Aims. In the native hip, the hip capsular ligaments tighten at the limits of range of hip motion and may provide a passive stabilizing force to protect the hip against
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6) in situ. These
implants were obtained from 54 patients (16 men and 38 women) with
a mean age of 67 years (33 to 88) who underwent revision for a variety
of reasons. Posterior
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of
We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm. 3. /yr for the femoral heads, and 0.04 mm. 3. /yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of <
15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm. 3. /yr, compared with 0 mm. 3. /yr for those with an anteversion of ≥15° (15 femoral heads, p <
0.001). Even under
Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by
We investigated the characteristics of patients
who achieved Japanese-style deep flexion (seiza-sitting) after total knee
replacement (TKR) and measured three-dimensional positioning and
the contact positions of the femoral and tibial components. Seiza-sitting
was achieved after surgery by 23 patients (29 knees) of a series
of 463 TKRs in 341 patients. Pre-operatively most of these patients
were capable of seiza-sitting, had a lower body mass index and a favourable
attitude towards the Japanese lifestyle (27 of 29 knees). According
to two-/three-dimensional image registration analysis in the seiza-sitting
position, flexion, varus and internal rotation angles of the tibial
component relative to the femoral component had means of 148° (. sd. 8.0),
1.9° (. sd. 3.2) and 13.4° (. sd. 5.9), respectively.
Femoral surface contact positions tended to be close to the posterior
edge of the tibial polyethylene insert, particularly in the lateral
compartment, but only 8.3% (two of 24) of knees showed femoral subluxation
over the posterior edge. The mean contact positions of the femoral
cam on the tibial post were located 7.8 mm (. sd. 1.5) proximal
to the lowest point of the polyethylene surface and 5.5 mm (. sd. 0.9)
medial to the centre of the post, indicating that the post-cam contact
position translated medially during seiza-sitting, but not proximally.
Collectively, the seiza-sitting position seems safe against component
dislocation, but the risks of posterior
We examined the relationship between the size
of the femoral cam in femoroacetabular impingement (FAI) and acetabular
pathomorphology to establish if pincer impingement exists in patients
with a femoral cam. CT scans of 37 symptomatic impinging hips with a femoral cam
were analysed in a three-dimensional study and were compared with
34 normal hips. The inclination and version of the acetabulum as
well as the acetabular rim angle and the bony acetabular coverage
were calculated. These measurements were correlated with the size
and shape of the femoral cams. While the size of the femoral cam varied characteristically,
the acetabular morphology of the two groups was similar in terms
of version (normal mean 23° (. sd. 7°); cam mean 22° (. sd. 9°)),
inclination (normal mean 57° (. sd. 5°); cam mean 56° (. sd. 5°)),
acetabular coverage (normal mean 41% (. sd. 5%); cam mean
42% (. sd. 4%)) and the mean acetabular rim angle (normal
mean 82° (. sd. 5°); cam mean 83° (. sd. 4°)). We found no correlation between acetabular morphology and the
severity of cam lesion and no evidence of either global or focal
over-coverage to support the diagnosis of ‘mixed’ FAI. The femoral
cam may provoke
Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients). Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 (. sd. 0.10) vs mean 1.30 (. sd. 0.08) p = 0.001). At operation the patients with pseudotumours had a greater reduction in the size of their femoral heads (p = 0.035) and subsequently had greater neck narrowing (mean 10.1% (. sd. 7.2) vs mean 3.8% (. sd. 3.2) p <
0.001). No female patient with a pre-operative HNR ≤ 1.3 developed a pseudotumour. We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and
We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the tibial component. Peak contact stresses at 90°, 120°, and 150° were 24.0, 33.9, and 28.8 MPa, respectively, for the Kirschner; 26.0, 32.4, and 22.1 MPa, respectively, for the Scorpio; and 34.1, 31.5, and 32.5 MPa, respectively, for the NexGen LPS Flex Fixed. With an internally rotated tibia, the contact stress increased significantly with all the fixed-bearing arthroplasties but not with the NexGen LPS Flex Mobile arthroplasty. The post-cam design should be modified in order to provide a larger contact area whilst avoiding any impingement and
The October 2023 Hip & Pelvis Roundup360 looks at: Femoroacetabular impingement syndrome at ten years – how do athletes do?; Venous thromboembolism in patients following total joint replacement: are transfusions to blame?; What changes in pelvic sagittal tilt occur 20 years after total hip arthroplasty?; Can stratified care in hip arthroscopy predict successful and unsuccessful outcomes?; Hip replacement into your nineties; Can large language models help with follow-up?; The most taxing of revisions – proximal femoral replacement for periprosthetic joint infection – what’s the benefit of dual mobility?
