Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

THE 5 YEAR MIGRATION OF THE BIRMINGHAM HIP RESURFACING AS ASSESSED BY RSA



Abstract

Introduction: Interest in hip resurfacing has recently been renewed by the introduction of metal-on-metal designs; it is being increasingly used for young patients, with over 30,000 implanted worldwide. The 5 year clinical results appear promising, but there are no long term data available. Radiostereometry (RSA) measures of implant migration have been able to predict implant failure; specifically large and continuous migration predicts aseptic loosening. We present the results of a five year RSA study examining the migration of the Birmingham Hip Resurfacing (BHR).

Methods: Twenty-four subjects with primary OA were implanted with the BHR device and with bone markers for RSA. RSA measurements were taken at 3, 6, 12, 24 and 60 months. The migration of the head and the tip of femoral component were measured in 3D.

Results: Preliminary analysis showed that the total 3D migration of the head and tip over five years was 0.32mm and 0.23mm respectively.

Discussion: A distal migration of more than 0.4mm over 2 years increases the likelihood of failure in conventional stems. The total migration of the BHR was approximately 0.3mm over a five year period, significantly less than cemented THR devices. The device is stable and this is promising for long-term survival.

Correspondence should be addressed to Ms Larissa Welti, Scientific Secretary, EFORT Central Office, Technoparkstrasse 1, CH-8005 Zürich, Switzerland