Vol. 102-B, Issue SUPP_2 | Pages 100 - 100
1 Feb 2020
Decker M Walzak M Khalili A Klassen R Teeter M McCalden R Lanting B
Full Access

Introduction

HXLPE acetabular liners were introduced to reduce wear-related complications in THA. However, post-irradiation thermal free radical stabilization can compromise mechanical properties, leave oxidation-prone residual free radicals, or both. Reports of mechanical failure of HXLPE acetabular liner rims raise concerns about thermal free radical stabilization and in vivo oxidization on implant properties. The purpose of this study is to explore the differences in the mechanical, physical and chemical properties of HXLPE acetabular liner rims after extended time in vivo between liners manufactured with different thermal free radical stabilization techniques.

Material and Methods

Remelted, single annealed and sequentially annealed retrieved HXLPE acetabular liners with in vivo times greater than 4.5 years were obtained from our implant retrieval laboratory. All retrieved liners underwent an identical sanitation and storage protocol. For mechanical testing, a total of 55 explants and 13 control liners were tested. Explant in vivo time ranged from 4.6 – 14.0 years and ex vivo time ranged from 0 – 11.6 years. Rim mechanical properties were tested by microindentation hardness testing using a Micromet II Vickers microhardness tester following ASTM standards. A subset of 16 explants with ex vivo time under one year along with five control liners were assessed for oxidation by FTIR, crystallinity by Raman spectroscopy, and evidence of microcracking by SEM.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 55 - 55
1 Feb 2020
Broberg J Howard J Lanting B Vasarhelyi E Yuan X McCalden R Naudie D Teeter M
Full Access

Introduction

Despite improvements in the survivorship of total knee replacements (TKR) over the years, patient satisfaction following TKR has not improved, with approximately 20% of patients recording dissatisfaction with their new knee joint. It is unclear why many patients feel this way, but it may relate in part to implant designs that do not provide a “natural” feeling knee. Implant manufacturers continue to introduce new concepts for implant design, which are essential for reaching the goal of a “normal” knee after TKR surgery. The Journey II TKR (Smith & Nephew) was developed with this goal in mind. Its anatomical design attempts to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Our objective is to examine patients receiving the Journey II TKR to measure the knee joint contact kinematics of the Journey II TKR compared to a non-anatomically designed implant by the same manufacturer. We hypothesize that the Journey II TKR will have more natural contact kinematics that differ from the non-anatomically designed implant.

Methods

A total of 28 individuals were recruited to receive a Journey II TKR, matching an existing prior cohort with a non-anatomical design from the same manufacturer (Legion TKR, Smith & Nephew). For both groups, a series of radiostereometric analysis (RSA) images were acquired at different knee flexion angles, ranging in 20° increments from 0° to 120°. Model-based RSA software (RSACore, Leiden, Netherlands) was used to obtain the 3D positions and orientations of the femoral and tibial implant components, which were in turn used to obtain kinematic measures (contact locations and magnitude of excursion) for each condyle. Results from the Journey II TKR group at 3 months post-operation were compared to the 2-year post-operative measurements from the Legion TKR group.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 10 - 10
1 Mar 2017
Sisko Z Teeter M Lanting B Howard J McCalden R Vasarhelyi E
Full Access

Purpose

Previous retrieval studies demonstrate increased tibial baseplate roughness leads to higher polyethylene backside wear in total knee arthroplasty (TKA). Micromotion between the polyethylene backside and baseplate is affected by the locking mechanism design and can further increase backside wear. This study's purpose was to examine modern locking mechanisms influence, in the setting of both polished and non-polished tibial baseplates, on backside tibial polyethylene damage and wear.