Computer-assisted 3D preoperative planning software has the potential to improve postoperative stability in total hip arthroplasty (THA). Commonly, preoperative protocols simulate two functional positions (standing and relaxed sitting) but do not consider other common positions that may increase postoperative impingement and possible dislocation. This study investigates the feasibility of simulating commonly encountered positions, and positions with an increased risk of impingement, to lower postoperative impingement risk in a CT-based 3D model. A robotic arm-assisted arthroplasty planning platform was used to investigate 11 patient positions. Data from 43 primary THAs were used for simulation. Sacral slope was retrieved from patient preoperative imaging, while angles of hip flexion/extension, hip external/internal rotation, and hip abduction/adduction for tested positions were derived from literature or estimated with a biomechanical model. The hip was placed in the described positions, and if impingement was detected by the software, inspection of the impingement type was performed.Aims
Methods
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size.Aims
Methods
The aims of the study were to report for a cohort aged younger than 40 years: 1) indications for HRA; 2) patient-reported outcomes in terms of the modified Harris Hip Score (HHS); 3) dislocation rate; and 4) revision rate. This retrospective analysis identified 267 hips from 224 patients who underwent an hip resurfacing arthroplasty (HRA) from a single fellowship-trained surgeon using the direct lateral approach between 2007 and 2019. Inclusion criteria was minimum two-year follow-up, and age younger than 40 years. Patients were followed using a prospectively maintained institutional database.Aims
Methods
The Exeter short stem was designed for patients with Dorr type A femora and short-term results are promising. The aim of this study was to evaluate the minimum five-year stem migration pattern of Exeter short stems in comparison with Exeter standard stems. In this case-control study, 25 patients (22 female) at mean age of 78 years (70 to 89) received cemented Exeter short stem (case group). Cases were selected based on Dorr type A femora and matched first by Dorr type A and then age to a control cohort of 21 patients (11 female) at mean age of 74 years (70 to 89) who received with cemented Exeter standard stems (control group). Preoperatively, all patients had primary hip osteoarthritis and no osteoporosis as confirmed by dual X-ray absorptiometry scanning. Patients were followed with radiostereometry for evaluation of stem migration (primary endpoint), evaluation of cement quality, and Oxford Hip Score. Measurements were taken preoperatively, and at three, 12, and 24 months and a minimum five-year follow-up.Aims
Methods
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Pelvic incidence (PI) is considered an important anatomical parameter for determining the sagittal balance of the spine. The contribution of an abnormal PI to hip osteoarthritis (OA) remains controversial. In this study, we aimed to investigate the relationship between PI and hip OA, and the difference in PI between hip OA without anatomical abnormalities (primary OA) and hip OA with developmental dysplasia of the hip (DDH-OA). In this study, 100 patients each of primary OA, DDH-OA, and control subjects with no history of hip disease were included. CT images were used to measure PI, sagittal femoral head coverage, α angle, and acetabular anteversion. PI was also subdivided into three categories: high PI (larger than 64.0°), medium PI (42.0° to 64.0°), and low PI (less than 42.0°). The anterior centre edge angles, posterior centre edge angles, and total sagittal femoral head coverage were measured. The correlations between PI and sagittal femoral head coverage, α angle, and acetabular anteversion were examined.Aims
Methods
Spinopelvic mobility plays an important role in functional acetabular component position following total hip arthroplasty (THA). The primary aim of this study was to determine if spinopelvic hypermobility persists or resolves following THA. Our second aim was to identify patient demographic or radiological factors associated with hypermobility and resolution of hypermobility after THA. This study investigated patients with preoperative posterior hypermobility, defined as a change in sacral slope (SS) from standing to sitting (ΔSSstand-sit) ≥ 30°. Radiological spinopelvic parameters, including SS, pelvic incidence (PI), lumbar lordosis (LL), PI-LL mismatch, anterior pelvic plane tilt (APPt), and spinopelvic tilt (SPT), were measured on preoperative imaging, and at
six weeks and a minimum of
one year postoperatively. The severity of bilateral hip osteoarthritis (OA) was graded using Kellgren-Lawrence criteria.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
