Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 53 - 53
1 Nov 2021
ten Heggeler R Schröder F de Graaf F Fluit R Becea D Verdonschot N Hoogeslag R
Full Access

Introduction and Objective

After anterior cruciate ligament reconstruction one of the risk factors for graft (re-)rupture is an increased posterior tibial slope (PTS). The current treatment for PTS is a high tibial osteotomy (HTO). This is a free-hand method, with 1 degree of tibial slope correction considered to be equal to 1 or even 1.67 mm of the anterior wedge resection. Error rates in the frontal plane reported in literature vary from 1 – 8.6 degrees, and in the sagittal plane outcomes in a range of 2 – 8 degrees are reported when planned on PTSs of 3 – 5 degrees. Therefore, the free-hand method is considered to have limited accuracy. It is expected that HTO becomes more accurate with patient specific saw guides (PSGs), with an accuracy margin reported in literature of 2 degrees. This proof of concept porcine cadaver case study aimed to investigate whether the use of PSGs improves the accuracy of HTO to less than 2 degrees. Secondly, the reproducibility of tibial slope measurement was evaluated.

Materials and Methods

Preoperative MRI images of porcine cadaver knees (n = 3) were used to create 3D anatomical bone models (Mimics, Materialise, Belgium). These 3D models were subsequently used to develop PSGs (3-Matic, Materialise, Belgium) to correct all tibias for 3 degrees PTS and 4 degrees varus. The PSG mediated HTOs were performed by an experienced orthopaedic surgeon, after which postoperative MRI images were obtained. 3D anatomical models of postoperative tibias were created, and tibial slopes were assessed on both pre- and postoperative tibias. The tibial slope was defined as the angle between the mechanical axis and 3D tibial reference plane in the frontal and sagittal plane. The accuracy of the PSG mediated HTO (median and range) was defined as the difference in all possible combinations of the preoperatively planned and postoperatively obtained tibial slopes. To ensure reproducibility, the pre- and postoperative tibial slopes were measured thrice by one observer. The intra-class correlation coefficients (ICCs) were subsequently calculated to assess the intra-rater reliability (SPSS, IBM Corp., Armonk, N.Y., USA).


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 125 - 125
1 Mar 2021
Eggermont F van der Wal G Westhoff P Laar A de Jong M Rozema T Kroon HM Ayu O Derikx L Dijkstra S Verdonschot N van der Linden YM Tanck E
Full Access

Patients with cancer and bone metastases can have an increased risk of fracturing their femur. Treatment is based on the impending fracture risk: patients with a high fracture risk are considered for prophylactic surgery, whereas low fracture risk patients are treated conservatively with radiotherapy to decrease pain. Current clinical guidelines suggest to determine fracture risk based on axial cortical involvement of the lesion on conventional radiographs, but that appears to be difficult. Therefore, we developed a patient-specific finite element (FE) computer model that has shown to be able to predict fracture risk in an experimental setting and in patients. The goal of this study was to determine whether patient-specific finite element (FE) computer models are better at predicting fracture risk for femoral bone metastases compared to clinical assessments based on axial cortical involvement on conventional radiographs, as described in current clinical guidelines.

45 patients (50 affected femurs) affected with predominantly lytic bone metastases who were treated with palliative radiotherapy for pain were included. CT scans were made and patients were followed for six months to determine whether or not they fractured their femur. Non-linear isotropic FE models were created with the patient-specific geometry and bone density obtained from the CT scans. Subsequently, an axial load was simulated on the models mimicking stance. Failure loads normalized for bodyweight (BW) were calculated for each femur. High and low fracture risks were determined using a failure load of 7.5 × BW as a threshold. Experienced assessors measured axial cortical involvement on conventional radiographs. Following clinical guidelines, patients with lesions larger than 30 mm were identified as having a high fracture risk. FE predictions were compared to clinical assessments by means of diagnostic accuracy values (sensitivity, specificity and positive (PPV) and negative predictive values (NPV)).

Seven femurs (14%) fractured during follow-up. Median time to fracture was 8 weeks. FE models were better at predicting fracture risk in comparison to clinical assessments based on axial cortical involvement (sensitivity 100% vs. 86%, specificity 74% vs. 42%, PPV 39% vs. 19%, and NPV 100% vs. 95%, for the FE computer model vs. axial cortical involvement, respectively). We concluded that patient-specific FE computer models improve fracture risk predictions of femoral bone metastases in advanced cancer patients compared to clinical assessments based on axial cortical involvement, which is currently used in clinical guidelines. Therefore, we are initiating a pilot for clinical implementation of the FE model.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 77 - 77
1 Mar 2021
Ataei A Eggermont F Baars M Linden Y Rooy J Verdonschot N Tanck E
Full Access

Patients with advanced cancer can develop bone metastases in the femur which are often painful and increase the risk of pathological fracture. Accurate segmentation of bone metastases is, amongst others, important to improve patient-specific computer models which calculate fracture risk, and for radiotherapy planning to determine exact radiation fields. Deep learning algorithms have shown to be promising to improve segmentation accuracy for metastatic lesions, but require reliable segmentations as training input. The aim of this study was to investigate the inter- and intra-operator reliability of manual segmentation of femoral metastatic lesions and to define a set of lesions which can serve as a training dataset for deep learning algorithms. F

CT-scans of 60 advanced cancer patients with a femur affected with bone metastases (20 osteolytic, 20 osteoblastic and 20 mixed) were used in this study. Two operators were trained by an experienced radiologist and then segmented the metastatic lesions in all femurs twice with a four-week time interval. 3D and 2D Dice coefficients (DCs) were calculated to quantify the inter- and intra-operator reliability of the segmentations. We defined a DC>0.7 as good reliability, in line with a statistical image segmentation study.

Mean first and second inter-operator 3D-DCs were 0.54 (±0.28) and 0.50 (±0.32), respectively. Mean intra-operator I and II 3D-DCs were 0.56 (±0.28) and 0.71 (±0.23), respectively. Larger lesions (>60 cm3) scored higher DCs in comparison with smaller lesions.

This study reveals that manual segmentation of metastatic lesions is challenging and that the current manual segmentation approach resulted in dissatisfying outcomes, particularly for lesions with small volumes. However, segmentation of larger lesions resulted in a good inter- and intra-operator reliability. In addition, we were able to select 521 slices with good segmentation reliability that can be used to create a training dataset for deep learning algorithms. By using deep learning algorithms, we aim for more accurate automated lesion segmentations which might be used in computer modelling and radiotherapy planning.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 9 - 9
1 Feb 2021
Soltanihafshejani N Bitter T Janssen D Verdonschot N
Full Access

Introduction

The fixation of press-fit orthopaedic devices depends on the mechanical properties of the bone that is in contact with the implants. During the press-fit implantation, bone is compacted and permanently deformed, finally resulting in the mechanical interlock between implant and bone. For the development and design of new devices, it is imperative to understand these non-linear interactions. One way to investigate primary fixation is by using computational models based on Finite Element (FE) analysis. However, for a successful simulation, a proper material model is necessary that accurately captures the non-linear response of the bone. In the current study, we combined experimental testing with FE modeling to establish a Crushable Foam model (CFM) to represent the non-linear bone biomechanics that influences implant fixation.

Methods

Mechanical testing of human tibial trabecular bone was done under uniaxial and confined compression configurations. We examined 62 human trabecular bone samples taken from 8 different cadaveric tibiae to obtain all the required parameters defining the CFM, dependent on local bone mineral density (BMD). The derived constitutive rule was subsequently applied using an in-house subroutine to the FE models of the bone specimens, to compare the model predictions against the experimental results.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_1 | Pages 50 - 50
1 Feb 2021
Sanchez E Schilling C Grupp T Giurea A Verdonschot N Janssen D
Full Access

Introduction

Cementless total knee arthroplasty (TKA) implants use an interference fit to achieve fixation, which depends on the difference between the inner dimensions of the implant and outer dimensions of the bone. However, the most optimal interference fit is still unclear. A higher interference fit could lead to a superior fixation, but it could also cause bone abrasion and permanent deformation during implantation. Therefore, this study aims to investigate the effect of increasing the interference fit from 350 µm to 700 µm on the primary stability of cementless tibial implants by measuring micromotions and gaps at the bone-implant interface when subjected to two loading conditions.

Methods

Two cementless e.motion® tibial components (Total Knee System, B. Braun) with different interference fit and surface coating were implanted in six pairs of relatively young human cadaver tibias (47–60 years). The Orthoload peak loads of gait (1960N) and squat (1935N) were applied to the specimens with a custom made load applicator (Figure 1A). The micromotions (shear displacement) and opening/closing gaps (normal displacement) were measured with Digital Image Correlation (DIC) in 6 different regions of interest (ROIs - Figure 1B). Two General Linear Mixed Models (GLMMs) were created with micromotions and interfacial gaps as dependent variables, bone quality, loading conditions, ROIs, and interference fit implants as independent variables, and the cadaver specimens as subject variables.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 99 - 99
1 Feb 2020
Schroder F Post C Simonis F Wagenaar F in'tVeld RH Verdonschot N
Full Access

Introduction

Instability, loosening, and patellofemoral pain belong to the main causes for revision of total knee arthroplasty (TKA). Currently, the diagnostic pathway requires various diagnostic techniques such as x-rays, CT or SPECT-CT to reveal the original cause for the failed knee prosthesis, but increase radiation exposure and fail to show soft-tissue structures around TKA. There is a growing demand for a diagnostic tool that is able to simultaneously visualize soft tissue structures, bone, and TKA without radiation exposure. MRI is capable of visualising all the structures in the knee although it is still disturbed by susceptibility artefacts caused by the metal implant. Low-field MRI (0.25T) results in less metal artefacts and offers the ability to visualize the knee in weight-bearing condition. Therefore, the aim of this study is to investigate the possibilities of low field MRI to image, the patellofemoral joint and the prosthesis to evaluate the knee joint in patients with and without complaints after TKA.

Method

Ten patients, eight satisfied and two unsatisfied with their primary TKA, (NexGen posterior stabilized, BiometZimmer) were included. The patients were scanned in sagittal, coronal, and transversal direction on a low field MRI scanner (G-scan Brio, 0.25T, Esaote SpA, Italy) in weight-bearing and non-weight-bearing conditions with T1, T2 and PD-weighted metal artefact reducing sequences (TE/TR 12–72/1160–7060, slice thickness 4.0mm, FOV 260×260×120m3, matrix size 224×216). Scans were analysed by two observers for:

- Patellofemoral joint: Caton-Descamps index and Tibial Tuberosity-Trochlear Groove (TT-TG) distance.

- Prosthesis malalignment: femoral component rotation using the posterior condylar angle (PCA) and tibial rotation using the Berger angle.

Significance of differences in parameters between weight-bearing and non-weight-bearing were calculated with the Wilcoxon rank test. To assess the reliability the inter and intra observer reliability was calculated with a two-way random effects model intra class correlation coefficient (ICC). The two unsatisfied patients underwent revision arthroplasty and intra-operative findings were compared with MRI findings.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 52 - 52
1 Feb 2020
Sadhwani S Picache D Janssen D de Ruiter L Rankin K Briscoe A Verdonschot N Shah A
Full Access

Introduction

Polyetheretherketone (PEEK) has been proposed as an implant material for femoral total knee arthroplasty (TKA) components. Potential clinical advantages of PEEK over standard cobalt chrome alloys include modulus of elasticity and subsequently reduced stress shielding potentially eliminating osteolysis, thermal conduction properties allowing for a more natural soft tissue environment, and reduced weight enabling quicker quadriceps recovery. Manufacturing advantages include reduced manufacturing and sterilization time, lower cost, and improved quality control. Currently, no PEEK TKA implants exist on the market. Therefore, evaluation of mechanical properties in a pre-clinical phase is required to minimize patient risk.

The objectives of this study include evaluation of implant fixation and determination of the potential for reduced stress shielding using the PEEK femoral TKA component.

Methods and Materials

Experimental and computational analysis was performed to evaluate the biomechanical response of the femoral component (Freedom Knee, Maxx Orthopedics Inc., Plymouth Meeting, PA; Figure 1).

Fixation strength of CoCr and PEEK components was evaluated in pull-off tests of cemented femoral components on cellular polyurethane foam blocks (Sawbones, Vashon Island, WA). Subsequent testing investigated the cemented fixation using cadaveric distal femurs. The reconstructions were subjected to 500,000 cycles of the peak load occurring during a standardized gait cycle (ISO 14243-1). The change from CoCr to PEEK on implant fixation was studied through computational analysis of stress distributions in the cement, implant, and the cement-implant interface. Reconstructions were analyzed when subjected to standardized gait and demanding squat loads.

To investigate potentially reduced stress shielding when using a PEEK component, paired cadaveric femurs were used to measure local bone strains using digital image correlation (DIC). First, standardized gait load was applied, then the left and right femurs were implanted with CoCr and PEEK components, respectively, and subjected to the same load. To verify the validity of the computational methodology, the intact and reconstructed femurs were replicated in FEA models, based on CT scans.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 81 - 81
1 Apr 2019
Bitter T Marra M Khan I Marriott T Lovelady E Verdonschot N Janssen D
Full Access

Introduction

Fretting corrosion at the taper interface of modular connections can be studied using Finite Element (FE) analyses. However, the loading conditions in FE studies are often simplified, or based on generic activity patterns. Using musculoskeletal modeling, subject-specific muscle and joint forces can be calculated, which can then be applied to a FE model for wear predictions. The objective of the current study was to investigate the effect of incorporating more detailed activity patterns on fretting simulations of modular connections.

Methods

Using a six-camera motion capture system, synchronized force plates, and 45 optical markers placed on 6 different subjects, data was recorded for three different activities: walking at a comfortable speed, chair rise, and stair climbing.

Musculoskeletal models, using the Twente Lower Extremity Model 2.0 implemented in the AnyBody modeling System™ (AnyBody Technology A/S, Aalborg, Denmark; figure1), were used to determine the hip joint forces. Hip forces for the subject with the lowest and highest peak force, as well as averaged hip forces were then applied to an FE model of a modular taper connection (Biomet Type-1 taper with a Ti6Al4V Magnum +9 mm adaptor; Figure 2). During the FE simulations, the taper geometry was updated iteratively to account for material removal due to wear. The wear depth was calculated based on Archard's Law, using contact pressures, micromotions, and a wear factor, which was determined from accelerated fretting experiments.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 135 - 135
1 Apr 2019
Post C Schroder FF Simonis FJJ Peters A Huis In't Veld R Verdonschot N
Full Access

Introduction

Fifteen percent of the primary total knee arthroplasties (TKA) fails within 20 years. Among the main causes for revision surgery are instability and patellofemoral pain. Currently, the diagnostic pathway requires various diagnostic techniques to reveal the original cause for the failed knee prosthesis and is therefore time consuming and inefficient.

Accordingly, there is a growing demand for a diagnostic tool that is able to simultaneously visualize soft tissue structures, bone and TKA. Magnetic resonance imaging (MRI) is capable of visualising all the structures in the knee although a trade- off needs to be made between metal artefact reducing capacities and image quality. Low-field MRI (0.25T) results in less metal artefacts and a lower image quality compared with high-field MRI (1.5T). The aim of this study is to develop a MRI imaging guide to image the problematic TKA and to evaluate this guide by comparing low-field and high-field MRI on a case study.

Method

Based on literature and current differential diagnostic pathways a guide to diagnose patellofemoral pain, instability, malposition and signs of infection or fracture with MRI was developed. Therefore, methods as Insall Salvati, patellar tilt angle and visibility of fluid and soft tissues were chosen. Visibility was scored on a VAS scale from 0 to 100mm (0mm zero visibility, 100mm excellent visibility).

Subsequently, this guide is used to analyse MRI scans made of a volunteer (female, 61 years, right knee) with primary TKA (Biomet, Zimmer) in sagittal, coronal and transversal direction with a FSE PD metal artefact reducing (MAR) sequence (TE/TR 12/1030ms, slice thickness 4.0mm, FOV 260×260×120mm3, matrix size 224×216) on low-field MRI (Esaote G-scan Brio, 0.25T) and with a FSE T1-weighted high bandwidth MAR sequence (TE/TR 6/500ms, slice thickness 3.0mm, FOV 195×195×100mm3, matrix size 320×224) on high-field MRI (Avanto 1.5T, Siemens).

Scans were analysed three times by one observer and the intra observer reliability was calculated with a two-way random effects model intra class correlation coefficient (ICC).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 18 - 18
1 Apr 2019
Schröder FF Huis In't Veld R Simonis FJJ Post C Vochteloo AJH Verdonschot N
Full Access

Introduction

Roentgen stereophotogrammetric analysis (RSA) is currently the gold standard to measure early prosthetic migration which can predict aseptic loosening. However, RSA has some limitations such as the need for perioperative placed markers and exposure to X-radiation during follow up. Therefore, this study evaluates if low field MRI could be an alternative for RSA. Low field MRI was chosen because it is less hampered by metal artifacts of the prosthesis than high field MRI.

Methods

3D models of both the tibial component of a total knee prosthesis (Genesis II, Smith and Nephew) and the porcine tibia were made. The tibial component was implanted in the tibial bone. Consequently, 17 acquisitions with the low field MRI scanner (Esaote G-scan 0.25T) in transverse direction with a 2D PD weighted metal artifact reducing sequence PD-XMAR (TE/TR 10/1020ms, slice thickness 3mm, FOV 180×180×120 mm³, matrix size 224×224) were made. The first five acquisitions were made without repositioning the cadaver, the second twelve after slightly repositioning the cadaver within limits that are expected to be encountered in a clinical setting. Hence, in these 17 acquisitions no prosthetic-bone motions were induced.

