Bone & Joint Open
Vol. 5, Issue 5 | Pages 394 - 400
15 May 2024
Nishi M Atsumi T Yoshikawa Y Okano I Nakanishi R Watanabe M Usui Y Kudo Y

Aims. The localization of necrotic areas has been reported to impact the prognosis and treatment strategy for osteonecrosis of the femoral head (ONFH). Anteroposterior localization of the necrotic area after a femoral neck fracture (FNF) has not been properly investigated. We hypothesize that the change of the weight loading direction on the femoral head due to residual posterior tilt caused by malunited FNF may affect the location of ONFH. We investigate the relationship between the posterior tilt angle (PTA) and anteroposterior localization of osteonecrosis using lateral hip radiographs. Methods. Patients aged younger than 55 years diagnosed with ONFH after FNF were retrospectively reviewed. Overall, 65 hips (38 males and 27 females; mean age 32.6 years (SD 12.2)) met the inclusion criteria. Patients with stage 1 or 4 ONFH, as per the Association Research Circulation Osseous classification, were excluded. The ratios of anterior and posterior viable areas and necrotic areas of the femoral head to the articular surface were calculated by setting the femoral head centre as the reference point. The PTA was measured using Palm’s method. The association between the PTA and viable or necrotic areas of the femoral head was assessed using Spearman’s rank correlation analysis (median PTA 6.0° (interquartile range 3 to 11.5)). Results. We identified a negative correlation between PTA and anterior viable areas (rho −0.477; p = 0.001), and no correlation between PTA and necrotic (rho 0.229; p = 0.067) or posterior viable areas (rho 0.204; p = 0.132). Conclusion. Our results suggest that residual posterior tilt after FNF could affect the anteroposterior localization of necrosis. Cite this article: Bone Jt Open 2024;5(5):394–400


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 891 - 897
1 Jul 2018
Teeter MG Lanting BA Naudie DD McCalden RW Howard JL MacDonald SJ

Aims. The aim of this study was to determine whether there is a difference in the rate of wear between acetabular components positioned within and outside the ‘safe zones’ of anteversion and inclination angle. Patients and Methods. We reviewed 100 hips in 94 patients who had undergone primary total hip arthroplasty (THA) at least ten years previously. Patients all had the same type of acetabular component with a bearing couple which consisted of a 28 mm cobalt-chromium head on a highly crosslinked polyethylene (HXLPE) liner. A supine radiostereometric analysis (RSA) examination was carried out which acquired anteroposterior (AP) and lateral paired images. Acetabular component anteversion and inclination angles were measured as well as total femoral head penetration, which was divided by the length of implantation to determine the rate of polyethylene wear. Results. The mean anteversion angle was 19.4° (-15.2° to 48°, . sd. 11.4°), the mean inclination angle 43.4° (27.3° to 60.5°, . sd. 6.6°), and the mean wear rate 0.055 mm/year (. sd. 0.060). Exactly half of the hips were positioned inside the ‘safe zone’. There was no difference (median difference, 0.012 mm/year; p = 0.091) in the rate of wear between acetabular components located within or outside the ‘safe zone’. When compared to acetabular components located inside the ‘safe zone’, the wear rate was no different for acetabular components that only achieved the target anteversion angle (median difference, 0.012 mm/year; p = 0.138), target inclination angle (median difference, 0.013 mm/year; p = 0.354), or neither target (median difference, 0.012 mm/year; p = 0.322). Conclusion. Placing the acetabular component within or outside the ‘safe zone’ did not alter the wear rate of HXLPE at long-term follow-up to a level that risked osteolysis. HXLPE appears to be a forgiving bearing material in terms of articular surface wear, but care must still be taken to position the acetabular component correctly so that the implant is stable. Cite this article: Bone Joint J 2018;100-B:891-7


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 881 - 885
1 Jul 2011
Cobb JP Davda K Ahmad A Harris SJ Masjedi M Hart AJ

Large-head metal-on-metal total hip replacement has a failure rate of almost 8% at five years, three times the revision rate of conventional hip replacement. Unexplained pain remains a feature of this type of arthroplasty. All designs of the femoral component of large-head metal-on-metal total hip replacements share a unique characteristic: a subtended angle of 120° defining the proportion of a sphere that the head represents. Using MRI, we measured the contact area of the iliopsoas tendon on the femoral head in sagittal reconstruction of 20 hips of patients with symptomatic femoroacetabular impingement. We also measured the articular extent of the femoral head on 40 normal hips and ten with cam-type deformities. Finally, we performed virtual hip resurfacing on normal and cam-type hips, avoiding overhang of the metal rim inferomedially. The articular surface of the femoral head has a subtended angle of 120° anteriorly and posteriorly, but only 100° medially. Virtual surgery in a normally shaped femoral head showed a 20° skirt of metal protruding medially where iliopsoas articulates. The excessive extent of the large-diameter femoral components may cause iliopsoas impingement independently of the acetabular component. This may be the cause of postoperative pain with these implants


Bone & Joint Open
Vol. 4, Issue 7 | Pages 523 - 531
11 Jul 2023
Passaplan C Hanauer M Gautier L Stetzelberger VM Schwab JM Tannast M Gautier E

Aims

Hyaline cartilage has a low capacity for regeneration. Untreated osteochondral lesions of the femoral head can lead to progressive and symptomatic osteoarthritis of the hip. The purpose of this study is to analyze the clinical and radiological long-term outcome of patients treated with osteochondral autograft transfer. To our knowledge, this study represents a series of osteochondral autograft transfer of the hip with the longest follow-up.

Methods

We retrospectively evaluated 11 hips in 11 patients who underwent osteochondral autograft transfer in our institution between 1996 and 2012. The mean age at the time of surgery was 28.6 years (8 to 45). Outcome measurement included standardized scores and conventional radiographs. Kaplan-Meier survival curve was used to determine the failure of the procedures, with conversion to total hip arthroplasty (THA) defined as the endpoint.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 79 - 86
1 Feb 2024
Sato R Hamada H Uemura K Takashima K Ando W Takao M Saito M Sugano N

Aims

This study aimed to investigate the incidence of ≥ 5 mm asymmetry in lower and whole leg lengths (LLs) in patients with unilateral osteoarthritis (OA) secondary to developmental dysplasia of the hip (DDH-OA) and primary hip osteoarthritis (PHOA), and the relationship between lower and whole LL asymmetries and femoral length asymmetry.

Methods

In total, 116 patients who underwent unilateral total hip arthroplasty were included in this study. Of these, 93 had DDH-OA and 23 had PHOA. Patients with DDH-OA were categorized into three groups: Crowe grade I, II/III, and IV. Anatomical femoral length, femoral length greater trochanter (GT), femoral length lesser trochanter (LT), tibial length, foot height, lower LL, and whole LL were evaluated using preoperative CT data of the whole leg in the supine position. Asymmetry was evaluated in the Crowe I, II/III, IV, and PHOA groups.


Bone & Joint Research
Vol. 1, Issue 3 | Pages 25 - 30
1 Mar 2012
Wroblewski BM Siney PD Fleming PA

Objectives. Metal-on-metal (MoM) hip resurfacing was introduced into clinical practice because it was perceived to be a better alternative to conventional total hip replacement for young and active patients. However, an increasing number of reports of complications have arisen focusing on design and orientation of the components, the generation of metallic wear particles and serum levels of metallic ions. The procedure introduced a combination of two elements: large-dimension components and hard abrasive particles of metal wear. The objective of our study was to investigate the theory that microseparation of the articular surfaces draws in a high volume of bursal fluid and its contents into the articulation, and at relocation under load would generate high pressures of fluid ejection, resulting in an abrasive water jet. Methods. This theoretical concept using MoM resurfacing components (head diameter 55 mm) was modelled mathematically and confirmed experimentally using a material-testing machine that pushed the head into the cup at a rate of 1000 mm/min until fully engaged. Results. The mathematical model showed the pattern but not the force of fluid ejection, the highest pressures were expected when the separation of the components was only a fraction of one millimetre. The experimental work confirmed the results; with the mean peak ejection pressure of 43 763 N/m. 2. equivalent to 306 mmHg or 5 psi. Conclusions. The mechanical effect of the high-pressure abrasive water jet is the likely cause of the spectrum of complications reported with metal-on-metal resurfacing. Investigating serum levels of metallic elements may not be the best method for assessing the local mechanical effects of the abrasive water jet


Bone & Joint Open
Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


Bone & Joint Open
Vol. 4, Issue 11 | Pages 853 - 858
10 Nov 2023
Subbiah Ponniah H Logishetty K Edwards TC Singer GC

Aims

Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon.

Methods

A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.


Bone & Joint Open
Vol. 4, Issue 3 | Pages 182 - 187
14 Mar 2023
Sheridan GA Hanlon M Welch-Phillips A Spratt K Hagan R O'Byrne JM Kenny PJ Kurmis AP Masri BA Garbuz DS Hurson CJ

Aims

Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings.

