Abstract
The purpose of this study is to identify clinical and radiographic results of 78 uncemented total hip arthroplasties using Metasul® metal on metal bearings with Wagner standard cup and proximal hydroxyapatite coated CLS stem.
Mean age was 39 years and average follow-up period was 11.7 years. Mean Harris hip score had improved from 51.4 points preoperatively to 95.2 points finally. There were 2 hips with progressive osteolysis around the acetabular cup. Of them, one hip was revised due to loosening of the cup, and the other was observed because of patient’s refusal to revise. In histopathologic findings on osteolytic area, a lot of macrophage phagocytizing metal debris and perivascular lymphocyte infiltration were found. Immunohistochemical analysis suggested delayed metal hypersensitivity. Serum cobalt levels in hips with osteolysis were not higher than those in hips without osteolysis.
Early osteolysis with sudden onset of groin pain in few hips remains a concern.
Correspondence should be addressed to ISTA Secretariat, PO Box 6564, Auburn, CA 95604, USA. Tel: 1-916-454-9884, Fax: 1-916-454-9882, Email: ista@pacbell.net