Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 118 - 118
2 Jan 2024
Stroobant L Verstraete M Onsem S Victor J Chevalier A
Full Access

Numerous papers present in-vivo knee kinematics data following total knee arthroplasty (TKA) from fluoroscopic testing. Comparing data is challenging given the large number of factors that potentially affect the reported kinematics. This paper aims at understanding the effect of following three different factors: implant geometry, performed activity and analysis method.

A total of 30 patients who underwent TKA were included in this study. This group was subdivided in three equal groups: each group receiving a different type of posterior stabilized total knee prosthesis. During single-plane fluoroscopic analysis, each patient performed three activities: open chain flexion extension, closed chain squatting and chair-rising. The 2D fluoroscopic data were subsequently converted to 3D implant positions and used to evaluate the tibiofemoral contact points and landmark-based kinematic parameters.

Significantly different anteroposterior translations and internal-external rotations were observed between the considered implants. In the lateral compartment, these differences only appeared after post-cam engagement. Comparing the activities, a significant more posterior position was observed for both the medial and lateral compartment in the closed chain activities during mid-flexion. A strong and significant correlation was found between the contact-points and landmarks-based analyses method. However, large individual variations were also observed, yielding a difference of up to 25% in anteroposterior position between both methods.

In conclusion, all three evaluated factors significantly affect the obtained tibiofemoral kinematics. The individual implant design significantly affects the anteroposterior tibiofemoral position, internal-external rotation and timing of post-cam engagement. Both kinematics and post-cam engagement additionally depend on the activity investigated, with a more posterior position and associated higher patella lever arm for the closed chain activities. Attention should also be paid to the considered analysis method and associated kinematics definition: analyzing the tibiofemoral contact points potentially yields significantly different results compared to a landmark-based approach.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 50 - 50
2 Jan 2024
Van Oevelen A Duquesne K Peiffer M Victor J Audenaert E
Full Access

Intra-articular cartilage pressure distribution in the knee joint is critical in the understanding of osteoarthritis. Combining personalized statistical modeling of the morphological characteristics with discrete element modeling enables patient-specific predictions of the pressure on the tibial plateau. However, modeling of the meniscus during gait is complicated by the dynamic nature of the structure. Nevertheless, the position of the meniscus has a substantial impact on intra-articular stress distribution. Therefore, the focus of this presentation will be on how modeling of meniscal movement during knee flexion improves insight in general meniscal kinematics for the use in tibiofemoral stress distribution calculations.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 55 - 55
2 Jan 2024
Stroobant L Jacobs E Arnout N Van Onsem S Burssens A Victor J
Full Access

7–20 % of the patients with a total knee arthroplasty (TKA) are dissatisfied without an indication for revision. Therapeutic options for this patient population with mostly a lack of quadriceps strength are limited. The purpose of this study is to evaluate the effect of six weeks low load resistance training with blood flow restriction (BFR) on the clinical outcome in these unhappy TKA patients.

Thirty-one unhappy TKA patients (of the scheduled fifty patients) without mechanical failure were included in this prospective study since 2022. The patients participate in a supervised resistance training combined with BFR, two times a week during nine weeks. Patients were evaluated by the Knee Osteoarthritis Outcome Score (KOOS), Knee Society Score: satisfaction (KSSs) and the Pain Catastrophizing Scale (PCS). Functionality was tested using the Six Minute Walk Test (6MWT) and the 30-Second Chair Stand Test (30CST). Follow-up took place at six weeks, three months and six months after the start.

Six weeks training with BFR provided statistically significant improvements in all the KOOS subscales compared to the baseline, especially for symptoms (55.1 (±15.4) versus 48.0 (±16.5); p<0.001), activities in daily living (50.3 (±21.1) versus 43.7 (±17.2); p<0.00) and quality of life (24.6 (±18.5) versus 17.3 (±13.0); p<0.001). The PCS reduced from 27.4 (±11.0) to 23.2 (±11.4) at six weeks (p<0.01), whereas the KSSs increased from 11.8 (±6.5) to 14.9 (±7.6) (p=0.021). Both the 6MWT and the 30CST improved statistically significant from respectively 319.7 (±15.0) to 341.6m (±106.5) (p<0.01) and 8.6 (±3.9) to 9.3 times (±4.5) (p<0.01).

Blood flow restriction appears to enhance the quality of life and functional performance of unhappy TKA patients. Based on these preliminary results, BFR seems to be a promising and valuable alternative for these TKA patients with limited therapeutic options.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_9 | Pages 12 - 12
17 Apr 2023
Van Oevelen A Burssens A Krähenbühl N Barg A Audenaert E Hintermann B Victor J
Full Access

Several emerging reports suggest an important involvement of the hindfoot alignment in the outcome of knee osteotomy. At present, studies lack a comprehensive overview. Therefore, we aimed to systematically review all biomechanical and clinical studies investigating the role of the hindfoot alignment in the setting of osteotomies around the knee.

A systematic literature search was conducted on multiple databases combining “knee osteotomy” and “hindfoot/ankle alignment” search terms. Articles were screened and included according to the PRISMA guidelines. A quality assessment was conducted using the Quality Appraisal for Cadaveric Studies (QUACS) - and modified methodologic index for non-randomized studies (MINORS) scales.

Three cadaveric, fourteen retrospective cohort and two case-control studies were eligible for review. Biomechanical hindfoot characteristics were positively affected (n=4), except in rigid subtalar joint (n=1) or talar tilt (n=1) deformity. Patient symptoms and/or radiographic alignment at the level of the hindfoot did also improve after knee osteotomy (n=13), except in case of a small pre-operative lateral distal tibia- and hip knee ankle (HKA) angulation or in case of a large HKA correction (>14.5°). Additionally, a pre-existent hindfoot deformity (>15.9°) was associated with undercorrection of lower limb alignment following knee osteotomy. The mean QUACS score was 61.3% (range: 46–69%) and mean MINORS score was 9.2 out of 16 (range 6–12) for non-comparative and 16.5 out of 24 (range 15–18) for comparative studies.

Osteotomies performed to correct knee deformity have also an impact on biomechanical and clinical outcomes of the hindfoot. In general, these are reported to be beneficial, but several parameters were identified that are associated with newly onset – or deterioration of hindfoot symptoms following knee osteotomy. Further prospective studies are warranted to assess how diagnostic and therapeutic algorithms based on the identified criteria could be implemented to optimize the overall outcome of knee osteotomy.

Remark: Aline Van Oevelen and Arne Burssens contributed equally to this work


Bone & Joint Research
Vol. 12, Issue 4 | Pages 285 - 293
17 Apr 2023
Chevalier A Vermue H Pringels L Herregodts S Duquesne K Victor J Loccufier M

Aims

The goal was to evaluate tibiofemoral knee joint kinematics during stair descent, by simulating the full stair descent motion in vitro. The knee joint kinematics were evaluated for two types of knee implants: bi-cruciate retaining and bi-cruciate stabilized. It was hypothesized that the bi-cruciate retaining implant better approximates native kinematics.

Methods

The in vitro study included 20 specimens which were tested during a full stair descent with physiological muscle forces in a dynamic knee rig. Laxity envelopes were measured by applying external loading conditions in varus/valgus and internal/external direction.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 101 - 101
1 Feb 2023
MacDessi SJ Oussedik S Abdel MP Victor J Pagnano MW Haddad FS


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 102 - 108
1 Feb 2023
MacDessi SJ Oussedik S Abdel MP Victor J Pagnano MW Haddad FS

Orthopaedic surgeons are currently faced with an overwhelming number of choices surrounding total knee arthroplasty (TKA), not only with the latest technologies and prostheses, but also fundamental decisions on alignment philosophies. From ‘mechanical’ to ‘adjusted mechanical’ to ‘restricted kinematic’ to ‘unrestricted kinematic’ — and how constitutional alignment relates to these — there is potential for ambiguity when thinking about and discussing such concepts. This annotation summarizes the various alignment strategies currently employed in TKA. It provides a clear framework and consistent language that will assist surgeons to compare confidently and contrast the concepts, while also discussing the latest opinions about alignment in TKA. Finally, it provides suggestions for applying consistent nomenclature to future research, especially as we explore the implications of 3D alignment patterns on patient outcomes.

Cite this article: Bone Joint J 2023;105-B(2):102–108.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_9 | Pages 7 - 7
1 Jun 2021
Herregodts S Verhaeghe M Gijsels S Herregodts J De Baets P Victor J
Full Access

Introduction

Robot systems have been successfully introduced to improve the accuracy and reduce severe iatrogenic soft tissue damage in knee arthroplasty. Unfortunately to perform complete a complete bone cut, the cutting tool has to slightly pass the edge of the bone. In the posterior zones were retractor protection is impossible this will lead to contact between the cutting tool and the soft tissue envelope. Therefore, complete soft tissue preservation cannot be guaranteed with the current commercial systems.

Methods

This study presents an alternative robotic controlled cutting technique to perform the bone resections during TKA by milling a slot with a long slender high-speed milling tool. The system is composed by a long milling tool driven by a high-speed motor and a protector covering the end of the cutter. The protector is rigidly connected to the motor by the support structure next to the mill, which moves behind the mill in the slot created by the cutter. The protector at the end of the cutter has four functions: providing mechanical support for the mill, preventing soft tissue to come into contact with the cutter, sensing the edge of the bone to accurately follow the shape of the bone and releasing the attached soft tissue. The edge of the bone is sensed by force feedback and with the help of a probing motion the adaptive algorithm enables the protector to follow the edge of the bone closely by compensating for small segmentation and registration errors. A pilot test to evaluate the concept was performed on three fresh frozen knees. The flatness of the resection, the iatrogenic soft tissue damage, the cutting time and the efficiency of the bone contour following algorithm was measured.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 87 - 97
1 Jan 2021
Burssens A De Roos D Barg A Welck MJ Krähenbühl N Saltzman CL Victor J

Aims

Patients with a deformity of the hindfoot present a particular challenge when performing total knee arthroplasty (TKA). The literature contains little information about the relationship between TKA and hindfoot alignment. This systematic review aimed to determine from both clinical and radiological studies whether TKA would alter a preoperative hindfoot deformity and whether the outcome of TKA is affected by the presence of a postoperative hindfoot deformity.

Methods

A systematic literature search was performed in the databases PubMed, EMBASE, Cochrane Library, and Web of Science. Search terms consisted of “total knee arthroplasty/replacement” combined with “hindfoot/ankle alignment”. Inclusion criteria were all English language studies analyzing the association between TKA and the alignment of the hindfoot, including the clinical or radiological outcomes. Exclusion criteria consisted of TKA performed with a concomitant extra-articular osteotomy and case reports or expert opinions. An assessment of quality was conducted using the modified Methodological Index for Non-Randomized Studies (MINORS). The review was performed according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines and registered in the PROSPERO database (CRD42019106980).


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1324 - 1330
3 Oct 2020
Herregodts S Verhaeghe M Paridaens R Herregodts J Vermue H Arnout N De Baets P Victor J

Aims

Inadvertent soft tissue damage caused by the oscillating saw during total knee arthroplasty (TKA) occurs when the sawblade passes beyond the bony boundaries into the soft tissue. The primary objective of this study is to assess the risk of inadvertent soft tissue damage during jig-based TKA by evaluating the excursion of the oscillating saw past the bony boundaries. The second objective is the investigation of the relation between this excursion and the surgeon’s experience level.

Methods

A conventional jig-based TKA procedure with medial parapatellar approach was performed on 12 cadaveric knees by three experienced surgeons and three residents. During the proximal tibial resection, the motion of the oscillating saw with respect to the tibia was recorded. The distance of the outer point of this cutting portion to the edge of the bone was defined as the excursion of the oscillating saw. The excursion of the sawblade was evaluated in six zones containing the following structures: medial collateral ligament (MCL), posteromedial corner (PMC), iliotibial band (ITB), lateral collateral ligament (LCL), popliteus tendon (PopT), and neurovascular bundle (NVB).


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 407 - 413
1 Apr 2020
Vermue H Lambrechts J Tampere T Arnout N Auvinet E Victor J

The application of robotics in the operating theatre for knee arthroplasty remains controversial. As with all new technology, the introduction of new systems might be associated with a learning curve. However, guidelines on how to assess the introduction of robotics in the operating theatre are lacking. This systematic review aims to evaluate the current evidence on the learning curve of robot-assisted knee arthroplasty. An extensive literature search of PubMed, Medline, Embase, Web of Science, and Cochrane Library was conducted. Randomized controlled trials, comparative studies, and cohort studies were included. Outcomes assessed included: time required for surgery, stress levels of the surgical team, complications in regard to surgical experience level or time needed for surgery, size prediction of preoperative templating, and alignment according to the number of knee arthroplasties performed. A total of 11 studies met the inclusion criteria. Most were of medium to low quality. The operating time of robot-assisted total knee arthroplasty (TKA) and unicompartmental knee arthroplasty (UKA) is associated with a learning curve of between six to 20 cases and six to 36 cases respectively. Surgical team stress levels show a learning curve of seven cases in TKA and six cases for UKA. Experience with the robotic systems did not influence implant positioning, preoperative planning, and postoperative complications. Robot-assisted TKA and UKA is associated with a learning curve regarding operating time and surgical team stress levels. Future evaluation of robotics in the operating theatre should include detailed measurement of the various aspects of the total operating time, including total robotic time and time needed for preoperative planning. The prior experience of the surgical team should also be evaluated and reported.

Cite this article: Bone Joint J 2020;102-B(4):407–413.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 276 - 279
1 Mar 2020
Oussedik S Abdel MP Victor J Pagnano MW Haddad FS

Dissatisfaction following total knee arthroplasty is a well-documented phenomenon. Although many factors have been implicated, including modifiable and nonmodifiable patient factors, emphasis over the past decade has been on implant alignment and stability as both a cause of, and a solution to, this problem. Several alignment targets have evolved with a proliferation of techniques following the introduction of computer and robotic-assisted surgery. Mechanical alignment targets may achieve mechanically-sound alignment while ignoring the soft tissue envelope; kinematic alignment respects the soft tissue envelope while ignoring the mechanical environment. Functional alignment is proposed as a hybrid technique to allow mechanically-sound, soft tissue-friendly alignment targets to be identified and achieved.

Cite this article: Bone Joint J 2020;102-B(3):276–279.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 53 - 53
1 Apr 2019
Van Onsem S Verstraete M Verrewaere D Van Der Straeten C Victor J
Full Access

Background

Under- or oversizing of either component of a total knee implant can lead to early component loosening, instability, soft tissue irritation or overstuffing of joint gaps. All of these complications may cause postoperative persistent pain or stiffness. While survival of primary TKA's is excellent, recent studies show that patient satisfaction is worse. Up to 20% of the patients are not satisfied with the outcome as and residual pain is still a frequent occurrence.

The goal of this study was therefore to evaluate if the sizing of the femoral component, as measured on a 3D-reconstructed projection, is related to patient reported outcome measures.

From our prospectively collected TKA outcome database, all patients with a preoperative CT and a postoperative X-ray of their operated knee were included in this study. Of these 43 patients, 26 (60,5%) were women and 17 (39,5%) were men. The mean age (+/−SD) was 74,6 +/− 9 years.

Methods

CT scans were acquired. All patients underwent TKA surgery in a single institution by one surgical team using the same bi- cruciate substituting total knee (Journey II BCS, Smith&Nephew, Memphis, USA). Using a recently released X-ray module in Mimics (Materialise NV, Leuven, Belgium), this module allows to align the post-operative bi-planar x-rays with the 3D- reconstructed pre-operative distal femur and to determine the 3D position of the bone and implant models using the CAD- file of the implant. This new technique was validated at our department and was found to have a sub-degree, sub-millimeter accuracy. Eleven zones of interest were defined. On the medial and the lateral condyle, the extension, mid-flexion and deep flexion facet were determined. Corresponding trochlear zones were defined and two zones were defined to evaluate the mediolateral width. In order to compare different sizes, elastic deforming mesh matching algorithms were implemented to transfer the selected surfaces from one implant to another. The orthogonal distances from the implant to the nearest bone were calculated. Positive values represent a protruding (oversized) femoral component, negative values an undersized femoral component. The figure shows the marked zones on the femoral implant. The KOOS subscores and KSS Satisfaction subscore were evaluated.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 55 - 55
1 Apr 2019
Van Onsem S Verstraete M Van Der Straeten C Victor J
Full Access

Background

Kinematic patterns in total knee arthroplasty (TKA) can vary considerably from the native knee. No study has shown a relation between a given kinematic pattern and patient satisfaction yet.

Questions

The purpose of this study was to test whether the kinematical pattern, and more specifically the anteroposterior translation during (1) open kinetic chain flexion-extension, (2) closed kinetic chain chair rising and (3) squatting, is related to the level of patient satisfaction after TKA.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 14 - 14
1 Nov 2018
Demey P Vluggen E Burssens A Leenders T Buedts K Victor J
Full Access

Hindfoot disorders are complex 3D deformities. Current literature has assessed their influence on the full leg alignment, but the superposition of the hindfoot on plain radiographs resulted in different measurement errors. Therefore, the aim of this study is to assess the hindfoot alignment on Weight-Bearing CT (WBCT) and its influence on the radiographic Hip-Knee-Ankle (HKA) angle. A retrospective analysis was performed on a study population of 109 patients (mean age of 53 years ± 14,49) with a varus or valgus hindfoot deformity. The hindfoot angle (HA) was measured on the WBCT while the HKA angle, and the anatomical tibia axis angle towards the vertical (TAX) were analysed on the Full Leg radiographs. The mean HA in the valgus hindfoot group was 9,19°±7.94, in the varus hindfoot group −7,29°±6.09. The mean TAX was 3,32°±2.17 in the group with a valgus hindfoot and 1,89°±2.63 in the group with a varus hindfoot, which showed to be statistically different (p<0.05). The mean HKA Angle was −1,35°±2.73 in the valgus hindfoot group and 0,4°±2.89 in the varus hindfoot group, which showed to be statistically different (p<0.05). This study demonstrates a higher varus in both the HKA and TAX in valgus hindfoot and a higher tibia valgus in varus hindfoot. This contradicts the previous assumption that a varus hindfoot is associated with a varus knee or vice versa. In clinical practice, these findings contribute to a better understanding of deformity corrections of both the hindfoot and the knee.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 15 - 15
1 Nov 2018
Van Oevelen A van Ovost E E De Mits S Bodere I Leenders T Clockaerts S Victor J Burssens A
Full Access

