Aims. To report our experience with
Aims. To present a surgically relevant update of trunnionosis. . Materials and Methods. Systematic review performed April 2017. Results. Trunnionosis accounts for approximately 2% of the revision total
hip arthroplasty (THA) burden. Thinner (reduced flexural rigidity)
and shorter
Taper corrosion has been widely reported to be problematic for modular total hip arthroplasty implants. A simple and systematic method to evaluate taper damage with sufficient resolution is needed. We introduce a semiquantitative grading system for modular femoral tapers to characterize taper corrosion damage. After examining a unique collection of retrieved cobalt-chromium (CoCr) taper sleeves (n = 465) using the widely-used Goldberg system, we developed an expanded six-point visual grading system intended to characterize the severity, visible material loss, and absence of direct component contact due to corrosion. Female taper sleeve damage was evaluated by three blinded observers using the Goldberg scoring system and the expanded system. A subset (n = 85) was then re-evaluated following destructive cleaning, using both scoring systems. Material loss for this subset was quantified using metrology and correlated with both scoring systems.Aims
Methods
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
There is increasing global awareness of adverse
reactions to metal debris and elevated serum metal ion concentrations
following the use of second generation metal-on-metal total hip
arthroplasties. The high incidence of these complications can be
largely attributed to corrosion at the head-neck interface. Severe
corrosion of the taper is identified most commonly in association
with larger diameter femoral heads. However, there is emerging evidence
of varying levels of corrosion observed in retrieved components
with smaller diameter femoral heads. This same mechanism of galvanic
and mechanically-assisted crevice corrosion has been observed in
metal-on-polyethylene and ceramic components, suggesting an inherent
biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers
clinicians and researchers must understand regarding corrosion of
the taper, and its relevance to current orthopaedic practice. Cite this article:
Aims. The aim of this study was to compare the design of the generic
OptiStem XTR femoral stem with the established Exeter femoral stem. Materials and Methods. We obtained five boxed, as manufactured, implants of both designs
at random (ten in total). Two examiners were blinded to the implant
design and independently measured the mass, volume,
Objectives. The use of ceramic femoral heads in total hip arthroplasty (THA) has increased due to their proven low bearing wear characteristics. Ceramic femoral heads are also thought to reduce wear and corrosion at the head-stem junction with titanium (Ti) stems when compared with metal heads. We sought to evaluate taper damage of ceramic compared with metal heads when paired with cobalt chromium (CoCr) alloy stems in a single stem design. Methods. This retrieval study involved 48 total hip arthroplasties (THAs) with CoCr V40
Aims. We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Methods. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time. Results. Metal ion levels were repeated in 33 patients. When comparing the results of serum metal ion levels over time, regardless of head size, there was a significant increase in both cobalt and chromium levels (p < 0.001). Two patients with larger head sizes had undergone revision arthroplasty with evidence of
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity. Ceramic bearing
couples are low wearing and their use minimises the risk of subsequent
further revision due to the production of wear debris. A total of
165 revision hip replacements were performed, in which a polyethylene-lined acetabular
component was revised to a new acetabular component with a ceramic
liner, while retaining the well-fixed femoral component. A titanium
sleeve was placed over the used femoral
Aims. Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. Patients and Methods. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients. Results. The patients presented with pain, swelling, stiffness, or instability
and an inflammatory mass was confirmed radiologically. Macroscopic
material deposition on the
Objectives. Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. Methods. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons. Results. Evidence of corrosion was observed in 55% of hips. The median Goldberg taper corrosion score was 2 (1 to 4) and the annual rate of material loss at the taper was 0.084 mm. 3. /year (0 to 0.239). The median
Objectives. Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Materials and Methods. Ti-6Al-4V
Aims. Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the
Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of larger diameter heads as they decrease
the risk of dislocation following primary and revision THR. Highly cross-linked
polyethylene has allowed us to increase femoral head size, without
a marked increase in wear. However, the thin polyethylene liners
necessary to accommodate larger heads may increase the risk of liner
fracture and larger heads have also been implicated in causing soft-tissue
impingement resulting in groin pain. Larger diameter heads also
impart larger forces on the femoral
We report complications from the use of modular components in 20 hip replacements in 18 patients. Fifteen complications (in 13 patients) were related to failure of a modular interface after operation. Femoral head detachment from its
In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.Aims
Methods
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
Whitehouse MR, Endo M, Zachara S, Nielsen TO,
Greidanus NV, Masri BA, Garbuz DS, Duncan CP.Adverse local tissue reactions in metal-on-polyethylene
total hip arthroplasty due to
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
Periprosthetic femoral fracture (PPF) is a major complication following total hip arthroplasty (THA). Uncemented femoral components are widely preferred in primary THA, but are associated with higher PPF risk than cemented components. Collared components have reduced PPF rates following uncemented primary THA compared to collarless components, while maintaining similar prosthetic designs. The purpose of this study was to analyze PPF rate between collarless and collared component designs in a consecutive cohort of posterior approach THAs performed by two high-volume surgeons. This retrospective series included 1,888 uncemented primary THAs using the posterior approach performed by two surgeons (PKS, JMV) from January 2016 to December 2022. Both surgeons switched from collarless to collared components in mid-2020, which was the only change in surgical practice. Data related to component design, PPF rate, and requirement for revision surgery were collected. A total of 1,123 patients (59.5%) received a collarless femoral component and 765 (40.5%) received a collared component. PPFs were identified using medical records and radiological imaging. Fracture rates between collared and collarless components were analyzed. Power analysis confirmed 80% power of the sample to detect a significant difference in PPF rates, and a Fisher’s exact test was performed to determine an association between collared and collarless component use on PPF rates.Aims
Methods
Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).Aims
Methods
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite coating in a series of 52 men (six bilateral) and 40 women (two bilateral), has been described previously. None required revision for aseptic loosening. We have extended the follow-up to 20 years with a minimum of 17 years. The mean age of the patients at total hip replacement was 58.9 years (19 to 84). Six patients were lost to follow-up, but were included up to their last clinical review. A total of 22 patients (22 hips) had died, all from causes unrelated to their surgery. There have been 43 re-operations for failure of the acetabular component. However, in 38 of these the stem was not revised since it remained stable and there was no associated osteolysis. Two of the revisions were for damage to the
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively.Aims
Methods
Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations. A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.Aims
Methods
Elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties are a suggested risk factor for developing cardiovascular complications including cardiomyopathy. Clinical studies assessing patients with MoM hips using left ventricular ejection fraction (LVEF) have found conflicting evidence of cobalt-induced cardiomyopathy. Global longitudinal strain (GLS) is an echocardiography measurement known to be more sensitive than LVEF when diagnosing early cardiomyopathies. The extent of cardiovascular injury, as measured by GLS, in patients with elevated blood cobalt levels has not previously been examined. A total of 16 patients with documented blood cobalt ion levels above 13 µg/l (13 ppb, 221 nmol/l) were identified from a regional arthroplasty database. They were matched with eight patients awaiting hip arthroplasty. All patients underwent echocardiography, including GLS, investigating potential signs of cardiomyopathy.Aims
Methods
The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up. Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening.Aims
Methods
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement. Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.Aims
Methods
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre. All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery.Aims
Methods
We describe the survival at ten years of 100 femoral components of the Freeman hip prosthesis. It is proximally hydroxyapatite (HA)-coated and was fixed without cement. Radiological assessment identified radiolucent lines (RLLs) and lytic lesions and was used to measure migration. The criterion of failure was revision or impending revision for aseptic femoral loosening. No femoral components were revised or are awaiting revision for aseptic loosening, giving 100% survival at ten years (95% confidence interval, 95.7 to 100), although 59 were at risk at ten years. Two components were revised for fracture of a ceramic head with damage to the
The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.Aims
Methods
Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Using the database of
a single academic centre, 126 revision THAs in 117 patients using a single
design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.Aims
Methods
The aim of this study was to evaluate fretting and corrosion in retrieved oxidized zirconium (OxZr; OXINIUM, Smith & Nephew, Memphis, Tennessee) femoral heads and compare the results with those from a matched cohort of cobalt-chromium (CoCr) femoral heads. A total of 28 OxZr femoral heads were retrieved during revision total hip arthroplasty (THA) and matched to 28 retrieved CoCr heads according to patient demographics. The mean age at index was 56 years (46 to 83) in the OxZr group and 70 years (46 to 92) in the CoCr group. Fretting and corrosion scores of the female taper of the heads were measured according to the modified Goldberg scoring method.Aims
Patients and Methods
The aim of this study was to report the initial results of the
Exeter V40 stem, which became available in 2000. A total of 540 total hip arthroplasties (THAs) were performed
in our unit using this stem between December 2000 and May 2002.
Our routine protocol is to review patients postoperatively and at
one, five, and ten years following surgery.Aims
Patients and Methods
This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current.Aims
Methods
Ceramic bearings have several desirable properties, such as resistance to wear, hardness, and biocompatibility, that favour it as an articulating surface in hip arthroplasty. However, ceramic fracture remains a concern. We have reviewed the contemporary literature, addressing the factors that can influence the incidence of ceramic bearing surface fracture. Cite this article:
Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR. Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.Aims
Patients and Methods
Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and mechanical failure rates with implantation of a short cementless humeral component in anatomical (TSA) and reverse shoulder arthroplasty (RSA). A total of 100 shoulder arthroplasties (35 TSA and 65 RSA) were evaluated at a mean of 3.8 years (3 to 8.3). The mean age at the time of surgery was 68 years (31 to 90). The mean body mass index was 32.7 kg/m2 (17.3 to 66.4).Aims
Patients and Methods
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.Aims
Patients and Methods
Displaced, comminuted acetabular fractures in the elderly are increasingly common, but there is no consensus on whether they should be treated non-surgically, surgically with open reduction and internal fixation (ORIF), or with acute total hip arthroplasty (THA). A combination of ORIF and acute THA, an approach called ’combined hip procedure’ (CHP), has been advocated and our aim was to compare the outcome after CHP or ORIF alone. A total of 27 patients with similar acetabular fractures (severe acetabular impaction with or without concomitant femoral head injury) with a mean age of 72.2 years (50 to 89) were prospectively followed for a minimum of two years. In all, 14 were treated with ORIF alone and 13 were treated with a CHP. Hip joint and patient survival were estimated. Operating times, blood loss, radiological outcomes, and patient-reported outcomes were assessed.Aims
Patients and Methods
Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.Aims
Materials and Methods
The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces. In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.Objectives
Methods
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated.Objectives
Methods
Patients with neuromuscular imbalance who require
total hip arthroplasty (THA) present particular technical problems
due to altered anatomy, abnormal bone stock, muscular imbalance
and problems of rehabilitation. In this systematic review, we studied articles dealing with THA
in patients with neuromuscular imbalance, published before April
2017. We recorded the demographics of the patients and the type
of neuromuscular pathology, the indication for surgery, surgical
approach, concomitant soft-tissue releases, the type of implant
and bearing, pain and functional outcome as well as complications
and survival. Recent advances in THA technology allow for successful outcomes
in these patients. Our review suggests excellent benefits for pain
relief and good functional outcome might be expected with a modest
risk of complication. Cite this article:
The use of large-diameter metal-on-metal (MoM)
components in total hip arthroplasty (THA) is associated with an increased
risk of early failure due to adverse local tissue reaction to metal
debris (ARMD) in response to the release of metal ions from the
bearing couple and/or head-neck taper corrosion. The aim of this
paper was to present a review of the incidence and natural history
of ARMD and the forms of treatment, with a focus on the need for
and extent of resection or debulking of the pseudotumour. An illustrative
case report is presented of a patient with an intra-pelvic pseudotumour
associated with a large diameter MoM THA, which was treated successfully
with revision of the bearing surface to a dual mobility couple and
retention of the well-fixed acetabular and femoral components. The
pseudotumour was left Cite this article:
Instability remains a challenging problem in both primary and
revision total hip arthroplasty (THA). Dual mobility components
confer increased stability, but there are concerns about the unique
complications associated with these designs, as well as the long-term
survivorship. We performed a systematic review of all English language articles
dealing with dual mobility THAs published between 2007 and 2016
in the MEDLINE and Embase electronic databases. A total of 54 articles
met inclusion criteria for the final analysis of primary and revision
dual mobility THAs and dual mobility THAs used in the treatment
of fractures of the femoral neck. We analysed the survivorship and
rates of aseptic loosening and of intraprosthetic and extra-articular
dislocation.Aims
Materials and Methods
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
Version of the femoral stem is an important factor
influencing the risk of dislocation after total hip replacement (THR)
as well as the position of the acetabular component. However, there
is no radiological method of measuring stem anteversion described
in the literature. We propose a radiological method to measure stem
version and have assessed its reliability and validity. In 36 patients
who underwent THR, a hip radiograph and CT scan were taken to measure
stem anteversion. The radiograph was a modified Budin view. This
is taken as a posteroanterior radiograph in the sitting position
with 90° hip flexion and 90° knee flexion and 30° hip abduction.
The angle between the stem-neck axis and the posterior intercondylar
line was measured by three independent examiners. The intra- and
interobserver reliabilities of each measurement were examined. The
radiological measurements were compared with the CT measurements
to evaluate their validity. The mean radiological measurement was
13.36° ( Cite this article:
Compared with primary total hip arthroplasty (THA), revision
surgery can be challenging. The cement-in-cement femoral revision
technique involves removing a femoral component from a well-fixed
femoral cement mantle and cementing a new stem into the original
mantle. This technique is widely used and when carried out for the
correct indications, is fast, relatively inexpensive and carries
a reduced short-term risk for the patient compared with the alternative
of removing well-fixed cement. We report the outcomes of this procedure
when two commonly used femoral stems are used. We identified 1179 cement-in-cement stem revisions involving
an Exeter or a Lubinus stem reported to the Swedish Hip Arthroplasty
Register (SHAR) between January 1999 and December 2015. Kaplan-Meier
survival analysis was performed.Aims
Patients and Methods
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.
Femoral stems with exchangeable (modular) necks were introduced
to offer surgeons an increased choice when determining the version,
offset and length of the femoral neck during total hip arthroplasty
(THA). It was hoped that this would improve outcomes and reduce
complications, particularly dislocation. In 2010, the Australian Orthopaedic
Association National Joint Replacement Registry (AOANJRR) first
reported an increased rate of revision after primary THA using femoral
stems with an exchangeable neck. The aim of this study was to provide
a more comprehensive up-to-date analysis of primary THA using femoral
stems with exchangeable and fixed necks. The data included all primary THA procedures performed for osteoarthritis
(OA), reported to the AOANJRR between 01 September 1999 and 31 December
2014. There were 9289 femoral stems with an exchangeable neck and
253 165 femoral stems with a fixed neck. The characteristics of
the patients and prostheses including the bearing surface and stem/neck
metal combinations were examined using Cox proportional hazard ratios
(HRs) and Kaplan-Meier estimates of survivorship. Aims
Materials and Methods
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
To analyse the effectiveness of debridement and implant retention
(DAIR) in patients with hip periprosthetic joint infection (PJI)
and the relationship to patient characteristics. The outcome was
evaluated in hips with confirmed PJI and a follow-up of not less
than two years. Patients in whom DAIR was performed were identified from our
hip arthroplasty register (between 2004 and 2013). Adherence to
criteria for DAIR was assessed according to a previously published
algorithm.Aims
Patients and Methods
The common recommended treatment for infected
total hip replacement is two-staged exchange including removal of
all components. However, removal of well-fixed femoral stems can
result in structural bone damage. We recently reported on an alternative
treatment of partial two-stage exchange used in selected cases,
in which a well-fixed femoral stem was left and only the acetabular
component removed, the joint space was debrided thoroughly, an antibiotic-laden
polymethylmethacrylate spacer was moulded using a bulb-type syringe
and placed in the acetabulum, intravenous antibiotics were administered
during the interval, and delayed re-implantation was performed.
In 19 patients treated with this technique from January 2000 to
January 2011, 89% were free of infection at a mean follow-up of
four years (2 to 11). Since then, disposable silicone moulds have
become available to fabricate spacers in separate femoral and head
units. The head spacer mould, which incorporates various neck taper adapter
options, greatly facilitates the technique of partial two-stage
exchange. We report our early experience using disposable silicone
head spacer moulds for partial two-stage exchange in seven patients
with infected primary hip replacements. Cite this article:
We present a series of 35 patients (19 men and
16 women) with a mean age of 64 years (36.7 to 75.9), who underwent
total hip replacement using the ESKA dual-modular short stem with
metal on-polyethylene bearing surfaces. This implant has a modular
neck section in addition to the modular head. Of these patients,
three presented with increasing post-operative pain due to pseudotumour
formation that resulted from corrosion at the modular neck-stem
junction. These patients underwent further surgery and aseptic lymphocytic
vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the
neck-stem taper. Blood cobalt and chromium levels were measured
at a mean of nine months (3 to 28) following surgery. These were
compared with the levels in seven control patients (three men and
four women) with a mean age of 53.4 years (32.1 to 64.1), who had
an identical prosthesis and articulation but with a prosthesis that
had no modularity at neck-stem junction. The mean blood levels of
cobalt in the study group were raised at 50.75 nmol/l (5 to 145)
compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important
source of metal ion release and pseudotumour formation requiring
revision surgery. Finite element modelling of the dual modular stem
demonstrated high stresses at the modular stem-neck junction. Dual
modular cobalt-chrome hip prostheses should be used with caution
due to these concerns.
We present a minimum 20-year follow-up study of 382 cemented
Exeter Universal total hip arthroplasties (350 patients) operated
on at a mean age of 66.3 years (17 to 94). All patients received the same design of femoral component, regardless
of the original diagnosis. Previous surgery had been undertaken
for 33 hips (8.6%). During the study period 218 patients with 236
hips (62%) died, 42 hips (11%) were revised and 110 hips (29%) in
96 patients were available for review. The acetabular components
were varied and some designs are now obsolete, however they were
all cemented.Aims
Patients and Methods
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck taper adaptor and corrosion products in the
tissues. Therefore, we believe corrosion products from the taper
caused by mismatch of the implant components led to pseudotumour
formation requiring revision.
This review examines the future of total hip arthroplasty, aiming to avoid past mistakes
High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA). We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015.Objectives
Methods
The June 2013 Hip &
Pelvis Roundup360 looks at: failure in metal-on-metal arthroplasty; minimal hip approaches; whether bisphosphonates improve femoral bone stock following arthroplasty; whether more fat means more operative time; surgical infection; vascularised fibular graft for osteonecrosis; subclinical SUFE; and dentists, hips and antibiotics.
The October 2015 Hip &
Pelvis Roundup360 looks at: Smoking and complications in arthroplasty; Smoking cessation beneficial in arthroplasty; Intermediate care and arthroplasty; Do we still need cell salvage?; Femoroacetabular impingement in the Japanese population; Trunnionosis or taperosis and geometry; Decontamination for staphylococcus aureus works!; Policeman or opportunity? Quality improvement with registries; Death rates higher in readmission to other hospitals
We investigated changes in the axial alignment of the ipsilateral
hip and knee after total hip arthroplasty (THA). We reviewed 152 patients undergoing primary THA (163 hips; 22
hips in men, 141 hips in women) without a pre-operative flexion
contracture. The mean age was 64 years (30 to 88). The diagnosis
was osteoarthritis (OA) in 151 hips (primary in 18 hips, and secondary
to dysplasia in 133) and non-OA in 12 hips. A posterolateral approach
with repair of the external rotators was used in 134 hips and an
anterior approach in 29 hips. We measured changes in leg length
and offset on radiographs, and femoral anteversion, internal rotation
of the hip and lateral patellar tilt on CT scans, pre- and post-operatively. Aims
Patients and Methods
In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately: Hip resurfacing, large head (>
36 mm) MoM
THA and MoM THA with traditional femoral head sizes.
Fracture of a ceramic component in total hip
replacement is a rare but potentially catastrophic complication.
The incidence is likely to increase as the use of ceramics becomes
more widespread. We describe such a case, which illustrates how
inadequate initial management will lead to further morbidity and
require additional surgery. We present the case as a warning that
fracture of a ceramic component should be revised to another ceramic-on-ceramic
articulation in order to minimise the risk of further catastrophic
wear.
Cementless femoral stems are currently preferred
for total hip replacement (THR) in the United States. Improvements
in stem design, instrumentation and surgical technique have made
this technology highly successful, reproducible, and applicable
to the vast majority of patients requiring a THR. However, there
are ongoing developments in some aspects of stem design that influence
clinical results, the incidence of complications and their inherent adaptability
in accommodating the needs of individual patients. Here we examine
some of these design features. Cite this article:
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.
We compared a modular neck system with a non-modular system in a cementless anatomical total hip replacement (THR). Each group consisted of 74 hips with developmental hip dysplasia. Both groups had the same cementless acetabular component and the same articulation, which consisted of a conventional polyethylene liner and a 28 mm alumina head. The mean follow-up was 14.5 years (13 to 15), at which point there were significant differences in the mean total Harris hip score (modular/non-modular: 98.6 (64 to 100)/93.8 (68 to 100)), the mean range of abduction (32° (15° to 40°)/28 (0° to 40°)), use of a 10° elevated liner (31%/100%), the incidence of osteolysis (27%/79.7%) and the incidence of equal leg lengths (≥ 6 mm, 92%/61%). There was no disassociation or fracture of the modular neck. The modular system reduces the need for an elevated liner, thereby reducing the incidence of osteolysis. It gives a better range of movement and allows the surgeon to make an accurate adjustment of leg length.
Large ceramic femoral heads offer several advantages
that are potentially advantageous to patients undergoing both primary
and revision total hip replacement. Many high-quality studies have
demonstrated the benefit of large femoral heads in reducing post-operative instability.
Ceramic femoral heads may also offer an advantage in reducing polyethylene wear
that has been reported Cite this article: Bone Joint J 2013;95-B, Supple A:63–6.
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article:
We report the long-term outcome of a modified second-generation cementing technique for fixation of the acetabular component of total hip replacement. An earlier report has shown the superiority of this technique assessed by improved survival compared with first-generation cementing. The acetabular preparation involved reaming only to the subchondral plate, followed by impaction of the bone in the anchorage holes. Between 1978 and 1993, 287 total hip replacements were undertaken in 244 patients with a mean age of 65.3 years (21 to 90) using a hemispherical Weber acetabular component with this modified technique for cementing and a cemented femoral component. The survival with acetabular revision for aseptic loosening as the endpoint was 99.1% (95% confidence interval 97.9 to 100 after ten years and 85.5% (95% confidence interval 74.7 to 96.2) at 20 years. Apart from contributing to a long-lasting fixation of the component, this technique also preserved bone, facilitating revision surgery when necessary.
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
An international faculty of orthopaedic surgeons
presented their work on the current challenges in hip surgery at
the London Hip Meeting which was attended by over
400 delegates. The topics covered included femoroacetabular impingement, thromboembolic
phenomena associated with hip surgery, bearing surfaces (including metal-on-metal
articulations), outcomes of hip replacement surgery and revision
hip replacement. We present a concise report of the current opinions
on hip surgery from this meeting with appropriate references to
the current literature.
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of pseudotumour formation and to identify
risk factors for their formation in a prospective cohort study. A total of 119 patients who underwent 120 MoM THRs with large-diameter
femoral heads between January 2005 and November 2007 were included
in the study. Outcome scores, serum metal ion levels, radiographs
and CT scans were obtained. Patients with symptoms or an identified
pseudotumour were offered MRI and an ultrasound-guided biopsy. There were 108 patients (109 hips) eligible for evaluation by
CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients
(39%) were diagnosed with a pseudotumour. The hips of 13 patients
(12%) were revised to a polyethylene acetabular component with small-diameter
metal head. Patients with elevated serum metal ion levels had a
four times increased risk of developing a pseudotumour. This study shows a substantially higher incidence of pseudotumour
formation and subsequent revisions in patients with MoM THRs than
previously reported. Because most revision cases were identified
only after an intensive screening protocol, we recommend close monitoring
of patients with MoM THR.
We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the head and polyethylene. The implants were retrieved at the time of revision surgery to correct leg-length discrepancy just 48 hours after the primary hip replacement. Appropriate analysis of the retrieved femoral head demonstrated loss of the Oxinium layer with exposure of the underlying substrate and transfer of titanium from the acetabular shell at the time of a reduction of the index total hip replacement. In addition, the level of damage to the polyethylene was extensive despite only 48 hours The purpose of this report is to highlight the care that is required at the time of reduction, especially with these hard femoral counter-faces such as Oxinium. To our knowledge, the damage occurring at the time of reduction has not been previously reported following the retrieval of an otherwise well-functioning hip replacement.
We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99). These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
Between 1986 and 1991 we implanted 331 consecutive Furlong hydroxyapatite-coated femoral components of a total hip replacement in 291 patients. A cemented acetabular prosthesis was used in 217 hips and a hydroxyapatite-coated component in 114. We describe the long-term clinical and radiological survival of the femoral component at a mean follow-up of 17.5 years (15 to 21). Only two patients (0.68%) were lost to follow-up. With revision of the femoral component for any reason as the endpoint, the survival at a mean of 17 years was 97.4% (95% confidence interval 94.1 to 99.5), and with revision for aseptic loosening as the endpoint it was 100%. The survival at a maximum of 21 years with revision of the femoral component for any reason as the endpoint was 97.4% (95% confidence interval 81.0 or 99.5). These results compare favourably with the best long-term results of cemented or uncemented femoral components used in total hip replacement.
The biological significance of cobalt-chromium wear particles from metal-on-metal hip replacements may be different to the effects of the constituent metal ions in solution. Bacteria may be able to discriminate between particulate and ionic forms of these metals because of a transmembrane nickel/cobalt-permease. It is not known whether wear particles are bacteriocidal. We compared the doubling time of coagulase negative staphylococcus, Doubling time halved in metal-on-metal (p = 0.003) and metal-on-polyethylene (p = 0.131) particulate debris compared with the control. Bacterial nickel/cobalt-transporters allow metal ions but not wear particles to cross bacterial membranes. This may be useful for testing the biological characteristics of different wear debris. This experiment also shows that metal-on-metal hip wear debris is not bacteriocidal.
The aim of this study was to assess whether a femoral component which retained the neck reduced the incidence of leg-length inequality following total hip arthroplasty. A retrospective review was undertaken of 130 consecutive primary total hip arthroplasties performed between April 1996 and April 2004 using such an implant. There were 102 suitable patients for the study. Standardised pre- and post-operative pelvic radiographs were measured by an independent investigator to the nearest millimetre. The leg-length inequality was reduced from a mean pre-operative value of −0.71 cm to a mean of 0.11 cm post-operatively. Of the 102 patients 24 (23.5%) had an equal leg-length post-operatively, and 95 (93.1%) had a leg-length inequality between −1 cm and 1 cm.
We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively porous-coated, distally-fixed femoral stem. Factors relating to the patients, the implant and the operations were compared with those with intact stems. Finite-element analysis was performed on two of the fractured implants. Factors associated with fracture of the stem were poor proximal bone support (type III–type IV; p = 0.001), a body mass index >
30; (p = 0.014), a smaller diameter of stem (<
13.5 mm; p = 0.007) and the use of an extended trochanteric osteotomy (ETO 4/5: p = 0.028). Finite-element analysis showed that the highest stresses on the stem occurred adjacent to the site of the fracture. The use of a strut graft wired over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stresses on the stem by 48%. We recommend the use of a strut allograft in conjunction with an extended trochanteric osteotomy in patients with poor proximal femoral bone stock.
We used an inverted shoulder arthroplasty in 43 consecutive patients with a mean age of 78 years (65 to 97) who had sustained a three- or four-part fracture of the upper humerus. All except two were reviewed with a mean follow-up of 22 months (6 to 58). The clinical outcome was satisfactory with a mean active anterior elevation of 97° (35° to 160°) and a mean active external rotation in abduction of 30° (0° to 80°). The mean Constant and the mean modified Constant scores were respectively 44 (16 to 69) and 66% (25% to 97%). Complications included three patients with reflex sympathetic dystrophy, five with neurological complications, most of which resolved, and one with an anterior dislocation. Radiography showed peri-prosthetic calcification in 36 patients (90%), displacement of the tuberosities in 19 (53%) and a scapular notch in ten (25%). Compared with conventional hemiarthroplasty, satisfactory mobility was obtained despite frequent migration of the tuberosities. However, long-term results are required before reverse shoulder arthroplasty can be recommended as a routine procedure in complex fractures of the upper humerus in the elderly.