Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 65 - 65
1 May 2017
Alzahrani M Cota A Alkhelaifi K Harvey E
Full Access

Background

Open reduction and internal fixation using plate osteosynthesis for midshaft clavicle fractures is often associated with hardware prominance. Although clinical studies have suggested a role for the use of thinner 2.7mm plates as a means of increasing cosmetic acceptability this still remains an area of controversy. We investigated the effect of plate size (2.7mm vs. 3.5mm), plate treatment (annealed vs. cold worked) and number of screws on the stiffness and yield point.

Methods

Twenty-four synthetic clavicles were randomly divided into four treatment groups - Synthes (Synthes, Paoli, PA) 2.7mm cold-worked calcaneal reconstruction plate with six or eight bicortical screws; 3.5mm LCP reconstruction plate (RP) and 3.5mm LCP pre-contoured superior-anterior clavicle plate (PCSA). After measuring the baseline stiffness of the intact specimens, all clavicles were plated, a wedge-shaped inferior defect was created and testing performed using a cantilever-bending model. Statistical analysis was performed using one-way ANOVA with Tukey's multiple comparison test with significance set at a P value <0.05.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 107 - 107
2 Jan 2024
Pastor T Zderic I Berk T Souleiman F Vögelin E Beeres F Gueorguiev B Pastor T
Full Access

Recently, a new generation of superior clavicle plates was developed featuring the variable-angle locking technology for enhanced screw positioning and optimized plate-to-bone fit design. On the other hand, mini-fragment plates used in dual plating mode have demonstrated promising clinical results. However, these two bone-implant constructs have not been investigated biomechanically in a human cadaveric model. Therefore, the aim of the current study was to compare the biomechanical competence of single superior plating using the new generation plate versus dual plating with low-profile mini-fragment plates. Sixteen paired human cadaveric clavicles were assigned pairwise to two groups for instrumentation with either a 2.7 mm Variable Angle Locking Compression Plate placed superiorly (Group 1), or with one 2.5 mm anterior plate combined with one 2.0 mm superior matrix mandible plate (Group 2). An unstable clavicle shaft fracture AO/OTA15.2C was simulated by means of a 5 mm osteotomy gap. All specimens were cyclically tested to failure under craniocaudal cantilever bending, superimposed with bidirectional torsion around the shaft axis and monitored via motion tracking. Initial stiffness was significantly higher in Group 2 (9.28±4.40 N/mm) compared to Group 1 (3.68±1.08 N/mm), p=0.003. The amplitudes of interfragmentary motions in terms of craniocaudal and shear displacement, fracture gap opening and torsion were significantly bigger over the course of 12500 cycles in Group 1 compared to Group 2; p≤0.038. Cycles to 2 mm shear displacement were significantly lower in Group 1 (22792±4346) compared to Group 2 (27437±1877), p=0.047. From a biomechanical perspective, low-profile 2.5/2.0 dual plates demonstrated significantly higher initial stiffness, less interfragmentary movements, and higher resistance to failure compared to 2.7 single superior variable-angle locking plates and can therefore be considered as a useful alternative for diaphyseal clavicle fracture fixation especially in unstable fracture configurations


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 18 - 18
11 Apr 2023
Kühl J Gorb S Klüter T Naujokat H Seekamp A Fuchs S
Full Access

Critical-sized bone defects can result from trauma, inflammation, and tumor resection. Such bone defects, often have irregular shapes, resulting in the need for new technologies to produce suitable implants. Bioprinting is an additive manufacturing method to create complex and individualised bone constructs, which can already include vital cells. In this study, we established an extrusion-based printing technology to produce osteoinductive scaffolds based on polycaprolactone (PCL) combined with calcium phosphate, which is known to induce osteogenic differentiation of stem cells. The model was created in python based on the signed distance functions. The shape of the 3D model is a ring with a diameter of 20 mm and a height of 10 mm with a spongiosa-like structure. The interconnected irregular pores have a diameter of 2 mm +/− 0.2 mm standard deviation. Extrusion-based printing was performed using the BIO X6. To produce the bioink, PCL (80 kDa) was combined with calcium phosphate nanopowder (> 150 nm particle size) under heating. After printing, 5 × 10. 6. hMSC were seeded on the construct using a rotating incubator. We were able to print a highly accurate ring construct with an interconnected pore structure. The PCL combined with calcium phosphate particles resulted in a precise printed construct, which corresponded to the 3D model. The bioink containing calcium phosphate nanoparticles had a higher printing accuracy compared to PCL alone. We found that hMSC cultured on the construct settled in close proximity to the calcium phosphate particles. The hMSC were vital for 22 days on the construct as demonstrated by life/dead staining. The extrusion printing technology enables to print a mechanically stable construct with a spongiosa-like structure. The porous PCL ring could serve as an outer matrix for implants, providing the construct the stability of natural bone. To extend this technology and to improve the implant properties, a biologised inner structure will be integrated into the scaffold in the future


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 99 - 99
14 Nov 2024
Umrath F Liang C Jud S Alexander D Danalache M
Full Access