Methods

A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1037 - 1043
1 Aug 2009
Krieg AH Speth BM Ochsner PE

Polyethylene wear of acetabular components is a key factor in the development of periprosthetic osteolysis and wear at the articular surface has been well documented and quantified, but fewer data are available about changes which occur at the backside of the liner. At revision surgery for loosening of the femoral component we retrieved 35 conventional modular acetabular liners of the same design. Linear and volumetric articular wear, backside volumetric change and the volume of the screw-head indentations were quantified. These volumes, clinical data and the results from radiological Ein Bild Röntgen Analyse migration analysis were used to identify potential factors influencing the volumetric articular wear and backside volumetric change. The rate of backside volumetric change was found to be 2.8% of the rate of volumetric articular wear and decreased with increasing liner size. Migrated acetabular components showed significantly higher rates of backside volumetric change plus screw-head indentations than those without migration. The backside volumetric change was at least ten times larger than finite-element simulation had suggested. In a stable acetabular component with well-anchored screws, the amount of backside wear should not cause clinical problems. Impingement of the screw-heads could produce more wear particles than those generated at the liner-shell interface. Because the rate of backside volumetric change is only 2.8% of the rate of volumetric articular wear and since creep is likely to contribute a significant portion to this, the debris generated by wear at the backside of the liner may not be sufficient to create a strong osteolytic response


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1344 - 1350
1 Oct 2012
Penny JO Ding M Varmarken JE Ovesen O Overgaard S

Radiostereometric analysis (RSA) can detect early micromovement in unstable implant designs which are likely subsequently to have a high failure rate. In 2010, the Articular Surface Replacement (ASR) was withdrawn because of a high failure rate. In 19 ASR femoral components, the mean micromovement over the first two years after implantation was 0.107 mm (. sd. 0.513) laterally, 0.055 mm (. sd. 0.204) distally and 0.150 mm (. sd. 0.413) anteriorly. The mean backward tilt around the x-axis was -0.08° (. sd. 1.088), mean internal rotation was 0.165° (. sd. 0.924) and mean varus tilt 0.238° (. sd. 0.420). The baseline to two-year varus tilt was statistically significant from zero movement, but there was no significant movement from one year onwards. We conclude that the ASR femoral component achieves initial stability and that early migration is not the mode of failure for this resurfacing arthroplasty


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1033 - 1044
1 Aug 2011
Periasamy K Watson WS Mohammed A Murray H Walker B Patil S Meek RMD

The ideal acetabular component is characterised by reliable, long-term fixation with physiological loading of bone and a low rate of wear. Trabecular metal is a porous construct of tantalum which promotes bony ingrowth, has a modulus of elasticity similar to that of cancellous bone, and should be an excellent material for fixation. Between 2004 and 2006, 55 patients were randomised to receive either a cemented polyethylene or a monobloc trabecular metal acetabular component with a polyethylene articular surface. We measured the peri-prosthetic bone density around the acetabular components for up to two years using dual-energy x-ray absorptiometry. We found evidence that the cemented acetabular component loaded the acetabular bone centromedially whereas the trabecular metal monobloc loaded the lateral rim and behaved like a hemispherical rigid metal component with regard to loading of the acetabular bone. We suspect that this was due to the peripheral titanium rim used for the mechanism of insertion


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1220 - 1221
1 Sep 2005
Wroblewski BM Siney PD Fleming PA

We report the results of our continued review of 11 total hip arthroplasties using 22.225 mm alumina ceramic femoral heads on a Charnley flanged stem, articulating with chemically cross-linked polyethylene. There was an initial bedding-in of up to 0.41 mm at the articular surface in the first two years. This had not progressed further, at a minimum follow-up of 15 years. Radiographically no femoral or acetabular component showed loosening or osteolysis


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 4 | Pages 512 - 516
1 May 2000
Miyanishi K Noguchi Y Yamamoto T Irisa T Suenaga E Jingushi S Sugioka Y Iwamoto Y

We have studied the correlation between the prevention of progressive collapse and the ratio of the intact articular surface of the femoral head, after transtrochanteric rotational osteotomy for osteonecrosis. We used probit analysis on 125 hips in order to assess the ratio necessary to prevent progressive radiological collapse over a ten-year period. The results show that a minimum postoperative intact ratio of 34% was required. This critical ratio may be useful for surgical planning and in assessing the natural history of the condition


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 281 - 288
1 Mar 1999
Reynolds D Lucas J Klaue K

We describe a little-known variety of hip dysplasia, termed ‘acetabular retroversion’, in which the alignment of the mouth of the acetabulum does not face the normal anterolateral direction, but inclines more posterolaterally. The condition may be part of a complex dysplasia or a single entity. Other than its retroversion, the acetabulum is sited normally on the side wall of the pelvis, and its articular surface is of normal extent and configuration. The retroverted orientation may give rise to problems of impingement between the femoral neck and anterior acetabular edge. We define the clinical and radiological parameters and discuss pathological changes which may occur in the untreated condition. A technique of management is proposed


Bone & Joint Research
Vol. 10, Issue 12 | Pages 830 - 839
15 Dec 2021
Robertson G Wallace R Simpson AHRW Dawson SP

Aims

Assessment of bone mineral density (BMD) with dual-energy X-ray absorptiometry (DXA) is a well-established clinical technique, but it is not available in the acute trauma setting. Thus, it cannot provide a preoperative estimation of BMD to help guide the technique of fracture fixation. Alternative methods that have been suggested for assessing BMD include: 1) cortical measures, such as cortical ratios and combined cortical scores; and 2) aluminium grading systems from preoperative digital radiographs. However, limited research has been performed in this area to validate the different methods. The aim of this study was to investigate the evaluation of BMD from digital radiographs by comparing various methods against DXA scanning.

Methods

A total of 54 patients with distal radial fractures were included in the study. Each underwent posteroanterior (PA) and lateral radiographs of the injured wrist with an aluminium step wedge. Overall 27 patients underwent routine DXA scanning of the hip and lumbar spine, with 13 undergoing additional DXA scanning of the uninjured forearm. Analysis of radiographs was performed on ImageJ and Matlab with calculations of cortical measures, cortical indices, combined cortical scores, and aluminium equivalent grading.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 767 - 774
1 Jul 2022
Nakashima Y Ishibashi S Kitamura K Yamate S Motomura G Hamai S Ikemura S Fujii M

Aims

Although periacetabular osteotomies are widely used for the treatment of symptomatic dysplastic hips, long-term surgical outcomes and patient-reported outcome measures (PROMs) are still unclear. Accordingly, we assessed hip survival and PROMs at 20 years after transpositional osteotomy of the acetabulum (TOA).

Methods

A total of 172 hips in 159 patients who underwent TOA were followed up at a mean of 21.02 years (16.6 to 24.6) postoperatively. Kaplan-Meier analysis was used to assess survivorship with an endpoint of total hip arthroplasty (THA). PROMs included the visual analogue scale (VAS) Satisfaction, VAS Pain, Oxford Hip Score (OHS), and Forgotten Joint Score-12 (FJS-12). Thresholds for favourable outcomes for OHS (≥ 42) and FJS-12 (≥ 51) were obtained using the receiver operating characteristic curve with VAS Satisfaction ≥ 50 and VAS Pain < 20 as anchors.


Bone & Joint Open
Vol. 2, Issue 11 | Pages 1004 - 1016
26 Nov 2021
Wight CM Whyne CM Bogoch ER Zdero R Chapman RM van Citters DW Walsh WR Schemitsch E

Aims

This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.

Methods

In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 665 - 671
1 Apr 2021
Osawa Y Seki T Okura T Takegami Y Ishiguro N Hasegawa Y

Aims

We compared the clinical outcomes of curved intertrochanteric varus osteotomy (CVO) with bone impaction grafting (BIG) with CVO alone for the treatment of osteonecrosis of the femoral head (ONFH).

Methods

This retrospective comparative study included 81 patients with ONFH; 37 patients (40 hips) underwent CVO with BIG (BIG group) and 44 patients (47 hips) underwent CVO alone (CVO group). Patients in the BIG group were followed-up for a mean of 12.2 years (10.0 to 16.5). Patients in the CVO group were followed-up for a mean of 14.5 years (10.0 to 21.0). Assessment parameters included the Harris Hip Score (HHS), Oxford Hip Score (OHS), Japanese Orthopaedic Association Hip-Disease Evaluation Questionnaire (JHEQ), complication rates, and survival rates, with conversion to total hip arthroplasty (THA) and radiological failure as the endpoints.


The Bone & Joint Journal
Vol. 103-B, Issue 9 | Pages 1488 - 1496
1 Sep 2021
Emara AK Zhou G Klika AK Koroukian SM Schiltz NK Higuera-Rueda CA Molloy RM Piuzzi NS

Aims

The current study aimed to compare robotic arm-assisted (RA-THA), computer-assisted (CA-THA), and manual (M-THA) total hip arthroplasty regarding in-hospital metrics including length of stay (LOS), discharge disposition, in-hospital complications, and cost of RA-THA versus M-THA and CA-THA versus M-THA, as well as trends in use and uptake over a ten-year period, and future projections of uptake and use of RA-THA and CA-THA.

