Abstract
Aim
We present the medium-term clinical results of a reverse total shoulder arthroplasty with a trabecular metal glenoid base plate.
Patients and Methods
We reviewed 125 consecutive primary reverse total shoulder arthroplasties (RTSA) implanted in 124 patients for rotator cuff arthropathy. There were 100 women and 24 men in the study group with a mean age of 76 years (58 to 89). The mean follow-up was 32 months (24 to 60). No patient was lost to follow-up.
Results
There were statistically significant improvements in the mean range of movement and Oxford Shoulder Score (p < 0.001). Kaplan-Meier survivorship at five years was 96.7% (95% confidence interval 91.5 to 98.7) with aseptic glenoid failure as the end point.
Radiologically, 63 shoulders (50.4%) showed no evidence of notching, 51 (40.8%) had grade 1 notching, ten (8.0%) had grade 2 notching and one (0.8%) had grade 4 notching. Radiolucency around the glenoid base plate was found in one patient (0.8%) and around the humeral stem in five (4.0%). In all, three RTSA (2.4%) underwent revision surgery for aseptic mechanical failure of the glenoid within 11 months of surgery due to malseating of the glenosphere.
Conclusion
The clinical results of this large independent single unit series are comparable to those from previous series of RTSA reported in the literature. A trabecular metal base plate is safe and effective in the medium-term.
Cite this article: Bone Joint J 2016;98-B:969–75.