Methods

Five TKA models were selected with different tibial baseplate and/or locking mechanism designs. Six retrieval tibial polyethylenes from each TKA model were matched based on time in vivo (TIV), age at TKA revision, BMI, gender, number of times revised, and revision reason. Two observers visually assessed each polyethylene. Primary outcomes were visual damage scores, individual visual damage modes, and linear wear rates determined on micro-computed tomography (micro-CT) scan in mm/year. Demographics were compared by one-way ANOVA. Damage scores, damage modes, and linear wear were analyzed by the Kruskal-Wallis test and Dunn's multiple comparisons test.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 46 - 46
1 Mar 2017
Teeter M Howard J Vasarhelyi E Yuan X McCalden R Naudie D
Full Access

Background

Patient specific instrumentation (PSI) for total knee replacement (TKR) has demonstrated mixed success in simplifying the operation, reducing its costs, and improving limb alignment. Evaluation of PSI with tools such as radiostereometric analysis (RSA) has been limited, especially for cut-through style guides providing mechanical alignment. The primary goal of the present study was to compare implant migration following TKR using conventional and PSI surgical techniques, with secondary goals to examine whether the use of PSI reduces operative time, instrumentation, and surgical waste.

Methods

The study was designed as a prospective, randomized controlled trial of 50 patients, with 25 patients each in the PSI and conventional groups, powered for the RSA analysis. Patients in the PSI group received an MRI and standing 3-foot x-rays to construct patient-specific cut-through surgical guides for the femur and tibia with a mechanical alignment. All patients received the same posterior-stabilized implant, with marker beads inserted in the bone around the implants to enable RSA imaging. Intraoperative variables such as time, number of instrumentation trays used, and mass of surgical waste were recorded. Patients underwent supine RSA exams at multiple time points (2&6 weeks, 3&6 months and yearly) with 6 months data currently available. Migration of the tibial and femoral components was calculated using model-based RSA software. WOMAC, SF-12, EQ5D, and UCLA outcome measures were recorded pre-operatively and post-operatively.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 66 - 66
1 Mar 2017
MacLean C Lanting B Vasarhelyi E Naudie D McAuley J Howard J McCalden R MacDonald S
Full Access

Background

The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations.

Methods

A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidized zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the center index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 55 - 55
1 Dec 2016
Lanting B Thoren J Yuan X McCalden R McAuley J MacDonald S Vasarhelyi E Howard J Naudie D Teeter M
Full Access

Adequate fixation of implant components is an important goal for all arthroplasty procedures. Aseptic loosening is one of the leading causes of revision surgery in total knee arthroplasty. Radiostereometric analysis (RSA) is an imaging technique to measure implant migration, with established migration thresholds for well-fixed, at risk, and unacceptably migrating components. The purpose of the present study was to examine the long-term fixation of a cemented titanium fixed bearing polished tibial baseplate.

Patients enrolled in a previous two-year prospective trial were recalled at ten years. All patients received a cemented, posterior-stabilised total knee replacement of the same design implanted by one of three surgeons. Of the original 35 patients, 16 were available for long-term follow-up, with one patient lost to follow-up, nine patients deceased, and a further nine patients unwilling to return to the clinic. Each patient underwent RSA imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes as well as maximum total point motion (MTPM) was compared between all time points (baseline, six weeks, three months, six months, one year, two years) up to the ten year follow-up visits. Outcome scores including the Knee Society Score (KSS), WOMAC, SF-12, and UCLA Activity Score were recorded.

At ten years, the mean migrations of the tibial component were less than 0.1 mm and 0.1 degree in all planes relative to the post-operative RSA exam. There was no significant difference in tibial component migration between time points. However, MTPM increased significantly over time (p = 0.002), from 0.23 ± 0.18 mm at six weeks to 0.42 ± 0.20 mm at ten years. At one year, 13 patients had an acceptable MTPM level, three patients had an ‘at risk’ level, and no patient had an ‘unacceptable’ level. No patients were revised at ten years. WOMAC and KSS were significantly improved (p < 0.0001) at the latest follow-up compared to pre-operatively, but there was no difference in SF-12. The median UCLA Activity Score at latest follow-up was six (range, two to eight).