Methods

The National Inpatient Sample was queried for primary THAs (2008 to 2017) which were categorized into RA-THA, CA-THA, and M-THA. Past and projected use, demographic characteristics distribution, income, type of insurance, location, and healthcare setting were compared among the three cohorts. In-hospital complications, LOS, discharge disposition, and in-hospital costs were compared between propensity score-matched cohorts of M-THA versus RA-THA and M-THA versus CA-THA to adjust for baseline characteristics and comorbidities.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 111 - 115
1 Jul 2021
Vakharia RM Mannino A Salem HS Roche MW Wong CHJ Mont MA

Aims

Although there is increasing legalization of the use of cannabis in the USA, few well-powered studies have evaluated the association between cannabis use disorder and outcomes following primary total hip arthroplasty (THA). Thus, the aim of this study was to determine whether patients who use cannabis and undergo primary THA have higher rates of in-hospital length of stay (LOS), medical complications, implant-related complications, and costs.

Methods

Using an administrative database, patients with cannabis use disorder undergoing primary THA were matched to a control group in a 1:5 ratio by age, sex, and various medical comorbidities. This yielded 23,030 patients (3,842 in the study group matched with 19,188 in the control group). The variables which were studied included LOS, 90-day medical complications, two-year implant-related complications, and 90-day costs of care. Mann-Whitney U tests were used to compare LOS and costs. Multivariate logistic regression analyses were used to calculate the odds ratios (ORs) of developing complications.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1231 - 1237
1 Jul 2021
Manninen E Lainiala O Karsikas M Reito A Jämsä P Eskelinen A

Aims

To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood.

Methods

Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 99 - 104
1 Jul 2020
Shah RF Bini S Vail T

Aims

Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction.

Methods

A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity).


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2020
Sodhi N Acuna A Etcheson J Mohamed N Davila I Ehiorobo JO Jones LC Delanois RE Mont MA

Aims

Earlier studies dealing with trends in the management of osteonecrosis of the femoral head (ONFH) identified an increasing rate of total hip arthroplasties (THAs) and a decreasing rate of joint-preserving procedures between 1992 and 2008. In an effort to assess new trends in the management of this condition, this study evaluated the annual trends of joint-preserving versus arthroplasties for patients aged < or > 50 years old, and the incidence of specific operative management techniques.

Methods

A total of 219,371 patients with ONFH were identified from a nationwide database between 1 January 2009 and 31 December 2015. The mean age was 54 years (18 to 90) and 105,298 (48%) were female. The diagnosis was made using International Classification of Disease, Ninth revision, Clinical Modification (ICD-9-CM) and Tenth Revision, Clinical Modification (ICD-10-CM) procedure codes. The percentage of patients managed using each procedure during each year was calculated and compared between years. The trends in the use of the types of procedure were also evaluated.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 432 - 439
1 Apr 2017
Weinberg DS Williamson DFK Millis MB Liu RW

Aims

Recently, there has been considerable interest in quantifying the associations between bony abnormalities around and in the hip joint and osteoarthritis (OA). Our aim was to investigate the relationships between acetabular undercoverage, acetabular overcoverage, and femoroacetabular impingement (FAI) with OA of the hip, which currently remain controversial.

Materials and Methods

A total of 545 cadaveric skeletons (1090 hips) from the Hamann-Todd osteological collection were obtained. Femoral head volume (FHV), acetabular volume (AV), the FHV/AV ratio, acetabular version, alpha angle and anterior femoral neck offset (AFNO) were measured. A validated grading system was used to quantify OA of the hip as minimal, moderate, or severe. Multiple linear and multinomial logistic regression were used to determine the factors that correlated independently with the FHV, AV, and the FHV/AV ratio.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims

There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs.

Patients and Methods

We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 712 - 719
1 Jun 2018
Batailler C Weidner J Wyatt M Dalmay F Beck M

Aims

The primary aim of this study was to define and quantify three new measurements to indicate the position of the greater trochanter. Secondary aims were to define ‘functional antetorsion’ as it relates to abductor function in populations both with and without torsional abnormality.

Patients and Methods

Three new measurements, functional antetorsion, posterior tilt, and posterior translation of the greater trochanter, were assessed from 61 CT scans of cadaveric femurs, and their reliability determined. These measurements and their relationships were also evaluated in three groups of patients: a control group (n = 22), a ‘high-antetorsion’ group (n = 22) and a ‘low-antetorsion’ group (n = 10).


Bone & Joint Research
Vol. 5, Issue 2 | Pages 66 - 72
1 Feb 2016
Gebhart JJ Weinberg DS Bohl MS Liu RW

Objectives

Sagittal alignment of the lumbosacral spine, and specifically pelvic incidence (PI), has been implicated in the development of spine pathology, but generally ignored with regards to diseases of the hip. We aimed to determine if increased PI is correlated with higher rates of hip osteoarthritis (HOA). The effect of PI on the development of knee osteoarthritis (KOA) was used as a negative control.

Methods

We studied 400 well-preserved cadaveric skeletons ranging from 50 to 79 years of age at death. Each specimen’s OA of the hip and knee were graded using a previously described method. PI was measured from standardised lateral photographs of reconstructed pelvises. Multiple regression analysis was performed to determine the relationship between age and PI with HOA and KOA.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 730 - 735
1 Jun 2016
Bsat S Frei H Beaulé PE

The acetabular labrum is a soft-tissue structure which lines the acetabular rim of the hip joint. Its role in hip joint biomechanics and joint health has been of particular interest over the past decade. In normal hip joint biomechanics, the labrum is crucial in retaining a layer of pressurised intra-articular fluid for joint lubrication and load support/distribution. Its seal around the femoral head is further regarded as a contributing to hip stability through its suction effect. The labrum itself is also important in increasing contact area thereby reducing contact stress. Given the labrum’s role in normal hip joint biomechanics, surgical techniques for managing labral damage are continuously evolving as our understanding of its anatomy and function continue to progress. The current paper aims to review the anatomy and biomechanical function of the labrum and how they are affected by differing surgical techniques.

Take home message: The acetabular labrum plays a critical role in hip function and maintaining and restoring its function during surgical intervention remain an essential goal.

Cite this article: Bone Joint J 2016;98-B:730–5.


The Bone & Joint Journal
Vol. 99-B, Issue 2 | Pages 175 - 183
1 Feb 2017
Morita D Hasegawa Y Okura T Osawa Y Ishiguro N

Aims

Transtrochanteric rotational osteotomy (TRO) is performed for young patients with non-traumatic osteonecrosis of the femoral head (ONFH) to preserve the hip. We aimed to investigate the long-term outcomes and the risk factors for failure 15 years after this procedure.

Patients and Methods

This study included 95 patients (111 hips) with a mean age of 40 years (21 to 64) who underwent TRO for ONFH. The mean follow-up was 18.2 years (3 to 26). Kaplan–Meier survivorship analyses were performed with conversion to total hip arthroplasty (THA) and radiological failure due to secondary collapse of the femoral head or osteoarthritic changes as the endpoint. Multivariate analyses were performed to assess risk factors for each outcome.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 33 - 41
1 Jan 2018
Matharu GS Hunt LP Murray DW Howard P Pandit HG Blom AW Bolland B Judge A

Aims

The aim of this study was to determine whether the rates of revision for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle components varied according to the year of the initial operation, and compare these with the rates of revision for other designs of MoM THA.

Patients and Methods

Data from the National Joint Registry for England and Wales included 36 mm MoM THAs with Pinnacle acetabular components which were undertaken between 2003 and 2012 with follow-up for at least five years (n = 10 776) and a control group of other MoM THAs (n = 13 817). The effect of the year of the primary operation on all-cause rates of revision was assessed using Cox regression and interrupted time-series analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1140 - 1146
1 Sep 2017
Shoji T Yamasaki T Izumi S Murakami H Mifuji K Sawa M Yasunaga Y Adachi N Ochi M

Aims

Our aim was to evaluate the radiographic characteristics of patients undergoing total hip arthroplasty (THA) for the potential of posterior bony impingement using CT simulations.

Patients and Methods

Virtual CT data from 112 patients who underwent THA were analysed. There were 40 men and 72 women. Their mean age was 59.1 years (41 to 76). Associations between radiographic characteristics and posterior bony impingement and the range of external rotation of the hip were evaluated. In addition, we investigated the effects of pelvic tilt and the neck/shaft angle and femoral offset on posterior bony impingement.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 741 - 746
1 Jun 2016
Hagio T Naito M Nakamura Y Muraoka K

Aims

Acetabular dysplasia is frequently associated with intra-articular pathology such as labral tears, but whether labral tears should be treated at the time of periacetabular osteotomy (PAO) remains controversial. The purpose of this study was to compare the clinical outcomes and radiographic corrections of PAO for acetabular dysplasia between patients with and without labral tears pre-operatively.

Patients and Methods

We retrospectively reviewed 70 hips in 67 patients with acetabular dysplasia who underwent PAO. Of 47 hips (45 patients) with labral tears pre-operatively, 27 (25 patients) underwent PAO alone, and were classified as the labral tear alone (LT) group, and 20 (20 patients) underwent combined PAO and osteochondroplasty, and were classified as the labral tear osteochondroplasty (LTO) group. The non-labral tear (NLT) group included 23 hips in 22 patients.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 865 - 871
1 Jul 2017
Amstutz HC Le Duff MJ Bhaurla SK

Aims

A contact patch to rim (CPR) distance of < 10 mm has been associated with edge-loading and excessive wear. However, not all arthroplasties with a low CPR distance show problems with wear. Therefore, CPR distance may not be the only variable affecting the post-operative metal ion concentrations.