The scans were segmented and registered with Mimics. Virtual translation and rotation of the prosthesis with respect to the bone between two scans were calculated using a Procrustes algorithm. The first five scans without repositioning were used to calculate the measurement error, the following twelve to calculate the precision of low field MRI to measure prosthetic migration. Results were expressed as the maximum total point motion, mean error and 95% CI and expressed in boxplots.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 125 - 125
1 Apr 2019
Sanchez E Schilling C Grupp TM Verdonschot N Janssen D
Full Access

Introduction

Although cementless press-fit femoral total knee arthroplasty (TKA) components are routinely used in clinical practice, the effect of the interference fit on primary stability is still not well understood. Intuitively, one would expect that a thicker coating and a higher surface roughness lead to a superior fixation. However, during implant insertion, a thicker coating can introduce more damage to the underlying bone, which could adversely influence the primary fixation. Therefore, in the current study, the effect of coating thickness and roughness on primary stability was investigated by measuring the micromotions at the bone-implant interface with experimental testing.

Methods

A previous experimental set-up was used to test 6 pairs of human cadaveric femurs (47–60 years, 5 females) implanted with two femoral component designs with either the standard e.motion (Total Knee System, B. Braun, Germany) interference fit of 350 µm (right femurs) or a novel, thicker interference fit of 700 µm (left femurs). The specimens were placed in a MTS machine (Figure 1) and subjected to the peak loads of normal gait (1960N) and squat (1935N), based on the Orthoload dataset for Average 75.

Varus/valgus moments were incorporated by applying the loads at an offset relative to the center of the implants, leading to a physiological mediolateral load distribution. Under these loads, micromotions at the implant-bone interface were measured using Digital Image Correlation (DIC) at different regions of interest (ROIs – Figure 1). In addition, DIC was used to measure opening and closing of the implant-bone interface in the same ROIs.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 93 - 93
1 Apr 2019
Anijs T Janssen D Verdonschot N
Full Access

Introduction

Aseptic loosening is the main reason for total knee arthroplasty (TKA) failure, responsible for more than 25% of the revision procedures, with most of the problems occurring with the tibial component. While early loosening can be attributed to failure of primary fixation, late implant loosening is associated with loss of fixation secondary to bone resorption due to altered physiological load transfer to the tibial bone. Several attempts have been made to investigate these changes in bone load transfer in biomechanical simulations and bone remodeling analyses, which can be useful to provide information on the effect of patient, surgery, or design-related factors. On the other hand, these factors have also been investigated in clinical studies of radiographic changes of bone density following TKA. In this study we made an overview of the knowledge obtained from these clinical studies, which can be used to inform clinical decision making and implant design choices.

Methods

A literature search was performed to identify clinical follow-up studies that monitored peri-prosthetic bone changes following TKA. Within these studies, effects of the following parameters on bone density changes were investigated: post-operative time, region of interest, alignment, body weight, systemic osteoporosis, implant design and cementation.

Moreover, we investigated the effect of bone density loss on implant survival. Results

A total of 19 studies was included in this overview, with a number of included patients ranging from 12 to 7,760. Most studies used DEXA (n=16), while a few studies performed analyses on calibrated digital radiographs (n=2), or computed tomography (n=1). Postoperative follow-up varied from 9 months to 10 years.

Studies consistently report the largest bone density reduction within the first postoperative year. Bone loss is mainly seen in the medial region. This has been attributed to the change in alignment following surgery, during which often the pre-operative varus knee is corrected to a more physiological alignment, resulting in a load shift towards the lateral compartment. Measurements in unoperated contralateral legs were performed in 3 cases, and two studies performed standardized DEXA measurements to provide information on systemic osteoporosis. While on the short term no changes were observed, significant negative correlations have been found between severity of osteoporosis and peri-prosthetic bone density. No clear effects of bodyweight and cementation on bone loss have been identified. Although some studies do find differences between implant types, the variation in the data makes it difficult to draw general conclusions from these findings.

Several studies reported no effect of bone loss on implant migration. In another study, a medial collapse was associated with a medial increase in density, suggesting that altered loading and increased stresses are responsible for both bone formation and the overload leading to collapse.


Bone & Joint Research
Vol. 7, Issue 6 | Pages 430 - 439
1 Jun 2018
Eggermont F Derikx LC Verdonschot N van der Geest ICM de Jong MAA Snyers A van der Linden YM Tanck E

Objectives

In this prospective cohort study, we investigated whether patient-specific finite element (FE) models can identify patients at risk of a pathological femoral fracture resulting from metastatic bone disease, and compared these FE predictions with clinical assessments by experienced clinicians.

Methods

A total of 39 patients with non-fractured femoral metastatic lesions who were irradiated for pain were included from three radiotherapy institutes. During follow-up, nine pathological fractures occurred in seven patients. Quantitative CT-based FE models were generated for all patients. Femoral failure load was calculated and compared between the fractured and non-fractured femurs. Due to inter-scanner differences, patients were analyzed separately for the three institutes. In addition, the FE-based predictions were compared with fracture risk assessments by experienced clinicians.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 14 - 14
1 Apr 2018
Bitter T Khan I Marriott T Lovelady E Verdonschot N Janssen D
Full Access

Introduction

Fretting corrosion of the modular taper junction in total hip arthroplasty has been studied in several finite element (FE) studies. Manufacturing tolerances can result in a mismatch between the femoral head and stem, which can influence the taper mechanics leading to possibly more wear. Using FE models the effect of these manufacturing tolerances on the amount of volumetric wear can be studied. The removal of material in the FE model was validated against experiments simulating the clinical fretting wear process, subsequently the mismatch and assembly force were varied to study the effect on the volumetric wear.

Methods

An FE model was developed in which the geometry can be updated to account for material removal due to wear. In this model the geometry was updated based on Archard's Law, using contact pressures, micromotions and a wear factor, which was determined based on accelerated fretting experiments. The linear wear was calculated using H=k*p*S. Where H is the linear wear depth in mm, k is a wear factor (mm3/Nmm), p is the contact pressure (MPa) and S is the sliding distance (mm). 10 million cycles were simulated using 50 virtual steps. Using this scaling and the measured volumetric wear from the experiments a wear factor of 2.7*10−5 was applied.

Based on general manufacturing tolerances the resulting mismatch in taper angles were determined to be ± 1.26°. Using this mismatch a tip fit (figure 1a) and base fit (Figure 1b) model were created. In combination with a perfect fit, meaning no mismatch, and two different assembly forces of 4 kN and 15 kN, 6 different situations were studied.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 78 - 78
1 Apr 2018
Srinivasan P Miller M Verdonschot N Mann K Janssen D
Full Access

INTRODUCTION

Mechanical overloading of the knee can occur during activities of daily living such as stair climbing, jogging, etc. In this finite element study we aim to investigate which parameters could detrimentally influence peri-implant bone in the tibial reconstructed knee. Bone quality and patient variables are potential factors influencing knee overloading (Zimmerman 2016).

METHODS

Finite element (FE) models of post-mortem retrieved tibial specimens (n=7) from a previous study (Zimmerman 2016) were created using image segmentation (Mimics Materialise v14) of CT scan data (0.6 mm voxel resolution). Tibial tray and polyethylene inserts were recreated from CT data and measurements of the specimens (Solidworks 2015). Specimens with varying implant geometry (keel/pegged) were chosen for this study. A cohesive layer between bone and cement was included to simulate the behavior of the bone–cement interface using experimentally obtained values. The FE models predict plasticity of bone according to Keyak (2005). Models were loaded to 10 body weight (BW) and then reduced to 1 BW to mimic experimental measurements. Axial FE bone strains at 1 BW were compared with experimental Digital Image Correlation (DIC) bone strains on cut sections of the specimens.

After validation of the FE models using strain data, models were rotated and translated to the coordinate system defined in Bergmann (2014). Four loading cases were chosen – walking, descending stairs, sitting down and jogging. Element strains were written to file for post-processing. The bone in all FE models was divided into regions of equal thickness (10 mm) for comparison of strains.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_20 | Pages 52 - 52
1 Dec 2017
Niu K Sluiter V Sprengers A Homminga J Verdonschot N
Full Access

Improving the accuracy of measuring 6 degree of freedom tibiofemoral kinematics is a crucial step in gait analysis, but skin-marker estimated kinematics are subject to soft tissue artefacts. Fluoroscopic systems have been reported to achieve high accurate kinematics, but their induced irradiation, limited field of view, and high cost hampers routine usage on large patient cohorts. The aim of this study is to assess the feasibility of measuring tibiofemoral kinematics using multi-channel A-mode ultrasound system in cadaver experiment and to assess its achievable accuracy.

A full cadaver was placed with its back on a surgery table while its legs were overhanging the edge of the table. Upper body was fixated and right leg was moved by means of pulling a rope. Two bone pins with optical markers were mounted to the femur and tibia separately to measure the ground truth of motion. Six custom holders containing 30 A-mode ultrasound transducers and 18 optical markers were mounted to six anatomical regions. By measuring the bone to ultrasound transducer distance and using the spatial information of the optical markers on the holders, 30 bone surface points were determined. The corresponding bones (femur and tibia) were registered to these acquired points after which the tibiofemoral kinematics were determined.

This study presents a multi-channel A-mode ultrasound system and the first results have shown its feasibility of reconstructing tibiofemoral kinematics in cadaver experiment. Although the reconstructed tibiofemoral kinematics is less accurate than a fluoroscopic system, it outperforms a skin-mounted markers system. Thus, this A-mode Ultrasound approach could provide a non-invasive and non-radiative method for measuring tibiofemoral kinematics, which may be used in clinic gait analysis or even computer-aided orthopaedic surgery.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 20 - 20
1 Apr 2017
Meijer M Boerboom A Stevens M Reininga I Janssen D Verdonschot N
Full Access

Background

Trabecular metal (TM) cones are designed to fill up major bone defects in total knee arthroplasty. Tibial components can be implanted in combination with a stem, but it is unclear if this is necessary after reconstruction with a TM cone. Implanting a stem may give extra stability, but may also have negative side-effects. Aim of this study was to investigate stability and strain distribution of a tibial plateau reconstruction with a TM cone while the tibal component is implanted with and without a stem, and whether prosthetic stability was influenced by bone mineral density (BMD).

Methods

Tibial revision arthroplasties were performed after reconstruction of an AORI 2B bone defect with TM cones. Plateaus were implanted in seven pairs of cadaveric tibiae; of each pair, one was implanted with and the other without stem. All specimens were loaded to one bodyweight alternating between the medial and lateral tibia plateau. Implant-bone micro motions, bone strains, BMD and correlations were measured and/or calculated.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 24 - 24
1 Apr 2017
Janssen D Bitter T Schreurs B Marriott T Khan I Verdonschot N
Full Access

Background

Fretting at modular junctions is thought to be a ‘mechanically assisted’ corrosion phenomenon, initiated by mechanical factors that lead to increased contact stresses and micromotions at the taper interface. We adopted a finite element approach to model the head-taper junction, to analyse the contact mechanics at the taper interface. We investigated the effect of assembly force and angle on contact pressures and micromotions, during loads commonly used to test hip implants, to demonstrate the importance of a good assembly during surgery.

Methods

Models of the Bimetric taper and adaptor were created, with elastic-plastic material properties based on material tests with the actual implant alloy. FE contact conditions were validated against push-on and pull-off experiments. The models were loaded according to ISO 7206-4 and −6, after being assembled at 2-4-15kN, both axially and at a 30° angle. Average micromotions and contact pressures were analysed, and a wear score was calculated based on the contact pressures and micromotions.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 40 - 40
1 Feb 2017
Berahmani S Hendriks M Janssen D Verdonschot N
Full Access

The primary stability of an uncemented femoral total knee replacement component is provided by press-fit forces at the bone-implant interface. This press-fit is achieved by resecting the bone slightly larger than the inner dimensions of the implant, resulting in a so-called interference fit. Previous animal studies have shown that an adequate primary stability is required to minimize micromotions at the bone-implant interface to achieve bone-ingrowth, which provides the secondary (long-term) fixation. It is assumed that during implantation a combination of elastic and plastic deformation and abrasion of the bone will occur, but little is known about what happens at the bone-implant interface and how much interference fit eventually is achieved. Purpose of this study was therefore to assess the actual and effective interference fit and the amount of bone damage during implantation of an uncemented femoral knee component.

In this study, five cadaveric distal femora were prepared and femoral knee components were implanted by an experienced surgeon. Micro-CT scans and conventional CT-scans were obtained pre- and post-implantation for geometrical measurements and to measure bone mineral density. In addition, the position of the implant with respect to the bone was determined by optical scanning of the reconstructions (Figure.1). By measuring the differences in surface geometry, assessments were made of the cutting error, the actual interference fit, the amount of bone damage, and the effective interference fit.

Our analysis showed an average cutting error of 0.67± 0.17 mm, which pointed mostly towards bone under-resections. We found an average actual AP interference fit of 1.48± 0.27 mm, which was close to the nominal value of 1.5 mm. We observed combinations of bone damage and elastic deformation in all bone specimens (Figure. 2), which showed a trend to be related with bone density. Higher bone density tended to lead to lower bone damage and higher elastic deformation (Figure. 3). The results of the current study indicate different factors that interact while implanting an uncemented femoral knee component. This knowledge can be used to fine-tune design criteria of femoral components and obtain adequate primary stability for all patients in a more predictable way.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 139 - 139
1 Feb 2017
Marra M Heesterbeek P van de Groes S Janssen D Koopman B Wymenga A Verdonschot N
Full Access

Introduction

Tibial slope was shown to majorly affect the outcomes of Total Knee Arthroplasty (TKA). More slope of the tibial component could help releasing a too tight flexion gap in cruciate-retaining (CR) TKA and is generally associated with a wider range of post-operative knee flexion. However, an excessive tibial slope could jeopardize the knee stability in flexion. The mechanism by which tibial slope affects the function of CR-TKA is not well understood. Moreover, it is not known whether the tibial bone resection should be performed by referencing the anterior cortex (AC) of the tibia or the center of the tibial plateau (CP) and whether the choice of either technique plays a role. The aim of this study was to investigate the effect of tibial slope on the position of tibiofemoral (TF) contact point, knee ligament forces, quadriceps muscle forces, and TF and patellofemoral (PF) joint contact forces during squat activity in CR-TKA.

Methods

A previously validated musculoskeletal model of CR-TKA was used to simulate a squat activity performed by a 86-year-old male subject wearing an instrumented prosthesis [1,2]. Marker data over four consecutive repetitions of a squat motion were tracked using a motion optimization algorithm. Muscle and joint forces and moments were calculated from an inverse-dynamic analysis, coupled with Force-Dependent Kinematics (FDK) to solve knee kinematics, ligament and contact forces simultaneously. The tibial slope in the postoperative case was 0 degree and constituted the reference case for our simulations. In addition, eight additional cases were simulated with −3, +3, +6, +9 degrees of tibial slope, four of them simulating an AC referencing technique and four a CP technique.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 46 - 46
1 Feb 2017
Bitter T Janssen D Schreurs B Marriott T Lovelady E Khan I Verdonschot N
Full Access

Introduction

Fretting corrosion of the modular taper junction in total hip arthroplasty has been studied in several finite element (FE) investigations. In FE analyses, different parameters can be varied to study micromotions and contact pressures at the taper interface. However, to truly study taper wear, the simulation of micromotions and contact pressures in non-adaptive FE models is insufficient, as over time these can change due to interfacial changes caused by the wear process.

In this study we developed an FE approach in which material removal during the wear process was simulated by adaptations to the taper geometry. The removal of material was validated against experiments simulating the clinical fretting wear process.

Method

Experimental test: An accelerated fretting screening test was developed that consistently reproduced fretting wear features observed in retrievals. Biomet Type-1 (4°) tapers and +9 mm offset adaptors were assembled with a 4 kN force (N=3). A custom head fixture was used to create an increased offset and torque. The stems were potted in accordance with ISO 7206–6:2013. The set-up was submerged in a 37°C PBS solution with a pH adjusted to 3 using HCL and NaCl concentration of 90gl−1. The components were cyclically loaded between 0.4 – 4 kN for 10 million cycles. After completion, the volumetric and linear wear was measured using a Talyrond-585 roundness measurement machine.

FE model: This was created to match the experimental set up (Figure 1). Taper geometry and experimental material data were obtained from the manufacturer (Zimmer Biomet). The coefficient of friction of the studied combination of components was based on previous experiments (Bitter, 2016). After each change in load the geometry was updated by moving nodes inwards perpendicular to the taper surface. Archard's Law (Archard, 1953) was used to calculate the wear with the following equation: H=k*p*S. Where H is the linear wear depth in mm, k is a wear factor (mm³/Nmm), p is the contact pressure (MPa) and S is the sliding distance (mm). The 10 million experimental cycles were simulated using a range of 5 to 200 computational cycles. For this purpose, the wear factor (k) was scaled for each simulation to match the volumetric wear found in the experiments.


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1086 - 1092
1 Aug 2016
de Vos MJ Wagener ML Hannink G van der Pluijm M Verdonschot N Eygendaal D

Aims

Revision total elbow arthroplasty (TEA) is often challenging. The aim of this study was to report on the clinical and radiological results of revision arthroplasty of the elbow with the Latitude TEA.