An adult acquired flatfoot deformity (AAFD) is a complex 3D deformity. Surgical correction consists of a medial calcaneal osteotomy (MCO) but shows limitations due to the current 2D assessment. Therefore, the aim is to determine the influence of an MCO on the longitudinal foot arch assessed by 2D and 3D weightbearing CT (WBCT). Seventeen patients with a mean age of 44,5 years (range 18–66 yrs) were retrospectively included. MCO was indicated in a stage II AAFD (N=15) and a post-traumatic valgus deformity (N=2). Pre- and post-operative imaging was obtained from a WBCT. The height of the longitudinal foot arch was measured as the distance from the navicular tuberositas to the floor (Navicular Height, NH) on 2D CT images (NH2D) and computed on 3D CT data (NH3D). Additionally, 3D assessment could compute the degree of exorotation (α) of the navicular bone towards the vertical axis. The mean pre-operative NH2D and NH3D were respectively 29.57mm ± 7.59 and 28.34mm ± 6.51. These showed to be statistically different from the mean post-operative NH2D and NH3D, respectively 31.62mm ± 6.69 and 31.67mm ± 6.47 (p < 0,001). A statistical difference was also found when comparing the mean degree of exorotation in pre- and post-operative, respectively: αpre=14.08° ± 4,92 and the αpost=19,88° ± 3.50 (p < 0,001). This study demonstrates a significant correction of the longitudinal foot arch after a MCO. The novelty is attributed to the accurate degree of rotation assessment using WBCT. This information could be assistive to optimise a pre-operative planning.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 4 - 4
1 Nov 2018
De Roos D Van den Bossche T Burssens A Victor J
Full Access

Patients with a hindfoot deformity impose a particular challenge when performing a total knee arthroplasty (TKA). This could be attributed to the lack of insights concerning the outcome towards the hindfoot alignment. Our objective was to perform a systematic review of the literature to investigate the influence of TKA on hindfoot alignment and vice-versa. In accordance with the Methodological Index For Non-Randomized Studies (MINORS) statement standards, we performed a systematic review. Electronic databases Pubmed, EMBASE, Web of Science, Google Scholar and Cochrane Library were searched to identify capable studies studying the influence between TKA and hindfoot malalignment. We indentified four prospective cohort studies, seven retrospective cohort studies and one case-control study. All twelve articles addressed the influence of TKA on hindfoot alignment. Seven out of nine studies which noticed an improvement of hindfoot alignment after TKA, found a significant improvement (p<0.05). Aditionally three of these studies reported a significant improvement only in valgus hindfeet (p<0.05). On the topic of hindfoot alignment influencing TKA, we identified two studies. These studies reported an impact of hindfoot alignment on the weightbearing and described that 87% of hindfeet remained in valgus alignment after TKA. Available data suggests that alignment in valgus hindfeet can improve after TKA, though long term results are not present. Contrary to last, improvement of hindfoot alignment is not expected in varus hindfeet. Furthermore hindfoot alignment deformity may cause a reduction of the long term survival of the knee prosthesis and therefore should be taken in to account.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 78 - 78
1 Apr 2018
Peiffer M Burssens A Verstraete M Boey H Clockaerts S Leenders T Victor J
Full Access

Background

A calcaneal medial osteotomy (CMO) is a surgical procedure frequently performed to correct a valgus alignment of the hindfoot. However currently little is known on its accurate influence on hindfoot alignment (HA).

Aim

To assess the influence of a CMO on HA in both 2D and 3D measurements using weightbearing CT (WBCT).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_20 | Pages 47 - 47
1 Dec 2017
Verstraete M Van Onsem S Victor J
Full Access

INTRODUCTION

Thorough understanding and feedback of the post-operative implant position relative to the pre-operative anatomy is missing in today's clinical practice. However, three dimensional insights in the local under or oversizing of the implant can provide important feedback to the surgeon. For the knee for instance, to identify a shift in the sagittal joint line that potentially links to mid-flexion instability or to identify zones at risk for soft tissue impingement.

Despite a proven inferior outcome, clinical post-operative implant evaluation remains primarily based on bi-planar, static 2D x-rays rather than 3D imaging. Along with the cost, a possible reason is the increased radiation dose and/or metal artifact scatter in computed tomography (CT) and/or magnetic resonance imaging (MRI). These detrimental effects are now avoided by using recently released x-ray processing software. This technique uses standard-of-care post-operative x-rays in combination with a pre-operative CT and 3D file of the implant to determine the implant position relative to the pre-operative situation. The accuracy of this new technique is evaluated in this paper using patient cases. Therefore, the obtained implant position is benchmarked against post-operative CT scans.

MATERIALS & METHODS

Retrospectively, 19 patients were selected who underwent total knee arthroplasty and received pre- and post-operative CT of their diseased knee. The CT scans were performed with a pixel size of 0.39 mm and slice spacing of 0.60 mm (Somatom, Siemens, München, Germany). All patients underwent TKA surgery using the same bi-cruciate substituting total knee (Journey II, Smith&Nephew, Memphis, USA). Following surgery, standard bi-planar standing x-rays of the operated knee was additionally performed as standard of care. To evaluate the implant position relative to the pre-operative situation, the 3D implants are first positioned on the post-operative CT slices. Using Mimics (Materialise NV, Leuven, Belgium), the pre-operative bone was subsequently automatically matched onto the post-operative scan to identify the implant location relative to the reconstructed pre-operative bone. This has been independently repeated by three observers to assess the inter-observer variability. Second, the post-operative bi-planar x-rays are combined with the reconstructed pre-operative bone and 3D file of the implant. This combination is performed using the 2D-to-3D conversion integrated in the recently launched X-ray module of Mimics. This module uses a contour based registration method to determine the implant and bone position using the post-operative x-rays. For both reconstruction methods, the implant position has been evaluated in six degrees of freedom using an automated Matlab routine; resulting in three translations and three rotations.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_20 | Pages 48 - 48
1 Dec 2017
Verstraete M Arnout N De Baets P Vancouillie T Van Hoof T Victor J
Full Access

INTRODUCTION

To assess and compare the effect of new orthopedic surgical procedures, in vitro evaluation remains critical during the pre-clinical validation. Focusing on reconstruction surgery, the ability to restore normal kinematics and stability is thereby of primary importance. Therefore, several simulators have been developed to study the kinematics and create controlled boundary conditions.

To simultaneously capture the kinematics in six degrees of freedom as outlined by Grood & Suntay, markers are often rigidly connected to the moving bone segments. The position of these markers can subsequently be tracked while their position relative to the bones is determined using computed tomography (CT) of the test specimen with the markers attached. Although this method serves as golden standard, it clearly lacks real-time feedback. Therefore, this paper presents the validation of a newly developed real-time framework to assess knee kinematics at the time of testing.

MATERIALS & METHODS

A total of five cadaveric fresh frozen lower limb specimens have been used to quantitatively assess the difference between the golden standard, CT based, method and the newly developed real-time method. A schematic of the data flow for both methods. Prior to testing, both methods require a CT scan of the full lower limb. During the tests, the proximal femur and distal tibia are necessarily resected to fit the knees in the test setup, thus also removing the anatomical landmarks needed to evaluate their mechanical axis. Subsequently, a set of three passive markers are rigidly attached to the femur and tibia, referred to as M3F and M3T respectively. For the CT based method, the marker positions are captured during the tests and a second CT scan is eventually performed to link the marker positions to the knee anatomy. Using in-house developed software, this allowed to offline evaluate the knee kinematics in six degrees of freedom by combining both CT datasets with the tracked marker positions. For the newly developed real-time method, a calibration procedure is first performed. This calibration aims to link the position of the 3D reconstructed bone and landmarks with the attached markers. A set of bone surface points is therefore registered. These surface points are obtained by tracking the position of a pen while touching the bone surface. The pen's position is thereby tracked by three rigidly attached markers, denoted M3P. The position of the pen tip is subsequently calculated from the known pen geometry. The iterative closest point (ICP) algorithm is then used to match the 3D reconstructed bone to the registered surface points. Two types of 3D reconstructions have therefore been considered. First, the original reconstructions were used, obtained from the CT data. Second, a modified reconstruction was used. This modification accounted for the finite radius (r = 1.0 mm) of the registration pen, by shifting the surface nodes 1.0 mm along the direction of the outer surface normal. During the tests, the positions of the femur and tibia markers are tracked and streamed in real-time to an in-house developed, Matlab based software framework (MathWorks Inc., Natick, Massachussets, USA). This software framework simultaneously calculates the bone positions and knee kinematics in six degrees of freedom, displaying this information to the surgeons and operators. To assess the accuracy, all knee specimens have been subjected to passive flexion-extension movement ranging from 0 to 120 degrees of flexion. For each degree of freedom, the average root mean square (RMS) difference between both measurement methods has been evaluated during this movement. In addition, the distribution of the registered surface points has been assessed along the principal directions of the uniformly meshed 3D reconstructions (average mesh size of 1.0 mm).


