Abstract
There remains concern with the use of constrained liners (CL) implanted at the time of acetabular cup revision in revision total hip replacement (rTHA). The aim of this study was to determine the implant survival in rTHA when a CL was implanted at the same time as acetabular cup revision.
We reviewed our institutional database to identify all consecutive rTHAs where a CL was implanted simultaneously at the time acetabular cup revision from 2001 to 2021. One-hundred and seventy-four revisions (173 patients) were included in the study. Mean follow-up of 8.7 years (range two – 21.7). The most common indications for rTHA were instability (35%), second-stage periprosthetic joint infection (26.4%), and aseptic loosening (17.2%).
Kaplan Meier Analysis was used to determine survival with all-cause re-revision and revision for cup aseptic loosening (fixation failure) as the endpoints.
A total of 32 (18.3%) patients underwent re-revision at a mean time of 2.9 years (range 0.1 – 14.1). The most common reasons for re-revision were instability (14), periprosthetic joint infection (seven), and loosening of the femoral component (four). Three (1.7%) required re-revision due to aseptic loosening of the acetabular component (fixation failure) at a mean of two years (0.1 – 5.1). Acetabular component survival free from re-revision due to aseptic loosening was 98.9% (95% CI 97.3 – 100) at five-years and 98.1% (95% CI 95.8 – 100) at 10-years. There were no acetabular component fixation failures in modern highly porous shells.
CLs implanted at the time acetabular cup revision in rTHA have a 98.1% 10-year survival free from acetabular cup aseptic loosening (fixation failure). There were no cup fixation failures in modern highly porous shells. Thus, when necessary, implanting a CL during revision of an acetabular component with stable screw fixation is safe with an extremely low risk of cup fixation failure.