Bone & Joint Research
Vol. 6, Issue 2 | Pages 90 - 97
1 Feb 2017
Rajfer RA Kilic A Neviaser AS Schulte LM Hlaing SM Landeros J Ferrini MG Ebramzadeh E Park S

Objectives. We investigated the effects on fracture healing of two up-regulators of inducible nitric oxide synthase (iNOS) in a rat model of an open femoral osteotomy: tadalafil, a phosphodiesterase inhibitor, and the recently reported nutraceutical, COMB-4 (consisting of L-citrulline, Paullinia cupana, ginger and muira puama), given orally for either 14 or 42 days. Materials and Methods. Unilateral femoral osteotomies were created in 58 male rats and fixed with an intramedullary compression nail. Rats were treated daily either with vehicle, tadalafil or COMB-4. Biomechanical testing of the healed fracture was performed on day 42. The volume, mineral content and bone density of the callus were measured by quantitative CT on days 14 and 42. Expression of iNOS was measured by immunohistochemistry. Results. When compared with the control group, the COMB-4 group exhibited 46% higher maximum strength (t-test, p = 0.029) and 92% higher stiffness (t-test, p = 0.023), but no significant changes were observed in the tadalafil group. At days 14 and 42, there was no significant difference between the three groups with respect to callus volume, mineral content and bone density. Expression of iNOS at day 14 was significantly higher in the COMB-4 group which, as expected, had returned to baseline levels at day 42. Conclusion. This study demonstrates an enhancement in fracture healing by an oral natural product known to augment iNOS expression. Cite this article: R. A. Rajfer, A. Kilic, A. S. Neviaser, L. M. Schulte, S. M. Hlaing, J. Landeros, M. G. Ferrini, E. Ebramzadeh, S-H. Park. Enhancement of fracture healing in the rat, modulated by compounds that stimulate inducible nitric oxide synthase: Acceleration of fracture healing via inducible nitric oxide synthase. Bone Joint Res 2017:6:–97. DOI: 10.1302/2046-3758.62.BJR-2016-0164.R2


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 97 - 97
4 Apr 2023
van Knegsel K Zderic I Kastner P Varga P Gueorguiev B Knobe M Pastor T
Full Access

Recently, a new suture was designed to minimize laxity in order to preserve consistent tissue approximation while improving footprint compression after tendon repair. The aims of this study were: (1) to compare the biomechanical competence of two different high strength sutures in terms of slippage and failure load, (2) to investigate the influence of both knots number and different media (air, saline and fat) on the holding capacity of the knots. Alternating surgical knots of two different high-strength sutures (group1: FibreWire; group2: DynaCord; n = 105) were tied on two roller bearings with 50N tightening force. Biomechanical testing was performed in each medium applying ramped monotonic tension to failure defined in terms of either knot slippage or suture rupture. For each group and medium, seven specimens with either 3, 4, 5, 6, or 7 knots each were tested, evaluating their knot slippage and ultimate load to failure. The minimum number of knots preventing slippage failure and thus resulting in suture rupture was determined in each group and medium, and taken as a criterium for better performance when comparing the groups. In each group and medium failure occurred via suture rupture in all specimens for the following minimum knot numbers: group1: air – 7, saline – 7, fat – 7; group2: air – 6; saline – 4; fat – 5. The direct comparison between the groups when using 7 knots demonstrated significantly larger slippage in group1 (6.5 ± 2.2 mm) versus group2 (3.5 ± 0.4 mm) in saline (p < 0.01) but not in the other media (p ≥0.52). Ultimate load was comparable between the two groups for all three media (p ≥ 0.06). The lower number of required knots providing sufficient repair stability, smaller slippage levels and identical suture strength, combined with the known laxity alleviation effect demonstrate advantages of DynaCord versus FibreWire


Full Access

Matrix metalloproteinase enzymes (MMPs) play a crucial role in the remodeling of articular cartilage, contributing also to osteoarthritis (OA) progression. The pericellular matrix (PCM) is a specialized space surrounding each chondrocyte, containing collagen type VI and perlecan. It acts as a transducer of biomechanical and biochemical signals for the chondrocyte. This study investigates the impact of MMP-2, -3, and -7 on the integrity and biomechanical characteristics of the PCM. Human articular cartilage explants (n=10 patients, ethical-nr.:674/2016BO2) were incubated with activated MMP-2, -3, or -7 as well as combinations of these enzymes. The structural degradative effect on the PCM was assessed by immunolabelling of the PCM's main components: collagen type VI and perlecan. Biomechanical properties of the PCM in form of the elastic moduli (EM) were determined by means of atomic force microscopy (AFM), using a spherical cantilever tip (2.5µm). MMPs disrupted the PCM-integrity, resulting in altered collagen type VI and perlecan structure and dispersed pericellular arrangement. A total of 3600 AFM-measurements revealed that incubation with single MMPs resulted in decreased PCM stiffness (p<0.001) when compared to the untreated group. The overall EM were reduced by ∼36% for all the 3 individual enzymes. The enzyme combinations altered the biomechanical properties at a comparable level (∼36%, p<0.001), except for MMP-2/-7 (p=0.202). MMP-induced changes in the PCM composition have a significant impact on the biomechanical properties of the PCM, similar to those observed in early OA. Each individual MMP was shown to be highly capable of selectively degrading the PCM microenvironment. The combination of MMP-2 and -7 showed a lower potency in reducing the PCM stiffness, suggesting a possible interplay between the two enzymes. Our study showed that MMP-2, -3, and -7 play a direct role in the functional and structural remodeling of the PCM. Acknowledgements: This work was supported by the Faculty of Medicine of the University of Tübingen (grant number.: 2650-0-0)


