Advertisement for orthosearch.org.uk
Results 1 - 100 of 319
Results per page:
The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims. The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components. Methods. Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components. Results. We found that the surface roughness of CoCr components rapidly increased during the first 1,000 wear cycles, then reached a steady state, but material loss from the surface continued at a rate of 1,778,000 μm. 3. per million cycles as carbides were removed from its matrix. These carbides formed third-body wear particles, leading to the formation of new scratches even as older scratches were worn away. In contrast, no scratching, loss of material, or other surface damage, when evaluated with one nanometer resolution, was found on the surface of the ceramic components after a 15 M wear cycle test. Conclusion. This study showed wear and loss of CoCr material from scratching and microabrasive wear in TKA. The material loss from the surface continued in a linear relationship with increasing cycles. We also found the absence of scratching and roughening of ceramic femoral components in simulated wear, suggesting an advantage in wear rate and avoiding metal sensitivity. This may have implications in the management of persistent pain after TKA. Cite this article: Bone Joint J 2021;103-B(6 Supple A):94–101


Aims. The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). Methods. A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints. Results. There was no clinically relevant difference in terms of tibial component migration, insert wear, and PROMs between the HXLPE and PE groups. The mean difference in tibial component migration (maximal total point migration (MTPM)) was 0.02 mm (95% confidence interval (CI) -0.07 to 0.11), which is below the value of 0.2 mm considered to be clinically relevant. Wear after five years for HXLPE was 0.16 mm (95% CI 0.05 to 0.27), and for PE was 0.23 mm (95% CI 0.12 to 0.35). The mean difference in wear rate was 0.01 mm/year (95% CI -0.02 to 0.05) in favour of the HXLPE group. Wear is mainly present on the medial side of the insert. Conclusion. There is no clinically relevant difference in tibial component migration and insert wear for up to five years between the HXLPE conventional PE inserts. For the implant studied, the potential advantages of a HXLPE insert remain to be proven under clinical conditions at longer-term follow-up. Cite this article: Bone Joint J 2023;105-B(5):518–525


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 129 - 137
1 Jun 2020
Knowlton CB Lundberg HJ Wimmer MA Jacobs JJ

Aims. A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients. Methods. In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component. Results. Volumetric wear rates on eight non-delaminated components measured 15.9 mm. 3. /year (standard error (SE) ± 7.7) on the total part, 11.4 mm. 3. /year (SE ± 6.4) on the medial side and 4.4 (SE ± 2.6) mm. 3. /year on the lateral side. Volumetric wear rates modelled from patient gait mechanics predicted 16.4 mm. 3. /year (SE 2.4) on the total part, 11.7 mm. 3. /year (SE 2.1) on the medial side and 4.7 mm. 3. /year (SE 0.4) on the lateral side. Measured and modelled wear volumes correlated significantly on the total part (p = 0.017) and the medial side (p = 0.012) but not on the lateral side (p = 0.154). Conclusion. In the absence of delamination, patient-specific knee mechanics during gait directly affect wear of the tibial component in TKA. Cite this article: Bone Joint J 2020;102-B(6 Supple A):129–137


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims. A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA). Patients and Methods. Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear. Results. Linear wear occurred at 0.015 mm/year (supine) and 0.220 mm/year (standing). Volumetric wear occurred at 10.3 mm. 3. /year (supine) and 39.3 mm. 3. /year (standing). Wear occurred primarily on the medial side of the joint. Weightbearing imaging greatly improved the reliability of measurement. Clinical precision of volumetric wear was 34 mm. 3. No significant associations were found between patient demographics or function scores and measured wear. Conclusion. In vivo volumetric wear of TKAs can be assessed at short-term follow-up using MBRSA. Cite this article: Bone Joint J 2019;101-B:1348–1355


Bone & Joint Research
Vol. 8, Issue 11 | Pages 563 - 569
1 Nov 2019
Koh Y Lee J Lee H Kim H Kang K

Objectives. Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty with isolated medial or lateral compartment osteoarthritis. However, polyethylene wear can significantly reduce the lifespan of UKA. Different bearing designs and materials for UKA have been developed to change the rate of polyethylene wear. Therefore, the objective of this study is to investigate the effect of insert conformity and material on the predicted wear in mobile-bearing UKA using a previously developed computational wear method. Methods. Two different designs were tested with the same femoral component under identical kinematic input: anatomy mimetic design (AMD) and conforming design inserts with different conformity levels. The insert materials were standard or crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). We evaluated the contact pressure, contact area, wear rate, wear depth, and volumetric wear under gait cycle loading conditions. Results. Conforming design inserts had the lower contact pressure and larger contact area. However, they also had the higher wear rate and volumetric wear. The improved wear performance was found with AMD inserts. In addition, the computationally predicted volumetric wear of crosslinked UHMWPE inserts was less than half that of standard UHMWPE inserts. Conclusion. Our results showed that increasing conformity may not be the sole predictor of wear performance; highly crosslinked mobile-bearing polyethylene inserts can also provide improvement in wear performance. These results provide improvements in design and materials to reduce wear in mobile-bearing UKA. Cite this article: Bone Joint Res 2019;8:563–569


Objectives. Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty for patients who require treatment of single-compartment osteoarthritis, especially for young patients. To satisfy this requirement, new patient-specific prosthetic designs have been introduced. The patient-specific UKA is designed on the basis of data from preoperative medical images. In general, knee implant design with increased conformity has been developed to provide lower contact stress and reduced wear on the tibial insert compared with flat knee designs. The different tibiofemoral conformity may provide designers the opportunity to address both wear and kinematic design goals simultaneously. The aim of this study was to evaluate wear prediction with respect to tibiofemoral conformity design in patient-specific UKA under gait loading conditions by using a previously validated computational wear method. Methods. Three designs with different conformities were developed with the same femoral component: a flat design normally used in fixed-bearing UKA, a tibia plateau anatomy mimetic (AM) design, and an increased conforming design. We investigated the kinematics, contact stress, contact area, wear rate, and volumetric wear of the three different tibial insert designs. Results. Conforming increased design showed a lower contact stress and increased contact area. In addition, increased conformity resulted in a reduction of the wear rate and volumetric wear. However, the increased conformity design showed limited kinematics. Conclusion. Our results indicated that increased conformity provided improvements in wear but resulted in limited kinematics. Therefore, increased conformity should be avoided in fixed-bearing patient-specific UKA design. We recommend a flat or plateau AM tibial insert design in patient-specific UKA. Cite this article: Y-G. Koh, K-M. Park, H-Y. Lee, K-T. Kang. Influence of tibiofemoral congruency design on the wear of patient-specific unicompartmental knee arthroplasty using finite element analysis. Bone Joint Res 2019;8:156–164. DOI: 10.1302/2046-3758.83.BJR-2018-0193.R1


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims. Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate. Results. Five patient-specific variables and seven implant-specific variables were evaluated for significant association with lower insert wear rate. Six were significant when controlling for other factors: greater patient age, female sex, shorter duration in vivo, polished tray, highly cross-linked PE (HXLPE), and constrained knee design. Conclusion. This study confirmed that knee wear rate increased with duration in vivo. Older patients and females had significantly lower wear rates. Polished modular tibial tray surfaces, HXLPE, and constrained TKA designs were device design factors associated with significantly reduced wear rate. Cite this article: Bone Joint J 2021;103-B(11):1695–1701


Bone & Joint Research
Vol. 13, Issue 5 | Pages 226 - 236
9 May 2024
Jürgens-Lahnstein JH Petersen ET Rytter S Madsen F Søballe K Stilling M

Aims. Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear. Methods. A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay. Results. At mean six years’ follow-up, the PE wear rate was 0.08 mm/year (95% confidence interval 0.06 to 0.09 mm/year). PE inlay external rotation was below the precision limit and did not influence PE wear. Varus knee alignment did not influence PE wear (p = 0.874), but increased tibial component total translation (p = 0.041). Conclusion. The PE inlay was well fixed and there was no relationship between PE stability and PE wear. The PE wear rate was low and similar in the medial and lateral compartments. Varus knee alignment did not influence PE wear. Cite this article: Bone Joint Res 2024;13(5):226–236


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1527 - 1534
3 Nov 2020
Orita K Minoda Y Sugama R Ohta Y Ueyama H Takemura S Nakamura H

Aims. Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA. Methods. A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy. Results. The total number of polyethylene wear particles obtained from the knees with E1 (mean 6.9, SD 4.0 × 10. 7. counts/knee) was greater than that obtained from those with ArCom (mean 2.2, SD 2.6 × 10. 7. counts/knee) (p = 0.001). The particle size (equivalent circle of diameter) from the knees with E1 was smaller (mean 0.5 μm, SD 0.1) than that of knees with ArCom (mean 1.5, SD 0.3 μm) (p = 0.001). The aspect ratio of particles from the knees with E1 (mean 1.3, SD 0.1) was smaller than that with ArCom (mean 1.4, SD 0.1) (p < 0.001 ). Conclusion. This is the first report of in vivo wear particle analysis of E1. E1 polyethylene did not reduce the number of in vivo polyethylene wear particles compared with ArCom in early clinical stage. Further careful follow-up of newly introduced E1 for TKA should be carried out. Cite this article: Bone Joint J 2020;102-B(11):1527–1534


Bone & Joint Research
Vol. 8, Issue 11 | Pages 535 - 543
1 Nov 2019
Mohammad HR Campi S Kennedy JA Judge A Murray DW Mellon SJ

Objectives. The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. Methods. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation. Results. The wear rate for 1900 resin machined (n = 19), 1050 machined (n = 21), and 1050 moulded bearings (n = 23) were 60 µm/year (. sd. 42), 76 µm/year (. sd. 32), and 57 µm/year (. sd. 30), respectively. There was no significant difference between 1900 machined and 1050 machined (p = 0.20), but 1050 moulded had significantly less wear than the 1050 machined (p = 0.05). Increasing femoral (p < 0.001) and tibial (p < 0.001) component size were associated with increasing wear. Conclusion. Wear rate is similar with 1050 and 1900 resin, but lower with moulded bearings than machined bearings. The currently used Phase 3 bearings wear rate is low (1050 moulded, 57 µm/year), but higher than the previously reported Phase 2 bearings (1900 moulded, 20 µm/year). This is unlikely to be due to the change in polyethylene but may relate to the minimally invasive approach used with the Phase 3. This approach, as well as improving function and thus increasing activity levels, may increase the risk of surgical errors, such as impingement or bearing overhang, which can increase wear. Surgeons should aim to use 4 mm thick bearings rather than 3 mm thick bearings in young patients, unless they are small and need conservative bone resections. Cite this article: Bone Joint Res 2019;8:535–543