Patients and Methods

We used multiple logistic regression to determine what variables differed between the patients who had high and low cobalt (CoS) and chromium (CrS) serum ion concentrations within a cohort of patients with low (< 10 mm) CPR distances. A total of 56 patients treated with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS ion studies performed more than one year after surgery. The mean age of the patients at the time of surgery was 51.7 years (29 to 70), with 38 women (68%) and 18 men (32%).


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 303 - 309
1 Mar 2017
Beaulé PE Bleeker H Singh A Dobransky J

Aims

Joint-preserving surgery of the hip (JPSH) has evolved considerably and now includes a number of procedures, including arthroscopy, surgical dislocation, and redirectional osteotomies of the femur and acetabulum. There are a number of different factors which lead to failure of JPSH. Consequently, it is of interest to assess the various modes of failure in order to continue to identify best practice and the indications for these procedures.

Patients and Methods

Using a retrospective observational study design, we reviewed 1013 patients who had undergone JPSH by a single surgeon between 2005 and 2015. There were 509 men and 504 women with a mean age of 39 years (16 to 78). Of the 1013 operations, 783 were arthroscopies, 122 surgical dislocations, and 108 peri-acetabular osteotomies (PAO). We analysed the overall failure rates and modes of failure. Re-operations were categorised into four groups: Mode 1 was arthritis progression or organ failure leading to total hip arthroplasty (THA); Mode 2 was an Incorrect diagnosis/procedure; Mode 3 resulted from malcorrection of femur (type A), acetabulum (type B), or labrum (type C) and Mode 4 resulted from an unintended consequence of the initial surgical intervention.


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 16 - 21
1 Jan 2017
Aprato A Leunig M Massé A Slongo T Ganz R

Aims

Several studies have reported the safety and efficacy of subcapital re-alignment for patients with slipped capital femoral epiphysis (SCFE) using surgical dislocation of the hip and an extended retinacular flap. Instability of the hip and dislocation as a consequence of this surgery has only recently gained attention. We discuss this problem with some illustrative cases.

Materials and Methods

We explored the literature on the possible pathophysiological causes and surgical steps associated with the risk of post-operative instability and articular damage. In addition, we describe supplementary steps that could be used to avoid these problems.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 31 - 37
1 Jan 2013
Zywiel MG Brandt J Overgaard CB Cheung AC Turgeon TR Syed KA

Symptomatic cobalt toxicity from a failed total hip replacement is a rare but devastating complication. It has been reported following revision of fractured ceramic components, as well as in patients with failed metal-on-metal articulations. Potential clinical findings include fatigue, weakness, hypothyroidism, cardiomyopathy, polycythaemia, visual and hearing impairment, cognitive dysfunction, and neuropathy. We report a case of an otherwise healthy 46-year-old patient, who developed progressively worsening symptoms of cobalt toxicity beginning approximately six months following synovectomy and revision of a fractured ceramic-on-ceramic total hip replacement to a metal-on-polyethylene bearing. The whole blood cobalt levels peaked at 6521 µg/l. The patient died from cobalt-induced cardiomyopathy. Implant retrieval analysis confirmed a loss of 28.3 g mass of the cobalt–chromium femoral head as a result of severe abrasive wear by ceramic particles embedded in the revision polyethylene liner. Autopsy findings were consistent with heavy metal-induced cardiomyopathy.

We recommend using new ceramics at revision to minimise the risk of wear-related cobalt toxicity following breakage of ceramic components.

Cite this article: Bone Joint J 2013;95-B:31–7.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 925 - 933
1 Jul 2016
Sidaginamale RP Joyce TJ Bowsher JG Lord JK Avery PJ Natu S Nargol AVF Langton DJ

Aims

We wished to investigate the influence of metal debris exposure on the subsequent immune response and resulting soft-tissue injury following metal-on-metal (MoM) hip arthroplasty. Some reports have suggested that debris generated from the head-neck taper junction is more destructive than equivalent doses from metal bearing surfaces.

Patients and Methods

We investigated the influence of the source and volume of metal debris on chromium (Cr) and cobalt (Co) concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using multiple regression analysis of prospectively collected data. A total of 199 explanted MoM hips (177 patients; 132 hips female) were analysed to determine rates of volumetric wear at the bearing surfaces and taper junctions.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 56 - 63
1 Apr 2012
Langton DJ Sidaginamale R Lord JK Nargol AVF Joyce TJ

Objectives

An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer.

Methods

Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular bearing surfaces and also the surfaces of the taper junctions. The relationship between taper wear rates and a number of variables, including bearing diameter and orientation of the acetabular component, was investigated.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 44 - 49
1 Jan 2016
Sheth NP Melnic CM Paprosky WG

Given the increasing number of total hip arthroplasty procedures being performed annually, it is imperative that orthopaedic surgeons understand factors responsible for instability. In order to treat this potentially complex problem, we recommend correctly classifying the type of instability present based on component position, abductor function, impingement, and polyethylene wear. Correct classification allows the treating surgeon to choose the appropriate revision option that ultimately will allow for the best potential outcome.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):44–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 725 - 729
1 Jun 2007
Ikemura S Yamamoto T Jingushi S Nakashima Y Mawatari T Iwamoto Y

Transtrochanteric curved varus osteotomy was designed to avoid some of the disadvantages of varus wedge osteotomy, such as post-operative leg-length discrepancy. In this retrospective study we investigated the leg-length discrepancy and clinical outcome after transtrochanteric curved varus osteotomy undertaken in patients with osteonecrosis of the femoral head. Between January 1993 and March 2004, this osteotomy was performed in 42 hips of 36 patients with osteonecrosis of the femoral head. There were 15 males and 21 females with a mean age at surgery of 34 years (15 to 68). The mean follow-up was 5.9 years (2.0 to 12.5). The mean pre-operative Harris hip score was 64.0 (43 to 85) points, which improved to a mean of 88.7 (58 to 100) points at final follow-up. The mean varus angulation post-operatively was 25° (12° to 38°) and the post-operative mean leg-length discrepancy was 13 mm (4 to 25). The post-operative leg-length discrepancy showed a strong correlation with varus angulation (Pearson’s correlation coefficient; r = 0.9530, p < 0.0001), which may be useful for predicting the leg-length discrepancy which can occur even after transtrochanteric curved varus osteotomy.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 40 - 48
1 Jan 2016
Matharu GS Mansour R Dada O Ostlere S Pandit HG Murray DW

Aims

The aims of this study were to compare the diagnostic test characteristics of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI) alone, and ultrasound combined with MARS-MRI for identifying intra-operative pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients undergoing revision surgery.

Methods

This retrospective diagnostic accuracy study involved 39 patients (40 MoMHRs). The time between imaging modalities was a mean of 14.6 days (0 to 90), with imaging performed at a mean of 5.3 months (0.06 to 12) before revision. The prevalence of intra-operative pseudotumours was 82.5% (n = 33).


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1045 - 1051
1 Aug 2013
Arthroplasty Society C

The purpose of this study was twofold: first, to determine whether the five-year results of hip resurfacing arthroplasty (HRA) in Canada justified the continued use of HRA; and second, to identify whether greater refinement of patient selection was warranted.

This was a retrospective cohort study that involved a review of 2773 HRAs performed between January 2001 and December 2008 at 11 Canadian centres. Cox’s proportional hazards models were used to analyse the predictors of failure of HRA. Kaplan–Meier survival analysis was performed to predict the cumulative survival rate at five years. The factors analysed included age, gender, body mass index, pre-operative hip pathology, surgeon’s experience, surgical approach, implant sizes and implant types. The most common modes of failure were also analysed.

The 2773 HRAs were undertaken in 2450 patients: 2127 in men and 646 in women. The mean age at operation was 50.5 years (sd 8.72; 18 to 82) and mean follow-up was 3.4 years (sd 2.1; 2.0 to 10.1). At the last follow-up a total of 101 HRAs (3.6%) required revision. Using revision for all causes of failure as the endpoint, Kaplan–Meier survival analysis showed a cumulative survival of 96.4% (95% confidence interval (CI) 96.1 to 96.9) at five years. With regard to gender, the five-year overall survival was 97.4% in men (95% CI 97.1 to 97.7) and 93.6% in women (95% CI 92.6 to 94.6). Female gender, smaller femoral components, specific implant types and a diagnosis of childhood hip problems were associated with higher rates of failure. The most common cause of failure was fracture of the femoral neck, followed by loosening of the femoral component.

The failure rates of HRA at five years justify the ongoing use of this technique in men. Female gender is an independent predictor of failure, and a higher failure rate at five years in women leads the authors to recommend this technique only in exceptional circumstances for women.

Cite this article: Bone Joint J 2013;95-B:1044–50.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 164 - 171
1 Feb 2011
Langton DJ Joyce TJ Jameson SS Lord J Van Orsouw M Holland JP Nargol AVF De Smet KA

We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p < 0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with < 1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.


Bone & Joint Research
Vol. 5, Issue 9 | Pages 370 - 378
1 Sep 2016
Munir S Oliver RA Zicat B Walter WL Walter WK Walsh WR

Objectives

This study aimed to characterise and qualitatively grade the severity of the corrosion particles released into the hip joint following taper corrosion.

