Introduction. Within articular cartilage, chondrocytes reside within the pericellular matrix (PCM), collectively constituting the microanatomical entity known as a chondron. The PCM functions as a pivotal protective shield and mediator of biomechanical and biochemical cues. In the context of Osteoarthritis (OA), enzymatic degradation of the PCM is facilitated by matrix metalloproteinases (MMPs). This study delves into the functional implications of PCM structural integrity decline on the biomechanical properties of chondrons and impact on Ca. 2+. signaling dynamics. Method. Chondrons isolated from human cartilage explants were incubated with activated MMP-2, -3, or -7. Structural degradation of the pericellular matrix (PCM) was assessed by immunolabelling (collagen type VI and perlecan, n=5).
Introduction. Tendon ruptures are a common injury and often require surgical intervention to heal. A refixation is commonly performed with high-strength suture material. However, slipping of the thread is unavoidable even at 7 knots potentially leading to reduced compression of the sutured tendon at its footprint. This study aimed to evaluate the biomechanical properties and effectiveness of a novel dynamic high-strength suture, featuring self-tightening properties. Method. Distal biceps tendon rupture tenotomies and subsequent repairs were performed in sixteen paired human forearms using either conventional or the novel dynamic high-strength sutures in a paired design. Each tendon repair utilized an intramedullary biceps button for radial fixation.
Objectives. We investigated the effects on fracture healing of two up-regulators of inducible nitric oxide synthase (iNOS) in a rat model of an open femoral osteotomy: tadalafil, a phosphodiesterase inhibitor, and the recently reported nutraceutical, COMB-4 (consisting of L-citrulline, Paullinia cupana, ginger and muira puama), given orally for either 14 or 42 days. Materials and Methods. Unilateral femoral osteotomies were created in 58 male rats and fixed with an intramedullary compression nail. Rats were treated daily either with vehicle, tadalafil or COMB-4.
Recently, a new suture was designed to minimize laxity in order to preserve consistent tissue approximation while improving footprint compression after tendon repair. The aims of this study were: (1) to compare the biomechanical competence of two different high strength sutures in terms of slippage and failure load, (2) to investigate the influence of both knots number and different media (air, saline and fat) on the holding capacity of the knots. Alternating surgical knots of two different high-strength sutures (group1: FibreWire; group2: DynaCord; n = 105) were tied on two roller bearings with 50N tightening force.
Matrix metalloproteinase enzymes (MMPs) play a crucial role in the remodeling of articular cartilage, contributing also to osteoarthritis (OA) progression. The pericellular matrix (PCM) is a specialized space surrounding each chondrocyte, containing collagen type VI and perlecan. It acts as a transducer of biomechanical and biochemical signals for the chondrocyte. This study investigates the impact of MMP-2, -3, and -7 on the integrity and biomechanical characteristics of the PCM. Human articular cartilage explants (n=10 patients, ethical-nr.:674/2016BO2) were incubated with activated MMP-2, -3, or -7 as well as combinations of these enzymes. The structural degradative effect on the PCM was assessed by immunolabelling of the PCM's main components: collagen type VI and perlecan.
Several emerging reports suggest an important involvement of the hindfoot alignment in the outcome of knee osteotomy. At present, studies lack a comprehensive overview. Therefore, we aimed to systematically review all biomechanical and clinical studies investigating the role of the hindfoot alignment in the setting of osteotomies around the knee. A systematic literature search was conducted on multiple databases combining “knee osteotomy” and “hindfoot/ankle alignment” search terms. Articles were screened and included according to the PRISMA guidelines. A quality assessment was conducted using the Quality Appraisal for Cadaveric Studies (QUACS) - and modified methodologic index for non-randomized studies (MINORS) scales. Three cadaveric, fourteen retrospective cohort and two case-control studies were eligible for review.
