A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
There remains concern with the use of constrained liners (CL) implanted at the time of acetabular
The increasing number of total hip arthroplasty (THA) used in young patients will inevitably lead to more revision procedures at younger ages, especially since the outcome of primary THA in young patients is already inferior compared to older patients. However, these data are lacking in literature. The aim of this study was to determine the survival of both acetabular and femoral components placed during primary and revision hip arthroplasty in patients under 55 years using Dutch Arthroplasty Register (LROI) data. All primary THA registered in the LROI between 2007–2018 in patients under 55 years were selected (n=25,682). Subsequent cup- and stem revision procedures were included. Kaplan-Meier survival analyses were used to estimate the survival probability of primary and revised cup- and stem components. Mean follow-up of primary cups and stems was 5.8 years (SD 3.2) and 5.9 years (SD 3.2), respectively. In total, 659
Dual mobility (DM) components are increasingly used to prevent and treat dislocation after total hip arthroplasty (THA). Intraprosthetic dissociation (IPD) is a known rare complication of these implants and has reportedly decreased with modern implants. The purpose of this paper is to report the diagnosis and treatment of modern DM IPD. 1453 DM components were implanted between 2010 and 2021. 695 in primary and 758 in revision THA. 49 hips sustained a dislocation of the large head and 5 sustained an IPD at presentation. 6 additional IPD occurred at the time of reduction of large head. The average age was 64, 54% were female and the mean follow-up was three years. Of the 11 IPD, 8 had a history of instability, 5 had abductor insufficiency, 4 had prior lumbar fusion, and 3 were conversions from fracture. The overall IPD incidence was 0.76%. Ten of the 11 DM IPD were missed at initial presentation or at the time of reduction, and all were discharged with presumed reduction. The mean time from IPD to surgical treatment was 3 weeks. One patient died with an IPD at 5 months. A DM head was reimplanted in six, two underwent revision of the acetabular component with exchange of DM head, and four were revised to a constrained liner. The re-revision rate was 55% at a mean 1.8 years. None of the patients who underwent
In total hip arthroplasty (THA), cementless cup without screw holes has the putative benefits of maximizing host bone contact and reducing osteolysis by eliminating channels to backside wear particles. However, supplemental trans-acetabular screws cannot be used. 74 hips in 60 patients receiving same model of cementless cup without screw holes (Depuy Duraloc 100 HA cup) from 6/1999 to 3/2003 were prospectively followed up. All patients were allowed to have immediate full weight bearing. Age at THA was 53 ± 13 (range 24–74) years. Osteonecrosis was the leading hip disease (45% of hips). Survivorship was assessed using revision of the cup as the end point. Radiological parameters, including lateral opening angle, vertical and horizontal migration distances of the cups were measured. Paired t-test was used to compare between the measurements in early postoperative period and at final follow up. 51 hips were assessed at minimum 20 years follow-up. The mean follow-up was 22.6 (range 21 – 25) years. All the cups were well fixed. There were two
In 2022, approximately 60% of inserted cups and stems in Sweden utilized cemented fixation. Two predominant brands, Refobacin Bone Cement R and Palacos R+G, both incorporating gentamicin, were employed in over 90% of primary cemented Total Hip Arthroplasties (THAs) between 2012 and 2022. This study investigates whether the choice between these cement types affects the risk of revision. The five most frequently used cemented cups and the three most common stems were studied. Inclusion criteria encompassed hips with non-tumour diagnoses, operated through a direct lateral or posterior incision, featuring a 28–36 mm metal or ceramic head. Outcomes were assessed for
Background. Modular dual mobility (MDM) acetabular components are often used to prevent dislocation in revision total hip arthroplasty (THA). As there is insufficient data on these components, the outcomes were evaluated in a cohort with a mean follow-up time of greater than five years. Methods. Using the database of a single academic center, 126 revision THAs (117 patients) with one MDM were retrospectively reviewed. There were 94 hips in 88 patients with a mean follow-up time of 5.5 years. Survivorship analysis was performed with the endpoints of dislocation, reoperation for dislocation,
Highly porous tantalum cups have been used in complex acetabular revisions for nearly 20 years but reports of long term results are limited. This study was designed to report ten year results of revision using a single porous tantalum cup design with special attention to re-operation for any reason, all-cause revision, and revision for aseptic loosening. Retrospective review of all revision THA cases performed from 1999–2006 using a highly porous tantalum acetabular component design with multiple screw holes and a cemented polyethylene liner (Zimmer Biomet, Warsaw, IN). Our institutional medical record and total joint registry were used to assess follow-up and xrays were reviewed. The Paprosky classification system was used to rate acetabular bone loss. Radiographic loosening was defined as new/progressive radiolucencies in all 3 acetabular zones, or cup migration (>2mm). Kaplan-Meier survivorship was used to assess survivorship free of
With the approval of our institute, we reviewed all the robot-assisted hip revision during October 2019 and August 2021. MAKO joint arthroplasty system was used to perform the hip revision surgery. Seventy-one robot-assisted hip revision cases were included.
