Aims. We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem
Objectives. An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Methods. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the
Objectives. This study aimed to characterise and qualitatively grade the severity of the corrosion particles released into the hip joint following
Aims. We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the
Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular
Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40
Aims. The aim of this study was to evaluate the suitability of the
Aims.
Aims. United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with modular
Aims. This study investigates head-neck
Aims. When performing revision total hip arthroplasty using diaphyseal-engaging titanium
Aims. There are limited published data detailing the volumetric material loss from
Aims. Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted
Aims. The aim of this study was to develop a novel computational model for estimating head/stem
Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless,
The Unified Classification System (UCS), or Vancouver system, is a validated and widely used classification system to guide the management of periprosthetic femoral fractures. It suggests that well-fixed stems (type B1) can be treated with fixation but that loose stems (types B2 and B3) should be revised. Determining whether a stem is loose can be difficult and some authors have questioned how to apply this classification system to polished
Aims. The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented
Aims. The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished
Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated
Objectives. Favourable results for collarless polished
Aims. The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless polished
Revision total hip arthroplasty (THA) is challenging
when there is severe loss of bone in the proximal femur. The purpose
of this study was to evaluate the clinical and radiographic outcomes
of revision THA in patients with severe proximal femoral bone loss
treated with a fluted,
Objectives. Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. Methods. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium,
Aims. The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). Patients and Methods. We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated
Aims. Although good clinical outcomes have been reported for monolithic
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. . We review the risks and causes of fracture of
Non-modular
Objectives. The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces. Methods. In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels. Results. The mean wear rate of the articulating surfaces of the heads and acetabular components of all six joints tested was found to be 6.1 mm. 3. /year (4.1 to 7.6). The mean wear rate of the femoral head
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished
We prospectively followed 191 consecutive collarless
polished
The early failure and revision of bimodular primary
total hip arthroplasty prostheses requires the identification of the
risk factors for material loss and wear at the
We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a
Polished,
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished,
We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.Objectives
Patients and Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article:
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
To evaluate the hypothesis that failed osteosynthesis of periprosthetic
Vancouver type B1 fractures can be treated successfully with stem
revision using a transfemoral approach and a cementless, modular,
tapered revision stem with reproducible rates of fracture healing,
stability of the revision stem, and clinically good results. A total of 14 patients (11 women, three men) with a mean age
of 72.4 years (65 to 90) undergoing revision hip arthroplasty after
failed osteosynthesis of periprosthetic fractures of Vancouver type
B1 were treated using a transfemoral approach to remove the well-fixed
stem before insertion of a modular, fluted titanium stem which obtained
distal fixation. These patients were clinically and radiologically
followed up for a mean 52.2 months (24 to 144).Aims
Patients and Methods
We report our experience of revision total hip
replacement (THR) using the Revitan curved modular titanium fluted revision
stem in patients with a full spectrum of proximal femoral defects.
A total of 112 patients (116 revisions) with a mean age of 73.4
years (39 to 90) were included in the study. The mean follow-up
was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died
but their data were included in the survival analysis, and four
patients (4 hips) were lost to follow-up. The clinical outcome,
proximal bone regeneration and subsidence were assessed for 101
hips. The mean Harris Hip Score was 88.2 (45.8 to 100) after five years
and there was an increase of the mean Barnett and Nordin-Score,
a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7).
Five stems had to be revised (4.3%), three (2.9%) showed subsidence,
five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%)
a periprosthetic infection. At the latest follow-up, the survival with revision of the stem
as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%)
and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic
fractures were not observed. We report excellent results with respect to subsidence, the risk
of fracture, and loosening after femoral revision using a modular
curved revision stem with distal cone-in-cone fixation. A successful
outcome depends on careful pre-operative planning and the use of
a transfemoral approach when the anatomy is distorted or a fracture
is imminent, or residual cement or a partially-secured existing
stem cannot be removed. The shortest appropriate stem should, in
our opinion, be used and secured with >
3 cm fixation at the femoral
isthmus, and distal interlocking screws should be used for additional
stability when this goal cannot be realised. Cite this article:
This study aimed to compare the change in health-related quality
of life of patients receiving a traditional cemented monoblock Thompson
hemiarthroplasty compared with a modern cemented modular polished-taper
stemmed hemiarthroplasty for displaced intracapsular hip fractures. This was a pragmatic, multicentre, multisurgeon, two-arm, parallel
group, randomized standard-of-care controlled trial. It was embedded
within the WHiTE Comprehensive Cohort Study. The sample size was
964 patients. The setting was five National Health Service Trauma
Hospitals in England. A total of 964 patients over 60 years of age who
required hemiarthroplasty of the hip between February 2015 and March
2016 were included. A standardized measure of health outcome, the
EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and
at four months following the operation.Aims
Patients and Methods
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
Aims. In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used
Aims. In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased
The February 2024 Hip & Pelvis Roundup. 360. looks at: Trial of vancomycin and cefazolin as surgical prophylaxis in arthroplasty; Is preoperative posterior femoral neck tilt a risk factor for fixation failure? Cemented versus uncemented hemiarthroplasty for displaced intracapsular fractures of the hip; Periprosthetic fractures in larger hydroxyapatite-coated stems: are collared stems a better alternative for total hip arthroplasty?; Postoperative periprosthetic fracture following hip arthroplasty with a polished