Patients and Methods

Between 2006 and 2010 we used the Latitude TEA for revision in 18 consecutive elbows (17 patients); mean age 53 years (28 to 80); 14 women. A Kudo TEA was revised in 15 elbows and a Souter-Strathclyde TEA in three.

Stability, range of movement (ROM), visual analogue score (VAS) for pain and functional scores, Elbow Functional Assessment Scale (EFAS), the Functional Rating Index of Broberg and Morrey (FRIBM) and the Modified Andrews’ Elbow Scoring System (MAESS) were assessed pre-operatively and at each post-operative follow-up visit (six, 12 months and biennially thereafter). Radiographs were analysed for loosening, fractures and dislocation. The mean follow-up was 59 months (26 to 89).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 37 - 37
1 May 2016
Berahmani S Janssen D Wolfson D Hendriks M Wright A Malefijt M Verdonschot N
Full Access

To achieve a long-lasting fixation of uncemented femoral knee implants, an adequate primary stability is required. Several factors, including the applied load, bone quality, surgical preparation, and implant characteristics affect the primary fixation. Recently, novel Attune® cementless femoral component has been proposed by DePuy Synthes (Warsaw, IN, USA). We aimed to compare the primary stability of this novel high-flex design against the conventional LCS® under different loading conditions (gait, deep knee bend (DKB), and high-flex loading), while accounting for the effect of bone quality and cut accuracy.

Six pairs of femora were prepared following the normal surgical procedure. Calibrated CT-scans and 3D-optical scans of the bones were obtained to measure bone mineral density (BMD) and bone cut accuracy, respectively. After implantation of the appropriate size implants (Left legs: Attune; right: LCS), a black-and-white speckle pattern was applied to each specimen (Fig.1B). The micromotion measurement was repeated three times in nine regions of interest (ROIs): the medial and lateral condyles from the posterior view; anterior, distal, and posterior regions from the medial and lateral views; the proximal tip of the anterior flange. The reconstructions were subjected to a gait load and a portion (around 50%) of the peak force of a DKB to prevent fracture of the proximal femur (Fig. 1A and Table. 1). The loads were derived from the Orthoload database using implant-specific inverse dynamics [1]. In addition, a sequence of DIC-images synchronized with the applied load was captured to find the relationship between micromotion and load. Afterwards, implants were pushed-off simulating 150° of flexion, while force-displacement graph was recorded.

BMD and bone cut accuracy were not significantly different between the groups. Under both loading conditions, Attune had a significantly lower micromotion (Table. 1). Cut accuracy was not a significant factor, and BMD was only significant for the comparison under the gait loading (not under DKB conditions). High-flex push-off force was not significantly different. However, Attune required a significantly higher load to reach a micromotion of 50 or 150 µm during the push-off test. Different relations between micromotion and applied load, depending on the loading configuration and implant design, were found (Fig. 2).

Our study has shown a clearly lower range of micromotion for the novel implant. Potential factors to explain the higher micromotion of LCS are parallel anterior and posterior bone cuts in the LCS versus the tapered bone cuts of the Attune. In addition, LCS has a less surface area in contact with bone due to the presence of a rim at the borders of the implant, which may have resulted in lower pre-stresses at the bone-implant interface.

Taking to account, the promising clinical outcome of LCS and also the lower range of micromotion of Attune, we suggest that the Attune has a potential to be at least as successful as the LCS system from a bone fixation point of view. However, further clinical evaluation of the Attune is necessary to assess its performance on the longer term.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 26 - 26
1 May 2016
Hanzlik J Day J Kurtz S Verdonschot N Janssen D
Full Access

Introduction

Initial large-scale clinical studies of porous tantalum implants have been generally promising with well-fixed implants and few cases of loosening [1–3]. An initial retrieval study suggests increased bone ingrowth in a modular tibial tray design compared to the monoblock design [4]. Since micromotion at the bone-implant interface is known to influence bone ingrowth [5], the goal of this study was to determine the effect of implant design, bone quality and activity type on micromotion at the bone-implant interface, through FE modeling.

Patients & Methods

Our case-specific FE model of bone was created from CT data (68 year-old female, right tibia, Fig-1). Isotropic properties of cortical and trabecular bone were derived from the calibrated CT data. Modular and monoblock porous tantalum tibial implants were virtually placed in the tibia following surgical guidelines. All models parts were 3D meshed with 4-noded tetrahedral elements (MSC.MARC-Mentat 2013, MSC Software Corporation, USA). Frictional contact was applied to the bone-tantalum interface (µ=0.88) and UHWMPE-Femoral condyle interface (µ=0.05) with all other interfaces bonded. Loading was applied to simulate walking, standing up and descending stairs. For each activity, a full load cycle [6] was applied to the femoral condyles in incremental steps. The direction and magnitude of micromotions were calculated by tracking the motions of nodes of the bone, projected onto the tibial tray. Micromotions were calculated parallel to the implant surface (shear), and perpendicularly (tensile). We report the maximum (resultant) micromotion that occurred during a cycle of each activity. The bone properties were varied to represent a range in BMD (−30%BMD, Norm, +30%BMD). We compared design type, bone quality and activity type considering micromotion below 40 µm to be favorable for bone ingrowth [5].


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 44 - 44
1 May 2016
Bitter T Janssen D Schreurs B Marriott T Khan I Verdonschot N
Full Access

Introduction

Fretting corrosion at the taper interface has been implicated as a possible cause of implant failure. Using in-vitro testing, fretting wear observed at tapers of retrieved implants may be reproduced (Marriott, EORS-2014). In order to reduce time and cost associated with experimental testing, a validated finite element method (FE) can be employed to study the mechanics at the taper. In this study we compared experimental and representative FE simulations of an accelerated fretting test set-up. Comparison was made by between the FE wear score and volumetric material loss from the testing.

Methods

Experimental test set-up: An accelerated wear test was developed that consistently reproduced fretting wear features observed in retrievals. Biomet stems with smooth 4° Type-1 tapers were combined with Ti6Al4V Magnum +9 mm adaptors using a 2 or 15 kN assembly force. The head was replaced with a custom head fixture to increase the offset and apply a torque at the taper interface. The stems were potted according to ISO 7206-6:2013. The set-up was submerged in a test medium containing PBS and 90gl-1 NaCl. The solution was pH adjusted to 3 using HCl and maintained at 37°C throughout the tests. For each assembly case, n=3 tests were cyclically loaded between 0.4–4 kN for 10 Million cycles. Volumetric wear measurements were performed using a Talyrond-365 roundness measurement machine. The FE model was created to replicate the experimental set up. Geometries and experimental material data were obtained from the manufacturer (Biomet). The same assembly forces of 2 and 15 kN were applied, and the same head fixture was used for similar offset and loading conditions. The 4 kN load was applied at the same angles in accordance with ISO 7206-6:2013. Micromotions and contact pressures were calculated, and based on these a wear score was determined by summation over all contact points.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 78 - 78
1 May 2016
Tomaszewski P Eijkenboom J Berahmani S Janssen D Verdonschot N
Full Access

INTRODUCTION

Total hip arthroplasty (THA) is a very successful orthopaedic treatment with 15 years implant survival reaching 95%, but decreasing age and increasing life expectancy of THA patients ask for much longer lasting solutions. Shorter and more flexible cementless stems are of high interest as these allow to maintain maximum bone stock and reduce adverse long-term bone remodeling.1 However, decreasing stem length and reducing implant stiffness might compromise the initial stability by excessively increasing interfacial stresses. In general, a good balance between implant stability and reduced stress shielding must be provided to obtain durable THA reconstruction.2

This finite element (FE) study aimed to evaluate primary stability and bone remodeling of a new design of short hip implant with solid and U-shaped cross-section.

MATERIALS AND METHODS

The long tapered Quadra-H stem and the short SMS implants (Medacta International, Castel San Pietro, Switzerland) were compared in this study (Figure 1). A FE model of a femur was based on calibrated CT data of an 81 year-old male (osteopenic bone quality). Both titanium alloy implants were assigned an elastic modulus of 105 GPa and the Poisson's ratios were set to 0.3. Initial stability simulations included the hip joint force and all muscle loads during a full cycle of normal walking as calculated in AnyBody software (Anybody Technology AS, Denmark), whereas the remodeling simulation used the peak loads from normal walking and stair climbing activities. Initial stability results are presented as micromotions on the implant surface with a threshold of 40 µm.3 Bone remodeling outcomes are represented in a form of simulated Dual X-ray Absorptiometry (DEXA) scans and the quantitative bone mineral density (BMD) changes in 7 periprosthetic zones.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 106 - 106
1 May 2016
de Ruiter L Janssen D Briscoe A Verdonschot N
Full Access

Introduction

A previous computational study on an all-polymer PEEK-on-UHMWPE total knee replacement implant showed improved periprosthetic bone loading, compared to a conventional implant [1]. That study used a simulated gait cycle to determine distal loading, but a patella was not included. Substantial distal decrease of bone remodeling stimulus was found, in accordance with previous reports [2], but it was not consistent with other clinical and post-mortem DEXA results, which found the largest loss of bone stock in the anterior region [3,4]. As patellofemoral forces are relatively low during gait compared to squatting, we simulated a deep squat, expecting that a high-demand activity would provide similar indications of bone loss as literature [3,4]. Consequently, we applied both high tibiofemoral and patellofemoral loads, to provide more insight in the potential benefits of a new PEEK-Optima® femoral component on periprosthetic bone stock.

Methods

We adopted a deep squat finite element model from Zelle et al. and included quasi-static deep flexion and load sharing at the posterior condyles [6]. A new implant design was inserted, with three variations in material properties: intact, CoCr and PEEK. The stiffness of the femoral elements was mapped from CT and applied to either the cut femur only (CoCr and PEEK) or the entire femoral construct (intact). The strain energy density (SED) was evaluated in the periprosthetic region as a measure for bone remodeling stimulus. To examine the effects of the entire exercise, SED values were integrated over all increments.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 99 - 99
1 May 2016
van de Groes S Kreemers-Van De Hei K Koeter S Verdonschot N
Full Access

Introduction

Special high-flexion prosthetic designs show a small increase in postoperative flexion compared to standard designs and some papers show increased anterior knee pain with these prosthesis. However, no randomised controlled trails have been published which investigate difference in postoperative complaints of anterior knee pain. To assess difference in passive and active postoperative flexion and anterior knee pain we performed a randomized clinical trial including the two extremes of knee arthroplasty designs, being a high flex posterior stabilized rotating platform prosthesis versus a traditional cruciate retaining fixed bearing prosthesis. We hypothesised that the HF-PS design would allow more flexion, due to increased femoral rollback with less anterior knee pain than the CR design. We specifically assessed the following hypotheses:

Patients have increased flexion after HF-PS TKA compared to CR TKA, both passive and active.

Patients show an increased femoral rollback in the HF-PS TKA as compared to the CR TKA.

Patients receiving a HF-PS TKA design report reduced anterior knee pain relative to those receiving the CR TKA.

Methods

In total 47 patients were randomly allocated to a standard cruciate retaining fixed bearing design (CR) in 23 patients and to a high-flexion posterior stabilized mobile bearing design (HF-PS) in 24 patients. Preoperative and one year postoperative we investigated active and passive maximal flexion. Furthermore, we used the VAS pain score at rest and during exercise and the Feller score to investigate anterior knee pain. A lateral roentgen photograph was used to measure femoral rollback during maximal flexion.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 43 - 43
1 Jan 2016
Berahmani S Janssen D Wolfson D De Waal Malefijt M Verdonschot N
Full Access

A durable biological fixation between implant and bone depends largely on the micro-motions [Pilliar et al., 1986]. Finite element analysis (FEA) is a numerical tool to calculate micro-motions during physiological loading. However, micromotions can be simulated and calculated in various ways. Generally, only a single peak force of an activity is applied, but it is also possible to apply discretized loads occurring during a continuous activity, offering the opportunity to analyze incremental micro-motions as well. Moreover, micro-motions are affected by the initial press-fit. We therefore aimed to evaluate the effect of different loading conditions and calculation methods on the micro-motions of an uncemented femoral knee component, while varying the interference-fit.

We created an FE model of a distal femur based on calibrated CT-scans. A Sigma® Cruciate-Retaining Porocoat® (DePuy Synthes, Leeds, UK) was placed following the surgical instructions. A range of interference-fits (0–100 µm) was applied, while other contact parameters were kept unchanged. Micro-motions were calculated by tracking the projection of implant nodes onto the bone surface. We defined three different micro-motions measures: micro-motions between consecutive increments of a full loading cycle (incremental), micro-motions for each increment relative to the initial position (reference), and the largest distance between projected displacements, occurring during a discretized full cycle (resulting) (Fig. 1A). Four consecutive cycles of normal gait and squat movements were applied, in different configurations. In the first configuration, incremental tibiofemoral and patellofemoral contact forces were applied, which were derived from Orthoload database using inverse dynamics [Fitzpatrick et al., 2012]. Secondly, we applied the same loads without the patellofemoral force, which is often used in experimental set-ups. Finally, only the peak tibiofemoral force was applied, as a single loading instance. We calculated the average of micro-motions of all nodes per increment to compare different calculation techniques. The percentage of area with resulting micro-motions less than 5 µm was also calculated.

The percentage of surface area was increased non-linearly when the interference fit changed from 0 to 100 µm particularly for squat movement. Tracking nodes over multiple cycles showed implant migration with interference-fits lower than 30µm (Fig. 1A). Loading configurations without the patellofemoral force, and with only the peak tibiofemoral force slightly overestimated and underestimated the resulting micro-motions of squat movement, respectively; although, the effect was less obvious for the gait simulation when no patella force was applied. Both incremental and reference micro-motions underestimated the resulting micro-motions (Fig. 1B). Interestingly, the reference micro-motions followed the pattern of the tibiofemoral contact force (Fig. 1B).

The calculation technique has a substantial effect on the micro-motions, which means there is a room for interpretation of micro-motions analyses. This furthermore stresses the importance of validation of the predicted micro-motions against experimental set-ups. In addition, the minor effect of loading configurations indicates that a simplified loading condition using only the peak tibiofemoral force is suitable for experimental studies. From a clinical perspective, the migration pattern of femoral components implanted with a low interference fit stresses the role of an adequate surgical technique, to obtain a good initial stability.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 140 - 140
1 Jan 2016
de Ruiter L Janssen D Briscoe A Verdonschot N
Full Access

Introduction

Conventional implant designs in total knee arthroplasty (TKA) are based on metal on UHMWPE bearing couples. Although this procedure is quite successful, early loosening is still a matter of concern. One of the causes for early failure is stress shielding, leading to loss of bone stock, periprosthetic bone fractures and eventually aseptic loosening of the component. The introduction of a polyetheretherketone (PEEK) on UHMWPE bearing couple could address this problem. With mechanical properties more similar to distal (cortical) bone it could allow stresses to be distributed more naturally in the distal femur. A potential adverse effect, however, is that the femoral component and the underlying cement mantle may be at risk of fracturing. Therefore, we analyzed the effect of a PEEK-Optima® femoral component on stress shielding and the integrity of the component and cement mantle, compared to a conventional Cobalt-Chromium (CoCr) alloy implant.

Methods

We created a Finite Element (FE) model of a reconstructed knee in gait, based on the ISO-14243-1 standard. The model consisted of an existing cemented cruciate retaining TKA design implanted on a distal femur, and a tibial load applicator, which together with the bone cement layer and the tibial implant is referred to as the tibial construct. The knee flexion angle was controlled by the femoral construct, consisting of the femoral implant, the bone cement and the distal femur. The tibial construct was loaded with an axial force, anterior-posterior (AP) force and a rotational torque, representing the ground reaction force, soft tissue constraints and internal/external rotation of the tibia, respectively. The integrity of the femoral component and cement mantle were expressed as a percentage of their yield stress. Stress shielding in the periprosthetic femur was evaluated by the strain energy (density) in the bone and compared to a model replicating an intact knee joint.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 97 - 97
1 Jan 2016
Verdonschot N Weerdesteyn V Vigneron L Damsgaard M Sitnik R Feikas T Carbone V Koopman B
Full Access

INTRODUCTION

The burden of Musculoskeletal (M-S) diseases and prosthetic revision operations is huge and increasing rapidly with the aging population. For patients that require a major surgical intervention, procedures are unsafe, uncertain in outcome and have a high complication rate. The goal of this project is to create an ICT-based patient-specific surgical navigation system that helps the surgeon safely reaching the optimal functional result for the patient and is a user friendly training facility for the surgeons. The purpose of this paper is to demonstrate the advancements in personalized musculoskeletal modeling for patients who require severe reconstructive surgery of the lower extremity.

METHODS

TLEMsafe is a European Project dedicated to generating semi-automated 3-D image-analyzing tools to simulate the musculoskeletal (M-S) system. The patient-specific parameters are fed into models with which the patient specific functional outcome can be predicted. Hence, we can analyze the functional effect e.g. due to placement of prosthetic components in a patient. Surgeons can virtually operate on the patient-specific model after which the model predicts the functional effects. Once the optimal plan is selected, this is fed into a computer navigation system (see figure 1).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 48 - 48
1 Jan 2016
Bitter T Janssen D Schreurs BW Marriott T Khan I Verdonschot N
Full Access

Introduction

Recent reports implicate fretting corrosion at the head-stem taper junction as a potential cause of failure of some large diameter metal-on-metal (MOM) devices. Fretting observed at modular junctions is thought to be a type of ‘mechanically assisted’ corrosion phenomenon, initiated by mechanical factors that lead to an increase in contact stresses and micromotions at the taper interface. These may include: intra-operative taper assembly, taper contamination by debris or body fluids, patient weight and ‘toggling’ of the head or increased frictional torque in a poorly functioning bearing.