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 894 - 903
1 Jul 2017
Bonnin MP Saffarini M Nover L van der Maas J Haeberle C Hannink G Victor J

Aims

The morphometry of the distal femur was largely studied to improve bone-implant fit in total knee arthroplasty (TKA), but little is known about the asymmetry of the posterior condyles. This study aimed to investigate the dimensions of the posterior condyles and the influence of externally rotating the femoral component on potential prosthetic overhang or under-coverage.

Patients and Methods

We analysed the shape of 110 arthritic knees at the time of primary TKA using pre-operative CT scans. The height and width of each condyle were measured at the posterior femoral cut in neutral position, and in 3º and 5º of external rotation, using both central and medial referencing systems. We compared the morphological characteristics with those of 14 TKA models.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 81 - 81
1 May 2017
Bonnin M de Kok A Verstraete M Van Hoof T Van der Straten C Victor J
Full Access

Background

The goal of this study was (1) to investigate the relationships between the bony contours of the knee and the popliteus tendon before and after TKA and (2) to analyse the influence of implant sizing. Our hypothesis was that an apparently well-sized prosthesis, will modify the position or the tracking of popliteus tendon.

Methods

4 fresh frozen cadavers were selected. The popliteus tendon was injected with contrast dye and a CT-scan was performed from full extension to full flexion with increments of 20°. Afterwards a total knee arthroplasty (TKA) was performed. Each cadaver received either a normal-sized, oversized, undersized or mobile bearing prosthesis. After TKA the limb was scanned again using the same protocol as before. 3D-reconstructions were created using Materialise Mimics software. These 3D-models were then imported into custom made Matlab software to measure and compare the deviation of the popliteus tendon before and after TKA.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 34 - 34
1 May 2017
Verstraete M De Coninck S Smis J De Baets P Victor J
Full Access

Background

A new knee simulator has been developed at Ghent University. This simulator provides the unique opportunity of evaluating the knee kinematics during activities of daily living. The simulator therefore controls the position of the ankle in the sagittal plane while keeping the hip at a fixed position. This approach provides full kinematic freedom to the knee. To evaluate and validate the performance of the simulator, the development of and comparison with a numerical simulation model is discussed in this paper.

Methods

Both a two and three dimensional simulation model have been developed using the AnyBody Modelling System (AMS). In the two dimensional model, the knee joint is represented by a hinge. Similarly, the ankle and hip joint are represented by a hinge joint and a variable amplitude quadriceps and hamstrings force is applied. In line with this simulation model, a hinge model was created that could be mounted in the UGent knee simulator to evaluate the performance of the simulated model. The hinge model thereby performs a cyclic motion under varying simulated muscle loads while recording the ankle reaction forces. In addition to the two dimensional model, a three dimensional model has been developed. More specifically, a model is built of a sawbone leg holding a posterior stabilised single radius total knee implant. The physical sawbone model contains simplified medial and lateral collateral ligaments. In line with the boundary conditions of the UGent knee simulator, the simulated hip contains a single rotational degree of freedom and the ankle holds four degrees of freedom (three rotations, single translation). In the simulations, the knee is modelled using the force-dependent kinematics (FDK) method built in the AMS. This leaves the knee with six degrees of freedom that are controlled by the ligament tension in combination with the applied quadriceps load and shape of the implant. The physical sawbone model goes through five cycles in the UGent simulator using while recording the kinematics of the femur and tibia using a set of markers rigidly attached to the femur and tibia bone. The position of the implant with respect to the markers was evaluated by CT-scanning the sawbone model.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 112 - 112
1 Apr 2017
Herregodts S De Baets P Victor J Verstraete M
Full Access

Background

Biomechanical joint contact pressure distribution measurements have proven to be a very valuable tool in orthopaedic research to investigate the influence of surgical techniques such as total knee arthroplasty (TKA) on the human knee joint. Quantification of the in vitro tibiofemoral and patellofemoral contact pressure distribution before and after the intervention are an important measure to evaluate the impact of the surgery. The K scan pressure sensor from Tekscan (South Boston USA) is a commonly reported device for these in vitro pressure measurements. Despite the large interest in the sensor, the effective measurement accuracy for in vitro biomechanical joint contact measurement still remains a big question and therefore the reliability of these measurements should be questioned.

Methods

Reliable contact pressure measurements can only be done if the sensor behaviour is fully understood. Therefore, a tailored multi-axial testing machine has been designed to profoundly investigate and characterise the sensor behaviour. This test setup is unique through its ability to apply a predefined tangential force or sliding velocity to the sensor's interface next to a normal force. Dynamic effects occurring in knee joint motion can thus be simulated while evaluating the effect on the contact pressure measurements.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 111 - 111
1 Apr 2017
Van Onsem S Lambrecht D Verstraete M Van Der Straeten C Victor J
Full Access

Introduction

Better functional outcomes, lower pain and better stability have been reported with knee designs which restore physiological knee kinematics. Also the ability of the TKA design to properly restore the physiological femoral rollback during knee flexion, has shown to be correlated with better restoration of the flexor/extensor mechanism, which is fundamental to the function of the human knee. The purpose of the study is to compare the kinematics of three different TKA designs, by evaluating knee motion during Activities of Daily Living. The second goal is to see if there is a correlation between the TKA kinematics and the patient reported outcomes.

Methods

Ten patients of each design, who are at least 6 months after their Total Knee Replacement, will be included in this study. Seven satisfied and 3 dissatisfied patients will be selected for each design. In this study 5 different movements will be analysed: flexion/extension; Sitting on and rising from a chair, Stair climbing, descending stairs, Flexion and extension open chain and squatting. These movements will be captured with a fluoroscope. The 2D images that are obtained, will be matched with the 3D implants. This 3D image will be processed with custom-made software to be able to analyse the movement. Tibio-femoral contact points of the medial and lateral condyles, tibio-femoral axial rotation, determination of the pivot-point will be analysed and described. After this analysis, a correlation between the kinematics and the KOOS and KSS will be investigated.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 29 - 29
1 Apr 2017
Verstraete M Van Onsem S Stevens C Herregodts S Arnout N Victor J
Full Access

Background

In-vitro testing of knee joints remains vital in the understanding of knee surgery and arthroplasty. However, based on the design philosophy of the original Oxford knee simulator, this in-vitro testing has mainly focused on squatting motion. As the activities of daily living might drastically differ from this type of motion, both from a kinematic and kinetic point of view, a new knee simulator is required that allows studying more random motion patterns. This paper describes a novel knee simulator that overcomes the limitations of traditional Oxford simulators, providing both kinematic and kinetic freedom with respect to the applied boundary conditions.

Methods

This novel test simulator keeps the hip at a fixed position, only providing a single rotational degree of freedom (DOF) in the sagittal plane. In addition, the ankle holds four DOF, including all rotational DOF and the translation along the medio-lateral axis. Combining these boundary conditions leaves five independent DOF to the knee; the knee flexion angle is actively controlled through the positioning of the ankle joint in the antero-posterior and proximal-distal direction. The specimens' quadriceps muscle is actively controlled, the medial and lateral hamstrings are passively loaded. To validate the performance of this simulator, two fresh frozen specimens have been tested during normal squatting and cycling. Their kinematic patterns have been compared to relevant literature data.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_5 | Pages 134 - 134
1 Mar 2017
Salvadore G Meere P Verstraete M Victor J Walker P
Full Access

INTRODUCTION

During TKA surgery, the usual goal is to achieve equal balancing between the lateral and medial side, which can be achieved by ligament releases or “pie crusting”. However little is known regarding a relationship between the balancing forces on the medial and lateral plateaus during TKA surgery, and the varus and valgus and rotational laxities when the TKA components are inserted. It seems preferable that the laxity after TKA is the same as for the normal intact knee. Hence the first aim of this study was to compare the laxity envelope of a native knee, with the same knee after TKA surgery. The second aim was to examine the relationship between the Varus-Valgus (VV) laxity and the contact forces on the tibial plateau.

METHODS

A special rig that reproduced surgical conditions and fit onto an operating table was designed (Figure 1) (Verstraete et al. 2015). The rig allows application of a constant varus/valgus moment, and an internal-external (IE) torque. A series of heel push tests under these loading conditions were performed on 12 non-arthritic half semibodies hip-to-toe cadaveric specimens. Five were used for method development. To measure laxities, the flexion angle, the VV and the IE angle were measured using a navigation system. After testing the native knee, a TKA was performed using the Journey II BCS implant, the navigation assuring correct alignments. Soft tissue balancing was achieved by measuring compressive forces on the lateral and medial condyles with an instrumented tibial trial (Orthosensor, Dania Beach, Florida). At completion of the procedure, the laxity tests were repeated for VV and IE rotation and the contact forces on the tibial plateau were recorded, for the full range of flexion.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 64 - 64
1 Mar 2017
Van Onsem S Van Der Straeten C Arnout N Deprez P Van Damme G Victor J
Full Access

Background

Total knee arthroplasty (TKA) is a proven and cost-effective treatment for osteoarthritis. Despite the good to excellent long-term results, some patients remain dissatisfied. Our study aimed at establishing a predictive model to aid patient selection and decision-making in TKA.