Introduction. Osteoarthritis (OA) often results from joint misloading, which affects chondrocyte calcium signaling through mechano-sensitive receptors such as Piezo1, -2, and TRPV4. Activation of Piezo1, especially under inflammatory conditions, can trigger premature chondrocyte apoptosis. Intra-articular glucocorticoid therapy, while beneficial against inflammation and pain in osteoarthritis, may induce oxidative stress and chondrotoxicity at higher doses. This study aims to assess the effects of glucocorticoids, particularly triamcinolone, on chondrocyte elasticity and mechanosignaling. Method. Chondrocytes isolated from articular condyles obtained from patients undergoing knee replacement surgery (n= 5) were cultured for 7 days in triamcinolone acetonide (TA) at different concentrations (0.2µM – 2mM). Cytoskeletal changes were assessed by F-actin labeling. Cell elasticity was measured using atomic force microscopy (AFM). Labeling cells (n=6 patients) with the calcium-sensitive dye (Fluo-4) enabled monitoring changes in intracellular calcium fluorescence intensity during guided single-cell mechanical indentation (500 nN) by AFM. Result. Cell exposure to 2 mM TA led to cell death and crystallization of TA in the cell culture media. However, the concentration of TA for intra-articular application is 46 times higher at 92.1 mM (40 mg/ml). The maximal pharmacological effect on viable cells was observed at 0.2 mM. AFM results showed a significant decrease of elasticity (p<0.001), alongside significantly higher calcium intensities both prior to and during mechanical stimulation in the TA-treated samples (p<0.05). Conclusion. Administration of TA significantly impacts the mechanical properties of chondrocytes, reducing cellular elasticity while simultaneously enhancing calcium-dependent mechanosensitivity. This data suggests a correlation between glucocorticoid-induced changes in cell elasticity and cell mechanosensitivity. Finding ways to minimize the effect of glucocorticoids on cell mechanosensitivity could help to make future therapies safer and reduce side effects


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 16 - 16
17 Nov 2023
Youssef A Pegg E Gulati A Mangwani J Brockett C Mondal S
Full Access

Abstract. Objectives. The fidelity of a 3D model created using image segmentation must be precisely quantified and evaluated for the model to be trusted for use in subsequent biomechanical studies such as finite element analysis. The bones within the ankle joint vary significantly in size and shape. The purpose of this study was to test the hypothesis that the accuracy and reliability of a segmented bone geometry is independent of the particular bone being measured. Methods. Computed tomography (CT) scan data (slice thickness 1 mm, pixel size 808±7 µm) from three anonymous patients was used for the development of the ankle geometries (consisting of the tibia, fibula, talus, calcaneus, and navicular bones) using Simpleware Scan IP software (Synopsys, Exeter, UK). Each CT scan was segmented 4 times by an inexperienced undergraduate, resulting in a total of 12 geometry assemblies. An experienced researcher segmented each scan once, and this was used as the ‘gold standard’ to quantify the accuracy. The solid bone geometries were imported into CAD software (Inventor 2023, Autodesk, CA, USA) for measurement of the surface area and volume of each bone, and the distances between bones (tibia to talus, talus to navicular, talus to calcaneus, and tibia to fibula) were carried out. The intra-class coefficient (ICC) was used to assess intra-observer reliability. Bland Altman plots were employed as a statistical measure for criteria validity (accuracy) [1]. Results. The average ICC score was 0.93, which is regarded as a high reliability score for an inexperienced user. The talus to navicular and talus to tibia separations, which had the smallest distances, showed a slight decrease in reliability and this was observed for all separations shorter than 2 mm. According to the Bland-Altman plots, more than 95% of the data points were inside the borders of agreement, which is an excellent indication of accuracy. The bias percentage (average error percentage) varied between 1% and 4% and was constant across all parameters, with the proportion rising for short distance separations. Conclusions. The current study demonstrates that an inexperienced undergraduate, with access to software manuals, can segment an ankle CT scan with excellent reliability. The present study also concluded that all five bones were segmented with high levels of accuracy, and this was not influenced by bone volume or type. The only factor found to influence the reliability was the magnitude of distance between bones, where if this was smaller than 2 mm it reduced the reliability, indicating the influence of CT scan resolution on the segmentation reliability. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 78 - 78
11 Apr 2023
Vind T Petersen E Lindgren L Sørensen O Stilling M
Full Access