We adopted a finite element approach to model the head-taper junction, to analyze the contact mechanics at the taper interface. We investigated the effect of assembly force and angle on contact pressures and micromotions, during loads commonly used to test hip implants.

Materials and methods

Models of the Biomet Type-1 taper, a 60 mm head and a taper adaptor were created. These models were meshed with a mesh size based on a mesh density convergence study. Internal mesh coarsening was applied to reduce computational cost.

Elastic-plastic material properties based on tensile tests were assigned to all titanium components. The contact conditions used in the FE analyses were validated against push-on and pull-off experiments, resulting in a coefficient of friction of 0.5.

To analyze micromotions at the taper-adaptor connection, the models were loaded with 2300N (ISO 7206-4) and 5340N (ISO 7206-6), after being assembled with 2-4-15 kN, axially and under a 30º angle. This ISO standard is commonly used to determine endurance properties of stemmed femoral components.

Micromotions and contact pressures were analyzed by scoring them to an average micromotion and average contact pressure for the surface area in contact.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 44 - 44
1 Jan 2016
Berahmani S Janssen D Wolfson D De Waal Malefijt M Verdonschot N
Full Access

Femoral knee implants have promising outcomes, although some high-flex designs have shown rather high loosening rates (Han et al., 2007). In uncemented implants, it is vital to limit micromotions at the implant-bone interface, to facilitate secondary fixation through bone ingrowth (kienapfel et al., 1999). Hence, it is essential to investigate how micromotions of different uncemented implants are affected by various loading conditions when a range of bone qualities as a patient-related factor is applied.

Using finite element (FE) analysis, we simulated implant-bone interface micromotions during four consecutive cycles of normal gait and squat movements. An FE model of a distal femur was generated based on calibrated CT-scans, after which Sigma® and LCS® Cruciate-Retaining Porocoat® components (DePuy Synthes, Leeds, UK) were implanted. Using a frictional contact algorithm (µ=0.95), an initial press-fit fixation was simulated, which was previously validated against experimental data. The micromotions were calculated by tracking the projection of implant nodes on the bone surface excluding overhang area. The applied loading patterns were based on discretized simulations, providing incremental loads for each activity based on implant-specific kinematics, which was derived from Orthoload database using inverse dynamics (Fitzpatrick et al., 2012). This provided the opportunity to calculate incremental micromotions, but also the resulting micromotions for each single cycle, for both activities. In addition, the percentage of implant surface area with resulting micromotions less than a defined threshold was calculated.

Regardless of the type of loading, in all simulations, the predicted micromotions were highest in the first cycle, suggesting settling of the implant during initial cycle. The Sigma®implant displayed a 30% larger area with micromotions below the threshold of 5 microns, for both loading conditions (Fig. 1A). The highest micromotions occurred at the anterior flange, regardless of type of activity or design. Squatting had a more detrimental effect on the primary stability, with smaller areas of low micromotions as compared to the gait load (Fig. 1B). Bone stiffness had a minor effect, which was more apparent for squatting (Fig. 1B).

We found acceptable low ranges of micromotions in both implant designs, although demanding activities such as squatting generated higher motions. In addition, LCS® experienced higher micromotions, probably caused by the smaller contact area at bone-implant interface compared with Sigma®. Nevertheless, the predicted micromotions were all below the clinically relevant threshold for bone ingrowth (<40 microns) (kienapfel et al., 1999). Furthermore, our simulated settling behavior stresses the necessity for simulating multiple loading cycles, rather than just a single cycle. The effect of bone stiffness was evident, but only to a limited extent. The main current limitation of our study is the utilization of an elastic material model for the bone which is probably the reason to predict a low range of micromotions. We are planning to make the material model more realistic, by including plasticity and viscoelastic bone behavior.


The Bone & Joint Journal
Vol. 97-B, Issue 10 | Pages 1338 - 1344
1 Oct 2015
te Stroet MAJ Keurentjes JC Rijnen WHC Gardeniers JWM Verdonschot N Slooff TJJH Schreurs BW

We present the results of 62 consecutive acetabular revisions using impaction bone grafting and a cemented polyethylene acetabular component in 58 patients (13 men and 45 women) after a mean follow-up of 27 years (25 to 30). All patients were prospectively followed. The mean age at revision was 59.2 years (23 to 82).

We performed Kaplan–Meier (KM) analysis and also a Competing Risk (CR) analysis because with long-term follow-up, the presence of a competing event (i.e. death) prevents the occurrence of the endpoint of re-revision.

A total of 48 patients (52 hips) had died or had been re-revised at final review in March 2011. None of the deaths were related to the surgery. The mean Harris hip score of the ten surviving hips in ten patients was 76 points (45 to 99).

The KM survivorship at 25 years for the endpoint ‘re-revision for any reason’ was 58.0% (95% confidence interval (CI) 38 to 73) and for ‘re-revision for aseptic loosening’ 72.1% (95% CI 51 to 85). With the CR analysis we calculated the KM analysis overestimates the failure rate with respectively 74% and 93% for these endpoints. The current study shows that acetabular impaction bone grafting revisions provide good clinical results at over 25 years.

Cite this article: Bone Joint J 2015;97-B:1338–44.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 681 - 688
1 May 2015
Wagener ML de Vos MJ Hannink G van der Pluijm M Verdonschot N Eygendaal D

Unlinked, linked and convertible total elbow arthroplasties (TEAs) are currently available. This study is the first to report the clinical results of the convertible Latitude TEA. This was a retrospective study of a consecutive cohort of 63 patients (69 primary TEAs) with a mean age of 60 years (23 to 87). Between 2006 and 2008 a total of 19 men and 50 women underwent surgery. The mean follow-up was 43 months (8 to 84). The range of movement, function and pain all improved six months post-operatively and either continued to improve slightly or reached a plateau thereafter. The complication rate is similar to that reported for other TEA systems. No loosening was seen. Remarkable is the disengagement of the radial head component in 13 TEAs (31%) with a radial head component implanted.

Implantation of both the linked and the unlinked versions of the Latitude TEA results in improvement of function and decreased pain, and shows high patient satisfaction at mid-term follow-up.

Cite this article: Bone Joint J 2015; 97-B:681–8.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 133 - 133
1 Jul 2014
O'Kane C Vrancken A O'Rourke D Janssen D Ploegmakers M Buma P Fitzpatrick D Verdonschot N
Full Access

Summary

Our statistical shape analysis showed that size is the primary geometrical variation factor in the medial meniscus. Shape variations are primarily focused in the posterior horn, suggesting that these variations could influence cartilage contact pressures.

Introduction

Variations in meniscal geometry are known to influence stresses and strains inside the meniscus and the articulating cartilage surfaces. This geometry-dependent functioning emphasizes that understanding the natural variation in meniscus geometry is essential for a correct selection of allograft menisci and even more crucial for the definition of different sizes for synthetic meniscal implants. Moreover, the design of such implants requires a description of 3D meniscus geometry. Therefore, the aim of this study was to quantify 3D meniscus geometry and to determine whether variation in medial meniscus geometry is size or shape driven.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 229 - 236
1 Feb 2014
deVos MJ Verdonschot N Luites JWH Anderson PG Eygendaal D

We determined the short-term clinical outcome and migration within the bone of the humeral cementless component of the Instrumented Bone Preserving (IBP) total elbow replacement in a series of 16 patients. There were four men and 12 women with a mean age at operation of 63 years (40 to 81). Migration was calculated using radiostereometric analysis. There were no intra-operative complications and no revisions. At two-year follow-up, all patients showed a significant reduction in pain and functional improvement of the elbow (both p < 0.001). Although ten components (63%) showed movement or micromovement during the first six weeks, 14 (88%) were stable at one year post-operatively. Translation was primarily found in the proximal direction (median 0.3 mm (interquartile range (IQR) -0.09 to 0.8); the major rotational movement was an anterior tilt (median 0.7° (IQR 0.4° to 1.6°)). One malaligned component continued to migrate during the second year, and one component could not be followed beyond three months because migration had caused the markers to break off the prosthesis.

This study shows promising early results for the cementless humeral component of the IBP total elbow replacement. All patients had a good clinical outcome, and most components stabilised within six months of the operation.

Cite this article: Bone Joint J 2014;96-B:229–36.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 238 - 238
1 Dec 2013
Berahmani S Janssen D Wolfson D De Waal Malefijt M Verdonschot N
Full Access

Pegs are often used in cementless total knee replacement (TKR) to improve fixation strength. Studies have demonstrated that interference fit, surface properties, bone mineral density (BMD) and viscoelasticity affect the performance of press-fit designs. These parameters also affect the insertion force and the bone damage occurring during insertion. We aimed to quantify the effect of the aforementioned parameters on the short-term fixation strength of cementless pegs.

6 mm holes were drilled in twenty-four human femora. BMD was measured using calibrated CT-scans, and randomly assigned to samples. Pegs were produced to investigate the effect of interference fit (diameters 6.5 and 7.6 mm), surface treatment (smooth and rough- porous-coating [friction coefficient: 1.4]) and bone relaxation (relaxation time 0 and 30 min) and interactions were studied using a DOE method. Two additional rough surfaced peg designs (diameters 6.2 and 7.3 mm) were included to scrutinize interference. Further, a peg based on the LCS Porocoat® (DePuy Synthes Joint Reconstruction, Leeds, UK) was added as a clinical baseline. In total seven designs were used (n = 10 for all groups). Pegs were inserted and extracted using an MTS machine (Figure 1), while recording force and displacement. Bone damage was defined as the difference between the cross-sectional hole area prior to and after the test.

BMD and interference fit were significant factors for insertion force. BMD had a significant positive correlation with pull-out force and subsequent analyses were therefore normalised for BMD.

Pull-out force increased significantly with interference for both surface coatings at time 0 (p < 0.05). However, after 30 minutes the effect remained significant for rough pegs only (p < 0.05-Figure 2A).

Pull-out force reduced significantly with roughness for both peg diameters at time 0 (p < 0.001). However, after 30 minutes the effect remained significant for small pegs only (p < 0.05-Figure 2A).

The time dependant interaction was only significant for smooth pegs in both diameters (p < 0.05-Figure 2A).

Additionally, the pull-out force increased with diameter in a non-linear manner for the rough pegs (Figure 2B). The two surface treatments were not significantly different to the clinical comparator. Interference fit was the only significant factor for bone damage.

BMD was significant for insertion and pull-out forces, reinforcing the need to account for this factor in biomechanical studies and clinical practice. This study also highlights the importance of time in studying bone interactions, with surface treatment and interference showing different interaction effects with relaxation time. Although smooth pegs initially have a higher pull-out force, this effect reduces over time whereas the pullout force for rough pegs is maintained. Smooth pegs also show time sensitivity in relation to interference and the benefit of increased interference reduces over time, whereas it is maintained in rough pegs. This may be explained by different damage (compressive and abrasive) mechanisms associated with different surface treatments.

In conclusion, BMD and interference fit are significant factors for initial fixation. Bone relaxation plays an important role as it reduces the initial differences between groups. Therefore, these findings should be strongly considered in the design development of cementless TKR.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 15 - 15
1 Dec 2013
Berahmani S Janssen D Wolfson D De Waal Malefijt M Verdonschot N
Full Access

The effect of an advanced porous surface morphology on the mechanical performance of an uncemented femoral knee prosthesis was investigated. Eighteen implants were inserted and then pushed-off from nine paired femurs (Left legs: advanced surface coating; right legs: Porocoat® surface coating as baseline). Bone mineral density (BMD) and anteroposterior dimension were measured, which both were not significantly different between groups. The insertion force was not significantly different, but push-off force was significantly higher in the advanced surface coating group (P = 0.007). BMD had direct relationship with the insertion force and push-off force (p < 0.001). The effect of surface morphology on implant alignment was very small. We suggest that the surface properties create a higher frictional resistance thereby providing a better inherent stability of implants featuring the advanced surface coating.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 241 - 241
1 Dec 2013
Bitter T Janssen D Schreurs BW Khan I Verdonschot N
Full Access

Introduction

Recent reports have implicated fretting corrosion at the head-stem taper junction as a potential cause of failure of some large diameter metal-on-metal (MOM) devices. While it has been suggested that larger MOM heads, may induce greater frictional torques at the taper connection, the exact mechanisms underlying fretting corrosion remain poorly understood.

It is likely that the onset of the corrosion process is caused by mechanical factors, such as contact stresses and micromotions occurring at the interface. These stresses and micromotions depend on the fixation of the head onto the stem and may be affected by blood, fat, bone debris or other contaminations. The fixation of the head is achieved intraoperatively through impaction.

To further study this phenomenon, we adopted a finite element approach in which we modeled the head-taper junction fixation mechanics. In this model, we analyzed the effect of impaction force on the micromotions occurring at the head-stem interface.

Materials and methods

We created a model of a BIOMET Type-1 taper and an adapter that is typically used for larger heads.

Titanium alloy material properties were assigned to both components, and frictional contact (μ = 0.5) was simulated between the adapter and the taper.

To ensure that the model accurately represented the contact mechanics, we first simulated experiments in which the head was assembled on the taper in a load-controlled manner, at different load (4 and 15 kN), after which it was disassembled axially. The disassembly loads predicted by the FEA simulations were then compared to the experimental values.

After ensuring a correct prediction of the disassembly loads, we used various impaction loads (2, 4, and 15 kN) to assemble the taper, after which a 2.3 kN load (ISO 7206-4) was applied to the adapter/taper assembly. This loading regime is commonly used to determine endurance properties of stemmed femoral components. Under these loading conditions, we then analyzed the contact stresses and micromotions, and the effect of impaction load on these quantities.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 239 - 239
1 Dec 2013
Berahmani S Janssen D Wolfson D De Waal Malefijt M Verdonschot N
Full Access

To achieve desirable outcomes in cementless total knee replacement (TKR), sufficient primary stability is essential. The primary stability inhibits excessive motions at the bone-implant interface, hence providing the necessary condition for osseointegration [1]. Primary stability for cementless TKR is provided by press-fit forces between the bone and implant. The press-fit forces depend on several factors including interference fit, friction between bone and implant surface, and the bone material properties. It is expected that bone mineral density (BMD) will affect the stability of cementless TKR [2]. However, the effect of BMD on the primary stability of cementless femoral knee component has not been investigated in vitro.

Phantom calibrated CT-scans of 9 distal femora were obtained after the surgical cuts were made by an experienced surgeon. Since the press-fit forces of the femoral component mainly occur in the Anteroposterior (AP) direction, the BMD was measured in the anterior and posterior faces for a depth of 5 mm; this depth was based on stress distributions from a Finite Element Analysis of the same implant design. In addition, four strain gauges were connected to different locations on the implant's outer surface and implant strain measured throughout as an indication of underlying bone strain. A cementless Sigma CR femoral component (DePuy Synthes Joint Reconstruction, Leeds, UK) was then implanted using an MTS machine. In order to simulate a ‘normal’ bone condition, the implanted bone was preconditioned for one hour at a cyclic load of 250–1500 N, and a rate of 1 Hz. Finally, the implants were pushed-off from the bone in a high-flex position. Forces and displacements were recorded both during insertion and push-off tests.

Strong correlations were found for insertion and push-off forces with BMD, R2 = 0.88 and R2 = 0.88, respectively (p < 0.001), so although implantation may be harder in patients with higher BMD, initial stability is also improved. A correlation was also found between final strain and push-off forces (R2 = 0.89, p < 0.01) and BMD also showed a strong reverse correlation with total bone relaxation (R2 = 0.76, p = 0.023). These results indicate that higher BMD induces higher bone strain, which can lead to improved fixation strength.

There is no consensus on the best fixation method for the TKR but some surgeons prefer a cementless design for young and active patients. The results of our study showed that the primary stability of a cementless femoral knee component is directly correlated with the bone mineral density. Therefore, patient selection based on bone quality may increase the likelihood of good osseointegration and adequate long-term fixation for cementless femoral knee components.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 513 - 513
1 Dec 2013
Ruiter L Janssen D Briscoe A Verdonschot N
Full Access

Introduction

Current clinical practice in total knee arthroplasty (TKA) is largely based on metal on polyethylene bearing couples. A potential adverse effect of the stiff metal femoral component is stress shielding, leading to loss of bone stock, periprosthetic bone fractures and eventually aseptic loosening of the component. The use of a polymer femoral component may address this problem. However, a more flexible material may also have consequences for the fixation of the femoral component. Concerns are raised about its expected potential to introduce local stress peaks on the interface.

The objective of this study was to analyze the effect of using a polyether-etherketone (PEEK-Optima®) femoral component on the cement-implant interface. We analyzed the interface stress distribution occurring during normal gait, and compared this to results of a standard CoCr component.