Cross-table lateral (CTL) radiographs are commonly used to measure acetabular component anteversion after total hip arthroplasty (THA). The CTL measurements may differ by > 10° from CT scan measurements but the reasons for this discrepancy are poorly understood. Anteversion measurements from CTL radiographs and CT scans are compared to identify spinopelvic parameters predictive of inaccuracy. THA patients (n = 47; 27 males, 20 females; mean age 62.9 years (SD 6.95)) with preoperative spinopelvic mobility, radiological analysis, and postoperative CT scans were retrospectively reviewed. Acetabular component anteversion was measured on postoperative CTL radiographs and CT scans using 3D reconstructions of the pelvis. Two cohorts were identified based on a CTL-CT error of ≥ 10° (n = 11) or < 10° (n = 36). Spinopelvic mobility parameters were compared using independent-samples Aims
Methods
The material and design of knee components can have a considerable effect on the contact characteristics of the tibial post. This study aimed to analyze the stress distribution on the tibial post when using different grades of polyethylene for the tibial inserts. In addition, the contact properties of fixed-bearing and mobile-bearing inserts were evaluated. Three different grades of polyethylene were compared in this study; conventional ultra high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE), and vitamin E-stabilized polyethylene (VEPE). In addition, tibial baseplates with a fixed-bearing and a mobile-bearing insert were evaluated to understand differences in the contact properties. The inserts were implanted in neutral alignment and with a 10° internal malrotation. The contact stress, von Mises stress, and equivalent plastic strain (PEEQ) on the tibial posts were extracted for comparison.Aims
Methods
There is inconsistent evidence on whether prior spinal fusion surgery adversely impacts outcomes following total hip arthroplasty (THA). We conducted a systematic review and meta-analysis to assess the association between pre-existing spinal fusion surgery and the rate of complications following primary THA. We searched MEDLINE, Embase, Web of Science, and Cochrane Library up to October 2019 for randomized controlled trials (RCTs) and observational studies comparing outcomes of dislocation, revision, or reasons for revision in patients following primary THA with or without pre-existing spinal fusion surgery. Furthermore, we compared short (two or less levels) or long (three or more levels) spinal fusions to no fusion. Summary measures of association were relative risks (RRs) (with 95% confidence intervals (CIs)).Aims
Methods
The optimum clearance between the bearing surfaces of hip arthroplasties is unknown. Theoretically, to minimize wear, it is understood that clearances must be low enough to maintain optimal contact pressure and fluid film lubrication, while being large enough to allow lubricant recovery and reduce contact patch size. This study aimed to identify the relationship between diametrical clearance and volumetric wear, through the analysis of retrieved components. A total of 81 metal-on-metal Pinnacle hips paired with 12/14 stems were included in this study. Geometrical analysis was performed on each component, using coordinate and roundness measuring machines. The relationship between their as-manufactured diametrical clearance and volumetric wear was investigated. The Mann-Whitney U test and unpaired Aims
Methods
The medially spherical GMK Sphere (Medacta International AG, Castel San Pietro, Switzerland) total knee arthroplasty (TKA) was previously shown to accommodate lateral rollback while pivoting around a stable medial compartment, aiming to replicate native knee kinematics in which some coronal laxity, especially laterally, is also present. We assess coronal plane kinematics of the GMK Sphere and explore the occurrence and pattern of articular separation during static and dynamic activities. Using pulsed fluoroscopy and image matching, the coronal kinematics and articular surface separation of 16 well-functioning TKAs were studied during weight-bearing and non-weight-bearing, static, and dynamic activities. The closest distances between the modelled articular surfaces were examined with respect to knee position, and proportions of joint poses exhibiting separation were computed.Objectives
Methods
It is important to consider sagittal pelvic rotation when introducing
the acetabular component at total hip arthroplasty (THA). The purpose
of this study was to identify patients who are at risk of unfavourable
pelvic mobility, which could result in poor outcomes after THA. A consecutive series of 4042 patients undergoing THA had lateral
functional radiographs and a low-dose CT scan to measure supine
pelvic tilt, pelvic incidence, standing pelvic tilt, flexed-seated
pelvic tilt, standing lumbar lordotic angle, flexed-seated lumbar
lordotic angle, and lumbar flexion. Changes in pelvic tilt from
