Abstract
To compare complications, survivorship and results in 2 groups of Furlong-HAP Active, one with ceramic-ceramic and the other with metal-XLPE friction pair
Prospective, non-randomized, comparison of 2 series of JRI uncemented prosthesis, implanted with identical protocol by 1 surgeon in 1 hospital from 2006 to 2014. Friction pair was ceramic (Biolox Forte or Delta) in 35 patients of 53.7+/−10.6 years (25–69) (21 males, 60%), and CrCo-XLPE in 65 cases of 69.0+/−8.9 years (42–81) (36 males, 55%); there were significant differences in age (p<0.00001) but not in sex (p=0.6565).
Head diameter: Ce-Ce with 19 of 28mm, 9 of 32 and 7 of 36mm; Me-PE with 63 of 28mm, 1 of 32 and 1 of 36. Follow-up averaged 10.5+/−3.1 years (1–15) in ceramic and 9.8+/−3.8 years (2–15) in XLPE group. Pearson, Fisher, Kolmogorov-Smirnov, Student, Mann-Whitney, calculated with the informatic tools Microsoft Excel 2007 and https://www.socscistatistics.com/tests/.
Complications in ceramic joints: 2 late infections (Fisher exact test=0.6101) (1 DAIR, 1 one-stage exchange); 1 dislocation (Fisher exact test=0.2549) (closed reduction); 1 Vancouver C fracture (ORIF) (Fisher exact test=0.6548). Complications in Me-XLPE joints: 2 late infections (Fisher=0.6101) (1 DAIR, 1 two-stage exchange); 7 dislocations (Fisher=0.2549) (2 early, open reduction) (5 late: 3 closed reduction, 1 cup revision, 1 constrained cup); 4 Vancouver B fractures (Fisher=0.6548) (2 intraoperative, cerclages; 2 late, exchange).
Final follow-up: Harris Hip Score averaged 93.2+/−13.7 (23–100) in ceramic and 94.3+/−8.7 (65–100) in XLPE joints (p=0.64552).
Wear: 0.06+/−0.38mm (0–1.5) in ceramic and 0.16+/−0.5mm (0–2) in Me-PE THAs (p=0.30302).
Osteolysis in Charnley-De Lee zones: 8 zones (6 patients) (17%) in ceramic cups, 25 zones (15 patients) (23%) in XLPE cups (p=0.980127).
Survivorship without any surgery or closed reduction after 15 years: 91.0% in ceramic joints, 83.8% in Me-XLPE joints.
Survivorship without component exchange after 15 years: 93.9% in ceramic joints, 93.6% in Me-XLPE joints.
At least after 10 years follow-up of Furlong-HAP Active, metal-XLPE and ceramic-ceramic joints present no significant differences in complications, clinical score, wear, acetabular osteolysis, or survivorship without component exchange. On the contrary, survivorship without any surgery or closed reduction is different because of the high rate of dislocation in 28mm metal-poly joints.