Bone & Joint Research
Vol. 8, Issue 2 | Pages 65 - 72
1 Feb 2019
Cowie RM Aiken SS Cooper JJ Jennings LM

Objectives. Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation. Methods. A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry. Results. There was no significant difference (p  >  0.05) between the wear rate of implants challenged with CS-BVF (3.3 mm. 3. /MC (95% confidence interval (CI) 1.8 to 4.8)) and the wear rate of those not challenged (2.8 mm. 3. /MC (95% CI 1.3 to 4.3)). However, scratching the cobalt-chrome (CoCr) significantly (p < 0.001) increased the wear rate (20.6 mm. 3. /MC (95% CI 15.5 to 25.7)). The mean surface roughness of implants challenged with CS-BVF was equivalent to negative controls both after damage simulation (p = 0.98) and at the conclusion of the study (p = 0.28). Conclusion. When used close to articulating surfaces, a low-hardness, high-purity CS-BVF had no influence on wear. When trapped between the articulating surfaces of a total knee arthroplasty, CS-BVF did not scratch the surface of CoCr femoral components, nor did it increase the wear of UHMWPE tibial inserts compared with undamaged negative controls. Cite this article: R. M. Cowie, S. S. Aiken, J. J. Cooper, L. M. Jennings. The influence of a calcium sulphate bone void filler on the third-body damage and polyethylene wear of total knee arthroplasty. Bone Joint Res 2019;8:65–72. DOI: 10.1302/2046-3758.82.BJR-2018-0146.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 367 - 373
1 Mar 2010
Kendrick BJL Longino D Pandit H Svard U Gill HS Dodd CAF Murray DW Price AJ

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (. sd. 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/− signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 470 - 475
1 Apr 2011
Kendrick BJL Simpson DJ Kaptein BL Valstar ER Gill HS Murray DW Price AJ

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1493 - 1497
1 Nov 2005
Price AJ Short A Kellett C Beard D Gill H Pandit H Dodd CAF Murray DW

Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique. In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year. The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 976 - 982
1 Nov 1998
Psychoyios V Crawford RW Murray DW O’Connor JJ

Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements. We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year. In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1057 - 1063
1 Aug 2013
Zeng Y Shen B Yang J Zhou ZK Kang PD Pei FX

The purpose of this study was to undertake a meta-analysis to determine whether there is lower polyethylene wear and longer survival when using mobile-bearing implants in total knee replacement when compared with fixed-bearing implants. Of 975 papers identified, 34 trials were eligible for data extraction and meta-analysis comprising 4754 patients (6861 knees). We found no statistically significant differences between the two designs in terms of the incidence of radiolucent lines, osteolysis, aseptic loosening or survival. There is thus currently no evidence to suggest that the use of mobile-bearing designs reduce polyethylene wear and prolong survival after total knee replacement. Cite this article: Bone Joint J 2013;95-B:1057–63


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 990 - 993
1 Nov 1998
Crawford R Sabokbar A Wulke A Murray DW Athanasou NA

We present a case in which the growth of an intraosseous cyst arising from the proximal tibiofibular joint appeared to have been increased by polyethylene wear particles from a medial unicompartmental knee replacement. Histological examination of the cyst wall showed a histiocytic response associated with numerous polyethylene wear particles. This case demonstrates that there is a direct communication between the joint cavity and the cyst. Such communication is probably through openings in the articular cartilage large enough to allow the passage of these particles


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1103 - 1110
1 Jun 2021
Tetreault MW Hines JT Berry DJ Pagnano MW Trousdale RT Abdel MP

Aims. This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA). Methods. From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded. Results. Survivorship free of any re-revision was 68% at ten years. For the indication of insert wear, survivorship free of any re-revision at ten years was 74%. Re-revisions were more frequent for index diagnoses other than wear (hazard ratio (HR) 1.9; p = 0.013), with ten-year survivorships of 69% for instability and 37% for insert fracture/dissociation. Following ITIE for wear, the most common reason for re-revision was aseptic loosening (33%, n = 7). For other indications, the most common reason for re-revision was recurrence of the original diagnosis. Mean Knee Society Scores improved from 54 (0 to 94) preoperatively to 77 (38 to 94) at ten years. Conclusion. After ITIE, the risk and reasons for re-revision correlated with preoperative indications. The best results were for polyethylene wear. For other diagnoses, the re-revision rate was higher and the failure mode was most commonly recurrence of the original indication for the revision TKA. Cite this article: Bone Joint J 2021;103-B(6):1103–1110


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791–1801


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 277 - 283
1 Mar 2023
Gausden EB Puri S Chiu Y Figgie MP Sculco TP Westrich G Sculco PK Chalmers BP

Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This failure mode, which we refer to as aseptic mechanical failure, was the most common revision indication. The mean time from primary TKA to revision for aseptic mechanical failure was five years (5 months to 11 years). Conclusion. In this series of nearly 8,000 primary TKAs performed with a specific implant, we identified a lower-than-expected mid-term survivorship and a high number of revisions with a unique presentation. This study, along with the recent recall of the implant, confirms the need for frequent monitoring of patients with Optetrak TKAs given the incidence of polyethylene failure, osteolysis, and component loosening. Cite this article: Bone Joint J 2023;105-B(3):277–283


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 47 - 55
1 Jan 2023
Clement ND Avery P Mason J Baker PN Deehan DJ

Aims. The aim of this study was to identify variables associated with time to revision, demographic details associated with revision indication, and type of prosthesis employed, and to describe the survival of hinge knee arthroplasty (HKA) when used for first-time knee revision surgery and factors that were associated with re-revision. Methods. Patient demographic details, BMI, American Society of Anesthesiologists (ASA) grade, indication for revision, surgical approach, surgeon grade, implant type (fixed and rotating), time of revision from primary implantation, and re-revision if undertaken were obtained from the National Joint Registry data for England, Wales, Northern Ireland, and the Isle of Man over an 18-year period (2003 to 2021). Results. There were 3,855 patient episodes analyzed with a median age of 73 years (interquartile range (IQR) 66 to 80), and the majority were female (n = 2,480, 64.3%). The median time to revision from primary knee arthroplasty was 1,219 days (IQR 579 to 2,422). Younger age (p < 0.001), decreasing ASA grade (p < 0.001), and indications for revision of sepsis (p < 0.001), unexplained pain (p < 0.001), non-polyethylene wear (p < 0.001), and malalignment (p < 0.001) were all associated with an earlier time to revision from primary implantation. The median follow-up was 4.56 years (range 0.00 to 17.52), during which there were 410 re-revisions. The overall unadjusted probability of re-revision for all revision HKAs at one, five, and ten years after surgery were 2.7% (95% confidence interval (CI) 2.2 to 3.3), 10.7% (95% CI 9.6 to 11.9), and 16.2% (95% CI 14.5 to 17.9), respectively. Male sex (p < 0.001), younger age (p < 0.001), revision for septic indications (p < 0.001) or implant fracture (p = 0.010), a fixed hinge (p < 0.001), or surgery performed by a non-consultant grade (p = 0.023) were independently associated with an increased risk of re-revision. Conclusion. There were several factors associated with time to first revision. The re-revision rate was 16.2% at ten years; however, the risk factors associated with an increased risk of re-revision could be used to counsel patients regarding their outcome. Cite this article: Bone Joint J 2023;105-B(1):47–55


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal


Bone & Joint Open
Vol. 3, Issue 5 | Pages 441 - 447
23 May 2022
Mikkelsen M Wilson HA Gromov K Price AJ Troelsen A

Aims. Treatment of end-stage anteromedial osteoarthritis (AMOA) of the knee is commonly approached using one of two surgical strategies: medial unicompartmental knee arthroplasty (UKA) or total knee arthroplasty (TKA). In this study we aim to investigate if there is any difference in outcome for patients undergoing UKA or TKA, when treated by high-volume surgeons, in high-volume centres, using two different clinical guidelines. The two strategies are ‘UKA whenever possible’ vs TKA for all patients with AMOA. Methods. A total of 501 consecutive AMOA patients (301 UKA) operated on between 2013 to 2016 in two high-volume centres were included. Centre One employed clinical guidelines for the treatment of AMOA allowing either UKA or TKA, but encouraged UKA wherever possible. Centre Two used clinical guidelines that treated all patients with a TKA, regardless of wear pattern. TKA patients were included if they had isolated AMOA on preoperative radiographs. Data were collected from both centres’ local databases. The primary outcome measure was change in Oxford Knee Score (OKS), and the proportion of patients achieving the patient-acceptable symptom state (PASS) at one-year follow-up. The data were 1:1 propensity score matched before regression models were used to investigate potential differences. Results. The matched cohort included 400 patients (mean age 67 years (SD 9.55), 213 (53%) female, mean BMI 30.2 kg/m. 2. , 337 (84%) American Society of Anesthesiologists grade ≤ 2). We found a mean adjusted difference in change score of 3.02 points (95% confidence interval (CI) 1.41 to 4.63; p < 0.001) and a significantly larger likeliness of achieving PASS (odds ratio 3.67 (95% CI 1.73 to 8.45); p = 0.001) both in favour of the UKA strategy. Conclusion. UKA and TKA are both good strategies for treating end-stage AMOA. However, when compared as a strategy, UKA achieved larger improvements in OKS, and were more likely to reach the PASS value at one-year follow-up. Cite this article: Bone Jt Open 2022;3(5):441–447


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 376 - 385
1 Mar 2022
Gramlich Y Hofmann L Kress S Ruckes C Kemmerer M Klug A Hoffmann R Kremer M