Methods

The 26 cases examined were CoC/ABG Modular (n = 13) and ASR/SROM (n = 13). Blood serum metal ion levels were collected before and after revision surgery. The haematoxylin and eosin tissue sections were graded on the presence of fibrin exudates, necrosis, inflammatory cells and corrosion products. The corrosion products were identified based on visible observation and graded on abundance. Two independent observers blinded to the clinical patient findings scored all cases. Elemental analysis was performed on corrosion products within tissue sections. X-Ray diffraction was used to identify crystalline structures present in taper debris.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 332 - 336
1 Mar 2011
Konan S Rayan F Meermans G Witt J Haddad FS

There have been considerable recent advances in the understanding and management of femoroacetabular impingement and associated labral and chondral pathology. We have developed a classification system for acetabular chondral lesions. In our system, we use the six acetabular zones previously described by Ilizaliturri et al. The cartilage is then graded on a scale of 0 to 4 as follows: grade 0, normal articular cartilage lesions; grade 1, softening or wave sign; grade 2, cleavage lesion; grade 3, delamination; and grade 4, exposed bone. The site of the lesion is further classed as A, B or C based on whether the lesion is less than one-third of the distance from the acetabular rim to the cotyloid fossa, one-third to two-thirds of the same distance and greater than two-thirds of the distance, respectively. In order to validate the classification system, six surgeons graded ten video recordings of hip arthroscopy.

Our findings showed a high intra-observer reliability of the classification system with an intraclass correlation coefficient of 0.81 and a high interobserver reliability with an intraclass correlation coefficient of 0.88.

We have developed a simple reproducible classification system for lesions of the acetabular cartilage, which it is hoped will allow standardised documentation to be made of damage to the articular cartilage, particularly that associated with femoroacetabular impingement.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1287 - 1295
1 Oct 2009
Langton DJ Sprowson AP Joyce TJ Reed M Carluke I Partington P Nargol AVF

There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months).

The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p < 0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations > 45° and anteversion angles of < 10° and > 20°. These levels were only increased in the BHR group when the acetabular component was implanted with an inclination > 55°.

A significant relationship was identified between the anteversion of the BHR acetabular component and the levels of Cr and Co (p < 0.05 for Co), with an increase observed at anteversion angles < 10° and > 20°. The median whole blood and serum Cr concentrations of the male ASR patients were significantly lower than those of the BHR men (p < 0.001). This indicates that reduced diametral clearance may equate to a reduction in metal ion concentrations in larger joints with satisfactory orientation of the acetabular component.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 32 - 35
1 Jan 2005
Diab M Clark JM Weis MA Eyre DR

In developmental dysplasia of the hip, a deficient acetabulum may be augmented by placing local autogenous iliac osseous graft, or the ilium itself, over the head of the femur with the expectation that the added bone will function as a bearing surface. We analysed this bone obtained en bloc during subsequent surgery which was performed for degenerative osteoarthritis in three patients at 6, 25 and 30 years after the initial augmentation procedure. In each patient, the augmentation comprised of red cancellous bone covered on its articulating surface by a distinct layer of white tissue. Microscopy of this tissue showed parallel rows of spindle-shaped cells lying between linearly arranged collagen bundles typical of joint capsule. Biochemical analysis showed type I collagen, the principal collagen of joint capsule and bone, with no significant quantity of type II collagen, the principal collagen of cartilage. While the added bone produced by acetabular augmentation was durable, histological and biochemical analyses suggested that it had not undergone cartilage metaplasia. The augmented acetabulum articulates with the head of the femur by means of an interposed hip joint capsule.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 11 - 16
1 Nov 2014
Khanna V Tushinski DM Drexler M Backstein DB Gross AE Safir OA Kuzyk PR

Cartilage defects of the hip cause significant pain and may lead to arthritic changes that necessitate hip replacement. We propose the use of fresh osteochondral allografts as an option for the treatment of such defects in young patients. Here we present the results of fresh osteochondral allografts for cartilage defects in 17 patients in a prospective study. The underlying diagnoses for the cartilage defects were osteochondritis dissecans in eight and avascular necrosis in six. Two had Legg-Calve-Perthes and one a femoral head fracture. Pre-operatively, an MRI was used to determine the size of the cartilage defect and the femoral head diameter. All patients underwent surgical hip dislocation with a trochanteric slide osteotomy for placement of the allograft. The mean age at surgery was 25.9 years (17 to 44) and mean follow-up was 41.6 months (3 to 74). The mean Harris hip score was significantly better after surgery (p < 0.01) and 13 patients had fair to good outcomes. One patient required a repeat allograft, one patient underwent hip replacement and two patients are awaiting hip replacement. Fresh osteochondral allograft is a reasonable treatment option for hip cartilage defects in young patients.

Cite this article: Bone Joint J 2014;96-B(11 Supple A):11–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 781 - 786
1 Jun 2010
Zhao G Yamamoto T Ikemura S Motomura G Mawatari T Nakashima Y Iwamoto Y

We investigated the factors related to the radiological outcome of a transtrochanteric curved varus osteotomy in patients with osteonecrosis of the hip. We reviewed 73 hips in 62 patients with a mean follow-up of 12.4 years (5 to 31.1). There were 28 men and 34 women, with a mean age of 33.3 years (15 to 68) at the time of surgery. The 73 hips were divided into two groups according to their radiological findings: group 1 showed progression of collapse and/or joint-space narrowing; group 2 had neither progressive collapse nor joint-space narrowing. Both of these factors and the radiological outcomes were analysed by a stepwise discriminant analysis. A total of 12 hips were categorised as group 1 and 61 as group 2. Both the post-operative intact ratio and the localisation of the necrotic lesion correlated with the radiological outcome. The cut-off point of the postoperative intact ratio to prevent the progression of collapse was 33.6%, and the cut-off point to prevent both the progression of collapse and joint-space narrowing was 41.9%.

The results of this study indicate that a post-operative intact ratio of 33.0% is necessary if a satisfactory outcome is to be achieved after this varus osteotomy.


Bone & Joint 360
Vol. 3, Issue 1 | Pages 40 - 41
1 Feb 2014
Ivory J

Metal-on-metal (MoM) hip resurfacing was developed in the 1990s by surgeons in Birmingham, UK, as a surgical solution to the problem of osteoarthritis in younger, more active patients. Early results were promising and the procedure gained in popularity. However, adverse reports of soft-tissue reaction and failure started to appear from 2008 onwards. Surgeons may be asked to write medico-legal reports on the surgical aspects of an individual case for claimant lawyers or in defence for the NHSLA or indemnity insurers. The purpose of this article is to cover some of the aspects of the operation that may be considered in such medico-legal reports.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1011 - 1016
1 Aug 2011
Langton DJ Jameson SS Joyce TJ Gandhi JN Sidaginamale R Mereddy P Lord J Nargol AVF

There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent National Joint Registry data have shown clear differences in the rates of failure of different designs of hip resurfacing. Our aim was to update the failure rates related to metal debris for the Articular Surface Replacement (ASR). A total of 505 of these were implanted.

Kaplan-Meier analysis showed a failure rate of 25% at six years for the ASR resurfacing and of 48.8% for the ASR total hip replacement (THR). Of 257 patients with a minimum follow-up of two years, 67 (26.1%) had a serum cobalt concentration which was greater than 7 μg/l. Co-ordinate measuring machine analysis of revised components showed that all patients suffering adverse tissue reactions in the resurfacing group had abnormal wear of the bearing surfaces. Six THR patients had relatively low rates of articular wear, but were found to have considerable damage at the trunion-taper interface. Our results suggest that wear at the modular junction is an important factor in the development of adverse tissue reactions after implantation of a large-diameter MoM THR.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 7 - 10
1 Nov 2013
Petrie J Sassoon A Haidukewych GJ

Most hip fractures treated with modern internal fixation techniques will heal. However, failures occasionally occur and require revision procedures. Salvage strategies employed during revision are based on whether the fixation failure occurs in the femoral neck, or in the intertrochanteric region. Patient age and remaining bone stock also influence decision making. For fractures in young patients, efforts are generally focused on preserving the native femoral head via osteotomies and repeat internal fixation. For failures in older patients, some kind of hip replacement is usually selected. Disuse osteopenia, deformity, bone loss, and stress-risers from previous internal fixation devices all pose technical challenges to successful reconstruction. Attention to detail is important in order to minimise complications. In the majority of cases, good outcomes have been reported for the various salvage strategies.

Cite this article: Bone Joint J 2013;95-B, Supple A:7–10.


Bone & Joint Research
Vol. 3, Issue 1 | Pages 1 - 6
1 Jan 2014
Yamada K Mihara H Fujii H Hachiya M

Objectives

There are several reports clarifying successful results following open reduction using Ludloff’s medial approach for congenital (CDH) or developmental dislocation of the hip (DDH). This study aimed to reveal the long-term post-operative course until the period of hip-joint maturity after the conventional surgical treatments.