Materials and methods

An FEA model was created, consisting of a femoral component cemented onto a femur, and a polyethylene tibial component. A standard loading regime was applied mimicking an adapted gait cycle, according to ISO14243-1. The implant-cement interface was modelled as a zero-thickness layer connecting the implant to the cement layer. Femoral flexion/extension was prescribed for the femur in a displacement controlled manner, while the joint loads were applied to pivoting nodes attached to the tibial construct, consistent with the ISO standard. Implant-cement interface properties were adopted from a previous study on CoCr interface debonding[1].


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 358 - 358
1 Mar 2013
Verdonschot N Van Der Ploeg B Tarala M Homminga J Janssen D
Full Access

Introduction

Many finite element (FE) studies have been performed in the past to assess the biomechanical performance of TKA and THA components. The boundary conditions have often been simplified to a few peak loads. With the availability of personalized musculoskeletal (MS) models we becomes possible to estimate dynamic muscle and prosthetic forces in a patient specific manner. By combining this knowledge with FE models, truly patient specific failure analyses can be performed.

In this study we applied this combined technique to the femoral part of a cementless THR and calculated the cyclic micro-motions of the stem relative to the bone in order to assess the potential for bone ingrowth.

Methods

An FE model of a complete femur with a CLS Spotorno stem inserted was generated. An ideal fit between the implant and the bone was modeled proximally, whereas distally an interface gap of 100μm was created to simulate a more realistic interface condition obtained during surgery. Furthermore, a gait analysis was performed on a young subject and fed into the Anybody™ MS modeling system. The anatomical data set (muscle attachment points) used by the Anybody™ system was morphed to the shape of the femoral reconstruction. In this way a set of muscle attachment points was obtained which was consistent with the FE model. The predicted muscle and hip contact forces by the Anybody™ modeling system were dynamic and divided into 37 increments including two stance phases and a swing phase of the right leg.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 357 - 357
1 Mar 2013
van de Groes S De Waal Malefijt M Verdonschot N
Full Access

Introduction

A few follow-up studies of high flexion total knee arthoplasties report disturbingly high incidences of femoral loosening. Finite element analysis showed a high risk for early loosening at the cement-implant interface at the anterior flange. However, femoral implant fixation is depending on two interfaces: cement-implant interface and the cement-bone interface. Due to the geometry of the distal femur, a part of the cement-bone interface consists of cement-cortical bone interface. The strength of the cement-bone interface is lower than the strength of the cement-implant interface.

The research questions addressed in this study were: 1) which interface is more prone to loosening and 2) what is the effect of different surgical preparation techniques on the risk for early loosening.

Materials & methods

To achieve data for the cement-(cortical)bone interface strength and the effects of different preparation techniques on interfacial strength, human cadaver interface stress tests were performed for different preparation techniques of the bony surface and the results were implemented in a finite element (FE) model as described before. The FE model consisted of a proximal tibia and fibula, TKA components, a quadriceps and patella tendon and a non-resurfaced patella. For use in this study, the distal femur was integrated in the FE model including cohesive interface elements and a 1 mm bone cement layer. In the model, the cement-bone interface was divided into two areas, representing cortical and cancellous bone. The posterior-stabilised PFC Sigma RP-F (DePuy, J&J, USA) was incorporated in the FE knee model following the surgical procedure provided by the manufacturer. A full weight-bearing squatting cycle was simulated (ROM = 50°-155°). The interface failure index was calculated.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 25 - 25
1 Sep 2012
Biemond L Hannink G Verdonschot N Buma P
Full Access

INTRODUCTION

Electron beam melting is a promising technique to produce surface structures for cementless implants. Biomimetic apatite coatings can be used to enhance bone ingrowth. The goal of this study was to evaluate bone ingrowth of an E-beam produced structure with biomimetic coating and compare this to an uncoated structure and a conventionally made implant surface.

METHODS


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 199 - 199
1 Sep 2012
van de Groes S Ypma J Spierings P Verdonschot N
Full Access

In the present study we describe the clinical results of the Scientific Hip Prosthesis® (SHP). With the goal of smoothening cement-bone interface stress peaks, the SHP was developed using shape optimization algorithms together with finite element modelling techniques. The resulting shape and cement stresses are seen in Figure 1. The introduction of the SHP prosthesis was performed in a stepwise fashion including a RSA study performed by Nivbrant et al1. RSA studies for prosthetic types that are in long-term use are of great value in predicting the survivorship related to the migration rate and pattern for that specific type of prosthesis. If a stem in a patient shows a much higher migration rate than the typical one, the stem may be identified as at high-risk for early loosening. The study of Nivbrant et al1 revealed unexpectedly high migration values and it was stated that the SHP stem was not the preferred stem to use despite the good Harris Hip Score and Pain score at two years follow-up.

In the present study the clinical results of a single surgeon study consisting of 171 hips with a follow-up of 5–12 years were evaluated. The mean follow-up was 8.2 years (5.0–12.0). The survival rate was 98.8% at ten years follow-up for aseptic loosening of the stem. The mean Harris Hip Score at 10 year follow-up was 89.2 ± 7.5. This study therefore indicates that a new prosthetic design may function clinically rather well, despite the relatively high migration rates which have been reported.

In case of a RSA study with a new prosthesis it may not be so evident what the expected “typical” migration rate or pattern is. So in order to predict early loosening the typical migration rate has to be known. Perhaps typical migration rates can be established using standardized cadaver migration experiments or computer simulation models techniques. Since these standardized tools are currently not available, the prediction of clinical survival of new prosthetic components remains a challenging task and the interpretation of migration rates with new designs should be considered with much caution.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 211 - 211
1 Sep 2012
Waanders D Janssen D Mann K Verdonschot N
Full Access

In cemented total hip arthroplasty, the cement-bone interface can be considerably degraded in less than one year in-vivo service (Figure 1). This makes the interface much weaker relative to the direct post-operative situation. Retrieval studies show that patients do, to a certain extent, not suffer from the degraded cement-bone interface itself. It is, however, unknown whether the degraded cement-bone interface affects other failure mechanisms in the cemented hip reconstruction. A good understanding of the mechanics of the cement-bone interface is therefore essential. The aim of this study was to investigate the mechanics of the cement-bone interface in the direct post-operative and degraded situation by the utilization of finite element analysis (FEA) and laboratory experiments. It was subsequently analyzed how the mechanics of the cement-bone interface affect failure of the cement mantle in terms of crack formation.

In order to investigate the mechanical response of the cement-bone interface, laboratory prepared (direct post-operative state) and postmortem (degraded state) specimens were loaded in various directions in the laboratory and FEA environment. From all specimens, multiple interface morphology parameters were documented, which were related to the interfacial response and subsequently converted to a numerical cohesive model. As a validation, this cohesive model was implemented into two FEA models of transverse sections of cemented hip reconstructions with distinct mechanical characteristics (Figure 2). Finally, the differences in fatigue crack formation in a complete hip reconstruction were determined by varying the cement-bone interface compliance (Figure 3).

When loaded in multiple directions, the interface compliance could not be related to the cement interdigitation depth (r2=0.08). However, compliance did correlate to the gap thickness between the bone and cement (r2=0.81) and the amount of interfacial contact (r2=0.50). Surprisingly, for the same amount of contact, the interface was more compliant in degraded state than in the direct post-operative state. The mechanical response of the experimental and FEA cement-bone interface tests could, independent on the direct post-operative or degraded state, successfully be described by a cohesive model. The cohesive model was even more confirmed by the successful reproduction of the mechanics of the retrieved transverse sections. When the cohesive model was implemented in a complete reconstruction, we found that a compliant cement-bone interface resulted in considerably more fatigue cracks in the cement mantle than a very stiff interface.

This study showed that an increased compliancy of the cement-bone interface results in an increase of cement cracks in the cement mantle. It is therefore crucial to minimize the interfacial gaps and, as a result, increase the amount of contact between the bone and cement to generate a stiff cement-bone interface. It is, unfortunately, unknown how this well fixed interface can be maintained. We finally conclude that the derived cohesive model of the cement-bone interface can be used for multiple applications in orthopaedics, including pre-clinical of implants and patient specific studies of failed cemented reconstructions.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 192 - 192
1 Sep 2012
Tomaszewski P Verdonschot N Bulstra S Verkerke G
Full Access

For amputated patients, direct attachment of upper leg prosthesis to the skeletal system by a percutaneous implant is an alternative solution to the traditional socket fixation. Currently available implants, the OPRA system (Integrum AB, Göteborg, Sweden) and the ISP Endo/Exo prosthesis (ESKA Implants AG, Lübeck, Germany) [1-2] allow overcoming common soft tissue problems of conventional socket fixation and provide better control of the prosthetic limb [3], higher mobility and comfort [2, 4]. However, restraining issues such as soft-tissue infections, peri-prosthetic bone fractures [3, 5–8] and considerable bone loss around the stem [9], which might lead to implant's loosening, are present. Finally, a long a residual limb is required for implant fitting.

In order to overcome the limiting biomechanical issues of the current designs, a new concept of the direct intramedullary fixation was developed. The aim was to restore the natural load transfer in the femur and allow implantations in short femur remnants (Figure 1). We hypothesize that the new design will reduce the peri-prosthetic bone failure risk and adverse bone remodeling.

Generic CT-based finite element models of an intact femoral bone and amputated bones implanted with 3 analyzed implants were created for the study. Models were loaded with two loading cases from a normal walking obtained from the experimental measurements with the OPRA device [10-11]. Periprosthetic bone failure risk was evaluated by considering the von Mises stress criterion [12-14]. Subsequently the strain adaptive bone remodeling theory was used to predict long-term changes in bone mineral density (BMD) around the implants. The bone mineral content (BMC) change was measured around implants and the results were visualized in the form of DXA scans.

The OPRA and the ISP implants induced the high stress concentration in the proximal region decreasing in the distal direction to values below physiological levels as compared with the intact bone. The stresses around the new design were more uniformly distributed along the cortex and resembled better the intact case. Consequently, the bone failure risk was reduced as compared to the OPRA and the ISP implants. The adaptive bone remodeling simulations showed high bone resorption around distal parts of the OPRA and the ISP implants in the distal end of the femur (on average −75% ISP to −78% OPRA after 60 months). The bone remodeling simulation did not reveal any bone loss around the new design, but more bone densification was seen (Figure 2). In terms of total bone mineral content (BMC) the OPRA and the ISP implants induced only a short-term bone densification in contrast to the new design, which provoked a steady increase in the BMC over the whole analyzed period (Figure 3).

In conclusion, we have seen that the new design offers much better bone maintenance and lower failure probability than the current osseointegrated trans-femoral prostheses. This positive outcome should encourage further developments of the presented concept, which in our opinion has a potential to considerably improve safety of the rehabilitation with the direct fixation implants and allow treatment of patients with short stumps.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 74 - 74
1 Sep 2012
Innocenti B van Jonbergen H Labey L Verdonschot N
Full Access

INTRODUCTION

Patellofemoral joint (PFJ) replacement is a successful treatment option for isolated patellofemoral osteoarthritis. With this approach only the involved joint compartment is replaced and the femoro-tibial joint remains intact. Minimizing periprosthetic bone loss, which may occur due to the stress shielding effect of the femoral component, is important to insure long-term outcomes. The objective of this study was to investigate, using finite element analyses, the effects of patellofemoral replacement on the expected stress distribution of the distal femur eventually leading to changes in bone density.

METHODS

MRI images of a healthy knee were acquired, segmented and reconstructed into a 3D physiological model of the bony and cartilaginous geometries of distal femur and patella with patellar tendon and insertion of the quadriceps tendon. This model was modified to include PFJ replacements with either a Journey PFJ or a Richards II PFJ prosthesis, and a Genesis II TKA (Smith&Nephew, Memphis, TN). The prosthetic components were incorporated in the intact model based on the manufacturer's instructions or previously described surgical techniques (Figure 1).

Cortical bone was modeled with orthotropic properties, while homogeneous linear isotropic elasticity was assumed for trabecular bone, cartilage, cement and femoral components materials. The patellar tendon was given Neo-Hookean behavior. UHMWPE patellar buttons for all designs were assigned non-linear elasto-plastic material.

The simulated motion consisted of a 10 second loaded squat, starting from 0° until a flexion angle of 120° matching experimental kinematics tests performed in previous in-vitro analysis on physiological cadaveric legs [1-2]. The patella model was constrained fixing the distal part of the patellar ligament and applying a quadriceps force distributed on the quadriceps insertion on the proximal surface of the patella.

During the dynamic simulation the average Von Mises stress was calculated in two regions of interest (ROI) defined in the femoral bone: one anterior and one proximal. The location of the ROIs was defined to fit the same regions as used in a previous bone mineral density analysis following patellofemoral arthroplasty (height 1cm, length 1cm).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1135 - 1142
1 Aug 2012
Derikx LC van Aken JB Janssen D Snyers A van der Linden YM Verdonschot N Tanck E

Previously, we showed that case-specific non-linear finite element (FE) models are better at predicting the load to failure of metastatic femora than experienced clinicians. In this study we improved our FE modelling and increased the number of femora and characteristics of the lesions. We retested the robustness of the FE predictions and assessed why clinicians have difficulty in estimating the load to failure of metastatic femora. A total of 20 femora with and without artificial metastases were mechanically loaded until failure. These experiments were simulated using case-specific FE models. Six clinicians ranked the femora on load to failure and reported their ranking strategies. The experimental load to failure for intact and metastatic femora was well predicted by the FE models (R2 = 0.90 and R2 = 0.93, respectively). Ranking metastatic femora on load to failure was well performed by the FE models (τ = 0.87), but not by the clinicians (0.11 < τ < 0.42). Both the FE models and the clinicians allowed for the characteristics of the lesions, but only the FE models incorporated the initial bone strength, which is essential for accurately predicting the risk of fracture. Accurate prediction of the risk of fracture should be made possible for clinicians by further developing FE models.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 255 - 255
1 Jun 2012
Zelle J Malefijt MDW Verdonschot N
Full Access

Introduction

High-flexion knee implants have been developed to accommodate a large range of motion (ROM > 120°) after total knee arthroplasty (TKA). In a recent follow-up study, Han et al. [1] reported a disturbingly high incidence of femoral loosening for high-flexion TKA. The femoral component loosened particularly at the implant-cement interface. Highly flexed knee implants may be more sensitive to femoral loosening as the knee load is high during deep knee flexion [2], which may result in increased tensile and/or shear stresses at the femoral implant fixation.

The objective of this study was to analyse the load-transfer mechanism at the femoral implant-cement interface during deep knee flexion (ROM = 155°). For this purpose, a three-dimensional finite element (FE) knee model was developed including high-flexion TKA components. Zero-thickness cohesive elements were used to model the femoral implant-cement interface. The research questions addressed in this study were whether high-flexion leads to an increased tensile and/or shear stress at the femoral implant-cement interface and whether this would lead to an increased risk of femoral loosening.

Materials & methods

The FE knee model utilized in this study has been described previously [3] and consisted of a proximal tibia and fibula, TKA components, a quadriceps and patella tendon and a non-resurfaced patella. For use in this study, the distal femur was integrated in the FE model including cohesive interface elements and a 1 mm bone cement layer. High-flexion TKA components of the posterior-stabilised PFC Sigma RP-F (DePuy, J&J, USA) were incorporated in the FE knee model following the surgical procedure provided by the manufacturer. A full weight-bearing squatting cycle was simulated (ROM = 50°-155°). The interface stresses calculated by the FE knee model were decomposed into tension, compression and shear components. The strength of the femoral implant-cement interface was determined experimentally using interface specimens to predict whether a local interface stress-state calculated by the FE knee model would lead to interface debonding.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 145 - 145
1 Jun 2012
Meijerink H Loon CV Malefijt MDW Kampen AV Verdonschot N
Full Access

Introduction

Within the reconstruction of unicondylar femoral bone defects with morselized bone grafts in revision total knee arthroplasty (TKA), a stem extension appears to be critical to obtain adequate mechanical stability. Whether the stability is still secured by this reconstruction technique in bicondylar defects has not been assessed. Long, rigid stem extensions have been advocated to maximize the stability in revision TKAs. The disadvantage of relatively stiff stem extensions is that bone resorption is promoted due to stress shielding. Therefore, we developed a relatively thin intramedullary stem which allowed for axial sliding movements of the articulating part relative to the intramedullary stem. The hypothesis behind the design is that compressive contact forces are directly transmitted to the distal femoral bone, whereas adequate stability is provided by the sliding intramedullary stem. A prototype was made of this new knee revision design and applied to the reconstruction of uncontained bicondylar femoral bone defects.

Materials and Methods

Five synthetic distal femora with a bicondylar defect were reconstructed with impacted bone grafting (IBG) and this new knee revision design. A custom-made screw connection between the stem and the intercondylar box was designed to lock or initiate the sliding mechanism, another screw (dis)connected the stem. A cyclically axial load of 500 N was applied to the prosthetic condyles to assess the stability of the reconstruction. Radiostereometry was used to determine the migrations of the femoral component with a rigidly connected stem, a sliding stem and no stem extension.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 106 - 106
1 Jun 2012
Janssen D Srinivasan P Scheerlinck T Verdonschot N
Full Access

Introduction

Hip resurfacing arthroplasty has gained popularity as an alternative for total hip arthroplasty. Usually, cemented fixation is used for the femoral component. However, each type of resurfacing design has its own recommended cementing technique.

In a recent investigation the effect of various cementing techniques on cement mantle properties was studied. This study showed distinct differences in cement mantle volume, filling index and morphology.

In this study, we investigated the effect of these cement mantle variations on the heat generation during polymerization, and its consequences in terms of thermal bone necrosis.