Aims. This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. Methods. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life. Results. No differences in mean age (69 years, 40 to 86), mean age adapted Charlson Comorbidity Index (3.1 (SD 1.4)), mean BMI (29.2 kg/m. 2. (SD 5.8)), or number of other implants were observed between groups. Significant improvements in outcome scores and pain levels were achieved for all groups, and there was no difference in quality of life (12-Item Short-Form Health Survey questionnaire (SF-12)). Mean cobalt and chromium ion levels were significantly higher for the GenuX versus the BPKS hinged TKA (GenuX vs BPKS: cobalt: 16.3 vs 9.4 µg/l; chromium: 9.5 vs 5.2 µg/l). The TiN-coated implants did not appear to confer improvement in the metal ion levels. Metal ion concentrations above 7 µg/l were detected in 81%(29/36) of GenuX patients versus 41% (11/27) in the BPKS group. No GenuX patients had normal levels under 2 µg/l, versus 22% of BPKS patients. No significant reduction in outcome scores was observed regardless of the metal ion levels, whereas higher work-related activity was correlated with higher chromium concentrations. Conclusion. Hinged TKA, using MoM hinges, resulted in critically high cobalt and chromium ion concentrations. The BPKS hinged TKA showed significantly lower metal ion concentrations compared with the GenuX TKA. No benefits were observed using TiN coating. The different weightbearing mechanics might influence the wear of the component materials. Higher workloads and physical activity could influence chromium levels. Cite this article: Bone Joint J 2022;104-B(3):376–385


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 151 - 157
1 Jun 2020
Gil D Atici AE Connolly RL Hugard S Shuvaev S Wannomae KK Oral E Muratoglu OK

Aims. We propose a state-of-the-art temporary spacer, consisting of a cobalt-chrome (CoCr) femoral component and a gentamicin-eluting ultra-high molecular weight polyethylene (UHMWPE) tibial insert, which can provide therapeutic delivery of gentamicin, while retaining excellent mechanical properties. The proposed implant is designed to replace conventional spacers made from bone cement. Methods. Gentamicin-loaded UHMWPE was prepared using phase-separated compression moulding, and its drug elution kinetics, antibacterial, mechanical, and wear properties were compared with those of conventional gentamicin-loaded bone cement. Results. Gentamicin-loaded UHMWPE tibial components not only eradicated planktonic Staphylococcus aureus, but also prevented colonization of both femoral and tibial components. The proposed spacer possesses far superior mechanical and wear properties when compared with conventional bone cement spacers. Conclusion. The proposed gentamicin-eluting UHMWPE spacer can provide antibacterial efficacy comparable with currently used bone cement spacers, while overcoming their drawbacks. The novel spacer proposed here has the potential to drastically reduce complications associated with currently used bone cement spacers and substantially improve patients’ quality of life during the treatment. Cite this article: Bone Joint J 2020;102-B(6 Supple A):151–157


Bone & Joint Open
Vol. 2, Issue 6 | Pages 397 - 404
1 Jun 2021
Begum FA Kayani B Magan AA Chang JS Haddad FS

Limb alignment in total knee arthroplasty (TKA) influences periarticular soft-tissue tension, biomechanics through knee flexion, and implant survival. Despite this, there is no uniform consensus on the optimal alignment technique for TKA. Neutral mechanical alignment facilitates knee flexion and symmetrical component wear but forces the limb into an unnatural position that alters native knee kinematics through the arc of knee flexion. Kinematic alignment aims to restore native limb alignment, but the safe ranges with this technique remain uncertain and the effects of this alignment technique on component survivorship remain unknown. Anatomical alignment aims to restore predisease limb alignment and knee geometry, but existing studies using this technique are based on cadaveric specimens or clinical trials with limited follow-up times. Functional alignment aims to restore the native plane and obliquity of the joint by manipulating implant positioning while limiting soft tissue releases, but the results of high-quality studies with long-term outcomes are still awaited. The drawbacks of existing studies on alignment include the use of surgical techniques with limited accuracy and reproducibility of achieving the planned alignment, poor correlation of intraoperative data to long-term functional outcomes and implant survivorship, and a paucity of studies on the safe ranges of limb alignment. Further studies on alignment in TKA should use surgical adjuncts (e.g. robotic technology) to help execute the planned alignment with improved accuracy, include intraoperative assessments of knee biomechanics and periarticular soft-tissue tension, and correlate alignment to long-term functional outcomes and survivorship


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 91 - 95
1 Jun 2020
Johnson, Jr. WB Engh, Jr. CA Parks NL Hamilton WG Ho PH Fricka KB

Aims. It has been hypothesized that a unicompartmental knee arthroplasty (UKA) is more likely to be revised than a total knee arthroplasty (TKA) because conversion surgery to a primary TKA is a less complicated procedure. The purpose of this study was to determine if there is a lower threshold for revising a UKA compared with TKA based on Oxford Knee Scores (OKSs) and range of movement (ROM) at the time of revision. Methods. We retrospectively reviewed 619 aseptic revision cases performed between December 1998 and October 2018. This included 138 UKAs that underwent conversion to TKA and 481 initial TKA revisions. Age, body mass index (BMI), time in situ, OKS, and ROM were available for all patients. Results. There were no differences between the two groups based on demographics or time to revision. The top reasons for aseptic TKA revision were loosening in 212 (44%), instability in 88 (18%), and wear in 69 (14%). UKA revision diagnoses were primarily for loosening in 50 (36%), progression of osteoarthritis (OA) in 50 (36%), and wear in 17 (12%). Out of a maximum 48 points, the mean OKS of the UKAs before revision was 23 (SD 9.3), which was significantly higher than the TKAs at 19.2 (SD 9.8; p < 0.001). UKA patients scored statistically better on nine of the 12 individual OKS questions. The UKA cases also had a larger pre-revision mean ROM (114°, SD 14.3°) than TKAs (98°, SD 25°) ; p < 0.001). Conclusion. At revision, the mean UKA OKSs and ROM were significantly better than those of TKA cases. This study suggests that at our institution there is a difference in preoperative OKS between UKA and TKA at the time of revision, demonstrating a revision bias. Cite this article: Bone Joint J 2020;102-B(6 Supple A):91–95


Bone & Joint Open
Vol. 2, Issue 12 | Pages 1082 - 1088
1 Dec 2021
Hippalgaonkar K Chandak V Daultani D Mulpur P Eachempati KK Reddy AG

Aims. Single-shot adductor canal block (ACB) after total knee arthroplasty (TKA) for postoperative analgesia is a common modality. Patients can experience breakthrough pain when the effect of ACB wears off. Local anaesthetic infusion through an intra-articular catheter (IAC) can help manage breakthrough pain after TKA. We hypothesized that combined ACB with ropivacaine infusion through IAC is associated with better pain relief compared to ACB used alone. Methods. This study was a prospective double-blinded placebo-controlled randomized controlled trial to compare the efficacy of combined ACB+ IAC-ropivacaine infusion (study group, n = 68) versus single-shot ACB+ intra-articular normal saline placebo (control group, n = 66) after primary TKA. The primary outcome was assessment of pain, using the visual analogue scale (VAS) recorded at 6, 12, 24, and 48 hours after surgery. Secondary outcomes included active knee ROM 48 hours after surgery and additional requirement of analgesia for breakthrough pain. Results. The study group (mean visual analogue scale (VAS) pain score of 5.5 (SD 0.889)) experienced significant reduction in pain 12 hours after surgery compared to the control group (mean VAS 6.62 (SD 1.356); mean difference = 1.12, 95% confidence interval (CI) -1.46 to 0.67; p < 0.001), and pain scores on postoperative day (POD) 1 and POD-2 were lower in the study group compared to the control group (mean difference in VAS pain = 1.04 (-1.39 to -0.68, 95% CI, p < 0.001). Fewer patients in the study group (0 vs 3 in the control group) required additional analgesia for breakthrough pain, but this was not statistically significant. The study group had significantly increased active knee flexion (mean flexion 86.4° (SD 7.22°)), compared to the control group (mean 73.86° (SD 7.88°), mean difference = 12.54, 95% CI 9.97 to 15.1; p < 0.014). Conclusion. Combined ACB+ ropivacaine infusion via IAC is a safe, reproducible analgesic modality after primary TKA, with superior analgesia compared to ACB alone. Further large volume trials are warranted to generate evidence on clinical significance on analgesia after TKA. Cite this article: Bone Jt Open 2021;2(12):1082–1088


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article: Bone Jt Open 2021;2(1):48–57


Aims

The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients.

Methods

A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1279 - 1285
1 Dec 2023
Baker JF Nadar AC Jouflas AC Smith LS Sachdeva S Yakkanti MR Malkani AL

Aims

The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

Methods

This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.


Aims

The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral.

Methods

The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 277 - 285
8 Apr 2024
Khetan V Baxter I Hampton M Spencer A Anderson A

Aims

The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery.

Methods

A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 916 - 923
1 Sep 2024
Fricka KB Wilson EJ Strait AV Ho H Hopper, Jr RH Hamilton WG Sershon RA

Aims

The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution.

Methods

Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).


Bone & Joint Research
Vol. 13, Issue 4 | Pages 149 - 156
4 Apr 2024
Rajamäki A Lehtovirta L Niemeläinen M Reito A Parkkinen J Peräniemi S Vepsäläinen J Eskelinen A

Aims

Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium.

Methods

Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia.


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1271 - 1278
1 Dec 2023
Rehman Y Korsvold AM Lerdal A Aamodt A

Aims

This study compared patient-reported outcomes of three total knee arthroplasty (TKA) designs from one manufacturer: one cruciate-retaining (CR) design, and two cruciate-sacrificing designs, anterior-stabilized (AS) and posterior-stabilized (PS).

Methods

Patients scheduled for primary TKA were included in a single-centre, prospective, three-armed, blinded randomized trial (n = 216; 72 per group). After intraoperative confirmation of posterior cruciate ligament (PCL) integrity, patients were randomly allocated to receive a CR, AS, or PS design from the same TKA system. Insertion of an AS or PS design required PCL resection. The primary outcome was the mean score of all five subscales of the Knee injury and Osteoarthritis Outcome Score (KOOS) at two-year follow-up. Secondary outcomes included all KOOS subscales, Oxford Knee Score, EuroQol five-dimension health questionnaire, EuroQol visual analogue scale, range of motion (ROM), and willingness to undergo the operation again. Patient satisfaction was also assessed.


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 269 - 276
1 Mar 2023
Tay ML Monk AP Frampton CM Hooper GJ Young SW

Aims

Unicompartmental knee arthroplasty (UKA) has higher revision rates than total knee arthroplasty (TKA). As revision of UKA may be less technically demanding than revision TKA, UKA patients with poor functional outcomes may be more likely to be offered revision than TKA patients with similar outcomes. The aim of this study was to compare clinical thresholds for revisions between TKA and UKA using revision incidence and patient-reported outcomes, in a large, matched cohort at early, mid-, and late-term follow-up.