Methods

A long-term follow-up beyond the age of hip-joint maturity was performed for 115 hips in 103 patients who underwent open reduction using Ludloff’s medial approach in our hospital. The mean age at surgery was 8.5 months (2 to 26) and the mean follow-up was 20.3 years (15 to 28). The radiological condition at full growth of the hip joint was evaluated by Severin’s classification.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 11 - 16
1 Nov 2013
Sierra RJ Mabry TM Sems SA Berry DJ

Total hip replacement (THR) after acetabular fracture presents unique challenges to the orthopaedic surgeon. The majority of patients can be treated with a standard THR, resulting in a very reasonable outcome. Technical challenges however include infection, residual pelvic deformity, acetabular bone loss with ununited fractures, osteonecrosis of bone fragments, retained metalwork, heterotopic ossification, dealing with the sciatic nerve, and the difficulties of obtaining long-term acetabular component fixation. Indications for an acute THR include young patients with both femoral head and acetabular involvement with severe comminution that cannot be reconstructed, and the elderly, with severe bony comminution. The outcomes of THR for established post-traumatic arthritis include excellent pain relief and functional improvements. The use of modern implants and alternative bearing surfaces should improve outcomes further.

Cite this article: Bone Joint J 2013;95-B, Supple A:11–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1143 - 1151
1 Sep 2008
Langton DJ Jameson SS Joyce TJ Webb J Nargol AVF

Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1320 - 1325
1 Oct 2013
Tamura S Nishii T Takao M Sakai T Yoshikawa H Sugano N

We investigated differences in the location and mode of labral tears between dysplastic hips and hips with femoroacetabular impingement (FAI). We also investigated the relationship between labral tear and adjacent cartilage damage. We retrospectively studied 72 symptomatic hips (in 68 patients: 19 men and 49 women) with radiological evidence of dysplasia or FAI on high-resolution CT arthrography. The incidence and location of labral tears and modes of tear associated with the base of the labrum (Mode 1) or body of the labrum (Mode 2) were compared among FAI, mildly dysplastic and severely dysplastic hips. The locations predominantly involved with labral tears were different in FAI and mild dysplastic hips (anterior and anterosuperior zones) and in severely dysplastic hips (anterosuperior and superior zones) around the acetabulum. Significant differences were observed in the prevalence of Mode 1 versus Mode 2 tears in FAI hips (72% (n = 13) vs 28% (n = 5)) and severe dysplastic hips (25% (n = 2) vs 75% (n = 6)). The frequency of cartilage damage adjacent to Mode 1 tears was significantly higher (42% (n = 14)) than that adjacent to Mode 2 tears (14% (n = 3)).

Hip pathology is significantly related to the locations and modes of labral tears. Mode 1 tears may be a risk factor for the development of adjacent acetabular cartilage damage.

Cite this article: Bone Joint J 2013;95-B:1320–5.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 314 - 319
1 Mar 2013
Masjedi M Nightingale CL Azimi DY Cobb JP

We examined the relationship between the size of the femoral cam in femoroacetabular impingement (FAI) and acetabular pathomorphology to establish if pincer impingement exists in patients with a femoral cam.

CT scans of 37 symptomatic impinging hips with a femoral cam were analysed in a three-dimensional study and were compared with 34 normal hips. The inclination and version of the acetabulum as well as the acetabular rim angle and the bony acetabular coverage were calculated. These measurements were correlated with the size and shape of the femoral cams.

While the size of the femoral cam varied characteristically, the acetabular morphology of the two groups was similar in terms of version (normal mean 23° (sd 7°); cam mean 22° (sd 9°)), inclination (normal mean 57° (sd 5°); cam mean 56° (sd 5°)), acetabular coverage (normal mean 41% (sd 5%); cam mean 42% (sd 4%)) and the mean acetabular rim angle (normal mean 82° (sd 5°); cam mean 83° (sd 4°)).

We found no correlation between acetabular morphology and the severity of cam lesion and no evidence of either global or focal over-coverage to support the diagnosis of ‘mixed’ FAI. The femoral cam may provoke edge loading but removal of any acetabular bearing surface when treating cam FAI might induce accelerated wear.

Cite this article: Bone Joint J 2013;95-B:314–19.


Bone & Joint 360
Vol. 1, Issue 3 | Pages 35 - 36
1 Jun 2012
Mason L


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1019 - 1024
1 Aug 2008
Cashin M Uhthoff H O’Neill M Beaulé PE

Damage to and repair of the acetabular labral-chondral complex are areas of clinical interest in the treatment of young adults with pain in the hip and in the prevention of degenerative arthritis of the hip. There are varying theories as to why most acetabular tears are located anterosuperiorly. We have studied the prenatal development of the human acetabular labral-chondral complex in 11 fetal hips, aged from eight weeks of gestation to term.

There were consistent differences between the anterior and posterior acetabular labral-chondral complex throughout all ages of gestation. The anterior labrum had a somewhat marginal attachment to the acetabular cartilage with an intra-articular projection. The posterior labrum was attached and continuous with the acetabular cartilage. Anteriorly, the labral-chondral transition zone was sharp and abrupt, but posteriorly it was gradual and interdigitated. The collagen fibres of the anterior labrum were arranged parallel to the labral-chondral junction, but at the posterior labrum they were aligned perpendicular to the junction.

We believe that in the anterior labrum the marginal attachment and the orientation of the collagen fibres parallel to the labral-chondral junction may render it more prone to damage than the posterior labrum in which the collagen fibres are anchored in the acetabular cartilage. The anterior intra-articular projection of the labrum should not be considered to be a pathological feature.


Bone & Joint Research
Vol. 1, Issue 12 | Pages 324 - 332
1 Dec 2012
Verhelst L Guevara V De Schepper J Van Melkebeek J Pattyn C Audenaert EA

The aim of this review is to evaluate the current available literature evidencing on peri-articular hip endoscopy (the third compartment). A comprehensive approach has been set on reports dealing with endoscopic surgery for recalcitrant trochanteric bursitis, snapping hip (or coxa-saltans; external and internal), gluteus medius and minimus tears and endoscopy (or arthroscopy) after total hip arthroplasty. This information can be used to trigger further research, innovation and education in extra-articular hip endoscopy.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1187 - 1192
1 Sep 2012
Rakhra KS Lattanzio P Cárdenas-Blanco A Cameron IG Beaulé PE

Advanced MRI cartilage imaging such as T1-rho (T1ρ) for the diagnosis of early cartilage degradation prior to morpholgic radiological changes may provide prognostic information in the management of joint disease. This study aimed first to determine the normal T1ρ profile of cartilage within the hip, and secondly to identify any differences in T1ρ profile between the normal and symptomatic femoroacetabular impingement (FAI) hip. Ten patients with cam-type FAI (seven male and three female, mean age 35.9 years (28 to 48)) and ten control patients (four male and six female, mean age 30.6 years (22 to 35)) underwent 1.5T T1ρ MRI of a single hip. Mean T1ρ relaxation times for full thickness and each of the three equal cartilage thickness layers were calculated and compared between the groups. The mean T1ρ relaxation times for full cartilage thickness of control and FAI hips were similar (37.17 ms (sd 9.95) and 36.71 ms (sd 6.72), respectively). The control group demonstrated a T1ρ value trend, increasing from deep to superficial cartilage layers, with the middle third having significantly greater T1ρ relaxation values than the deepest third (p = 0.008). The FAI group demonstrated loss of this trend. The deepest third in the FAI group demonstrated greater T1ρ relaxation values than controls (p = 0.028).

These results suggest that 1.5T T1ρ MRI can detect acetabular hyaline cartilage changes in patients with FAI.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 10 - 19
1 Jan 2013
Bedi A Kelly BT Khanduja V

The technical advances in arthroscopic surgery of the hip, including the improved ability to manage the capsule and gain extensile exposure, have been paralleled by a growth in the number of conditions that can be addressed. This expanding list includes symptomatic labral tears, chondral lesions, injuries of the ligamentum teres, femoroacetabular impingement (FAI), capsular laxity and instability, and various extra-articular disorders, including snapping hip syndromes. With a careful diagnostic evaluation and technical execution of well-indicated procedures, arthroscopic surgery of the hip can achieve successful clinical outcomes, with predictable improvements in function and pre-injury levels of physical activity for many patients.

This paper reviews the current position in relation to the use of arthroscopy in the treatment of disorders of the hip.

Cite this article: Bone Joint J 2013;95-B:10–19.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 290 - 296
1 Mar 2012
Jayakumar P Ramachandran M Youm T Achan P

Hip arthroscopy is particularly attractive in children as it confers advantages over arthrotomy or open surgery, such as shorter recovery time and earlier return to activity. Developments in surgical technique and arthroscopic instrumentation have enabled extension of arthroscopy of the hip to this age group. Potential challenges in paediatric and adolescent hip arthroscopy include variability in size, normal developmental change from childhood to adolescence, and conditions specific to children and adolescents and their various consequences. Treatable disorders include the sequelae of traumatic and sports-related hip joint injuries, Legg–Calve–Perthes’ disease and slipped capital femoral epiphysis, and the arthritic and septic hip. Intra-articular abnormalities are rarely isolated and are often associated with underlying morphological changes.