supine-to-standing positions and supine-to-flexed-seated positions
were determined. A change in pelvic tilt of 13° between positions was
deemed unfavourable as it alters functional anteversion by 10° and
effectively places the acetabular component outside the safe zone
of orientation.Aims
Patients and Methods
Total ankle arthroplasty (TAA) surgery is complex and attracts a wide variety of complications. The literature lacks consistency in reporting adverse events and complications. The aim of this article is to provide a comprehensive analysis of each of these complications from a literature review, and to compare them with rates from our Unit, to aid clinicians with the process of informed consent. A total of 278 consecutive total ankle arthroplasties (251 patients), performed by four surgeons over a six-year period in Wrightington Hospital (Wigan, United Kingdom) were prospectively reviewed. There were 143 men and 108 women with a mean age of 64 years (41 to 86). The data were recorded on each follow-up visit. Any complications either during initial hospital stay or subsequently reported on follow-ups were recorded, investigated, monitored, and treated as warranted. Literature search included the studies reporting the outcomes and complications of TAA implants.Aims
Patients and Methods
It is not known whether change in patient-reported outcome measures (PROMs) over time can be predicted by factors present at surgery, or early follow-up. The aim of this study was to identify factors associated with changes in PROM status between two-year evaluation and medium-term follow-up. Patients undergoing Birmingham Hip Resurfacing completed the Veteran’s Rand 36 (VR-36), modified Harris Hip Score (mHHS), Tegner Activity Score, and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) at two years and a minimum of three years. A change in score was assessed against minimal clinically important difference (MCID) and patient-acceptable symptom state (PASS) thresholds. Binary logistic regression was used to assess the relationship between patient factors and deterioration in PASS status between follow-ups.Aims
Patients and Methods
Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
This study of patients with osteoarthritis (OA) of the hip aimed to: 1) characterize the contribution of the hip, spinopelvic complex, and lumbar spine when moving from the standing to the sitting position; 2) assess whether abnormal spinopelvic mobility is associated with worse symptoms; and 3) identify whether spinopelvic mobility can be predicted from static anatomical radiological parameters. A total of 122 patients with end-stage OA of the hip awaiting total hip arthroplasty (THA) were prospectively studied. Patient-reported outcome measures (PROMs; Oxford Hip Score, Oswestry Disability Index, and Veterans RAND 12-Item Health Survey Score) and clinical data were collected. Sagittal spinopelvic mobility was calculated as the change from the standing to sitting position using the lumbar lordosis angle (LL), sacral slope (SS), pelvic tilt (PT), pelvic-femoral angle (PFA), and acetabular anteinclination (AI) from lateral radiographs. The interaction of the different parameters was assessed. PROMs were compared between patients with normal spinopelvic mobility (10° ≤ ∆PT ≤ 30°) or abnormal spinopelvic mobility (stiff: ∆PT < ± 10°; hypermobile: ∆PT > ± 30°). Multiple regression and receiver operating characteristic (ROC) curve analyses were used to test for possible predictors of spinopelvic mobility.Aims
Patients and Methods
Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty for patients who require treatment of single-compartment osteoarthritis, especially for young patients. To satisfy this requirement, new patient-specific prosthetic designs have been introduced. The patient-specific UKA is designed on the basis of data from preoperative medical images. In general, knee implant design with increased conformity has been developed to provide lower contact stress and reduced wear on the tibial insert compared with flat knee designs. The different tibiofemoral conformity may provide designers the opportunity to address both wear and kinematic design goals simultaneously. The aim of this study was to evaluate wear prediction with respect to tibiofemoral conformity design in patient-specific UKA under gait loading conditions by using a previously validated computational wear method. Three designs with different conformities were developed with the same femoral component: a flat design normally used in fixed-bearing UKA, a tibia plateau anatomy mimetic (AM) design, and an increased conforming design. We investigated the kinematics, contact stress, contact area, wear rate, and volumetric wear of the three different tibial insert designs.Objectives
Methods
We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless THA using a delta ceramic femoral head and liner. At
each follow-up, clinical and radiological outcomes were recorded.