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 12 - 12
17 Apr 2023
Van Oevelen A Burssens A Krähenbühl N Barg A Audenaert E Hintermann B Victor J
Full Access

Several emerging reports suggest an important involvement of the hindfoot alignment in the outcome of knee osteotomy. At present, studies lack a comprehensive overview. Therefore, we aimed to systematically review all biomechanical and clinical studies investigating the role of the hindfoot alignment in the setting of osteotomies around the knee. A systematic literature search was conducted on multiple databases combining “knee osteotomy” and “hindfoot/ankle alignment” search terms. Articles were screened and included according to the PRISMA guidelines. A quality assessment was conducted using the Quality Appraisal for Cadaveric Studies (QUACS) - and modified methodologic index for non-randomized studies (MINORS) scales. Three cadaveric, fourteen retrospective cohort and two case-control studies were eligible for review. Biomechanical hindfoot characteristics were positively affected (n=4), except in rigid subtalar joint (n=1) or talar tilt (n=1) deformity. Patient symptoms and/or radiographic alignment at the level of the hindfoot did also improve after knee osteotomy (n=13), except in case of a small pre-operative lateral distal tibia- and hip knee ankle (HKA) angulation or in case of a large HKA correction (>14.5°). Additionally, a pre-existent hindfoot deformity (>15.9°) was associated with undercorrection of lower limb alignment following knee osteotomy. The mean QUACS score was 61.3% (range: 46–69%) and mean MINORS score was 9.2 out of 16 (range 6–12) for non-comparative and 16.5 out of 24 (range 15–18) for comparative studies. Osteotomies performed to correct knee deformity have also an impact on biomechanical and clinical outcomes of the hindfoot. In general, these are reported to be beneficial, but several parameters were identified that are associated with newly onset – or deterioration of hindfoot symptoms following knee osteotomy. Further prospective studies are warranted to assess how diagnostic and therapeutic algorithms based on the identified criteria could be implemented to optimize the overall outcome of knee osteotomy. Remark: Aline Van Oevelen and Arne Burssens contributed equally to this work


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 20 - 20
1 May 2017
Fletcher J Evans S
Full Access

Background. Devices are frequently used to gain sufficient purchase in a bone so that the bone itself can be manipulated to move or rupture soft tissue attachments. During hip surgery, several different extraction corkscrews are available to remove the femoral head, each with a different screw design with no evidence to suggest which is most effective. Additionally during the use of corkscrew devices, often due to the low bone density, stripping of the screw threads out of the femoral head can occur prior to its extraction, thus requiring reinsertion. The aims of this project were to measure the primary pullout and reinsertional forces of five commercially available corkscrews. Methods. Polyurethane bone models covering a spectrum of osteoporotic to normal bone densities (0.08 gcm-3, 0.16 gcm-3 and 0.32 gcm-3) were used in axial tensile testing at two insertional depths to assess the maximum pullout force of these corkscrews (Zimmer, Stryker, Medacta, JRI, Depuy). Results. There are significant differences between the pullout forces produced throughout the different densities at both insertional depths; 0.08gcm-3 (p=0.002), 0.16 gcm-3 (p<0.001) and 0.32 gcm-3 (p=0.006). Reinsertional testing on all corkscrews demonstrated a reduction in the pullout force of approximately 70%, underlining the benefit of effective insertion and extraction on the first attempt. Conclusion. In all bone model densities, consistent differences in the pullout forces generated were identified. All the corkscrews generate forces greater than the estimated forces required to yield the soft tissues attached to a femoral head. Further work into optimising the corkscrew design, especially regarding the torque generated during insertion, will improve the efficacy of future corkscrew use. This information should be transferable to other situations where manipulation of a bone is needed at the temporary or permanent expense of the attached soft tissues. Level of Evidence. V: Biomechanical study