Methods

Using data from our prospective arthroplasty outcome database, 113 patients were included. Pre- and postoperatively, the patients completed 107 questions in 5 questionnaires: KOOS, OKS, PCS, EQ-5D and KSS. First, outcome parameters were compared between the satisfied and dissatisfied group. Secondly, we developed a new prediction tool using regression analysis. Each outcome score was analysed with simple regression. Subsequently, the predictive weight of individual questions was evaluated applying multiple linear regression. Finally, 10 questions were retained to construct a new prediction tool.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 71 - 71
1 Mar 2017
Verstraete M Meere P Salvadore G Victor J Walker P
Full Access

Introduction

A correct balancing of the knee following TKA surgery is believed to minimize instability and improve patient satisfaction. In that respect, trial components containing force sensors can be used. These force sensors provide insight in the medial/lateral force ratio as well as absolute contact forces. Although this method finds clinical application already, the target values for both the force magnitude and ratio under surgical conditions remain uncertain.

Methods

A total of eight non-arthritic cadaveric knees have been tested mimicking surgical conditions. Therefore, the specimens are mounted in a custom knee simulator (Verstraete et al., 2015). This simulator allows to test full lower limb specimens, providing kinematic freedom throughout the range of motion. Knee flexion is obtained by lifting the femur (thigh pull). Knee kinematics are simultaneously recorded by means of a navigation system and based on the mechanical axis of the femur and tibia.

In addition, the load transferred through the medial and lateral compartment of the knee is monitored. Therefore, a 2.4 mm thick sawing blade is used to machine a slot in the tibia perpendicular to the mechanical axis, at the location of the tibial cut in TKA surgery. A complete disconnection was thereby assured between the tibial plateau and the distal tibia. To fill the created gap, custom 3D printed shims were inserted (Fig. 1). Through their specific geometry, these shims create a load deviation between two pressure pads (Tekscan type 4011 sensor) seated on the medial and lateral side. Following the insertion of the shims, the knee was closed before performing the kinematic and kinetic tests.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 49 - 49
1 Feb 2017
Bonnin M Saffarini M Victor J
Full Access

Purpose

Analysis of the morphology of the distal femur, and by extension of the femoral components in total knee arthroplasty (TKA), has been related to the aspect ratio, which represents the width of the femur. Little is known about variations in trapezoidicity (i.e whether the femur is more rectangular or more trapezoidal). This study aimed to quantify additional morphological characteristics of the distal femur and identify anatomical features associated with higher risks of over- or under-sizing of components in TKA.

Method

We analyzed the shape of 114 arthritic knees at the time of primary TKA using the pre-operative CT scans. The maximum AP dimension was measured. The mediolateral dimensions were measured on the theoretical distal resection slice at three levels: the posterior region (MLP), the central region (MLC) and the anterior region (MLA) (Fig 1). The ‘aspect’ ratio (MLC/AP) ratio quantified how wide or narrow the shape is. The ‘trapezoidicity’ ratio (MLP/MLA) ratio quantified how rectangular or trapezoidal the shape is. We also quantified the medial and lateral ‘narrowing angles’ in the anterior and central zones (α and β) (Fig 2).

The post-operative prosthetic overhang was calculated from CT-scan.

We compared the morphological characteristics with those of twelve TKA models scanned using a three-dimensional optical scanning machine (ATOS II, GOM mbH, Braunschweig, Germany) and its photogrammetric analysis software (TRITOP, GOM mbH, Braunschweig, Germany).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 98 - 98
1 Jan 2017
Bonnin M Saffarini M de KoK A Verstraete M Van Hoof T Van der Straten C Victor J
Full Access

To determine the mechanisms and extents of popliteus impingements before and after TKA and to investigate the influence of implant sizing. The hypotheses were that (i) popliteus impingements after TKA may occur at both the tibia and the femur and (ii) even with an apparently well-sized prosthesis, popliteal tracking during knee flexion is modified compared to the preoperative situation.

The location of the popliteus in three cadaver knees was measured using computed tomography (CT), before and after implantation of plastic TKA replicas, by injecting the tendon with radiopaque liquid. The pre- and post-operative positions of the popliteus were compared from full extension to deep flexion using normosized, oversized and undersized implants (one size increments).

At the tibia, TKA caused the popliteus to translate posteriorly, mostly in full extension: 4.1mm for normosized implants, and 15.8mm with oversized implants, but no translations were observed when using undersized implants. At the femur, TKA caused the popliteus to translate laterally at deeper flexion angles, peaking between 80º-120º: 2.0 mm for normosized implants and 2.6 mm with oversized implants. Three-dimensional analysis revealed prosthetic overhang at the postero-superior corner of normosized and oversized femoral components (respectively, up to 2.9 mm and 6.6 mm).

A well-sized tibial component modifies popliteal tracking, while an undersized tibial component maintains more physiologic patterns. Oversizing shifts the popliteus considerably throughout the full arc of motion. This study suggests that both femoro- and tibio-popliteus impingements could play a role in residual pain and stiffness after TKA.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 53 - 53
1 Jan 2017
Verstraete M Meere P Salvadore G Victor J Walker P
Full Access

A correct ligament loading following TKA surgery is believed to minimize instability and improve patient satisfaction. The evaluation of the ligament stress or strain is however impractical in a surgical setting. Alternatively, tibial trial components containing force sensors have the potential to indirectly assess the ligament loading. These instrumented components quantify the medial and lateral forces in the tibiofemoral joint. Although this method finds clinical application already, the target values for both the force magnitude and medial / lateral force ratio under surgical conditions remain uncertain.

A total of eight non-arthritic cadaveric knees have been tested mimicking surgical conditions. Therefore, the specimens are mounted in a custom knee simulator. This simulator allows to test full lower limb specimens, providing kinematic freedom throughout the range of motion. Knee flexion is obtained by lifting the femur (thigh pull). Knee kinematics are simultaneously recorded by means of a navigation system and based on the mechanical axis of the femur and tibia.

In addition, the load transferred through the medial and lateral compartment of the knee is monitored. Therefore, a 2.4 mm thick sawing blade is used to machine a slot in the tibia perpendicular to the mechanical axis, at the location of the tibial cut in TKA surgery. A complete disconnection was thereby assured between the tibial plateau and the distal tibia. To fill the created gap, custom 3D printed shims were inserted. Through their specific geometry, these shims create a load deviation between two Tekscan pressure pads on the medial and lateral side. Following the insertion of the shims, the knee was closed before performing the kinematic and kinetic tests.

Seven specimens showed a limited varus throughout the range of motion (ranging from 1° to 7° varus). The other knee was in valgus (4° valgus). Amongst varus knees, the results were very consistent, indicating high loads in full extension. Subsequently, the loads decrease as the knee flexes and eventually vanishes on the lateral side. This leads to consistently high compartmental load ratios (medial load / total load) in flexion.

In full extension the screw-home mechanism results in increased loads, both medially and laterally. Upon flexion, the lateral loads disappear. This is attributed to slackening of the lateral collateral ligament, in turn linked to the femoral rollback and slope of the lateral compartment. The isometry of the medial collateral ligament contributes on the other hand to the near-constant load in the medial compartment. The above particularly applies for varus knees. The single valgus knee tested indicated a higher load transmission by the lateral compartment, potentially attributed to a contracture of the lateral structures.

With respect to TKA surgery, these findings are particularly relevant when considering anatomically designed implants. For those implants, this study concludes that a tighter medial compartment reflects that of healthy varus knees. Be aware however that in full extension, higher and up to equal loads can be acceptable for the medial and lateral compartment.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 96 - 96
1 Jan 2017
Salvadore G Verstraete M Meere P Victor J Walker P
Full Access

During TKA surgery, the usual goal is to achieve equal balancing between the lateral and medial side, which can be achieved by ligament releases or “pie crusting”. However little is known regarding a relationship between the balancing forces on the medial and lateral plateaus during TKA surgery, and the varus and valgus and rotational laxities when the TKA components are inserted. It seems preferable that the laxity after TKA is the same as for the normal intact knee. Hence the first aim of this study was to compare the laxity envelope of a native knee, with the same knee after TKA surgery. The second aim was to examine the relationship between the Varus-Valgus (VV) laxity and the contact forces on the tibial plateau.

A special rig that reproduced surgical conditions and fit onto an operating table was designed (Verstraete et al. 2015). The rig allows application of a constant varus/valgus moment, and an internal-external (IE) torque. A series of heel push tests under these loading conditions were performed on 12 non-arthritic half semibodies hip-to-toe cadaveric specimens. Five were used for method development. To measure laxities, the flexion angle, the VV and the IE angle were measured using a navigation system. After testing the native knee, a TKA was performed using the Journey II BCS implant, the navigation assuring correct alignments. Soft tissue balancing was achieved by measuring compressive forces on the lateral and medial condyles with an instrumented tibial trial (Orthosensor, Dania Beach, Florida). At completion of the procedure, the laxity tests were repeated for VV and IE rotation and the contact forces on the tibial plateau were recorded, for the full range of flexion.

