures, cup size, inner head size, outer head size, and perioperative complications, including infection, dislocation, mechanical failure and reoperation. Visual analogue pain scale (VAS), modified Harris Hip Score (mHHS) and SF-12 questionnaires were collected prospectively to assess functional outcomes after THA revision with a dual mobility component. Results. Fifteen consecutive patients (16 hips) underwent revision surgery utilizing a dual mobility component. Six hips were indicated for failed metal-on-metal THA and 10 for failed HR. The mean follow-up was 20 months (range, 12–29 months) and the mean VAS scores decreased from 8.9 preoperatively to 3.8 postoperatively (p < 0.01). The mean mHHS score increased from 26.9 preoperatively to 57.8 postoperatively (p < 0.05). A statistically significant improvement in the mean SF-12 scores was also noted (p < 0.05). Complications consisted of two patients with residual chronic pain. No dislocations, fractures, or infections were observed. Conclusion. Single component revision THA for failed MOM THA or HR utilizing a dual mobility device is an effective and relatively simple procedure for a complex problem


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 130 - 130
1 Dec 2013
Morapudi S
Full Access

Introduction:. The clinical significance of serum metal ion levels alone in patients with stemmed MoM hip arthroplasty remains uncertain. This study aims to measure the metal ion levels in patients with unilateral and bilateral hip arthroplasty to see whether or not the patients with bilateral hips have higher metal ions. Patients and Methods. All the patients with bilateral MoM hip arthroplasty were identified from the research database. These patients were then matched with those in a similar age group who had a unilateral hip arthroplasty, but same University of California Los Angeles (UCLA) activity score. The UCLA activity scores along with other hip scores have been previously gathered for all the patients. Both sets of patients then had measurement of serum cobalt and chromium levels. The results were analyzed and compared. Results:. There were a total of 53 patients with bilateral hips and 53 matched patients with unilateral hips. The average serum Cobalt and Chromium levels were 34 nmol/L (range <10–76) and 23 nmol/L (range <10–104) respectively in the bilateral group; and 14 nmol/L (range <10–45) and 21 nmol/L (range <10–57) in the unilateral group. Given that the MHRA (UK) threshold levels deemed clinically significant are 120 nmol/L and 135 nmol/L for Cobalt and Chromium respectively; the results obtained here are not clinically significant. Conclusions:. One would normally expect, with mathematical reasoning, a patient with bilateral hip arthroplasty to have higher serum metal ion levels than a patient with unilateral arthroplasty. However, this study goes to show that it is not the case


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 423 - 423
1 Apr 2004
McMinn D
Full Access

Development of the metal/ metal hip resurfacing began in 1989 with the first patient implantation in February 1991. In the first three years three methods of implant fixation were employed in a pilot study of 119 resurfacings and the optimum fixation was identified as Hydroxyapatite Coated (HA) uncemented cups and cemented femoral components. From March 1994 until February 2001 the author has performed 1503 hybrid fixed resurfacing implants, 294 of the McMinn type (HA on Smooth Metal Cup) and 1209 Birmingham Hip resurfacings (BHR; HA on Porous Metal Cup). Mean age was 52.9 years, diagnosis – Osteoarthritis 78%; Dysplasia 7%; Inflammatory 2%; Avascular necrosis 5%; other 8%. In the 294 McMinn Resurfacings 8 have failed and 7 have been revised. Reasons for failure; 5 cup loosenings, two infections and one femoral head collapse in a patient with pre-existing avascular necrosis. High wear of the metal/ metal bearings, with altered metallurgy, is now presenting as a cause for concern with these McMinn resurfacings performed in 1996 where the metal was Solution Heat Treated (SHT) and Hot Isostatically Pressed (HIP). Patients from 1996 are exhibiting asymptomatic radiographic osteolysis and at revision metallosis is seen with high wear measured on the retrieved components. Laboratory studies have shown that SHT and HIP alters the carbide microstructure and increases wear on abrasive wear testing. In the 1209 BHR’s 7 failures have occurred requiring revision. Reasons for revision – 4 femoral neck fractures, 2 infections and 1 collapsed femoral head from avascular necrosis. Implant fixation is reliable as judged by very low migration on an R.S.A. study. Bearing wear is very low as judged from wear analysis of explants as would be expected from the historically proven “As Cast” metal microstructure. Function is good as judged by the Harris Hip Score (mean 96.3 at 3years). Cumulative survival on all Hybrid fixed implants is 97% at 7–8 years. Hip resurfacing, using metal on metal bearings with historically proven metallurgy using hybrid fixation, offers a viable treatment alternative for the young and more active patient with hip arthritis


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 3 - 3
1 Mar 2008
Davies A Willert H Campbell P Case C Learmonth I
Full Access

Metal-on-metal bearing surfaces have been reintroduced for use in total hip replacement, despite concerns regarding the potential risks posed by metallic by-products. We have compared periprosthetic tissues from metal-on-metal and metal-on-polyethylene hip replacements at revision surgery with control tissues at primary arthroplasty.

Tissues were obtained from 9 control, 25 contemporary metal-on-metal, 9 CoCr-on-polyethylene and 10 titanium-on-polyethylene hip replacement arthroplasties. Each was processed for routine histology with Haematoxylin and Eosin. Quantitative stereological analysis was performed at the light microscopic level.

Metal-on-metal sections showed more surface ulceration and this was correlated with the density of inflammation in the deeper tissues layers. Metal-on-metal tissues displayed a pattern of well-demarcated tissue layers, which were rarely seen in metal-on-polyethylene cases. In metal-on-polyethylene cases, the inflammation was predominantly histiocytic. Metal-on-metal cases by contrast showed a lymphocytic infiltrate with abundant plasma cells. Metal-on-metal tissues showed a striking pattern of peri-vascular inflammation with prominent lymphocytic cuffs especially deep to areas of surface ulceration. Levels of inflammation were higher in cases revised for failure than in those retrieved at autopsy or exploratory surgery. Total replacement and surface replacement designs of metal-on-metal arthroplasty showed similar histological changes. Plasma cells were not seen in any of the metal-on-polyethylene cases. The differences between the patterns of inflammation and cellular infiltration seen in metal-on-metal and metal-on-polyethylene tissues were highly statistically significant.

The pattern and type of inflammation in periprosthetic tissues from metal-on-metal and metal-on-poly-ethylene arthroplasties is very different. Our findings support the conclusion that metal-on-metal articulations are capable of generating a form of immunological response to metallic wear debris that has not been described previously. The incidence and clinical implications of these immunological responses in failed metal-on-metal joints are unknown.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 333 - 333
1 Mar 2004
Amstutz HC BeaulŽ PE Gruen TA Le Duff MJ
Full Access

Aims: To review the medium term clinical and X-Ray results of DDH in SA, using the Conserve Plusª. Methods: 28 hips (26 patients, avg. age 43.2 years) with DDH were treated with hybrid MMSA. 81% were female. 42% had prior osteotomies. 25% had acetabular cysts that were grafted and 64% had femoral cysts (> 1cm in size) & 1 was grafted. A posterior approach was used in all but one case. Results: Mean F/U was 4.1 years (3.3– 6.5). UCLA hip scores were 3.3, 6.0, 5.7 and 4.5 pre-op and 9.2, 9.3, 9.3 and 7.1 at last F/U for pain, walking, function and activity. SF-12 scores normalized. Range of motion was 121.3û, 73.7û and 83.7û at last F/U for Flexion, abductionadduction, and rotation arcs. 2 hips were converted to THA: 1 for femoral neck fracture at 2 months (neck osteopenia and cystic degeneration) and 1 for recurrent subluxations due to poor offset and impingement due to prior multiple osteotomies. 4 technical complications successfully resolved: 1 trochanteric non-union, 1 post-op dislocation (closed reduction), 1 component mismatch (acetabular component exchange) and 1 femoral nerve palsy. X-ray analysis: 2 pts. presented complete radiolucencies around the metaphyseal stem and 1 patient was symptomatic and will require revision). Conclusions: SA is stable, except for 1 subluxation due to impingement with poor off-set compared to 7–10% with THA. The socket þxation is secure with grafts incorporating whereas prior SA with PE were poor due primarily to socket loosening. Femoral bone preparation is critical to avoid fractures and loosening.


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_II | Pages 158 - 158
1 Feb 2003
Amstutz H Beaulé P Gruen T Le Duff M
Full Access

To review short to medium term results of a metalon-metal (M/M) hybrid surface arthroplasty (SA) for a young and active patient population. The first 300 hips (of 564) in 263 patients underwent M/M hybrid surface arthroplasty (cementless acetabular and cemented femoral components). Demographics: mean age 48.4; 75% males, 25% females; 141 Charnley Class A; 109 Class B and 13 Class C. Diagnosis at surgery: OA 67%, DDH 10.3%, ON 8.3%, Post-traumatic 6.7%, Inflammatory arthritis and Rheumatoid Disease 4%, SCFE and LCP 3.4%, Melorheostosis 0.3%. Mean follow-up 3.6 years (2.5–5.9). Average UCLA hip scores post-op: pain 9.4, walking 9.5, function 9.4 and activity 7.7. Average Harris Hip Score was 92.8. The SF-12 physical and mental components were respectively, 31.4 and 47.2 pre-op and 49.9 and 52.9 post-op. DEXA data suggests preservation or restoration of neck BMD. The experience with SA of all cemented metal/ UHMWPE bearing demonstrated failure rates of 15%–33% at 3 years. At longer follow-up, the preliminary experience is encouraging (3.6% failure rate) and eventual conversions to THR are facilitated with unipolar heads. The technique preserves femoral and ace-tabular bone, dislocation is rare, and acetabular fixation secure. Initial femoral fixation is critical as the fixation area is small, especially with osteopenia and cystic degeneration. The percent of potentially adverse radiographic changes was much greater in the first 100 cases, during the time of development of instrumentation, technique, and bearing optimisation


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims. The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Methods. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis. Results. Mean follow-up was 20.9 years (19.3 to 22.4). Registry data revealed that 19 hips (8.1%) had been revised and 26 patients (12%) had died from causes unrelated to the BHR. Among the remaining 189 hips, 61% were available for clinical follow-up at 20 years (n = 115) and 70% of patients had biochemical follow-up (n = 132). The cumulative implant survival rate at 20 years for male patients was 96.5% (95% confidence interval (CI) 93.5 to 99.6), and for female patients 87% (95% CI 79.7 to 94.9). The difference was statistically significant (p = 0.029). The mean Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, and Forgotten Joint Score were 45 (29 to 48), 89 (43 to 100), and 84 (19 to 100), respectively. The mean scores for each of the five domains of the EuroQol five-dimension three-level questionnaire were 1.2, 1.0, 1.2, 1.3, and 1.1, and mean overall score 82.6 (50 to 100). Ultrasound showed no pseudotumour. Mean cobalt and chromium levels were 32.1 nmol/l (1 to 374) and 45.5 nmol/l (9 to 408), respectively. Conclusion. This study shows that BHRs provide excellent survivorship and functional outcomes in young male patients. At 20 years, soft-tissue imaging and serum metal ion studies suggest that a metal-on-metal resurfacing implant can be well tolerated in a group of young patients. Cite this article: Bone Joint J 2023;105-B(9):946–952


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 403 - 404
1 Apr 2004
Philippe T
Full Access

Our experience with metal/metal desings in France date from 1994. The goal of this study is to communicate our primary and comparative clinical results of 90 implants followed a minimum of 6 years. Material and method: We have reviewed the first 94 patients from our first year using metal/metal.90 case files were complete and we compare thes patients (group 1) to a control group of 90 patients receiving ceramic/poly couple, operated in the same time period (group two) with the same hybrid TH system. Average follow up: Five years and eight months. Results and discussion: 4 of the 94 were unavailable for follow up at six years. None of these patients experienced clinical problems relared to the metal/metal. The clinical results (Merle and Harris score) are all good or excellent and there is no difference between the 2 groups. The x-ray results: For acetabular cup the two groups are identical and for femoral component (Barrack evaluation) all two groups have the same quality of cementation and maintain the cement mantle integrity achieved post-op. We decided therefore to examine specially the RLL changes in zone seven (comparison 3 month x-ray/latest follow-up x-ray); significant differences between the two groups were found in this study: The lower incidence of zone seven radiolucent changes in group one indicates a better clinical tolerance in the metal/metal configuration (p=0.0256). Considering the results reported here and the advanced studies, we can confirm that metal/metal systems are not the unique answer to PE-wear and THA longevity, but it is one solution that is available today and seems trustable


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 361 - 361
1 Mar 2004
Amstutz HC BeaulŽ PE Gruen TA Le Duff MJ
Full Access

Aims: Review short to medium term results of a metal-on-metal (M/M) hybrid surface arthroplasty (SA) for a young and active patient population. Methods: The þrst 400 hips (of 533) in 355 patients underwent M/M hybrid surface arthroplasty. Mean age 48.2; 73% males, 27% females; 198 Charnley Class A; 139 Class B and 18 Class C. Diagnosis at surgery: OA 64%, DDH 11%, ON 9%, Post-traumatic 8%, Inßammatory arthritis 3%, SCFE and LCP 4%, Melorheostosis 0.3%. Results: Mean follow-up 3.9 years (2.6–6.5). Average UCLA hip scores post-op: pain 9.4, walking 9.5, function 9.4, and activity 7.7. The SF-12 physical and mental components were respectively, 31.2 and 46.8 pre-op and 50.1 and 53.0 post-op. 12 hips were converted to THA (8 in the þrst 102): 2 neck fractures, 7 femoral loosenings, 1 socket protrusio due to over-reaming, 1 subluxation due to impingement and 1 sepsis. Conclusions: Clinical results are excellent despite very high activity levels. The experience with SA of all cemented metal/UHMWPE bearing demonstrated failure rates of 15%–33% at 3 years. At longer follow-up, the preliminary experience is encouraging (3% failure rate). Dislocation is rare, and acetab-ular þxation secure. Initial femoral þxation is critical as the þxation area is small, especially with osteopenia and cystic degeneration. The percent of complications and potentially adverse radiographic changes was much greater in the þrst 100 cases, during the time of development of instrumentation, technique, and bearing optimization


The authors did not receive grants or outside funding in support of their research or preparation of this manuscript. They did not receive any payments or make agreements to provide such benefit from a commercial entity. No commercial entity paid or directed, or agreed to pay or direct, any benefits to any research fund, foundation, educational institution, or other non profit organization with which the authors are affiliated or associated. Background: We observed early osteolysis and loosening of the rough surface-cement stem with a second generation metal-on-metal articulation. This study was to investigate the possible etiologic role of stem loosening and osteolysis by examination of the surface of the rough blast cement and related periprosthetic tissues that have been retrieved at revision. Methods: We retrospectively analyzed 39 hips (37 patients) with use of metasul metal-on-metal total hip arthroplasty. Of the femoral stem, rough surface-cement stem used in 18 hips and cementless Ti-alloy stem in 21 hips. The mean duration of follow-up was 10.2 years. Of these eight rough blast cement stem were revised, seven in loosening and one in recurrent dislocation. By using energy disperse spectroscope and back scattered electron image, histologic studies were performed to the samples of periprosthetic tissues. Skin patch tests for metal hypersensitivity were done to select patients. Results: All of cementless stems and cups showed excellent results at the last follow-up. However, eight cement hips were revised. Light microscopy showed polishing effect on retrieval femoral stem affected by the rotational force. EDS and BSE image revealed that there were abundant cement and related metal particles with size of 5-10μm. However there were few metal particles and had greater size (20–100μm) in periprosthetic tissues. Histologic finding shows perivascular infiltration of lymphocytes and accumulation of macrophages No relation was found between skin patch test and loosening. Conclusion: These findings raise the possibility that early osteolysis and loosening in patients with metal-on-metal hip replacement were associated with rough blast surface cement stem. These mixed particles such as Fe and Zr maybe trigger lymphocytic reactivity suggestive delayed type hypersensitive reaction. This study suggests that cement stem which have rough blasted surface should be considered in metal-on-metal total hip arthroplasty. Level of Evidence: Therapeutic level III-1(case-control study). See instructions to authors for a complete description of level of evidence


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 7 - 7
1 Apr 2019
Paulus A Dirmeier S Hasselt S Kretzer P Bader R Jansson V Utzschneider S
Full Access

