Advertisement for orthosearch.org.uk
Results 1 - 20 of 28
Results per page:
Bone & Joint Open
Vol. 5, Issue 1 | Pages 37 - 45
19 Jan 2024
Alm CE Karlsten A Madsen JE Nordsletten L Brattgjerd JE Pripp AH Frihagen F Röhrl SM

Aims. Despite limited clinical scientific backing, an additional trochanteric stabilizing plate (TSP) has been advocated when treating unstable trochanteric fractures with a sliding hip screw (SHS). We aimed to explore whether the TSP would result in less post operative fracture motion, compared to SHS alone. Methods. Overall, 31 patients with AO/OTA 31-A2 trochanteric fractures were randomized to either a SHS alone or a SHS with an additional TSP. To compare postoperative fracture motion, radiostereometric analysis (RSA) was performed before and after weightbearing, and then at four, eight, 12, 26, and 52 weeks. With the “after weightbearing” images as baseline, we calculated translations and rotations, including shortening and medialization of the femoral shaft. Results. Similar migration profiles were observed in all directions during the course of healing. At one year, eight patients in the SHS group and 12 patients in the TSP group were available for analysis, finding a clinically non-relevant, and statistically non-significant, difference in total translation of 1 mm (95% confidence interval -4.7 to 2.9) in favour of the TSP group. In line with the migration data, no significant differences in clinical outcomes were found. Conclusion. The TSP did not influence the course of healing or postoperative fracture motion compared to SHS alone. Based on our results, routine use of the TSP in AO/OTA 31-A2 trochanteric fractures cannot be recommended. The TSP has been shown, in biomechanical studies, to increase stability in sliding hip screw constructs in both unstable and intermediate stable trochanteric fractures, but the clinical evidence is limited. This study showed no advantage of the TSP in unstable (AO 31-A2) fractures in elderly patients when fracture movement was evaluated with radiostereometric analysis. Cite this article: Bone Jt Open 2024;5(1):37–45


Bone & Joint Research
Vol. 6, Issue 4 | Pages 216 - 223
1 Apr 2017
Ang BFH Chen JY Yew AKS Chua SK Chou SM Chia SL Koh JSB Howe TS

Objectives. External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). Methods. A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness. Results. The mean axial stiffness was very similar for UEF (528 N/mm) and ESS-LCP (525 N/mm), while it was slightly lower for ET-LCP (469 N/mm). One-way analysis of variance (ANOVA) testing in all three groups demonstrated no significant difference (F(2,12) = 2.057, p = 0.171). There was a significant difference in mean torsional stiffness between the UEF (0.512 Nm/degree), the ESS-LCP (0.686 Nm/degree) and the ET-LCP (0.639 Nm/degree), as determined by one-way ANOVA (F(2,12) = 6.204, p = 0.014). A Tukey post hoc test revealed that the torsional stiffness of the ESS-LCP was statistically higher than that of the UEF by 0.174 Nm/degree (p = 0.013). No catastrophic failures were observed. Conclusion. Using the LCP as an external fixator may provide a viable and attractive alternative to the traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness. Cite this article: B. F. H. Ang, J. Y. Chen, A. K. S. Yew, S. K. Chua, S. M. Chou, S. L. Chia, J. S. B. Koh, T. S. Howe. Externalised locking compression plate as an alternative to the unilateral external fixator: a biomechanical comparative study of axial and torsional stiffness. Bone Joint Res 2017;6:216–223. DOI: 10.1302/2046-3758.64.2000470


Bone & Joint Open
Vol. 5, Issue 2 | Pages 147 - 153
19 Feb 2024
Hazra S Saha N Mallick SK Saraf A Kumar S Ghosh S Chandra M

Aims

Posterior column plating through the single anterior approach reduces the morbidity in acetabular fractures that require stabilization of both the columns. The aim of this study is to assess the effectiveness of posterior column plating through the anterior intrapelvic approach (AIP) in the management of acetabular fractures.

Methods

We retrospectively reviewed the data from R G Kar Medical College, Kolkata, India, from June 2018 to April 2023. Overall, there were 34 acetabulum fractures involving both columns managed by medial buttress plating of posterior column. The posterior column of the acetabular fracture was fixed through the AIP approach with buttress plate on medial surface of posterior column. Mean follow-up was 25 months (13 to 58). Accuracy of reduction and effectiveness of this technique were measured by assessing the Merle d’Aubigné score and Matta’s radiological grading at one year and at latest follow-up.