The average of the varus-valgus and the IE laxity envelope is plotted for the native (yellow), the TKA (pink) and the overlap between the two (orange). The average for six specimens of the contact force ratio (medial/medial+lateral force) during the varus and valgus test is plotted as a function of the laxity for each flexion angle.

The Journey II implant replicated the VV laxity of the native knee except for up to 3 degrees more valgus in high flexion. For the IE, the TKA was equal in internal rotation, but up to 5 degrees more constrained in varus in mid range. Plotting contact force ratio against VV laxity, as expected during the varus test the forces were clustered in a 0.85–0.95 ratio, implying predominant medial force with likely lateral lift-off. For the valgus test, the force ratio is more spread out, with all the values below 0.6. This could be due to the different stiffness of the MCL and LCL ligaments which are stressed during the VV test. During both tests the laxity increases progressively with flexion angle. Evidently the geometry knee reproduces more lateral laxity at higher flexion as in the anatomic situation.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 52 - 52
1 Jan 2017
Bonnin M Saffarini M Bossard N Victor J
Full Access

Analysis of the morphology of the distal femur, and by extension of the femoral components in total knee arthroplasty (TKA), has largely been related to the aspect ratio, which represents the width of the femur. Little is known about variations in trapezoidicity (i.e. whether the femur is more rectangular or more trapezoidal). This study aimed to quantify additional morphological characteristics of the distal femur and identify anatomical features associated with higher risks of over- or under-sizing of components in TKA.

We analysed the shape of 114 arthritic knees at the time of primary TKA using the pre- operative CT scans. The aspect ratio and trapezoidicity ratio were quantified, and the post- operative prosthetic overhang was calculated. We compared the morphological characteristics with those of 12 TKA models.

There was significant variation in both the aspect ratio and trapezoidicity ratio between individuals. Femoral trapezoidicity was mostly due to an inward curve of the medial cortex. Overhang was correlated with the aspect ratio (with a greater chance of overhang in narrow femurs), trapezoidicity ratio (with a greater chance in trapezoidal femurs), and the tibio- femoral angle (with a greater chance in valgus knees).

This study shows that rectangular/trapezoidal variability of the distal femur cannot be ignored. Most of the femoral components which were tested appeared to be excessively rectangular when compared with the bony contours of the distal femur. These findings suggest that the design of TKA should be more concerned with matching the trapezoidal/ rectangular shape of the native femur.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 107 - 107
1 May 2016
Verstraete M Salvadore G Victor J Meere P Walker P
Full Access

Total knee arthroplasty can largely impact the functioning of a knee. To minimize the impact of surgery and increase patient satisfaction, it is believed that restoring knee stability and control of the laxity has the potential to improve surgical outcome. In that respect, it is hypothesized that a well-balanced knee restores the native knee's laxity and stability, whereas unbalanced conditions result in an increased laxity and instability. This study intends to precisely evaluate knee laxity and stability in a cadaveric model in order to improve the clinical evaluation of the knee laxity under surgical conditions. This paper provides insight in the design considerations and methodology of a novel knee simulator and the preliminary results

In a first phase, a new knee simulator has therefore been developed. This simulator allows quantifying the knee kinematics and surgical feel at the time of surgery in a laboratory environment. More specifically, full lower limb specimens can be mounted in the simulator. This overcomes the need for disarticulation at the hip and ankle, often reported in cadaveric testing. The latter is believed to potentially release the tension in the knee and should therefore be avoided. Note that in respect to surgical conditions no muscle activation is considered for this simulator.

To facilitate a repeatable and unbiased evaluation of the knee kinematics, it is important that the knee simulator provides full kinematic freedom to the tested knee specimen. To obtain six degrees of freedom, a dedicated hip and ankle setup has been created (figure 1). The hip setup constrains the hip joint to a single axis hinge joint around the femoral head center. The remaining five degrees of freedom are built into the ankle setup. More specifically, the ankle setup has two translational degrees of freedom and full rotational freedom. The translational freedom is provided along the specimen's proximal-distal axis and medio-lateral axis. The rotational freedom is provided at a single point, using a ball in socket joint located along the mechanical axis of the tibia. The translation along the proximal-distal axis is thereby actively controlled by the operator, simulating heel push conditions. In addition to studying the neutral path kinematics, the presented simulator allows evaluating the laxity boundaries throughout the range of motion. Therefore, a constant internal/external torque can be applied to the tibia. Alternatively, a constant varus/valgus moment can be simulated.

Second, following the design and construction of this simulator, a set of ten cadaveric knees has been tested on this simulator, both before and after TKA surgery. For the native knees, the results of these tests confirm the kinematic freedom provided to the tested knee. In addition, the laxity envelope around the neutral path can be realistically evaluated and quantified.

Conclusion

Design and evaluation of new knee simulator that allows synchronous studying of the knee kinematics, contact loads and tensile forces, under neutral conditions and extreme varus/valgus moment or internal/external tibial torque.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 108 - 108
1 May 2016
Verstraete M Herregodts S De Baets P Victor J
Full Access

Introduction

For the evaluation of new orthopaedic implants, cadaveric testing remains an attractive solution. However, prior to cadaveric testing, the performance of an implant can be evaluated using numerical simulations. These simulations can provide insight in the kinematics and contact forces associated with a specific implant design and/or positioning.

Methods

Both a two and three dimensional simulation model have been created using the AnyBody Modelling System (AMS). In the two dimensional model, the knee joint is represented by a hinge. Similarly, the ankle and hip joint are represented by a hinge joint and a variable amplitude quadriceps force is applied to a rigid bar connected to the tibia (Figure 1a). In line with this simulation model, a hinge model was created that could be mounted in the UGent knee simulator to evaluate the performance of the simulated model. The hinge model thereby performs a cyclic motion under varying quadriceps load while recording the ankle reaction forces.

In addition to the two dimensional model, a three dimensional model has been developed (Figure 1b). More specifically, a model is built of a sawbone leg holding a posterior stabilized single radius total knee implant. The physical sawbone model contains simplified medial and lateral collateral ligaments. In line with the boundary conditions of the UGent knee simulator, the simulated hip contains a single rotational degree of freedom and the ankle holds four degrees of freedom (three rotations, single translation). In the simulations, the knee is modelled using the force-dependent kinematics (FDK) method built in the AMS. This leaves the knee with six degrees of freedom that are controlled by the ligament tension in combination with the applied quadriceps load and shape of the implant. The physical sawbone model goes through five cycles in the UGent simulator using while recording the kinematics of the femur and tibia using a set of markers rigidly attached to the femur and tibia bone. The position of the implant with respect to the markers was evaluated by CT-scanning the sawbone model.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 103 - 103
1 May 2016
Van Onsem S Lambrecht D Verstraete M Van Der Straeten C Victor J
Full Access

Introduction

Better functional outcomes, lower pain and better stability have been reported with knee designs which restore physiological knee kinematics. Also the ability of the TKA design to properly restore the physiological femoral rollback during knee flexion, has shown to be correlated with better restoration of the flexor/extensor mechanism (appropriate flexor/extensor muscle lever arm, sufficient quadriceps force to extend the knee under load and limited patello-femoral force), which is fundamental to the function of the human knee. The purpose of the study is to compare the kinematics of three different TKA designs, by evaluating knee motion during Activities of Daily Living. The second goal is to see if there is a correlation between the TKA kinematics and the patient reported outcomes.

Methods

Ten patients who are at least 6 months after their Total Knee Replacement are included in this study. Seven satisfied and 3 dissatisfied patients are selected for this design. In this study 5 different movements are being analysed: flexion/extension; Sitting on and rising from a chair, Stair climbing, descending stairs, Flexion and extension open chain and squatting. These movements will be captured with a fluoroscope. The 2D images that are obtained, are matched with the 3D implants. (see figure 1 and 2.) This 3D image is processed with custom-made software to be able to analyse the movement (figure 3.). Tibio-femoral contactpoints of the medial and lateral condyles, tibio-femoral axial rotation, determination of the pivot-point are analysed and described. After this analysis, a correlation between the kinematics and the KOOS and KSS is investigated.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 105 - 105
1 May 2016
Verstraete M Van Onsem S Stevens C Arnout N Victor J
Full Access

For evaluating the impact of knee surgery, cadaveric knee simulators are commonly applied. However, most of the knee simulators are based on the Oxford type as originally described by Zavatsky (Zavatsky, J. of Biomechanics, 1997). These simulators mainly focus on the squatting motion. Although a wide range of flexion angles can be examined while performing this motion, the significance for activities of daily living is limited.

To that extent a new knee simulator has recently been developed at Ghent University. In this simulator, the ankle motion is dynamically controlled in the sagittal plane; both in the proximal/distal direction and the anterior/posterior direction. As a result, this simulator allows simulating random motion patterns, e.g. cycling, stair ascent and descent, … The ankle translation is unrestrained in the medial/lateral direction. In addition, all rotational degrees of freedom are unrestrained at the ankle, resulting in four degrees of freedom at the ankle. The hip adds one rotational degree of freedom being the rotation in the sagittal plane. This leaves 5 degrees of freedom (DOF) to the knee; the sixth being flexion/extension that is controlled by the actuators at the ankle. During the simulation of different motion patterns, the quadriceps and hamstring force are actively controlled to mimic realistic conditions obtained through musculoskeletal simulations.

