Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Trauma

Reamed versus unreamed femoral nails

A RANDOMISED, PROSPECTIVE TRIAL



Download PDF

Abstract

We performed a randomised, prospective trial to evaluate the use of unreamed titanium nails for femoral fractures. Of 48 patients with 50 femoral fractures 45 were followed to union; 23 with an unreamed and 22 with a reamed nail. The study was stopped early because of a high rate of implant failure.

The fractures in the unreamed group were slower to unite (39.4 weeks) than those in the reamed group (28.5 weeks; p = 0.007). The time to union was over nine months in 57% of the unreamed group and in 18% of the reamed group.

In the unreamed group 14 secondary procedures were required in ten patients to enhance healing compared with three in three patients in the reamed group. Six implants (13%) failed, three in each group. Four of these six fractures showed evidence of delayed union.

To achieve quicker union and fewer implant failures we recommend the use of reamed nails of at least 12 mm in diameter for female patients and 13 mm in males.


Correspondence should be sent to Dr M. G. Clatworthy at 34 Manawa Road, Remuera, Auckland 5, New Zealand.

For access options please click here