The tibial baseplate demonstrated solid fixation at ten years. No patients had an unacceptable MTPM level at one year and no patients were revised at ten years, supporting the use of RSA to predict long-term loosening risk. The low level of tibial baseplate migration found in the present study correlates to the low rate of revision for this implant as reported in individual studies and in joint replacement registries.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 84 - 84
1 Dec 2016
Nyland M Lanting B Somerville L Vasarhelyi E Naudie D McAuley J McCalden R MacDonald S Howard J
Full Access

Infection following total hip arthroplasty (THA) represents a devastating complication and is one of the main causes for revision surgery. This complication may be treated by irrigation and debridement with head and polyethylene exchange (IDHPE) or a two-stage revision (2SR). Previous studies have reported on the eradication success rates but few have reported patient outcome scores. The purpose of this study was to report patient outcome scores for both IDHPE and 2SR and compare these to a non-infected matched cohort. We hypothesised that both cohorts would have worse outcomes than the control group, and that those who failed an initial IDHPE and required a 2SR would have a worse outcome than those treated initially with a 2SR.

A retrospective review identified 137 patients from our institutional arthroplasty database who had an infected primary THA between 1986–2013. We excluded patients with less than one-year follow-up. Mean follow-up was 60 months (12–187 months). A control cohort was identified and matched according to age and Charlton Comorbidity Index (CCI). Harris Hip Scores, Short Form 12 and WOMAC scores were compared between our control group and our infected cohort.

Sixty-eight patients were treated with a 2SR and 69 patients were treated with an IDHPE. There was a 59% success rate in eradicating the infection with an IDHPE. All of the 28 patients who failed an IDHPE later went on to a 2SR. Outcome scores for the 2SR cohort were significantly worse than the non-infected controls (p0.05). There was no difference in outcome scores when comparing our 2SR cohort to our failed IDHPE (p>0.05).

Previous studies have focused on eradication rates. However, it is important to consider patient outcome scores when deciding the best treatment. Infected patients treated with a successful IDHPE had similar outcomes to non-infected patients. Patients that failed IDHPE and went onto 2SR had similar outcomes to those that had a 2SR alone. IDHPE should still be considered in the treatment algorithm of infected THA.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_21 | Pages 59 - 59
1 Dec 2016
Sisko Z Teeter M Lanting B Howard J McCalden R Naudie D MacDonald S Vasarhelyi E
Full Access

Previous retrieval studies demonstrate increased tibial baseplate roughness leads to higher polyethylene backside wear in total knee arthroplasty (TKA). Micromotion between the polyethylene backside and tibial baseplate is affected by the locking mechanism design and can further increase backside wear. The purpose of this study was to examine modern locking mechanisms, in the setting of both roughened and polished tibial baseplates, on backside tibial polyethylene wear.

Five TKA models were selected, all with different tibial baseplate and/or locking mechanism designs. Six retrieval tibial polyethylenes from each TKA model were matched based on time in vivo (TIV), age at TKA revision, BMI, gender, number of times revised, and revision reason. Two observers scored each polyethylene backside according to a visual damage score and individual damage modes. Primary outcomes were mean damage score and individual damage modes. Demographics were compared by one-way ANOVA. Damage scores and modes were analysed by the Kruskal-Wallis test and Dunn's multiple comparisons test.

There were no differences among the groups based on TIV (p=0.962), age (p=0.651), BMI (p=0.951), gender, revision number, or reason for revision. There was a significant difference across groups for mean total damage score (p=0.029). The polished tibial design with a partial peripheral capture locking mechanism and anterior constraint demonstrated a significantly lower score compared to one of the roughened tibial designs with a complete peripheral-rim locking mechanism (13.0 vs. 22.1, p=0.018). Otherwise, mean total damage scores were not significant between groups. As far as modes of wear, there were identifiable differences among the groups based on abrasions (p=0.005). The polished design with a tongue-in-groove locking mechanism demonstrated a significantly higher score compared to both groups with roughened tibial baseplates (5.83 vs. 0.83, p=0.024 and 5.83 vs. 0.92, p=0.033). Only the two designs with roughened tibial baseplates demonstrated dimpling (5.67 and 8.67) which was significant when compared against all other groups (p0.99). No other significant differences were identified when examining burnishing, cold flow, scratching, or pitting. No polyethylene components exhibited embedded debris or delamination.

