3D STRAIN MAPPING OF SUPERFICIAL COLLATERAL LIGAMENT

The International Society for Technology in Arthroplasty (ISTA), 27th Annual Congress. PART 4.



Abstract

It is nowadays widely recognized that patient satisfaction following knee arthroplasty strongly depends on ligament balancing. To obtain this balancing, the occurring ligament strain is assumed to play a crucial role. To measure this strain, a method is described in this paper that allows full field 3D evaluation of the strains. The latter is preferred over traditional measurement techniques, e.g. displacement transducers or strain gauges, as human soft tissue is not expected to deform uniformly due to its highly inhomogeneous and anisotropic properties.

To facilitate full field strain measurements, the 3D digital image correlation (DIC) technique was adopted. This technique was previously validated by our research group on human tissue. First, a high contrast speckle pattern was applied on the sMCL. Therefore, the specimens are first coated with a small layer of methylene blue. Following, a random white speckle pattern is applied. During knee flexion, two cameras simultaneously take pictures of the deforming region at predefined flexion angles. Using dedicated software, the captured images are eventually combined and result in 3D full field strains and displacements.

Using this method, the strain distribution was studied in six cadaveric knees during flexion extension movement. Therefore, the femur was rigidly fixed in a custom test rig. The tibia was left unconstrained, allowing the six degrees of freedom in the knee. A load was applied to all major muscles in physiological directions of each muscle by attaching a series of calibrated weights (Farahmand et al., J Orthop Res., 1998;16(1)). The direction of the pulling cables was controlled using a digital inclinometer for each specimen. As a result, a statically balanced muscle loading of the knee was obtained.

From these cadaveric experiments, it is observed that on average the sMCL behaves isometrically between 0° and 90° of flexion. However, high regional differences in strain distribution are observed from the full field measurements. The proximal region of the sMCL experiences relatively high strains upon flexion. These strains are positive (tension) in the anterior part and negative (compression) in the posterior region. In contrast, the distal region remains approximately isometric upon knee flexion (see Figure 1).

It is accordingly concluded that the sMCL behaves isometric, though large regional differences are observed. The proximal region experiences higher strains. Furthermore, the DIC technique provided valuable insights in the deformation of the sMCL. This technique will therefore be applied to study the impact of knee arthroplasty in the near future.

Caption with figure 1: Full field strain distribution in the sMCL's longitudinal direction for specimen in 45° (a) and 90° (b) of knee flexion