The Pivot-shift test is a clinical test for knee instability for patinets with Anterior cruciate ligament (ACL), however the test has low inter-observer reliability. Dynamic radiostereometry (dRSA) imaging is a highly precise method for objective evaluation of joint kinematics. The purpose of the study was to quantify precise knee kinematics during Pivot-shift test by use of the non-invasive dynamic RSA imaging. Eight human donor legs with hemipelvis were evaluated. Ligament lesion intervention of the ACL was performed during arthroscopy and anterolateral ligament (ALL) section was performed as a capsular incision. Pivot-shift test examination was recorded with dRSA on ligament intact knees, ACL-deficient knees and ACL+ALL-deficient knees. A Pivot-shift pattern was identifyable after ligament lesion as a change in tibial posterior drawer velocity from 7.8 mm/s in ligament intact knees, to 30.4 mm/s after ACL lesion, to 35.1 mm/s after combined ACL-ALL lesion. The anterior-posterior drawer excursion increased from 2.8 mm in ligament intact knees, to 7.2 mm after ACL lesion, to 7.6 mm after combined lesion. Furthermore a change in tibial rotation was found, with increasing external rotation at the end of the pivot-shift motion going from intact to ACL+ALL-deficient knees. This experimental study demonstrates the feasibility of RSA to objectively quantify the kinematic instability patterns of the knee during the Pivot-shift test. The dynamic parameters found through RSA displayed the kinematic changes from ACL to combined ACL-ALL ligament lesion


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_14 | Pages 25 - 25
1 Dec 2022
Spina G Napoleone F Mancuso C Gasparini G Mercurio M Familiari FF
Full Access

Magnetic resonance imaging (MRI) is the gold standard for the diagnosis of the pathologies affecting the glenohumeral joint and the rotator cuff diseases. MRI allows to highlight anatomic discontinuities of both muscles and tendons. However, MRI diagnostic accuracy has not proven to be highly sensitive in distinguishing between a partial-thickness tear and a full-thickness rotator cuff tear. The purpose of this study was to determine if MRI under axial traction can be helpful in increasing MRI sensitivity to identify partial-thickness rotator cuff tears. The study included 10 patients (4 males and 6 females) who had clinical examination and MRI suggesting a partial-thickness rotator cuff tear. They were candidates for shoulder arthroscopy because of persistent symptoms after at least three months of conservative treatment. The patients underwent a new MRI (under axial traction: MRI-AT) with a 4-kg weight applied to the affected arm. Then the patients underwent arthroscopy to confirm the diagnosis. Patients with a suspected full-thickness rotator cuff tear were excluded from the study. Patients’ average age was 52.4 years, and the dominant side was affected in 77.7% of the cases. Preoperative Constant-Murley Score was 57. MRI-AT showed that 3 patients were affected by a complete tear of the rotator cuff, 3 patients by a partial-thickness rotator cuff tear and 4 patients had no lesion. The analysis of data showed that: under axial traction the subacromial space increased by 0,2 mm (P value = 0,001075), the superior glenohumeral space decreased by 2.4 mm (P value = 0,07414), the inferior glenohumeral space increased by 0.3 mm (P value = 0,02942), the acromial angle decreased by 1.9° (P value = 0,0002104) and the acromion-glenohumeral angle decreased by 0.3° (P-value = 0,01974). Two experienced evaluators analyzed previous standard MRI and MRI-AT scans in a double-blinded fashion, with inter-rater evaluation of all the images and measures. Intraclass correlation coefficient (ICC) has been utilized to assess the reliability of the measures performed by different operators. ICC always resulted in more than 0.7, showing a high concordance among values in the same group. A comparative evaluation between standard MRI and MRI-AT has been conducted to highlight possible discrepancies and this has been compared to intraoperative findings. Concordance of the values was 89% between standard MRI and MRI-AT and 100% between MRI under axial traction and intraoperative findings. This study showed a high correlation between the diagnosis achieved with MRI-AT and the intraoperative arthroscopic findings. The use of MRI-AT in clinical practice may improve the diagnostic sensitivity of this method to detect a partial-thickness rotator cuff tear


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 39 - 39
4 Apr 2023
Lim W Lie D Chou S Lie H Yew A
Full Access

This study aims to investigate the mechanical properties of a rotator cuff tear repaired with a polypropylene interposition graft in an ovine infraspinatus ex-vivo model. Twenty fresh shoulders from skeletally mature sheep were used in this study. A tear size of 20 mm from the tendon joint was created in the infraspinatus tendon to simulate a large tear in fifteen specimens. This was repaired with a polypropylene mesh used as an interposition graft between the ends of the tendon. Eight specimens were secured with mattress stitches while seven were secured to the remnant tendon on the greater tuberosity side by continuous stitching. Remaining five specimens with an intact tendon served as a control group. All specimens underwent cyclic loading with a universal testing machine to determine the ultimate failure load and gap distance. Gap distance increased with progressive cyclic loading through 3000 cycles for all repaired specimens. Mean gap distance after 3000 cycles for both continuous and mattress groups are 1.7 mm and 4.2 mm respectively (P = .001). Significantly higher mean ultimate failure load was also observed with 549.2 N in the continuous group, 426.6 N in the mattress group and 370 N in the intact group. The use of a polypropylene mesh as an interposition graft for large irreparable rotator cuff tears is biomechanically suitable and results in a robust repair that is comparable to an intact rotator cuff tendon. When paired with a continuous suturing technique, it demonstrates significantly resultant superior biomechanical properties that may potentially reduce re-tear rates after repairing large or massive rotator cuff tears


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 19 - 19
17 Apr 2023
Niessen L Wendlandt R Schulz A
Full Access