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 23 - 23
1 Mar 2021
Schopper C Zderic I Menze J Muller D Rocci M Knobe M Shoda E Richards G Gueorguiev B Stoffel K
Full Access

Femoral neck fractures account for half of all hip fractures and are recognized as a major public health problem associated with a high socioeconomic burden. Whilst internal fixation is preferred over arthroplasty for physiologically younger patients, no consensus exists about the optimal fixation device yet. The recently introduced implant Femoral Neck System (FNS) (DePuy Synthes, Zuchwil, Switzerland) was developed for dynamic fixation of femoral neck fractures and provides angular stability in combination with a minimally invasive surgical technique. Alternatively, the Hansson Pin System (HPS) (Swemac, Linköping, Sweden) exploits the advantages of internal buttressing. However, the obligate peripheral placement of the pins, adjacent to either the inferior or posterior cortex, renders the instrumentation more challenging. The aim of this study was to evaluate the biomechanical performance of FNS versus HPS in a Pauwels II femoral neck fracture model with simulated posterior comminution. Forty-degree Pauwels II femoral neck fractures AO 31-B2.1 with 15° posterior wedge were simulated in fourteen paired fresh-frozen human cadaveric femora, followed by instrumentation with either FNS or HPS in pair-matched fashion. Implant positioning was quantified by measuring the shortest distances between implant and inferior cortex (DI) as well as posterior cortex (DP) on anteroposterior and axial X-rays, respectively. Biomechanical testing was performed in 20° adduction and 10° flexion of the specimens in a novel setup with simulated iliopsoas muscle tension. Progressively increasing cyclic loading was applied until construct failure. Interfragmentary femoral head-to-shaft movements, namely varus deformation, dorsal tilting and rotation around the neck axis were measured by means of motion tracking and compared between the two implants. In addition, varus deformation and dorsal tilting were correlated with DI and DP. Cycles to 5/10° varus deformation were significantly higher for FNS (22490±5729/23007±5496) versus HPS (16351±4469/17289±4686), P=0.043. Cycles to 5/10° femoral head dorsal tilting (FNS: 10968±3052/12765±3425; HPS: 12244±5895/13357±6104) and cycles to 5/10° rotation around the femoral neck axis (FNS: 15727±7737/24453±5073; HPS: 15682±10414/20185±11065) were comparable between the implants, P≥0.314. For HPS, the outcomes for varus deformation and dorsal tilting correlated significantly with DI and DP, respectively (P=0.025), whereas these correlations were not significant for FNS (P≥0.148). From a biomechanical perspective, by providing superior resistance against varus deformation and performing in a less sensitive way to variations in implant placement, the angular stable Femoral Neck System can be considered as a valid alternative to the Hansson Pin System for the treatment of Pauwels II femoral neck fractures


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 7 - 7
1 Apr 2017
Karakaşlı A Ertem F Demirkıran N Bektaş Y Havıtçıoğlu H
Full Access

Background. Currently about 4–6% of all femur fractures consist of distal femoral fractures. Different methods and implants have been used for the surgical treatment of distal femoral fractures, including intramedullary nails. Retrograde nail. By contrast with antegrade nails, surgical approach or retrograde nailing exposes the knee joint which may lead to tendency of infection and increased knee pain. Present study aims to compare the biomechanical behaviour of distal angular condyler femoral intramedullary nail (DACFIN), retrograde nail and plate fixation. Methods. Fifteen 4th generation Saw bones were used to evaluate the biomechanical differences between the groups (Group 1: Plate fixation, Group 2: Retrograde nailing, Group 3: DACFIN; (n=5)). Biomechanical test was performed by using an electromechanical test device Shimadzu (AG-IS 5kN, Japan). Displacement values were recorded by using a Non-contact Video Extensometer (DVE-101/201, Shimadzu, Japan) during the loading each femur with 5 cycles of 500 N at a rate of 10 N/s to determine axial stiffness. The faliure stiffness was measured by axial load to each constructat a displacement rate of 5 mm/min. Torsional loading applied to all groups in amount of 6 Nm of torque with a velocity of 18 degrees/min. Results. The mean torsion stiffness value of Group 3 (6.33 Nm/degree) was signifacantly higher than Group 1 (1.18 Nm/degree) and Group 2 (2.11Nm/degree), p<0.05). The failure stiffness, Group 3 (1725 N/mm) was significantly higher than Group 1 (1275 N/ mm) and Group 2 (1290 N/mm). However, In axial stiffness, the mean value of Group 2 (2554 N/mm) was higher than Group 3 (1822 N/mm), and signifantly higher than Group 1(468 N/mm), p<0.05). Conclusions. DACFIN is more stiffer than retrograde nail and plate fixation during torsional and failure load conditions. But in axial stiffness retrograde nail was stiffer. DACFIN provide intramedullary femur condyle fracture fixations without open knee joint. Level of evidence. Level 5. Disclosure. Authors declare that there is no conflict of interest related to the present study


