Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIV | Pages 24 - 24
1 Jul 2012
Mehmood S Gill H Murray D Glyn-Jones S
Full Access

Stem geometry is known to influence the outcome in THA; however it is unknown whether the material properties, stiffness in particular can influence the stem stability and outcome. The aim of this study was to measure the influence of stem material properties on micromotion and migration using Roentgen Stereophotogrammetric Analysis (RSA) system.

41 patients were implanted with a collarless polished tapered (CPT) femoral stem (Zimmer, Warsaw, Indiana), which was made of either cobalt-chromium (CoCr) (n=21) or stainless steel (n=20). RSA was used to measure dynamically inducible micromotion (DIMM: difference in stem position in going from double-leg stance (DLS) to single leg stance (SLS)), prosthesis bending (difference in the head-tip distance when going from DLS to SLS), and mean migration of the head, tip and the cement restrictor. DIMM and bending were measured at 3 months, migration at 6, 12 and 24 months. All analyses were carried out using SPSS for windows (v.15.0.0, Chicago. IL, USA). Results were reported as mean ± 95% confidence interval (CI) and regarded as significant when p < 0.05.

Preliminary analysis showed that total head DIMM was significantly (p = 0.02) greater for CoCr (0.97mm ± 0.6mm) than stainless steel (0.27mm ± 0.6mm). The mean stem bending for CoCr was 0.08mm (± 0.06mm) and for stainless steel 0.15mm (± 0.06mm) (p =0.77). Both implants heads migrated posteriorly, medially and distally. The mean subsidence for the cobalt-chromium and stainless steel stems was 1.02mm (± 0.19mm) (p < 0.001) and 1.12mm (± 0.34mm) (p=0.001) (p= 0.07) at 24 months.

It was interesting to note that the dynamically induced micromotion was greater for the stiffer stem, however there were no differences in terms of overall migration, indicating that survival (in terms of loosening) should be the similar for both steel and CoCr versions of this implant.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 80 - 80
1 May 2012
Thomas GER Simpson DJ Gill HS Glyn-Jones S Beard DJ Murray DW EPOS study group
Full Access

Femoral stem varus has been associated with poorer results. We report the incidence of varus/valgus malalignment of the Exeter polished, double taper design in a multicentre prospective study. The surgical outcomes at a minimum of five year and complication rates are also reported.

A multicentre prospective study of 987 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and femoral stem malalignment. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of dislocation and revision.

The incidence of varus and valgus malignment were 7.1% and 2.6% respectively. There was no significant difference in OHS between neutral and malaligned femoral stems at 5 years (neutral, mean = 40.2; varus, mean 39.3, p = 0.465; valgus, mean = 40.9, p = 0.605). There was no significant difference in dislocation rate between the groups (p = 0.66). There was also no significant difference in revision rate (p = 0.34).

This study provides evidence that the Exeter stem is extremely tolerant of varus and valgus malalignment, both in terms of outcome and complication rate.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 546 - 546
1 Nov 2011
Grammatopoulos G Pandit H Taylor A Whitwell D Glyn-Jones S Gundle R McLardy-Smith P Murray D Gill H
Full Access

Introduction: Metal-on-metal-hip-resurfacing-arthroplasty (MoMHRA) has been associated with the development of inflammatory pseudotumours(IP), especially in females. IPs have been linked to wear debris, which can be related to metal-ion blood levels. Acetabular component position has been shown to influence wear. We have identified an optimum component orientation minimising IP risk around an inclination/anteversion of 40°/20°±10°. Our aim was to see if this optimal position results in lower metal ions and to identify the boundary of an optimal placement zone for low wear.

Methods: A cohort of 104 patients(60M:44F) with unilateral MoMHRA was studied. Blood tests were obtained at a mean follow up of 3.9 years and serum Co/Cr levels were measured(ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. Radiographic cup inclination/anteversion were measured using EBRA. The differences in ion levels between different orientation zones were investigated. Three orientation zones were defined centered on 40°/20°: Z1-within ±5°, Z2-outside ±5°/within ±10° and Z3-within ±10°.