A promising application of Mesenchymal stem cells (MSCs) is the treatment of non-unions. Substituting bone grafts, MSCs are directly injected into the fracture gap. High cell viability seems to be a prerequisite for therapeutic success. Administration of the MSCs via injection creates shear stresses possibly damaging or destroying the cells. Aim of this study was to investigate the effect of the injection process on cell viability. MSCs were isolated and cultivated from femoral tissue of five subjects undergoing arthroplasty. Prior to injection, the cells were identified as MSCs. After dissolving to a concentration of 1 Million cells/ml, 1 ml of the suspension was injected through a cannula of 200 mm length and 2 mm diameter (14 G) with flow rates of 38 and 100 ml/min. The viability of the MSCs at different flow rates was evaluated by staining to detect the healthy cell fraction. It was analyzed statistically against a control group via the Kruskal-Wallis-test and for equivalence via the TOST procedure. Significance level was set to 5 %, equivalence margin to 20 %. The healthy cell fraction of the control group was 85.88 ± 2.98 %, 86.04 ± 2.53 % at 38 ml/min and 85.48 ± 1.64 % at 100 ml/min. There was no significant difference between the fraction of healthy cells (p = 0.99) for different volume flows, but a significant equivalence between the control group and the two volume flows (38 ml/min: p = 0.002, 100 ml/min: p = 0.001). When injecting MSC solutions, e.g. into a non-union, the viability of the injected cells does not deterioriate significant. The injecting technique is therefore feasible


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 84 - 84
4 Apr 2023
Gehweiler D Pastor T Beeres F Kastner P Migliorini F Nebelung S Scaglioni M Souleiman F Link B Babst R Gueorguiev B Knobe M
Full Access

Helical plates potentially bypass the medial neurovascular structures of the thigh. Recently, two plate designs (90°- and 180°-helix) proved similar biomechanically behaviour compared to straight plates. Aims of this study were: (1) Feasibility of MIPO-technique with 90°- and 180°-helical plates on the femur, (2) Assessment of distances to adjacent anatomical structures at risk, (3) Comparison of these distances to using medial straight plates instead, (4) Correlation of measurements performed in anatomic dissection with CT-angiography. MIPO was performed in ten cadaveric femoral pairs using either a 90°-helical 14-hole-LCP (Group1) or a 180°-helical 15-hole-LCP-DF (Group2). CT angiography was used to evaluate the distances between the plates and the femoral arteries as well as the distances between the plates and the perforators. Subsequently, the specimens were dissected, and the distances were determined again manually. Finally, all helical plates were removed, and all measurements were repeated after application of straight medial plates (Group3). Closest overall distances between plates and femoral arteries were 15 mm (11 − 19 mm) in Group1, 22 mm (15 − 24 mm) in Group2 and 6 mm (1 − 8 mm) in Group3 with a significant difference between Group1 and Group3 (p < 0.001). Distances to the nearest perforators were 24 mm (15 − 32 mm) in Group1 and 2 mm (1 − 4 mm) in Group2. Measurement techniques (visual after surgery and CT-angiography) demonstrated a strong correlation of r. 2. = 0.972 (p < 0.01). MIPO with 90°- and 180°-helical plates is feasible and safe. Attention must be paid to the medial neurovascular structures with 90°-helical implants and to the proximal perforators with 180°-helical implants. Helical implants can avoid medial neurovascular structures compared to straight plates although care must be taken during their distal insertion. Measurements during anatomical dissection correlate with CT-angiography


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 121 - 121
4 Apr 2023
Kale S Mehra S Gunjotikar A Patil R Dhabalia P Singh S
Full Access

Osteochondromas are benign chondrogenic lesions arising on the external surface of the bone with aberrant cartilage (exostosis) from the perichondral ring that may contain a marrow cavity also. In a few cases, depending on the anatomical site affected, different degrees of edema, redness, paresthesia, or paresis can take place due to simple contact or friction. Also, depending on their closeness to neurovascular structures, the procedure of excision becomes crucial to avoid recurrence. We report a unique case of recurrent osteochondroma of the proximal humerus enclosing the brachial artery which makes for an important case and procedure to ensure that no relapse occurs. We report a unique case of a 13-year-old female who had presented with a history of pain and recurrent swelling for 5 years. The swelling size was 4.4 cm x 3.7 cm x 4 cm with a previous history of swelling at the same site operated in 2018. CT reports were suggestive of a large well defined broad-based exophytic diaphyseal lesion in the medial side of the proximal humerus extending posteriorly. Another similar morphological lesion measuring approximately 9 mm x 7 mm was noted involving the posterior humeral shaft. The minimal distance between the lesion and the brachial artery was 2 mm just anterior to the posterio-medial growth. Two intervals were made, first between the tumor and the neurovascular bundle and the other between the anterior tumor and brachial artery followed by exostosis and cauterization of the base. Proper curettage and excision of the tumor was done after dissecting and removing the soft tissue, blood vessels, and nerves so that there were very less chances of relapse. Post-operative X-ray was done and post 6 months of follow-up, there were no changes, and no relapse was observed. Thus, when presented with a case of recurrent osteochondroma of the proximal humerus, osteochondroma could also be in proximity to important vasculature as in this case enclosing the brachial artery. Thus, proper curettage and excision should be done in such cases to avoid recurrence