Introduction. It is well-known that wear debris generated by metal-on-metal hip replacements leads to aseptic loosening. This process starts in the local tissue where an inflammatory reaction is induced, followed by an periprosthetic osteolysis. MOM bearings generate particles as well as ions. The influence of both in human bodies is still the subject of debate. For instance hypersensitivity and high blood metal ion levels are under discussion for systemic reactions or pseudotumors around the hip replacement as a local reaction. The exact biopathologic mechanism is still unknown. The aim of this study was to investigate the impact of local injected metal ions and metal particles. Material and Methods. We used an established murine inflammation model with Balb/c mice and generated three groups. Group PBS (control group, n=10) got an injection of 50µl 0.1 vol% PBS-suspension, Group MI (Metal-ion, n=10) got an injection of 50µl metal ion suspension at a concentration of 200µg/l and Group MP (Metal-particles, n=10) got an injection of 50µl 0.1 vol% metal particle suspension each in the left knee. After incubation for 7 days the mice were euthanized and the extraction of the left knee ensued. Followed by immunhistochemical treatment with markers of inflammation that implied TNFα, IL-6, IL-1β, CD 45, CD 68, CD 3, we counted the positive cells in the synovial layer in the left knees by light microscopy, subdivided into visual fields 200× magnified. The statistical analysis was done with Kruskal-Wallis test and a post hoc Bonferroni correction. Results. The Group with metal particles showed significantly elevated inflammatory markers (TNFα, IL-6, IL-1β, CD 68, CD 45) compared to all other groups. Interestingly, CD 3 as a marker for T-lymphocytes showed no increased levels in all groups. The metal ion group showed significant elevated CD 45 expressions compared to the control group. Conclusion. The results clearly demonstrate that especially metal wear particles lead to an intensive inflammatory reaction. The tissue formations in the metal particle group show an osseous destructive behavior in previously demonstrated results, comparable to pseudotumors. But, in this study, the expression of the immunohistological markers CD 3, CD 45 and CD 68 indicate that the tissue consists of leucocytes and macrophages, whereas lymphocytes could not be detected. This might be due to an acute inflammatory reaction, whereas the adaptive immune response by T-lymphocytes seems not (yet) to be activated. Overall it must be stated that solid metal wear particles are responsible for local inflammatory reactions, whereas it is still unknown whether wear particles corrode in vivo and release a potentially high level of locally toxic metal ions


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1018 - 1024
1 Aug 2018
Ando W Yasui H Yamamoto K Oinuma K Tokunaga H Inaba Y Kobayashi N Aihara M Nakanishi R Ohzono K

Aims. The purpose of this study was to compare two different types of metal-on-metal (MoM) bearing for total hip arthroplasty (THA): one with a large femoral head (38 mm to 52 mm) and the other with a conventional femoral head (28 mm or 32 mm). We compared clinical outcome, blood metal ion levels, and the incidence of pseudotumour in the two groups. Patients and Methods. Between December 2009 and December 2011, 62 patients underwent MoM THA with a large femoral head (Magnum group) and 57 patients an MoM THA with a conventional femoral head (conventional group). Clinical outcome was assessed using the Harris Hip score, University of California, Los Angeles (UCLA) activity score and EuroQol-5D (EQ-5D). Blood metal ion levels were measured and MRI scans were analyzed at a minimum of five years postoperatively. Results. No acetabular component was implanted with more than 50° of inclination in either group. The Harris Hip Score, UCLA activity score, and EQ-5D improved postoperatively in both groups; no significant clinical differences were noted between the groups. The blood cobalt ion levels in the conventional group continued to rise postoperatively to five years while reaching a plateau at two years postoperatively in the Magnum group. At five years, the mean cobalt ion level of 1.16 μg/l (. sd. 1.32) in the Magnum group was significantly lower than the 3.77 μg/l (. sd. 9.80) seen in the conventional group (p = 0.0015). The incidence of moderate to severe pseudotumour was 4.7% in the Magnum group and 20.6% in the conventional group. There were no dislocations in the Magnum group and two in the conventional group. One patient in the Magnum group underwent revision for pseudotumour at 4.7 years postoperatively. Conclusion. At five years, a well-positioned large head MoM THA has a significantly lower level of metal ion release and a lower incidence of moderate to severe pseudotumour than a MoM bearing of conventional size. Cite this article: Bone Joint J 2018;100-B:1018–24


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 768 - 773
1 Jun 2012
Wang Q Zhang XL Chen YS Shen H Shao JJ

In this prospective study a total of 80 consecutive Chinese patients with Crowe type I or II developmental dysplasia of the hip were randomly assigned for hip resurfacing arthroplasty (HRA) or total hip replacement (THR).

Three patients assigned to HRA were converted to THR, and three HRA patients and two THR patients were lost to follow-up. This left a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips) who underwent THR. The mean follow-up was 59.4 months (52 to 70) in the HRA group and 60.6 months (50 to 72) in the THR group. There was no failure of the prosthesis in either group. Flexion of the hip was significantly better after HRA, but there was no difference in the mean post-operative Harris hip scores between the groups. The mean size of the acetabular component in the HRA group was significantly larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was no difference in the mean abduction angle of the acetabular component between the two groups.

Although the patients in this series had risk factors for failure after HRA, such as low body weight, small femoral heads and dysplasia, the clinical results of resurfacing in those with Crowe type I or II hip dysplasia were satisfactory. Patients in the HRA group had a better range of movement, although neck-cup impingement was observed. However, more acetabular bone was sacrificed in HRA patients, and it is unclear whether this will have an adverse effect in the long term.


Cemented total hip arthroplasty yields reliable results in short to medium term studies, but aseptic loosening remains a problem in long-term follow up, especially in young and active patients. Aseptic loosening has been related to wear, and in order to minimize wear various alternatives to the traditional metal on polyethylene have been proposed. Both ceramic on polyethylene (COP) and metal on metal (MOM) have been shown to produce less wear than metal on polyethylene (MOP). In order to study the effect of the bearing, we have utilized identical stems and cups while comparing the different bearings. Methods and material. 396 hips were randomized to MOP, COP or MOM using a cemented triple tapered polished stem (MS-30; Sulzer Orthopedics) with a cemented polyethylene cup (Weber; Sulzer, Orthopedics) and a 28 millimeter head. For the MOP and COP articulations, a all-polyethylene cup was used with a Protasul™ metal head or a Sulox™ alumina head, whereas a polyethylene cup with metal insert was used for the MOM articulations (Weber Polyethylene Cup with Metasul™ Insert) with a Metasul™ metal head. Harris Hip Score (HHS) and radiological evaluation was performed after two, five and seven years. Ethical approval was obtained. Results. HHS was available for 338 hips after seven years. The HHS in the MOP group (116 hips) was 93.7 (SD 9.0), 93.5 in the COP group (112 hips) (SD 8.8), and 91.0 (SD 13.4) in the MOM group (110 hips). Radiographic evaluation was available for 335 hips. Radiolucencies around the stem larger than one millimeter were found in five of 115 MOP hips, seven of 111 COP hips and in seven of 110 MOM hips. Periacetabular radiolucencies identified as larger than one millimeter were found in none of the 116 MOP hips, five of the 112 COP hips and in 19 of the 110 MOM hips. Ten revisions were performed. In the MOP group there were three revisions (infection, dislocation, pain); one in the COM group (infection), and six in the MOM group (three infections, two aseptic loosening and one septic loosening). Discussion. Clinically, the three bearing types perform equally well at seven years, and there are few signs of impending failure for the stem. However, there are an alarmingly high proportion of cups showing signs of loosening. Two hips have been revised in the MOM group for aseptic loosening as opposed to none in the two other groups, which adds to the concern. The MOM articulation used in this study should be monitored closely for signs of aseptic loosening


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 21 - 21
1 May 2016
Rodgers W Buchele C Kaddick C Schroeder D
Full Access

Introduction. Clinically relevant attributes of an orthopedic bearing material include its strength, oxidative stability, and wear resistance. Recent reductions in bearing wear and oxidation have been realized by crosslinking (HXLPE), and through the incorporation of α-tocopherol (VE). VE infusion has improved the oxidative stability of HXLPE in vitro [1] and in vivo (as evidenced by retrievals) [2]. However, concern has been raised that adding VE may increase the frictional torque of bearings, potentially increasing the risk of trunnionosis [3]. This study compares the 3D frictional torques of HXLPE vs Vitamin-E HXPLE, against 28 mm and 36 mm metal and ceramic heads. Methods. Samples were made from prints for commercially available Ringloc liners (Biomet, IN). The HXLPE group was made from ArCom XL, and the VE-HXLPE was made from E1 HXLPE (Biomet, IN). Femoral heads were of cobalt chrome (ASTM F1537), or ceramic (Biolox Delta™). Testing was performed at EndoLab®, Germany. An ISO 14242–1 compliant six-station simulator and 3D gait cycle was utilized. During the cycle, the abduction/adduction range was −4°/+7°, the flexion/extension was +25°/-18°, and the external/internal rotation was −10°/+2°. Testing was performed at 37±2 °C, at 1 Hz, and with a maximum dynamic load of 3.0 kN. Lubrication medium was calf serum, EDTA, and antibiotics diluted in DI water (30 g/l of protein). Measurements were averaged across 5 cycles after 120 completed cycles of motion, and after 200 cycles. Analysis was performed using Minitab with multiple 2-way ANOVAs, with a p=0.05 significance threshold. Results. There was a statistically significant effect from head size (p≤0.0005). 36 mm heads exhibited higher torque (4.25±0.31 Nm, 200 cycles) than the 28 mm heads (2.90±0.08 Nm). There was no statistically significant effect (p≥0.409, all outputs) from head type (metal vs ceramic). In the 36 mm groups, there was no statistically significant effect (p≥0.300, all outputs) from the liner material (HXLPE vs VE-HXLPE). Within the metal head groups, the inclusion of VE statistically significantly decreased (p≤0.018, all outputs) the average mean frictional torques by 8.07±4.6%. Conclusion. A larger head size increased the frictional torque. In the metal head groups the infusion of VE decreased the frictional torque, and VE did not statistically significantly change the torque in the ceramic groups. A previous report found that VE increased torque [3], but evidence herein does not support that conclusion. Limitations of the previous study include; use of a simplified rotation torque method, non-physiologic lubrication, lack of statistical analysis, unknown implant design, and unknown VE concentration and processing methods. Based on these results, no adverse clinical effects are anticipated due to VE-associated changes in frictional torque. Short-to-mid-term clinical results have not shown any adverse effects [4]