Background. Devices are frequently used to gain sufficient purchase in a bone so that the bone itself can be manipulated to move or rupture soft tissue attachments. During hip surgery, several different extraction corkscrews are available to remove the femoral head, each with a different screw design with no evidence to suggest which is most effective. Additionally during the use of corkscrew devices, often due to the low bone density, stripping of the screw threads out of the femoral head can occur prior to its extraction, thus requiring reinsertion. The aims of this project were to measure the primary pullout and reinsertional forces of five commercially available corkscrews. Methods. Polyurethane bone models covering a spectrum of osteoporotic to normal bone densities (0.08 gcm-3, 0.16 gcm-3 and 0.32 gcm-3) were used in axial tensile testing at two insertional depths to assess the maximum pullout force of these corkscrews (Zimmer, Stryker, Medacta, JRI, Depuy). Results. There are significant differences between the pullout forces produced throughout the different densities at both insertional depths; 0.08gcm-3 (p=0.002), 0.16 gcm-3 (p<0.001) and 0.32 gcm-3 (p=0.006). Reinsertional testing on all corkscrews demonstrated a reduction in the pullout force of approximately 70%, underlining the benefit of effective insertion and extraction on the first attempt. Conclusion. In all bone model densities, consistent differences in the pullout forces generated were identified. All the corkscrews generate forces greater than the estimated forces required to yield the soft tissues attached to a femoral head. Further work into optimising the corkscrew design, especially regarding the torque generated during insertion, will improve the efficacy of future corkscrew use. This information should be transferable to other situations where manipulation of a bone is needed at the temporary or permanent expense of the attached soft tissues. Level of Evidence. V:
Femoral neck fractures account for half of all hip fractures and are recognized as a major public health problem associated with a high socioeconomic burden. Whilst internal fixation is preferred over arthroplasty for physiologically younger patients, no consensus exists about the optimal fixation device yet. The recently introduced implant Femoral Neck System (FNS) (DePuy Synthes, Zuchwil, Switzerland) was developed for dynamic fixation of femoral neck fractures and provides angular stability in combination with a minimally invasive surgical technique. Alternatively, the Hansson Pin System (HPS) (Swemac, Linköping, Sweden) exploits the advantages of internal buttressing. However, the obligate peripheral placement of the pins, adjacent to either the inferior or posterior cortex, renders the instrumentation more challenging. The aim of this study was to evaluate the biomechanical performance of FNS versus HPS in a Pauwels II femoral neck fracture model with simulated posterior comminution. Forty-degree Pauwels II femoral neck fractures AO 31-B2.1 with 15° posterior wedge were simulated in fourteen paired fresh-frozen human cadaveric femora, followed by instrumentation with either FNS or HPS in pair-matched fashion. Implant positioning was quantified by measuring the shortest distances between implant and inferior cortex (DI) as well as posterior cortex (DP) on anteroposterior and axial X-rays, respectively.
Background. Currently about 4–6% of all femur fractures consist of distal femoral fractures. Different methods and implants have been used for the surgical treatment of distal femoral fractures, including intramedullary nails. Retrograde nail. By contrast with antegrade nails, surgical approach or retrograde nailing exposes the knee joint which may lead to tendency of infection and increased knee pain. Present study aims to compare the biomechanical behaviour of distal angular condyler femoral intramedullary nail (DACFIN), retrograde nail and plate fixation. Methods. Fifteen 4th generation Saw bones were used to evaluate the biomechanical differences between the groups (Group 1: Plate fixation, Group 2: Retrograde nailing, Group 3: DACFIN; (n=5)).
Background.
This research aims to enhance the control of intricate musculoskeletal spine models, a critical tool for comprehending both healthy and pathological spinal conditions. State-of-the-art musculoskeletal spine models incorporate segments for all vertebra, each possessing 3 degrees-of-freedom (DOF). Manually defining the posture with this amount of DOFs presents a significant challenge. The prevalent method of equally distributing the spine's overall rotation among the vertebrae often proves to be an inadequate assumption, particularly when dealing with the entire spine. We have engineered a comprehensive non-linear spine rhythm and the requisite tools for its implementation in widely utilized musculoskeletal modelling software (1). The rhythm controls lateral bending, axial rotation, and flexion/extension. The mathematical and implementation details of the rhythm are beyond this abstract, but it's noteworthy that the implementation accommodates non-linear rhythms. This means, for example, that one set of rhythm coefficients is used for flexion and another for extension. The rhythm coefficients, which distinguish the movement along the spine, were derived from a review of spine literature. The values for spine and vertebra range-of-motion (ROM) vary significantly in published studies, and no complete dataset was found in any single study. Consequently, the rhythm presented here is a composite, designed to provide the most consistent and average set of rhythm coefficients.Introduction
Method
Glenohumeral joint injuries frequently result in shoulder instability. However, the biomechanical effect of cartilage loss on shoulder stability remains unknown. The aim of the current study was to investigate biomechanically the effect of two severity stages of cartilage loss in different dislocation directions on shoulder stability.
Joint dislocation was provoked for 11 human cadaveric glenoids in seven different dislocation directions between 3 o'clock (anterior) to 9 o'clock (posterior) dislocation. Shoulder stability ratio (SSR) and concavity gradient were assessed in intact condition, and after 3 mm and 6 mm simulated cartilage loss. The influence of cartilage loss on SSR and concavity gradient was statistically evaluated.