Bone & Joint Open
Vol. 3, Issue 2 | Pages 165 - 172
21 Feb 2022
Kuwahara Y Takegami Y Tokutake K Yamada Y Komaki K Ichikawa T Imagama S

Aims

Postoperative malalignment of the femur is one of the main complications in distal femur fractures. Few papers have investigated the impact of intraoperative malalignment on postoperative function and bone healing outcomes. The aim of this study was to investigate how intraoperative fracture malalignment affects postoperative bone healing and functional outcomes.

Methods

In total, 140 patients were retrospectively identified from data obtained from a database of hospitals participating in a trauma research group. We divided them into two groups according to coronal plane malalignment of more than 5°: 108 had satisfactory fracture alignment (< 5°, group S), and 32 had unsatisfactory alignment (> 5°, group U). Patient characteristics and injury-related factors were recorded. We compared the rates of nonunion, implant failure, and reoperation as healing outcomes and Knee Society Score (KSS) at three, six, and 12 months as functional outcomes. We also performed a sub-analysis to assess the effect of fracture malalignment by plates and nails on postoperative outcomes.


Aims

Our objective was to conduct a systematic review and meta-analysis, to establish whether differences arise in clinical outcomes between autologous and synthetic bone grafts in the operative management of tibial plateau fractures.

Methods

A structured search of MEDLINE, EMBASE, the online archives of Bone & Joint Publishing, and CENTRAL databases from inception until 28 July 2021 was performed. Randomized, controlled, clinical trials that compared autologous and synthetic bone grafts in tibial plateau fractures were included. Preclinical studies, clinical studies in paediatric patients, pathological fractures, fracture nonunion, or chondral defects were excluded. Outcome data were assessed using the Risk of Bias 2 (ROB2) framework and synthesized in random-effect meta-analysis. The Preferred Reported Items for Systematic Review and Meta-Analyses guidance was followed throughout.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 294 - 298
1 Feb 2021
Hadeed MM Prakash H Yarboro SR Weiss DB

Aims

The aim of this study was to determine the immediate post-fixation stability of a distal tibial fracture fixed with an intramedullary nail using a biomechanical model. This was used as a surrogate for immediate weight-bearing postoperatively. The goal was to help inform postoperative protocols.

Methods

A biomechanical model of distal metaphyseal tibial fractures was created using a fourth-generation composite bone model. Three fracture patterns were tested: spiral, oblique, and multifragmented. Each fracture extended to within 4 cm to 5 cm of the plafond. The models were nearly-anatomically reduced and stabilized with an intramedullary nail and three distal locking screws. Cyclic loading was performed to simulate normal gait. Loading was completed in compression at 3,000 N at 1 Hz for a total of 70,000 cycles. Displacement (shortening, coronal and sagittal angulation) was measured at regular intervals.


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 530 - 538
1 Apr 2020
Rollick NC Gadinsky NE Klinger CE Kubik JF Dyke JP Helfet DL Wellman DS

Aims

Dual plating of distal femoral fractures with medial and lateral implants has been performed to improve construct mechanics and alignment, in cases where isolated lateral plating would be insufficient. This may potentially compromise vascularity, paradoxically impairing healing. This study investigates effects of single versus dual plating on distal femoral vascularity.

Methods

A total of eight cadaveric lower limb pairs were arbitrarily assigned to either 1) isolated lateral plating, or 2) lateral and medial plating of the distal femur, with four specimens per group. Contralateral limbs served as matched controls. Pre- and post-contrast MRI was performed to quantify signal intensity enhancement in the distal femur. Further evaluation of intraosseous vascularity was done with barium sulphate infusion with CT scan imaging. Specimens were then injected with latex medium and dissection was completed to assess extraosseous vasculature.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 8 | Pages 1155 - 1160
1 Nov 2003
Yang KH Byun YS