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 85 - 85
4 Apr 2023
Wulfhorst M Büssemaker H Meinshausen A Herbster M Döring J Mai V Lohmann C Kautz A Laube T Wyrwa R Schnabelrauch M Bertrand J
Full Access

The implantation of endoprosthesis is a routine procedure in orthopaedics. Endoprosthesis are mainly manufactured from ceramics, polymers, metals or metal alloys. To ensure longevity of the implants they should be as biocompatible as possible and ideally have antibacterial properties, to avoid periprosthetic joint infections (PJI). Various antibacterial implant materials have been proposed, but have so far only been used sporadically in patients. PJI is one of the main risk factors for revision surgeries. The aim of the study was to identify novel implant coatings that both exhibit antibacterial properties whilst having optimal biocompatibility. Six different novel implant coatings and surface modifications (EBM TiAl6V4, strontium, TiCuN, TiNbN, gentamicin phosphate (GP), gentamicin phosphate+cationic polymer (GP+CP)) were compared to standard CoCrMo-alloy. The coatings were further characterized with regard to the surface roughness. E. coli and S. capitis were cultured on the modified surfaces to investigate the antibacterial properties. To quantify bacterial proliferation the optical density (OD) was measured and viability was determined using colony forming units (CFU). Murine bone marrow derived macrophages (BMMs) were cultured on the surfaces and differentiated into osteoblasts to quantify the mineralisation using the alizarin red assay. All novel coatings showed reduced bacterial proliferation and viability compared to standard CoCrMo-alloy. A significant reduction was observed for GP and GP+CP coated samples compared to CoCrMo (OD. GP,E.coli. = 0.18±0.4; OD. GP+CP,E.coli. = 0.13±0.3; p≤0.0002; N≥7-8). An increase in osteoblast-mediated mineralisation was observed on all surfaces tested compared to CoCrMo. Furthermore, GP and GP+CP coated samples showed a statistically significant increase (M. GP. = 0.21±0.1; M. GP+CP. = 0.25±0.2; p<0.0001; N≥3-6). The preliminary data indicates that the gentamicin containing surfaces have the most effective antibacterial property and the highest osseointegrative capacity. The use of antibiotic coatings on prostheses could reduce the risk of PJI while being applied on osseointegrative implant surfaces


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 124 - 124
1 Mar 2021
Jelsma J Schotanus M Kleinveld H Grimm B Heyligers I
Full Access

An increase in metal ion levels is seen after implantation of all MoM hip prosthesis due to release from the surface directly, more so during articulation and corrosion of the bearing surfaces. The bearing surfaces in MoM prosthesis consist of cobalt, chromium and molybdenum. Several case-reports of cobalt toxicity due to a MoM prosthesis have been published in the last decade. Cobalt intoxication may lead to a variety of symptoms: neuro-ocular toxicity (tinnitus, vertigo, deafness, blindness, convulsions, headaches and peripheral neuropathy), cardiotoxicity and thyroid toxicity. Nausea, anorexia and unexplained weight loss have been described. Systemic effects from metal ions even with well functioning implants or with ion concentrations lower than those associated with known adverse effects may exist and warrant investigation. The aim of this study is to investigate self-reported systemic complaints in association with cobalt ion concentrations in patients with any type of MoM hip prosthesis. A cohort study was conducted. Patients with both unilateral and bilateral, resurfacing and large head metal on metal total hip arthroplasties were included for the current study. Blood metal ion concentrations (cobalt and chromium) were measured by inductively coupled plasma mass spectrometry (ICP-MS). Based on the known cobalt toxicity symptoms of case-reports and toxicology reports a new non-validated questionnaire was developed. questions were subdivided in general questions/symptoms, vestibular symptoms, neurological symptoms, emotional health and cardio- and thyroid toxicity symptoms. Independent samples T test, Fishers Exact Test and Pearsons (R) correlation were used. Analysis was performed on two groups; a low cobalt ion concentration group and a high cobalt ion concentration group A total of 62 patients, 36 (58%) men and 26 (42%) women, were included with a mean age at surgery of 60.8 ± 9.3 years (41.6 – 78.1) and a mean follow up of 6.3 ± 1.4years (3.7 – 9.6). In these patients a total of 71 prosthesis were implanted: 53 unilateral and 9 bilateral. Of these, 44 were resurfacing and 27 large head metal on metal (LHMoM) total hip arthroplasties. Mean cobalt and chromium ion concentrations were 104 ± 141 nmol/L (9 – 833) and 95 ± 130nmol/L (6 – 592), respectively. Based on the different thresholds (120 – 170 or 220 nmol/L) the low cobalt ion concentration group consisted of 44 (71%), 51 (82%) or 55 (89%) subjects respectively. No differences were found in general characteristics, independently of the threshold. The composite score of vestibular symptoms (vision, hearing, tinnitus, dizziness) was significantly higher (p < .050) in all high cobalt ion concentrations groups, independent of the threshold value This study aimed to detect a trend in self-reported systemic complaints in patients with metal-on-metal hip arthroplasty due to raised cobalt ion concentrations. Vestibular symptoms were more common in high cobalt ion concentration groups independent of the three threshold levels tested. The upper limit of acceptable cobalt ion concentrations remains uncertain. With regards to proactively inquired, self-reported symptoms the threshold where effects may be present could be lower than values currently applied in clinical follow-up. It is unknown what exposure to elevated metal ion concentrations for a longer period of time causes with aging subjects. Further research with a larger cohort and a more standardized questionnaire is necessary to detect previously undiscovered or under-reported effects


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_17 | Pages 28 - 28
24 Nov 2023
De Vecchi E Balzano V Bottagisio M Gavioli L
Full Access

Aim. Antibacterial activity of coatings based on metal and metal oxide nanoparticles (NPs) often depends on materials and biotic targets resulting in a material-specific killing activity of selected Gram-positive and Gram-negative bacteria, including drug-resistant strains. In this perspective, the NPs loading amount, the relative elemental concentration inside the nanogranular building blocks and the deposition method are of paramount importance when the goal is to widen the antimicrobial spectrum, but at the same time to avoid high levels of metal content to limit undesired toxic effects. Aim of the present study was evaluation of the antimicrobial properties of two multielement nanogranular coatings composed of Titanium-Silver and Copper and of Magnesium-Silver and Copper. Method. Ti-Ag-Cu and Mg-Ag-Cu NPs were deposited on circular cover glasses (VWR) by Supersonic Cluster Beam Deposition. Biofilm-producer strains of Staphylococcus aureus (methicillin susceptible and resistant), Staphylococcus epidermidis (methicillin susceptible and resistant), Escherichia coli (fully susceptible and producer of extended spectrum beta lactamases), and Pseudomonas aeruginosa (susceptible and multidrug-resistant) were selected. The abilities of the selected strains to adhere, colonize and produce biofilm on the discs coated with Ti-Ag-Cu or Mg-Ag-Cu NPs were compared to uncoated circular cover glasses which were used as growth control. Cytotoxicity was also evaluated in order to assess the biocompatibility of the newly synthesized NPs. Results. In comparison to uncoated controls, both coatings showed significant anti-adhesive properties against S. aureus, S. epidermidis, and E. coli. Reduction in adhesion to Mg-Ag-Cu coated discs was observed also for P. aeruginosa isolates, although differences vs uncoated controls did not reach statistical significance. Biofilm formation was reduced on discs coated with Mg-Ag-Cu compared to Ti-Ag-Cu and, again, coatings had a milder effect on P. aeruginosa, probably due to its exceptional capability of attachment and matrix production. These results were confirmed by the evaluation of bacterial colonization on nanoparticles-coated discs by means of confocal laser scanning microscopy. A viability of 95.8% and 89.4% of cells cultured in the presence of Ti-Ag-Cu and Mg-Ag-Cu discs, respectively, when compared to negative controls was observed, thus excluding cytotoxic effects on eukaryotic cells. Conclusions. The newly synthesized Ti-Ag-Cu and Mg-Ag-Cu coatings are able to limit bacterial adhesion colonization and biofilm production, thus highlighting the safe use of multi-element families of NPs as new strategies against bacterial attachment to the surface of biomedical implants. However, further studies addressing activity against P. aeruginosa and including a wide number of isolates are warranted


