Introduction: Local problems of metal on metal (MOM) hip arthroplasty such as pseudotumours, neck thinning and osteolysis maybe related to concentrations of cobalt and chromium ions in the synovial fluid. There is little reported on these values. Our aim was to determine the range of metal ion levels in synovial fluid, and to investigate the relationship between these samples and simultaneous blood samples.
Methods: Synovial fluid and whole blood samples were taken from 30 consecutive patients at the time of revision surgery for a painful MOM hip. Aspirated fluid was not visibly contaminated with blood. Impants were in situ for a mean period of 31 months. All had normal renal function. Samples were analysed using ICP mass spectrometry and compared with 10 samples from patients without implants.
Results: The mean (and range) of synovial fluid metal ion levels were 1965 ug/l (30 to 13618) and 6265 ug/l (11 to 81630) for Cobalt and Chromium respectively. There was a good correlation between synovial and blood levels for both cobalt (R=0.65, p = 0.0001) and chromium (R = 0.59, p = 0.006).
Discussion and Conclusion: Metal ions in synovial fluid from MOM hips are generated from wear of the bearing surfaces, the correlation with blood metal ion levels, shown in this study, suggest that blood levels may be used as surrogate marker for hip wear rate. Our range of synovial fluid metal ion levels may be useful for those conducting in vitro studies on the biocompatibility of MOM hips.