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 33 - 33
14 Nov 2024
Fallahy M Shaker F Ghanbari F Aslani MA Mohammadi S Behrouzieh S
Full Access

Introduction. Knee Osteoarthritis (KOA) is a prevalent joint disease requiring accurate diagnosis and prompt management. The condition occurs due to cartilage deterioration and bone remodeling. Ultrasonography has emerged as a promising modality for diagnosing KOA. Medial meniscus extrusion (MME), characterized by displacement of medial meniscus beyond the joint line has been recognized as a significant marker of KOA progression. This study aimed to explore potentials Ultrasound findings in timely detection of MME and compare it to magnetic resonance imaging (MRI) as a reference standard. Method. A comprehensive literature search was performed in 4 databases from inception to May 1 2024. Two independent reviewers, initiated screening protocols and selected the articles based on inclusion and exclusion criteria and then extracted the data. Meta-analysis was conducted using R 4.3.2 packages mada and metafor. Result. A total of 2500 articles from 4 databases was retrieved; however, following the application of inclusion and exclusion criteria 23 articles were finally extracted. These studies collectively encompassed a total of 777 patients with mean age of 53.2±7.4. The mean BMI calculated for patients was 28.31 ± 2.45. All patients underwent non-weight bearing knee ultrasonography in supine position with 0° flexion. The reported medial meniscus extrusion was 2.58 mm for articles using MRI and 2.65 mm for those using Ultrasound (MD: 0.05 ± 0.12, P= 0.65, I. 2. : 54%). Our meta-analysis revealed insignificant difference between US and MRI. (SMD: 0.03, 95% CI: -0.18 _0.23, P= 0.77, I. 2. : 56%) Meta analysis for diagnostic accuracy measures yielded a pooled sensitivity and specificity of 90.8% and 77% (95% CI: 84.2% – 94.8%, 35.5% – 95.3%, respectively, I. 2. : 44%). Conclusion. Our results indicate a close alignment in the accuracy of measurements obtained using Ultrasound modality. The narrow range suggests a minimal discrepancy in MME values between MRI and ultrasound, highlighting their comparable precision in diagnostic assessments


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 80 - 80
2 Jan 2024
Lerma-Juárez M Escudero-Duch C Serrano-Yamba R Moreno-García A Yus C Arruebo M Vilaboa N
Full Access

We have developed plasmonic fibrin-based hydrogels that incorporate gold nanoparticles which transduce incident near-infrared (NIR) light into heat. Human adenovirus serotype type-5 vectors encoding a firefly luciferase (fLuc) coding sequence driven by a heat-inducible promoter were incorporated into the hydrogels. Transmission electronic microscopic analysis revealed that the adenoviral vectors were associated to the fibrin fibers. In vitro experiments in which human cells were cultured with plasmonic hydrogels showed that the adenoviral vectors can diffuse from the hydrogels, transduce the cells, and stimulate heat-induced transgene expression upon NIR irradiation. The hydrogels were implanted in 4.2 mm drill hole defects generated in the humerus of male rabbits. Three days after implantation, the defects were NIR-irradiated. Six h later, the animals were euthanized and samples from the bone defect zone were processed for immunohistochemical analyses using a specific fLuc antibody. The results showed strong expression of fLuc in tissues surrounding the implants of NIR-irradiated rabbits, while non- irradiated animals exhibited negligible expression. We next aimed to use the temperature increase to induce the production of transgenic bone morphogenetic protein 6 (BMP-6), using safe gene switches that can provide tighter control of in vivo transgene expression than heat-inducible promoters. These switches are only activated by heat in the presence of rapamycin and maintain a high level of targeted transgene expression for several days after heat activation. Adenoviral vectors encoding the safe switches that control the expression of BMP-6 were incorporated to the composites. The resulting NIR-responsive hydrogels were implanted in the bone defects generated in rabbits and used as a platform to transduce host cells, generate local hyperthermia and stimulate BMP-6 production. Acknowledgements: This research was supported by grants RTI2018-095159-B-I00 and PID2021-126325OB-I00 (MCIN/AEI/10.13039/501100011033 and “ERDF A way of making Europe”), by grant P2022/BMD- 7406 (Regional Government of Madrid). M.A.L-J. is the recipient of predoctoral fellowship PRE2019-090430 (MCIN/AEI/10.13039/501100011033)


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 42 - 42
4 Apr 2023
Benca E van Knegsel K Zderic I Caspar J Strassl A Hirtler L Fuchssteiner C Gueorguiev B Windhager R Widhalm H Varga P
Full Access

