Objectives. Our study aimed to examine if a mobile-bearing
There remains a lack of data on the reliability of methods to
estimate tibial coverage achieved during total knee replacement.
In order to address this gap, the intra- and interobserver reliability
of a three-dimensional (3D) digital templating method was assessed
with one symmetric and one asymmetric prosthesis design. A total of 120 template procedures were performed according to
specific rotational and over-hang criteria by three observers at
time zero and again two weeks later. Total and sub-region coverage
were calculated and the reliability of the templating and measurement
method was evaluated.Objectives
Methods
Objectives. The purpose of this study was to examine the effect of posterior
cruciate ligament (PCL) retention, PCL recession, and PCL excision
during cruciate-retaining total knee replacement. Methods. A total of 3018 anatomic graduated component
Objective. Mortality rates reported by the National Joint Registry for England
and Wales (NJR) were higher following cemented
The aim of this study was to examine whether tourniquet use can improve perioperative blood loss, early function recovery, and pain after primary total knee arthroplasty (TKA) in the setting of multiple-dose intravenous tranexamic acid. This was a prospective, randomized clinical trial including 180 patients undergoing TKA with multiple doses of intravenous tranexamic acid. One group was treated with a tourniquet during the entire procedure, the second group received a tourniquet during cementing, and the third group did not receive a tourniquet. All patients received the same protocol of intravenous tranexamic acid (20 mg/kg) before skin incision, and three and six hours later (10 mg/kg). The primary outcome measure was perioperative blood loss. Secondary outcome measures were creatine kinase (CK), CRP, interleukin-6 (IL-6), visual analogue scale (VAS) pain score, limb swelling ratio, quadriceps strength, straight leg raising, range of motion (ROM), American Knee Society Score (KSS), and adverse events.Aims
Methods
Posterior condylar offset (PCO) and posterior tibial slope (PTS) are critical factors in total knee arthroplasty (TKA). A computational simulation was performed to evaluate the biomechanical effect of PCO and PTS on cruciate retaining TKA. We generated a subject-specific computational model followed by the development of ± 1 mm, ± 2 mm and ± 3 mm PCO models in the posterior direction, and -3°, 0°, 3° and 6° PTS models with each of the PCO models. Using a validated finite element (FE) model, we investigated the influence of the changes in PCO and PTS on the contact stress in the patellar button and the forces on the posterior cruciate ligament (PCL), patellar tendon and quadriceps muscles under the deep knee-bend loading conditions.Objectives
Methods
The length of the tourniquet time during total knee arthroplasty (TKA) is related to the incidence of post-operative deep vein thrombosis (DVT). Our aim in this study was to investigate the effect of the early release of the tourniquet on the incidence of DVT in patients undergoing TKA. A total of 200 patients who underwent TKA between November 2015 and November 2016 were prospectively enrolled. The tourniquet was inflated before surgery and released immediately after the introduction of the components (early release group). This group was compared with a retrospective cohort of 200 primary TKAs, in which the tourniquet was released after the dressings had been applied (late release group). The presence of a DVT was detected using bilateral lower limb ultrasonography. Peri-operative clinical and follow-up data were collected for analysis.Objectives
Methods
Malrotation of the femoral component can result in post-operative complications in total knee arthroplasty (TKA), including patellar maltracking. Therefore, we used computational simulation to investigate the influence of femoral malrotation on contact stresses on the polyethylene (PE) insert and on the patellar button as well as on the forces on the collateral ligaments. Validated finite element (FE) models, for internal and external malrotations from 0° to 10° with regard to the neutral position, were developed to evaluate the effect of malrotation on the femoral component in TKA. Femoral malrotation in TKA on the knee joint was simulated in walking stance-phase gait and squat loading conditions.Objectives
Materials and Methods
A patient-centred approach, usually achieved through shared decision
making, has the potential to help improve decision making around
knee arthroplasty surgery. However, such an approach requires an
understanding of the factors involved in patient decision making.
This review’s objective is to systematically examine the qualitative literature
surrounding patients’ decision making in knee arthroplasty. A systematic literature review using Medline and Embase was conducted
to identify qualitative studies that examined patients’ decision
making around knee arthroplasty. An aggregated account of what is
known about patients’ decision making in knee arthroplasties is
provided.Objectives
Methods
Because there have been no standard methods to determine pre-operatively
the thickness of resection of the proximal tibia in unicompartmental
knee arthroplasty (UKA), information about the relationship between
the change of limb alignment and the joint line elevation would
be useful for pre-operative planning. The purpose of this study
was to clarify the correlation between the change of limb alignment
and the change of joint line height at the medial compartment after
UKA. A consecutive series of 42 medial UKAs was reviewed retrospectively.
These patients were assessed radiographically both pre- and post-operatively
with standing anteroposterior radiographs. The thickness of bone
resection at the proximal tibia and the distal femur was measured
radiographically. The relationship between the change of femorotibial
angle (δFTA) and the change of joint line height, was analysed.Objectives
Methods
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
This systematic review and meta-analysis was conducted to determine
the mid- to long-term clinical outcomes for a medial-pivot total
knee replacement (TKR) system. The objectives were to synthesise
available survivorship, Knee Society Scores (KSS), and reasons for
revision for this system. A systematic search was conducted of two online databases to
identify sources of survivorship, KSS, and reasons for revision.
Survivorship results were compared with values in the National Joint
Registry of England, Wales, and Northern Ireland (NJR).Objectives
Methods
We are entering a new era with governmental bodies
taking an increasingly guiding role, gaining control of registries,
demanding direct access with release of open public information
for quality comparisons between hospitals. This review is written
by physicians and scientists who have worked with the Swedish Knee
Arthroplasty Register (SKAR) periodically since it began. It reviews
the history of the register and describes the methods used and lessons
learned. Cite this article:
The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score.Objectives
Methods
Lower limb muscle power is thought to influence outcome following
total knee replacement (TKR). Post-operative deficits in muscle
strength are commonly reported, although not explained. We hypothesised
that post-operative recovery of lower limb muscle power would be
influenced by the number of satellite cells in the quadriceps muscle at
time of surgery. Biopsies were obtained from 29 patients undergoing TKR. Power
output was assessed pre-operatively and at six and 26 weeks post-operatively
with a Leg Extensor Power Rig and data were scaled for body weight.
Satellite cell content was assessed in two separate analyses, the
first cohort (n = 18) using immunohistochemistry and the second
(n = 11) by a new quantitative polymerase chain reaction (q-PCR)
protocol for Pax-7 (generic satellite cell marker) and Neural Cell
Adhesion Molecule (NCAM; marker of activated cells).Objectives
Methods