Materials and methods

Two FEA models of resurfacing reconstructions were created based on CT-data of in vitroimplantations (Fig 1). The two models had distinct differences with respect to the amount of cement that was used for fixation. The first model was based on an implantation with low-viscosity cement, with anchoring holes drilled in the bone, and suction applied to maximize cement penetration. The second model was based on an implantation with medium viscosity cement smeared onto the bone, with no holes and no suction, leading to a thin cement layer.

Thermal analyses were performed of the polymerization process, simulating three different types of bone cement: Simplex P (Stryker), CMW3 (DePuy J&J) and Osteobond (Zimmer), with distinct differences in polymerization characteristics. The polymerization kinematics were based on data reported previously.

During the polymerization simulations the cement and bone temperature were monitored. Based on the local temperature and time of exposure, the occurrence of thermal bone necrosis was predicted. The total volume of necrotic bone was calculated for each case.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 425 - 426
1 Nov 2011
Janssen D Waanders D Mann K Verdonschot N
Full Access

The stability of cemented hip implants relies on the fixation of the cement mantle within the bone cavity. This fixation has been investigated in experiments with cement-bone interface specimens, which have shown that the cement-bone interface is much more compliant than is commonly assumed. Other studies demonstrated that the mechanical response of the interface is dependent on penetration of the cement into the bone. It is, however, unclear how cement penetration exactly affects the stiffness and strength of the cement-bone interface. We therefore used finite element (FE) models of cement-bone specimens to study the effect of cement penetration depth on the micromechanical behavior of the interface.

The FE models were created based on micro computed tomography (micro CT) data of two small cement-bone interface specimens (8x8x4 mm). The specimens had distinct differences with respect to interface morphology. In these models we varied the penetration depth, with six different penetration levels for each model. We then incrementally deformed each model in tension and in shear, until failure of the models. Failure was simulated to occur in the bone and cement when the local ultimate tensile stress was exceeded, by locally reducing the material stiffness to near zero. From the resulting force-displacement curves we established the apparent tensile stiffness and strength for each of the models.

Our results indicated that the strength and stiffness of the cement-bone interface increased with increasing cement penetration depth, both in tension and in shear. However, after reaching a certain penetration depth, both strength and stiffness did not further increase. This depth was dependent on the specific interface morphology. We furthermore found that the strength of the models was higher in shear than in tension. After failure of the models, damage was mainly found in the cement, rather than in the bone.

The FE-based techniques developed for the current study are suitable for exploration of a variety of aspects that may affect the cement-bone interface micromechanics, such as biological changes to the bone and variations of cement material properties.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 455 - 455
1 Nov 2011
Zelle J De Waal Malefijt M Verdonschot N
Full Access

High-flexion knee replacements have been developed to accommodate a large range of motion (ROM > 120°) after total knee arthroplasty (TKA). Femoral rollback or posterior translation of the femoral condyles during knee flexion is essential to maximise ROM and to avoid bone-implant impingement during deep knee flexion. The posterior cruciate ligament (PCL) has been described as the main contributor to femoral rollback. In posterior-stabilised TKA designs the PCL is substituted by a post-cam mechanism. The main objective of this study was to analyse the mechanical interaction between the PCL and a highflexion cruciate-retaining knee replacement during deep knee flexion. For this purpose, the mechanical performance of the high-flexion cruciate-retaining TKA design was evaluated and compared with two control designs including a highflexion posterior-stabilised design.

Materials & Methods: Prosthetic knee kinematics and kinetics were computed using a three-dimensional dynamic finite element (FE) model of the knee joint. The FE knee model consisted of a distal femur, a proximal tibia and fibula, a quadriceps and patella tendon, a non-resurfaced patella, TKA components and a posterior cruciate ligament in case cruciate-retaining designs were evaluated. Tibio-femoral and patello-femoral contact were defined in the FE knee model and the polyethylene insert was modelled as a non-linear elastic-plastic material. Three different rotating platform TKA systems were analysed in this study: the high-flexion cruciate-retaining PFC Sigma CR150, the high-flexion posterior-stabilised PFC Sigma RP-F and the conventional cruciate-retaining PFC Sigma RP (Depuy, J& J, UK). Both the polyethylene stress characteristics and the tibio-femoral contact locations were evaluated during a squatting movement (ROM = 50° – 150°).

Results: During deep knee flexion (ROM > 120°), the high-flexion cruciate-retaining TKA design showed a lower peak contact stress (74.7 MPa) than the conventional cruciate-retaining design (96.5 MPa). The posterior-stabilized high-flexion TKA design demonstrated the lowest peak contact stress at the condylar contact interface (54.2 MPa), although the post was loaded higher (77.4 MPa). All three TKA designs produced femoral rollback in the normal flexion range (ROM ≤ 120°), whereas the cruciate-retaining designs showed a paradoxical anterior movement of the femoral condyles during high-flexion.

Discussion: PCL retention is a challenging surgical aim and affects the prosthetic knee load and kinematics as shown in this study. In addition, for adequate functioning the PCL should not be too tight or too lax after surgery. Hence, we investigated the effect of PCL laxity on the prosthetic performance and the best-balanced PCL was used in our simulations. Although PCL balancing is not an issue for posterior-stabilized TKA, we found the tibial post to be loaded relatively high for this implant type.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 463 - 464
1 Nov 2011
Heesterbeek P Keijsers N Verdonschot N Wymenga A
Full Access

Instability is a major cause for revision surgery in total knee replacement (TKR). With a balanced gap technique, the ligaments are theoretically balanced. However, there is concern that ligament releases needed to align the leg may cause instability. Furthermore, no information is available about the relationship between the amount of varus-valgus laxity directly after implantation and at a later postoperative interval. This prospective clinical study investigated whether ligament releases necessary during total knee replacement (TKR) led to a higher varus-valgus laxity during peroperative examination and after 6 months.

In this prospective cohort study, in 49 patients a primary TKR was implanted using a balanced gap technique. Varus and valgus laxity of the knee was assessed in extension and flexion (70 degrees) per-operative (before and after implant) with a navigation system and post-operative with standardised stress radiographs (both methods 15 Nm stress applied).

Knees were catalogued according to ligament releases performed during surgery: no releases, lateral releases, medial releases with posteromedial condyle (PMC), and medial releases with superficial medial collateral ligament (SMCL). ANOVA was used to test between release groups.

At surgery, before and after implantation of the prosthesis, there was no difference in varus or valgus laxity in extension and flexion between knees that did not need a ligament release (n=22), knees with lateral release (n=5), knees with medial SMCL releases (n=15) and knees with medial PMC releases (n=7). Six months after TKR, varus or valgus laxity in extension and flexion was not significantly different between the release categories.

In conclusion, ligament releases of the SMCL, PMC, and lateral structures performed during a balanced gap technique in TKR do not lead to an increased varus-valgus laxity in extension and flexion at 6 months after surgery. Therefore, routine releases of these structures to achieve neutral leg alignment can safely be performed without causing increased varus-valgus laxity. The results of this study suggest that the reported high incidence of revisions for ligament instability after TKR is not likely to be caused by routine ligament releases when a balanced gap technique is used. Apparently, there is not a ligament instability problem as long as the gaps are properly filled with prosthesis components. We believe that the conclusion of this study would also be valid when bone referenced techniques are applied instead of tensors, as long as the gaps created are balanced.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 446 - 447
1 Nov 2011
Walschot L Aquarius R Schreurs B Verdonschot N Buma P
Full Access

Bone impaction grafting (BIG) is a surgical technique for the restoration of bone stock loss with impaction of autograft or allograft bone particles (BoP). The goal of a series in-vitro and in-vivo experiments was to assess the suitability of deformable pure Ti (titanium) particles (TiP, FONDEL MEDICAL BV, Rotterdam, The Netherlands) for application as a full bone graft substitute in cemented revision total hip arthroplasty. TiP are highly porous (interconnective porosity before impaction 85 to 90%). In-vitro acetabular reconstructions were made in Sawbones (SAWBONES EUROPE, Malmö, Sweden) to evaluate migration by roentgen stereo photogrammetric analysis and shear force resistance by a lever out experiment. In-vitro femoral TiP reconstructions (SAWBONES, Malmö, Sweden) were used to evaluate micro-particle release and subsidence. Mature Dutch milk goats were used for two in-vivo experiments.

A non-loaded femoral defect model was used to compare osteoconduction of bioceramic coated TiP with BoP and ceramic particles (CeP).

Acetabular defects (AAOS type 3) were reconstructed in 10 goats using a metal mesh with impacted TiP acting as a full bone graft substitute in combination with a cemented polyethylene cup and a downsized cemented Exeter femoral stem (STRYKER BENOIST, Girard, France).

Blood samples were taken for toxicological analysis.

In-vitro: TiP were as deformable as BoP and created an entangled graft layer (porosity after impaction 70 to 75%). Acetabular TiP reconstructions were more stable and resistant to subsidence and shear force than BoP reconstructions (lever-out moment 56 ± 12 Nm respectively 12 ± 4 Nm, p < 0.001). After initial setting, femoral subsidence rates were smaller than seen in femoral bone impaction grafting (0.45 ± 0.04 mm after 300 000 loading cycles). Impaction generated 1.3 mg particles/g TiP (particle Ø 0.7–2 000 μm, tri-modal size distribution). In-vivo: Bioceramic coated (10 −40 μm) TiP showed bone ingrowth rates comparable to BoP and CeP. Reconstructed acetabular defects showed rapid bone ingrowth into the layer of TiP. Serum titanium concentrations slowly increased from 0.60 ± 0.28 parts per billion (ppb) preoperatively to 1.06 ± 0.70 ppb at fifteen weeks postoperatively (p = 0.04).

Mechanical studies showed very good initial mechanical properties of TiP reconstructed defects. The in-vitro study showed micro-particle generation, but in the short-term goat studies, histology showed very few particles and no negative biological effects were found. The in-vivo acetabular study showed very favorable bone ingrowth characteristics into the TiP layer and a much thinner interface with the cement layer compared to similar defects reconstructed with BoP or mixtures of BoP with CeP. Further analysis in a human pilot study should proof that TiP is an attractive and safe alternative for allograft bone in impaction grafting revision arthroplasty.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 432 - 432
1 Nov 2011
Heesterbeek P Keijsers N Verdonschot N Wymenga A
Full Access

Balancing the PCL in a PCL-retaining total knee replacement (TKR) is important, but sometimes difficult to execute in an optimal manner. Due to the orientation of the PCL it is conceivable that flexion gap distraction will lead to anterior movement of the tibia relative to the femur. This tibio-femoral repositioning influences the tibio-femoral contact point, which on its turn affects the kinematics of the TKR. So far, the amount of tibiofemoral repositioning during flexion gap distraction is unknown which leads to uncertain kinematic effects after surgery. The goal of this study was to quantitatively describe the parameters of the flexion gap (gap height, anterior tibial translation and femoral rotation) and their relationship while the knee is distracted during implantation of a PCL-retaining TKR with the use of computer navigation. Furthermore, the effect of PCL elevation angle on the flexion gap parameters was determined.

In 50 knees, during a ligament-guided TKR procedure, the flexion gap was distracted with a double-spring tensor with 100 and 200 N after the tibia had been cut. The flexion gap height, anterior tibial translation and femoral rotation were measured intra-operatively using a CT-free navigation system. PCL elevation was calculated based on the femoral and tibial insertion sites as indicated by the surgeon with the pointer of the navigation system.

To identify a relationship between flexion gap height increase and anterior tibial translation, the ratio between anterior translation and gap height increase was determined for each patient between 100 and 200 N.

The mean gap height increased 2.2 mm (SD 0.96) and mean increase in anterior tibial translation was 4.2 mm (SD 1.6). Hence, on average, for each mm increase in gap height, the tibia moved 1.9 mm (SD 0.96) in anterior direction. Knees with a steep PCL showed significantly more AP translation for each mm gap height increase (gap/AP-ratio was 1 : 2.31 (SD 0.63)) compared to knees with a flat PCL (gap/AP-ratio was 1 : 1.73 (SD 0.50)).

The increase in femur (exo)rotation was on average 0.60° (SD 1.4).

With a tensioned PCL the tibia will move anteriorly on average 1.9 mm for every extra mm that the flexion gap is increased. The flexion gap dynamics can be explained in part by the orientation of the PCL: the greater the elevation angle, the more anterior tibial displacement during distraction of the flexion gap. The surgeon must be aware that distraction of the flexion gap influences the tibiofemoral contact point. The tibio-femoral contact point will move posteriorly and stresses in the PCL will rise and produce limited flexion and pain. In case of a conforming insert AP-movement will be limited but high PE stresses may be introduced that can lead to wear. This information may be helpful in selecting the optimal soft tissue balancing procedure and the optimal PE insert thickness in PCL retaining TKR.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 90 - 90
1 Mar 2010
Verdonschot N van Tienen T Buma P
Full Access

Meniscectomy, induces osteoarthritis. Options for repair of a damaged meniscus are an allograft meniscus, an implant made of natural scaffold materials (the collagen meniscus implant; CMI) or an implant made of polymers.

Allograft menisci and the CMI are already clinically used for a considerably number of years. In this educational lecture the focus is on a comparison between the three implant types and the status of a tissue-engineered meniscus.

The allograft meniscus is already used for at least ten years. It is intended for the younger patient with a previous total meniscectomy, with moderate cartilage degeneration and with a good alignment of the knee. The clinical outcome is based on function and pain scores. In this lecture the functional scores, the survival rate and the histology of allograft menisci will be highlighted.

The CMI meniscus implant is intended for a different patient group. To enable implantation of the CMI the rim of the native meniscus should be intact. Patient series that should demonstrate the efficacy of this type of implant are still small and are mainly of the inventors of the implant. In general patients tolerated the implant well. Tissue ingrowth and remodelling into a fibro-cartilaginous tissue was found in animals and patients.

Polymers may be a good alternative for the allograft and CMI implant. Previously they were used to guide vascularized new repair tissue through an ingrowth channel to the avascular lesion. We developed a porous polymer meniscus scaffold with properties to allow tissue infiltration and regeneration of a neomeniscus. It was implanted in dog knees and compared with total meniscectomy. The tissue infiltration and redifferentiation in the scaffold, the stiffness of the scaffold, and the articular cartilage degeneration were evaluated.

Three months after implantation, the implant was completely filled with fibrovascular tissue. After 6 months, the central areas of the implant contained cartilage-like tissue with abundant collagen type II and proteoglycans in their matrix. The foreign-body reaction remained limited to a few giant cells in the implant. The compression modulus of the implant-tissue construct still differed significantly from that of the native meniscus, even at 6 months. Cartilage degeneration was observed both in the meniscectomy group and in the implant group.

The improved properties of these polymer implants resulted in a faster tissue infiltration and in phenotypical differentiation into tissue resembling that of the native meniscus. However, the material characteristics of the implant need to be improved to prevent degeneration of the articular cartilage.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 124 - 124
1 Mar 2010
Zelle J Barink M Verdonschot N
Full Access

Introduction: Recently, high-flexion knee implants have been developed to provide for a large range of motion (ROM > 120°) after total knee arthroplasty (TKA). High-flexion knee implants are more likely subjected to large knee loads than conventional implants since knee joint forces increase with larger flexion angles. Highly conforming knee replacements are designed to minimise polyethylene peak stresses during (deep) knee flexion.

The Birmingham Knee Replacement (BKR, Jointmedica, UK) is a newly designed knee replacement which combines a high conformity during the complete ROM with the principles of rotating platform and high-flexion TKA. The main objective of this study was to analyze the mechanical performance of the BKR during its full ROM (0°–155°) and investigate whether its high conformity could be maintained during high-flexion. In addition, the BKR polyethylene loading computed in this study was compared with other mobile bearings.

Materials & methods: TKA performance was analyzed using a three-dimensional dynamic finite element (FE) model of the knee joint. The FE knee model consisted of a distal femur, a proximal tibia and fibula, a quadriceps and patella tendon, a non-resurfaced patella and TKA components. Tibio-femoral and patello-femoral contact were defined in the knee model. Three different posterior stabilised rotating platform TKAs were subsequently incorporated: the high-flexion BKR, the high-flexion PFC Sigma RP-F and the standard PFC Sigma RP (Depuy, J& J, USA). The polyethylene insert was modelled as a non-linear elastic-plastic material in each TKA system. Polyethylene loading parameters as well as the tibio-femoral contact point locations were computed during an entire flexion movement (0°–155°).

Results: In the normal flexion range (flexion ≤ 120°) the three knee implants behaved very similar except for the polyethylene loading at the post. At 120° of flexion, the contact stress at the dish was ±45 MPa for all implants whereas the maximal post-cam contact stress came down to 26.7 MPa for the BKR which was half the amount of contact stress experienced by both PFC Sigma implants. During high-flexion (flexion > 120°), the contact stress difference at the post between the BKR and the PFC Sigma RP-F became smaller and came down to 37.9 MPa and 60.7 MPa, respectively. The total amount of plastic deformation at maximal flexion (155°) was smaller for the BKR (577 mm3) in comparison with the Sigma RP-F (2256 mm3). Femoral rollback was negligible for the BKR in the high-flexion range in comparison with the Sigma RP-F (1.9 mm).

Discussion: A comparison between different geometrical models using finite element techniques is jeopardised by differences in element distribution within the various models. These differences may affect calculated parameters such as peak stress values. However, in this study the models were very similar which would indicate that the differences in stress patterns found are due to design differences rather than model artefacts.