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 112 - 112
1 Apr 2017
Herregodts S De Baets P Victor J Verstraete M
Full Access

Background. Biomechanical joint contact pressure distribution measurements have proven to be a very valuable tool in orthopaedic research to investigate the influence of surgical techniques such as total knee arthroplasty (TKA) on the human knee joint. Quantification of the in vitro tibiofemoral and patellofemoral contact pressure distribution before and after the intervention are an important measure to evaluate the impact of the surgery. The K scan pressure sensor from Tekscan (South Boston USA) is a commonly reported device for these in vitro pressure measurements. Despite the large interest in the sensor, the effective measurement accuracy for in vitro biomechanical joint contact measurement still remains a big question and therefore the reliability of these measurements should be questioned. Methods. Reliable contact pressure measurements can only be done if the sensor behaviour is fully understood. Therefore, a tailored multi-axial testing machine has been designed to profoundly investigate and characterise the sensor behaviour. This test setup is unique through its ability to apply a predefined tangential force or sliding velocity to the sensor's interface next to a normal force. Dynamic effects occurring in knee joint motion can thus be simulated while evaluating the effect on the contact pressure measurements. Results. The change in contact friction coefficient by insertion of the sensor in the joint is quantified. Different interface conditions (dry, lubricated with PTFE spray, lubricated with surgical lubricant) have been evaluated to obtain the best sliding conditions and to minimise the undesired sensor accuracy deteriorating effects. Conclusion. An optimal calibration procedure is put forward and side-effects that deteriorate the measurement accuracy are quantified. The provided knowledge facilitates orthopaedic biomechanics researchers to optimally perform joint contact measurements and to estimate the effective measurement accuracy. In addition, the presented rig provides the opportunity to study the inherent knee kinematics and prosthesis shape optimisation through the inherent degrees of freedom in the rig


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 82 - 82
11 Apr 2023
Souleiman F Zderic I Pastor T Varga P Helfen T Richards G Gueorguiev B Theopold J Osterhoff G Hepp P
Full Access

Glenohumeral joint injuries frequently result in shoulder instability. However, the biomechanical effect of cartilage loss on shoulder stability remains unknown. The aim of the current study was to investigate biomechanically the effect of two severity stages of cartilage loss in different dislocation directions on shoulder stability. Joint dislocation was provoked for 11 human cadaveric glenoids in seven different dislocation directions between 3 o'clock (anterior) to 9 o'clock (posterior) dislocation. Shoulder stability ratio (SSR) and concavity gradient were assessed in intact condition, and after 3 mm and 6 mm simulated cartilage loss. The influence of cartilage loss on SSR and concavity gradient was statistically evaluated. Between intact state and 6 mm cartilage loss, both SSR and concavity gradient decreased significantly in every dislocation direction (p≤0.038), except the concavity gradient in 4 o'clock dislocation direction (p=0.088). Thereby, anterior-inferior dislocation directions were associated with the highest loss of SSR and concavity gradient of up to 59.0% and 49.4%, respectively, being significantly higher for SSR compared to all other dislocation directions (p≤0.04). The correlations between concavity gradient and SSR for pooled dislocation directions were significant for all three conditions of cartilage loss (p<0.001). From a biomechanical perspective, articular cartilage of the glenoid contributes significantly to the concavity gradient, correlating strongly with the associated loss in glenohumeral joint stability. The highest effect of cartilage loss was observed in anterior-inferior dislocation directions, suggesting that surgical intervention should be considered for recurrent shoulder dislocations in the presence of cartilage loss.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 41 - 41
4 Apr 2023
Benca E Zderic I van Knegsel K Caspar J Hirtler L Fuchssteiner C Strassl A Gueorguiev B Widhalm H Windhager R Varga P
Full Access

Odontoid fracture of the second cervical vertebra (C2) is the most common spinal fracture type in elderly patients. However, very little is known about the biomechanical fracture mechanisms, but could play a role in fracture prevention and treatment. This study aimed to investigate the biomechanical competence and fracture characteristics of the odontoid process.

A total of 42 human C2 specimens (14 female and 28 male, 71.5 ± 6.5 years) were scanned via quantitative computed tomography, divided in 6 groups (n = 7) and subjected to combined quasi-static loading at a rate of 0.1 mm/s until fracturing at inclinations of −15°, 0° and 15° in sagittal plane, and −50° and 0° in transverse plane. Bone mineral density (BMD), specimen height, fusion state of the ossification centers, stiffness, yield load, ultimate load, and fracture type according to Anderson and d'Alonzo were assessed.