The goal of the study was to describe the features of the aseptic loosening of the cup in CoC THR and to determine factors that affect the time to revision. It is a retrospective study including all patients who had a revision of CoC THR for aseptic failure fixation of the cup, between 2007 and 2017. 55 patients (27 women, 28 men) (56 hips) were included in the study. Eight hips (13 %) had also a stem exchange. At the primary T HR, the mean age of the patients was 47.9 years (17–72), 28 press fit cups had screws, the mean diameter of the cup was 51.2 mm (46–62) and the mean inclination was 52° (37–67). Clinical and radiological data were retrospectively recorded by an observer different to the initial operators. The mean age of the patients at the revision was 55.4 years (26–84). The mean time to revision was 90.1 months (14–240), and was significantly greater in patients aged less than 52 years, in cups without screws and with a 28mm head. The trend curve of the time to
To compare complications, survivorship and results in 2 groups of Furlong-HAP Active, one with ceramic-ceramic and the other with metal-XLPE friction pair. Prospective, non-randomized, comparison of 2 series of JRI uncemented prosthesis, implanted with identical protocol by 1 surgeon in 1 hospital from 2006 to 2014. Friction pair was ceramic (Biolox Forte or Delta) in 35 patients of 53.7+/−10.6 years (25–69) (21 males, 60%), and CrCo-XLPE in 65 cases of 69.0+/−8.9 years (42–81) (36 males, 55%); there were significant differences in age (p<0.00001) but not in sex (p=0.6565). Head diameter: Ce-Ce with 19 of 28mm, 9 of 32 and 7 of 36mm; Me-PE with 63 of 28mm, 1 of 32 and 1 of 36. Follow-up averaged 10.5+/−3.1 years (1–15) in ceramic and 9.8+/−3.8 years (2–15) in XLPE group. Pearson, Fisher, Kolmogorov-Smirnov, Student, Mann-Whitney, calculated with the informatic tools Microsoft Excel 2007 and . https://www.socscistatistics.com/tests/. . Complications in ceramic joints: 2 late infections (Fisher exact test=0.6101) (1 DAIR, 1 one-stage exchange); 1 dislocation (Fisher exact test=0.2549) (closed reduction); 1 Vancouver C fracture (ORIF) (Fisher exact test=0.6548). Complications in Me-XLPE joints: 2 late infections (Fisher=0.6101) (1 DAIR, 1 two-stage exchange); 7 dislocations (Fisher=0.2549) (2 early, open reduction) (5 late: 3 closed reduction, 1
Iliopsoas impingement occurs in 4% to 30% of patients after undergoing total hip arthroplasty (THA). Despite a relatively high incidence, there are few attempts at modelling impingement between the iliopsoas and acetabular component, and no attempts at modelling this in a representative cohort of subjects. The purpose of this study was to develop a novel computational model for quantifying the impingement between the iliopsoas and acetabular component and validate its utility in a case-controlled investigation. This was a retrospective cohort study of patients who underwent THA surgery that included 23 symptomatic patients diagnosed with iliopsoas tendonitis, and 23 patients not diagnosed with iliopsoas tendonitis. All patients received postoperative CT imaging, postoperative standing radiography, and had minimum six months’ follow-up. 3D models of each patient’s prosthetic and bony anatomy were generated, landmarked, and simulated in a novel iliopsoas impingement detection model in supine and standing pelvic positions. Logistic regression models were implemented to determine if the probability of pain could be significantly predicted. Receiver operating characteristic curves were generated to determine the model’s sensitivity, specificity, and area under the curve (AUC).Aims
Methods
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
A revision for periprosthetic joint infection (PJI) in total hip arthroplasty (THA) has a major effect on the patient’s quality of life, including walking capacity. The objective of this case control study was to investigate the histological and ultrastructural changes to the gluteus medius tendon (GMED) in patients revised due to a PJI, and to compare it with revision THAs without infection performed using the same lateral approach. A group of eight patients revised due to a PJI with a previous lateral approach was compared with a group of 21 revised THAs without infection, performed using the same approach. The primary variables of the study were the fibril diameter, as seen in transmission electron microscopy (TEM), and the total degeneration score (TDS), as seen under the light microscope. An analysis of bacteriology, classification of infection, and antibiotic treatment was also performed.Aims