Results: There was a wide range of cup placements. Females had significantly (p< 0.001) smaller components(mean:51, 44–60) than males(mean: 56, 52–64). Cr levels, but not Co, were higher in females(p=0.002) and those with small femoral components(< 50mm, p =0.03). Patients with cups within Z1 (n=13) had significantly lower Co(p=0.005) and Cr(p=0.001). Males with cups within Z3(n=27) had lower ion levels in comparison to those outside, which were significantly lower for Co(p=0.049) but not Cr(p=0.084). Females had similar ion levels within and out of Z3(Cr/Co: p=0.83/0.84). Co levels were significantly lower in Z1(n=13) in comparison to Z2(n=33)(p=0.048) but Cr levels were not different (p=0.06).

Discussion: MoMHRA cups placed within ±5° of the optimum(40°/20°) had significantly lower metal ions indicating lower wear within this narrow zone. This safe zone, could extend to ±10° for males only. The narrower safe zone coupled with smaller components implanted are possible factors contributing to the increased IP incidence seen in females.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 550 - 550
1 Nov 2011
Glyn-Jones S Roques A Esposito C Walter W Tuke M Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty-induced pseudotumours are a serious complication, which occur in 1.8% of patients who undergo this procedure in our institution. The aim of this study was to measure the 3D in vivo wear on the surface of resurfacing components revised for pseudotumour, compared to a control group.

Method: Thirty-six hip resurfacing implants were divided into two groups; 18 patients with a clinical and histopathological diagnosis of pseudotumour and 18 controls (revised for femoral neck fracture and infection). Three dimensional contactless metrology (Redlux Ltd) was used to scan the surface of the femoral and acetabular components, to a resolution of 20 nanometers. The location, depth and area of the wear scar were determined for each component. A separate blinded analysis to determine the presence of absence of impingement was performed by one of the authors.

Results: The volumetric wear rate for femoral component of the pseudotumour group was 3.29 mm3/yr (SD5.7) and 0.79 mm3/yr (SD1.2) for the control group (p=0.005). In the pseudotumour group, the volumetric wear rate of the acetabular component was 2.5 mm3/yr (SD6.9) compared to 0.36 mm3/yr (SD0.80) for the control group (p=0.008). Edge-wear was detected in 89% of acetabular components in the pseudotumour group and 21% of those in the control group (p=0.01). Anterior or posterior edge-wear, consistent with impingement was present on the femoral components of 73% of patients in the pseudotumour group and 22% in the control group (p=0.01).

Discussion: This work demonstrates that implants revised for pseudotumour have significantly higher volumetric wear rates than controls. They also have a significantly higher incidence of edge-wear than controls. We suggest that a significant proportion of pseudotumours are associated with high concentrations of metal wear debris; however a minority may result from a hypersensitivity reaction to metal ions.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 547 - 547
1 Nov 2011
Thomas G Simpson D Taylor A Whitwell D Gibbons C Gundle R Mclardy-smith P Gill H Glyn-jones S Murray D
Full Access

Introduction: The use of highly cross-linked polyethylene (HXLPE) is now commonplace for total hip arthroplasty, however there is no long-term data to support its use. Hip simulator studies suggest that the wear rate of some types of HXLPE is ten times less than conventional polyethylene (UHMWPE). The outcomes of hip simulator studies are not always reproduced in vivo and there is some evidence that HXLPE wear may increase between 5 and 7 years.

Method: A prospective double blind randomised control trial was conducted using Radiostereometric Analysis (RSA). Fifty-four subjects were randomised to receive hip replacements with either UHMWPE liners or HXLPE liners. All subjects received a cemented CPT stem and uncemented Trilogy acetabular component (Zimmer, Warsaw, IN, USA). The 3D penetration of the head into the socket was determined to a minimum of 7 years.

Results: The total liner penetration was significantly different at 7 years (p=0.005) with values of 0.33 mm (SE 0.05 mm) for the HXLPE group and 0.55 mm (SE 0.05 mm) for the UHMWPE group. The steady state wear rate from 1 year onwards was significantly lower for HXLPE (0.005 mm/yr, SE 0.007 mm/yr) than for UHMWPE (0.037 mm/yr, SE 0.009 mm/yr) (p=0.007). The direction of wear was supero-lateral.