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 31 - 31
1 Apr 2022
Langton D Bhalekar R Joyce T Shyam N Nargol M Pabbruwe M Su E Nargol A
Full Access

Cobalt chrome alloy is commonly used in joint replacement surgery. However, it is recognised that some patients develop lymphocyte mediated delayed type hypersensitivity (DTH) responses to this material, which may result in extensive bone and soft tissue destruction. Phase 1. United Kingdom: From an existing database, we identified extreme phenotype patient groups following metal on metal (MoM) hip resurfacing or THR: ALVAL with low wearing prostheses; ALVAL with high wear; no ALVAL with high wear; and asymptomatic patients with implants in situ for longer than ten years. Class I and II HLA genotype frequency distributions were compared between these patients’ groups, and in silico peptide binding studies were carried out using validated methodology. Phase 2. United Kingdom: We expanded the study to include more patients, including those with intermediary phenotypes to test whether an algorithm could be developed incorporating “risk genotypes”, patient age, sex and metal exposure. This model was trained in phase 3. Phase 3. United Kingdom, Australia, United States. Patients from other centres were invited to give DNA samples. The data set was split in two. 70% was used to develop machine learning models to predict failure secondary to DTH. The predictions were tested using the remaining blinded 30% of data, using time-dependent AUROCs, and integrated calibration index performance statistics. A total of 606 DNA samples, from 397 males and 209 female patients, were typed. This included 176 from patients with failed prostheses, and 430 from asymptomatic patients at a mean of >10 years follow up. C-index and ROC(t) scores suggested a high degree of discrimination, whilst the IBS indicated good calibration and further backed up the indication of high discriminatory ability. At ten years, the weighted mean survival probability error was < 4%. At present, there are no tests in widespread clinical use which use a patient's genetic profile to guide implant selection or inform post-operative management. The algorithm described herein may address this issue and we suggest that the application may not be restricted to the field of MoM hip arthroplasty


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_16 | Pages 83 - 83
1 Nov 2018
Paulus AC Ebinger K Haßelt S Kretzer JP Bader R Utzschneider S
Full Access

The biological reaction in metallosis and pseudotumor generation after metal on metal total hip arthroplasty or corroding metal implants remains unsettled. Clinically, still lethal cases appear with massive bone loss and metal ions are suspected to be responsible for this inflammatory reaction, solid metal wear particles instead are usually not observed in the common literature. The aim of this study was to compare the biological reactions of metal ions and metal wear particles in a murine in vivo model. Metal ions (CoCr), metal particles (CoCr), polyethylene particles (UHMWPE) and phosphate buffered saline (PBS) were injected into the left knee joint of female BALB/c mice. 7 days after injection, the microcirculation was observed using intravital fluorescence microscopy, followed by euthanasia of the animals. After the assessment of the knee diameter, the knees underwent histological evaluations of the synovial layer. Throughout all recorded data, CoCr particles caused higher inflammatory reactions compared to metal ions and UHMWPE particles. The mice treated with the solid particles showed enlarged knee diameters, more intensive leukocyte–endothelial cell interactions and an elevated functional capillary density. Pseudotumor-like tissue formations in the synovial layer of the mice were only seen after the exposition to solid CoCr particles. Even if the focus of several national guidelines concerning metallosis and pseudotumor generation is on metal ions, the present data reveal that solid CoCr particles have the strongest inflammatory activity compared with metal ions and UHMWPE particles in vivo


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 149 - 149
1 Jul 2014
Slagis S Skrepnik N Wild J Robertson M Nielsen B Skrepnik T Eberle R
Full Access

Summary. Management of metal on metal hip replacements can be accomplished with a simple algorithm including easily available metal ion levels and hip MRI with metal artifact reducing software. After revision serum metal ion levels can be expected to fall rapidly. Introduction. Metallic ion release may be related to bearing surface wear and thus serves as an indicator of the in-vivo performance of metal on metal articulations. The purpose of this prospective, controlled study was to compare new large head metal on metal hip components with established modular metal on metal and metal on polyethylene and to determine their effects on serum metal levels before and after revision. Patients & Methods. We performed a multi-surgeon, prospective, controlled trial to compare clinical, radiographic, and metal ion concentration in serum (cobalt and chrome) results across multiple devices including the Large Head ASR XL System (MoM-1), the Ultamet Advanced Modularity System (MoM-2), and as the control the Pinacle Acetabular Cup System with polyethylene liner (MoP). One hundred and fifty-one consecutive patients undergoing THA were enrolled in the study: MoM-1 n=97; MoM-2 n=22; MoP n=32. Clinical, radiographic, and venous blood assessments were performed pre-operatively, and post-operatively at 6 months, 1 year and 2 years, and after revision (1,3,6,12 months). All serum ion concentrations are reported in nmol/L. We are following metal ion levels after revision and have developed an algorithm to diagnose and manage patients with MoM THA. Results. MoM-1 patients had significantly increased average cobalt and chromium levels. Clinical scores improved after surgery in all groups and continued to improve in MoM-2 and MoP patients after 2 years but decreased slightly in the MoM-1 patients at 2 years. Average cup inclination angle did not differ significantly between the groups: MoM-1 50.2, MoM-2 47.8, and MoP 51.7. In the MoM-1 group 11 patients (11%) had significantly elevated ion levels (MoM-1 Outliers). Nine hips (9.3%) in 8 MoM-1 outlier patients required revision. Metal ion levels were not significantly different between MoM-2 and MoP groups. Metal ion levels after revision in the MoM-1 group decreased rapidly but at one year post-operatively have still not returned to an equivalent baseline comparable to the MoM-2 and MoP groups. All revisions were in the MoM-1 group. Chromium levels decreased more slowly than Cobalt levels. Discussion. To our knowledge this is the only data in the literature prospectively comparing ion levels among groups and reporting post revision ion levels. Average serum ion levels were elevated at all post-operative samples in the MoM-1 group but this was due to significantly elevated levels in a subset of outliers who required revision. Excluding the outliers there is not a significant difference in post-operative ion levels between the groups. There was no radiographic evidence of component malposition or aseptic loosening in any of the groups. Control groups (MoM-2, MoP) performed comparatively across all variables. We present an algorithm to diagnose and manage patients with metal on metal THA and offer evidence that metal ion levels do decrease after revision but still remain abnormally elevated at one-year post revision compared to the control group. A significant portion of MoM-1 performs comparatively to the controls in terms of ions


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 239 - 239
1 May 2009
MacDonald S Bourne RB Chess D McCalden RW Rorabeck CH Thompson A
Full Access

We performed a prospective, randomised, blinded clinical trial comparing metal versus polyethylene bearing surfaces in patients receiving a THA. Forty-one patients were randomised to receive a metal (twenty-three) or a polyethylene (eighteen) insert with identical femoral and acetabular components. The metal bearing was a 28mm low carbon on high carbon couple. Patients were evaluated pre-operatively, at three, six, twelve months and annually thereafter, including an evaluation of erythrocyte and urine cobalt, chromium, and titanium, outcome measures (WOMAC, SF-12, Harris Hip Score) and radiographs. No patients were lost to follow-up. At an average 7.2 (range 6.1 – 7.8) years follow-up there were no differences in any outcome measures or radiographic findings. Patients receiving metal liners had significantly elevated metal ion measurements. At most recent follow-up, compared to the polyethylene control group, patients receiving a metal on metal bearing had erythrocyte cobalt levels were eleven times elevated (median 1.4 μg/L (metal) vs 0.12 μg/L (poly), p< .001). Urine cobalt levels were thirty-nine times elevated (median 11.4 μg/L/day (metal) vs 0.29 μg/day (poly), p< .001) and urine chromium levels were twenty-eight times elevated (median 4.75 μg/day (metal) vs 0.17 μg/day (poly), p< .001). Additionally the metal ion levels reached an early steady state level and did not decrease over time. (Erythrocyte Co (μg/L): 1.29 (6 mos), 1.20 (1 yr), 1.0 (2 yr), 1.10 (3 yr), 1.35 (4 yr), 1.40 (5 yr)). Both cobalt and chromium ion measurements were significantly elevated in the blood and urine of the patients randomised to receive the metal on metal THA at all time intervals. Contrary to previous reports, in 34% of patients with a metal liner, metal ion (erythrocyte cobalt) elevation was still increasing at latest follow-up, and in the overall metal on metal patient cohort, metal ion median levels were not decreasing over time. As in polyethylene wear data, metal ion results are not necessarily generalizable, but are bearing design specific


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 135 - 135
1 Apr 2019
Lage L
Full Access

