header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

INCREASED WEAR AND EDGE-LOADING IN METAL-ON-METAL HIP RESURFACING IMPLANTS REVISED DUE TO PSEUDOTUMOURS



Abstract

Introduction: Pseudotumours (soft-tissue masses relating to the hip joint) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated serum and hip aspirate metal ion levels, suggesting that pseudotumours occur when there is increased wear. This study aimed to quantify in vivo wear of implants revised for pseudotumours and a control group of implants revised for other reasons of failure.

Methods: A total of 30 contemporary MoMHRA implants in two groups were investigated in this Institutional Review Board approved study:

  1. 8 MoMHRA implants revised due to pseudotumour;

  2. 22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection).

The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.

Results: In comparison with the non-pseudotumour implant group, the pseudotumour implant group was associated with:

  1. significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and

  2. significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001.

Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:

  1. 21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and

  2. 14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008.

Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).

Discussion: Significantly greater linear wear rates of the MoMHRA implants revised due to pseudotumour support the in vivo elevated metal ion concentrations in patients with pseudotumours. This study provides the first direct evidence to confirm that pseudotumour is associated with increased wear at the MoM articulation. Furthermore, edge-loading with the loss of fluid film lubrication may be the dominant wear generation mechanism in patients with pseudotumour.

Correspondence should be addressed to: EFORT Central Office, Technoparkstrasse 1, CH – 8005 Zürich, Switzerland. Tel: +41 44 448 44 00; Email: office@efort.org

Author: Young-Min Kwon, United Kingdom

E-mail: young-min.kwon@ndorms.ox.ac.uk