Methods

Analyses were performed on propensity score-matched patient cohorts of TKAs and UKAs (2:1) registered in the New Zealand Joint Registry between 1 January 1999 and 31 December 2019 with an Oxford Knee Score (OKS) response at six months (n, TKA: 16,774; UKA: 8,387), five years (TKA: 6,718; UKA: 3,359), or ten years (TKA: 3,486; UKA: 1,743). Associations between OKS and revision within two years following the score were examined. Thresholds were compared using receiver operating characteristic analysis. Reasons for aseptic revision were compared using cumulative incidence with competing risk.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 635 - 640
1 Jun 2023
Karczewski D Siljander MP Larson DR Taunton MJ Lewallen DG Abdel MP

Aims

Knowledge on total knee arthroplasties (TKAs) in patients with a history of poliomyelitis is limited. This study compared implant survivorship and clinical outcomes among affected and unaffected limbs in patients with sequelae of poliomyelitis undergoing TKAs.

Methods

A retrospective review of our total joint registry identified 94 patients with post-polio syndrome undergoing 116 primary TKAs between January 2000 and December 2019. The mean age was 70 years (33 to 86) with 56% males (n = 65) and a mean BMI of 31 kg/m2 (18 to 49). Rotating hinge TKAs were used in 14 of 63 affected limbs (22%), but not in any of the 53 unaffected limbs. Kaplan-Meier survivorship analyses were completed. The mean follow-up was eight years (2 to 19).


Aims

Achievement of accurate microbiological diagnosis prior to revision is key to reducing the high rates of persistent infection after revision knee surgery. The effect of change in the microorganism between the first- and second-stage revision of total knee arthroplasty for periprosthetic joint infection (PJI) on the success of management is not clear.

Methods

A two-centre retrospective cohort study was conducted to review the outcome of patients who have undergone two-stage revision for treatment of knee arthroplasty PJI, focusing specifically on isolated micro-organisms at both the first- and second-stage procedure. Patient demographics, medical, and orthopaedic history data, including postoperative outcomes and subsequent treatment, were obtained from the electronic records and medical notes.


Aims

The tibial component of total knee arthroplasty can either be an all-polyethylene (AP) implant or a metal-backed (MB) implant. This study aims to compare the five-year functional outcomes of AP tibial components to MB components in patients aged over 70 years. Secondary aims are to compare quality of life, implant survivorship, and cost-effectiveness.

Methods

A group of 130 patients who had received an AP tibial component were matched for demographic factors of age, BMI, American Society of Anesthesiologists (ASA) grade, sex, and preoperative Knee Society Score (KSS) to create a comparison group of 130 patients who received a MB tibial component. Functional outcome was assessed prospectively by KSS, quality of life by 12-Item Short-Form Health Survey questionnaire (SF-12), and range of motion (ROM), and implant survivorships were compared. The SF six-dimension (6D) was used to calculate the incremental cost effectiveness ratio (ICER) for AP compared to MB tibial components using quality-adjusted life year methodology.


Bone & Joint Open
Vol. 4, Issue 5 | Pages 393 - 398
25 May 2023
Roof MA Lygrisse K Shichman I Marwin SE Meftah M Schwarzkopf R

Aims

Revision total knee arthroplasty (rTKA) is a technically challenging and costly procedure. It is well-documented that primary TKA (pTKA) have better survivorship than rTKA; however, we were unable to identify any studies explicitly investigating previous rTKA as a risk factor for failure following rTKA. The purpose of this study is to compare the outcomes following rTKA between patients undergoing index rTKA and those who had been previously revised.

Methods

This retrospective, observational study reviewed patients who underwent unilateral, aseptic rTKA at an academic orthopaedic speciality hospital between June 2011 and April 2020 with > one-year of follow-up. Patients were dichotomized based on whether this was their first revision procedure or not. Patient demographics, surgical factors, postoperative outcomes, and re-revision rates were compared between the groups.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 907 - 915
1 Sep 2024
Ross M Zhou Y English M Sharplin P Hirner M

Aims

Knee osteoarthritis (OA) is characterized by a chronic inflammatory process involving multiple cytokine pathways, leading to articular cartilage degeneration. Intra-articular therapies using pharmaceutical or autologous anti-inflammatory factors offer potential non-surgical treatment options. Autologous protein solution (APS) is one such product that uses the patient’s blood to produce a concentrate of cells and anti-inflammatory cytokines. This study evaluated the effect of a specific APS intra-articular injection (nSTRIDE) on patient-reported outcome measures compared to saline in moderate knee OA.

Methods

A parallel, double-blinded, placebo-controlled randomized controlled trial was conducted, where patients with unilateral moderate knee OA (Kellgren-Lawrence grade 2 or 3) received either nSTRIDE or saline (placebo) injection to their symptomatic knee. The primary outcome was the difference in Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) total score at 12 months post-intervention. Secondary outcomes included WOMAC component scores, Knee injury and Osteoarthritis Outcome Score (KOOS), and visual analogue scale (VAS) scores at all follow-up timepoints (three, six, and 12 months).


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 622 - 634
1 Jun 2023
Simpson CJRW Wright E Ng N Yap NJ Ndou S Scott CEH Clement ND

Aims

This systematic review and meta-analysis aimed to compare the influence of patellar resurfacing following cruciate-retaining (CR) and posterior-stabilized (PS) total knee arthroplasty (TKA) on the incidence of anterior knee pain, knee-specific patient-reported outcome measures, complication rates, and reoperation rates.

Methods

A systematic review of MEDLINE, PubMed, and Google Scholar was performed to identify randomized controlled trials (RCTs) according to search criteria. Search terms used included: arthroplasty, replacement, knee (Mesh), TKA, prosthesis, patella, patellar resurfacing, and patellar retaining. RCTs that compared patellar resurfacing versus unresurfaced in primary TKA were included for further analysis. Studies were evaluated using the Scottish Intercollegiate Guidelines Network assessment tool for quality and minimization of bias. Data were synthesized and meta-analysis performed.


Bone & Joint Open
Vol. 5, Issue 5 | Pages 401 - 410
20 May 2024
Bayoumi T Burger JA van der List JP Sierevelt IN Spekenbrink-Spooren A Pearle AD Kerkhoffs GMMJ Zuiderbaan HA

Aims

The primary objective of this registry-based study was to compare patient-reported outcomes of cementless and cemented medial unicompartmental knee arthroplasty (UKA) during the first postoperative year. The secondary objective was to assess one- and three-year implant survival of both fixation techniques.

Methods

We analyzed 10,862 cementless and 7,917 cemented UKA cases enrolled in the Dutch Arthroplasty Registry, operated between 2017 and 2021. Pre- to postoperative change in outcomes at six and 12 months’ follow-up were compared using mixed model analyses. Kaplan-Meier and Cox regression models were applied to quantify differences in implant survival. Adjustments were made for patient-specific variables and annual hospital volume.


Bone & Joint Open
Vol. 3, Issue 8 | Pages 656 - 665
23 Aug 2022
Tran T McEwen P Peng Y Trivett A Steele R Donnelly W Clark G

Aims

The mid-term results of kinematic alignment (KA) for total knee arthroplasty (TKA) using image derived instrumentation (IDI) have not been reported in detail, and questions remain regarding ligamentous stability and revisions. This paper aims to address the following: 1) what is the distribution of alignment of KA TKAs using IDI; 2) is a TKA alignment category associated with increased risk of failure or poor patient outcomes; 3) does extending limb alignment lead to changes in soft-tissue laxity; and 4) what is the five-year survivorship and outcomes of KA TKA using IDI?

Methods

A prospective, multicentre, trial enrolled 100 patients undergoing KA TKA using IDI, with follow-up to five years. Alignment measures were conducted pre- and postoperatively to assess constitutional alignment and final implant position. Patient-reported outcome measures (PROMs) of pain and function were also included. The Australian Orthopaedic Association National Joint Arthroplasty Registry was used to assess survivorship.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1126 - 1131
1 Oct 2022
Hannon CP Kruckeberg BM Pagnano MW Berry DJ Hanssen AD Abdel MP

Aims

We have previously reported the mid-term outcomes of revision total knee arthroplasty (TKA) for flexion instability. At a mean of four years, there were no re-revisions for instability. The aim of this study was to report the implant survivorship and clinical and radiological outcomes of the same cohort of of patients at a mean follow-up of ten years.

Methods

The original publication included 60 revision TKAs in 60 patients which were undertaken between 2000 and 2010. The mean age of the patients at the time of revision TKA was 65 years, and 33 (55%) were female. Since that time, 21 patients died, leaving 39 patients (65%) available for analysis. The cumulative incidence of any re-revision with death as a competing risk was calculated. Knee Society Scores (KSSs) were also recorded, and updated radiographs were reviewed.


Bone & Joint Research
Vol. 12, Issue 5 | Pages 313 - 320
8 May 2023
Saiki Y Kabata T Ojima T Kajino Y Kubo N Tsuchiya H

Aims

We aimed to assess the reliability and validity of OpenPose, a posture estimation algorithm, for measurement of knee range of motion after total knee arthroplasty (TKA), in comparison to radiography and goniometry.

Methods

In this prospective observational study, we analyzed 35 primary TKAs (24 patients) for knee osteoarthritis. We measured the knee angles in flexion and extension using OpenPose, radiography, and goniometry. We assessed the test-retest reliability of each method using intraclass correlation coefficient (1,1). We evaluated the ability to estimate other measurement values from the OpenPose value using linear regression analysis. We used intraclass correlation coefficients (2,1) and Bland–Altman analyses to evaluate the agreement and error between radiography and the other measurements.


Bone & Joint Open
Vol. 4, Issue 9 | Pages 682 - 688
6 Sep 2023
Hampton M Balachandar V Charalambous CP Sutton PM

Aims

Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA.

Methods

A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.


Bone & Joint Open
Vol. 5, Issue 8 | Pages 681 - 687
19 Aug 2024
van de Graaf VA Shen TS Wood JA Chen DB MacDessi SJ

Aims

Sagittal plane imbalance (SPI), or asymmetry between extension and flexion gaps, is an important issue in total knee arthroplasty (TKA). The purpose of this study was to compare SPI between kinematic alignment (KA), mechanical alignment (MA), and functional alignment (FA) strategies.

Methods

In 137 robotic-assisted TKAs, extension and flexion stressed gap laxities and bone resections were measured. The primary outcome was the proportion and magnitude of medial and lateral SPI (gap differential > 2.0 mm) for KA, MA, and FA. Secondary outcomes were the proportion of knees with severe (> 4.0 mm) SPI, and resection thicknesses for each technique, with KA as reference.