The current study therefore indicates that the BKR benefits from its high conformity during the full ROM. Hence, the BKR demonstrated relatively low polyethylene stresses. The quadriceps efficiency during deep knee flexion may be lower in case of the BKR since the femoral rollback was negligible at these flexion angles. Whether this phenomenon is of any clinical relevance is unknown.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 443 - 443
1 Sep 2009
van Aken J Verdonschot N Huizenga H Kooloos J Tanck E
Full Access

Bone metastases occur in about 15% of all cancer cases. Pathological fractures that result from these tumours most frequently occur in the femur. It is extremely difficult to determine the fracture risk with the current X-ray methods, even for experienced physicians. The purpose of this study was to assess whether the use of a predictive finite element model could improve the prediction of strength in comparison to an clinical assessment.

Eight human cadaver femora, with and without simulated metastases, were CT-scanned. A solid calibration phantom was included in each scan. From the scans, eight finite element (FE) models were generated using brick elements. The non-linear mechanical properties were based on bone density. After scanning, laboratory experiments were performed. The femora were loaded under compression until failure. During the experiments the failure forces and the course of failure were registered. These experiments were simulated in the FE-models, in which plastic deformation simulated failure of the bones. Six experienced physicians, were asked to rank the femora on strength using X-rays (AP and ML) and additional information on gender and age.

The results showed a strong Pearson’s correlation (r2 = 0.92) between the experimental failure force and predicted failure force. The Spearman’s rank correlations between experiment and predictions ranged between ρ=0.58 and ρ=0.8 for the physicians, whereas it was significantly higher (ρ=0.92) for the FE-model

This study showed that femur specific FE models better predicted femoral failure risk under axial loading than experienced physicians. When the model is further improved by adding, for example, other loading conditions, it can be clinically implemented to predict in vivo fracture risk for patients suffering, for example, bone metastases or osteoporosis.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 456 - 456
1 Sep 2009
Zelle J Barink M De Waal Malefijt M Verdonschot N
Full Access

Recently, high-flexion knee implants have been developed to provide for a large range of motion after total knee arthroplasty. Since knee forces increase with larger flexion angles, it is commonly assumed that high-flex-ion implants are subjected to large loads in the highflexion range (flexion > 120°). However, high-flexion studies often do not consider thigh-calf contact which occurs during high-flexion activities such as squatting and kneeling. We hypothesized that thigh-calf contact is substantial and has a reducing effect on the prosthetic knee loading during deep knee flexion.

The effect of thigh-calf contact on the loading of a knee implant was evaluated using a three-dimensional dynamic finite element knee model. The knee model consisted of a distal femur, a proximal tibia and fibula, a patella, high-flexion components of the PFC Sigma RP-F (Depuy, Warsaw, USA) and a quadriceps and patella tendon. Using this knee model, a squatting movement was simulated including thigh-calf contact characteristics of a typical subject which have been described in an earlier study.

Thigh-calf contact considerably reduced the implant loading during deep knee flexion. At maximal flexion (155°), the compressive knee force decreased from 4.9 to 2.9 times bodyweight. The maximal joint forces shifted from occurring at maximal flexion angle to the flexion angle at which thigh-calf contact initiated (±130°). The maximal polyethylene contact stress at the tibial post decreased from 49.3 to 28.1 MPa at maximal flexion.

This study confirms that thigh-calf contact reduces the knee loading during high-flexion. Both the joint forces and the polyethylene stresses reduced considerably when thigh-calf contact was included.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 462 - 462
1 Sep 2009
Aquarius R Walschot L Buma P Schreurs BW Verdonschot N
Full Access

We investigated the feasibility of using porous titanium particles (TiP) to reconstruct femoral bone defects in revision hip replacement surgery in stead of using morzelised bone grafts. Questions regarding handling, initial stability and titanium particle release were addressed.

Seven composite femurs (Sawbones) were reamed and filled, stepwise, with 32 grams of large (Ø 3.15 – 4 mm) and 9 grams of smaller (Ø 2.8 – 3.15 mm) pure, 85% porous TiP. Subsequently an Exeter stem was cemented into the graft layer. All reconstructions were loaded axially (0–3000 N) for 300,000 loading cycles at 2 Hz. Subsidence of the stem was measured with radio stereometric analysis (RSA) and possible titanium particle release was measured using the laser diffraction technique.

The TiP were impacted into a > 3 mm (SD 1.43 mm) thick, highly entangled, graft layer. An average cement mantle of > 2 mm (SD 0.86 mm) was measured and little cement penetration was observed. The average subsidence of only 0.45 mm (SD 0.04 mm) was measured after 300 000 loading cycles. Most titanium particles were found directly after impaction. Most of these particles (87%) were smaller than 10 μm and could therefore be potentially harmful since they can induce osteolysis.

We can conclude that:

A graft layer of impacted TiP can be constructed,

The graft layer is stable enough to initially support a cemented Exeter stem,

Titanium particles are released during impaction.

These data warrant further animal tests to assess the biological response to these released impaction particles. Also, animal tests should clarify possible particle release upon loading and its effects.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1148 - 1153
1 Sep 2009
Schreurs BW Keurentjes JC Gardeniers JWM Verdonschot N Slooff TJJH Veth RPH

We present an update of the clinical and radiological results of 62 consecutive acetabular revisions using impacted morsellised cancellous bone grafts and a cemented acetabular component in 58 patients, at a mean follow-up of 22.2 years (20 to 25). The Kaplan-Meier survivorship for the acetabular component with revision for any reason as the endpoint was 75% at 20 years (95% confidence interval (CI) 62 to 88) when 16 hips were at risk. Excluding two revisions for septic loosening at three and six years, the survivorship at 20 years was 79% (95% CI 67 to 93). With further exclusions of one revision of a well-fixed acetabular component after 12 years during a femoral revision and two after 17 years for wear of the acetabular component, the survivorship for aseptic loosening was 87% at 20 years (95% CI 76 to 97). At the final review 14 of the 16 surviving hips had radiographs available. There was one additional case of radiological loosening and four acetabular reconstructions showed progressive radiolucent lines in one or two zones.

Acetabular revision using impacted large morsellised bone chips (0.5 cm to 1 cm in diameter) and a cemented acetabular component remains a reliable technique for reconstruction, even when assessed at more than 20 years after surgery.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_II | Pages 290 - 290
1 May 2009
Caruana J Janssen D Verdonschot N Hua J Blunn G
Full Access

Introduction: Finite element (FE) simulation of damage accumulation in the femoral cement mantle is widely used to predict failure of hip prostheses. It is often assumed that the stem-cement interface remains bonded, although debonding is thought to affect cement stress and damage. Rough stems may reduce subsidence, but have been reported to have a detrimental effect on implant survival. Other factors thought to influence cement damage include stem design and orientation and cement thickness. This study investigates the effect of cement mantle thickness and stem malpositioning on cement damage around a smooth, collared implant, and the extent to which this is affected by debonding of the stem-cement interface.

Method: Three FE meshes were built to represent proximal femora with Stanmore Hip prostheses implanted into a thick (2.5 mm) and a thin (1.0 mm) cement mantle, and another thin (1.0 mm) mantle with the implant tilted in varus to achieve a minimal thickness of 0.1 mm laterally. Each model consisted of 4304 eight-noded brick elements with frictional contact at the stem-cement interface. Two analyses were run for each model, in which the stem-cement interface was (a) fully bonded, and (b) fully debonded, with Coulomb frictional contact using a friction coefficient of 0.5. Standardised femur geometry and elastic properties were used. Creep and non-linear damage accumulation in the cement mantle under cyclic loading was modelled using subroutines developed by Stolk et al. (2003). Boundary conditions were applied representing a peak stair-climbing load.

Results: Bonded cases showed extensive cracking around the tip in all cases. Debonded cases had 4–8 times less cracking, which was much more focused at the tip; only the poorly-centralised mantle showed extensive damage elsewhere, in the very thin lateral region. When bonded, the thick mantle had least cracks and the poorly-centralised mantle had most; in the debonded cases, there was no major difference between thick, thin, and poorly-centralised mantles. For each cement mantle geometry, peak maximum principal cement stress was consistently lower in the debonded case than in the bonded case.

Discussion: Our results show greater, more widely distributed cracking in bonded than debonded cement mantles, in contrast with previous studies involving collarless implants. For a collared stem, calcar contact prevents subsidence, allowing cement stress relaxation. A possible explanation for our result is that debonding enhances the stress relaxation process, reducing and redistributing interfacial and shear stresses; thus reducing damage rates. In contrast, a debonded collarless stem subsides continuously, sustaining high cement stress levels and damage rates. These results may explain the disappointing clinical performance of some rough-surfaced prostheses. Our results suggest that bonding might increase both cement damage and its sensitivity to cement thickness. Similar results for all debonded cement mantles indicate that cement thickness may be less critical than previously thought for smooth, collared prostheses. Bonding should not be assumed in FE studies of smooth stems which clinically are likely to debond; cement damage simulation should be extended to incorporate the debonding process.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 91 - 91
1 Mar 2009
Scheerlinck T Janssen D van Aken J Verdonschot N
Full Access

Introduction: It is generally accepted that the cement mantle surrounding femoral hip implants should be at least 2–3 mm thick. To achieve that goal, manufactures or surgeons often undersize the stem compared to the broach. However, some implants, such as the Charnley-Kerboul stem, are typically cemented line-to-line i.e. with a broach and stem of the same size. Despite their “minimal” cement mantle, these stems are very successful. This apparent contradiction is known as the “French Paradox”[1]. We used a finite element analysis (FEA) model to investigate the effect of these different cementation philosophies on cement crack propagation and rotational stem stability.

Material and Methods: Based on a CT-scan image of a Charnley-Kerboul plastic stem replica[2], twelve FEA models were created. By decreasing the stem size (4 stems), the average cement mantle thickness increased (1.71–3.77mm). However, the incidence of cement mantle defects (< 1mm) and areas of thin cement (< 2mm) decreased (defects: 34.7–0.0%; thin cement: 40.7–0.0%). The amount of cortical bone support was varied (3 times) between 18.4 and 72.2%. All models were alternately loaded with a cyclic torque load (25.8Nm) and a transversal load (400N) in a ratio of 9:1 for two million cycles. The model predicted fatigue crack formation within the cement and rotational stem stability.

Results: Overall, increasing implant size and increasing the amount of cortical bone support to the cement, improved resistance to accumulated cement damage and rotational stem stability. In both models with undersized stems, more cement cracks and full thickness (FT) cement fractures appeared after less loading cycles than in both models with canal-filling stems. Worst results were obtained with a severely undersized implant surrounded by a thick cement mantle that was poorly supported by cortical bone (first FT crack after < 100 000 cycles, > 220 initiated cracks and 0.6° of implant rotation after 2 million cycles). Best results were obtained with the maximal canal-filling stem surrounded by a thin and deficient cement mantle that was well supported by cortical bone (no FT cracks, < 10 initiated cracks and 0.3° of implant rotation after 2 million cycles).

Conclusion: This study emphasizes the importance of an adequate cementation technique that aims at pressurizing cement up to the cortical bone. This protects the cement mantle against fatigue fracture and stabilises the implant especially if the stem is undersized. From a mechanical point of view, canal-filling stems make sense. They limited the formation of cement cracks and improved rotational stability to the implant. This could explain the excellent results obtained by implants that are cemented line-to-line.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 156 - 156
1 Mar 2008
Banks S Ginsel B Mitchell K Verdonschot N Hodge W
Full Access

Contemporary knee implants use a variety of methods to control tibiofemoral motions. Posterior stabilized implants have a post and cam to force the femur posterior with flexion. Most posterior cruciate retaining designs rely solely on this ligament and symmetric tibial surfaces to control tibiofemoral translations. However, many studies have demonstrated poor control of tibiofemoral motion in PCL retaining knees. One strategy to augmenting PCL function is to provide a gait-congruent lateral articulation providing definitive stability in extension while allowing lateral condylar translation in deep flexion. It is unknown whether this design strategy, essentially substituting for the ACL, allows the PCL to function more normally.

Fifteen knees in ten patients with a fixed-bearing, PCL retaining, lateral pivot arthroplasty were observed during maximum flexion kneeling and lunging using fluoroscopy. The tibial insert provides a fully conforming lateral articulation from 0°–70° flexion, allowing lateral AP translation at greater flexion. Recruited on the basis of combined KSS scores > 180 points, patients averaged 72 years, 27.5 BMI, and 12 months post-op. Shape matching techniques were used to determine the 3D pose of the implant components.

Skeletal flexion during kneeling averaged 134° (117°–156°) with 11° tibial internal rotation. Medial condylar contact was 3mm posterior, and lateral contact was 11 mm posterior to the tibial AP midpoint. Skeletal flexion during lunging averaged 122° (106°–146°) with 11° tibialinternal rotation. Medial condylar contact was 1mm posterior, and lateral condylar contact was 9mm posterior to the tibial AP midpoint.

Knees with lateral pivot arthroplasty exhibited flexion comparable to the best reported results in North American patients. Tibial rotation was statistically greater than has been reported for symmetric posterior stabilized or PCL retaining implants for the same activities. Posterior translation of the condyles with flexion beyond the range of full articular congruity is consistent with relatively normal PCL function.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 188 - 189
1 Mar 2008
Verdonschot N Willems MM Stungo B Slomsikowsky M Gibbons P Kriek H Revie I
Full Access

Most minimal invasive surgical (MIS) systems use traditional implant systems combined with new instrumentation. In this study we analyzed a THR system that basically implies that all components are implanted through the femoral neck. The cemented femoral component consists of a highly polished tapered design. The acetabular component is made of Alumina and has an outside diameter of 20 mm. The purpose of this study was to investigate the range of motion, the wear characteristics, the fatigue characteristics of the femoral neck and the stability of the femoral component.

The range of motion of the MIS prosthetic system was calculated with a mathematical model that enabled calculation of prosthetic impingement angles. To assess the wear properties, four pairs of Zirconia heads on alumina acetabuli were tested in a hip simulato. To assess the probability of femoral neck fracture, 3 components were tested according to ISO7206. The stability of the femoral components were tested in five fresh cadaver using dynamic loading conditions. After this test, the load was increased until reconstructive failure occurred.

The ROM was in the order of 100 degrees of flexion and at least 30 degrees in other directions. The bearings showed remarkably low wear with a maximum of 0.02 mm3. All three stems survived the ISO-fatigue test. During the dynamic experiments the specimens did not fail, and no macroscopic damage was detected. Migration was only minor and stabilized during testing. The post-testing failure loads varied between 4.1 and 5.5 kN.

The ROM, stem-neck strength and wear properties of the system seem acceptable. The stability of the femoral component was satisfying; but the post-testing strength may be similar to loads that are applied on the hip at a falling accident. We conclude that these results are encouraging and warrant further studies to develop this system.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 156 - 157
1 Mar 2008
Barink M Verdonschot N De Waal Malefijt M Van Kampen A
Full Access

It is impossible to determine the effect of a single parameter in clinical or in-vitro knee research. There are also parameters which can not or hardly be determined. These disadvantages can be overcome with a model. The objective of this study was to create a dynamic FE model of a human knee joint after TKA which is applicable to a variety of research question.

The knee model consisted of a femur, tibia and patella, collateral ligaments and a PCL, combined with a CKS cruciate retaining total knee prosthesis. The patella was not resurfaced. An axialload of 150 N and a quadriceps-force of 81N was applied. The model was validated by the model prediction of joint laxities at different flexion-angles and the calculation of the knee kinematics during flexion-extension.

The predicted varus-valgus laxity at different flexion angles was in between 0 and 6.3 degrees. Laxity values decreased towards extension and towards 90 degrees of flexion. The AP test at 20, 30 and 90 degrees of flexion showed a anterior laxity of 3.1, 4.3 and 2 mm, respectively. The posterior laxity was 5.7 mm, but could only be determined at 90 degrees. The model predicted reasonable kinematics, which were identical for two consecutive flexion-extension movements.

The model predictions were well in agreement with reported values, which were measured experimentally. Differences could be well explained by ligament structures which were (still) omitted with in the model. This dynamic model, in which ligaments were actually modelled as bands, combined all major structures within the knee joint. It was well able to predict laxities and kinematics and turned out to be very stable, mathematically. With this model we will be able to address effects of prosthetic and surgical parameters on the stability and kinematics of the knee joint.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 68 - 68
1 Mar 2006
Schreurs B Arts J Verdonschot N Buma P Slooff J Gardeniers J
Full Access

Introduction: The purpose of this study was to evaluate the long-term clinical and radiological outcome of instrumented femoral revisions after failed total hip arthroplasties using the impaction bone grafting technique with morsellized bone chips in combination with a cemented polished stem.