Between intact state and 6 mm cartilage loss, both SSR and concavity gradient decreased significantly in every dislocation direction (p≤0.038), except the concavity gradient in 4 o'clock dislocation direction (p=0.088). Thereby, anterior-inferior dislocation directions were associated with the highest loss of SSR and concavity gradient of up to 59.0% and 49.4%, respectively, being significantly higher for SSR compared to all other dislocation directions (p≤0.04). The correlations between concavity gradient and SSR for pooled dislocation directions were significant for all three conditions of cartilage loss (p<0.001).
From a biomechanical perspective, articular cartilage of the glenoid contributes significantly to the concavity gradient, correlating strongly with the associated loss in glenohumeral joint stability. The highest effect of cartilage loss was observed in anterior-inferior dislocation directions, suggesting that surgical intervention should be considered for recurrent shoulder dislocations in the presence of cartilage loss.
Odontoid fracture of the second cervical vertebra (C2) is the most common spinal fracture type in elderly patients. However, very little is known about the biomechanical fracture mechanisms, but could play a role in fracture prevention and treatment. This study aimed to investigate the biomechanical competence and fracture characteristics of the odontoid process. A total of 42 human C2 specimens (14 female and 28 male, 71.5 ± 6.5 years) were scanned via quantitative computed tomography, divided in 6 groups (n = 7) and subjected to combined quasi-static loading at a rate of 0.1 mm/s until fracturing at inclinations of −15°, 0° and 15° in sagittal plane, and −50° and 0° in transverse plane. Bone mineral density (BMD), specimen height, fusion state of the ossification centers, stiffness, yield load, ultimate load, and fracture type according to Anderson and d'Alonzo were assessed. While the lowest values for stiffness, yield, and ultimate load were observed at load inclination of 15° in sagittal plane, no statistically significant differences could be observed among the six groups (p = 0.235, p = 0.646, and p = 0.505, respectively). Evaluating specimens with only clearly distinguishable fusion of the ossification centers (n = 26) reveled even less differences among the groups for all mechanical parameters. BMD was positively correlated with yield load (R² = 0.350, p < 0.001), and ultimate load (R² = 0.955, p < 0.001), but not with stiffness (p = 0.070). Type III was the most common fracture type (23.5%). These biomechanical outcomes indicate that load direction plays a subordinate role in traumatic fractures of the odontoid process in contrast to BMD which is a strong determinant of stiffness and strength. Thus, odontoid fractures appear to result from an interaction between load magnitude and bone quality.
Tibiocalcaneal arthrodesis with a retrograde intramedullary nail is an established procedure considered as a salvage in case of severe arthritis and deformity of the ankle and subtalar joints [1]. Recently, a significant development in hindfoot arthrodesis with plates has been indicated. Therefore, the aim of this study was to compare a plate specifically developed for arthrodesis of the hindfoot with an already established nail system [2] Sixteen paired human cadaveric lower legs with removed forefoot and cut at mid-tibia were assigned to two groups for tibiocalcaneal arthrodesis using either a hindfoot arthrodesis nail or an arthrodesis plate. The specimens were tested under progressively increasing cyclic loading in dorsiflexion and plantar flexion to failure, with monitoring via motion tracking. Initial stiffness was calculated together with range of motion in dorsiflexion and plantar flexion after 200, 400, 600, 800, and 1000 cycles. Cycles to failure were evaluated based on 5° dorsiflexion failure criterionIntroduction
Method
This study investigates the relationships between Intervertebral Disc (IVD) morphology and biomechanics using patient-specific (PS) finite element (FE) models and poromechanical simulations. 169 3D lumbar IVD shapes from the European project MySpine (FP7-269909), spanning healthy to Pfirrmann grade 4 degeneration, were obtained from MRIs. A Bayesian Coherent Point Drift algorithm aligned meshes to a previously validated structural FE mesh of the IVD. After mesh quality analyses and Hausdorff distance measurements, mechanical simulations were performed: 8 and 16 hours of sleep and daytime, respectively, applying 0.11 and 0.54 MPa of pressure on the upper cartilage endplate (CEP). Simulation results were extracted from the anterior (ATZ) and posterior regions (PTZ) and the center of the nucleus pulposus (CNP). Data mining was performed using Linear Regression, Support Vector Machine, and eXtreme Gradient Boosting techniques. Mechanical variables of interest in DD, such as pore fluid velocity (FLVEL), water content, and swelling pressure, were examined. The morphological variables of the simulated discs were used as input features. Local morphological variables significantly impacted the local mechanical response. The local disc heights, respectively in the mid (mh), anterior (ah), and posterior (ph) regions, were key factors in general. Additionally, fluid transport, reflected by FLVEL, was greatly influenced (r2 0.69) by the shape of the upper and lower cartilage endplates (CEPs). This study suggests that disc morphology affects Mechanical variables of interest in DD. Attention should be paid to the antero-posterior distribution and local effects of disc heights. Surprisingly, the CEP morphology remotely affected the fluid transport in NP volumes around mid-height, and mechanobiological implications shall be explored. In conclusion, patient-specific IVD modeling has strong potential to unravel important correlations between IVD phenotypes and local tissue regulation.
