header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Hip

USE OF CEMENTLESS ELASTIC ACETABULAR CUPS WITH NON-HIGHLY CROSS-LINKED POLYETHYLENE IN TOTAL HIP REPLACEMENT

British Hip Society meeting (BHS) March 2016



Abstract

Press-fit fixation of uncemented acetabular components is increasingly being used for total hip replacement (THR) surgery. This study was aimed to analyse the survival of an established, un-cemented flexible metal backed cup with non-highly cross-linked polyethylene in primary and simple revision THR.

We assessed the outcome of 1703 consecutive cementless elastic acetabular cups (Atlas, FH France) with non-highly cross-linked polyethylene (GUR1030) in total hip replacement surgery. There were 1582 patients (513 males and 1079 females) with mean age of 74 years (36–97 years). Seventeen senior operating surgeons accounted for 88.5% of cases from a single arthroplasty unit. Survival analysis was calculated using the Kaplan Meier Estimator. At 184 months post surgery the survivorship was 92.4% (95% CI 88.3–94.7) with cup revision for any cause as an endpoint. Three patients died within first month of surgery and 268 patients (17.13%) died due to unrelated causes.

Overall, 59 patients had revisions (3.5%), Thirty eight patients (2.3%) were due to acetabular causes such as: aseptic loosening in 6 patients (0.35%); hip instability in 13 patients (0.76%); and deep infection in 11 patients (0.64%).

We believe the cementless elastic acetabular cup on in total hip replacement is tolerant of technique and patient variation and continues to give excellent long-term results and its continued widespread use can be recommended.