The Bone & Joint Journal
Vol. 106-B, Issue 10 | Pages 1059 - 1066
1 Oct 2024
Konishi T Hamai S Tsushima H Kawahara S Akasaki Y Yamate S Ayukawa S Nakashima Y

Aims

The Coronal Plane Alignment of the Knee (CPAK) classification has been developed to predict individual variations in inherent knee alignment. The impact of preoperative and postoperative CPAK classification phenotype on the postoperative clinical outcomes of total knee arthroplasty (TKA) remains elusive. This study aimed to examine the effect of postoperative CPAK classification phenotypes (I to IX), and their pre- to postoperative changes on patient-reported outcome measures (PROMs).

Methods

A questionnaire was administered to 340 patients (422 knees) who underwent primary TKA for osteoarthritis (OA) between September 2013 and June 2019. A total of 231 patients (284 knees) responded. The ­Knee Society Score 2011 (KSS 2011), Knee injury and Osteoarthritis Outcome Score-12 (KOOS-12), and Forgotten Joint Score-12 (FJS-12) were used to assess clinical outcomes. Using preoperative and postoperative anteroposterior full-leg radiographs, the arithmetic hip-knee-ankle angle (aHKA) and joint line obliquity (JLO) were calculated and classified based on the CPAK classification. To investigate the impact on PROMs, multivariable regression analyses using stepwise selection were conducted, considering factors such as age at surgery, time since surgery, BMI, sex, implant use, postoperative aHKA classification, JLO classification, and changes in aHKA and JLO classifications from preoperative to postoperative.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 101 - 108
6 Feb 2024
Jang SJ Kunze KN Casey JC Steele JR Mayman DJ Jerabek SA Sculco PK Vigdorchik JM

Aims

Distal femoral resection in conventional total knee arthroplasty (TKA) utilizes an intramedullary guide to determine coronal alignment, commonly planned for 5° of valgus. However, a standard 5° resection angle may contribute to malalignment in patients with variability in the femoral anatomical and mechanical axis angle. The purpose of the study was to leverage deep learning (DL) to measure the femoral mechanical-anatomical axis angle (FMAA) in a heterogeneous cohort.

Methods

Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A DL workflow was created to measure the FMAA and validated against human measurements. To reflect potential intramedullary guide placement during manual TKA, two different FMAAs were calculated either using a line approximating the entire diaphyseal shaft, and a line connecting the apex of the femoral intercondylar sulcus to the centre of the diaphysis. The proportion of FMAAs outside a range of 5.0° (SD 2.0°) was calculated for both definitions, and FMAA was compared using univariate analyses across sex, BMI, knee alignment, and femur length.


Bone & Joint Open
Vol. 4, Issue 12 | Pages 923 - 931
4 Dec 2023
Mikkelsen M Rasmussen LE Price A Pedersen AB Gromov K Troelsen A

Aims

The aim of this study was to describe the pattern of revision indications for unicompartmental knee arthroplasty (UKA) and total knee arthroplasty (TKA) and any change to this pattern for UKA patients over the last 20 years, and to investigate potential associations to changes in surgical practice over time.

Methods

All primary knee arthroplasty surgeries performed due to primary osteoarthritis and their revisions reported to the Danish Knee Arthroplasty Register from 1997 to 2017 were included. Complex surgeries were excluded. The data was linked to the National Patient Register and the Civil Registration System for comorbidity, mortality, and emigration status. TKAs were propensity score matched 4:1 to UKAs. Revision risks were compared using competing risk Cox proportional hazard regression with a shared γ frailty component.


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 817 - 825
1 Aug 2024
Borukhov I Ismailidis P Esposito CI LiArno S Lyon J McEwen PJ

Aims

This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height.

Methods

Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.


Bone & Joint Open
Vol. 5, Issue 9 | Pages 758 - 765
12 Sep 2024
Gardner J Roman ER Bhimani R Mashni SJ Whitaker JE Smith LS Swiergosz A Malkani AL

Aims

Patient dissatisfaction following primary total knee arthroplasty (TKA) with manual jig-based instruments has been reported to be as high as 30%. Robotic-assisted total knee arthroplasty (RA-TKA) has been increasingly used in an effort to improve patient outcomes, however there is a paucity of literature examining patient satisfaction after RA-TKA. This study aims to identify the incidence of patients who were not satisfied following RA-TKA and to determine factors associated with higher levels of dissatisfaction.

Methods

This was a retrospective review of 674 patients who underwent primary TKA between October 2016 and September 2020 with a minimum two-year follow-up. A five-point Likert satisfaction score was used to place patients into two groups: Group A were those who were very dissatisfied, dissatisfied, or neutral (Likert score 1 to 3) and Group B were those who were satisfied or very satisfied (Likert score 4 to 5). Patient demographic data, as well as preoperative and postoperative patient-reported outcome measures, were compared between groups.


Bone & Joint Open
Vol. 4, Issue 4 | Pages 262 - 272
11 Apr 2023
Batailler C Naaim A Daxhelet J Lustig S Ollivier M Parratte S

Aims

The impact of a diaphyseal femoral deformity on knee alignment varies according to its severity and localization. The aims of this study were to determine a method of assessing the impact of diaphyseal femoral deformities on knee alignment for the varus knee, and to evaluate the reliability and the reproducibility of this method in a large cohort of osteoarthritic patients.

Methods

All patients who underwent a knee arthroplasty from 2019 to 2021 were included. Exclusion criteria were genu valgus, flexion contracture (> 5°), previous femoral osteotomy or fracture, total hip arthroplasty, and femoral rotational disorder. A total of 205 patients met the inclusion criteria. The mean age was 62.2 years (SD 8.4). The mean BMI was 33.1 kg/m2 (SD 5.5). The radiological measurements were performed twice by two independent reviewers, and included hip knee ankle (HKA) angle, mechanical medial distal femoral angle (mMDFA), anatomical medial distal femoral angle (aMDFA), femoral neck shaft angle (NSA), femoral bowing angle (FBow), the distance between the knee centre and the top of the FBow (DK), and the angle representing the FBow impact on the knee (C’KS angle).


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 808 - 816
1 Aug 2024
Hall AJ Cullinan R Alozie G Chopra S Greig L Clarke J Riches PE Walmsley P Ohly NE Holloway N

Aims

Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice.

Methods

This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 610 - 621
1 Jun 2023
Prodromidis AD Chloros GD Thivaios GC Sutton PM Pandit H Giannoudis PV Charalambous CP

Aims

Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods

A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.


Bone & Joint Research
Vol. 11, Issue 7 | Pages 494 - 502
20 Jul 2022
Kwon HM Lee J Koh Y Park KK Kang K

Aims

A functional anterior cruciate ligament (ACL) or posterior cruciate ligament (PCL) has been assumed to be required for patients undergoing unicompartmental knee arthroplasty (UKA). However, this assumption has not been thoroughly tested. Therefore, this study aimed to assess the biomechanical effects exerted by cruciate ligament-deficient knees with medial UKAs regarding different posterior tibial slopes.

Methods

ACL- or PCL-deficient models with posterior tibial slopes of 1°, 3°, 5°, 7°, and 9° were developed and compared to intact models. The kinematics and contact stresses on the tibiofemoral joint were evaluated under gait cycle loading conditions.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 672 - 679
1 Jun 2022
Tay ML Young SW Frampton CM Hooper GJ

Aims

Unicompartmental knee arthroplasty (UKA) has a higher risk of revision than total knee arthroplasty (TKA), particularly for younger patients. The outcome of knee arthroplasty is typically defined as implant survival or revision incidence after a defined number of years. This can be difficult for patients to conceptualize. We aimed to calculate the ‘lifetime risk’ of revision for UKA as a more meaningful estimate of risk projection over a patient’s remaining lifetime, and to compare this to TKA.

Methods

Incidence of revision and mortality for all primary UKAs performed from 1999 to 2019 (n = 13,481) was obtained from the New Zealand Joint Registry (NZJR). Lifetime risk of revision was calculated for patients and stratified by age, sex, and American Society of Anesthesiologists (ASA) grade.


Bone & Joint Open
Vol. 3, Issue 6 | Pages 495 - 501
14 Jun 2022
Keohane D Sheridan GA Masterson E

Aims

Total knee arthroplasty (TKA) is a common and safe orthopaedic procedure. Zimmer Biomet's NexGen is the second most popular brand of implant used in the UK. The primary cause of revision after the first year is aseptic loosening. We present our experience of using this implant, with significant concerns around its performance with regards early aseptic loosening of the tibial component.

Methods

A retrospective, single-surgeon review was carried out of all of the NexGen Legacy Posterior Stabilized (LPS) TKAs performed in this institute. The specific model used for the index procedures was the NexGen Complete Knee System (Legacy Knee-Posterior Stabilized LPS-Flex Articular Surface, LPS-Flex Femoral Component Option, and Stemmed Nonaugmentable Tibial Component Option).


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 500 - 507
1 Apr 2010
Lu Y Huang C Chang T Ho F Cheng C Huang C

Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R. 2. = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (. sd. 0.76) in the MB group and 0.66 mm (. sd. 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores. . For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (. sd.  0.014). This was significantly smaller than the wear rate of 0.11 mm/year (. sd. 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant


Bone & Joint Open
Vol. 3, Issue 6 | Pages 470 - 474
7 Jun 2022
Baek J Lee SC Ryu S Kim J Nam CH

Aims

The purpose of this study was to compare the clinical outcomes, mortalities, implant survival rates, and complications of total knee arthroplasty (TKA) in patients with or without hepatitis B virus (HBV) infection over at least ten years of follow-up.

Methods

From January 2008 to December 2010, 266 TKAs were performed in 169 patients with HBV (HBV group). A total of 169 propensity score–matched patients without HBV were chosen for the control group in a one-to-one ratio. Then, the clinical outcomes, mortalities, implant survival rates, and complications of TKA in the two groups were compared. The mean follow-up periods were 11.7 years (10.5 to 13.4) in the HBV group and 11.8 years (11.5 to 12.4) in the control group.


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 663 - 671
1 Jun 2022
Lewis E Merghani K Robertson I Mulford J Prentice B Mathew R Van Winden P Ogden K

Aims

Platelet-rich plasma (PRP) intra-articular injections may provide a simple and minimally invasive treatment for early-stage knee osteoarthritis (OA). This has led to an increase in its adoption as a treatment for knee OA, although there is uncertainty about its efficacy and benefit. We hypothesized that patients with early-stage symptomatic knee OA who receive multiple PRP injections will have better clinical outcomes than those receiving single PRP or placebo injections.

