Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 76 - 76
1 Mar 2013
Holsgrove T Petheram T Miles AW Timperley AJ
Full Access

250 words max Long polished cemented femoral stems, such as the Exeter Hip Revision stem, are one option available to the revision hip arthroplasty surgeon. When proximal bone stock is compromised, distal fixation is often relied upon for stability of the femoral component. In such circumstances, torsional forces can result in debonding and loosening. This study compared the torsional behaviour of a cemented polished and featureless (plain) stem with cemented, polished stems featuring fins or flutes. Nine torsional tests were carried out on each of these three different stem designs. The finned stem construct was significantly stiffer than the fluted stem (mean 24.5 Nm/deg v 17.5 Nm/deg). The plain stem mean stiffness was less than the featured stems (13 Nm/deg), but wide variability lead to no statistically significant difference. The maximum torque of the finned (30.5 Nm) and fluted stems (29 Nm) was significantly higher than the plain stem (10.5 Nm); with no significance to the difference between the finned and fluted stems. Distal stem features may provide a more reliable and greater resistance to torque in polished, cemented revision hip stems. Finned stem features may also increase the stiffness of the construct. Consideration should thus be given to the incorporation of distal stem features in the design of revision hip stems


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 15 - 15
1 Mar 2013
Petheram T Bone M Joyce T Partington P
Full Access

Recent NICE guidance recommends use of a well proven cemented femoral stem for hip hemiarthroplasty in management of fractured neck of femur. The Exeter Trauma Stem (ETS) has been designed based on the well proven Exeter hip stem. It has a double taper polished stem design, proclaimed to share geometry and surface finish with the Exeter hip. This study investigated the surface finish of the two stems in order to investigate the hypothesis that they were different. Two ETS and two Exeter stems were examined using a profilometer with a sensitivity of one nanometer. Macroscopic visual inspection showed that the two Exeter stems had significantly smoother surface finish than the ETS stems. The roughness average (RA) values on the ETS stems were approximately an order of magnitude higher than those of the Exeter stems, mean of 0.235μm compared with 0.025μm (p<0.0001). This difference in surface finish has implications for the biomechanical functioning of the stem. Previous change of the Exeter stem to a matt surface-finish in 1976 resulted in a significant increase in stem failure rates and an understanding of the importance of the polished surface-finish in order to function within a taper-slip philosophy. By changing the surface finish in the ETS stem, longevity of the implant may similarly be affected. Clinical results have yet to be published demonstrating this. We recommend the manufacturer reconsiders the surface finish of the ETS stem to ensure it functions as well as the Exeter primary stem with which it shares a design philosophy


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 8 | Pages 1182 - 1188
1 Nov 2000
Barker DS Wang AW Yeo MF Nawana NS Brumby SA Pearcy MJ Howie DW

We studied the effect of the surface finish of the stem on the transfer of load in the proximal femur in a sheep model of cemented hip arthroplasty. Strain-gauge analysis and corresponding finite-element (FE) analysis were performed to assess the effect of friction and creep at the cement-stem interface. No difference was seen between the matt and polished stems. FE analysis showed that the effects of cement creep and friction at the stem-cement interface on femoral strain were small compared with the effect of inserting a cemented stem


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 665 - 669
1 Jul 1997
Verdonschot N Huiskes R

It has been suggested that the endurance of cemented femoral reconstructions in total hip arthroplasty is affected by the creep of acrylic cement, but it is not known to what extent cement creeps under loading conditions in vivo, or how this affects load transfer. We have simulated the long-term creep properties of acrylic cement in finite-element models of femoral stem constructs and analysed their effects. We investigated whether subsidence rates measured in vivo could be explained by creep of acrylic cement, and if polished, unbonded, stems accommodated creep better than bonded stems. Our findings showed that polished prostheses subsided only about 50 μm as a result of cement creep. The long-term prosthetic subsidence rates caused by creep of acrylic cement are therefore very small and do not explain the excessive migration rates which have sometimes been reported. Cement creep did, however, relax cement stresses and create a more favourable stress distribution at the interfaces. These trends were found around both the bonded and unbonded stems. Our results did not confirm that polished, unbonded, stems accommodated creep better than bonded stems in terms of cement and interface stress patterns


Bone & Joint 360
Vol. 10, Issue 4 | Pages 49 - 51
1 Aug 2021
Evans JT Welch M Whitehouse MR


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 31 - 31
1 Jan 2017
Morellato K Fink B Grupp T Cristofolini L
Full Access