This review presents the current concepts of hip arthroscopy in the paediatric and adolescent patient, covering clinical assessment and investigation, indications and results of the experience to date, as well as technical challenges and future directions.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 916 - 920
1 Jul 2005
Ebied A Hoad-Reddick DA Raut V

A reduced femoral offset in total hip replacement has been thought to be disadvantageous. We reviewed the results of 54 consecutive primary total hip replacements in 49 patients (mean age of 68 years) performed between August 1990 and December 1994, with a mean follow-up of 8.8 years (sd 2.1). The mean pre-operative femoral offset for these hips was 41 mm (sd 7.4). All patients received a low-offset Charnley stem and a polyethylene cup inserted by a single surgeon.

At their latest follow-up, surviving patients had a significant improvement in the performance of their hip. Three had undergone revision, one each for deep infection, recurrent dislocation and late pain with subluxation. No hips had been revised, or were at risk of revision, for aseptic loosening. The mean annual linear rate of wear was 0.2 mm (sd 0.08) for the whole group. There was no correlation between the pre-operative femoral offset and the post-operative rate of wear. Our survivorship estimate was 91% when revision for any reason was taken as an end-point, but 100% if aseptic loosening was considered as the end-point. Our study demonstrates that a low-offset femoral stem can produce good, medium-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 173 - 178
1 Feb 2012
Malizos KN Papasoulis E Dailiana ZH Papatheodorou LK Varitimidis SE

The introduction of a trabecular tantalum rod has been proposed for the management of early-stage osteonecrosis of the femoral head but serves as a single-point of support of the necrotic lesion. We describe a technique using two or three 4.2 mm (or later 4.7 mm) tantalum pegs for the prevention of collapse of the necrotic lesion. We prospectively studied 21 patients (26 hips) with non-traumatic osteonecrosis of the femoral head treated in this manner. Of these, 21 patients (24 hips) were available for radiological and clinical evaluation at a mean follow-up of 46 months (18 to 67). Radiological assessment showed that only eight hips deteriorated according to the Association Research Circulation Osseous classification, and four hips according to the Classification of the Japanese Investigation Committee of Health and Welfare. Functional improvement was obtained with an improvement in the mean Harris hip score from 65.2 (33.67 to 95) to 88.1 (51.72 to 100), the mean Merle D’Aubigné-Postel score from 13 (6 to 18) to 16 (11 to 18), a mean visual analogue score for pain from 5.2 (0 to 9.5) to 2.6 (0 to 7), and the mean Short-Form 36 score from 80.4 (56.8 to 107.1) to 92.4 (67.5 to 115.7). Of these 24 hips followed for a minimum of 18 months, three were considered as failures at the final follow-up, having required total hip replacement. One of the hips without full follow-up was also considered to be a failure. In more than two-thirds of the surviving hips a satisfactory clinical outcome was achieved with promising radiological findings. The estimated mean implant survival was 60 months (95% confidence interval 53.7 to 66.3).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 746 - 754
1 Jun 2012
Jameson SS Baker PN Mason J Porter ML Deehan DJ Reed MR

Modern metal-on-metal hip resurfacing has been widely performed in the United Kingdom for over a decade. However, the literature reports conflicting views of the benefits: excellent medium- to long-term results with some brands in specific subgroups, but high failure rates and local soft-tissue reactions in others. The National Joint Registry for England and Wales (NJR) has collected data on all hip resurfacings performed since 2003. This retrospective cohort study recorded survival time to revision from a resurfacing procedure, exploring risk factors independently associated with failure. All patients with a primary diagnosis of osteoarthritis who underwent resurfacing between 2003 and 2010 were included in the analyses. Cox’s proportional hazard models were used to analyse the extent to which the risk of revision was related to patient, surgeon and implant covariates.

A total of 27 971 hip resurfacings were performed during the study period, of which 1003 (3.59%) underwent revision surgery. In the final adjusted model, we found that women were at greater risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007), but the risk of revision was independent of age. Of the implant-specific predictors, five brands had a significantly greater risk of revision than the Birmingham Hip Resurfacing (BHR) (ASR: HR = 2.82, p < 0.001, Conserve: HR = 2.03, p < 0.001, Cormet: HR = 1.43, p = 0.001, Durom: HR = 1.67, p < 0.001, Recap: HR = 1.58, p = 0.007). Smaller femoral head components were also significantly more likely to require revision (≤ 44 mm: HR = 2.14, p < 0.001, 45 to 47 mm: HR = 1.48, p = 0.001) than medium or large heads, as were operations performed by low-volume surgeons (HR = 1.36, p <  0.001). Once these influences had been removed, in 4873 male patients < 60 years old undergoing resurfacing with a BHR, the five-year estimated risk of revision was 1.59%.

In summary, after adjustment for a range of covariates we found that there were significant differences in the rate of failure between brands and component sizes. Younger male patients had good five-year implant survival when the BHR was used.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 732 - 737
1 Jun 2011
Kim K Cho Y Ramteke AA Yoo M

Reconstructive acetabular osteotomy is a well established and effective procedure in the treatment of acetabular dysplasia. However, the dysplasia is frequently accompanied by intra-articular pathology such as labral tears. We intended to determine whether a concomitant hip arthroscopy with peri-acetabular rotational osteotomy could identify and treat intra-articular pathology associated with dysplasia and thereby produce a favourable outcome. We prospectively evaluated 43 consecutive hips treated by combined arthroscopy and acetabular osteotomy. Intra-operative arthroscopic examination revealed labral lesions in 38 hips. At a mean follow-up of 74 months (60 to 97) the mean Harris hip score improved from 72.4 to 94.0 (p < 0.001), as did all the radiological parameters (p < 0.001). Complications included penetration of the joint by the osteotome in one patient, a fracture of the posterior column in another and deep-vein thrombosis in one further patient. This combined surgical treatment gave good results in the medium term.

We suggest that arthroscopy of the hip can be performed in conjunction with peri-acetabular osteotomy to provide good results in patients with symptomatic dysplasia of the hip, and the arthroscopic treatment of intra-articular pathology may alter the progression of osteoarthritis.


Bone & Joint Research
Vol. 1, Issue 7 | Pages 131 - 144
1 Jul 2012
Papavasiliou AV Bardakos NV

Over recent years hip arthroscopic surgery has evolved into one of the most rapidly expanding fields in orthopaedic surgery. Complications are largely transient and incidences between 0.5% and 6.4% have been reported. However, major complications can and do occur. This article analyses the reported complications and makes recommendations based on the literature review and personal experience on how to minimise them.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1169 - 1177
1 Sep 2011
Underwood R Matthies A Cann P Skinner JA Hart AJ

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component.

There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 738 - 745
1 Jun 2011
Davda K Lali FV Sampson B Skinner JA Hart AJ

We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging.

Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 629 - 633
1 May 2010
Haviv B Singh PJ Takla A O’Donnell J

This study evaluates the outcome of arthroscopic femoral osteochondroplasty for cam lesions of the hip in the absence of additional pathology other than acetabular chondral lesions. We retrospectively reviewed 166 patients (170 hips) who were categorised according to three different grades of chondral damage. The outcome was assessed in each grade using the modified Harris Hip Score (MHHS) and the Non-Arthritic Hip Score (NAHS).

Overall, at the last follow-up (mean 22 months, 12 to 72), the mean MHHS had improved by 15.3 points (95% confidence interval (CI), 8.9 to 21.7) and the mean NAHS by 15 points (95% CI, 9.4 to 20.5). Significantly better results were observed in hips with less severe chondral damage. Microfracture in limited chondral lesions showed superior results.

Arthroscopic femoral osteochondroplasty for cam impingement with microfracture in selected cases is beneficial. The outcome correlates with the severity of acetabular chondral damage.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 589 - 594
1 May 2009
Allen D Beaulé PE Ramadan O Doucette S

Femoroacetabular impingement is a cause of hip pain in adults and is potentially a precursor of osteoarthritis. Our aim in this study was to determine the prevalence of bilateral deformity in patients with symptomatic cam-type femoroacetabular impingement as well as the presence of associated acetabular abnormalities and hip pain. We included all patients aged 55 years or less seen by the senior author for hip pain, with at least one anteroposterior and lateral pelvic radiograph available. All patients with dysplasia and/or arthritis were excluded. A total of 113 patients with a symptomatic cam-impingement deformity of at least one hip was evaluated. There were 82 men and 31 women with a mean age of 37.9 years (16 to 55).

Bilateral cam-type deformity was present in 88 patients (77.8%) while only 23 of those (26.1%) had bilateral hip pain. Painful hips had a statistically significant higher mean alpha angle than asymptomatic hips (69.9° vs 63.1°, p < 0.001). Hips with an alpha angle of more than 60° had an odds ratio of being painful of 2.59 (95% confidence interval 1.32 to 5.08, p = 0.006) compared with those with an alpha angle of less than 60°. Of the 201 hips with a cam-impingement deformity 42% (84) also had a pincer deformity.