Orthopaedic implants, such as femoral heads, sockets and stems, are manufactured with a high degree of smoothness and very low form error in order to function as low wear bearings. The surfaces are subject to both wear and damage during in vivo use. Articulating surfaces naturally wear during normal use. Aseptic loosening associated with osteolysis and release of wear particles is the main reason for revision of total hip arthroplasty (THA). Damage of femoral heads is well known to increase the wear rate at the articulating surface and is vulnerable to scratching during the maneuver of positioning the femoral component into the acetabulum component either in primary as in revision total hip arthroplasties. The findings emphasize the importance of achieving and maintaining good surface finish of the femoral head component. The author presents a very simple and “zero cost” method of preventing scratching of the femoral head of any kind of total hip prosthesis (ceramic on ceramic, ceramic on poly, metal on metal, metal on poly and even metal on ceramic) when the reduction of the femoral head prosthesis is done inside the new acetabular component with metal, ceramic liner or poly liner with metal back (where the scratching can also occur) as one of the final stages of the surgical procedure which can be crucial to the long survival of the hip prosthesis. A short one minute video on an e-poster will show how this can be done being an easy, reproducible, safe and reliable technique to prevent femoral head scratching


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 183 - 183
1 May 2011
Nevelos J Bhimji S Dong N Macintyre J Coustance A Streicher R
Full Access

It is accepted that larger diameter heads are more difficult to dislocate due to the increased distance the head has to travel to come out of the cup. Currently larger femoral heads are being used for their resistance to dislocation however, there remains little reporting on the effect of design of cup on jump distance. Monoblock metal on metal cups, which were designed for hip resurfacing are typically less than a hemisphere internally in order to increase the range of motion (ROM) needed when the femoral neck is retained. This does however also reduce the jump distance. We investigated several designs of cup with a variety of head sizes in order to compare ROM using a computer range of motion tool and a two dimensional jump distance with the cup at 45 degrees inclination. Jump distances were calculated for: Internally hemispheric cups in 28, 32 and 36mm bearing diameters; 28, 40 and 44mm polyethylene liners which were hemispheric but with an additional 2mm cylinder and a 0.7mm chamfer at the equator (Trident, Stryker, Mahwah, USA); 38, 48 and 54mm monoblock metal on metal resurfacing cups with a 3.5mm offset (BHR, Smith and Nephew, Memphis, USA); 40, 48, 58 dual mobility cups with an anatomic rim (Restoration ADM, Stryker, Mahwah, USA). Range of motion modeling was carried out using custom-written software according to a previously published method2 with 5 degrees of pelvic tilt and a standard femoral component. For the present study, range of motion was assessed on a standard stem with a 132° neck angle. Inclination of the cup was set to 45° and anteversion to 20°. For each implant tested, the total ROM was computed in flexion/extension, ab/adduction, and int/external rotation. Components tested for range of motion were: Trident 32, 36, 40 and 44mm Internal Diameter; Hemispheric 28 and 32mm Internal Diameter cups; MITCH TRH MoM Monoblock Resurfacing Cup (Stryker EMEA, Montreux, Switzerland) 46mm cup bearing diameter with a 2.75mm offset bore; Dual Mobility 40, 46 and 58mm cups. The metal on metal monoblock cups had a very high range of motion but a 48mm head has only a similar jump distance to a hemispheric 36mm design. The designs with the cylinder and chamfer have a markedly higher jump distance than their hemispheric equivalents but slightly reduced ROM. Interestingly, the dual mobility design has almost double the jump distance of an equivalently sized metal on metal resurfacing type cup and a higher jump distance than an equivalent head size in a conventional unipolar design. The dual mobility design has similar ROM to a 40mm head in the hemisphere with cylinder and chamfer design. ROM is slightly higher in the hemispheric and sub-hemispheric designs but this model does not take into account bony or soft tissue impingement. The role of design of ace-tabular component has a great effect on the range of motion and jump distance of bearings


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXI | Pages 10 - 10
1 May 2012
R. DS L. M P. R S. G
Full Access

Introduction. Bearing surfaces used for total hip arthroplasty must have characteristics including bio-compatibility, low friction and low wear rate. Bearing combinations are generally characterised as Soft on Hard/Hard. In general, all newer bearing combinations have reduced wear but may present with other issues that impact on patient outcomes. Materials. The Australian Orthopaedic Association – National Joint Replacement Registry classifies bearing surfaces into six categories. These are metal on polyethylene, ceramic on polyethylene, metal on metal, ceramic on ceramic, ceramic on metal and a sixth category relating to a small number of procedures where the bearing surface is yet to be classified. 147,422 conventional total hip arthroplasty procedures have been recorded by the Registry between 1 September 1999 and 31 December 2008 and analysis has been performed of the cumulative percentage revision in relation to bearing surface. Results. In general, metal on polyethylene has the lowest risk of revision compared to all other bearing surfaces and metal on metal has the highest revision rate. The risk of revision, however, does vary depending on head size, with larger head sizes having a lower risk of revision with the exception of metal on metal. Conclusion. When deciding which bearing surface is suitable for patients it must be emphasised that wear reduction is only one of several considerations when choosing the most appropriate combination


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 545 - 545
1 Nov 2011
Jones HW Wimhurst J Macnair R Derbishire B Chirodian N Toms A Cahir J
Full Access

Introduction: Although good mid-term results have been reported with some metal on metal hip replacements, reported complications due to metal on metal (MOM) related reactions are a cause for concern. We have assessed the clinical outcome and MRI metallic artefact reduction sequence (MARS) findings in a consecutive series of patients with a large head metal on metal hip replacement. Methods: 62 ASR XL Corail THRs and 17 ASR resurfacings were performed at our hospital between 2005 and 2008. All patients were reviewed and assessed with an Oxford hip score (OHS), a plain radiograph and a MRI imaging was obtained on 76 (96%) hips. Implant position was assessed using Wrightington cup orientation software. Results: At a mean follow up of 32 months, 9 (15%) ASR XL Corail THRs, and 2(12%) ASR resurfacings had been revised. 10 revisions were performed for MRI confirmed MOM related pathology. Histology confirmed a MOM reaction in all 10 cases. Of the 76 hips that were MRI scanned, 27 (36%) had typical features of a MOM reaction. These were classified as mild in 10 (13%), moderate in 13 (17%) and severe in 4 (5%). 78 patients completed an OHS and the mean score was 21. The mean OHS was 29 pre-operatively in those that had been revised, 25 in patients with abnormal MRI findings and 20 in those with a normal MRI. 10 patients with abnormal MRIs had a near perfect OHS (15 or less). Conclusions: The ASR XL Corail THR has an unacceptably high early failure rate. MARS MRI is able to detect metal debris related soft tissue pathology around metal on metal THRs. These lesions are sometimes asymptomatic. We suggest that MARS MRI evaluation should form part of the routine evaluation of all metal on metal THRs, and in particular of this implant


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 26 - 26
1 Oct 2020
Hamilton WG Robertson RN Cororaton AD Ho H Hopper RH
Full Access

Introduction. The rise of the anterior approach (AA) in primary surgery has led to enthusiasm for using this approach in revision surgery, especially because head and liner exchanges have been accompanied by a high dislocation rate when the posterior approach (PA) is used. The aim of this study was to compare the institutional dislocation rate comparing the PA and AA in isolated head and liner exchange. Methods. A retrospective institutional database query was done to identify all aseptic head and liner exchanges between the years 2010- June, 2020. 186 hips were identified with an average age of 64.8+/−10.8 yrs (27.9–87.6) and average BMI of 28.7+/−6.3 kg/m2 (16.2–52). The reason for revision was polyethylene wear and osteolysis in 105 hips (56.5%), adverse tissue reaction to metal on metal (MOM) in 43 hips (23.1%), and recurrent instability in 38 hips (20.4%). The approach used for the revision surgery was PA in 128 hips (68.8%) and AA in 58 hips (31.2%). The mean follow-up for this cohort is 1.9±2 yrs (0.0–8.6). Results. The overall dislocation rate for the entire cohort was 28/186 (15%). There was no difference in the dislocation rate whether the revision surgery was performed by the PA 19/128 (15%) or the AA 9/58 (16%) (p=0.9). With the numbers available, there was no difference in dislocation rates when comparing reason for revision: poly wear/osteolysis 12/105 (11%), MOM 8/43 (19%) or recurrent instability 8/38 (21%) (p=0.28). The head size used in the exchange did not influence the dislocation rate; 28 mm-2/11 (18%), 32 mm-6/56 (11%), 36 mm-16/91 (18%), 40 mm-4/26 (15%) (p=0.89). Conclusion. The use of either the PA or AA did not alter the dislocation rate in our cohort. While there was a trend towards a higher dislocation rate when liner exchange was performed for MOM or recurrent instability, no statistical difference was shown with the numbers available. The head size used in the revision did not influence the rate of dislocation