In this study, five cadaveric experiments have been performed on the simulator. While mounting the cadaveric specimens in the test rig, the initial alignment remains crucial. Whilst the rig leaves 5 DOF to the knee, it is important to restore the anatomical position of the hip and ankle. To minimize the impact of the mounting procedure, cadaver specific 3D printed guides are used to assure the alignment of the cadaver in the test rig. As a result, the kinematics are more likely to represent physiological conditions. These kinematics have been evaluated in accordance to the methodology described by Grood&Suntay (Grood & Suntay, Transactions of the ASME, 1983). Therefore, a CT scan of the examined knee is combined with motion tracking data from rigidly attached markers on both the femur and the tibia. The cadaveric knees have been subjected to a variety of motion patterns, i.e. squatting and cycling. The squatting experiments provide evidence that the knee simulator creates adequate boundary conditions as the kinematic patterns coincide with literature reportings. The cycling experiments however significantly differ from the squatting patterns. Most noteworthy is the difference in terms of internal/external rotation for these native knees (Figure 1). This internal/external rotations is highly fluctuating from flexion to extension. This is understood as the quadriceps force is not constant during the extension phase, representing physiological conditions.

Conclusion

Significant difference in knee kinematics between squatting and cycling indicates the importance of testing a variety of conditions. Furthermore, this reveals the need to study clinically relevant motion patterns, selected from patient reported outcomes.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 106 - 106
1 May 2016
Verstraete M Van Onsem S Biebouw S Cortens W Arnout N Victor J
Full Access

Total knee arthroplasty aims at restoring the function of the native knee. An important aspect at this point are the knee kinematics, as it can be assumed that following TKA surgery these should resemble the native conditions. The use of cadaveric testing is since long an important step in the development and validation of implant designs and surgical techniques. However, this cadaveric testing has primarily focused on squatting under load bearing conditions. The main research question of this paper is therefore to evaluate the impact of TKA surgery on the knee kinematics under a range of boundary conditions.

A set of five cadaveric knees have been tested in a newly developed and validated knee simulator at Ghent University. In contrast to other simulators, this simulator allows simulating a wide range of conditions as it facilitates a controlled movement of the ankle in the sagittal plane under continuously variable hamstring and quadriceps loading. In the framework of this study, two different motion patterns have been studied. First, the knees were subjected to a traditional squatting motion maintaining constant quadriceps loading. Second, the knees were tested while performing a cycling movement with a highly variable quadriceps load during the extension phase. For both cases, the studied motion patterns have been repeated five times. Following the evaluation of the native knee kinematics, TKA surgery was performed using a single radius implant. During surgery, the implant alignment has been controlled using computer navigation. Subsequently, the same boundary conditions have been applied and the kinematics again recorded.

Focusing on the native knee, the measured kinematic patterns for the squatting motion significantly differ from the ones observed for the cycling movement for similar flexion angles. This is attributed to a difference in quadriceps loading. However, following TKA surgery, the kinematic patterns are remarkably comparable between the squatting and cycling experiments. These observations suggest that the TKA design considered in this study displays a highly constrained behavior. More specifically, the design appears to favor the squatting behavior. Further study is however required to thoroughly evaluate this observation for other implant designs and a wider range of motion patterns.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 47 - 47
1 May 2016
Bonnin M De Kok A Verstraete M Van Hoof T Van Der Straeten C Victor J
Full Access

Goals of the study

(1) to investigate the relationships between the bony contours of the knee and the Popliteus Tendon (PT) in the healthy knee and after implantation of a TKA and (2) to analyze the influence of implant sizing.

Hypothesis

With an apparently well-sized TKA, the position of the PT during knee flexion is modified compared with the preoperative situation.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 98 - 98
1 Jan 2016
Verstraete M Van Der Straeten C Opsomer G De Lepeleere B Victor J
Full Access

An accurate evaluation of the mechanical properties of human tissue is key to understanding and successfully simulating (parts of) human joints. Due to the rapid post-mortem decay, however, the cadavers are usually frozen or embalmed. The main aim of this paper is to quantitatively compare the impact of both techniques on the biomechanical properties. To that extent, the Achilles tendons of seven cadavers have been tested. For each cadaver, one of the Achilles tendons was tested after being frozen for at maximum two weeks, whilst the other tendon was tested following a Thiel embalming process.

All specimens were gripped in custom made clamps and subjected to uniaxial tensile loading. The specimens were scanned using a micro-CT to determine their cross-sectional area, which allowed transferring the applied forces to stresses. During the tensile tests, the specimens’ elongation was measured both using the digital image correlation (DIC) technique and using linear variable displacement transducers (LVDT's) mounted across the grips. The former allowed to assess the severity of slip in the grips. As is well described in literature, the obtained stress-strain relationship is not linear (Figure 1). Accordingly, the following bilinear relationship was fitted through the data points using a least squares fit:

s = E0 e     e <= ê

s = E0 ê + E (e - ê)     e > ê

As a result, the stress-strain response is sub-divided in two regions: a toe-region (e <= ê) with a low slope and stiffness (E0) and a linear elastic region (e > ê) with a higher stiffness (E). Both stiffness values were subsequently compared between the fresh frozen and Thiel embalmed group. Given the non-normal distribution of the test data, the non-parametric Wilcoxon signed rank test was used to assess the statistical significance of the obtained results.

No statistically significant difference was observed between the stiffness of the toe-region (e <= ê) obtained from Thiel embalmed and fresh frozen specimens (p-value = 0.249). In the contrary, the stiffness of linear elastic region (e > ê) was significantly different between both groups (p-value = 0.046 – see Figure 2). An average, the Thiel embalmed specimens displayed a 36% higher stiffness compared to the fresh frozen specimens. The latter contrasts the findings of other studies reported in literature, which report a decrease of the stiffness following Thiel embalming. To the authors’ opinion, this discrepancy could either be attributed to a difference in testing protocol (embalming time, donor factors, …) or tissue perfusion kinetics (Achilles tendon is relatively massive).

In conclusion, this study has demonstrated that Thiel embalming significantly alters the biomechanical properties of tendons. Specimens that underwent Thiel embalming should therefore not be considered for determining input parameters for advanced numerical models.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 49 - 57
1 Jan 2016
Bonnin MP Saffarini M Bossard N Dantony E Victor J

Aims

Analysis of the morphology of the distal femur, and by extension of the femoral components in total knee arthroplasty (TKA), has largely been related to the aspect ratio, which represents the width of the femur. Little is known about variations in trapezoidicity (i.e. whether the femur is more rectangular or more trapezoidal). This study aimed to quantify additional morphological characteristics of the distal femur and identify anatomical features associated with higher risks of over- or under-sizing of components in TKA.

Methods

We analysed the shape of 114 arthritic knees at the time of primary TKA using the pre-operative CT scans. The aspect ratio and trapezoidicity ratio were quantified, and the post-operative prosthetic overhang was calculated. We compared the morphological characteristics with those of 12 TKA models.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 99 - 99
1 Jan 2016
Verstraete M Van Der Straeten C Victor J
Full Access

To evaluate the impact of a knee prosthesis on the soft-tissue envelope or knee kinematics, cadaveric lower extremities are often mounted in a custom test rig, e.g. Oxford knee rig. Using such test rig, the knee is tested while performing a squatting motion. However, such motion is of limited daily-life relevance and clinical practices has shown that squatting commonly causes problems for knee patients. As a result, a new test rig was developed that allows a random, controlled movement of the ankle relative to the hip in the sagittal plane.

Mounting the specimen in the test rig, restricts five degrees of freedom (DOF) at the hip; only the rotation in the sagittal plane is not restrained (Figure 1). On the other hand, at the ankle, only two degrees of freedom are restrained, namely the movement in the sagittal plane. The ankle has thus three rotational degrees of freedom, all rotation axis intersect in a single point: the center of the ankle. In addition, the out-of-plane translational movement of the ankle remains free. This is achieved by means of a linear bearing. The other translational degrees of freedom, in the sagittal plane, are controlled by two actuators. As a result, the knee has five degrees of freedom left; flexion-extension is controlled. This represents typical closed chain applications, such as cycling.

In a first step, the knee kinematics have been evaluated under un-loaded conditions (no quadriceps or hamstring forces applied). To evaluate the knee kinematics, an infrared camera system (OptiTrack, NaturalPoint Inc, USA) is used. Therefore, three infrared markers are placed on the femur and tibia respectively. In addition, markers are placed on the test rig itself, to evaluate the accuracy of the applied motion. All markers are tracked using eight infrared cameras. At the ankle, a 2D circular motion with a radius of 100 mm was applied. Based on the 3D motion analysis, it was demonstrated that the control system has an accuracy of ± 0.5 mm.

