Bone & Joint Research
Vol. 10, Issue 2 | Pages 105 - 112
1 Feb 2021
Feng X Qi W Fang CX Lu WW Leung FKL Chen B

Aims

To draw a comparison of the pullout strengths of buttress thread, barb thread, and reverse buttress thread bone screws.

Methods

Buttress thread, barb thread, and reverse buttress thread bone screws were inserted into synthetic cancellous bone blocks. Five screw-block constructs per group were tested to failure in an axial pullout test. The pullout strengths were calculated and compared. A finite element analysis (FEA) was performed to explore the underlying failure mechanisms. FEA models of the three different screw-bone constructs were developed. A pullout force of 250 N was applied to the screw head with a fixed bone model. The compressive and tensile strain contours of the midsagittal plane of the three bone models were plotted and compared.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1648 - 1655
1 Nov 2021
Jeong S Hwang K Oh C Kim J Sohn OJ Kim JW Cho Y Park KC

Aims

The incidence of atypical femoral fractures (AFFs) continues to increase. However, there are currently few long-term studies on the complications of AFFs and factors affecting them. Therefore, we attempted to investigate the outcomes, complications, and risk factors for complication through mid-term follow-up of more than three years.

Methods

From January 2003 to January 2016, 305 patients who underwent surgery for AFFs at six hospitals were enrolled. After exclusion, a total of 147 patients were included with a mean age of 71.6 years (48 to 89) and 146 of whom were female. We retrospectively evaluated medical records, and reviewed radiographs to investigate the fracture site, femur bowing angle, presence of delayed union or nonunion, contralateral AFFs, and peri-implant fracture. A statistical analysis was performed to identify the significance of associated factors.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1686 - 1694
1 Nov 2021
Yang H Kwak W Kang SJ Song E Seon J

Aims

To determine the relationship between articular cartilage status and clinical outcomes after medial opening-wedge high tibial osteotomy (MOHTO) for medial compartmental knee osteoarthritis at intermediate follow-up.

Methods

We reviewed 155 patients (155 knees) who underwent MOHTO from January 2008 to December 2016 followed by second-look arthroscopy with a mean 5.3-year follow-up (2.0 to 11.7). Arthroscopic findings were assessed according to the International Cartilage Repair Society (ICRS) Cartilage Repair Assessment (CRA) grading system. Patients were divided into two groups based on the presence of normal or nearly normal quality cartilage in the medial femoral condyle: good (second-look arthroscopic) status (ICRS grade I or II; n = 70), and poor (second-look arthroscopic) status (ICRS grade III or IV; n = 85) groups at the time of second-look arthroscopy. Clinical outcomes were assessed using the International Knee Documentation Committee (IKDC) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 36-Item Short Form survey.


Bone & Joint Open
Vol. 2, Issue 10 | Pages 796 - 805
1 Oct 2021
Plumarom Y Wilkinson BG Willey MC An Q Marsh L Karam MD

Aims

The modified Radiological Union Scale for Tibia (mRUST) fractures score was developed in order to assess progress to union and define a numerical assessment of fracture healing of metadiaphyseal fractures. This score has been shown to be valuable in predicting radiological union; however, there is no information on the sensitivity, specificity, and accuracy of this index for various cut-off scores. The aim of this study is to evaluate sensitivity, specificity, accuracy, and cut-off points of the mRUST score for the diagnosis of metadiaphyseal fractures healing.

Methods

A cohort of 146 distal femur fractures were retrospectively identified at our institution. After excluding AO/OTA type B fractures, nonunions, follow-up less than 12 weeks, and patients aged less than 16 years, 104 sets of radiographs were included for analysis. Anteroposterior and lateral femur radiographs at six weeks, 12 weeks, 24 weeks, and final follow-up were separately scored by three surgeons using the mRUST score. The sensitivity and specificity of mean mRUST score were calculated using clinical and further radiological findings as a gold standard for ultimate fracture healing. A receiver operating characteristic curve was also performed to determine the cut-off points at each time point.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1611 - 1618
1 Oct 2021
Kavarthapu V Budair B

Aims

In our unit, we adopt a two-stage surgical reconstruction approach using internal fixation for the management of infected Charcot foot deformity. We evaluate our experience with this functional limb salvage method.