Comminuted and displaced fractures of the inferior pole of the patella are not easy to reduce and it is difficult to fix the fragments soundly enough to allow early movement of the knee. We have evaluated the clinical effectiveness of the separate vertical wiring technique in acute comminuted fractures of the inferior pole of the patella. A biomechanical study was also performed using ten pairs of embalmed cadaver knees. A four-part fracture was made on the inferior pole of the patella and fixed by two separate vertical wires on one side and two pull-out sutures after partial patellectomy on the other. The ultimate load to failure in the first group was significantly higher than in the second (250.1± 109.7 N v 69.7 ± 18.9 N, p< 0.002), as was the stiffness (279.9 ± 76.4 N/mm v 23.2 ± 11.4 N/mm, p< 0.001). The separate wire technique was used in 25 patients with comminuted fractures of the inferior pole of the patella who were followed up for a mean period of 22 months (10 to 50). All the fractures healed at a mean of seven weeks (6 to 10). No breakage of a wire or infection occurred. The mean grading at the final follow-up was 29.5 points (27 to 30) using the Böstman method. This technique preserved the length of the patella, fixed the comminuted fragments of the inferior pole and avoided long-term immobilisation of the knee


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 973 - 983
1 Jul 2018
Schmal H Froberg L S. Larsen M Südkamp NP Pohlemann T Aghayev E Goodwin Burri K

Aims

The best method of treating unstable pelvic fractures that involve the obturator ring is still a matter for debate. This study compared three methods of treatment: nonoperative, isolated posterior fixation and combined anteroposterior stabilization.

Patients and Methods

The study used data from the German Pelvic Trauma Registry and compared patients undergoing conservative management (n = 2394), surgical treatment (n = 1345) and transpubic surgery, including posterior stabilization (n = 730) with isolated posterior osteosynthesis (n = 405) in non-complex Type B and C fractures that only involved the obturator ring anteriorly. Calculated odds ratios were adjusted for potential confounders. Outcome criteria were intraoperative and general short-term complications, the incidence of nerve injuries, and mortality.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 95 - 100
1 Jan 2018
Evers J Fischer M Zderic I Wähnert D Richards RG Gueorguiev B Raschke MJ Ochman S

Aims

The aim of this study was to investigate the effect of a posterior malleolar fragment (PMF), with < 25% ankle joint surface, on pressure distribution and joint-stability. There is still little scientific evidence available to advise on the size of PMF, which is essential to provide treatment. To date, studies show inconsistent results and recommendations for surgical treatment date from 1940.

Materials and Methods

A total of 12 cadaveric ankles were assigned to two study groups. A trimalleolar fracture was created, followed by open reduction and internal fixation. PMF was fixed in Group I, but not in Group II. Intra-articular pressure was measured and cyclic loading was performed.


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 128 - 133
1 Jan 2017
Socci AR Casemyr NE Leslie MP Baumgaertner MR

Aims

The aim of this paper is to review the evidence relating to the anatomy of the proximal femur, the geometry of the fracture and the characteristics of implants and methods of fixation of intertrochanteric fractures of the hip.

Materials and Methods

Relevant papers were identified from appropriate clinical databases and a narrative review was undertaken.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 114 - 121
1 Jan 2014
Pekmezci M McDonald E Buckley J Kandemir U

We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (sd 33 900) vs 12 800 cycles (sd 6100); p = 0.007) and mean axial stiffness (220 N/mm (sd 80) vs 70 N/mm (sd 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (sd 0.9) vs 5.1 Nm/° (sd 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (sd 25 600) vs 52 400 cycles (sd 22 500); p = 0.04).

The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients.

Cite this article: Bone Joint J 2014;96-B:114–21.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1378 - 1384
1 Oct 2014
Weiser L Korecki MA Sellenschloh K Fensky F Püschel K Morlock MM Rueger JM Lehmann W

It is becoming increasingly common for a patient to have ipsilateral hip and knee replacements. The inter-prosthetic (IP) distance, the distance between the tips of hip and knee prostheses, has been thought to be associated with an increased risk of IP fracture. Small gap distances are generally assumed to act as stress risers, although there is no real biomechanical evidence to support this.

The purpose of this study was to evaluate the influence of IP distance, cortical thickness and bone mineral density on the likelihood of an IP femoral fracture.

A total of 18 human femur specimens were randomised into three groups by bone density and cortical thickness. For each group, a defined IP distance of 35 mm, 80 mm or 160 mm was created by choosing the appropriate lengths of component. The maximum fracture strength was determined using a four-point bending test.