Discussion: We have previously demonstrated that the penetration in the first year is creep-dominated, from one year onwards the majority of penetration is probably due to wear. This study confirms the predictions from hip simulator studies which suggest that the wear rate of this HXLPE approaches that of metal-on-metal and ceramic-on-ceramic articulations. HXLPE may have the potential to reduce the need of revision surgery, due to wear debris induced osteolysis. It may also enable surgeons to use larger couples, thus reducing the risk of impingement and dislocation.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 546 - 546
1 Nov 2011
Grammatopoulos G Pandit H Taylor A Whitwell D Glyn-Jones S Gundle R McLardy-Smith P Gill H Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty(MoMHRA) is an alternative option to THR in the treatment of young adults with OA. A recognised MoMHRA complication is the development of an inflammatory pseudotumour(IP). Diagnosis is made with the aid of US and/or MRI. To-date, no radiographic indication of the presence of IP has been identified. Neck thinning is a recognised phenomenon in MoMHRA hips not associated with any adverse clinical events. Its pathogenesis is considered multi-factorial. Our aim was to establish whether excessive neck narrowing is associated with the presence of a pseudotumour.

Methods: Twenty-seven hips (26 patients) with IP confirmed clinically, radiologically, intra-operatively and histologically were matched for sex, age, pre-operative diagnosis, component size and follow-up with an asymptomatic MoMHRA cohort (Control n=60). For all patients, prosthesis-neck-ratio(PNR) was measured on plain AP pelvic radiographs post-operatively and at follow-up as previously described and validated.

Results: All IP patients (4M:23F) and all (12M:48F) but two controls had a posterior approach at the time of MoMHRA. Post-operatively, there was no difference in the PNR between the two groups (p=0.19). At an average follow up of 3.5 years (range:0.7–8.3), IP patients(mean 1.26, 1.10–1.79) had a significantly higher (p< 0.0001) PNR in comparison to their controls(mean 1.14, 1.03–1.35). Greater neck narrowing occurred in both genders. IP necks had narrowed by an average of 8% (range:3–23). The degree of neck narrowing was correlated with length of survival of implant (p=0.001).

Discussion: This study shows a strong association between IP and neck narrowing. Processes such as impingement and increased wear are considered to be involved in the pathogenesis of both IP and neck narrowing. Furthermore, the presence of an IP, could lead to altered vascularity via a mass effect and further contribute to neck narrowing. Neck narrowing in symptomatic MOMHRA patients should alert surgeons of the possible presence of IP.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 298 - 298
1 Jul 2011
Glyn-Jones S Pandit H Doll H McLardy-Smith P Gundle R Gibbons M Athanasou N Ostlere S Whitwell D Taylor A Gill R Murray D
Full Access

Metal on metal hip resurfacing (MMHR) is a popular procedure for the treatment of osteoarthritis in young patients. Several centres have observed masses, arising from around these devices, we call these inflammatory pseudotumours. They are locally invasive and may cause massive soft tissue destruction. The aim of this study was to determine the incidence and risk factors for pseudotumours that are serious enough to require revision surgery.

In out unit, 1,419 MMHRs were performed between June 1999 and November 2008. All revisions were identified, including all cases revised for pseudotumour. Pseudotumour diagnosis was made by histological examination of samples from revision. A Kaplan-Meier survival analysis was performed, Cox regression analysis was used to estimate the independent effects of different factors.

The revision rate for pseudotumour increased with time and was 4% (95% CI: 2.2% to 5.8%) at eight years. Female gender was a strong risk factor: at eight years the revision rate for pseudotumours in men was 0.5% (95% CI 0% to 1.1%), in women over 40 it was 6% (95% CI 2.3% to 10.1%) and in women under 40 it was 25% (95% CI 7.3% to 42.9%) (p< 0.001). Other factors associated with an increase in revision rate were, small components (p=0.003) and dysplasia (p=0.019), whereas implant type was not (p=0.156).

We recommend that resurfacings are undertaken with caution in women, especially those younger than 40 years of age, but they remain a good option in men. Further work is required to understand the patho-aetiology of pseudotumours so that this severe complication can be avoided.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 221 - 222
1 May 2011
Kwon Y Glyn-Jones S Simpson D Kamali A Counsell L Mclardy-Smith P Beard D Gill H Murray D
Full Access

Introduction: Pseudotumours (soft-tissue masses relating to the hip joint) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated serum and hip aspirate metal ion levels, suggesting that pseudotumours occur when there is increased wear. This study aimed to quantify in vivo wear of implants revised for pseudotumours and a control group of implants revised for other reasons of failure.