Total damage scores were similar between all groups except when comparing one of the polished TKA design to one of the roughened designs. The other TKA model with a roughened tibial baseplate had similar damage scores to the polished designs, likely due to its updated locking mechanism. Dimpling wear patterns were specific for roughened tibial baseplates while abrasive wear patterns were identified in the design with a tongue-in-groove locking mechanism. Our study showed even in the setting of a roughened tibial baseplate, modern locking mechanisms decrease backside wear similar to that of other current generation TKA designs.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 59 - 59
1 Nov 2016
Goyal P Yuan X Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Lanting B Howard J
Full Access

Studies that have previously examined the relationship between inclination angle and polyethylene wear have shown increased wear of conventional polyethylene with high inclination angles. To date, there have been no long term in vivo studies examining the correlation between cup position and polyethylene wear with highly crosslinked polyethylene.

An institutional arthroplasty database was used to identify patients who had metal-on-highly crosslinked polyethylene primary total hip arthroplasty (THA) using the same component design with a minimum follow up of 10 years ago. A modified RSA examination setup was utilised, recreating standard anteroposterior (AP) and cross-table lateral exams in a single biplane RSA acquisition. Three dimensional head penetration was measured using the centre index method. The same radiographs were used to measure inclination angle and anteversion. Spearman correlation was used to show an association between the parameters of acetabular position and wear rate.

A total of 43 hips were included for analysis in this study. Average follow-up was 12.3 ± 1.2 years. The average linear wear rate was calculated to be 0.066 ± 0.066 mm/year. Inclination angle was not correlated with polyethylene wear rate (p=0.82). Anteversion was also not correlated with polyethylene wear rate (p=0.11). There was no statistical difference between wear rates of hips within Lewinnek's “safe zone” and those outside this “safe zone” (p=0.11). Males had a higher wear rate of 0.094 ± 0.089 mm/year compared to females with a wear rate of 0.046 ± 0.032 mm/year (p=0.045).

At long term follow up of greater than 10 years, highly cross linked polyethylene has very low wear rates. This excellent tribology is independent of acetabular position, but gender did impact wear rates. Due to the low wear rates, follow-up of even longer term is suggested to examine variables affecting wear.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 87 - 87
1 Nov 2016
Matz J Morden D Teeter M McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J Lanting B
Full Access

Complications involving the patellofemoral joint are a source of anterior knee pain, instability, and dysfunction following total knee arthroplasty. “Overstuffing” the patello-femoral joint refers to an increase in the thickness of the patellofemoral joint after a total knee replacement compared to the preoperative thickness. While biomechanical studies have indicated that overstuffing the patellofemoral joint may lead to adverse clinical outcomes, limited clinical evidence exists to support this notion. The purpose of this study is to evaluate the effect of changing the thickness of the patellafemoral joint on functional outcomes following total knee arthroplasty.

Our institutional arthroplasty database was used to identify 1347 patients who underwent a primary total knee arthroplasty between 2006 and 2012 with the same component design. Standard preoperative and postoperative anteroposterior, lateral, and skyline radiographs were collected and measured for patello-femoral overstuffing. These measurements included anterior patellar displacement, anterior femoral offset, and anteroposterior femoral size. These measurements were correlated with patient outcome data using WOMAC, KSS scores, and postoperative range of motion. Multiple linear regression analysis was used to assess the association between stuffing and functional outcomes.

A total of 1031 patients who underwent total knee arthroplasty were included. Increased anterior patellar displacement, a measure of patellofemoral joint thickness, was associated with decreased WOMAC scores (p=0.02). Anterior femoral offset (p=0.210) and anteroposterior femoral size (p=0.091) were not significantly associated with patient functional outcomes. Postoperative range of motion (ROM) was not associated with patellofemoral stuffing (p=0.190).