Methods
A randomized trial was designed to compare the outcome of ceramic-on-ceramic with ceramic on conventional polyethylene. These patients have been followed for 15 years. 58 hips in 57 patients under 60 years of age were randomized into one of two groups. Patients were blinded to the type of hip they received. Both groups of patients were treated routinely with prophylactic peri-operative antibiotics and low molecular weight Heparin. All patients were seen at six weeks, three months and annually after surgery. Clinical and radiologic assessment was carried out at each visit. Fifty-eight hips were available for analysis, 28 in the CoP group and 29 patients in the CoC group. Mean age of both groups was less than 45 years. There were seven revisions (16%) among the 58 patients enrolled in the study. In the CoP group four patients underwent revision with head and liner exchange for eccentric polyethylene wear 16 years post-implantation. In the CoC group one patient had a
Different pathologies, deformities, bone defects, previous surgeries and polyethylene wear limit the survival of total hip arthroplasty (THA) in young patients. We compare preoperative status and outcome in 171 young and very young patients (207 hips) who underwent a ceramic-on-ceramic THA. Sixty-three (77 hips) were less than 30 years old (group 1) and 108 (130) were between 30 and 40 years old. Mean follow-up was 11.3 years. Two-way ANOVA with repeated measures were used to analyse clinical and radiological changes. Juvenile rheumatoid arthritis (JRA) was the most frequent diagnosis in group 1 and avascular necrosis in group 2. Charnley class type C, low activivity, previous surgery and osteoporotic bone were more frequent in group 1 patients. There were 2
We analysed the results of different strategies in the revision of primary uncemented acetabular components reported to the Norwegian Arthroplasty Register. The aim was to compare the risk of further acetabular revision after isolated liner exchange and complete component revision. The results of exchanging well-fixed components were also compared with those of exchanging loose acetabular components. The period studied was between September 1987 and April 2005. The following groups were compared: group 1, exchange of liner only in 318 hips; group 2, exchange of well-fixed components in 398; and group 3, exchange of loose components in 933. We found that the risk of a further
Although cement in cement acetabular revision is a recognised option in the presence of a well-fixed cement mantle, partial cement mantle retention is not normally recommended or practiced. However, when revising a cemented acetabular cup it is not infrequent to be faced with loose superolateral cement but well-fixed medial cement. Removal of the well-fixed cement can be time consuming and destructive. An alternative would be to retain this cement and incorporate it into the reconstruction. This study assesses the practice and results of partial cement mantle retention (PCR) at acetabular revision. We retrospectively identified a cohort of 28 hips in 26 patients using the PCR technique from 1. st. January 2000 to 1. st. January 2013. This represented 3.3% of
Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Using the database of
a single academic centre, 126 revision THAs in 117 patients using a single
design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.Aims
Methods
Press-fit fixation of uncemented acetabular components is increasingly being used for total hip replacement (THR) surgery. This study was aimed to analyse the survival of an established, un-cemented flexible metal backed cup with non-highly cross-linked polyethylene in primary and simple revision THR. We assessed the outcome of 1703 consecutive cementless elastic acetabular cups (Atlas, FH France) with non-highly cross-linked polyethylene (GUR1030) in total hip replacement surgery. There were 1582 patients (513 males and 1079 females) with mean age of 74 years (36–97 years). Seventeen senior operating surgeons accounted for 88.5% of cases from a single arthroplasty unit. Survival analysis was calculated using the Kaplan Meier Estimator. At 184 months post surgery the survivorship was 92.4% (95% CI 88.3–94.7) with