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 88 - 88
1 Mar 2010
Fisher J
Full Access

Young and active patients require bearing materials that can last up to 200 million walking steps, ten fold greater than conventional polyethylene bearings. Cross linked polyethylene provides reduced wear rate compared to conventional polyethylene, and further advantage is gained from using ceramic femoral heads. However in polyethylene bearings wear increases with the head diameter, and there is currently little opportunity to use head sizes greater then 36mm diameter. There is evidence of polyethylene fracture with steeply positioned cups. Ceramic on ceramic bearings provide substantially lower wear rates than polyethylene bearings. Steep cups, lateralised heads or neck impingement can lead to head contact on superior rim of the cup and stripe wear, but this still results in very low wear rates. Recently developed ceramic matrix composites Biolox Delta provide greater resistance to stripe wear. In a few patients stripe wear may lead to squeaking. Metal on metal bearings also provide substantially lower wear than polyethylene bearings. However there remains concern about elevated metal ion levels in a few patients and resultant risk of hypersensitivity reactions. In metal on metal bearings larger head sizes and reduced diametrical clearance can lead to reduced wear. Increased wear is associated with steep cups and lateralised heads resulting in rim wear. Ceramic on metal bearings have been introduced recently as the first differential hard on hard bearings. These bearings show substantial reduction in wear, corrosive wear mechanisms, metal ion levels in laboratory simulators and initial clinical studies have shown a reduction in metal ion levels in vivo compared to metal on metal bearings


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 145 - 145
1 Mar 2010
Yoo M Cho Y Kim K Chun Y Rhyu K Roh J Kang C
Full Access

The purpose of this study is to evaluate short to mid-term clinical and radiological results of metal on metal resurfacing arthroplasty in osteonecrosis of the femoral head (ONFH). 185 hips of 169 patients who underwent metal on metal resurfacing arthroplasty using Birmingham Hip Resurfacing system (Midland Medical Technololgies, Birmingham, UK) between December 1998 and May 2005 were available for this study and all cases were followed up over 3 years. All preoperative diagnoses were ONFH. The extents of necrotic area were analyzed by preoperative MRI scanning. Their mean age at the time of operation was 37.7(range, 16–67) years old and mean period of follow-up was 88(range, 36–113) months. For the clinical assessments, Harris hip scores, UCLA activity scores, pain and ROM were evaluated. Radiological changes such as radiolucencies around the stem, impingement sign, neck narrowing, osteolysis around head and neck junction, loosening of implants, heterotopic ossifications were evaluated in the serial antero-posterior, translateral radiographs of the hip joint. Preoperative necrotic area was average 42.7(range, 11.5–60) %. Clinically, the average Harris hip score was improved from 85.2 points to 97.1 points at final follow-up. Average UCLA activity scores at the last follow-up was 8.8 and almost of the patients showed high activity and returned to their original job. ROM were very satisfactory. Radiologically, the mean inclination of acetabular component was 48.0°. There were no radiolucent lines around the acetabular components, but 3 cases showed radiolucent lines around the stem of femoral components. Osteolytic lesions were noticed in 10 cases around head-neck junction. Nine hips had impingement signs around the head-neck junction. There was no case which showed evidence of stress shielding. Moderate neck narrowing were shown in 3 cases. There were 6 cases of heterotopic ossification. One hip had a revision surgery to a total hip arthroplasty using big metal ball because of loosening of acetabular component. There was no patient complained limb length discrepancy and no infection, dislocation, thigh pain. The midterm performance of metal on metal resurfacing arthroplasty in ONFH was very excellent in the aspects of pain relief, ROM of hip joints, rehabilitation and return to preoperative activity and minimization of common complications of conventional total hip arthroplasty. There was no mechanical failure related to the osteonecrosis and we can conclude that performing resurfacing arthorplasty in osteonecrosis less than 50% of extent can be justified. However, performing resurfacing arthroplasties in osteonecrosis of femoral heads needs meticulous surgical techniques and longer learning curve to prevent early failure. Even though our midterm follow-up study revealed excellent results, more long-term follow-up studies are mandatory to determine the survivorship and to verify the problems related to the increased serum metal ion and metal ion toxicity after resurfacing arthroplasty


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_III | Pages 259 - 259
1 Nov 2002
Kuster M
Full Access

The main problem of modern total hip replacement is the reduction of wear debris. Hence, new tribological partners such as ceramic on ceramic, metal on highly crosslinked polyethylene and metal on metal have evolved. Of these new combinations metal on metal has the longest history. The early problems of high friction using a “micro-fit” between acetabulum and femoral head have been solved by introducing an optimal clearance between the head and the cup to allow for small deformations of the acetabulum during activities without locking. The annual wear rate of metal on metal combinations has been shown to be extremely low ranging from 2 to 5 micrometers/year only. A further advantage of Metasul may be the “wearing in of small scratches” as well as forgiving slight malpositions of the acetabulum, which is not the case in ceramic – ceramic combinations. However, Metasul should not be implanted in patients with renal failure or severe allergies. Metal-metail pairing has proven a valuable alternative in young and active patients over the last 10 years


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 224 - 224
1 Sep 2012
Ebreo D Khan A El-Meligy M Armstrong C Peter V
Full Access

INTRODUCTION. The advantages of large diameter metal on metal total hip arthroplasty (MoM THA) and hip resurfacing arthroplasty are decreased wear rate, preservation of bone stock, anatomical restoration and enhanced stability. Large amounts of metal wear particles and metal ions are released which may induce adverse reactions including local soft tissue toxicity, hypersensitivity reactions, bone loss and risk of carcinogenesis. Aseptic loosening can be the result of a peri-prosthetic osteolysis generated as a result of a biological response to particulate wear debris. No reports in the literature exist as to whether circulating levels of Chromium (Cr) and Cobalt (Co) decrease upon removal of a symptomatic large diameter MoM implant or whether levels remain high due to the effect of metal ions debris left behind in the soft tissues after revision surgery. PATIENTS AND METHODS. Between June 2006 and June 2009 we undertook 44 revision surgeries of both large head MoM THAs (femoral head diameter 38mm) and metal-on-metal hip resurfacings for suspected metallosis. Mean time from original implant to revision was 4 years, 8 months (1yr 4mo–7yr 9mo). The mean follow up evaluation was 2 years and 2 months (1yr 2mo–4 years). Blood samples were taken for whole blood Cr and serum Co according to a recognised protocol and compared with reference levels indicated by the Medicines and Healthcare Regulatory Agency recommendation of less than 7ppb for Cr (130nM/L) and Co (119nM/L). RESULTS. 42 patients were found to have histological evidence of either metal allergy, metal toxicity or foreign body reaction. 2 patients had evidence of infection with no features of metal reaction. 3 patients suffered early dislocation requiring closed reduction. 1 patient had infective complications necessitating Girdlestones. 11 patients were lost to follow up, 8 patients were diagnosed pre operatively on Co and Cr levels in urine or synovial fluid aspirate alone. 23 patients had pre revision blood or serum metal ion level results available for direct comparison. Median serum Co level pre revision was 176.6nM/L, falling post revision to 5.1nM/L (p=<0.001∗). The median whole blood Cr level pre revision was 117nM/L and 19nm/L post revision (p=<0.001∗). Mean Oxford Hip Score was 23.7. DISCUSSION. This study demonstrates that at greater than one year post removal of a large diameter MoM hip implant for the indication of symptomatic metallosis or metal hypersensitivity, metal ion levels fall to almost normal levels and that outcome of revision surgery in terms of patient satisfaction is not adversely affected


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 351 - 352
1 Mar 2004
Dorn U Neumann D Berka J
Full Access

Aims: A polyethylene free, metal on metal acetabular system (Hofer-Imhof cup; Lubrimet¨ metal on metal articulation made of CoCr-forge alloy) was designed in an effort to improve total hip arthroplasty longevity. This study was undertaken to review the clinical performance of this implant and to determine if early ace-tabular loosening or revision and wear and osteolysis were prevalent. The mid term results (mean follow up period 62,7 months) are presented in this study. Methods: In this prospective, randomized study a minimum of 55 months follow up results involving the þrst 100 implanted metal liner total hip arthroplasties are presented. The mean follow up was 62,7 months. Between April 1995 and November 1996 ninety-eight patients (100 hips) had a total hip replacement consisting of a titanium cementless self reaming, parabolic cup, a cementless titanium stem and the Lubrimet¨ metal on metal articulation. 98 patients (100 hips) had complete clinical and radiographic data 55 to 89 months after the operation. One acetabular and one femoral component had to be revised due to aseptic loosening without showing macroscopic evidence of metallosis and no histological evidence of excessive metal wear. As a none device related orthopedic complication one acetabular component required revision surgery due to a peri-prosthetic fracture. Conclusions: The mid-term results of the Hofer-Imhof Lubrimet¨Metal-on-metal articulation are encouraging and so the system may represent a viable alternative for total hip arthroplasty in younger higher demand patients