Methods

We conducted a retrospective analysis of prospectively collected data of all patients with infected Charcot foot deformity who underwent two-stage reconstruction with internal fixation between July 2011 and November 2019, with a minimum of 12 months’ follow-up.


Bone & Joint Research
Vol. 8, Issue 7 | Pages 313 - 322
1 Jul 2019
Law GW Wong YR Yew AK Choh ACT Koh JSB Howe TS

Objectives

The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling.

Methods

A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until cut-out occurred. The medial migration distance (MMD) was recorded at the end of the testing cycles.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


Bone & Joint 360
Vol. 9, Issue 3 | Pages 31 - 34
1 Jun 2020


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 512 - 521
1 May 2019
Carter TH Duckworth AD White TO

Abstract

The medial malleolus, once believed to be the primary stabilizer of the ankle, has been the topic of conflicting clinical and biomechanical data for many decades. Despite the relevant surgical anatomy being understood for almost 40 years, the optimal treatment of medial malleolar fractures remains unclear, whether the injury occurs in isolation or as part of an unstable bi- or trimalleolar fracture configuration. Traditional teaching recommends open reduction and fixation of medial malleolar fractures that are part of an unstable injury. However, there is recent evidence to suggest that nonoperative management of well-reduced fractures may result in equivalent outcomes, but without the morbidity associated with surgery. This review gives an update on the relevant anatomy and classification systems for medial malleolar fractures and an overview of the current literature regarding their management, including surgical approaches and the choice of implants.

Cite this article: Bone Joint J 2019;101-B:512–521.


Bone & Joint 360
Vol. 10, Issue 2 | Pages 33 - 37
1 Apr 2021


Bone & Joint 360
Vol. 9, Issue 6 | Pages 22 - 27
1 Dec 2020


Bone & Joint 360
Vol. 9, Issue 4 | Pages 15 - 17
1 Aug 2020


Bone & Joint Open
Vol. 1, Issue 7 | Pages 376 - 382
10 Jul 2020
Gill JR Vermuyten L Schenk SA Ong JCY Schenk W

Aims

The aim of this study is to report the results of a case series of olecranon fractures and olecranon osteotomies treated with two bicortical screws.

Methods

Data was collected retrospectively for all olecranon fractures and osteotomies fixed with two bicortical screws between January 2008 and December 2019 at our institution. The following outcome measures were assessed; re-operation, complications, radiological loss of reduction, and elbow range of flexion-extension.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 394 - 399
1 Mar 2020
Parker MJ Cawley S

Aims

A lack of supporting clinical studies have been published to determine the ideal length of intramedullary nail in fixation of trochanteric fractures of the hip. Nevertheless, there has been a trend to use shorter intramedullary nails for the internal fixation of trochanteric hip fractures. Our aim was to determine if the length of nail affected the outcome.

Methods

We randomized 229 patients with a trochanteric hip fracture between two implants: a ‘standard’ nail of 220 mm and a shorter nail of 175 mm, which had decreased proximal angulation (4° vs 7°) and a reduced diameter at the level of the lesser trochanter. Patients were followed up for one year by a nurse blinded to the type of implant used to determine if there were differences in mobility and pain with two nail designs. Pain was assessed on a scale of 1 (none) to 8 (severe and constant) and mobility on a scale of 1 (full mobility) to 9 (immobile).


The Bone & Joint Journal
Vol. 102-B, Issue 4 | Pages 530 - 538
1 Apr 2020
Rollick NC Gadinsky NE Klinger CE Kubik JF Dyke JP Helfet DL Wellman DS

Aims

Dual plating of distal femoral fractures with medial and lateral implants has been performed to improve construct mechanics and alignment, in cases where isolated lateral plating would be insufficient. This may potentially compromise vascularity, paradoxically impairing healing. This study investigates effects of single versus dual plating on distal femoral vascularity.