The fracture force of all three groups was similar (p = 0.498). There was a highly significant correlation between the cortical area and the fracture strength (r = 0.804, p <  0.001), whereas bone density showed no influence.

This study suggests that the IP distance has little influence on fracture strength in IP femoral fractures: the thickness of the cortex seems to be the decisive factor.

Cite this article: Bone Joint J 2014;96-B:1378–84.


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 398 - 404
1 Mar 2015
Fang C Lau TW Wong TM Lee HL Leung F

The spiral blade modification of the Dynamic Hip Screw (DHS) was designed for superior biomechanical fixation in the osteoporotic femoral head. Our objective was to compare clinical outcomes and in particular the incidence of loss of fixation.

In a series of 197 consecutive patients over the age of 50 years treated with DHS-blades (blades) and 242 patients treated with conventional DHS (screw) for AO/OTA 31.A1 or A2 intertrochanteric fractures were identified from a prospectively compiled database in a level 1 trauma centre. Using propensity score matching, two groups comprising 177 matched patients were compiled and radiological and clinical outcomes compared. In each group there were 66 males and 111 females. Mean age was 83.6 (54 to 100) for the conventional DHS group and 83.8 (52 to 101) for the blade group.

Loss of fixation occurred in two blades and 13 DHSs. None of the blades had observable migration while nine DHSs had gross migration within the femoral head before the fracture healed. There were two versus four implant cut-outs respectively and one side plate pull-out in the DHS group. There was no significant difference in mortality and eventual walking ability between the groups. Multiple logistic regression suggested that poor reduction (odds ratio (OR) 11.49, 95% confidence intervals (CI) 1.45 to 90.9, p = 0.021) and fixation by DHS (OR 15.85, 95%CI 2.50 to 100.3, p = 0.003) were independent predictors of loss of fixation.

The spiral blade design may decrease the risk of implant migration in the femoral head but does not reduce the incidence of cut-out and reoperation. Reduction of the fracture is of paramount importance since poor reduction was an independent predictor for loss of fixation regardless of the implant being used.

Cite this article: Bone Joint J 2015;97-B:398–404.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1535 - 1539
1 Nov 2014
Tonne BM Kempton LB Lack WD Karunakar MA

The purpose of this study was to describe the radiological characteristics of a previously unreported finding: posterior iliac offset at the sacroiliac joint and to assess its association with pelvic instability as measured by initial displacement and early implant loosening or failure. Radiographs from 42 consecutive patients with a mean age of 42 years (18 to 77; 38 men, four women) and mean follow-up of 38 months (3 to 96) with Anteroposterior Compression II injuries, were retrospectively reviewed. Standardised measurements were recorded for the extent of any diastasis of the pubic symphysis, widening of the sacroiliac joint, static vertical ramus offset and a novel measurement (posterior offset of the ilium at the sacroiliac joint identified on axial CT scan). Pelvic fractures with posterior iliac offset exhibited greater levels of initial displacement of the anterior pelvis (anterior sacroiliac widening, pubic symphysis diastasis and static vertical ramus offset, p < 0.001,0.034 and 0.028, respectively). Pelvic fractures with posterior ilium offset also demonstrated higher rates of implant loosening regardless of fixation method (p = 0.05). Posterior offset of the ilium was found to be a reliable and reproducible measurement with substantial inter-observer agreement (kappa = 0.70). Posterior offset of the ilium on axial CT scan is associated with greater levels of initial pelvic displacement and early implant loosening.

Cite this article: Bone Joint J 2014;96-B:1535–9.


Bone & Joint Research
Vol. 2, Issue 5 | Pages 79 - 83
1 May 2013
Goffin JM Pankaj P Simpson AHRW Seil R Gerich TG

Objectives

Because of the contradictory body of evidence related to the potential benefits of helical blades in trochanteric fracture fixation, we studied the effect of bone compaction resulting from the insertion of a proximal femoral nail anti-rotation (PFNA).

Methods

We developed a subject-specific computational model of a trochanteric fracture (31-A2 in the AO classification) with lack of medial support and varied the bone density to account for variability in bone properties among hip fracture patients.