While the lowest values for stiffness, yield, and ultimate load were observed at load inclination of 15° in sagittal plane, no statistically significant differences could be observed among the six groups (p = 0.235, p = 0.646, and p = 0.505, respectively). Evaluating specimens with only clearly distinguishable fusion of the ossification centers (n = 26) reveled even less differences among the groups for all mechanical parameters. BMD was positively correlated with yield load (R² = 0.350, p < 0.001), and ultimate load (R² = 0.955, p < 0.001), but not with stiffness (p = 0.070). Type III was the most common fracture type (23.5%).

These biomechanical outcomes indicate that load direction plays a subordinate role in traumatic fractures of the odontoid process in contrast to BMD which is a strong determinant of stiffness and strength. Thus, odontoid fractures appear to result from an interaction between load magnitude and bone quality.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 138 - 138
2 Jan 2024
Muñoz-Moya E Ruiz C Piella G Noailly J
Full Access

This study investigates the relationships between Intervertebral Disc (IVD) morphology and biomechanics using patient-specific (PS) finite element (FE) models and poromechanical simulations.

169 3D lumbar IVD shapes from the European project MySpine (FP7-269909), spanning healthy to Pfirrmann grade 4 degeneration, were obtained from MRIs. A Bayesian Coherent Point Drift algorithm aligned meshes to a previously validated structural FE mesh of the IVD. After mesh quality analyses and Hausdorff distance measurements, mechanical simulations were performed: 8 and 16 hours of sleep and daytime, respectively, applying 0.11 and 0.54 MPa of pressure on the upper cartilage endplate (CEP). Simulation results were extracted from the anterior (ATZ) and posterior regions (PTZ) and the center of the nucleus pulposus (CNP). Data mining was performed using Linear Regression, Support Vector Machine, and eXtreme Gradient Boosting techniques. Mechanical variables of interest in DD, such as pore fluid velocity (FLVEL), water content, and swelling pressure, were examined. The morphological variables of the simulated discs were used as input features.

Local morphological variables significantly impacted the local mechanical response. The local disc heights, respectively in the mid (mh), anterior (ah), and posterior (ph) regions, were key factors in general. Additionally, fluid transport, reflected by FLVEL, was greatly influenced (r2 0.69) by the shape of the upper and lower cartilage endplates (CEPs).

This study suggests that disc morphology affects Mechanical variables of interest in DD. Attention should be paid to the antero-posterior distribution and local effects of disc heights. Surprisingly, the CEP morphology remotely affected the fluid transport in NP volumes around mid-height, and mechanobiological implications shall be explored. In conclusion, patient-specific IVD modeling has strong potential to unravel important correlations between IVD phenotypes and local tissue regulation.

Acknowledgments: European Commission: Disc4All-MSCA-2020-ITN-ETN GA: 955735; O-Health-ERC-CoG-2021-101044828


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 25 - 25
17 Apr 2023
Kwak D Bae T Kim I
Full Access

The objective of this study was to analyze the biomechanical effect of an implanted ACL graft by determining the tunnel position according to the aspect ratio (ASR) of the distal femur during flexion-extension motion.

To analyze biomechanical characteristics according to the ASR of the knee joint, only male samples were selected to exclude the effects of gender and 89 samples were selected for measurement. The mean age was 50.73 years, and the mean height was 165.22 cm. We analyzed tunnel length, graft bending angle, and stress of the graft according to tunnel entry position and aspect ratio (ratio of antero-posterior depth to medio-lateral width) of the articular surface for the distal femur during single-bundle outside-in anterior cruciate ligament reconstruction surgery. We performed multi-flexible-body dynamic analyses with wherein four ASR (98, 105, 111, and 117%) knee models.

The various ASRs were associated with approximately 1-mm changes in tunnel length. The graft bending angle increased when the entry point was far from the lateral epicondyle and was larger when the ASR was smaller. The graft was at maximum stress, 117% ASR, when the tunnel entry point was near the lateral epicondyle. The maximum stress value at a 5-mm distance from the lateral epicondyle was 3.5 times higher than the 15-mm entry position and, the cases set to 111% and 105% ASR, showed 1.9 times higher stress values when at a 5-mm distance compared with a 15-mm distance. In the case set at 98% ASR, the low-stress value showed a without-distance difference from the lateral epicondyle.

Our results suggest that there is no relationship between the ASR and femoral tunnel length, A smaller ASR causes a higher graft bending angle, and a larger ASR causes greater stress in the graft.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 39 - 39
4 Apr 2023
Lim W Lie D Chou S Lie H Yew A
Full Access

This study aims to investigate the mechanical properties of a rotator cuff tear repaired with a polypropylene interposition graft in an ovine infraspinatus ex-vivo model.