Methods

A double-blinded, randomized placebo-controlled trial was performed with three groups receiving either placebo injections (Normal Saline), one PRP injection followed by two placebo injections, or three PRP injections. Each injection was given one week apart. Outcomes were prospectively collected prior to intervention and then at six weeks, three months, six months, and 12 months post-intervention. Primary outcome measures were Knee Injury and Osteoarthritis Outcome Score (KOOS) and EuroQol five-dimension five-level index (EQ-5D-5L). Secondary outcomes included visual analogue scale for pain and patient subjective assessment of the injections.


Aims

The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 613 - 619
2 May 2022
Ackerman IN Busija L Lorimer M de Steiger R Graves SE

Aims

This study aimed to describe the use of revision knee arthroplasty in Australia and examine changes in lifetime risk over a decade.

Methods

De-identified individual-level data on all revision knee arthroplasties performed in Australia from 2007 to 2017 were obtained from the Australian Orthopaedic Association National Joint Replacement Registry. Population data and life tables were obtained from the Australian Bureau of Statistics. The lifetime risk of revision surgery was calculated for each year using a standardized formula. Separate calculations were undertaken for males and females.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1221 - 1227
1 Sep 2012
Kim Y Park J Kim J

Despite many claims of good wear properties following total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral component, there are conflicting clinical results. We hypothesised that there would be no difference in either the mid-term clinical and radiological outcomes or the characteristics of the polyethylene wear particles (weight, size and shape) in patients using an OxZr or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent bilateral TKR, receiving an OxZr femoral component in one knee and a CoCr femoral component in the other. The mean follow-up was 7.5 years (6 to 8). Following aspiration, polyethylene wear particles were analysed using thermogravimetric methods and scanning electron microscopy. At the most recent follow-up, the mean Knee Society score, Western Ontario and McMaster Universities Osteoarthritis Index score, range of movement and satisfaction score were not significantly different in the two groups. The mean weight, size, aspect ratio and roundness of the aspirated wear particles were similar for each femoral component. Survivorship of the femoral, tibial and patellar components was 100% in both groups. In the absence of evidence of an advantage in the medium term we cannot justify the additional expense of an OxZr femoral component.


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1162 - 1167
1 Sep 2018
Metcalfe AJ Ahearn N Hassaballa MA Parsons N Ackroyd CE Murray JR Robinson JR Eldridge JD Porteous AJ

Aims. This study reports on the medium- to long-term implant survivorship and patient-reported outcomes for the Avon patellofemoral joint (PFJ) arthroplasty. Patients and Methods. A total of 558 Avon PFJ arthroplasties in 431 patients, with minimum two-year follow-up, were identified from a prospective database. Patient-reported outcomes and implant survivorship were analyzed, with follow-up of up to 18 years. Results. Outcomes were recorded for 483 implants (368 patients), representing an 86% follow-up rate. The median postoperative Oxford Knee Score (0 to 48 scale) was 35 (interquartile range (IQR) 25.5 to 43) and the median Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC, 0 to 100 scale) was 35 (IQR 25 to 53) at two years. There were 105 revisions, 61 (58%) for progression of osteoarthritis. All documented revisions were to primary knee systems without augmentation. The implant survival rate was 77.3% (95% confidence interval (CI) 72.4 to 81.7, number at risk 204) at ten years and 67.4% (95% CI 72.4 to 81.7 number at risk 45) at 15 years. Regression analysis of explanatory data variable showed that cases performed in the last nine years had improved survival compared with the first nine years of the cohort, but the individual operating surgeon had the strongest effect on survivorship. Conclusion. Satisfactory long-term results can be obtained with the Avon PFJ arthroplasty, with maintenance of patient-reported outcome measures (PROMs), satisfactory survival, and low rates of loosening and wear. Cite this article: Bone Joint J 2018;100-B:1162–7


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 996 - 1002
1 Aug 2017
Brown TS Van Citters DW Berry DJ Abdel MP

Advances in polyethylene (PE) in total hip arthroplasty have led to interest and increased use of highly crosslinked PE (HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest improved wear characteristics for HXLPE inserts over conventional PE in TKA. Short-term results from registry data and few clinical trials are promising. Our aim is to present a review of the history of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical complications, and a thorough review of the available biomechanical and clinical data. Cite this article: Bone Joint J 2017;99-B:996–1002


Bone & Joint Open
Vol. 3, Issue 3 | Pages 211 - 217
1 Mar 2022
Hsu C Chen C Wang S Huang J Tong K Huang K

Aims

The Coronal Plane Alignment of the Knee (CPAK) classification is a simple and comprehensive system for predicting pre-arthritic knee alignment. However, when the CPAK classification is applied in the Asian population, which is characterized by more varus and wider distribution in lower limb alignment, modifications in the boundaries of arithmetic hip-knee-ankle angle (aHKA) and joint line obliquity (JLO) should be considered. The purposes of this study were as follows: first, to propose a modified CPAK classification based on the actual joint line obliquity (aJLO) and wider range of aHKA in the Asian population; second, to test this classification in a cohort of Asians with healthy knees; third, to propose individualized alignment targets for different CPAK types in kinematically aligned (KA) total knee arthroplasty (TKA).

Methods

The CPAK classification was modified by changing the neutral boundaries of aHKA to 0° ± 3° and using aJLO as a new variable. Radiological analysis of 214 healthy knees in 214 Asian individuals was used to assess the distribution and mean value of alignment angles of each phenotype among different classifications based on the coronal plane. Individualized alignment targets were set according to the mean lateral distal femoral angle (LDFA) and medial proximal tibial angle (MPTA) of different knee types.


Bone & Joint Open
Vol. 2, Issue 12 | Pages 1075 - 1081
17 Dec 2021
Suthar A Yukata K Azuma Y Suetomi Y Yamazaki K Seki K Sakai T Fujii H

Aims

This study aimed to investigate the relationship between changes in patellar height and clinical outcomes at a mean follow-up of 7.7 years (5 to 10) after fixed-bearing posterior-stabilized total knee arthroplasty (PS-TKA).

Methods

We retrospectively evaluated knee radiographs of 165 knees, which underwent fixed-bearing PS-TKA with patella resurfacing. The incidence of patella baja and changes in patellar height over a minimum of five years of follow-up were determined using Insall-Salvati ratio (ISR) measurement. We examined whether patella baja (ISR < 0.8) at final follow-up affected clinical outcomes, knee joint range of motion (ROM), and Knee Society Score (KSS). We also assessed inter- and intrarater reliability of ISR measurements and focused on the relationship between patellar height reduction beyond measurement error and clinical outcomes.


Bone & Joint Open
Vol. 2, Issue 9 | Pages 737 - 744
1 Sep 2021
Øhrn F Lian ØB Tsukanaka M Röhrl SM

Aims

Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA.

Methods

A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m2 (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 235 - 241
1 Feb 2022
Stone B Nugent M Young SW Frampton C Hooper GJ

Aims

The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime risk of revision may be more meaningful to patients when gauging risks, especially in younger patients. We aimed to assess the lifetime risk of revision for patients in different age categories at the time of undergoing primary TKA.

Methods

The New Zealand Joint Registry database was used to obtain revision rates, mortality, and the indications for revision for all primary TKAs performed during an 18-year period between January 1999 and December 2016. Patients were stratified into age groups at the time of the initial TKA, and the lifetime risk of revision was calculated according to age, sex, and the American Society of Anesthesiologists (ASA) grade. The most common indications for revision were also analyzed for each age group.


Aims

Nearly 99,000 total knee arthroplasties (TKAs) are performed in UK annually. Despite plenty of research, the satisfaction rate of this surgery is around 80%. One of the important intraoperative factors affecting the outcome is alignment. The relationship between joint obliquity and functional outcomes is not well understood. Therefore, a study is required to investigate and compare the effects of two types of alignment (mechanical and kinematic) on functional outcomes and range of motion.

Methods

The aim of the study is to compare navigated kinematically aligned TKAs (KA TKAs) with navigated mechanically aligned TKA (MA TKA) in terms of function and ROM. We aim to recruit a total of 96 patients in the trial. The patients will be recruited from clinics of various consultants working in the trust after screening them for eligibility criteria and obtaining their informed consent to participate in this study. Randomization will be done prior to surgery by a software. The primary outcome measure will be the Knee injury and Osteoarthritis Outcome Score The secondary outcome measures include Oxford Knee Score, ROM, EuroQol five-dimension questionnaire, EuroQol visual analogue scale, 12-Item Short-Form Health Survey (SF-12), and Forgotten Joint Score. The scores will be calculated preoperatively and then at six weeks, six months, and one year after surgery. The scores will undergo a statistical analysis.


Bone & Joint Open
Vol. 3, Issue 1 | Pages 35 - 41
9 Jan 2022
Buchalter DB Nduaguba A Teo GM Kugelman D Aggarwal VK Long WJ

Aims

Despite recent literature questioning their use, vancomycin and clindamycin often substitute cefazolin as the preoperative antibiotic prophylaxis in primary total knee arthroplasty (TKA), especially in the setting of documented allergy to penicillin. Topical povidone-iodine lavage and vancomycin powder (VIP) are adjuncts that may further broaden antimicrobial coverage, and have shown some promise in recent investigations. The purpose of this study, therefore, is to compare the risk of acute periprosthetic joint infection (PJI) in primary TKA patients who received cefazolin and VIP to those who received a non-cephalosporin alternative and VIP.

Methods

This was a retrospective cohort study of 11,550 primary TKAs performed at an orthopaedic hospital between 2013 and 2019. The primary outcome was PJI occurring within 90 days of surgery. Patients were stratified into two groups (cefazolin vs non-cephalosporin) based on their preoperative antibiotic. All patients also received the VIP protocol at wound closure. Bivariate and multiple logistic regression analyses were performed to control for potential confounders and identify the odds ratio of PJI.