A Kaplan-Meier analysis was undertaken to estimate survival.Aims
Patients and Methods
We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants. Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.Aims
Patients and Methods
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a large femoral head (38 mm to 52 mm) and the other with
a conventional femoral head (28 mm or 32 mm). We compared clinical
outcome, blood metal ion levels, and the incidence of pseudotumour in
the two groups. Between December 2009 and December 2011, 62 patients underwent
MoM THA with a large femoral head (Magnum group) and 57 patients
an MoM THA with a conventional femoral head (conventional group).
Clinical outcome was assessed using the Harris Hip score, University
of California, Los Angeles (UCLA) activity score and EuroQol-5D
(EQ-5D). Blood metal ion levels were measured and MRI scans were
analyzed at a minimum of five years postoperatively.Aims
Patients and Methods
The aim of this study was to evaluate two hypotheses. First,
that disruption of posterior bundle of the medial collateral ligament
(PMCL) has to occur for the elbow to subluxate in cases of posteromedial
rotatory instability (PMRI) and second, that ulnohumeral contact
pressures increase after disruption of the PMCL. Six human cadaveric elbows were prepared on a custom-designed
apparatus which allowed muscle loading and passive elbow motion
under gravitational varus. Joint contact pressures were measured
sequentially in the intact elbow (INTACT), followed by an anteromedial
subtype two coronoid fracture (COR), a lateral collateral ligament (LCL)
tear (COR + LCL), and a PMCL tear (COR + LCL + PMCL).Aims
Materials and Methods
The purpose of this study was to compare the clinical and radiographic
outcomes of total ankle arthroplasty (TAA) in patients with pre-operatively
moderate and severe arthritic varus ankles to those achieved for
patients with neutral ankles. A total of 105 patients (105 ankles), matched for age, gender,
body mass index, and follow-up duration, were divided into three
groups by pre-operative coronal plane tibiotalar angle; neutral
(<
5°), moderate (5° to 15°) and severe (>
15°) varus deformity.
American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot
score, a visual analogue scale (VAS), and Short Form (SF)-36 score
were used to compare the clinical outcomes after a mean follow-up period
of 51 months (24 to 147).Aims
Patients and Methods
The aim of this study was to report the mid-term clinical outcome
of cemented unlinked J-alumina ceramic elbow (JACE) arthroplasties
when used in patients with rheumatoid arthritis (RA). We retrospectively reviewed 87 elbows, in 75 patients with RA,
which was replaced using a cemented JACE total elbow arthroplasty
(TEA) between August 2003 and December 2012, with a follow-up of
96%. There were 72 women and three men, with a mean age of 62 years
(35 to 79). The mean follow-up was nine years (2 to 14). The clinical condition
of each elbow before and after surgery was assessed using the Mayo
Elbow Performance Index (MEPI, 0 to 100 points). Radiographic loosening
was defined as a progressive radiolucent line of >1 mm that was
completely circumferential around the prosthesis.Aims
Patients and Methods
We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. Aims
Patients and Methods
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
In this Cite this article:
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article: