Aims. We evaluated a large database with
In a prospective study we assessed the causes of
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD).
Aims. The risk of
Aims. The timing of when to remove a circular frame is crucial; early removal results in refracture or deformity, while late removal increases the patient morbidity and delay in return to work. This study was designed to assess the effectiveness of a staged reloading protocol. We report the incidence of
Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the risk of revision (hazard ratio (HR) 0.68 (95% confidence interval (CI) 0.50 to 0.92)), while the angulated/offset liner increased the risk of revision for
Aims. Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. Methods. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or
The case reports of four patients with breakage of the intramedullary nail of the femur are presented. In all four patients the nail broke after consolidation of the fracture of the femoral shaft. All the nails broke at the same place: the junction between the round cross-section and the cloverleaf cross-section.
Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This failure mode, which we refer to as aseptic
Aims. Periprosthetic joint infection (PJI) is a challenging complication of any arthroplasty procedure. We reviewed our use of static antibiotic-loaded cement spacers (ABLCSs) for staged management of PJI where segmental bone loss, ligamentous instability, or soft-tissue defects necessitate a static construct. We reviewed factors contributing to their failure and techniques to avoid these complications when using ABLCSs in this context. Methods. A retrospective analysis was conducted of 94 patients undergoing first-stage revision of an infected knee prosthesis between September 2007 and January 2020 at a single institution. Radiographs and clinical records were used to assess and classify the incidence and causes of static spacer failure. Of the 94 cases, there were 19 primary total knee arthroplasties (TKAs), ten revision TKAs (varus-valgus constraint), 20 hinged TKAs, one arthrodesis (nail), one failed spacer (performed elsewhere), 21 distal femoral endoprosthetic arthroplasties, and 22 proximal tibial arthroplasties. Results. A total of 35/94 patients (37.2%) had spacer-related complications, of which 26/35 complications (74.3%) were because of
Aims. We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for
Aims. Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. Patients and Methods. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1). Results. The median follow-up was 132 months (interquartile range (IQR) 44 to 189). In all, 20 stems required revision: eight for infection, five for structural failure, five for aseptic loosening, and two for tumour progression. Mechanical survivorship at five, ten, and 15 years was 84%, 78%, and 78%, respectively.
Aims. Debate continues regarding the optimum management of periprosthetic distal femoral fractures (PDFFs). This study aims to determine which operative treatment is associated with the lowest perioperative morbidity and mortality when treating low (Su type II and III) PDFFs comparing lateral locking plate fixation (LLP-ORIF) or distal femoral arthroplasty (DFA). Methods. This was a retrospective cohort study of 60 consecutive unilateral (PDFFs) of Su types II (40/60) and III (20/60) in patients aged ≥ 60 years: 33 underwent LLP-ORIF (mean age 81.3 years (SD 10.5), BMI 26.7 (SD 5.5); 29/33 female); and 27 underwent DFA (mean age 78.8 years (SD 8.3); BMI 26.7 (SD 6.6); 19/27 female). The primary outcome measure was reoperation. Secondary outcomes included perioperative complications, calculated blood loss, transfusion requirements, functional mobility status, length of acute hospital stay, discharge destination and mortality. Kaplan-Meier survival analysis was performed. Cox multivariate regression analysis was performed to identify risk factors for reoperation after LLP-ORIF. Results. Follow-up was at mean 3.8 years (1.0 to 10.4). One-year mortality was 13% (8/60). Reoperation was more common following LLP-ORIF: 7/33 versus 0/27 (p = 0.008). Five-year survival for reoperation was significantly better following DFA; 100% compared to 70.8% (95% confidence interval (CI) 51.8% to 89.8%, p = 0.006). There was no difference for the endpoint
Aims. There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. Methods. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality. Results. The cumulative incidence of septic failure was 8% (95% confidence interval (CI) 3.5 to 15) at one year, 13.8% (95% CI 7.6 to 22) at two years, and 19.7% (95% CI 12 to 28.6) at five and ten years of follow-up. A femoral bone defect worse than Paprosky IIIA (hazard ratio (HR) 13.58 (95% CI 4.86 to 37.93); p < 0.001) and obesity (BMI ≥ 30 kg/m. 2. ; HR 3.88 (95% CI 1.49 to 10.09); p = 0.005) were significantly associated with septic failure. Instability and periprosthetic fracture were the most common reasons for
Aims. Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or
Aims. Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. Methods. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up. Results. The implant was revised in ten shoulders (5.2%) with a median time to revision of 21.2 months (interquartile range (IQR) 9.9 to 41.8). The Kaplan-Meier survivorship estimate at seven years was 95.9% (95% confidence interval (CI) 91.7 to 98; 35 RTSAs at risk) for aseptic
Aims. The aim of this study was to describe implant and patient-reported outcome in patients with a unilateral transfemoral amputation (TFA) treated with a bone-anchored, transcutaneous prosthesis. Methods. In this cohort study, all patients with a unilateral TFA treated with the Osseointegrated Prostheses for the Rehabilitation of Amputees (OPRA) implant system in Sahlgrenska University Hospital, Gothenburg, Sweden, between January 1999 and December 2017 were included. The cohort comprised 111 patients (78 male (70%)), with a mean age 45 years (17 to 70). The main reason for amputation was trauma in 75 (68%) and tumours in 23 (21%). Patients answered the Questionnaire for Persons with Transfemoral Amputation (Q-TFA) before treatment and at two, five, seven, ten, and 15 years’ follow-up. A prosthetic activity grade was assigned to each patient at each timepoint. All mechanical complications, defined as fracture, bending, or wear to any part of the implant system resulting in removal or change, were recorded. Results. The Q-TFA scores at two, five, seven, and ten years showed significantly more prosthetic use, better mobility, fewer problems, and an improved global situation, compared with baseline. The survival rate of the osseointegrated implant part (the fixture) was 89% and 72% after seven and 15 years, respectively. A total of 61 patients (55%) had mechanical complications (mean 3.3 (SD 5.76)), resulting in exchange of the percutaneous implant parts. There was a positive relationship between a higher activity grade and the number of mechanical complications. Conclusion. Compared with before treatment, the patient-reported outcome was significantly better and remained so over time. Although osseointegration and the ability to transfer loads over a 15-year period have been demonstrated, a large number of
Aims. Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and
Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42%
The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of proximal fixation was recorded from
intra-operative fluoroscopic images, and technical inadequacy for
each point was defined. All patients were followed to final follow-up
and
The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA). Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.Aims
Methods
Aims. The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of
Aims. The use of a noninvasive growing endoprosthesis in the management
of primary bone tumours in children is well established. However,
the efficacy of such a prosthesis in those requiring a revision
procedure has yet to be established. The aim of this series was
to present our results using extendable prostheses for the revision
of previous endoprostheses. Patients and Methods. All patients who had a noninvasive growing endoprosthesis inserted
at the time of a revision procedure were identified from our database.
A total of 21 patients (seven female patients, 14 male) with a mean
age of 20.4 years (10 to 41) at the time of revision were included.
The indications for revision were
The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.Aims
Methods
There are concerns regarding complications and longevity of total elbow arthroplasty (TEA) in young patients, and the few previous publications are mainly limited to reports on linked elbow devices. We investigated the clinical outcome of unlinked TEA for patients aged less than 50 years with rheumatoid arthritis (RA). We retrospectively reviewed the records of 26 elbows of 21 patients with RA who were aged less than 50 years who underwent primary TEA with an unlinked elbow prosthesis. The mean patient age was 46 years (35 to 49), and the mean follow-up period was 13.6 years (6 to 27). Outcome measures included pain, range of motion, Mayo Elbow Performance Score (MEPS), radiological evaluation for radiolucent line and loosening, complications, and revision surgery with or without implant removal.Aims
Methods
The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance. A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.Aims
Methods
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis. We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.Aims
Methods
Proximal humeral fractures are the third most common fracture among the elderly. Complications associated with fixation include screw perforation, varus collapse, and avascular necrosis of the humeral head. To address these challenges, various augmentation techniques to increase medial column support have been developed. There are currently no recent studies that definitively establish the superiority of augmented fixation over non-augmented implants in the surgical treatment of proximal humeral fractures. The aim of this systematic review and meta-analysis was to compare the outcomes of patients who underwent locking-plate fixation with cement augmentation or bone-graft augmentation versus those who underwent locking-plate fixation without augmentation for proximal humeral fractures. The search was carried out according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses guidelines. Articles involving patients with complex proximal humeral fractures treated using open reduction with locking-plate fixation, with or without augmentation, were considered. A meta-analysis of comparative studies comparing locking-plate fixation with cement augmentation or with bone-graft augmentation versus locking-plate fixation without augmentation was performed.Aims
Methods
Aim. We present the medium-term clinical results of a reverse total
shoulder arthroplasty with a trabecular metal glenoid base plate. Patients and Methods. We reviewed 125 consecutive primary reverse total shoulder arthroplasties
(RTSA) implanted in 124 patients for rotator cuff arthropathy. There
were 100 women and 24 men in the study group with a mean age of
76 years (58 to 89). The mean follow-up was 32 months (24 to 60).
No patient was lost to follow-up. Results. There were statistically significant improvements in the mean
range of movement and Oxford Shoulder Score (p <
0.001). Kaplan-Meier
survivorship at five years was 96.7% (95% confidence interval 91.5
to 98.7) with aseptic glenoid failure as the end point. . Radiologically, 63 shoulders (50.4%) showed no evidence of notching,
51 (40.8%) had grade 1 notching, ten (8.0%) had grade 2 notching
and one (0.8%) had grade 4 notching. Radiolucency around the glenoid
base plate was found in one patient (0.8%) and around the humeral
stem in five (4.0%). In all, three RTSA (2.4%) underwent revision surgery
for aseptic
Cite this article:
The August 2023 Children’s orthopaedics Roundup360 looks at: DDH: What can patients expect after open reduction?; Femoral head deformity associated with hip displacement in non-ambulatory cerebral palsy; Bony hip reconstruction for displaced hips in patients with cerebral palsy: is postoperative immobilization indicated?; Opioid re-prescriptions after ACL reconstruction in adolescents are associated with subsequent opioid use disorder; Normative femoral and tibial lengths in a modern population of USA children; Retrospective analysis of associated anomalies in 636 patients with operatively treated congenital scoliosis; Radiological hip shape and patient-reported outcome measures in healed Perthes’ disease; Significantly displaced adolescent posterior sternoclavicular joint injuries.
The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
The aims of this study were to determine the success of a reconstruction algorithm used in major acetabular bone loss, and to further define the indications for custom-made implants in major acetabular bone loss. We reviewed a consecutive series of Paprosky type III acetabular defects treated according to a reconstruction algorithm. IIIA defects were planned to use a superior augment and hemispherical acetabular component. IIIB defects were planned to receive either a hemispherical acetabular component plus augments, a cup-cage reconstruction, or a custom-made implant. We used national digital health records and registry reports to identify any reoperation or re-revision procedure and Oxford Hip Score (OHS) for patient-reported outcomes. Implant survival was determined via Kaplan-Meier analysis.Aims
Methods
Prior cost-effectiveness analyses on osseointegrated prosthesis for transfemoral unilateral amputees have analyzed outcomes in non-USA countries using generic quality of life instruments, which may not be appropriate when evaluating disease-specific quality of life. These prior analyses have also focused only on patients who had failed a socket-based prosthesis. The aim of the current study is to use a disease-specific quality of life instrument, which can more accurately reflect a patient’s quality of life with this condition in order to evaluate cost-effectiveness, examining both treatment-naïve and socket refractory patients. Lifetime Markov models were developed evaluating active healthy middle-aged male amputees. Costs of the prostheses, associated complications, use/non-use, and annual costs of arthroplasty parts and service for both a socket and osseointegrated (OPRA) prosthesis were included. Effectiveness was evaluated using the questionnaire for persons with a transfemoral amputation (Q-TFA) until death. All costs and Q-TFA were discounted at 3% annually. Sensitivity analyses on those cost variables which affected a change in treatment (OPRA to socket, or socket to OPRA) were evaluated to determine threshold values. Incremental cost-effectiveness ratios (ICERs) were calculated.Aims
Methods
The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral. The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.Aims
Methods
This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component. We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.Aims
Methods
With up to 40% of patients having patellofemoral joint osteoarthritis (PFJ OA), the two arthroplasty options are to replace solely the patellofemoral joint via patellofemoral arthroplasty (PFA), or the entire knee via total knee arthroplasty (TKA). The aim of this study was to assess postoperative success of second-generation PFAs compared to TKAs for patients treated for PFJ OA using patient-reported outcome measures (PROMs) and domains deemed important by patients following a patient and public involvement meeting. MEDLINE, EMBASE via OVID, CINAHL, and EBSCO were searched from inception to January 2022. Any study addressing surgical treatment of primary patellofemoral joint OA using second generation PFA and TKA in patients aged above 18 years with follow-up data of 30 days were included. Studies relating to OA secondary to trauma were excluded. ROB-2 and ROBINS-I bias tools were used.Aims
Methods
The sacroiliac joint (SIJ) is the only mechanical connection between the axial skeleton and lower limbs. Following iliosacral resection, there is debate on whether reconstruction of the joint is necessary. There is a paucity of data comparing the outcomes of patients undergoing reconstruction and those who are not formally reconstructed. A total of 60 patients (25 females, 35 males; mean age 39 years (SD 18)) undergoing iliosacral resection were reviewed. Most resections were performed for primary malignant tumours (n = 54; 90%). The mean follow-up for surviving patients was nine years (2 to 19).Aims
Methods
The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.Aims
Methods
The aim of this study was to determine the prevalence of depressive and anxiety disorders prior to total hip (THA) and total knee arthroplasty (TKA) and to assess their impact on the rates of any infection, revision, or reoperation. Between January 2000 and March 2019, 21,469 primary and revision arthroplasties (10,011 THAs; 11,458 TKAs), which were undertaken in 15,504 patients at a single academic medical centre, were identified from a 27-county linked electronic medical record (EMR) system. Depressive and anxiety disorders were identified by diagnoses in the EMR or by using a natural language processing program with subsequent validation from review of the medical records. Patients with mental health diagnoses other than anxiety or depression were excluded.Aims
Methods
Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up. This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.Aims
Methods
Total femoral arthroplasty (TFA) is a rare procedure used in cases of significant femoral bone loss, commonly from cancer, infection, and trauma. Low patient numbers have resulted in limited published work on long-term outcomes, and even less regarding TFA undertaken for non-oncological indications. The aim of this study was to evaluate the long-term clinical outcomes of all TFAs in our unit. Data were collected retrospectively from a large tertiary referral revision arthroplasty unit’s database. Inclusion criteria included all patients who underwent TFA in our unit. Preoperative demographics, operative factors, and short- and long-term outcomes were collected for analysis. Outcome was defined using the Musculoskeletal Infection Society (MSIS) outcome reporting tool.Aims
Methods
Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.Aims
Methods
Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components. We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.Aims
Methods
Despite limited clinical scientific backing, an additional trochanteric stabilizing plate (TSP) has been advocated when treating unstable trochanteric fractures with a sliding hip screw (SHS). We aimed to explore whether the TSP would result in less post operative fracture motion, compared to SHS alone. Overall, 31 patients with AO/OTA 31-A2 trochanteric fractures were randomized to either a SHS alone or a SHS with an additional TSP. To compare postoperative fracture motion, radiostereometric analysis (RSA) was performed before and after weightbearing, and then at four, eight, 12, 26, and 52 weeks. With the “after weightbearing” images as baseline, we calculated translations and rotations, including shortening and medialization of the femoral shaft.Aims
Methods
There is little information in the literature about the use of dual-mobility (DM) bearings in preventing re-dislocation in revision total hip arthroplasty (THA). The aim of this study was to compare the use of DM bearings, standard bearings, and constrained liners in revision THA for recurrent dislocation, and to identify risk factors for re-dislocation. We reviewed 86 consecutive revision THAs performed for dislocation between August 2012 and July 2019. A total of 38 revisions (44.2%) involved a DM bearing, while 39 (45.3%) and nine (10.5%) involved a standard bearing and a constrained liner, respectively. Rates of re-dislocation, re-revision for dislocation, and overall re-revision were compared. Radiographs were assessed for the positioning of the acetabular component, the restoration of the centre of rotation, leg length, and offset. Risk factors for re-dislocation were determined by Cox regression analysis. The modified Harris Hip Scores (mHHSs) were recorded. The mean age of the patients at the time of revision was 70 years (43 to 88); 54 were female (62.8%). The mean follow-up was 5.0 years (2.0 to 8.75).Aims
Methods
This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes? Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.Aims
Methods
The purpose of this study was to compare reoperation and revision rates of double plating (DP), single plating using a lateral locking plate (SP), or distal femoral arthroplasty (DFA) for the treatment of periprosthetic distal femur fractures (PDFFs). All patients with PDFF primarily treated with DP, SP, or DFA between 2008 and 2022 at a university teaching hospital were included in this retrospective cohort study. The primary outcome was revision surgery for failure following DP, SP, or DFA. Secondary outcome measures included any reoperation, length of hospital stay, and mortality. All basic demographic and relevant implant and injury details were collected. Radiological analysis included fracture classification and evaluation of metaphyseal and medial comminution.Aims
Methods
Conventional cemented acetabular components are
reported to have a high rate of failure when implanted into previously
irradiated bone. We recommend the use of a cemented reconstruction
with the addition of an acetabular reinforcement cross to improve
fixation. We reviewed a cohort of 45 patients (49 hips) who had undergone
irradiation of the pelvis and a cemented total hip arthroplasty
(THA) with an acetabular reinforcement cross. All hips had received
a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour
or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3:
45 to 60; mean: 49.57, 32 to 72). . The mean follow-up after THA was 51 months (17 to 137). The cumulative
probability of revision of the acetabular component for a mechanical
reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60
months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One
hip was revised for
Postoperative complication rates remain relatively high after adult spinal deformity (ASD) surgery. The extent to which modifiable patient-related factors influence complication rates in patients with ASD has not been effectively evaluated. The aim of this retrospective cohort study was to evaluate the association between modifiable patient-related factors and complications after corrective surgery for ASD. ASD patients with two-year data were included. Complications were categorized as follows: any complication, major, medical, surgical, major mechanical, major radiological, and reoperation. Modifiable risk factors included smoking, obesity, osteoporosis, alcohol use, depression, psychiatric diagnosis, and hypertension. Patients were stratified by the degree of baseline deformity (low degree of deformity (LowDef)/high degree of deformity (HighDef): below or above 20°) and age (Older/Younger: above or below 65 years). Complication rates were compared for modifiable risk factors in each age/deformity group, using multivariable logistic regression analysis to adjust for confounders.Aims
Methods
The purpose of this study was to assess whether
the use of a joint-sparing technique such as curettage and grafting was
successful in eradicating giant cell tumours of the proximal femur,
or whether an alternative strategy was more appropriate. Between 1974 and 2012, 24 patients with a giant cell tumour of
the proximal femur were treated primarily at our hospital. Treatment
was either joint sparing or joint replacing. Joint-sparing treatment
was undertaken in ten patients by curettage with or without adjunctive
bone graft. Joint replacement was by total hip replacement in nine patients
and endoprosthetic replacement in five. All 11 patients who presented
with a pathological fracture were treated by replacement. Local recurrence occurred in five patients (21%): two were treated
by hip replacement, three by curettage and none with an endoprosthesis.
Of the ten patients treated initially by curettage, six had a successful
outcome without local recurrence and required no further surgery.
Three eventually needed a hip replacement for local recurrence and
one an endoprosthetic replacement for
This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with ≤ 50% acetabular bleeding host bone
contact were compared with a control group of 49 patients with >
50% to 85% bleeding host bone contact. All patients were treated
with the same type of trabecular metal acetabular revision shell.
The mean age at revision was 62.4 years (42 to 80) and the mean
follow-up of both groups was 72.4 months (60 to 102). Clinical,
radiological and functional outcomes were assessed. There were four
(7.5%)
Giant cell tumours (GCTs) of the proximal femur are rare, and there is no consensus about the best method of filling the defect left by curettage. In this study, we compared the outcome of using a fibular strut allograft and bone cement to reconstruct the bone defect after extended curettage of a GCT of the proximal femur. In a retrospective study, we reviewed 26 patients with a GCT of the proximal femur in whom the bone defect had been filled with either a fibular strut allograft (n = 12) or bone cement (n = 14). Their demographic details and oncological and nononcological complications were retrieved from their medical records. Limb function was assessed using the Musculoskeletal Tumor Society (MSTS) score.Aims
Methods
Both the femoral and tibial component are usually cemented at revision total knee arthroplasty (rTKA), while stems can be added with either cemented or press-fit (hybrid) fixation. The aim of this study was to compare the long-term stability of rTKA with cemented and press-fitted stems, using radiostereometric analysis (RSA). This is a follow-up of a randomized controlled trial, initially involving 32 patients, of whom 19 (nine cemented, ten hybrid) were available for follow-up ten years postoperatively, when further RSA measurements were made. Micromotion of the femoral and tibial components was assessed using model-based RSA software (RSAcore). The clinical outcome was evaluated using the Knee Society Score (KSS), the Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (pain and satisfaction).Aims
Methods
A modular layered acetabular component (metal-polyethylene-ceramic) was developed in Japan for use in alumina ceramic-on-ceramic total hip replacement. Between May 1999 and July 2000, we performed 35 alumina ceramic-on-ceramic total hip replacements in 30 consecutive patients, using this layered component and evaluated the clinical and radiological results over a mean follow-up of 5.8 years (5 to 6.5). A total of six hips underwent revision, one for infection, two for dislocation with loosening of the acetabular component, two for alumina liner fractures and one for component dissociation with pelvic osteolysis. There were no fractures of the ceramic heads, and no loosening of the femoral or acetabular component in the unrevised hips was seen at final follow-up. Osteolysis was not observed in any of the unrevised hips. The survivorship analysis at six years after surgery was 83%. The layered acetabular component in our experience, has poor durability because of unexpected
Debridement, antibiotics, and implant retention (DAIR) is a widely accepted form of surgical treatment for patients with an early periprosthetic joint infection (PJI) after primary arthroplasty. The outcome of DAIR after revision arthroplasty, however, has not been reported. The aim of this study was to report the success rate of DAIR after revision arthroplasty with a follow-up of two years. This retrospective study, conducted at the Sint Maartenskliniek, Nijmegen, the Netherlands, included 88 patients who underwent DAIR within 90 days of revision total hip or total knee arthroplasty between 2012 and 2019. Details of the surgical procedures and PJI were collected. Univariate analysis and a subgroup analysis of the culture-positive group were performed. Kaplan-Meier survivorship curves were constructed.Aims
Methods
During revision total hip replacement using morcellised compacted bone allograft, 16 patients were randomised to receive a graft which had been rinsed in either an ibandronate solution or in saline. Patients were assessed by dual energy x-ray absorptiometry after operation and at 3, 6, 12 and 24 months. A region of interest between the tip of the femoral stem and the distal plastic plug was chosen to measure the changes in bone density over time. The study was double-blinded. In all the control patients the bone density decreased during the first three months and then remained constant at this lower level. A large proportion of the mass of the bone graft was lost. In contrast, all patients with grafts treated with bisphosphonate showed a slight increase in bone density. The difference between the groups was highly significant at all points in time. We conclude that rinsing the graft in a bisphosphonate solution prevents its resorption and may therefore reduce the risk of
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge. We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108). The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no
The cemented Oxford unicompartmental knee arthroplasty (OUKA) features two variants: single and twin peg OUKA. The purpose of this study was to assess the stability of both variants in a worst-case scenario of bone defects and suboptimal cementation. Single and twin pegs were implanted randomly allocated in 12 pairs of human fresh-frozen femora. We generated 5° bone defects at the posterior condyle. Relative movement was simulated using a servohydraulic pulser, and analyzed at 70°/115° knee flexion. Relative movement was surveyed at seven points of measurement on implant and bone, using an optic system.Aims
Methods
The aim of this study was to compare the clinical and radiological outcomes of patients with early-onset scoliosis (EOS), who had undergone spinal fusion after distraction-based spinal growth modulation using either traditional growing rods (TGRs) or magnetically controlled growing rods (MCGRs). We undertook a retrospective review of skeletally mature patients who had undergone fusion for an EOS, which had been previously treated using either TGRs or MCGRs. Measured outcomes included sequential coronal T1 to S1 height and major curve (Cobb) angle on plain radiographs and any complications requiring unplanned surgery before final fusion.Aims
Methods
Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study. In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.Aims
Methods
Deficiency of the abductor mechanism is a well-recognised
cause of pain and limping after total hip replacement (THR). This
can be found incidentally at the time of surgery, or it may arise
as a result of damage to the superior gluteal nerve intra-operatively,
or after surgery owing to
Peri-acetabular tumour resections and their subsequent
reconstruction are among the most challenging procedures in orthopaedic
oncology. Despite the fact that a number of different pelvic endoprostheses
have been introduced, rates of complication remain high and long-term
results are mostly lacking. . In this retrospective study, we aimed to evaluate the outcome
of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis
after a type 2 or type 2/3 internal hemipelvectomy. A total of 19 patients (11M:8F) with a mean age of 48 years (14
to 72) were included, most of whom had been treated for a primary
bone tumour (n = 16) between 2003 and 2009. After a mean follow-up
of 39 months (28 days to 8.7 years) seven patients had died. After
a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive,
of whom 11 were disease-free. Complications occurred in 15 patients.
Three had recurrent dislocations and three experienced aseptic loosening.
There were no
Total hip arthroplasty (THA) patients undergoing or having a prior lumbar spine fusion (LSF) have an increased risk of mechanical complications. The aim of this registry-based, retrospective comparative cohort study is to assess the longer term survival of THA in patients who have undergone a LSF during a 17-year period (2000 to 2017). A registry-based population study was conducted on 679 patients who underwent both THA and LSF surgeries. Patients were identified from the regional arthroplasty data base and cross linked to patients with LSF from the regional hospital discharge database between 2000 and 2017. Demographic data, diagnosis leading to primary THA, primary implant survival, perioperative complications, number and causes of failure, and patients requiring revision arthroplasty were collated and compared. For comparison, data from 67,919 primary THAs performed during the same time time period were also retrieved and analyzed.Aims
Methods
To investigate whether idiopathic osteonecrosis of the femoral head (ONFH) is related to impaired osteoblast activities. We cultured osteoblasts isolated from trabecular bone explants taken from the femoral head and the intertrochanteric region of patients with idiopathic ONFH, or from the intertrochanteric region of patients with osteoarthritis (OA), and compared their viability, mineralization capacity, and secretion of paracrine factors.Aims
Methods
Advances in surgical technique and implant design may influence the incidence and mechanism of failure resulting in revision total hip arthroplasty (rTHA). The purpose of the current study was to characterize aetiologies requiring rTHA, and to determine whether temporal changes existed in these aetiologies over a ten-year period. All rTHAs performed at a single institution from 2009 to 2019 were identified. Demographic information and mode of implant failure was obtained for all patients. Data for rTHA were stratified into two time periods to assess for temporal changes: 2009 to 2013, and 2014 to 2019. Operative reports, radiological imaging, and current procedural terminology (CPT) codes were cross-checked to ensure the accurate classification of revision aetiology for each patient.Aims
Methods
Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship.
The aim of this study was to measure the effect of hospital case volume on the survival of revision total knee arthroplasty (RTKA). This is a retrospective analysis of Scottish Arthroplasty Project data, a nationwide audit which prospectively collects data on all arthroplasty procedures performed in Scotland. The primary outcome was RTKA survival at ten years. The primary explanatory variable was the effect of hospital case volume per year on RTKA survival. Kaplan-Meier survival curves were plotted with 95% confidence intervals (CIs) to determine the lifespan of RTKA. Multivariate Cox proportional hazards were used to estimate relative revision risks over time. Hazard ratios (HRs) were reported with 95% CI, and p-value < 0.05 was considered statistically significant.Aims
Methods
We have previously reported cryoablation-assisted joint-sparing surgery for osteosarcoma with epiphyseal involvement. However, it is not clear whether this is a comparable alternative to conventional joint arthroplasty in terms of oncological and functional outcomes. A total of 22 patients who had localized osteosarcoma with epiphyseal involvement around the knee and underwent limb salvage surgery were allocated to joint preservation (JP) group and joint arthroplasty (JA) group. Subjects were followed with radiographs, Musculoskeletal Tumor Society (MSTS) score, and clinical evaluations at one, three, and five years postoperatively.Aims
Methods
This is a retrospective review of the results of the Acclaim total elbow replacement in 11 older patients aged ≥ 65 years with primary osteoarthritis of the elbow, with a mean follow-up of 57.6 months (30 to 86.4). Significant reductions in pain and improvement in range of movement and function were recorded. Radiological review revealed two patients with 1 mm lucencies in a single zone, and one patient with 1 mm lucencies in two zones. No components required revision. There were no deep infections, dislocations or
We wanted to evaluate the effects of a bone anabolic agent (bone morphogenetic protein 2 (BMP-2)) on an anti-catabolic background (systemic or local zoledronate) on fixation of allografted revision implants. An established allografted revision protocol was implemented bilaterally into the stifle joints of 24 canines. At revision surgery, each animal received one BMP-2 (5 µg) functionalized implant, and one raw implant. One group (12 animals) received bone graft impregnated with zoledronate (0.005 mg/ml) before impaction. The other group (12 animals) received untreated bone graft and systemic zoledronate (0.1 mg/kg) ten and 20 days after revision surgery. Animals were observed for an additional four weeks before euthanasia.Aims
Methods
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
We used a biodegradable mesh to convert an acetabular defect into a contained defect in six patients at total hip replacement. Their mean age was 61 years (46 to 69). The mean follow-up was 32 months (19 to 50). Before clinical use, the strength retention and hydrolytic in vitro degradation properties of the implants were studied in the laboratory over a two-year period. A successful clinical outcome was determined by the radiological findings and the Harris hip score. All the patients had a satisfactory outcome and no
We reviewed ten patients with metastatic carcinoma and impending pathological fractures of the subtrochanteric region of the femur who had prophylactic insertion of the Russell-Taylor reconstruction nail. There were no operative complications or long-term
Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.Aims
Methods
1. The results of twenty-eight "hinge" arthroplasties of the knee are described. 2. Failure was most often due to breaking of the prosthesis. 3. The design of the prosthesis has been modified as a result of experience in cases of
The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes. A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).Aims
Methods
Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method. We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.Aims
Methods
We describe the use of intramedullary reconstruction nails in the treatment of 14 patients with pathological subtrochanteric fractures and coexisting metastases in the femoral shaft. After nailing, all patients were free from pain and regained mobility. They were followed up clinically and radiologically until death from the primary disease. There were no
We examined the mechanical properties of Vicryl (polyglactin 910) mesh in vitro and assessed its use in vivo as a novel biomaterial to attach tendon to a hydroxyapatite-coated metal implant, the interface of which was augmented with autogenous bone and marrow graft. This was compared with tendon re-attachment using a compressive clamp device in an identical animal model. Two- and four-ply sleeves of Vicryl mesh tested to failure under tension reached 5.13% and 28.35% of the normal ovine patellar tendon, respectively. Four-ply sleeves supported gait in an ovine model with 67.05% weight-bearing through the operated limb at 12 weeks, without evidence of
We compared the mechanical properties of carbon fibre composite bone plates with those of stainless steel and titanium. The composite plates have less stiffness with good fatigue properties. Tissue culture and small animal implantation confirmed the biocompatibility of the material. We also present a preliminary report on the use of the carbon fibre composite plates in 40 forearm fractures. All fractures united, 67% of them showing radiological remodelling within six months. There were no refractures or
The causes of
We have carried out a retrospective review of 11 Souter-Strathclyde primary total elbow arthroplasties in ten patients with osteoarthritis, over a period of nine years. The diagnosis was primary osteoarthritis in nine elbows and post-traumatic arthritis in two. The mean follow-up was 68 months (15 to 117). Although no patient was symptomatic, radiological review revealed evidence of loosening affecting three humeral and two ulnar components, one of which subsequently failed and was revised at 97 months. There were no dislocations, deep infections or
The Stanmore hinged total knee replacement was introduced in 1969 for severe destructive arthropathy of the knee, and the results of one hundred consecutive operations are presented after an average interval of two and a half years. Insertion of the prosthesis relieved pain in 94 per cent, improved the range of movement in 67 per cent, invariably restored stability and corrected valgus or varus deformity. Gross flexion contractures were improved but not always fully corrected. Serious complications were few, though of three cases of deep infection two came to amputation and one to fibrous ankylosis. There were no
1. A series of 100 intracapsular fractures of the femoral neck treated with the Charnley compression screw is reported. 2. There were six deaths, and in patients reviewed long enough (seventy-five) there were eleven undisplaced fractures, with successful union in 8l·8 per cent, and sixty-four displaced fractures, with a union rate of 59·6 per cent. 3. If we exclude six failures of operative technique, there remain fifty-eight patients in whom the fracture was judged radiologically to have united in 65·5 per cent. 4. The Charnley compression screw is an effective method of internal fixation in such fractures, and non-union with this method is usually due to causes other than
We reviewed the clinical and radiological results of 131 patients who underwent acetabular revision for aseptic loosening with impacted bone allograft and a cemented acetabular component. The mean follow-up was 51.7 months (24 to 156). The mean post-operative Merle D’Aubigné and Postel scores were 5.7 points (4 to 6) for pain, 5.2 (3 to 6) for gait and 4.5 (2 to 6) for mobility. Radiological evaluation revealed migration greater than 5 mm in four acetabular components. Radiological failure matched clinical failure. Asymptomatic radiolucent lines were observed in 31 of 426 areas assessed (7%). Further revision was required in six patients (4.5%), this was due to infection in three and
We reviewed 351 cemented revisions of femoral stems performed for aseptic loosening. At a mean of six years after the revision operation 72.1% of the patients were pain-free and 21.4% had only mild or occasional discomfort. The latest radiographs showed definite stem loosening in 10 (2.8%) and 20 hips (5.7%) had required rerevision, only nine (2.6%) of which were for
We studied 100 patients who had undergone endoprosthetic replacement of the proximal humerus between 1976 and 1998. The outcome was assessed with regard to the survivorship of the patients, the salvaged limbs and the prostheses. Function was determined in the 47 surviving patients, of whom 30 were assessed using the Musculo-Skeletal Tumour Society (MSTS) rating scale and 38 completed the Toronto Extremity Salvage Score (TESS) questionnaire. The median age of the patients was 34 years (10 to 80). The mean follow-up period for surviving patients was nine years (2 to 20). The mean MSTS score at follow-up was 79% and the mean TESS score was 72%. The length of bone which was resected influenced the functional outcome. Abduction of the shoulder was to 45° in most patients. The overall survival was 42% at ten years and that of the limb without amputation 93%. The survival of the prostheses using
Ten patients underwent endoprosthetic replacement of the distal humerus for bone tumours over a period of 30 years. There were eight primary and two secondary tumours in four men and six women with a mean age of 47.5 years (15 to 76). The mean follow-up was eight years (9 months to 31 years). Four patients required further surgery, three having revision for aseptic loosening; two of these and one other later needing a rebushing. There were no cases of postoperative nerve palsy, infection, local recurrence or
Of 135 consecutive patients admitted with subtrochanteric fractures of the femur, 104 were studied; half had internal fixation with a Zickel nail and half had a nail-plate device. Fractures treated with a Zickel nail had significantly less blood loss during operation and took less time to walk with full weight-bearing, but the incidence of technical error at operation was higher. Errors included malalignment of the cross-pin, penetration of the femoral cortex and comminution of the greater trochanter. There was
1. Two hundred and seventeen low-friction arthroplasties performed between November 1962 and April 1969 in 203 patients with failed previous operations have been analysed. 2. The technical details of operation in relation to these conversion problems have been outlined. 3. The quality of the results in relation to pain, mobility and ability to walk has been assessed before operation and one year afterwards, with small numbers at three and five years. 4. Pain was completely relieved or was minimal in 96·3 per cent of the patients, and in no case was the pain worse. 5. The total range of movement was 100 degrees or more in 98·5 per cent. No hip lost movement after conversion. 6. The failures were principally due to infection and to technical difficulties. There were no