The objective of this study was to analyze the biomechanical effect of an implanted ACL graft by determining the tunnel position according to the aspect ratio (ASR) of the distal femur during flexion-extension motion. To analyze biomechanical characteristics according to the ASR of the knee joint, only male samples were selected to exclude the effects of gender and 89 samples were selected for measurement. The mean age was 50.73 years, and the mean height was 165.22 cm. We analyzed tunnel length, graft bending angle, and stress of the graft according to tunnel entry position and aspect ratio (ratio of antero-posterior depth to medio-lateral width) of the articular surface for the distal femur during single-bundle outside-in anterior cruciate ligament reconstruction surgery. We performed multi-flexible-body dynamic analyses with wherein four ASR (98, 105, 111, and 117%) knee models. The various ASRs were associated with approximately 1-mm changes in tunnel length. The graft bending angle increased when the entry point was far from the lateral epicondyle and was larger when the ASR was smaller. The graft was at maximum stress, 117% ASR, when the tunnel entry point was near the lateral epicondyle. The maximum stress value at a 5-mm distance from the lateral epicondyle was 3.5 times higher than the 15-mm entry position and, the cases set to 111% and 105% ASR, showed 1.9 times higher stress values when at a 5-mm distance compared with a 15-mm distance. In the case set at 98% ASR, the low-stress value showed a without-distance difference from the lateral epicondyle. Our results suggest that there is no relationship between the ASR and femoral tunnel length, A smaller ASR causes a higher graft bending angle, and a larger ASR causes greater stress in the graft.
This study aims to investigate the mechanical properties of a rotator cuff tear repaired with a polypropylene interposition graft in an ovine infraspinatus ex-vivo model. Twenty fresh shoulders from skeletally mature sheep were used in this study. A tear size of 20 mm from the tendon joint was created in the infraspinatus tendon to simulate a large tear in fifteen specimens. This was repaired with a polypropylene mesh used as an interposition graft between the ends of the tendon. Eight specimens were secured with mattress stitches while seven were secured to the remnant tendon on the greater tuberosity side by continuous stitching. Remaining five specimens with an intact tendon served as a control group. All specimens underwent cyclic loading with a universal testing machine to determine the ultimate failure load and gap distance. Gap distance increased with progressive cyclic loading through 3000 cycles for all repaired specimens. Mean gap distance after 3000 cycles for both continuous and mattress groups are 1.7 mm and 4.2 mm respectively (P = .001). Significantly higher mean ultimate failure load was also observed with 549.2 N in the continuous group, 426.6 N in the mattress group and 370 N in the intact group. The use of a polypropylene mesh as an interposition graft for large irreparable rotator cuff tears is biomechanically suitable and results in a robust repair that is comparable to an intact rotator cuff tendon. When paired with a continuous suturing technique, it demonstrates significantly resultant superior biomechanical properties that may potentially reduce re-tear rates after repairing large or massive rotator cuff tears.
