We treated 108 patients with a pertrochanteric femoral fracture using either the dynamic hip screw or the proximal
The October 2023 Trauma Roundup. 360. looks at: Intramedullary nailing versus sliding hip screw in trochanteric fracture management: the INSITE randomized clinical trial; Five-year outcomes for patients with a displaced fracture of the distal tibia; Direct anterior versus anterolateral approach in hip joint hemiarthroplasty; Proximal humerus fractures: treat them all nonoperatively?; Tranexamic acid administration by prehospital personnel; Locked plating versus nailing for proximal tibia fractures: a multicentre randomized controlled trial; A retrospective review of the rate of septic knee arthritis after retrograde
Aims. Restoration of proximal medial femoral support is the keystone in the treatment of intertrochanteric fractures. None of the available implants are effective in constructing the medial femoral support. Medial sustainable nail (MSN-II) is a novel cephalomedullary nail designed for this. In this study, biomechanical difference between MSN-II and proximal
Retrograde
We describe a simple method of inserting the distal screws in a locked
Objectives. Because of the contradictory body of evidence related to the
potential benefits of helical blades in trochanteric fracture fixation,
we studied the effect of bone compaction resulting from the insertion
of a proximal
We performed a randomised, prospective trial to evaluate the use of unreamed titanium nails for femoral fractures. Of 48 patients with 50 femoral fractures 45 were followed to union; 23 with an unreamed and 22 with a reamed nail. The study was stopped early because of a high rate of implant failure. The fractures in the unreamed group were slower to unite (39.4 weeks) than those in the reamed group (28.5 weeks; p = 0.007). The time to union was over nine months in 57% of the unreamed group and in 18% of the reamed group. In the unreamed group 14 secondary procedures were required in ten patients to enhance healing compared with three in three patients in the reamed group. Six implants (13%) failed, three in each group. Four of these six fractures showed evidence of delayed union. To achieve quicker union and fewer implant failures we recommend the use of reamed nails of at least 12 mm in diameter for female patients and 13 mm in males.
We describe a method of closed, unlocked nailing for femoral fractures using ultrasound instead of an image intensifier. Radiography was used only to confirm that the guide wire had been passed into the intramedullary canal of both fragments. The method succeeded in 26 of 30 cases. The failures all occurred in fractures which could not be reduced within 20 minutes. The operating time in those nailed successfully with ultrasound control was not different from the time for 30 control cases using conventional methods with an image intensifier.
The December 2023 Children’s orthopaedics Roundup. 360. looks at: A comprehensive nonoperative treatment protocol for developmental dysplasia of the hip in infants; How common are refractures in childhood?;
Objectives. The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling. Methods. A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal
Proximal femur fractures treatment can involve anterograde nailing with a single or double cephalic screw. An undesirable failure for this fixation is screw cut-out. In a single-screw nail, a tip-apex distance (TAD) greater than 25 mm has been associated with an increased risk of cut-out. The aim of the study was to examine the role of TAD as a risk factor in a cephalic double-screw nail. A retrospective study was conducted on 112 patients treated for intertrochanteric femur fracture with a double proximal screw nail (Endovis BA2; EBA2) from January to September 2021. The analyzed variables were age, sex, BMI, comorbidities, fracture type, side, time of surgery, quality of reduction, pre-existing therapy with bisphosphonate for osteoporosis, screw placement in two different views, and TAD. The last follow-up was at 12 months. Logistic regression was used to study the potential factors of screw cut-out, and receiver operating characteristic curve to identify the threshold value.Aims
Methods
The use of fluoroscopy in orthopaedic surgery creates risk of radiation exposure to surgeons. Appropriate personal protective equipment (PPE) can help mitigate this. The primary aim of this study was to assess if current radiation protection in orthopaedic trauma is safe. The secondary aims were to describe normative data of radiation exposure during common orthopaedic procedures, evaluate ways to improve any deficits in protection, and validate the use of electronic personal dosimeters (EPDs) in assessing radiation dose in orthopaedic surgery. Radiation exposure to surgeons during common orthopaedic trauma operations was prospectively assessed using EPDs and thermoluminescent dosimeters (TLDs). Normative data for each operation type were calculated and compared to recommended guidelines.Aims
Methods
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles (. sd. 33 900) vs 12
800 cycles (. sd. 6100); p = 0.007) and mean axial stiffness
(220 N/mm (. sd. 80) vs 70 N/mm (. sd. 18);
p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (. sd. 0.9) vs 5.1
Nm/° (. sd. 1.5); p = 0.008). In addition, in the nail group
the mode of failure was either cut-out of the distal screws or breakage
of nails, and in the locking plate group breakage of the plate was always
the mode of failure. Locking nail constructs had significantly longer
mean fatigue life than non-locking nail constructs (78 900 cycles
(. sd. 25 600) vs 52 400 cycles (. sd. 22
500); p = 0.04). The new locking retrograde
Periprosthetic fractures (PPFs) around the knee are challenging injuries. This study aims to describe the characteristics of knee PPFs and the impact of patient demographics, fracture types, and management modalities on in-hospital mortality. Using a multicentre study design, independent of registry data, we included adult patients sustaining a PPF around a knee arthroplasty between 1 January 2010 and 31 December 2019. Univariate, then multivariable, logistic regression analyses were performed to study the impact of patient, fracture, and treatment on mortality.Aims
Methods