Stem geometry is known to influence the outcome in THA; however it is unknown whether the material properties, stiffness in particular can influence the stem stability and outcome. The aim of this study was to measure the influence of stem material properties on micromotion and migration using Roentgen Stereophotogrammetric Analysis (RSA) system. 41 patients were implanted with a collarless polished tapered (CPT) femoral stem (Zimmer, Warsaw, Indiana), which was made of either cobalt-chromium (CoCr) (n=21) or stainless steel (n=20). RSA was used to measure dynamically inducible micromotion (DIMM: difference in stem position in going from double-leg stance (DLS) to single leg stance (SLS)), prosthesis bending (difference in the head-tip distance when going from DLS to SLS), and mean migration of the head, tip and the cement restrictor. DIMM and bending were measured at 3 months, migration at 6, 12 and 24 months. All analyses were carried out using SPSS for windows (v.15.0.0, Chicago. IL, USA). Results were reported as mean ± 95% confidence interval (CI) and regarded as significant when p < 0.05. Preliminary analysis showed that total head DIMM was significantly (p = 0.02) greater for CoCr (0.97mm ± 0.6mm) than stainless steel (0.27mm ± 0.6mm). The mean stem bending for CoCr was 0.08mm (± 0.06mm) and for stainless steel 0.15mm (± 0.06mm) (p =0.77). Both implants heads migrated posteriorly, medially and distally. The mean subsidence for the cobalt-chromium and stainless steel stems was 1.02mm (± 0.19mm) (p < 0.001) and 1.12mm (± 0.34mm) (p=0.001) (p= 0.07) at 24 months. It was interesting to note that the dynamically induced micromotion was greater for the stiffer stem, however there were no differences in terms of overall migration, indicating that survival (in terms of loosening) should be the similar for both steel and CoCr versions of this implant.
Femoral stem varus has been associated with poorer results. We report the incidence of varus/valgus malalignment of the Exeter polished, double taper design in a multicentre prospective study. The surgical outcomes at a minimum of five year and complication rates are also reported. A multicentre prospective study of 987 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and femoral stem malalignment. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of dislocation and revision. The incidence of varus and valgus malignment were 7.1% and 2.6% respectively. There was no significant difference in OHS between neutral and malaligned femoral stems at 5 years (neutral, mean = 40.2; varus, mean 39.3, p = 0.465; valgus, mean = 40.9, p = 0.605). There was no significant difference in dislocation rate between the groups (p = 0.66). There was also no significant difference in revision rate (p = 0.34). This study provides evidence that the Exeter stem is extremely tolerant of varus and valgus malalignment, both in terms of outcome and complication rate.
Metal on metal hip resurfacing (MMHR) is a popular procedure for the treatment of osteoarthritis in young patients. Several centres have observed masses, arising from around these devices, we call these inflammatory pseudotumours. They are locally invasive and may cause massive soft tissue destruction. The aim of this study was to determine the incidence and risk factors for pseudotumours that are serious enough to require revision surgery. In out unit, 1,419 MMHRs were performed between June 1999 and November 2008. All revisions were identified, including all cases revised for pseudotumour. Pseudotumour diagnosis was made by histological examination of samples from revision. A Kaplan-Meier survival analysis was performed, Cox regression analysis was used to estimate the independent effects of different factors. The revision rate for pseudotumour increased with time and was 4% (95% CI: 2.2% to 5.8%) at eight years. Female gender was a strong risk factor: at eight years the revision rate for pseudotumours in men was 0.5% (95% CI 0% to 1.1%), in women over 40 it was 6% (95% CI 2.3% to 10.1%) and in women under 40 it was 25% (95% CI 7.3% to 42.9%) (p<
0.001). Other factors associated with an increase in revision rate were, small components (p=0.003) and dysplasia (p=0.019), whereas implant type was not (p=0.156). We recommend that resurfacings are undertaken with caution in women, especially those younger than 40 years of age, but they remain a good option in men. Further work is required to understand the patho-aetiology of pseudotumours so that this severe complication can be avoided.
8 MoMHRA implants revised due to pseudotumour; 22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection). The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.
significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001. Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:
21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and 14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008. Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).
Roentgen Stereophotogrammetric Analysis (RSA) can predict long-term outcome of prostheses by measuring migration over time. The Exeter femoral stem is a double-tapered highly polished implant and has been shown to subside within the cement mantle in 2 year RSA studies. It has a proven track record in terms of long-term survivorship and low revision rates. Several studies have demonstrated excellent clinical outcomes following its implantation but this is the first study to assess stem migration at 10 years, using RSA. This is a single-centre study involving 20 patients (mean age: 63 years, SD=7) undergoing primary total hip replacement for degenerative osteoarthritis using the lateral (Hardinge) approach. RSA radiographs were taken with the patient bearing full weight post-operatively, at 3, 6, 12 months and at 2, 5 and 10 years follow-up. The three-dimensional migration of the Exeter femoral stem was determined. The mean Oxford Hip Score at 10 years was 43.4 (SD=4.6) and there were no revisions. The stems subsided and rotated internally during a 10-year period. The mean migrations of the head and tip of the femoral stem in all three anatomic directions (antero-posterior, medio-lateral &
supero-distal) were 0.69 mm posterior, 0.04 mm lateral and 1.67 mm distal for the head and 0.20 mm anterior, 0.02 mm lateral and 1.23 mm distal for the tip. The total migration at 10 years was 1.81 mm for the head and 1.25 mm for the tip. The Exeter femoral stem exhibits migration which is a complex combination of translation and rotation in three dimensions. Comparing our 10 year with our previous 2 year migration results, the Exeter stems show continued, but slow distal migration and internal rotation. The subsidence continues to compress the cement and bone-cement interface which maintains secure fixation in the long term.
We report the clinical, radiographic and histologic features of these cases.
There were no infections in this series. Three subjects required revision surgery.
This study suggests that resurfacing arthoplasty can also induce a local hypersensitivity reaction in response to metal wear debris. It therefore raises new concerns regarding the long-term safety of this procedure.
The Charnley Elite femoral component was first introduced in 1992 as a new design variant of the original Charnley femoral component (De Puy, Leeds, UK) with modified neck and stem geometry. The original component had undergone few changes in nearly forty years and has excellent long-term results. Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA)1. Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is indicated as stems of this type are still being implanted. One hundred Charnley Elite stems, implanted in our centre between 1994 and 1997 were included in a prospective, cross-sectional follow-up study. Outcome measures include validated clinical scores (Charnley hip score, Harris hip score and Oxford hip score) and radiological scores (Gruen classification) as well as revision rates over the past 10 years. The clinical follow-up supports the RSA predictions of early failure of the Charnley Elite femoral stem.
Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA). Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is required as stems of this type are still being implanted.
Preliminary clinical scores in the patients who had not undergone any subsequent surgery were adequate (Oxford Hip Score mean average of 23.9). Thirteen percent of radiographs analysed had evidence of loosening, giving an overall loosening rate of 14% at 8 years.