Screw fixation is an established method for anterior cruciate ligament (ACL) reconstruction, although with a high rate of implant-related complications. An allograft system for implant fixation in ACL reconstruction, the Shark Screw ACL (surgebright GmbH) could overcome some of the shortcomings of bioabsorbable screws, such as foreign body reaction, need for implant removal and imaging artefacts. However, it needs to provide sufficient mechanical stability. Therefore, the aim of this study was to investigate the biomechanical stability, especially graft slippage, of the novel allograft system versus a conventional bioabsorbable interference screw (BioComposite Interference Screw; Arthrex Inc.) for tibial implant fixation in ACL reconstruction. Twenty-four paired human proximal tibiae (3 female, 9 male, 72.7 ± 5.6 years) underwent ACL reconstruction. The quadrupled semitendinosus and gracilis tendon graft were fixed in one specimen of each pair using the allograft fixation system Shak Screw ACL and the contralateral one using an interference screw. All specimens were cyclically loaded at 1 Hz with peak load levels monotonically increased from 50 N at a rate of 0.1 N/cycle until catastrophic failure. Relative movements of the graft versus the tibia were captured with a stereographic optical motion tracking system (Aramis SRX; GOM GmbH). The two fixation methods did not demonstrate any statistical difference in ultimate load at graft slippage (p = 0.24) or estimated survival at slippage (p = 0.06). Both, the ultimate load and estimated survival until failure were higher in the interference screw (p = 0.04, and p = 0.018, respectively). Graft displacement at ultimate load reached values of up to 7.2 mm (interference screw) and 11.3 mm (Shark Screw ACL). The allograft screw for implant fixation in ACL reconstruction showed similar behavior in terms of graft slippage compared to the conventional metal interference screw but underperformed in terms of ultimate load. However, the ultimate load may not be considered a direct indicator of clinical failure


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 15 - 15
4 Apr 2023
Luk J
Full Access

Many factors have been reported to affect the functional survival of OCA transplants, including chondrocyte viability at time of transplantation, rate and extent of allograft bone integration, transplantation techniques, and postoperative rehabilitation protocols and adherence. The objective of this study was to determine the optimal subchondral bone drilling technique by evaluating the effects of hole diameter on the material properties of OCAs while also considering total surface area for potential biologic benefits for cell and vascular ingrowth. Using allograft tissues that would be otherwise discarded in combination with deidentified diagnostic imaging (MRI and CT), a model of a large shell osteochondral allograft was recreated using LS-PrePost and FEBio based on clinically relevant elastic material properties for cortical bone, trabecular bone, cartilage, and hole ingrowth tissue. The 0.8 mesh size model consisted of 4 mm trabecular bone, 4 mm cortical bone, and 3 mm cartilage sections that summed to a cross-sectional area of 1600 mm2 (40 mm x 40 mm). Holes were modeled to be 4mm deep in relation to clinical practice where holes are drilled from the deep margin of subchondral trabecular bone to the cortical subchondral bone plate. To test the biomechanic variations between drill hole sizes, models with hole sizes pertinent to standard-of-care commercially available orthopaedic drill sizes of 1.1mm, 2.4 mm, or 4.0 mm holes were loaded across the top surface over a one second duration and evaluated for effective stress, effective strain, 1st principal strain, and 3rd principal strain in compressive conditions. Results measured effective stress and strain and 1st and 3rd principal strain increased with hole depth. The results of the present FEA modeling study indicate that the larger 4.0 mm diameter holes were associated with greater stresses and strains within OCA shell graft, which may render the allograft at higher risk for mechanical failure. Based on these initial results, the smaller diameter 2.4 mm and 1.1 mm holes will be further investigated to determine optimal number, configuration, and depth of subchondral drilling for OCA preparation for transplantation


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 16 - 16
4 Apr 2023
Luk J
Full Access

Despite the growing success of OCA transplantation in treating large articular cartilage lesions in multiple joints, revisions and failures still occur. While preimplantation subchondral drilling is intended to directly decrease allograft bioburden and has been associated with significant improvements in outcomes after OCA transplantation, the effects of size, number, and spacing of subchondral bone drill sites have not been fully evaluated. This study aimed to investigate the effects of drill size with or without pulse-lavage of OCA subchondral bone by quantifying remnant marrow elements using histomorphometry. With IRB and ACUC approvals, human and canine OCAs were acquired for research purposes. Portions of human tibial plateau OCAs acquired from AATB-certified tissue banks that would otherwise be discarded were recovered and sectioned into lateral and medial hemiplateaus (n=2 each) with a thickness of 7 mm. Canine femoral condyles and tibial plateaus were split into lateral and medial components with a thickness of 7 mm (n=8). Using our clinical preimplantation preparation protocol, holes were drilled into the subchondral bone of each condyle and hemiplateau OCA using either 1.6 mm OD or 3.2 mm OD drill bits from the cut surface to the cortical subchondral bone plate. One femoral condyle and one hemiplateau per drill bit size were pulse-lavaged while the corresponding OCAs were not. The mean total %-fill remaining marrow elements for each treatment group was calculated. Little to no quantifiable bone marrow element retention was noted to remain within the subchondral bone of human or canine OCA specimens after subchondral drilling of allograft bone with either drill bit size evaluated and with or without pulse-lavage. The %-fill was consistent across zones, ranging from 1-5%. This project was designed to provide a preliminary histologic evaluation of the effects of drill size on OCA preimplantation preparation efficacy based on amount of remaining bone marrow elements in human and canine femoral condyle and tibial plateau specimens. Based on these initial findings, choice of drill bit size for OCA subchondral drilling may need to be based on the associated biomechanical effects rather than effects on donor bone marrow element removal


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 30 - 30
14 Nov 2024
Schröder M Gens L Arens D Giger N Gehweiler D Nehrbass D Zderic I Zeiter S Stoddart M Wehrle E
Full Access

Introduction. Immunomodulation represents a novel strategy to improve bone healing in combination with low doses of bone morphogenetic growth factors like BMP-2. This study aims to investigate the effect and timing of monoclonal anti-IL-1ß antibody administration with 1μg BMP-2 on bone healing over 14 weeks in a rat femur segmental defect model. Method. 2 mm femoral defects were created in 22-27 weeks-old female Fischer F344 rats, internally fixed with a plate (animal license: GR/19/2022) using established protocols for analgesia and anesthesia. Animals (n=4/group) received either a collagen sponge, a collagen sponge+1μg BMP-2 (InductOs, Medtronic) or a collagen sponge+1μg BMP-2 with a monoclonal anti-IL-1ß antibody (BioXCell, 10 mg/ml), administered intravenously under anesthesia every third day until day 15, from day 0 or 3. In vivo micro-CT was performed after surgery and at 2, 3, 4, 6, 8, 10 and 14-weeks post-OP. Mechanical properties of the operated femurs were assessed by 4-point bending (Instron5866) and compared to contralateral femurs (one-way ANOVA, GraphPad Prism8). Histopathological analysis was performed semi-quantitatively on Giemsa-Eosin-stained sections (Olympus BX63) using a six-grade severity grading scale. Result. Operated femurs with BMP-2 reached an average stiffness of 91±37% of contralateral femurs, femurs in IL-1ß groups 105±11% (day 0) and 111±12% (day 3). Administration of anti-IL-1ß+1μg BMP-2 led to faster cortical bridging (3/4 femurs bridged by week 4 for day 0, 4/4 for day 3) than 1μg BMP-2 alone (0/4 by week 4). Micro-CT results confirmed histopathological evaluation, as collagen sponge alone led to non-union, complete bicortical bridging was observed for 3/4 femurs in the BMP-2 group and for 4/4 femurs in the IL-1β groups after 14 weeks. Conclusion. Anti-IL-1ß had a beneficial effect on late fracture healing with faster cortical bridging and new bone formation than 1μg BMP-2 alone. Acknowledgments. AO foundation. We thank Andrea Furter, Alisa Hangartner and Thomas Krüger for technical support


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 125 - 125
14 Nov 2024
Mungalpara N Kim S Baker H Lee C Shakya A Chen K Athiviraham A Koh J Elhassan B Maassen NH Amirouche F
Full Access

Introduction. Treatment strategies for irreparable Massive Rotator Cuff Tears (MRCTs) are debatable, especially for younger, active patients. Superior Capsular Reconstruction (SCR) acts as a static stabilizer, while Lower Trapezius Transfer (LTT) serves as a dynamic stabilizer. This study compares the biomechanical effectiveness of SCR and LTT, hypothesizing that their combination will enhance shoulder kinematics. Methods. Eight human shoulders from donors aged 55-75 (mean = 63.75 years), balanced for gender, averaging 219.5 lbs, were used. Rotator cuff and deltoid tendons were connected to force sensors through a pulley system, with the deltoid linked to a servohydraulic motor for dynamic force measurement. Results. From intact to MRCT, deltoid force was reduced by 28% (p = 0.023). LTT increased deltoid force by 27.25 (p = 0.166). SCR decreased deltoid force by 34% (p = 0.208). Combining LTT with SCR increased deltoid force by 32.57% compared to SCR (p = 0.023) and decreased it by 13.6% compared to LTT alone (p = 0.017). Combined LTT and SCR reduced deltoid force by 20.9% from the control (p = 0.001). Subacromial contact pressure rose by 15% in MRCT over intact, but LTT decreased it by 7.6%, achieving nearly 50% correction. SCR increased subacromial space volume, raising pressure by 6.5%. The humeral head translation (HHT) increases with MRCT, reaching 3.33 mm (SD = 0.95) at 0 degrees, compared to 2.24 mm (SD = 0.78) in the intact. LTT and the combined LTT + SCR significantly reduce HHT, with combined LTT + SCR achieving HHT of 2.24 mm (SD = 0.63) at 0 degrees, comparable to the control. Conclusion. Notable changes in deltoid force were observed. LTT outperformed the combined SCR and LTT in reducing deltoid force and subacromial peak pressure. Both SCR and LTT corrected HTT, with LTT being more effective. However, combining SCR and LTT optimally corrected HHT


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 102 - 102
1 Mar 2021
Tazawa R Minehara H Matsuura T Kawamura T Uchida K Inoue G Saito W Takaso M
Full Access