Methods: A total of 30 contemporary MoMHRA implants in two groups were investigated in this Institutional Review Board approved study:

8 MoMHRA implants revised due to pseudotumour;

22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection).

The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.

Results: In comparison with the non-pseudotumour implant group, the pseudotumour implant group was associated with:

significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and

significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001.

Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:

21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and

14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008.

Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).

Discussion: Significantly greater linear wear rates of the MoMHRA implants revised due to pseudotumour support the in vivo elevated metal ion concentrations in patients with pseudotumours. This study provides the first direct evidence to confirm that pseudotumour is associated with increased wear at the MoM articulation. Furthermore, edge-loading with the loss of fluid film lubrication may be the dominant wear generation mechanism in patients with pseudotumour.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 220 - 220
1 May 2011
Thomas G Simpson D Gill H McLardy-Smith P Murray D Glyn-Jones S
Full Access

Introduction: The use of second generation highly cross-linked polyethylene (HXLPE) is now commonplace for total hip arthroplasty, however there is no long-term data to support its use. Hip simulator studies suggest that the wear rate of HXLPE is ten times less than conventional polyethylene (UHMWPE). The outcomes of hip simulator studies are not always reproducible in vivo. Long term clinical data is required, as there is emerging clinical data, which suggests that some types of second generation HXLPE may have increased wear after 5 years.

Method: A prospective double blind randomised control trial was conducted using Radiostereometric analysis (RSA). Fifty-four subjects were randomised to receive hip replacements with either UHMWPE liners or HXLPE liners. All subjects received a cemented CPT stem and uncemented Trilogy acetabular component (Zimmer, Warsaw, IN, USA). The 3D penetration of the head into the socket was determined to a minimum of 7 years.

Results: The total liner penetration was significantly different at 7 years (p=0.01) with values of 0.33mm (SD 0.17mm) for the HXLPE group and 0.51mm (SD 0.14mm) for the UHMWPE group. The steady state wear rate from 1 year onwards was significantly lower for HXLPE (0.003 mm/yr, SD 0.04 mm/yr) than for UHMWPE (0.03 mm/yr, SD 0.03 mm/yr) (p=0.01). The direction of wear was in the antero-medial direction in both groups.

Conclusion: We have previously demonstrated that the penetration in the first year is creep-dominated, from one year onwards the majority of penetration is due to wear. The wear rate of this second generation HXLPE approaches that of metal on metal bearings. Second-generation HXLPE may have the potential to reduce the risk of revision surgery, due to wear debris induced osteolysis.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 221 - 221
1 May 2011
Glyn-Jones S Roques A Esposito C Gill H Walter W Tuke M Murray D
Full Access

Introduction: Metal on metal hip resurfacing arthroplasty-induced pseudotumours are a serious complication, which occur in 4% of patients who undergo this procedure. The aim of this study was to measure the 3D in vivo wear on the surface of resurfacing components revised for pseudotumour, compared to a control group.

Method: Thirty-nine hip resurfacing implants were examined; these were sourced from our institutions prosthesis retrieval bank. They were divided into two groups; 22 patients with a clinical and histopathological diagnosis of pseudotumour and 17 controls. Patient demographics and time to revision were known. Three dimensional contactless metrology (Redlux™ Ltd) was used to scan the surface of the femoral and acetabular components, to a resolution of 20 nanometers. The location, depth and area of the wear scar was determined for each component. Volumetric wear was determined, along with the presence of absence of edge-loading. A separate blinded analysis to determine the presence of absence of impingement was performed by one of the authors. ANOVA was used to test for differences in wear and Fishers Exact test was used to compare the incidence of edge-loading between the groups.

Results: The volumetric wear rate for femoral component of the pseudotumour group was 4.7mm3/yr (SD3.5) and 1.7 mm3/yr (SD1.5) for the control group (p=0.03). In the pseudotumour group, the volumetric wear rate of the acetabular component was 3.5 mm3/yr (SD3.6) compared to 0.02 mm3/yr (SD0.07) for the control group (p=0.01). Edge-loading was detected in 74% of acetabular components in the pseudotumour group and 22% of those in the control group (p=0.01). Anterior or posterior edge-loading, consistent with impingement was present on the femoral components of 73% of patients in the pseudotumour group and 22% in the control group (p=0.01).