Bone & Joint Research
Vol. 6, Issue 11 | Pages 623 - 630
1 Nov 2017
Suh D Kang K Son J Kwon O Baek C Koh Y

Objectives. Malalignment of the tibial component could influence the long-term survival of a total knee arthroplasty (TKA). The object of this study was to investigate the biomechanical effect of varus and valgus malalignment on the tibial component under stance-phase gait cycle loading conditions. Methods. Validated finite element models for varus and valgus malalignment by 3° and 5° were developed to evaluate the effect of malalignment on the tibial component in TKA. Maximum contact stress and contact area on a polyethylene insert, maximum contact stress on patellar button and the collateral ligament force were investigated. Results. There was greater total contact stress in the varus alignment than in the valgus, with more marked difference on the medial side. An increase in ligament force was clearly demonstrated, especially in the valgus alignment and force exerted on the medial collateral ligament also increased. Conclusion. These results highlight the importance of accurate surgical reconstruction of the coronal tibial alignment of the knee joint. Varus and valgus alignments will influence wear and ligament stability, respectively in TKA. Cite this article: D-S. Suh, K-T. Kang, J. Son, O-R. Kwon, C. Baek, Y-G. Koh. Computational study on the effect of malalignment of the tibial component on the biomechanics of total knee arthroplasty: A Finite Element Analysis. Bone Joint Res 2017;6:623–630. DOI: 10.1302/2046-3758.611.BJR-2016-0088.R2


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 101 - 104
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

Previous studies of failure mechanisms leading to revision total knee replacement (TKR) performed between 1986 and 2000 determined that many failed early, with a disproportionate amount accounted for by infection and implant-associated factors including wear, loosening and instability. Since then, efforts have been made to improve implant performance and instruct surgeons in best practice. Recently our centre participated in a multi-centre evaluation of 844 revision TKRs from 2010 to 2011. The purpose was to report a detailed analysis of failure mechanisms over time and to see if failure modes have changed over the past 10 to 15 years. Aseptic loosening was the predominant mechanism of failure (31.2%), followed by instability (18.7%), infection (16.2%), polyethylene wear (10.0%), arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to failure was 5.9 years (ten days to 31 years), 35.3% of all revisions occurred at less than two years, and 60.2% in the first five years. With improvements in implant and polyethylene manufacture, polyethylene wear is no longer a leading cause of failure. Early mechanisms of failure are primarily technical errors. In addition to improving implant longevity, industry and surgeons must work together to decrease these technical errors. All reports on failure of TKR contain patients with unexplained pain who not infrequently have unmet expectations. Surgeons must work to achieve realistic patient expectations pre-operatively, and therefore, improve patient satisfaction post-operatively. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):101–4


Bone & Joint Open
Vol. 2, Issue 11 | Pages 921 - 925
9 Nov 2021
Limberg AK Wyles CC Taunton MJ Hanssen AD Pagnano MW Abdel MP

Aims

Varus-valgus constrained (VVC) devices are typically used in revision settings, often with stems to mitigate the risk of aseptic loosening. However, in at least one system, the VVC insert is compatible with the primary posterior-stabilized (PS) femoral component, which may be an option in complex primary situations. We sought to determine the implant survivorship, radiological and clinical outcomes, and complications when this VVC insert was coupled with a PS femur without stems in complex primary total knee arthroplasties (TKAs).

Methods

Through our institution’s total joint registry, we identified 113 primary TKAs (103 patients) performed between 2007 and 2017 in which a VVC insert was coupled with a standard cemented PS femur without stems. Mean age was 68 years (SD 10), mean BMI was 32 kg/m2 (SD 7), and 59 patients (50%) were male. Mean follow-up was four years (2 to 10).


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1578 - 1585
1 Oct 2021
Abram SGF Sabah SA Alvand A Price AJ

Aims

To compare rates of serious adverse events in patients undergoing revision knee arthroplasty with consideration of the indication for revision (urgent versus elective indications), and compare these with primary arthroplasty and re-revision arthroplasty.

Methods

Patients undergoing primary knee arthroplasty were identified in the national Hospital Episode Statistics (HES) between 1 April 1997 to 31 March 2017. Subsequent revision and re-revision arthroplasty procedures in the same patients and same knee were identified. The primary outcome was 90-day mortality and a logistic regression model was used to investigate factors associated with 90-day mortality and secondary adverse outcomes, including infection (undergoing surgery), pulmonary embolism, myocardial infarction, and stroke. Urgent indications for revision arthroplasty were defined as infection or fracture, and all other indications (e.g. loosening, instability, wear) were included in the elective indications cohort.


The Bone & Joint Journal
Vol. 103-B, Issue 9 | Pages 1514 - 1525
1 Sep 2021
Scott CEH Holland G Gillespie M Keenan OJ Gherman A MacDonald DJ Simpson AHRW Clement ND

Aims

The aims of this study were to investigate the ability to kneel after total knee arthroplasty (TKA) without patellar resurfacing, and its effect on patient-reported outcome measures (PROMs). Secondary aims included identifying which kneeling positions were most important to patients, and the influence of radiological parameters on the ability to kneel before and after TKA.

Methods

This prospective longitudinal study involved 209 patients who underwent single radius cruciate-retaining TKA without patellar resurfacing. Preoperative EuroQol five-dimension questionnaire (EQ-5D), Oxford Knee Score (OKS), and the ability to achieve four kneeling positions were assessed including a single leg kneel, a double leg kneel, a high-flexion kneel, and a praying position. The severity of radiological osteoarthritis (OA) was graded and the pattern of OA was recorded intraoperatively. The flexion of the femoral component, posterior condylar offset, and anterior femoral offset were measured radiologically. At two to four years postoperatively, 151 patients with a mean age of 70.0 years (SD 9.44) were included. Their mean BMI was 30.4 kg/m2 (SD 5.36) and 60 were male (40%). They completed EQ-5D, OKS, and Kujala scores, assessments of the ability to kneel, and a visual analogue scale for anterior knee pain and satisfaction.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims. We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty. Materials and Methods. A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared. Results. The loss of polyethylene volume was 122 mm. 3. (standard deviation (. sd. ) 87) in the OxZr group and 170 mm. 3. (. sd. 96) in the CoCr group (p = 0.033). The volume loss in the OxZr group was also lower in the medial (72 mm. 3 . (. sd. 67) versus 92 mm. 3 . (. sd. 60); p = 0.096) and lateral (49 mm. 3 . (. sd. 36) versus 79 mm. 3 . (. sd. 61); p = 0.096) compartments separately, but these differences were not significant. Conclusion. Our results corroborate earlier findings from in vitro testing and visual retrieval analysis which suggest that polyethylene volume loss is lower with OxZr femoral components. Since both OxZr and CoCr are hard surfaces that would be expected to create comparable amounts of polyethylene creep, the differences in volume loss may reflect differences in the in vivo wear of these inserts. Cite this article: Bone Joint J 2017;99-B:793–8


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1367 - 1372
1 Aug 2021
Plancher KD Brite JE Briggs KK Petterson SC

Aims

The patient-acceptable symptom state (PASS) is a level of wellbeing, which is measured by the patient. The aim of this study was to determine if the proportion of patients who achieved an acceptable level of function (PASS) after medial unicompartmental knee arthroplasty (UKA) was different based on the status of the anterior cruciate ligament (ACL) at the time of surgery.

Methods

A total of 114 patients who underwent UKA for isolated medial osteoarthritis (OA) of the knee were included in the study. Their mean age was 65 years (SD 10). No patient underwent a bilateral procedure. Those who had undergone ACL reconstruction during the previous five years were excluded. The Knee injury Osteoarthritis Outcome Score Activities of Daily Living (KOOS ADL) function score was used as the primary outcome measure with a PASS of 87.5, as described for total knee arthroplasty (TKA). Patients completed all other KOOS subscales, Lysholm score, the Western Ontario and McMaster Universities Osteoarthritis Index, and the Veterans Rand 12-item health survey score. Failure was defined as conversion to TKA.


The Bone & Joint Journal
Vol. 103-B, Issue 9 | Pages 1505 - 1513
1 Sep 2021
Stockton DJ Schmidt AM Yung A Desrochers J Zhang H Masri BA Wilson DR

Aims

Anterior cruciate ligament (ACL) rupture commonly leads to post-traumatic osteoarthritis, regardless of surgical reconstruction. This study uses standing MRI to investigate changes in contact area, contact centroid location, and tibiofemoral alignment between ACL-injured knees and healthy controls, to examine the effect of ACL reconstruction on these parameters.

Methods

An upright, open MRI was used to directly measure tibiofemoral contact area, centroid location, and alignment in 18 individuals with unilateral ACL rupture within the last five years. Eight participants had been treated nonoperatively and ten had ACL reconstruction performed within one year of injury. All participants were high-functioning and had returned to sport or recreational activities. Healthy contralateral knees served as controls. Participants were imaged in a standing posture with knees fully extended.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 145 - 149
1 Jun 2021
Crawford DA Passias BJ Adams JB Berend KR Lombardi AV

Aims

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.


Bone & Joint Open
Vol. 2, Issue 8 | Pages 576 - 582
2 Aug 2021
Fuchs M Kirchhoff F Reichel H Perka C Faschingbauer M Gwinner C

Aims

Current guidelines consider analyses of joint aspirates, including leucocyte cell count (LC) and polymorphonuclear percentage (PMN%) as a diagnostic mainstay of periprosthetic joint infection (PJI). It is unclear if these parameters are subject to a certain degree of variability over time. Therefore, the aim of this study was to evaluate the variation of LC and PMN% in patients with aseptic revision total knee arthroplasty (TKA).

Methods

We conducted a prospective, double-centre study of 40 patients with 40 knee joints. Patients underwent joint aspiration at two different time points with a maximum period of 120 days in between these interventions and without any events such as other joint aspirations or surgeries. The main indications for TKA revision surgery were aseptic implant loosening (n = 24) and joint instability (n = 11).