The current study demonstrated that functional outcomes are adversely affected by patellofemoral overstuffing. Based on these results, caution is encouraged against increasing the thickness of the patellofemoral joint, particularly on the patellar side of the joint.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 64 - 64
1 Nov 2016
Teeter M Naudie D McCalden R Yuan X MacDonald S
Full Access

The philosophy of cemented total hip arthroplasty (THA) femoral components has become polarised. At one extreme are polished, collarless, tapered devices that are expected to subside; at the other extreme are roughened, non-tapered implants with a collar designed not to subside. Radiostereometric analysis (RSA) allows the accurate measurement of implant movement and has been extensively used for measurement of the in vivo migration of implants. The degree of migration as measured by RSA during the first years after surgery has been shown to correlate with the long-term performance of cemented femoral implants. The purpose of this study was to review the two-year RSA results of two different designs of primary cemented THA stems.

Data from two previous prospective RSA trials with two-year follow-up were pooled. The first group included 36 patients who received a Spectron (Smith & Nephew, Memphis, USA) cemented stem. The second group included 13 patients who received an Exeter (Stryker, Mahwah, USA) cemented stem, and 15 patients who received a CPCS (Smith & Nephew, Memphis, USA) cemented stem. All patients underwent RSA examinations shortly post-operation, at 6 weeks, 3 months, 6 months, 1 years, and 2 years. Migration and rotation of the femoral stems was measured at each time point relative to the post-operative exam, and compared between the two groups.

There was no difference in age at surgery (Spectron 78 ± 6 years, Exeter/CPCS 77 ± 5 years, p = 0.43), BMI (Spectron and Exeter/CPCS 28 ± 5 kg/m2, p = 0.92), or percentage of male patients (Spectron 23% male, Exeter/CPCS 21% male) between the implant groups. Subsidence was significantly greater (p < 0.0001) at all time points from three months to two years for the Exeter and CPCS stems (0.94 ± 0.39 mm at two years) compared to the Spectron stem (0.05 ± 0.16 mm at two years). There was no significant difference between the stem types for medial-lateral translation (p = 0.07) or anterior-posterior translation (p = 0.49), or for anterior-posterior tilt (p = 0.15), internal-external rotation (p = 0.89), or varus-valgus rotation (p = 0.05).

Implant material, design, and surface finish are all factors in the long-term performance of cemented femoral hip implants. In this study, both femoral stem designs had a magnitude of sub-sidence that was within the limits of what is considered to be safe with respect to long-term performance. The continuous subsidence of the Exeter and CPCS stems is consistent with previous reports in the literature.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 57 - 57
1 Nov 2016
Lanting B Tan S Lau A Teeter M Del Balso C McCalden R MacDonald S Vasarhelyi E McAuley J Naudie D Howard J
Full Access

Trunnionosis in modular hip arthroplasty has recently been recognised to be clinically important. Gaining an understanding of how the material interface at the head-trunnion affects the tribology at the modular junctions has current clinical implications as well as an implication on future implant selection and material choice. This matched-cohort study aims to compare tribocorrosion between ceramic and cobalt-chromium trunnions and to investigate other factors that contribute to the difference in tribocorrosion if present.

All hip prostheses retrieved between 1999 and 2015 at one centre were reviewed. Fifty two ceramic heads were retrieved, and these were matched to a cobalt-chromium cohort according to taper design, head size, neck length and implantation time in that order. The two cohorts were similar in male:female ratio (p=0.32) and body mass index (p=0.15) though the ceramic group was younger than the cobalt-chromium group (56.6 (+/−)13.5 years for ceramic group vs 66.3 (+/−14.4) years for cobalt-chromium group; p=0.001). There was no significant difference in the reasons for revision between the two groups (p=0.42). The femoral head trunnions were examined by two independent observers using a previously published 4-point scoring technique. The trunnions were divided into three zones: apex, middle and base. The observers were blinded to clinical and manufacturing data where possible.

Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (p<0.001), middle zone (p<0.001) and in the combined score (p<0.001). In a subgroup analysis by head size, ceramic heads had a lower fretting and corrosion score at 28mm head diameter (p<0.001). Within the ceramic group, taper design had a significant effect on fretting and corrosion in the apex zone (p=0.04). Taper design also had a similar effect in the cobalt-chromium group in the apex zone (p=0.03). For the ceramic trunnions, the largest effect was contributed by the difference between the 11/13 taper and the 12/14 taper. For the cobalt-chromium trunnions, the largest effect was contributed by the difference between the 5 degree 38′ 37″ taper and type 1 taper.

Ceramic head trunnions showed a significantly lower fretting and corrosion score as compared to cobalt-chromium trunnions. Ceramic heads had a lower score than cobalt-chromium heads at 28mm head diameter. Taper design had an effect on fretting and corrosion within each cohort.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 62 - 62
1 Nov 2016
MacLean C Vasarhelyi E Lanting B Naudie D Somerville L McCalden R McAuley J MacDonald S Howard J Yuan X Teeter M
Full Access

The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations.

A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidised zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the centre index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts.

A total of 68 patients underwent RSA analysis. Fifteen patients with a ceramic femoral head component and 14 patients with an Oxinium femoral head component along with the same number of matched patients with cobalt chrome femoral head component were included in the analysis. The time in vivo for the Oxinium (5.17 +/− 0.96 years), Oxinium matched cohort (5.13 +/− 0.72 years), ceramic (5.15 +/− 0.76 years) and ceramic matched cohort (5.36 +/− 0.63 years) were comparable. The demographics of all bearing surface cohorts were similar. The paired comparison between the Oxinium and cobalt chrome cohorts (0.33 vs. 0.29 mm/year, p=0.284) and ceramic vs cobalt chrome cohorts (0.26 vs. 0.20 mm/year, p=0.137) did not demonstrate a significant difference in wear rate. The independent groups analysis revealed a significantly higher wear rate of Oxinium (0.33 mm/year) compared to cobalt chrome (0.24 mm/year) (p = 0. 038). There were no differences in HHS and WOMAC scores between the Oxinium and cobalt chrome cohorts (HHS: p = 0.71, WOMAC: p=0.08) or the ceramic and cobalt chrome cohorts (HHS: p=0.15, WOMAC: p=023).

This study presents evidence of a greater wear rate (mm/year) of the Oxinium femoral head component compared to a cobalt chrome femoral head component. This difference was not demonstrated in the ceramic femoral head component. Despite this difference, there were no clinical differences as measured by the HHS and WOMAC. Future research should focus on factors that may contribute to the higher wear rate seen in the Oxinium cohort.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_20 | Pages 58 - 58
1 Nov 2016
Neely S Berta D Ralley F Lanting B Vasarhelyi E McAuley J McCalden R MacDonald S Naudie D Howard J
Full Access

Total joint arthroplasty is commonly associated with post-operative anemia. Blood conservation programs have been developed to optimise patients prior to surgery. Epoetin Alfa (Eprex) or intravenous (IV) iron transfusions are two modalities that can be used pre-operatively to optimise hemoglobin and ferritin levels. There are, however, potential complications and increased costs associated with their use. Oral iron is a less costly option for those undergoing surgery but requires more time to take effect. There are no studies to date that examine the effects of an early screening program utilising oral iron supplementation prior to total joint arthroplasty. The purpose of this study is to evaluate the effect of implementing early pre-operative oral iron supplementation on patients prior total joint arthroplasty.

A retrospective review of patients undergoing total joint arthroplasty was performed using our institution clinical informatics database. We identified all patients seen in pre-admission clinic (PAC) between Jan 1, 2009 and March 31, 2010 representing our control group. We then identified all patients seen in PAC between October 1, 2012 and December 31, 2013. Patients in this cohort received screening blood work when booked for surgery, and oral iron supplementation was given to patients with hemoglobin of less than 135g/L or ferritin less than 100ug/L, thus representing our treatment group. Patients undergoing revision, uni-compartment knee arthroplasty and bilateral arthroplasties were excluded from the study. Pearson Chi-Square tests were used to calculate significance between groups with main outcomes including pre-admission hemoglobin, and pre-operative requirements for Eprex or IV iron.