Discussion: This work demonstrates that implants revised for pseudotumour have significantly higher volumetric wear rates than controls. They also have a significantly higher incidence of edge-loading and impingement than controls. Edge-loading significantly increases wear. We suggest that pseudotumours are caused by high concentrations of metal wear debris, which have been shown to have a toxic effect on osteocytes and macrophages. This is the one of the first studies to demonstrate a clear link between pseudotumours and increased bearing surface wear. It is also the first to demonstrate that edge-loading, due to impingement, occurs in a significant number of patients who develop this condition. Improved implantation techniques and resurfacing designs may help avoid this serious complication of hip resurfacing.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_II | Pages 172 - 172
1 May 2011
Gill H Grammatopoulos G Pandit H Glyn-Jones S Whitwell D Mclardy-Smith P Taylor A Gundle R Murray D
Full Access

Introduction: Metal-on-metal hip resurfacing arthroplasty (MoMHRA) has gained popularity as an alternative to THR for younger patients with osteoarthritis. A growing concern has been the association of MoMHRA with the development of inflammatory pseudotumours (IP), especially in women. These have been linked to metal-on-metal wear, which can be related to metal ion concentrations. Elevated metal wear debris levels may result from impingement, rim contact and edge loading. Head-neck ratio (HNR) is a predetermining factor for range of movement and impingement. Neck thinning is a recognised phenomenon post-MoMHRA and we have found an association of IP with increased neck thinning based on a case control study. Our aims were to identify HNR changes a hip undergoes when resurfaced and at follow up; and whether greater neck thinning at follow-up could be associated with the presence of elevated metal ions.

Methods: A cohort of 91 patients (57M:34F) with unilateral MoMHRAs were included in this study. Blood tests were obtained at a mean follow up of 3.9 years (range 1.7–7 years) and serum (Co:Cr) ion levels were measured (ICPMS). High metal ion concentrations were defined as Co> 4.1ppb and Cr> 5.2ppb. For all patients, head-neck ratio (HNR) was measured on plain anterio-posterior pelvic radiographs pre-operatively, immediately post-operatively and at follow-up.

Results: Female patients had significantly bigger HNR pre-op (mean=1.35, range:1.22–1.64) compared to males(mean=1.22, range:1.05–1.38) (p< 0.01). Immediately post-op, female HNRs (mean: 1.26, range: 1.14–1.34) were not different to male patients(mean=1.24, range=1.11–1.38) (p=0.11). At follow-up HNR was once again significantly bigger (p< 0.01) in females (mean=1.35, range: 1.21–1.49), compare to males (mean=1.27, range:1.11–1.38). HNR alterations with operation (p=0.00) and at follow-up (p< 0.01) were significantly bigger in female patients. Furthermore, there was a significant correlation between high ion levels and HNR change at follow-up for both Co (p=0.02) and Cr (p< 0.01).

Conclusion: This study identified gender-specific changes in HNR that resurfaced hips undergo, not previously documented. Female hips have greater HNR pre-operatively, compared to male hips, and appear to be biomechanically disadvantaged when resurfaced. A decrease in HNR with resurfacing could result in impingement and lead to processes, known to be more prevalent in females, such as neck thinning, increased wear and IP development. In addition, we highlight a correlation between high ion levels and greater neck thinning at follow-up. Increased neck thinning in symptomatic MoMHRA hips could be secondary to increased wear and should be investigated further radiologically for the presence of IP.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 26 - 26
1 Jan 2011
Gulati A Glyn-Jones S Simpson D Palan J Beard D Gill H McLardy-Smith P Gundle R Murray D
Full Access

Roentgen Stereophotogrammetric Analysis (RSA) can predict long-term outcome of prostheses by measuring migration over time. The Exeter femoral stem is a double-tapered highly polished implant and has been shown to subside within the cement mantle in 2 year RSA studies. It has a proven track record in terms of long-term survivorship and low revision rates. Several studies have demonstrated excellent clinical outcomes following its implantation but this is the first study to assess stem migration at 10 years, using RSA.

This is a single-centre study involving 20 patients (mean age: 63 years, SD=7) undergoing primary total hip replacement for degenerative osteoarthritis using the lateral (Hardinge) approach. RSA radiographs were taken with the patient bearing full weight post-operatively, at 3, 6, 12 months and at 2, 5 and 10 years follow-up. The three-dimensional migration of the Exeter femoral stem was determined.