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 450 - 454
1 Apr 2018
Chalmers BP Mehrotra KG Sierra RJ Pagnano MW Taunton MJ Abdel MP

Aims. Primary (or spontaneous) and secondary osteonecrosis of the knee can lead to severe joint degeneration, for which either total or unicompartmental arthroplasty may be considered. However, there are limited studies analyzing outcomes of unicompartmental knee arthroplasties (UKAs) for osteonecrosis involving an isolated compartment of the knee. The aims of this study were to analyze outcomes of UKAs for osteonecrosis with specific focus on 1) survivorship free of any revision or reoperation, 2) risk factors for failure, 3) clinical outcomes, and 4) complications. Patients and Methods. A total of 45 patients underwent 46 UKAs for knee osteonecrosis between 2002 and 2014 at our institution (The Mayo Clinic, Rochester, Minnesota). Twenty patients (44%) were female; the mean age of the patients was 66 years, and mean body mass index (BMI) was 31 kg/m. 2. Of the 46 UKAs, 44 (96%) were medial UKAs, and 35 (76%) were fixed-bearing design. Mean mechanical axis postoperatively was 1.5° varus (0° to 5° varus); 41 UKAs (89%) were performed for primary osteonecrosis. Mean follow-up was five years (2 to 12). Results. Survivorship free of any revision in the cohort was 89% (95% CI 77 to 99) and 76% (95% CI 53 to 99) at five and ten years, respectively. In patients undergoing UKA for primary osteonecrosis survivorship free of any revision was 93% (95% CI 83 to 100)at both five and ten years. Secondary osteonecrosis was a significant risk factor for poorer survivorship free of any revision or reoperation (hazard ratio 7.7, p = 0.03). Three medial UKAs (6.5%) were converted to total knee arthroplasties (TKAs): two for lateral compartment degeneration and one for development of lateral osteonecrosis. No implants were revised for loosening, fracture, or wear. Knee Society scores improved from a mean of 60 (44 to 72) preoperatively to a mean of 94 postoperatively (82 to 100) (p < 0.001). There were no surgical complications. Conclusion. When done for primary osteonecrosis of the knee, UKA resulted in reliable clinical improvement, minimal complications, and durable estimated implant survivorship free of revision at ten years. UKA done for secondary osteonecrosis was substantially less durable at mid-term follow-up. Progression of knee degeneration, rather than implant failure or loosening, was most common indication for conversion to TKA. Cite this article: Bone Joint J 2018;100-B:450–4


Bone & Joint Open
Vol. 2, Issue 8 | Pages 638 - 645
1 Aug 2021
Garner AJ Edwards TC Liddle AD Jones GG Cobb JP

Aims

Joint registries classify all further arthroplasty procedures to a knee with an existing partial arthroplasty as revision surgery, regardless of the actual procedure performed. Relatively minor procedures, including bearing exchanges, are classified in the same way as major operations requiring augments and stems. A new classification system is proposed to acknowledge and describe the detail of these procedures, which has implications for risk, recovery, and health economics.

Methods

Classification categories were proposed by a surgical consensus group, then ranked by patients, according to perceived invasiveness and implications for recovery. In round one, 26 revision cases were classified by the consensus group. Results were tested for inter-rater reliability. In round two, four additional cases were added for clarity. Round three repeated the survey one month later, subject to inter- and intrarater reliability testing. In round four, five additional expert partial knee arthroplasty surgeons were asked to classify the 30 cases according to the proposed revision partial knee classification (RPKC) system.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 131 - 136
1 Jun 2021
Roof MA Sharan M Merkow D Feng JE Long WJ Schwarzkopf RS

Aims

It has previously been shown that higher-volume hospitals have better outcomes following revision total knee arthroplasty (rTKA). We were unable to identify any studies which investigated the effect of surgeon volume on the outcome of rTKA. We sought to investigate whether patients of high-volume (HV) rTKA surgeons have better outcomes following this procedure compared with those of low-volume (LV) surgeons.

Methods

This retrospective study involved patients who underwent aseptic unilateral rTKA between January 2016 and March 2019, using the database of a large urban academic medical centre. Surgeons who performed ≥ 19 aseptic rTKAs per year during the study period were considered HV and those who performed < 19 per year were considered LV. Demographic characteristics, surgical factors, and postoperative outcomes were compared between the two groups.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 51 - 58
1 Jun 2021
Yang J Heckmann ND Nahhas CR Salzano MB Ruzich GP Jacobs JJ Paprosky WG Rosenberg AG Nam D

Aims

Recent total knee arthroplasty (TKA) designs have featured more anatomical morphologies and shorter tibial keels. However, several reports have raised concerns about the impact of these modifications on implant longevity. The aim of this study was to report the early performance of a modern, cemented TKA design.

Methods

All patients who received a primary, cemented TKA between 2012 and 2017 with a minimum two-year follow-up were included. The implant investigated features an asymmetrical tibial baseplate and shortened keel. Patient demographic details, Knee Society Scores (KSS), component alignment, and the presence of radiolucent lines at final follow-up were recorded. Kaplan-Meier analyses were performed to estimate survivorship.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1358 - 1366
2 Aug 2021
Wei C Quan T Wang KY Gu A Fassihi SC Kahlenberg CA Malahias M Liu J Thakkar S Gonzalez Della Valle A Sculco PK

Aims

This study used an artificial neural network (ANN) model to determine the most important pre- and perioperative variables to predict same-day discharge in patients undergoing total knee arthroplasty (TKA).

Methods

Data for this study were collected from the National Surgery Quality Improvement Program (NSQIP) database from the year 2018. Patients who received a primary, elective, unilateral TKA with a diagnosis of primary osteoarthritis were included. Demographic, preoperative, and intraoperative variables were analyzed. The ANN model was compared to a logistic regression model, which is a conventional machine-learning algorithm. Variables collected from 28,742 patients were analyzed based on their contribution to hospital length of stay.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 351 - 356
1 Mar 2009
Parratte S Argenson JA Pearce O Pauly V Auquier P Aubaniac J

We retrospectively reviewed 35 cemented unicompartmental knee replacements performed for medial unicompartmental osteoarthritis of the knee in 31 patients ≤50 years old (mean 46, 31 to 49). Patients were assessed clinically and radiologically using the Knee Society scores at a mean follow-up of 9.7 years (5 to 16) and survival at 12 years was calculated. The mean Knee Society Function Score improved from 54 points (25 to 64) pre-operatively to 89 (80 to 100) post-operatively (p < 0.0001). Six knees required revision, four for polyethylene wear treated with an isolated exchange of the tibial insert, one for aseptic loosening and one for progression of osteoarthritis. The 12-year survival according to Kaplan-Meier was 80.6% with revision for any reason as the endpoint. Despite encouraging clinical results, polyethylene wear remains a major concern affecting the survival of unicompartmental knee replacement in patients younger than 50


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1261 - 1269
1 Jul 2021
Burger JA Zuiderbaan HA Sierevelt IN van Steenbergen L Nolte PA Pearle AD Kerkhoffs GMMJ

Aims

Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs.

Methods

A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 74 - 80
1 Jun 2021
Deckey DG Rosenow CS Verhey JT Brinkman JC Mayfield CK Clarke HD Bingham JS

Aims

Robotic-assisted total knee arthroplasty (RA-TKA) is theoretically more accurate for component positioning than TKA performed with mechanical instruments (M-TKA). Furthermore, the ability to incorporate soft-tissue laxity data into the plan prior to bone resection should reduce variability between the planned polyethylene thickness and the final implanted polyethylene. The purpose of this study was to compare accuracy to plan for component positioning and precision, as demonstrated by deviation from plan for polyethylene insert thickness in measured-resection RA-TKA versus M-TKA.

Methods

A total of 220 consecutive primary TKAs between May 2016 and November 2018, performed by a single surgeon, were reviewed. Planned coronal plane component alignment and overall limb alignment were all 0° to the mechanical axis; tibial posterior slope was 2°; and polyethylene thickness was 9 mm. For RA-TKA, individual component position was adjusted to assist gap-balancing but planned coronal plane alignment for the femoral and tibial components and overall limb alignment remained 0 ± 3°; planned tibial posterior slope was 1.5°. Mean deviations from plan for each parameter were compared between groups for positioning and size and outliers were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 137 - 144
1 Jun 2021
Lachiewicz PF Steele JR Wellman SS

Aims

To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period.

Methods

This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 59 - 66
1 Jun 2021
Abhari S Hsing TM Malkani MM Smith AF Smith LS Mont MA Malkani AL

Aims

Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment.

Methods

A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 9 | Pages 1158 - 1163
1 Sep 2006
Tai CC Cross MJ

We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 864 - 871
3 May 2021
Hunt LP Matharu GS Blom AW Howard PW Wilkinson JM Whitehouse MR

Aims

Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing.

Methods

A retrospective observational study of the National Joint Registry (NJR) was performed. All primary TKRs for osteoarthritis alone performed between 1 April 2003 and 31 December 2016 were eligible (n = 842,072). Patellar resurfacing during TKR was performed in 36% (n = 305,844). The primary outcome was all-cause revision surgery. Secondary outcomes were the number of excess all-cause revisions associated with using TKRs without (versus with) patellar resurfacing, and the risk of re-revision after secondary patellar resurfacing.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 610 - 618
1 Apr 2021
Batailler C Bordes M Lording T Nigues A Servien E Calliess T Lustig S

Aims

Ideal component sizing may be difficult to achieve in unicompartmental knee arthroplasty (UKA). Anatomical variants, incremental implant size, and a reduced surgical exposure may lead to over- or under-sizing of the components. The purpose of this study was to compare the accuracy of UKA sizing with robotic-assisted techniques versus a conventional surgical technique.

Methods

Three groups of 93 medial UKAs were assessed. The first group was performed by a conventional technique, the second group with an image-free robotic-assisted system (Image-Free group), and the last group with an image-based robotic arm-assisted system, using a preoperative CT scan (Image-Based group). There were no demographic differences between groups. We compared six parameters on postoperative radiographs to assess UKA sizing. Incorrect sizing was defined by an over- or under-sizing greater than 3 mm.


Bone & Joint Research
Vol. 3, Issue 10 | Pages 289 - 296
1 Oct 2014
van IJsseldijk EA Harman MK Luetzner J Valstar ER Stoel BC Nelissen RGHH Kaptein BL

Introduction. Wear of polyethylene inserts plays an important role in failure of total knee replacement and can be monitored in vivo by measuring the minimum joint space width in anteroposterior radiographs. The objective of this retrospective cross-sectional study was to compare the accuracy and precision of a new model-based method with the conventional method by analysing the difference between the minimum joint space width measurements and the actual thickness of retrieved polyethylene tibial inserts. . Method. Before revision, the minimum joint space width values and their locations on the insert were measured in 15 fully weight-bearing radiographs. These measurements were compared with the actual minimum thickness values and locations of the retrieved tibial inserts after revision. . Results. The mean error in the model-based minimum joint space width measurement was significantly smaller than the conventional method for medial condyles (0.50 vs 0.94 mm, p < 0.01) and for lateral condyles (0.06 vs 0.34 mm, p = 0.02). The precision (standard deviation of the error) of the methods was similar (0.84 vs 0.79 mm medially and both 0.46 mm laterally). The distance between the true minimum joint space width locations and the locations from the model-based measurements was less than 10 mm in the medial direction in 12 cases and less in the lateral direction in 13 cases. Conclusion. The model-based minimum joint space width measurement method is more accurate than the conventional measurement with the same precision. Cite this article: Bone Joint Res 2014;3:289–96