We reviewed the seven- to ten-year results of our previously reported prospective randomised controlled trial comparing total hip replacement and hemiarthroplasty for the treatment of displaced intracapsular fracture of the femoral neck. Of our original study group of 81 patients, 47 were still alive. After a mean follow up of nine years (7 to 10) overall mortality was 32.5% and 51.2% after total hip replacement and hemiarthroplasty, respectively (p = 0.09). At 100 months postoperatively a significantly greater proportion of hemiarthroplasty patients had died (p = 0.026). Three hips dislocated following total hip replacement and none after hemiarthroplasty. In both the total hip replacement and hemiarthroplasty groups a deterioration had occurred in walking distance (p = 0.02 and p <
0.001, respectively). One total hip replacement required revision compared with four hemiarthroplasties which were revised to total hip replacements. All surviving patients with a total hip replacement demonstrated wear of the cemented polyethylene component and all hemiarthroplasties had produced acetabular erosion. There was lower mortality (p = 0.013) and a trend towards superior function in patients with a total hip replacement in the medium term.
We have reviewed the
Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) Cite this article:
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
This study evaluated the feasibility of using published data from more than one register to define the performance of different hip implants. In order to obtain estimates of performance for specific types of hip system from different register, we analysed data from the annual reports of five national and one Italian regional register. We extracted the number of implants and rates of implant survival at different periods of follow-up. Our aim was to assess whether estimates of cumulative survival rate were comparable with data from registers from different countries, and our conclusion was that such a comparison could only be performed incompletely.
Percutaneous epiphysiodesis using transphyseal
screws (PETS) has been developed for the treatment of lower limb discrepancies
with the aim of replacing traditional open procedures. The goal
of this study was to evaluate its efficacy and safety at skeletal
maturity. A total of 45 consecutive patients with a mean skeletal
age of 12.7 years (8.5 to 15) were included and followed until maturity.
The mean efficacy of the femoral epiphysiodesis was 35% (14% to 87%)
at six months and 66% (21% to 100%) at maturity. The mean efficacy
of the tibial epiphysiodesis was 46% (18% to 73%) at six months
and 66% (25% to 100%) at maturity. In both groups of patients the
under-correction was significantly reduced between six months post-operatively
and skeletal maturity. The overall
In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.
We report ten-year clinical and radiological
follow-up data for the Sigma Press Fit Condylar total knee replacement system
(Sigma PFC TKR). Between October 1998 and October 1999 a total of
235 consecutive PFC Sigma TKRs were carried out in 203 patients.
Patients were seen at a specialist nurse-led clinic seven to ten
days before admission and at six and 18 months, three, five and
eight to ten years after surgery. Data were recorded prospectively
at each clinic visit. Radiographs were obtained at the five- and
eight- to ten-year follow-up appointments. Of the 203 patients,
147 (171 knees) were alive at ten years and 12 were lost to follow-up.
A total of eight knees (3.4%) were revised, five for infection and
three to change the polyethylene insert. The survival at ten years
with an endpoint of revision for any reason was 95.9%, and with
an endpoint of revision for aseptic failure was 98.7%. The mean
American Knee Society Score (AKSS) was 79 (10 to 99) at eight to
ten years, compared with 31 (2 to 62) pre-operatively. Of 109 knee
with radiographs reviewed, 47 knees had radiolucent lines but none
showed evidence of loosening. Cite this article:
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the tibial component at primary TKR. Cite this article:
An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated.Objectives
Methods
To review the current best surgical practice and detail a multi-disciplinary
approach that could further reduce joint replacement infection. Review of relevant literature indexed in PubMed.Objectives
Methods
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
The December 2012 Wrist &
Hand Roundup360 looks at: the imaging of scaphoid fractures; splinting to help Dupuytren’s disease; quality of life after nerve transfers; early failure of Moje thumbs; electra CMCJ arthroplasty; proximal interphalangeal joint replacement; pronator quadratus repair in distal radius fractures; and osteoporosis and wrist fractures.
The December 2012 Trauma Roundup360 looks at: whether tranexamic acid stops bleeding in trauma across the board; antibiotic beads and VAC; whether anaesthetic determines the outcome in surgery for distal radial fractures; high complications in surgery on bisphosphonate-hardened bone; better outcomes but more dislocations in femoral neck fractures; the mythical hip fracture; plate augmentation in nonunion surgery; and SIGN intramedullary nailing and infections.
We have investigated whether the use of laminar-flow theatres and space suits reduced the
We conducted a longitudinal study including patients with the same type of primary hybrid total hip replacement and evaluated patient activity and femoral osteolysis at either five or ten years post-operatively. Activity was measured using the University of California, Los Angeles scale. The primary outcome was the radiological assessment of femoral osteolysis. Secondary outcomes were revision of the femoral component for aseptic loosening and the patients’ quality of life. Of 503 hip replacements in 433 patients with a mean age of 67.7 years (30 to 91), 241 (48%) were seen at five and 262 (52%) at ten years post-operatively. Osteolytic lesions were identified in nine of 166 total hip replacements (5.4%) in patients with low activity, 21 of 279 (7.5%) with moderate activity, and 14 of 58 (24.1%) patients with high activity. The risk of osteolysis increased with participation in a greater number of sporting activities. In multivariate logistic regression adjusting for age, gender, body mass index and the inclination angle of the acetabular component, the adjusted odds ratio for osteolysis comparing high In conclusion, of patients engaged in high activity, 24% had developed femoral osteolysis five to ten years post-operatively.