Objectives. Total hip replacement is increasingly being conducted in younger and more active patients, so surgeons often use bearing surfaces with improved wear characteristics, such as ceramic on ceramic. The primary objective of this study was to determine if survivorship for a BIOLOX® delta ceramic on delta ceramic couple used with the PROCOTYL® L acetabular cup is significantly different from all other cementless cups in a large arthroplasty registry. The secondary objective of this study was to analyze patient reported outcomes measures (PROMs) of the subject cup with a minimum five year follow-up. Methods. Patient demographics and survivorship data was collected from the National Joint Registry of England, Wales, Northern Ireland, and the Isle of Man (NJR) database for all total hip replacements performed with the PROCOTYL® L cup used in combination with a delta-on-delta articulation, as well as for all other cementless cups. Survivorship data was compared for all revisions and cup revisions only and data was adjusted to exclude metal on metal articulations. The hazard ratio of the subject system to all cementless cups was also calculated with the Cox Proportional Hazards model. Patients with the subject components implanted for a minimum of five years completed Oxford Hip, EQ-5D, and EQ VAS score questionnaires. Results. The patient demographic data collected for the subject components and all cementless cups is provided in Figure 1. Six-year survivorship for the subject cup (98.6%) was similar to survivorship for all cementless cup revisions in the NJR database (98.5%), as seen in Figure 2. When the cup alone was revised, six-year survivorship of the subject cup (98.6%) and all NJR cementless cups (98.5%) was also similar. However, the subject cup survivorship remained at 98.6% from 4 to 6 years post-implantation, while survivorship for all cementless cups decreased slightly from years 4 to 6. The similarities between the revision risk of the subject system and all cementless cups in the NJR can be seen in the Cox Proportional Hazards model for revision risk ratios provided in Figure 3. Patients with the subject cup implanted for an average of 5.88 years reported Oxford Hip, EQ-5D, and EQ VAS scores of 39.60 ± 10.78, 0.801 ± 0.259, and 75.49 ± 19.25, respectively. Conclusions. The subject acetabular cup with a ceramic on ceramic articulation exhibited similar survivorship to all other cementless acetabular cups, excluding those with metal on metal bearings, in the NJR. Patients implanted with the subject system for an average of 5.88 years reported what are considered satisfactory Oxford Hip, EQ-5D, and EQ VAS scores. This survivorship and PROMs data is the first report of mid-term outcomes with the subject components


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 47 - 47
1 Nov 2015
Tucker K
Full Access

Introduction. Mix and Match (M&M) describes the use of components from more than one manufacturer in a total hip replacement (THR) The NJR has records of over 90,000 instances where this practice, which is contrary to the advice of most manufacturers and regulators, has been followed. Patients, Materials and Method. The NJR database 2003–13 was interrogated and the types of M&M were grouped using head size, bearing characteristics and use of cement. Results. Total Cohort. M&M. HARD ON SOFT. Metal head on poly socket. 345,618. 79,110. Ceramic Head on a poly socket. 42,176. 7,886. HARD ON HARD. Metal on Metal (all types). 31,609. 2,131. Ceramic on Ceramic. 87,156. 3,861. In a large number of cases, the M&M metal on plastic groups outshone their non-mix and matched controls. M&M “Head on Stem” has been shown to give inferior results. In the Ceramic on poly there is no significant difference in the performance between groups. In the Hard on Hard Metal on Metal groups the results are generally poor with some M&M doing better and some worse. In the case of ceramic on ceramic there is no significant difference so far. There have been outlier reports to MHRA for both M&M and non M&M groups. Discussion. This is the first time such a study, with such large numbers has ever been undertaken. The weakness of the study is the heavy reliance on “Revision” being the sole discriminator for failure, particularly in view of the sensitivity and potential incidence of under reporting of revision in NJR. Conclusion. It is suggested that BHS endorses the use of M&M where the results show it to be advantageous to patients and in revisions when it is appropriate. Surgeons should be aware of the results in the groups where there has not been any advantage of M&M


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_III | Pages 300 - 300
1 Nov 2002
Jakim I Velkes S
Full Access

Introduction: Historically Surface Replacement of the Hip exhibited a high failure rate due to femoral head loosening attributed to polyethylene wear debri, originating from the acetabulum. A metal on metal bearing resurfacing prosthesis has been developed to address this problem. The authors’ early experience using this prosthesis is presented. Material and methods: 48 patients underwent metal on metal hip Resurfacing Arthrolplasty (Cormet 2000R Corin U.K.) of the hip between 1999–2001. 43 patients suffered from osteoarthritis and 3 from avascular necrosis. 29 males and 19 females underwent the procedure with a mean age of 60 years (45–74). In 20 hips a posterior approach was used and in 28 our saggital trochenteric osteotomy. A cementless prosthesis was implanted in 30 patients and the femoral prosthesis was cemented in 18 patients. Results: All but two patients had an improved hip score with 95 % of the patients reporting a good to excellent result. There were 2 femoral neck fractures one early and one late. One patient had severe heterotrophic ossification and one a transient partial sciatic nerve palsy. Discussion: Conservative Hip Arthroplasty with Femoral Head and acetabular resurfacing is an attractive concept. Violation of the upper femoral canal is prevented and bone stock is preserved. Improved applications of metallurgical and tribological principles increases the predictability of metal on metal bearing surface function and prevents the catastrophic polyethylene wear previously observed in hip resurfacing procedures. The principles of metal on metal bearing surfaces, Resurfacing Arthroplasty of the hip and the early clinical results and complications are discussed


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 68 - 68
1 Mar 2017
Oladokun A Bryant M Hall R Neville A
Full Access

Introduction. Fretting corrosion at the Head-Neck taper interface of Large Metal on Metal (MoM), Metal on Polymer (MoP) and Ceramic on Ceramic (CoC) total hip arthroplasty (THA) remains a clinical concern. Ceramic femoral heads have gained a lot of attention more recently as a possible way to mitigate/reduce the dissolution of Cobalt Chromium ions. The objective of this study is to assess the fretting corrosion currents emanating from four material combinations for which Ti6Al4V and Co28Cr6Mo are the neck components of Co28Cr6Mo and BIOLOX®delta femoral heads at three different cyclic loads. Method. 12/14 Ti6Al4V and Co28Cr6Mo spigots (designed to geometrically represent the stem) were impacted against Ø36mm Co28Cr6Mo and BIOLOX®delta femoral heads with a static force of 2kN as shown in Figure 1. The tapers were immersed in 25% v/v diluted Foetal Bovine Serum, PBS balance and 0.03% Sodium Azide at room temperature. In-situ electrochemistry was facilitated using a 3-eletrode cell arrangement whereby the neck components were the working electrode, Ag/AgCl was the reference electrode and a platinum counter electrode completed the cell. All combinations were held at a potential of 0V vs. Ag/AgCl and the cyclic load applied unto each couple were 1kN, 3kN and 5kN at 1Hz consecutively (see Figure 2). The fretting corrosion currents were converted into cumulative charge transferred (Q) by integrating the wear enhanced corrosion current. Results and Discussion. Bergmann et al.1 plotted the loading profile of a patient weighing 1000N doing various daily living activities. In their study, the range in loading cycles vary from 1kN (standing on one leg) to ∼9kN (stumbling). For this study, Figure 2 shows the sinusoidal loading profile used and the corresponding charge transferred as a result of wear enhanced corrosion. The results reveal an increase in the cumulative charge for all four combinations as cyclic load increases. While for all combinations, no negligible amount of cumulative charge was measured at 1kN, no significant difference was observed at 3kN and at 5kN, the charge transferred from both MoM and CoM fretting couples where Ti6Al4V is the neck component were significantly lower than the couples with Co28Cr6Mo neck (see figure 3). The BIOLOX®delta – Ti6Al4V couple was observed to generate the least wear enhanced corrosion current. This, we observe, is due to thick agglomerated oxides resulting from wear and corrosion products which can adhere to the anodic fretting interface (see figure 3). Conclusion. This study reveals that for both MoM and CoM combinations, the charge transferred through wear enhanced corrosion of Ti6Al4V prove to be significantly lower than the combinations with Co28Cr6Mo alloy. Furthermore, this study proposes that the agglomeration of wear and corrosion products (oxides) can lead to the reduction of fretting corrosion currents at modular fretting interfaces as seen in combinations involving Ti6Al4V alloys. This is relevant as titanium alloys are known to form thick oxides at fretting contacts. For figures/tables, please contact authors directly.