Segmental bone transport (SBT) using an external fixator is currently a standard treatment for large-diameter bone defects at the donor site with low morbidity. However, long-term application of the device is needed for bone healing. In addition, patients who received SBT treatment sometimes fail to show bone repair and union at the docking site, and require secondary surgery. The objective of this study was to investigate whether a single injection of recombinant human bone morphogenetic protein 2 (rhBMP-2)-loaded artificial collagen-like peptide gel (rhBMP-2/ACG) accelerates consolidation and bone union at the docking site in a mouse SBT model. Six-month-old C57BL/6J mice were reconstructed by SBT with external fixator that has transport unit, and a 2.0-mm bone defect was created in the right femur. Mice were divided randomly into four treatment groups with eight mice in each group, Group CONT (immobile control), Group 0.2mm/d, Group 1.0mm/d, and Group BMP-2. Mice in Group 0.2mm/d and 1.0mm/d, bone segment was moved 0.2 mm per day for 10 days and 1.0 mm per day for 2 days, respectively. Mice in Group BMP-2 received an injection of 2.0 μg of rhBMP-2 dissolved in ACG into the bone defect site immediately after the defect-creating surgery and the bone segment was moved 1.0 mm/day for 2 days. All animals were sacrificed at eight weeks after surgery. Consolidation at bone defect site and bone union at docking site were evaluated radiologically and histologically. At the bone defect site, seven of eight mice in Group 0.2mm/d and two of eight mice in Group 1.0mm/d showed bone union. In contrast, all mice in Group CONT showed non-union at the bone defect site. At the docking site, four of eight mice in Group 0.2 mm/d and three of eight mice in Group 1.0 mm/d showed non-union. Meanwhile, all mice in Group BMP-2 showed bone union at the bone defect and docking sites. Bone volume and bone mineral content were significantly higher in Group 0.2mm/d and Group BMP-2 than in Group CONT. HE staining of tissue from Group 0.2mm/d and Group BMP-2 showed large amounts of longitudinal trabecular bone and regenerative new bone at eight weeks after surgery at the bone defect site. Meanwhile, in Group CONT and Group 1.0mm/d, maturation of regenerative bone at the bone defect site was poor. Differences between groups were analyzed using one-way ANOVA and a subsequent Bonferroni's post-hoc comparisons test. P < 0.05 was considered significant. rhBMP-2/ACG combined with SBT may be effective for enhancing bone healing in large bone defects without the need for secondary procedures


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 85 - 85
1 Dec 2020
Stefanov A Ivanov S Zderic I Baltov A Rashkov M Gehweiler D Richards G Gueorguiev B Enchev D
Full Access

Treatment of comminuted intraarticular calcaneal fractures remains controversial and challenging. Anatomic reduction with stable fixation has demonstrated better outcomes than nonoperative treatment of displaced intraarticular fractures involving the posterior facet and anterior calcaneocuboid joint (CCJ) articulating surface of the calcaneus. The aim of this study was to investigate the biomechanical performance of three different methods for fixation of comminuted intraarticular calcaneal fractures. Comminuted calcaneal fractures, including Sanders III-AB fracture of the posterior facet and Kinner II-B fracture of the CCJ articulating calcaneal surface, were simulated in 18 fresh-frozen human cadaveric lower legs by means of osteotomies. The ankle joint, medial soft tissues and midtarsal bones along with the ligaments were preserved. The specimens were randomized according to their bone mineral density to 3 groups for fixation with either (1) 2.7 mm variable-angle locking anterolateral calcaneal plate in combination with one 4.5 mm and one 6.5 mm cannulated screw (Group 1), (2) 2.7 mm variable-angle locking lateral calcaneal plate (Group 2), or (3) interlocking calcaneal nail with 3.5 mm screws in combination with 3 separate 4.0 mm cannulated screws (Group 3). All specimens were biomechanically tested until failure under axial loading with the foot in simulated midstance position. Each test commenced with an initial quasi-static compression ramp from 50 N to 200 N, followed by progressively increasing cyclic loading at 2Hz. Starting from 200 N, the peak load of each cycle increased at a rate of 0.2 N/cycle. Interfragmentary movements were captured by means of optical motion tracking. In addition, mediolateral X-rays were taken every 250 cycles with a triggered C-arm. Varus deformation between the tuber calcanei and lateral calcaneal fragments, plantar gapping between the anterior process and tuber fragments, displacement at the plantar aspect of the CCJ articular calcaneal surface, and Böhler angle were evaluated. Varus deformation of 10° was reached at significantly lower number of cycles in Group 2 compared to Group 1 and Group 3 (P ≤ 0.017). Both cycles to 10° plantar gapping and 2 mm displacement at the CCJ articular calcaneal surface revealed no significant differences between the groups (P ≥ 0.773). Böhler angle after 5000 cycles (1200 N peak load) had significantly bigger decrease in Group 2 compared to both other groups (P ≤ 0.020). From biomechanical perspective, treatment of comminuted intraarticular calcaneal fractures using variable-angle locked plate with additional longitudinal screws or interlocked nail in combination with separate transversal screws seems to provide superior stability as opposed to variable-angle locked plating only