Twenty fresh shoulders from skeletally mature sheep were used in this study. A tear size of 20 mm from the tendon joint was created in the infraspinatus tendon to simulate a large tear in fifteen specimens. This was repaired with a polypropylene mesh used as an interposition graft between the ends of the tendon. Eight specimens were secured with mattress stitches while seven were secured to the remnant tendon on the greater tuberosity side by continuous stitching. Remaining five specimens with an intact tendon served as a control group. All specimens underwent cyclic loading with a universal testing machine to determine the ultimate failure load and gap distance.

Gap distance increased with progressive cyclic loading through 3000 cycles for all repaired specimens. Mean gap distance after 3000 cycles for both continuous and mattress groups are 1.7 mm and 4.2 mm respectively (P = .001). Significantly higher mean ultimate failure load was also observed with 549.2 N in the continuous group, 426.6 N in the mattress group and 370 N in the intact group.

The use of a polypropylene mesh as an interposition graft for large irreparable rotator cuff tears is biomechanically suitable and results in a robust repair that is comparable to an intact rotator cuff tendon. When paired with a continuous suturing technique, it demonstrates significantly resultant superior biomechanical properties that may potentially reduce re-tear rates after repairing large or massive rotator cuff tears.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 27 - 27
1 Dec 2020
Gueorguiev B Zderic I Blauth M Weber A Koch R Dauwe J Schader J Stoffel K Finkemeier C Hessmann M
Full Access

Unstable distal tibia fractures are challenging injuries requiring surgical treatment. Intramedullary nails are frequently used; however, distal fragment fixation problems may arise, leading to delayed healing, malunion or nonunion. Recently, a novel angle-stable locking nail design has been developed that maintains the principle of relative construct stability, but introduces improvements expected to reduce nail toggling, screw migration and secondary loss of reduction, without the requirement for additional intraoperative procedures.

The aim of this study was to investigate the biomechanical competence of a novel angle-stable intramedullary nail concept for treatment of unstable distal tibia fractures, compared to a conventional nail in a human cadaveric model under dynamic loading.

Ten pairs of fresh-frozen human cadaveric tibiae with a simulated AO/OTA 42-A3.1 fracture were assigned to 2 groups for reamed intramedullary nailing using either a conventional (non-angle-stable) Expert Tibia Nail with 3 distal screws (Group 1) or the novel Tibia Nail Advanced system with 2 distal angle-stable locking low-profile screws (Group 2). The specimens were biomechanically tested under conditions including quasi-static and progressively increasing combined cyclic axial and torsional loading in internal rotation until failure of the bone-implant construct, with monitoring by means of motion tracking.

Initial axial construct stiffness, although being higher in Group 2, did not significantly differ between the 2 nail systems, p=0.29. In contrast, initial torsional construct stiffness was significantly higher in Group 2 compared to Group 1, p=0.04. Initial nail toggling of the distal tibia fragment in varus and flexion was lower in Group 2 compared to Group 1, being significant in flexion, p=0.91 and p=0.03, respectively. After 5000 cycles, interfragmentary movements in terms of varus, flexion, internal rotation, axial displacement and shear displacement at the fracture site were all lower in Group 2 compared to Group 1, with flexion and shear displacement being significant, p=0.14, p=0.04, p=0.25, p=0.11 and p=0.04, respectively. Cycles to failure until both interfragmentary 5° varus and 5° flexion were significantly higher in Group 2 compared to Group 1, p=0.04.

From a biomechanical perspective, the novel angle-stable intramedullary nail concept has the potential of achieving a higher initial axial and torsional relative stability and maintaining it with a better resistance towards loss of reduction under dynamic loading, while reducing the number of distal locking screws, compared to conventional locking in intramedullary nailed unstable distal tibia fractures.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 35 - 35
11 Apr 2023
Pastor T Knobe M Ciric D Zderic I van de Wall B Rompen I Visscher L Link B Babst R Richards G Gueorguiev B Beeres F
Full Access

Implant removal after clavicle plating is common. Low-profile dual mini-fragment plate constructs are considered safe for fixation of diaphyseal clavicle fractures. The aim of this study was to investigate: (1) the biomechanical competence of different dual plate designs from stiffness and cycles to failure, and (2) to compare them against 3.5mm single superoanterior plating.

Twelve artificial clavicles were assigned to 2 groups and instrumented with titanium matrix mandible plates as follows: group 1 (G1) (2.5mm anterior+2.0mm superior) and group 2 (G2) (2.0mm anterior+2.0mm superior). An unstable clavicle shaft fracture (AO/OTA15.2C) was simulated. Specimens were cyclically tested to failure under craniocaudal cantilever bending, superimposed with torsion around the shaft axis and compared to previous published data of 6 locked superoanterior plates tested under the same conditions (G3).