The mean Oxford Hip Score at 10 years was 43.4 (SD=4.6) and there were no revisions. The stems subsided and rotated internally during a 10-year period. The mean migrations of the head and tip of the femoral stem in all three anatomic directions (antero-posterior, medio-lateral & supero-distal) were 0.69 mm posterior, 0.04 mm lateral and 1.67 mm distal for the head and 0.20 mm anterior, 0.02 mm lateral and 1.23 mm distal for the tip. The total migration at 10 years was 1.81 mm for the head and 1.25 mm for the tip.

The Exeter femoral stem exhibits migration which is a complex combination of translation and rotation in three dimensions. Comparing our 10 year with our previous 2 year migration results, the Exeter stems show continued, but slow distal migration and internal rotation. The subsidence continues to compress the cement and bone-cement interface which maintains secure fixation in the long term.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 390
1 Jul 2010
Kwon Y Ostlere S Mclardy-Smith P Gundle R Whitwell D Gibbons C Taylor A Pandit H Glyn-Jones S Athanasou N Beard D Gill H Murray D
Full Access

Introduction: Despite the satisfactory short-term implant survivorship of MoM hip resurfacing arthroplasty, symptomatic abnormal periprosthetic soft-tissue masses relating to the hip joint, ‘pseudotumours’, are being increasingly reported. These were found be locally destructive, requiring revision surgery in 75% of patients. Asymptomatic pseudotumours have not been previously investigated.

Methods: The aims were: (1) to investigate the prevalence of asymptomatic pseudotumours; and (2) to investigate their potential association with the level of metal ions. A total of 160 hips in 123 patients with a mean age 56 years (range 33–73) were evaluated at a mean follow-up of 61 months (range 13–88). Radiographs and OHS were assessed. Patients with a cystic or solid mass detected on the ultrasound/MRI had an aspiration or biopsy performed. Cobalt and chromium levels were analysed using Inductively-Coupled Plasma Spectrometer.

Results: Pseudotumours were found in 6 patients (5F: 1M). In 80% of bilateral cases, it was found in both sides. Histological examination showed extensive necrosis and diffuse lymphocyte infiltration. The presence of pseudotumour was associated with higher serum cobalt (9.2 μg/L vs. 1.9μg/L, p< 0.001) and chromium levels (12.0μg/L vs. 2.1μg/L, p< 0.001); higher hip aspirate cobalt (1182 μg/L vs. 86.2μg/L, p=0.003) and chromium levels (883μg/L vs. 114.8μg/L, p=0.006); and with inferior OHS (23 vs. 14 p=0.08).

Discussion: The prevalence of asymptomatic pseudotumour (5%) was higher than previously reported for the symptomatic pseudotumours (1%). There was a sixfold elevation of serum and a twelve-fold elevation of hip aspirate levels of cobalt and chromium in patients with pseudotumours. This suggests that pseudotumours may be a biological consequence of the large amount of metal debris generated in vivo. The association between pseudotumour and elevated metal ion levels might theoretically be explained by either systemic hypersensitivity responses to metal ions or local cytotoxic effects due to a high level of metal ions.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 406 - 406
1 Sep 2009
Pandit H Glyn-Jones S Gundle R Whitwell D Gibbons C Ostlere S Athanasou N Gill H McLardy-Smith P Murray D
Full Access

Introduction: We report on a group of 20 metal-on-metal resurfaced hips (17 patients) presenting with a soft tissue mass associated with various symptoms; these masses we termed pseudotumours. All patients underwent plane radiography; CT, MRI and ultrasound investigations were also performed for some patients. Where samples were available histology was performed.

Methods: All patients in this series were female. Presentation was variable; the most common symptom was pain or discomfort in the hip region. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. Fourteen of the 20 cases (70%) have so far required revision to a conventional hip replacement and their symptoms have either settled completely or improved substantially since the revision surgery. Two of the three bilateral cases have asymptomatic pseudotumours on the opposite side.

