CEMENTLESS ACETABULAR CUP WITHOUT SCREW HOLES: A MINIMUM 20-YEAR FOLLOW-UP STUDY

The International Hip Society (IHS) 2024 Closed Meeting, Athens, Greece, 15–18 May 2024.



Abstract

In total hip arthroplasty (THA), cementless cup without screw holes has the putative benefits of maximizing host bone contact and reducing osteolysis by eliminating channels to backside wear particles. However, supplemental trans-acetabular screws cannot be used.

74 hips in 60 patients receiving same model of cementless cup without screw holes (Depuy Duraloc 100 HA cup) from 6/1999 to 3/2003 were prospectively followed up. All patients were allowed to have immediate full weight bearing. Age at THA was 53 ± 13 (range 24–74) years. Osteonecrosis was the leading hip disease (45% of hips). Survivorship was assessed using revision of the cup as the end point. Radiological parameters, including lateral opening angle, vertical and horizontal migration distances of the cups were measured. Paired t-test was used to compare between the measurements in early postoperative period and at final follow up.

51 hips were assessed at minimum 20 years follow-up. The mean follow-up was 22.6 (range 21 – 25) years. All the cups were well fixed. There were two cup revisions. Conventional polyethylene (PE) was used in both hips; osteolysis occurred 17 and 18 years later. Both cups were well fixed but were revised, one due to cup mal-positioning, one due to need in upsizing the articulation. 14 other hips were revised but these cups were well fixed and not revised; 9 loosened stems (most were cemented Elite plus stems), 5 PE wear and osteolysis (all were conventional PE). At 20 years, the survivorship of cups was 96.1%. Changes in lateral opening angle, vertical and horizontal migration distances were 0.44±1.59°, 0.01±1.52mm and -0.32±1.47mm respectively, without statistical significance.

This study provided evidence of excellent long-term survivorship of cementless cup without screw holes. Immediate postoperative weight-bearing also did not lead to cup migration in the long-term.