WEAR OF HIGH PERFORMANCE BEARINGS IN THE HIP INFLUENCE OF MATERIAL, HEAD SIZE AND COMPONENT POSITION



Abstract

Young and active patients require bearing materials that can last up to 200 million walking steps, ten fold greater than conventional polyethylene bearings.

Cross linked polyethylene provides reduced wear rate compared to conventional polyethylene, and further advantage is gained from using ceramic femoral heads. However in polyethylene bearings wear increases with the head diameter, and there is currently little opportunity to use head sizes greater then 36mm diameter. There is evidence of polyethylene fracture with steeply positioned cups.

Ceramic on ceramic bearings provide substantially lower wear rates than polyethylene bearings. Steep cups, lateralised heads or neck impingement can lead to head contact on superior rim of the cup and stripe wear, but this still results in very low wear rates. Recently developed ceramic matrix composites Biolox Delta provide greater resistance to stripe wear. In a few patients stripe wear may lead to squeaking.

Metal on metal bearings also provide substantially lower wear than polyethylene bearings. However there remains concern about elevated metal ion levels in a few patients and resultant risk of hypersensitivity reactions. In metal on metal bearings larger head sizes and reduced diametrical clearance can lead to reduced wear. Increased wear is associated with steep cups and lateralised heads resulting in rim wear.

Ceramic on metal bearings have been introduced recently as the first differential hard on hard bearings. These bearings show substantial reduction in wear, corrosive wear mechanisms, metal ion levels in laboratory simulators and initial clinical studies have shown a reduction in metal ion levels in vivo compared to metal on metal bearings.