Conclusions: We estimate that about 1% of patients develop a pseudotumour in the first five postoperative years after a hip resurfacing. The cause of these pseudotumours is unknown and is probably multi-factorial, further work is required to define this; they may be manifestations of a metal sensitivity response. We are concerned that with time the incidence of these pseudotumours will increase.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 25 - 25
1 Mar 2009
Glyn-Jones S Beard D Murray D Gill H
Full Access

Introduction: Interest in hip resurfacing has recently been renewed by the introduction of metal-on-metal designs; it is being increasingly used for young patients, with over 30,000 implanted worldwide. The 5 year clinical results appear promising, but there are no long term data available. Radiostereometry (RSA) measures of implant migration have been able to predict implant failure; specifically large and continuous migration predicts aseptic loosening. We present the results of a five year RSA study examining the migration of the Birmingham Hip Resurfacing (BHR).

Methods: Twenty-four subjects with primary OA were implanted with the BHR device and with bone markers for RSA. RSA measurements were taken at 3, 6, 12, 24 and 60 months. The migration of the head and the tip of femoral component were measured in 3D.

Results: Preliminary analysis showed that the total 3D migration of the head and tip over five years was 0.32mm and 0.23mm respectively.

Discussion: A distal migration of more than 0.4mm over 2 years increases the likelihood of failure in conventional stems. The total migration of the BHR was approximately 0.3mm over a five year period, significantly less than cemented THR devices. The device is stable and this is promising for long-term survival.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 398 - 398
1 Jul 2008
Glyn-Jones S Pandit H Whitwell D Athanasou N Gibbons M
Full Access

Purpose of study: We report the results of a prospective case series of 10 patients who developed tumour-like masses following resurfacing arthroplasty

Method: Ten subjects were referred to the tumour service at the Nuffield Orthopaedic Centre with symptomatic masses around the hip, all had previously received a resurfacing arthroplasty.

We report the clinical, radiographic and histologic features of these cases.

Results: MRI and ultrasound scanning was preformed, which demonstrated masses with solid and cystic components.

Biopsy was performed and subsequent histological examination revealed a profound plasma-cell lymphocytic response associated with metal wear debris.

There were no infections in this series.

Three subjects required revision surgery.

Conclusion: Over 50,000 resurfacing arthroplasties have been implanted worldwide over the past ten years. Although the early clinical results are encouraging little is known about the long term consequences of large head metal on metal bearing surfaces. Despite this, these devices are being widely marketed and are often implanted in younger patients. Resurfacing arthroplasties are associated with high serum and urine metal ion concentrations, metal particles have also been shown to migrate along the lymphatic system. In addition, there is now evidence that high local metal ion concentrations can induce haempoietic cancers.

This study suggests that resurfacing arthoplasty can also induce a local hypersensitivity reaction in response to metal wear debris. It therefore raises new concerns regarding the long-term safety of this procedure.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 166 - 166
1 Mar 2008
Hauptfleisch J Glyn-Jones S Gill H McLardy-Smith P Murray D
Full Access

The Charnley Elite femoral component was first introduced in 1992 as a new design variant of the original Charnley femoral component (De Puy, Leeds, UK) with modified neck and stem geometry. The original component had undergone few changes in nearly forty years and has excellent long-term results. Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA)1. Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is indicated as stems of this type are still being implanted.

One hundred Charnley Elite stems, implanted in our centre between 1994 and 1997 were included in a prospective, cross-sectional follow-up study. Outcome measures include validated clinical scores (Charnley hip score, Harris hip score and Oxford hip score) and radiological scores (Gruen classification) as well as revision rates over the past 10 years.

The clinical follow-up supports the RSA predictions of early failure of the Charnley Elite femoral stem.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 368 - 368
1 Oct 2006
Gill H Polgar K Glyn-Jones S McLardy-Smith P Murray D
Full Access

Introduction: The design philosophy of polished tapered THR stems, such as the Exeter, intend for them to migrate distally within the cement mantle. In addition it is likely that micromotion occurs as a result of functional activity. The pattern of induced stresses will be a function of stem geometry & surface finish, as well as applied loading. Aim: To investigate the stresses induced in the cement mantle of a polished tapered THR stem during functional activity.

Method: Using Roentgen Stereophotogrammetric Analysis (RSA) dynamically induced micro-motion (DIMM) was measured in 21 patients implanted with Exeter stems. DIMM was measured as the difference in stem position in going from double to single leg stance on the operated limb. All subjects were measured 3 months post-operatively. A finite element (FE) model of the femur, including all muscles was used to investigate the stress distribution within the cement; contact was modelled with sliding elements allowing separation. The model was validated by comparison to the DIMM measurements.

Results: The Exeter stem demonstrated significant DIMM(p < 0.017), the average motions are given in the table below. The FE model, with sliding contacts was able to predict similar distal migration of the head. The peak minimum principal stress in the mantle was approx 33MPa and occurred in the proximal medial region. Movements occurred at the stem/cement interface.

Discussion and Conclusion: It is possible to measure DIMM in the Exeter stem and combining this with FE modelling the mechanism of stress transfer between the stem and mantle can be investigated in a manner that can be validated.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 45 - 45
1 Mar 2006
Glyn-Jones S Gill R McLardy-Smith P Murray D
Full Access

Introduction Polyethylene wear debris is an important cause of failure in cemented total hip arthroplasty. As a result of the biological response to debris at the bone-cement interface, osteolysis and subsequent failure occurs in both femoral and acetabular components. Most acetabular components and liners are made of ultra high molecular weight polyethylene (UHMWPE). Cross-linking UHMWPE has been shown to significantly reduce abrasive wear in hip simulator studies. The wear rates measured in vitro do not always correlate with the wear rates measured in clinical studies[1]. Some new polyethylenes have shown catastrophic wear in clinical studies despite encouraging hip simulator study results[2]. The aim of this study was to compare the wear of standard UHMWPE to that of cross-linked UHMWPE (Longevity, Zimmer, Warsaw, USA)

Patients and Methods This was a prospective, double blind, randomised control trial. 50 subjects were recruited, all of whom received the cemented CPT stem and uncemented Trilogy liner (Zimmer, UK). Subjects were randomised to receive either a standard Trilogy liner or a Longevity liner at the time of operation. Both liners are identical in appearance. All liners were of a neutral configuration. RSA was used to measure linear wear. This was calculated by measuring the distance between the centre of the femoral head and the centre of the acetabular liner. The preliminary results of the study are presented.

Results Both groups underwent significant wear over two years. The two year linear wear of the cross-linked UHMWPE was 0.3mm (+/− 0.06mm, p< 0.001). The two year linear wear of the standard UHMWPE was 0.39mm (+/− 0.04mm, p< 0.001). No significant difference existed between the two groups (p=0.24). Both cohorts had around 0.15 to 0.2 mm of measured wear per year. Cross-linked UHMWPE therefore underwent less wear than standard UHMWPE at two years, however this difference was not statistically significant.

Discussion This study suggests that Longevity UHMWPE has similar wear properties to standard UHMWPE in the first two years following implantation. This does not correlate with in vitro hip simulator studies of Longevity polyethylene, which show a significantly lower wear rate than standard UHMWPE. It suggests that hip simulator studies may be of little value in predicting in vivo wear rates and that all new types of polyethylene should be evaluated clinically and radiologically prior to general release. Whether both cohorts continue to wear at similar rates will only be revealed through continued observation.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 73 - 73
1 Mar 2006
Hauptfleisch J Glyn-Jones S Beard D Gill H McLardy-Smith P Murray D
Full Access

Introduction: The Charnley Elite femoral component was first introduced in 1992 as a new design variant of the original Charnley femoral component (De Puy, Leeds, UK) with modified neck and stem geometry. The original component had undergone few changes in nearly forty years and has excellent long-term results.

Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA). Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is required as stems of this type are still being implanted.

Materials and method: One hundred Charnley Elite stems, implanted in our centre between 1994 and 1997 were included in a prospective, cross-sectional follow-up study. Outcome measures include validated clinical scores (Charnley hip score, Harris hip score and Oxford hip score) and radiological scores (Gruen classification) as well as revision rates over the past 10 years.

Results: The preliminary analysis results are given. The mean time to follow-up was 8.28 years. 20 patients have died due to causes unrelated to their operations. 10 patients had stem revisions: 9 for aseptic loosening and 1 for a peri-prosthetic fracture. This indicates a significant 10% failure rate of the prosthesis in less than 10 years.

Preliminary clinical scores in the patients who had not undergone any subsequent surgery were adequate (Oxford Hip Score mean average of 23.9).

Thirteen percent of radiographs analysed had evidence of loosening, giving an overall loosening rate of 14% at 8 years.

Discussion and conclusion: The clinical follow-up supports the RSA predictions of early failure of the Charnley Elite femoral stem.