METHODOLOGY FOR EVALUATING KNEE LAXITY AND STABILITY UNDER SURGICAL CONDITIONS

The International Society for Technology in Arthroplasty (ISTA), 28th Annual Congress, 2015. PART 4.



Abstract

Total knee arthroplasty can largely impact the functioning of a knee. To minimize the impact of surgery and increase patient satisfaction, it is believed that restoring knee stability and control of the laxity has the potential to improve surgical outcome. In that respect, it is hypothesized that a well-balanced knee restores the native knee's laxity and stability, whereas unbalanced conditions result in an increased laxity and instability. This study intends to precisely evaluate knee laxity and stability in a cadaveric model in order to improve the clinical evaluation of the knee laxity under surgical conditions. This paper provides insight in the design considerations and methodology of a novel knee simulator and the preliminary results

In a first phase, a new knee simulator has therefore been developed. This simulator allows quantifying the knee kinematics and surgical feel at the time of surgery in a laboratory environment. More specifically, full lower limb specimens can be mounted in the simulator. This overcomes the need for disarticulation at the hip and ankle, often reported in cadaveric testing. The latter is believed to potentially release the tension in the knee and should therefore be avoided. Note that in respect to surgical conditions no muscle activation is considered for this simulator.

To facilitate a repeatable and unbiased evaluation of the knee kinematics, it is important that the knee simulator provides full kinematic freedom to the tested knee specimen. To obtain six degrees of freedom, a dedicated hip and ankle setup has been created (figure 1). The hip setup constrains the hip joint to a single axis hinge joint around the femoral head center. The remaining five degrees of freedom are built into the ankle setup. More specifically, the ankle setup has two translational degrees of freedom and full rotational freedom. The translational freedom is provided along the specimen's proximal-distal axis and medio-lateral axis. The rotational freedom is provided at a single point, using a ball in socket joint located along the mechanical axis of the tibia. The translation along the proximal-distal axis is thereby actively controlled by the operator, simulating heel push conditions. In addition to studying the neutral path kinematics, the presented simulator allows evaluating the laxity boundaries throughout the range of motion. Therefore, a constant internal/external torque can be applied to the tibia. Alternatively, a constant varus/valgus moment can be simulated.

Second, following the design and construction of this simulator, a set of ten cadaveric knees has been tested on this simulator, both before and after TKA surgery. For the native knees, the results of these tests confirm the kinematic freedom provided to the tested knee. In addition, the laxity envelope around the neutral path can be realistically evaluated and quantified.

Conclusion

Design and evaluation of new knee simulator that allows synchronous studying of the knee kinematics, contact loads and tensile forces, under neutral conditions and extreme varus/valgus moment or internal/external tibial torque.