The evaluation of the knee kinematics in accordance to Grood and Suntay (J. of Biomechanical Engineering, 1983), additionally requires the evaluation of the knee anatomy. To that extent, the cadaveric specimen has been visualized using a CT scan, with the infrared markers in place. From these CT images, a 3D reconstruction has been created (Mimics, Materialise, Belgium). Subsequently, custom software has been developed that combines the CT data with the motion analysis data (Matlab, The MathWorks Inc., USA). As a result, knee motion is visualized in 3D (Figure 2.a) and clinical relevant kinematic parameters can be derived (Figure 2.b).

In conclusion, the presented test rig and analysis framework is ready to evaluate more complex knee kinematics with reasonable accuracy and stability of the control loops. Future research will however primarily focus on the evaluation and validation of the impact of forces applied onto the specimen.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 100 - 100
1 Jan 2016
Verstraete M Luyckx T De Roo K Bellemans J Victor J
Full Access

It is nowadays widely recognized that patient satisfaction following knee arthroplasty strongly depends on ligament balancing. To obtain this balancing, the occurring ligament strain is assumed to play a crucial role. To measure this strain, a method is described in this paper that allows full field 3D evaluation of the strains. The latter is preferred over traditional measurement techniques, e.g. displacement transducers or strain gauges, as human soft tissue is not expected to deform uniformly due to its highly inhomogeneous and anisotropic properties.

To facilitate full field strain measurements, the 3D digital image correlation (DIC) technique was adopted. This technique was previously validated by our research group on human tissue. First, a high contrast speckle pattern was applied on the sMCL. Therefore, the specimens are first coated with a small layer of methylene blue. Following, a random white speckle pattern is applied. During knee flexion, two cameras simultaneously take pictures of the deforming region at predefined flexion angles. Using dedicated software, the captured images are eventually combined and result in 3D full field strains and displacements.

Using this method, the strain distribution was studied in six cadaveric knees during flexion extension movement. Therefore, the femur was rigidly fixed in a custom test rig. The tibia was left unconstrained, allowing the six degrees of freedom in the knee. A load was applied to all major muscles in physiological directions of each muscle by attaching a series of calibrated weights (Farahmand et al., J Orthop Res., 1998;16(1)). The direction of the pulling cables was controlled using a digital inclinometer for each specimen. As a result, a statically balanced muscle loading of the knee was obtained.

From these cadaveric experiments, it is observed that on average the sMCL behaves isometrically between 0° and 90° of flexion. However, high regional differences in strain distribution are observed from the full field measurements. The proximal region of the sMCL experiences relatively high strains upon flexion. These strains are positive (tension) in the anterior part and negative (compression) in the posterior region. In contrast, the distal region remains approximately isometric upon knee flexion (see Figure 1).

It is accordingly concluded that the sMCL behaves isometric, though large regional differences are observed. The proximal region experiences higher strains. Furthermore, the DIC technique provided valuable insights in the deformation of the sMCL. This technique will therefore be applied to study the impact of knee arthroplasty in the near future.

Caption with figure 1: Full field strain distribution in the sMCL's longitudinal direction for specimen in 45° (a) and 90° (b) of knee flexion


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 88 - 88
1 Jan 2016
Van Der Straeten C Van Onsem S Callens R Slabbaert M Victor J
Full Access

Background

The use of Computed Tomography (CT) as a medical imaging tool has widespread applications in the field of knee surgery. Surgeons use a CT scan in a conventional way during the pre-operative stage, to plan the position of the femoral component in the horizontal plane. In the post-operative stage, the use of a CT scan is a routine tool in the evaluation of failed TKA as rotational malalignment of the femoral component has been determined as a cause of poor clinical outcome after TKA.

Aim

How accurately can we measure the different angles with importance for alignment on a 3D-image in comparison to a standard CT, 2D, image.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 89 - 89
1 Jan 2016
Van Der Straeten C Van Onsem S Victor J
Full Access

Introduction

Total Knee Arthroplasty (TKA) is a proven successful and cost-effective method to relieve pain and improve joint function and quality of life in patients with advanced knee arthritis. However, after a TKA, only 75 to 89% of patients are satisfied. Since patient satisfaction is one of the main objectives of elective orthopaedic surgery, it is important to investigate the reasons for dissatisfaction and develop remedial strategies.

Objectives

The aims of the current project are to investigate patient satisfaction after TKA and establish physical, mental and social determinants of patient satisfaction and overall socio-economic costs associated with unsatisfactory outcome. The global objective is to identify patients most likely to benefit from primary TKA, establish optimal evidence-based indications and timing for TKA, and address the necessity to educate patients preoperatively towards realistic expectations or propose alternative therapies. Based on the results, a composite score for patient selection will be developed using objective and subjective parameters. Cut-off values for acceptable indications for TKA will be proposed.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 91 - 91
1 Jan 2016
Van Der Straeten C Doyen B Dutordoir C Goedertier W Pirard S Victor J
Full Access

INTRODUCTION

Meniscal tears are very common and treated surgically by suturing or partial or total meniscectomy. After meniscectomy, the tibiofemoral contact area is decreased whih leads to higher contact stresses associated with clinical symproms and a faster progression of tibiofemoral osteoarthritis. Besides meniscus allograft transplantation, artificial implants have been developed to replace the menisci after meniscectomy.

AIM

We investigated the short- and medium-term clinical results and survivorship of two artificial meniscus implants used as a treatment for post-meniscectomy pain in young to middle-aged patients: the anchored polyurethane degradable Actifit® (2007–2013) and the non-anchored polycarbonate-urethane NUSurface® meniscal implants (2011–2013).


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 318 - 323
1 Mar 2015
Verdonk P Dhollander A Almqvist KF Verdonk R Victor J

The treatment of osteochondral lesions is of great interest to orthopaedic surgeons because most lesions do not heal spontaneously. We present the short-term clinical outcome and MRI findings of a cell-free scaffold used for the treatment of these lesions in the knee. A total of 38 patients were prospectively evaluated clinically for two years following treatment with an osteochondral nanostructured biomimetic scaffold. There were 23 men and 15 women; the mean age of the patients was 30.5 years (15 to 64). Clinical outcome was assessed using the Knee Injury and Osteoarthritis Outcome Score (KOOS), the Tegner activity scale and a Visual Analgue scale for pain. MRI data were analysed based on the Magnetic Resonance Observation of Cartilage Repair Tissue (MOCART) scoring system at three, 12 and 24 months post-operatively. There was a continuous significant clinical improvement after surgery. In two patients, the scaffold treatment failed (5.3%) There was a statistically significant improvement in the MOCART precentage scores. The repair tissue filled most of the defect sufficiently. We found subchondral laminar changes in all patients. Intralesional osteophytes were found in two patients (5.3%). We conclude that this one-step scaffold-based technique can be used for osteochondral repair. The surgical technique is straightforward, and the clinical results are promising. The MRI aspects of the repair tissue continue to evolve during the first two years after surgery. However, the subchondral laminar and bone changes are a concern.

Cite this article: Bone Joint J 2015; 97-B:318–23.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_1 | Pages 75 - 75
1 Feb 2015
Victor J
Full Access

Osteotomy is one of the oldest orthopaedic interventions and has evolved significantly over the years. The procedure is well established as a biomechanical solution in the treatment of arthritis and instability of the knee. The operation is technically demanding and carries risks of neurovascular injury, inadequate fixation and under- or overcorrection. These technical problems have given osteotomy significant headwind in the orthopaedic community. The relative success of knee arthroplasty (uni or total) in the past decade has fed the perception that this procedure is the only remaining treatment to be trusted for patients with knee arthritis. However, both registry data and single center studies often show disappointing results for knee arthroplasty in the young, active and demanding patient population. Osteotomy has a significant role for these patients, provided they have unicompartmental arthritis with constitutional malalignment. Also, more complex deformities as seen in the post-traumatic setting often need a biomechanical approach based upon osteotomy principles.

Recently, technology was developed to allow the surgeon perform a three-dimensional evaluation of the deformity and prediction of postoperative alignment. Patient specific guides with a broad fit on the femur or tibia can guide the osteotomy and fixation accurately, within 2 degrees of accuracy. With this technological approach, a new dawn for osteotomy appears on the horizon.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_1 | Pages 124 - 124
1 Feb 2015
Victor J
Full Access

Ligament release in knee arthroplasty can be limited with accurate placement of implants, correct sizing and anatomic geometry of the implants. Still, in some instances a release of the medial collateral ligament can be necessary. In the proposed minimally invasive technique, multiple punctures in the MCL are made, using an 19-gauge needle, in order to progressively stretch the MCL until a correct ligament balance is achieved. This technique requires no additional soft tissue dissection and can even be performed percutaneously when necessary.

We analyzed 61 cases with varus deformity that were intraoperatively treated using this technique. In 4 other cases the technique was used as a percutaneous procedure in order to correct postoperative medial tightness. The procedure was considered successful when 2–4mm mediolateral joint line opening was obtained in extension and 2–6mm in flexion. In 62 cases (95%) a progressive correction of medial tightness was achieved according to the above described criteria. We therefore consider needle puncturing as an effective and safe technique for progressive correction of MCL tightness during minimally invasive TKA.