Displacement (mm) after 5000 cycles was highest in G3 (10.7±0.8) followed by G2 (8.5±1.0) and G1 (7.5±1.0), respectively. Both outcomes were significantly higher in G3 as compared to both G1 and G2 (p≤0.027). Cycles to failure were highest in G3 (19536±3586) followed by G1 (15834±3492) and G2 (11104±3177), being significantly higher in G3 compared to G2 (p=0.004). Failure was breakage of one or two plates at the level of the osteotomy in all specimens. One G1 specimen demonstrated failure of the anterior plate. Both plates in other G1 specimens. Majority of G2 had fractures in both plates. No screw pullout or additional clavicle fractures were observed among specimens.

Low-profile 2.0/2.0 dual plates demonstrated similar initial stiffness compared to 3.5mm single plates, however, had significantly lower failure endurance. Low-profile 2.5/2.0 dual plates showed significant higher initial stiffness and similar resistance to failure compared to 3.5mm single locked plates and can be considered as a useful alternative for diaphyseal clavicle fracture fixation. These results complement the promising results of several clinical studies.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 137 - 137
11 Apr 2023
Quinn A Pizzolato C Bindra R Lloyd D Saxby D
Full Access

There is currently no commercially available and clinically successful treatment for scapholunate interosseous ligament rupture, the latter leading to the development of hand-wrist osteoarthritis. We have created a novel biodegradable implant which fixed the dissociated scaphoid and lunate bones and encourages regeneration of the ruptured native ligament. To determine if scaphoid and lunate kinematics in cadaveric specimens were maintained during robotic manipulation, when comparing the native wrist with intact ligament and when the implant was installed.

Ten cadaveric experiments were performed with identical conditions, except for implant geometry that was personalised to the anatomy of each cadaveric specimen. Each cadaveric arm was mounted upright in a six degrees of freedom robot using k-wires drilled through the radius, ulna, and metacarpals. Infrared markers were attached to scaphoid, lunate, radius, and 3rd metacarpal. Cadaveric specimens were robotically manipulated through flexion-extension and ulnar-radial deviation by ±40° and ±30°, respectively.

The cadaveric scaphoid and lunate kinematics were examined with 1) intact native ligament, 2) severed ligament, 3) and installed implant.

Digital wrist models were generated from computed tomography scans and included implant geometry, orientation, and location. Motion data were filtered and aligned relative to neutral wrist in the digital models of each specimen using anatomical landmarks. Implant insertion points in the scaphoid and lunate over time were then calculated using digital models, marker data, and inverse kinematics. Root mean squared distance was compared between severed and implant configurations, relative to intact.

Preliminary data from five cadaveric specimens indicate that the implant reduced distance between scaphoid and lunate compared to severed configuration for all but three trials.

Preliminary results indicate our novel implant reduced scapho-lunate gap caused by ligament transection. Future analysis will reveal if the implant can achieve wrist kinematics similar to the native intact wrist.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 42 - 42
4 Apr 2023
Benca E van Knegsel K Zderic I Caspar J Strassl A Hirtler L Fuchssteiner C Gueorguiev B Windhager R Widhalm H Varga P
Full Access

Screw fixation is an established method for anterior cruciate ligament (ACL) reconstruction, although with a high rate of implant-related complications. An allograft system for implant fixation in ACL reconstruction, the Shark Screw ACL (surgebright GmbH) could overcome some of the shortcomings of bioabsorbable screws, such as foreign body reaction, need for implant removal and imaging artefacts. However, it needs to provide sufficient mechanical stability. Therefore, the aim of this study was to investigate the biomechanical stability, especially graft slippage, of the novel allograft system versus a conventional bioabsorbable interference screw (BioComposite Interference Screw; Arthrex Inc.) for tibial implant fixation in ACL reconstruction.

Twenty-four paired human proximal tibiae (3 female, 9 male, 72.7 ± 5.6 years) underwent ACL reconstruction. The quadrupled semitendinosus and gracilis tendon graft were fixed in one specimen of each pair using the allograft fixation system Shak Screw ACL and the contralateral one using an interference screw. All specimens were cyclically loaded at 1 Hz with peak load levels monotonically increased from 50 N at a rate of 0.1 N/cycle until catastrophic failure. Relative movements of the graft versus the tibia were captured with a stereographic optical motion tracking system (Aramis SRX; GOM GmbH).

The two fixation methods did not demonstrate any statistical difference in ultimate load at graft slippage (p = 0.24) or estimated survival at slippage (p = 0.06). Both, the ultimate load and estimated survival until failure were higher in the interference screw (p = 0.04, and p = 0.018, respectively). Graft displacement at ultimate load reached values of up to 7.2 mm (interference screw) and 11.3 mm (Shark Screw ACL).

The allograft screw for implant fixation in ACL reconstruction showed similar behavior in terms of graft slippage compared to the conventional metal interference screw but underperformed in terms of ultimate load. However, the ultimate load may not be considered a direct indicator of clinical failure.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 26 - 26
11 Apr 2023
Kowalski E Pelegrinelli A Ryan N Dervin G Lamontagne M
Full Access

This study examined pre-operative measures to predict post-operative biomechanical outcomes in total knee arthroplasty (TKA) patients.

Twenty-eight patients (female=12/male=16, age=63.6±6.9, BMI=29.9±7.4 kg/m2) with knee osteoarthritis scheduled to undergo TKA were included. All surgeries were performed by the same surgeon (GD) with a subvastus approach. Patients visited the gait lab within one-month prior to surgery and 12 months following surgery. At the gait lab, patients completed the knee injury and osteoarthritis outcome score (KOOS), a timed up and go (TUG), maximum knee flexion and extension strength evaluation, and a walking task. Variables of interest included the five KOOS sub-scores, TUG time, maximum knee flexion and extension strength normalized to body weight, walking speed, and peak knee biomechanics variables (flexion angle, abduction moment, power absorption). A Pearson's correlation was used to identify significantly correlated variables which were then inputted into a multiple regression.

No assumption violations for the multiple regression existed for any variables. Pre-operative knee flexion and extension strength, TUG time, and age were used in the multiple regression. The multiple regression model statistically significantly predicted peak knee abduction moment, post-operative walking speed, and post-operative knee flexion strength. All four variables added statistically significantly to the prediction p<.05.

Pre-operative KOOS values did not correlate with any biomechanical indicators of post-operative success. Age, pre-operative knee flexion and extension strength, and TUG times predicted peak knee abduction moment, which is associated with medial knee joint loading. These findings stress the importance of pre-surgery condition, as stronger individuals achieved better post-operative biomechanical outcomes. Additionally, younger patients had better outcomes, suggesting that surgeons should not delay surgery in younger patients. This delay in surgery may prevent patients from achieving optimal outcomes. Future studies should utilize a hierarchical multiple regression to identify which variables are most predictive.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 20 - 20
1 Nov 2021
Gueorguiev B
Full Access

Nonunions occur in situations with interrupted fracture healing process and indicate conditions where the fracture has no potential to heal without further intervention. Per definition, no healing is detected nine months post operation and there is no visible progress of healing over the last three months. The classification of nonunions as hypertrophic, oligotrophic, atrophic and pseudoarthosis, as well as aseptic or septic, identifies mechanical and biological requirements for fracture healing that have not been met. The overall treatment strategy comprises identification and elimination of the problems. However, current clinical methods to determine the state of healing are based on highly subjective radiographic evaluation or clinical examination.

A data collection telemetric system for objective continuous measurement of the load carried by a bridging smart implant was developed to assess the mechanical stability and monitor bone healing in complicated fracture situations. The first results from a clinical trial show that the system is capable to offer early warning of nonunions or poor fracture healing.

Nonunions are often multifactorial in nature and not just related to a biomechanical problem. Their successful treatment requires consideration of both biological and mechanical aspects. Disturbed vascularity and stability are the most important factors. Infection could be another complicating factor resulting in unpredictable long-time treatment. New technologies for monitoring of fracture healing in addition to radiographic evaluation and clinical examination seem to be promising for early detection of nonunions.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 67 - 67
1 Nov 2018
Bouaicha S Ernstbrunner L Jud L Meyer D Snedeker J Bachmann E
Full Access

Tear pattern and tendon involvement are risk factors for the development of a pseudoparalytic shoulder. However, some patients have similar tendon involvement but significantly different active forward flexion. In these cases, it remains unclear why some patients suffer from pseudoparalysis and others with the same tear pattern show good active range of motion. Moment arms (MA) and force vectors of the RC and the deltoid muscle play an important role in the muscular equilibrium to stabilize the glenohumeral joint. Biomechanical and clinical analyses were conducted calculating different MA-ratios of the RC and the deltoid muscle using computer rigid body simulation and a retrospective radiographic investigation of two cohorts with and without pseudoparalysis and massive RC tears. Idealized MAs were represented by two spheres concentric to the joints centre of rotation either spanning to the humeral head or deltoid origin of the acromion. Individual ratios of the RC /deltoid MAs on antero-posterior radiographs using the newly introduced Shoulder Abduction Moment (SAM) Index was compared between the pseudoparalytic and non-pseudoparalytic patients. Decrease of RC activity and improved glenohumeral stability (+14%) was found in simulations for MA ratios with larger diameters of the humeral head which also were consequently beneficial for the (remaining) RC. Clinical investigation of the MA-ratio showed significant risk of having pseudoparalysis in patients with massive tears and a SAM Index <0.77 (OR=11). The SAM index, representing individual biomechanical characteristics of shoulder morphology has an impact on the presence or absence of pseudoparalysis in shoulders with massive RC tears