Most patients with cam-type femoroacetabular impingement had bilateral deformities and there was an associated acetabular deformity in 84 of 201 patients (42%). This information is important in order to define the natural history of these deformities, and to determine treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 570 - 573
1 May 2008
Civinini R D’Arienzo M Innocenti M

We reviewed the long-term results at ten to 12 years of 118 total hip replacements in 109 patients using a second-generation hemispherical cementless acetabular component (Reflection) designed to address the problem of backside wear. Five patients (five hips) died and six patients (seven hips) were lost to follow-up. The remaining 98 patients (106 hips) had a mean age of 62.9 years (34.0 to 86.2) A rate of revision for aseptic loosening of 0.9%, and predictable results were found with respect to radiological evidence of fixation, lack of pain, walking ability, range of movement and function. One component was revised for aseptic loosening, and of the 101 hips (95.2%) that did not have a revision, minor osteolytic lesions of the pelvis were seen in six (5.9%). Kaplan-Meier survival analysis for the total cohort of 118 hips revealed a 96.4% survival at both ten (95% confidence interval 90 to 98) and 12 years (95% confidence interval 86 to 98).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 310 - 315
1 Mar 2009
Olsen M Davis ET Waddell JP Schemitsch EH

We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip.

The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° (sd 2.6). The stem-shaft angle after operation differed from that planned by a mean of 2.8° (sd 2.0) and in 86% of cases the final angle measured within ± 5° of that planned. We had no instances of notching of the neck or varus alignment of the implant in our series. A learning curve was observed in the time taken for navigation, but not for accurate placement of the implant.

Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1291 - 1297
1 Oct 2008
De Haan R Pattyn C Gill HS Murray DW Campbell PA De Smet K

We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55°.

There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p < 0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55°, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1158 - 1163
1 Sep 2008
De Haan R Campbell PA Su EP De Smet KA

We have reviewed 42 patients who had revision of metal-on-metal resurfacing procedures, mostly because of problems with the acetabular component. The revisions were carried out a mean of 26.2 months (1 to 76) after the initial operation and most of the patients (30) were female.

Malpositioning of the acetabular component resulted in 27 revisions, mostly because of excessive abduction (mean 69.9°; 56° to 98°) or insufficient or excessive anteversion. Seven patients had more than one reason for revision. The mean increase in the diameter of the component was 1.8 mm (0 to 4) when exchange was needed.

Malpositioning of the components was associated with metallosis and a high level of serum ions. The results of revision of the femoral component to a component with a modular head were excellent, but four patients had dislocation after revision and four required a further revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 314 - 318
1 Mar 2008
Boyer T Dorfmann H

Between 1985 and 2000, 120 patients underwent arthroscopic management for primary synovial chondromatosis of the hip. We report the outcome of 111 patients with a mean follow-up of 78.6 months (12 to 196). More than one arthroscopy was required in 23 patients (20.7%), and 42 patients (37.8%) went on to require open surgery. Outcomes were evaluated in greater detail in 69 patients (62.2%) treated with arthroscopy alone, of whom 51 (45.9%) required no further treatment and 18 (16.2%) required further arthroscopies. Of the 111 patients, 63 (56.7%) had excellent or good outcomes. At the most recent follow-up, 22 patients (19.8%) had undergone total hip replacement.

Hip arthroscopy proved beneficial for patients diagnosed with primary synovial chondromatosis of the hip, providing good or excellent outcomes in more than half the cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1036 - 1041
1 Aug 2007
Knahr K Pospischill M Köttig P Schneider W Plenk H

Two Durasul highly crosslinked polyethylene liners were exchanged during revision surgery four and five years after implantation, respectively. The retrieved liners were evaluated macroscopically and surface analysis was performed using optical and electron microscopy. A sample of each liner was used to determine the oxidation of the material by Fourier transform infrared spectroscopy. Samples of the capsule were examined histologically.

The annual wear rate was found to be 0.010 and 0.015 mm/year, respectively. Surface analysis showed very little loss of material caused by wear. Histological evaluation revealed a continuous neosynovial lining with single multinucleated foreign-body giant cells. Our findings showed no unexpected patterns of wear on the articulating surfaces up to five years after implantation and no obvious failure of material.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 727 - 729
1 Jun 2006
Kim WY Hutchinson CE Andrew JG Allen PD

Excessive acetabular cover secondary to a retroverted acetabulum causes pincer impingement, which may cause early osteoarthritis of the hip. Our aim was to determine if there was a relationship between acetabular version and osteoarthritis of the hip. Using image processing and analysis software we studied 117 CT images of the hip in patients aged less than 65 years who had undergone a CT virtual colonoscopy. The mean CT joint space of the 18 hips with acetabular retroversion was narrower compared with the 99 hips with normal acetabular alignment (p < 0.0001). A correlation of r = 0.46 (p < 0.01) was found between right hip acetabular version and the mean right hip joint space and of r = 0.31 (p = 0.02) between left hip acetabular version and the mean left hip joint space. Acetabular retroversion is associated with radiological evidence of osteoarthritis of the hip. An understanding of the mechanical basis of osteoarthritis of the hip allows early treatment of the underlying structural abnormality and prevents progression of the degenerative condition.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 586 - 590
1 May 2007
Sauvé P Mountney J Khan T De Beer J Higgins B Grover M

Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings.

We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring metal-on-metal THRs for more than 30 years. The levels of cobalt and chromium were elevated by five and three times, respectively compared with those in our reference groups. Metal-on-metal articulations appear to be the source of metal ions throughout the life of the prosthesis. In three patients who had undergone revision of a previous metal-on-metal THR to a metal-on-polyethylene replacement the levels of metal ions were within the normal range. The elevations of cobalt and chromium ions seen in our study were comparable with those in patients with modern metal-on-metal THRs.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 35 - 39
1 Jan 2006
Beaulé PE Campbell PA Hoke R Dorey F

During hip resurfacing arthroplasty, excessive valgus positioning or surgical technique can result in notching of the femoral neck. Although mechanical weakening and subsequent fracture of the femoral neck are well described, the potential damage to the retinacular vessels leading to an ischaemic event is relatively unknown. Using laser Doppler flowmetry, we measured the blood flow in 14 osteoarthritic femoral heads during routine total hip replacement surgery, before and after notching of the femoral neck. In ten hips there was a reduction in blood flow of more than 50% from the baseline value after simulated notching of the femoral neck. Our results suggest that femoral head vascularity in the osteoarthritic state is similar to the non-arthritic state, where damage to the extraosseous vessels can predispose to avascular necrosis. Surgeons who perform resurfacing arthroplasty of the hip should pay careful attention to these vessels by avoiding excessive dissection around the femoral neck and/or notching.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 21 - 25
1 Jan 2007
Khan A Yates P Lovering A Bannister GC Spencer RF

We determined the effect of the surgical approach on perfusion of the femoral head during hip resurfacing arthroplasty by measuring the concentration of cefuroxime in bone samples from the femoral head. A total of 20 operations were performed through either a transgluteal or an extended posterolateral approach.

The concentration of cefuroxime in bone was significantly greater when using the transgluteal approach (mean 15.7 mg/kg; 95% confidence interval 12.3 to 19.1) compared with that using the posterolateral approach (mean 5.6 mg/kg; 95% confidence interval 3.5 to 7.8; p < 0.001). In one patient, who had the operation through a posterolateral approach, cefuroxime was undetectable.

Using cefuroxime as an indirect measure of blood flow, the posterolateral approach was found to be associated with a significant reduction in the blood supply to the femoral head during resurfacing arthroplasty compared with the transgluteal approach.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 877 - 882
1 Jul 2006
Hasegawa M Sudo A Uchida A

A modular layered acetabular component (metal-polyethylene-ceramic) was developed in Japan for use in alumina ceramic-on-ceramic total hip replacement. Between May 1999 and July 2000, we performed 35 alumina ceramic-on-ceramic total hip replacements in 30 consecutive patients, using this layered component and evaluated the clinical and radiological results over a mean follow-up of 5.8 years (5 to 6.5). A total of six hips underwent revision, one for infection, two for dislocation with loosening of the acetabular component, two for alumina liner fractures and one for component dissociation with pelvic osteolysis. There were no fractures of the ceramic heads, and no loosening of the femoral or acetabular component in the unrevised hips was seen at final follow-up. Osteolysis was not observed in any of the unrevised hips. The survivorship analysis at six years after surgery was 83%. The layered acetabular component in our experience, has poor durability because of unexpected mechanical failures including alumina liner fracture and component dissociation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1158 - 1164
1 Sep 2007
Lusty PJ Watson A Tuke MA Walter WL Walter WK Zicat B

We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of < 15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p < 0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 471 - 477
1 Apr 2005
Jacobsen S Sonne-Holm S Søballe K Gebuhr P Lund B

In a longitudinal case-control study, we followed 81 subjects with dysplasia of the hip and 136 control subjects without dysplasia for ten years assessing radiological evidence of degeneration of the hip at admission and follow-up. There were no cases of subluxation in the group with dysplasia. Neither subjects with dysplasia nor controls had radiological signs of ongoing degenerative disease at admission. The primary radiological discriminator of degeneration of the hip was a change in the minimum joint space width over time. There were no significant differences between these with dysplasia and controls in regard to age, body mass index or occupational exposure to daily repeated lifting at admission.

We found no significant differences in the reduction of the joint space width at follow-up between subjects with dysplasia and the control subjects nor in self-reported pain in the hip. The association of subluxation and/or associated acetabular labral tears with dysplasia of the hip may be a conditional factor for the development of premature osteoarthritis in mildly to moderately dysplastic hips.