Bone & Joint Research
Vol. 1, Issue 6 | Pages 118 - 124
1 Jun 2012
Grawe B Le T Williamson S Archdeacon A Zardiackas L

Objectives

We aimed to further evaluate the biomechanical characteristics of two locking screws versus three standard bicortical screws in synthetic models of normal and osteoporotic bone.

Methods

Synthetic tubular bone models representing normal bone density and osteoporotic bone density were used. Artificial fracture gaps of 1 cm were created in each specimen before fixation with one of two constructs: 1) two locking screws using a five-hole locking compression plate (LCP) plate; or 2) three non-locking screws with a seven-hole LCP plate across each side of the fracture gap. The stiffness, maximum displacement, mode of failure and number of cycles to failure were recorded under progressive cyclic torsional and eccentric axial loading.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1406 - 1409
1 Oct 2013
Wähnert D Lange JH Schulze M Gehweiler D Kösters C Raschke MJ

The augmentation of fixation with bone cement is increasingly being used in the treatment of severe osteoporotic fractures. We investigated the influence of bone quality on the mechanics of augmentation of plate fixation in a distal femoral fracture model (AO 33 A3 type). Eight osteoporotic and eight non-osteoporotic femoral models were randomly assigned to either an augmented or a non-augmented group. Fixation was performed using a locking compression plate. In the augmented group additionally 1 ml of bone cement was injected into the screw hole before insertion of the screw. Biomechanical testing was performed in axial sinusoidal loading. Augmentation significantly reduced the cut-out distance in the osteoporotic models by about 67% (non-augmented mean 0.30 mm (sd 0.08) vs augmented 0.13 mm (sd 0.06); p = 0.017). There was no statistical reduction in this distance following augmentation in the non-osteoporotic models (non-augmented mean 0.15 mm (sd 0.02) vs augmented 0.15 mm (sd 0.07); p = 0.915). In the osteoporotic models, augmentation significantly increased stability (p = 0.017).

Cite this article: Bone Joint J 2013;95-B:1406–9.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 983 - 987
1 Jul 2013
Soliman O Koptan W Zarad A

In Neer type II (Robinson type 3B) fractures of the distal clavicle the medial fragment is detached from the coracoclavicular ligaments and displaced upwards, whereas the lateral fragment, which is usually small, maintains its position. Several fixation techniques have been suggested to treat this fracture. The aim of this study was to assess the outcome of patients with type II distal clavicle fractures treated with coracoclavicular suture fixation using three loops of Ethibond. This prospective study included 14 patients with Neer type II fractures treated with open reduction and coracoclavicular fixation. Ethibond sutures were passed under the coracoid and around the clavicle (UCAC loop) without making any drill holes in the proximal or distal fragments. There were 11 men and three women with a mean age of 34.57 years (29 to 41). Patients were followed for a mean of 24.64 months (14 to 31) and evaluated radiologically and clinically using the Constant score. Fracture union was obtained in 13 patients at a mean of 18.23 weeks (13 to 23) and the mean Constant score was 96.07 (91 to 100). One patient developed an asymptomatic fibrous nonunion at one year. This study suggests that open reduction and internal fixation of unstable distal clavicle fractures using UCAC loops can provide rigid fixation and lead to bony union. This technique avoids using metal hardware, preserves the acromioclavicular joint and provides adequate stability with excellent results.

Cite this article: Bone Joint J 2013;95-B:983–7.


Bone & Joint Research
Vol. 1, Issue 6 | Pages 104 - 110
1 Jun 2012
Swinteck BJ Phan DL Jani J Owen JR Wayne JS Mounasamy V

Objectives

The use of two implants to manage concomitant ipsilateral femoral shaft and proximal femoral fractures has been indicated, but no studies address the relationship of dynamic hip screw (DHS) side plate screws and the intramedullary nail where failure might occur after union. This study compares different implant configurations in order to investigate bridging the gap between the distal DHS and tip of the intramedullary nail.

Methods

A total of 29 left synthetic femora were tested in three groups: 1) gapped short nail (GSN); 2) unicortical short nail (USN), differing from GSN by the use of two unicortical bridging screws; and 3) bicortical long nail (BLN), with two angled bicortical and one unicortical bridging screws. With these findings, five matched-pairs of cadaveric femora were tested in two groups: 1) unicortical long nail (ULN), with a longer nail than USN and three bridging unicortical screws; and 2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally rotated 90°/sec until failure.