Implant removal after clavicle plating is common. Low-profile dual mini-fragment plate constructs are considered safe for fixation of diaphyseal clavicle fractures. The aim of this study was to investigate: (1) the biomechanical competence of different dual plate designs from stiffness and cycles to failure, and (2) to compare them against 3.5mm single superoanterior plating. Twelve artificial clavicles were assigned to 2 groups and instrumented with titanium matrix mandible plates as follows: group 1 (G1) (2.5mm anterior+2.0mm superior) and group 2 (G2) (2.0mm anterior+2.0mm superior). An unstable clavicle shaft fracture (AO/OTA15.2C) was simulated. Specimens were cyclically tested to failure under craniocaudal cantilever bending, superimposed with torsion around the shaft axis and compared to previous published data of 6 locked superoanterior plates tested under the same conditions (G3). Displacement (mm) after 5000 cycles was highest in G3 (10.7±0.8) followed by G2 (8.5±1.0) and G1 (7.5±1.0), respectively. Both outcomes were significantly higher in G3 as compared to both G1 and G2 (p≤0.027). Cycles to failure were highest in G3 (19536±3586) followed by G1 (15834±3492) and G2 (11104±3177), being significantly higher in G3 compared to G2 (p=0.004). Failure was breakage of one or two plates at the level of the osteotomy in all specimens. One G1 specimen demonstrated failure of the anterior plate. Both plates in other G1 specimens. Majority of G2 had fractures in both plates. No screw pullout or additional clavicle fractures were observed among specimens. Low-profile 2.0/2.0 dual plates demonstrated similar initial stiffness compared to 3.5mm single plates, however, had significantly lower failure endurance. Low-profile 2.5/2.0 dual plates showed significant higher initial stiffness and similar resistance to failure compared to 3.5mm single locked plates and can be considered as a useful alternative for diaphyseal clavicle fracture fixation. These results complement the promising results of several clinical studies.
There is currently no commercially available and clinically successful treatment for scapholunate interosseous ligament rupture, the latter leading to the development of hand-wrist osteoarthritis. We have created a novel biodegradable implant which fixed the dissociated scaphoid and lunate bones and encourages regeneration of the ruptured native ligament. To determine if scaphoid and lunate kinematics in cadaveric specimens were maintained during robotic manipulation, when comparing the native wrist with intact ligament and when the implant was installed. Ten cadaveric experiments were performed with identical conditions, except for implant geometry that was personalised to the anatomy of each cadaveric specimen. Each cadaveric arm was mounted upright in a six degrees of freedom robot using k-wires drilled through the radius, ulna, and metacarpals. Infrared markers were attached to scaphoid, lunate, radius, and 3rd metacarpal. Cadaveric specimens were robotically manipulated through flexion-extension and ulnar-radial deviation by ±40° and ±30°, respectively. The cadaveric scaphoid and lunate kinematics were examined with 1) intact native ligament, 2) severed ligament, 3) and installed implant. Digital wrist models were generated from computed tomography scans and included implant geometry, orientation, and location. Motion data were filtered and aligned relative to neutral wrist in the digital models of each specimen using anatomical landmarks. Implant insertion points in the scaphoid and lunate over time were then calculated using digital models, marker data, and inverse kinematics. Root mean squared distance was compared between severed and implant configurations, relative to intact. Preliminary data from five cadaveric specimens indicate that the implant reduced distance between scaphoid and lunate compared to severed configuration for all but three trials. Preliminary results indicate our novel implant reduced scapho-lunate gap caused by ligament transection. Future analysis will reveal if the implant can achieve wrist kinematics similar to the native intact wrist.
Screw fixation is an established method for anterior cruciate ligament (ACL) reconstruction, although with a high rate of implant-related complications. An allograft system for implant fixation in ACL reconstruction, the Shark Screw ACL (surgebright GmbH) could overcome some of the shortcomings of bioabsorbable screws, such as foreign body reaction, need for implant removal and imaging artefacts. However, it needs to provide sufficient mechanical stability. Therefore, the aim of this study was to investigate the biomechanical stability, especially graft slippage, of the novel allograft system versus a conventional bioabsorbable interference screw (BioComposite Interference Screw; Arthrex Inc.) for tibial implant fixation in ACL reconstruction. Twenty-four paired human proximal tibiae (3 female, 9 male, 72.7 ± 5.6 years) underwent ACL reconstruction. The quadrupled semitendinosus and gracilis tendon graft were fixed in one specimen of each pair using the allograft fixation system Shak Screw ACL and the contralateral one using an interference screw. All specimens were cyclically loaded at 1 Hz with peak load levels monotonically increased from 50 N at a rate of 0.1 N/cycle until catastrophic failure. Relative movements of the graft versus the tibia were captured with a stereographic optical motion tracking system (Aramis SRX; GOM GmbH). The two fixation methods did not demonstrate any statistical difference in ultimate load at graft slippage (p = 0.24) or estimated survival at slippage (p = 0.06). Both, the ultimate load and estimated survival until failure were higher in the interference screw (p = 0.04, and p = 0.018, respectively). Graft displacement at ultimate load reached values of up to 7.2 mm (interference screw) and 11.3 mm (Shark Screw ACL). The allograft screw for implant fixation in ACL reconstruction showed similar behavior in terms of graft slippage compared to the conventional metal interference screw but underperformed in terms of ultimate load. However, the ultimate load may not be considered a direct indicator of clinical failure.