Methods: Thirty-three consecutive femoral reconstructions were performed between November 1991 and February 1996 using the X-Change femoral impaction system with fresh frozen morsellized bone grafts and a cemented polished Exeter stem. All patients were prospectively followed. The learning curve with this new technique is included in this report. This technique was used in twenty-four women and nine men; the average age at surgery was sixty-three years (range 33–82). Femoral bone stock defects were classified according to the Endoklinik classification as grade 1 in three hips, grade 2 in fourteen hips, grade 3 in twelve hips and grade 4 in four hips. At a minimal eight years follow-up no patient was lost to follow-up, but eight patients died (at 0.5, 3.5, 3.5, 7.0, 7.0, 7.5 and 9.0 after reconstruction). All were followed until death, none of these deaths was related to the surgery, and none had a re-revision

Results: No femoral reconstruction was re-rerevised at a mean follow-up of 10.4 years (range 8 to 13 years). However, there were three femoral fractures during follow-up (at 3, 6 and 22 months), all at the level of tip of the prostheses. All healed after plating, all femoral implants were left in situ. The average Harris hip score improved from 49 prior to surgery to 85 at review (68–100). The average migration of the stem within the cement mantle was three mm (0–14 mm), most migration was seen in the first year. Radiologically, there were no failures. With an endpoint of femoral revision for any reason, with endpoint aseptic loosening or with endpoint radiological loosening the survival rate using the Kaplan-Meier analysis was hundred per cent in all situations (one-sided 95% C.I. 100–91.3 %)

Conclusions Femoral revision using bone impaction grafting with fresh frozen bone grafts and a cemented polished stem showed an excellent survival at eight to thirteen years follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1561 - 1567
1 Nov 2005
Janssen D Aquarius R Stolk J Verdonschot N

The Capital Hip implant was a Charnley-based system which included a flanged and a roundback stem, both of which were available in stainless steel and titanium. The system was withdrawn from the market because of its inferior performance. However, all four of the designs did not produce poor rates of survival. Using a simulated-based, finite-element analysis, we have analysed the Capital Hip system. Our aim was to investigate whether our simulation was able to detect differences which could account for the varying survival between the Capital Hip designs, thereby further validating the simulation.

We created finite-element models of reconstructions with the flanged and roundback Capital Hips. A loading history was applied representing normal walking and stair-climbing, while we monitored the formation of fatigue cracks in the cement.

Corresponding to the clinical findings, our simulation was able to detect the negative effects of the titanium material and the flanged design in the Capital Hip system. Although improvements could be made by including the effect of the roughness of the surface of the stem, our study increased the value of the model as a predictive tool for determining failure of an implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 656 - 663
1 May 2005
Toms AD McClelland D Chua L de Waal Malefijt M Verdonschot N Jones RS Kuiper J

Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray.

We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement.

Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 4 | Pages 492 - 497
1 May 2004
Schreurs BW Bolder SBT Gardeniers JWM Verdonschot N Slooff TJJH Veth RPH

This study presents the clinical and radiological results of 62 consecutive acetabular revisions in 58 patients, at a mean of 16.5 years follow-up (15 to 20). The Kaplan-Meier survivorship for the cup with end-point revisions for any reason, was 79% at 15 years (95% confidence interval (CI); 67 to 91). Excluding two revisions for septic loosening at three and six years, and one revision of a well-fixed cup after 12 years in the course of a femoral revision, the survivorship was 84% at 15 years (95% CI; 73 to 95). At review there were no additional cases of loosening, although seven acetabular reconstructions showed radiolucent lines in one or two zones. Acetabular revision using impacted large morsellised bone chips (0.7 cm to 1.0 cm) and a cemented cup, is a reliable technique of reconstruction, when assessed at more than 15 years.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 428 - 428
1 Apr 2004
Willems M Gibbons P Revie I Verdonschot N
Full Access

The increasing success rates of total hip replacements (THR) have led to a younger patient population with an increased probability for revision. The survival of revised components is improved by a good bone quality. This has led to an increased interest in bone preserving THR designs. A novel type of THR was developed of which the femoral component is cemented in the neck. The load carrying area of this prosthesis is reduced in comparison with conventional cemented implants. Whether an adequate stability can be achieved was biomechanically evaluated during simulated normal walking and chair rising. In addition, the failure behaviour was investigated.

Bone mineral density (BMD) was measured in 5 fresh frozen proximal human cadaver femora. The femoral heads were resected and a 20 mm diameter canal was created in the femoral necks. Bone cement was pressurised in this canal and the polished, taper-shaped prosthesis was subsequently introduced centrally. A servohydraulic testing machine was used to apply dynamic loads up to 1.8 kN to the prosthetic head. Radiostereophotogrammetric analysis was used to measure rotations and translations between prosthesis and bone. In addition, the constructions were loaded until failure in a displacement-controlled test.

During the dynamic experiments, the femoral necks did not fail, and no macroscopical damage was detected. The initial stability of the implant did not seem to be sensitive to bone quality. Maximal values were found for normal walking with a mean rotation of about 0.2 degrees and a mean translation of about 120 microns. These motions stabilised during testing. The failure loads in this study varied between 4.1 and 5.5 kN, higher failure loads were associated with higher BMD values. Most specimens showed subtrochanteric spiral fractures.

In conclusion, the stability of the prosthetic device may be adequate under dynamic, physiological loading conditions. The static failure loads were relatively low and require further optimisation of the prosthetic implant.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 420 - 420
1 Apr 2004
Barink M van de Groes S Verdonschot N de Waal Malefijt M
Full Access

Malfunctioning of Total Knee Replacements is often related to patella-femoral problems. As the patella groove guides the patella during flexion, the difference between anatomic- and prosthetic groove geometry may be of major influence concerning patella-femoral problems. This study focusses on the orientation or direction of the femoral patella groove, relative to the mechanical axis of the femur. Literature shows a controversy in measured groove orientation: Eckhoff et al. (1996) have measured a lateral groove, and Feinstein et al. (1996) have measured a medial groove, relative to the mechanical axis. Current femoral knee components have a lateral, or neutral directed patella groove. As most TKA surgical techniques subscribe an exorotation of the femoral component during implantation, the prosthetic in vivo situation will show a lateral groove. The objectives were to clarify the described controversy and to determine whether there is a difference in anatomic- and prosthetic groove orientation, which might cause patella-femoral problems.

The patella groove orientation of 100 human femora was measured using a 3-D measurement system. A spherical measurement probe was moved through the groove, starting at the notch and finishing at the cartilage edge, to simulate patella motion. The patella groove angle was defined as the angle between the mechanical axis and the measured groove points, in the frontal plane. A medial patella groove angle of 1.8±2.6° was measured. An implanted situation of a femoral component with neutral groove showed a lateral groove angle of 1.3°. An implanted situation of a femoral component with assymmetrical groove showed a lateral groove angle of 2.6°. The authors measured a medial oriented patella groove. This anatomical groove orientation is in contradiction with current femoral knee component design and surgical practice, because that results in a lateral oriented groove. This difference in anatomic- and prosthetic groove orientation may be a cause of patella-femoral problems.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 222 - 223
1 Mar 2004
Schreurs B Bolder S Buma P Verdonschot N
Full Access

Bone impaction grafting of the femur is associated with more complications when segmental defects are present. The effect of segmental defect repair on initial stem stability was studied in an in vitro study with fresh frozen goat femora. A standardized medial segmental defect was reconstructed using a cortical strut or a metal mesh. As controls we used intact femora and femora with a non-reconstructed defect. In all four groups impacted bone grafting was performed in combination with a cemented Exeter stem. Each group contained five femora. Reconstructions were dynamically loaded up to 1500N. Migration was measured with Roentgen Stereo-photogrammetric Analysis. All cases with a non-reconstructed segmental defect failed into excessive varus rotation. None of the femora with a reconstructed defect failed. Cortical struts and metal meshes were equally effective in creating a stable stem construction (varus rotation 2.89±2.27 and 2.27±0.57, respectively). Reconstructions with a metal mesh were more reproducible, although the obtained stability was significantly lower (p< 0.01) when compared to impaction grafting in an intact femur (varus rotation 0.58±0.36).

Besides, structural grafts may negatively influence the revascularization of the underlying impacted grafts in contrast to an open wire mesh. So, an in vivo study of 12 goats was done. A standardized medial wall defect was reconstructed with a strut or a mesh in six goats per group. In all femora impaction grafting was performed in combination with a cemented Exeter stem. After six weeks the femora were harvested. A high rate of peri-prosthetic fractures was found (43% and 29% for the strut and mesh groups, respectively). Histological and micro-radiological examination showed different revascularization patterns for both reconstruction techniques. In the strut group revascularized graft was found at the edges of the defect. In the mesh group fibrous tissue and blood vessels penetrated through the mesh and a superficial zone of revascularized grafts was found. Segmental defect reconstruction with a strut reduced the amount of revascularized grafts medially behind the strut (p=0.004). This may interfere with the stability of the stem in the first period after surgery and the incorporation of the impacted grafts on the long-term.

We would recommend segmental defect reconstruction with a mesh. A regime of unloading and long-stem prostheses should be used, irrespective of the reconstruction technique


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 368 - 368
1 Mar 2004
Boonstra M Jenniskens A Barink M Kooloos J Verdonschot N De Waal Malefijt M
Full Access

Aims: The purpose of this study is to validate a novel, quantitative functional test for TKA patients using commonly used subjective questionnaires as standard. Methods: Electrogoniometry was used to get information about the stability of rising from a chair. Thirteen pre Ðoperative TKA patients were measured while rising from a chair at two different heights. Using phaseplots (relationship between joint angle and joint velocity), the relative phase between the hip and knee was calculated. Instability was deþned as the standard deviation over ten trials for every patient and each height. An independent physician administered the WOMAC, KSS and SF-36 questionnaires. The instability factor of rising was correlated (using Spearmanñs Rho) to the outcomes of the questionnaires. Results: The stability of rising correlated moderately high with the questionnaires. Best correlations were calculated for instability and the WOMAC and SF-36 physical functioning (0.45 and 0.56, respectively). The worst correlation was with the KSS pain score (0.05). In general, the correlations for the lowest chair height were slightly higher. Conclusions: As there is no gold standard for functional evaluation of the TKA, a chair rise protocol has been developed. In this study the stability of rising correlates moderately high with the WOMAC, KSS and SF-36 questionnaires. The stability of rising from a lower chair seems to discriminate better than rising from a higher chair. This test will be used as a quantitative functional follow up of TKA patients.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 367 - 367
1 Mar 2004
Arts J Schreurs B Verdonschot N Buma P
Full Access

Aims: The effects of bone graft washing and bone graft particle size on initial cup stability in an acetabular model were studied. Methods: From fresh human femoral heads large (8–12 mm) and small (2–4) grafts were produced. Four different groups were tested; large washed, large unwashed, small washed and small unwashed. An AAOS type 3 combined defect was reconstructed in a synthetic acetabular model using a metal mesh and screws. Bone grafts were impacted in a standardized way using a hammer and metal impactors. Cement was pressurized and a PE cup was inserted. Cups were dynamically loaded with 1500 N and 3000 N for 15 minutes using an MTS-device, RSA was used to calculate 3-D cup migrations. Next a lever-out test of the cup was done in a displacement (16¡/minute) controlled test. After testing CT scans of the cups were made to quantify cement layer roughness, as a measurement of cement penetration. Results: Cups reconstructed with large washed bone grafts migrated signiþcantly less during mechanical compression and withstood signiþcantly higher lever-out moments. Washing improved cup stability for both small and large grafts, however only signiþcantly for the larger bone grafts. A signiþcantly higher cement penetration depth was found for all large bone grafts, washing did not inßuence cement penetration depth. Discussion: For acetabular revisions with bone impaction grafting we recommend to use large washed bone grafts.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_I | Pages 12 - 13
1 Jan 2004
Stolk J Verdonschot N Huiskes R
Full Access

Proper pre-clinical testing of cemented THA implants may help to prevent bad implants from entering the market. Within the frame of a multinational EU-program, a finite element (FE) simulation was developed, for FE-based pre-clinical testing of cemented THA stems against the damage accumulation failure scenario. The simulation allows monitoring of cement crack formation and implant migration in cemented THA reconstructions. The current study is concerned with the clinical validation of the test. The damage accumulation failure scenario was simulated for four cemented hip stems, with well-known survival rates. The question was: Can the FE simulation rank the stems according to their clinical survival rates?

Four stems were analysed: the Lubinus SPII, the Exeter, the Charnley and the Mueller Curved. The Swedish hip register [1] reports survival rates of 4, 5, 8 and 13%, respectively, at 10 years after surgery. Four FE models were created, representing cemented THA reconstructions with the four stems in composite femurs. The stem-cement interfaces were unbonded (m = 0.25). A loading history was applied to the models, representing 20 million cycles of alternating walking and stair climbing. Using a 3D continuum damage mechanics approach, the damage accumulation and creep processes in the cement, and subsequent prosthetic migration were monitored.

The Mueller C. produced a considerably higher number of cement cracks than the other three stems. Cracks were formed around the entire stem. The cracked zones often extended over the thickness of the mantle. The Charnley performed better, with a lower number of cracks. Proximo-distal damage pathways were formed, although at a much lower rate than around the Mueller C. The Exeter performed better. Full thickness crack zones were produced only in the proximo-medial region. The Lubinus performed best, with the lowest number of cement cracks. No full thickness cracks were formed. Concerning migration, the Exeter migrated more than the other stems. From the collared implants, the Lubinus SPII showed the lowest migration values.

When considering the number of cement cracks produced in the simulation, the ranking of the stems would be, from superior to inferior: Lubinus SPII, Exeter, Charnley, Mueller Curved. This ranking corresponds to a ranking based on clinical survival rates. The stems behaved according to their design concepts, with the highest migration values for the Exeter stem. In conclusion, the FE simulations produced a clinically valid ranking of four cemented THA implants. This corroborates the use of the FE simulation for pre-clinical testing purposes.


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 3 | Pages 436 - 444
1 Apr 2000
van Loon CJM de Waal Malefijt MC Buma P Stolk PWT Verdonschot N Tromp AM Huiskes R Barneveld A

The properties of impacted morsellised bone graft (MBG) in revision total knee arthroplasty (TKA) were studied in 12 horses. The left hind metatarsophalangeal joint was replaced by a human TKA. The horses were then randomly divided into graft and control groups. In the graft group, a unicondylar, lateral uncontained defect was created in the third metatarsal bone and reconstructed using autologous MBG before cementing the TKA. In the control group, a cemented TKA was implanted without the bone resection and grafting procedure. After four to eight months, the animals were killed and a biomechanical loading test was performed with a cyclic load equivalent to the horse’s body-weight to study mechanical stability. After removal of the prosthesis, the distal third metatarsal bone was studied radiologically, histologically and by quantitative and micro CT.

Biomechanical testing showed that the differences in deformation between the graft and the control condyles were not significant for either elastic or time-dependent deformations. The differences in bone mineral density (BMD) between the graft and the control condyles were not significant. The BMD of the MBG was significantly lower than that in the other regions in the same limb. Micro CT showed a significant difference in the degree of anisotropy between the graft and host bone, even although the structure of the area of the MBG had trabecular orientation in the direction of the axial load. Histological analysis revealed that all the grafts were revascularised and completely incorporated into a new trabecular structure with few or no remnants of graft. Our study provides a basis for the clinical application of this technique with MBG in revision TKA.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 6 | Pages 1052 - 1057
1 Nov 1999
Giesen EBW Lamerigts NMP Verdonschot N Buma P Schreurs BW Huiskes R

The use of impacted, morsellised bone grafts has become popular in revision total hip arthroplasty (THA). The initial stability of the reconstruction and the effectiveness of any subsequent process of revitalisation and incorporation will depend on the mechanical integrity of the graft. Our aim in this study was to document the time-dependent mechanical properties of the morsellised graft. This information is useful in clinical application of the graft, in studies of migration of the implant and in the design of the joint.

We used 16 specimens of impacted, morsellised cancellous bone from the sternum of goats to assess the mechanical properties by confined compression creep tests. Consideration of the graft material as a porous, permeable solid, filled with fluid, allowed determination of the compressive modulus of the matrix, and its permeability to fluid flow.

In all specimens the compression tests showed large, irreversible deformations, caused by flow-independent creep behaviour as a result of rolling and sliding of the bone chips. The mean permeability was 8.82 *10−12 m4/Ns (SD 43%), and the compressive modulus was 38.7 MPa (SD 34%). No correlation was found between the apparent density and the permeability or between the apparent density and the compressive modulus. The irreversible deformations in the graft could be captured by a creep law, for which the parameters were quantified.

We conclude that in clinical use the graft is bound to be subject to permanent deformation after operation. The permeability of the material is relatively high compared with, for example, human cartilage. The confined compression modulus is relatively low compared with cancellous bone of the same apparent density. Designs of prostheses used in revision surgery must accommodate the viscoelastic and permanent deformations in the graft without causing loosening at the interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 665 - 669
1 Jul 1997
Verdonschot N Huiskes R

It has been suggested that the endurance of cemented femoral reconstructions in total hip arthroplasty is affected by the creep of acrylic cement, but it is not known to what extent cement creeps under loading conditions in vivo, or how this affects load transfer. We have simulated the long-term creep properties of acrylic cement in finite-element models of femoral stem constructs and analysed their effects. We investigated whether subsidence rates measured in vivo could be explained by creep of acrylic cement, and if polished, unbonded, stems accommodated creep better than bonded stems.

Our findings showed that polished prostheses subsided only about 50 μm as a result of cement creep. The long-term prosthetic subsidence rates caused by creep of acrylic cement are therefore very small and do not explain the excessive migration rates which have sometimes been reported. Cement creep did, however, relax cement stresses and create a more favourable stress distribution at the interfaces. These trends were found around both the bonded and unbonded stems. Our results did not confirm that polished, unbonded, stems accommodated creep better than bonded stems in terms of cement and interface stress patterns.