Full Access

General Orthopaedics

Improvement of Primary Stability of Uncemented Femoral Knee Prosthesis by an Advanced Rough CoCrMo Porous Surface Coating: A Cadaveric Study

International Society for Technology in Arthroplasty (ISTA)



Abstract

The effect of an advanced porous surface morphology on the mechanical performance of an uncemented femoral knee prosthesis was investigated. Eighteen implants were inserted and then pushed-off from nine paired femurs (Left legs: advanced surface coating; right legs: PorocoatĀ® surface coating as baseline). Bone mineral density (BMD) and anteroposterior dimension were measured, which both were not significantly different between groups. The insertion force was not significantly different, but push-off force was significantly higher in the advanced surface coating group (P = 0.007). BMD had direct relationship with the insertion force and push-off force (p < 0.001). The effect of surface morphology on implant alignment was very small. We suggest that the surface properties create a higher frictional resistance thereby providing a better inherent stability of implants featuring the advanced surface coating.


*Email: