Aims

The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance.

Methods

A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.

Results

The mean follow-up was six years (2 to 14). PRNR started after a mean of 7.5 weeks (SD 2.1) and progressed significantly during the first two years, by the end of which the bone resorption stabilized. PRNR was detected in 81% (n = 45) of patients in Group 1 and 88% (n = 36) in Group 2. The final mean PRNR was 3.0 mm (SD 2.3) in Group 1 and 3.7 mm (SD 2.5) in Group 2. The mean MEPS, QuickDASH, and pASES-E were 95.9 (SD 11.5), 4.4 (SD 9.2), and 94.8 (SD 10.9) in Group 1 and 92.2 (SD 16.2), 9.9 (SD 21.5), and 90.8 (SD 15) in Group 2, respectively. No significant differences were observed between groups in the clinical and radiological outcomes. No correlations were found between PRNR and the clinical results.

Conclusion

PRNR after press-fit RHA is a common radiological finding that develops in the first 24 months before stabilizing definitively. PRNR does not affect the clinical results or implant survival in the mid term.

Cite this article: Bone Joint J 2023;105-B(8):905–911.