In our control group, we identified 354 patients (25.6%) with hemoglobin less than 130 g/L at time of pre-admission clinic. In our treatment group, this number dropped significantly to only 16.4% of patients (p<0.005).

Implementation of an early screening program using oral iron supplementation resulted in a decrease in the number of patients with hemoglobin lower than 130 g/L at the time of pre-admission clinic. There was also a significant decrease in the use of Eprex and IV iron pre-operatively in the patients in the early screening program. These results encourage the use of early oral iron supplementation for patients with hemoglobin less than 135 g/L or ferritin less than 100ug/L in order to optimise patients prior to total joint arthroplasty.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 29 - 29
1 Sep 2012
MacDonald SJ Somerville L Naudie DD McAuley J McCalden R Bourne RB
Full Access

Purpose

Unicompartmental versus total knee arthroplasty has been a debated topic for decades. The purpose of this study was to compare the survivorship and clinical outcomes of a large primary total knee arthroplasty versus unicompartmental knee arthroplasty cohort.

Method

A consecutive series of 6352 TKAs and 296 UKAs with a minimum of one year follow-up were evaluated. Pre-operative scores, latest scores, and change in clinical outcome scores (KSCRS, SF12, WOMAC) were compared and tested for significance using the students t-test.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 115 - 115
1 Sep 2012
MacDonald SJ Somerville L Howard J Naudie DD McAuley J McCalden R Bourne R
Full Access

Purpose

The pros and cons of general anesthesia versus spinal anesthesia in total hip arthroplasty has been a long debated topic. The purpose of this study was to compare the surgical times, blood loss and transfusion requirements between anesthetic types in patients undergoing primary total hip arthroplasty.

Method

A consecutive series of 1600 THA procedures with complete preoperative and postoperative data were evaluated. Twenty eight percent of procedures were performed with a general anesthetic (GA), 67% with a spinal anesthetic (SP) and 5% with a combination of the two. Outcomes were compared and tested for significance using the Independent Samples Kruskal Wallis or Pearson Chi-Square analysis.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 24 - 24
1 Sep 2012
McCalden R Martin R Somerville L Bourne RB MacDonald SJ
Full Access

Purpose

The purpose of this study was to compare the outcomes and complications of normal weight, class one obese (BMI=30-34.9 kg/m2) and superobese (BMI≥50 kg/m2) primary total knee replacement (TKR) patients.

Method

Between January 1995 and December 2005, 4104 primary TKR were performed in our centre. Patients were stratified for their obesity level according the World Health Organisation (WHO) and current surgical literature classification. Sixty-seven superobese patients (group one) were identified. They were matched with normal weight (group two) and class one obese (group three) primary TKR patients for sex, age, side, preoperative diagnosis, implants used, time since surgery and preoperative SF-12 mental component summary score (MCS). Patient clinical outcomes (WOMAC, SF-12 and Knee Society scores), radiographs, complications, readmissions, and revisions rates were compared for normal BMI, class one obese and superobese patients.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 27 - 27
1 Sep 2012
McAuley JP Lyons M Howard J McCalden R Naudie DD Bourne RB MacDonald SJ
Full Access

Purpose

The patella provides a mechanical advantage to the knee extensor mechanism. Patellectomy, performed for trauma or patellofemoral arthrosis, does not preclude the development of tibiofemoral arthrosis. Total knee arthroplasty is the mainstay of treatment for tibiofemoral arthrosis. The purpose of this study was to evaluate the outcomes of total knee arthoplasty in patients who previously underwent patellectomy.

Method

A retrospective analysis was completed on a prospectively collected database to identify all patients who underwent total knee arthroplasty following a previous patellectomy. Sixty-one total knee arthroplasties in 57 patients were identified. Patient demographics as well as functional outcome scores, including WOMAC and Knee Society Scores, were evaluated.