Methods

A total of eight cadaveric lower limb pairs were arbitrarily assigned to either 1) isolated lateral plating, or 2) lateral and medial plating of the distal femur, with four specimens per group. Contralateral limbs served as matched controls. Pre- and post-contrast MRI was performed to quantify signal intensity enhancement in the distal femur. Further evaluation of intraosseous vascularity was done with barium sulphate infusion with CT scan imaging. Specimens were then injected with latex medium and dissection was completed to assess extraosseous vasculature.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 443 - 449
1 Apr 2018
Kalsbeek JH van Walsum ADP Vroemen JPAM Janzing HMJ Winkelhorst JT Bertelink BP Roerdink WH

Aims

The objective of this study was to investigate bone healing after internal fixation of displaced femoral neck fractures (FNFs) with the Dynamic Locking Blade Plate (DLBP) in a young patient population treated by various orthopaedic (trauma) surgeons.

Patients and Methods

We present a multicentre prospective case series with a follow-up of one year. All patients aged ≤ 60 years with a displaced FNF treated with the DLBP between 1st August 2010 and December 2014 were included. Patients with pathological fractures, concomitant fractures of the lower limb, symptomatic arthritis, local infection or inflammation, inadequate local tissue coverage, or any mental or neuromuscular disorder were excluded. Primary outcome measure was failure in fracture healing due to nonunion, avascular necrosis, or implant failure requiring revision surgery.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1385 - 1391
1 Oct 2019
Nicholson JA Gribbin H Clement ND Robinson CM

Aims

The primary aim of this study was to determine if delayed clavicular fixation results in a greater risk of operative complications and revision surgery.

Patients and Methods

A retrospective case series was undertaken of all displaced clavicular fractures that underwent plate fixation over a ten-year period (2007 to 2017). Patient demographics, time to surgery, complications, and mode of failure were collected. Logistic regression was used to identify independent risk factors contributing towards operative complications. Receiver operating characteristic (ROC) curve analysis was used to determine if a potential ‘safe window’ exists from injury to delayed surgery. Propensity score matching was used to construct a case control study for comparison of risk.


The Bone & Joint Journal
Vol. 99-B, Issue 11 | Pages 1496 - 1501
1 Nov 2017
Bali N Aktselis I Ramasamy A Mitchell S Fenton P

Aims

There has been an evolution recently in the management of unstable fractures of the ankle with a trend towards direct fixation of a posterior malleolar fragment. Within these fractures, Haraguchi type 2 fractures extend medially and often cannot be fixed using a standard posterolateral approach. Our aim was to describe the posteromedial approach to address these fractures and to assess its efficacy and safety.

Patients and Methods

We performed a review of 15 patients with a Haraguchi type 2 posterior malleolar fracture which was fixed using a posteromedial approach. Five patients underwent initial temporary spanning external fixation. The outcome was assessed at a median follow-up of 29 months (interquartile range (IQR) 17 to 36) using the Olerud and Molander score and radiographs were assessed for the quality of the reduction.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1226 - 1231
1 Sep 2015
Nakamura R Komatsu N Murao T Okamoto Y Nakamura S Fujita K Nishimura H Katsuki Y

The objective of this study was to validate the efficacy of Takeuchi classification for lateral hinge fractures (LHFs) in open wedge high tibial osteotomy (OWHTO). In all 74 osteoarthritic knees (58 females, 16 males; mean age 62.9 years, standard deviation 7.5, 42 to 77) were treated with OWHTO using a TomoFix plate. The knees were divided into non-fracture (59 knees) and LHF (15 knees) groups, and the LHF group was further divided into Takeuchi types I, II, and III (seven, two, and six knees, respectively). The outcomes were assessed pre-operatively and one year after OWHTO. Pre-operative characteristics (age, gender and body mass index) showed no significant difference between the two groups. The mean Japanese Orthopaedic Association score was significantly improved one year after operation regardless of the presence or absence of LHF (p = 0.0015, p < 0.001, respectively). However, six of seven type I cases had no LHF-related complications; both type II cases had delayed union; and of six type III cases, two had delayed union with correction loss and one had overcorrection. These results suggest that Takeuchi type II and III LHFs are structurally unstable compared with type I.

Cite this article: Bone Joint J 2015;97-B:1226–31.


Bone & Joint 360
Vol. 8, Issue 2 | Pages 23 - 26
1 Apr 2019