The stem and the rasp for cemented arthroplasty are typically designed to obtain a cement mantle 2–5 mm thick. However, sometimes a line-to-line cementation is preferred, where the femoral cavity is prepared with the same dimension as the actual stem. There are contrasting reports [1,2] about the suitability of this technique to withstand the long-term fatigue loads. While the theoretical geometry allows no space for the cement, a sort of cement mantle is formed as the cement penetrates in the spongy bone. The scopes of this study were: 1) developing a dedicated in vitro method to test line-to-line cementation; 2) assessing if a short, polished hip stem designed for a standard cementation can be safely cemented line-to-line. In order to perform long-term mechanical in vitro tests, composite bones must be used, as cadaveric bones cannot withstand millions of loading cycles [3]. For this study, the Sawbones Mod. 3406-4 were chosen: they feature an open-cell polyurethane core simulating low-density spongy bone. Post-implantation x-rays confirmed that a relevant cement-bone interdigitation was obtained. Four femurs were prepared with a CoreHip (Aesculap) with regular cement mantle (Regular). Another 4 femurs were rasped to the same rasp size, and implanted with line-to-line cementation with a larger stem (Line-to-line). The implanted femurs were subjected to an accelerated test derived from a validated protocol [3] which replicates the most demanding motor tasks of 24 years of patient activity. Implant elastic micromotions and permanent migrations were measured throughout the test. The implants were then sectioned and treated with dye penetrants to highlight the cement cracks. Elastic and permanent motions did not show any loosening trend, and never exceeded few micrometers. As expected, some damage was visible in the cement mantles after test completion, for both types of implantation (similar to retrieved cement mantles surrounding stable implants [3]. The cement damage was similar in all specimens. No sign of major disruption was visible, neither within the Regular nor in the Line-to-line specimens: in fact, the cracks were limited in length, did not seem to cross the entire mantle thickness, and did not result in any loose cement fragments. The cracks in the line-to-line implants showed the same position and distribution compared to those found in the regular implants, but were slightly longer in some specimens. This in vitro study confirmed the feasibility of simulating line-to-line cementation in vitro. Our results suggest that a stem designed for a regular cement mantle could induce slightly more damage when implanted line-to-line, but no significant trend toward loosening


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_30 | Pages 8 - 8
1 Aug 2013
Shaw C Badhesha J Ayana G Abu-Rajab R
Full Access

The Exeter Stem (Howmedica, UK) has been in use for over 35 years. Over the years it has undergone several modifications with the most recent being a highly polished, tapered stem in 1986. The manufacturers quote a rate of 0.0006%. In the current literature there are 16 (or less) instances of fractures of the Orthinox stem. We present a case of fracture of an Orthinox Exeter Stem 9 years after insertion. Our patient, BB, presented, aged 62, with symptoms & signs consistent with OA right hip. THR was performed through a lateral approach utilising a trochanteric osteotomy. A size 0 37.5 stem was inserted. Radiographs were very satisfactory. She suffered a post operative DVT/PTE from which she recovered uneventfully. She was independently mobile at 6 month review and was discharged at the 2 year stage pain free. Aged 71, BB presented to outpatient clinic with a several month history of generalised groin pain. She had a Trendelenberg gait. Considerable pain was experienced on axial compression of the limb. Radiographs revealed a midstem fracture with cement loosening proximally. No trauma was reported. She underwent revision surgery through a posterior approach. Acetabular component was rigidly fixed. This was revised to a pressfit Trident (Zimmer, UK) cup with screws & polyethylene liner. An extended trochanteric osteotomy was used to remove the broken stem. An uncemented Restoration (Stryker, UK) stem was inserted with a 28mm head. Post-operative recovery was unremarkable and at 6 months osteotomy has healed. The stem was sent to Stryker UK Laboratories for analysis. They reported the stem broke in fatigue with the origin on the antero-lateral surface. No material or manufacturing defects seen. Dimensionally correct. Fracture may be due to abnormal bending stresses secondary to proximal loosening and firm distal fixation. Our case demonstrates a set of circumstances that led to inevitable fatigue and stem fracture. The method of failure should reinforce the radiograph appearances that may cause concern or be acted upon


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 6 | Pages 973 - 978
1 Nov 1996
Malkani AL Voor MJ Fee KA Bates CS

We have tested the axial and torsional stability of femoral components after revision arthroplasty in a cadaver model, using impacted morsellised cancellous graft and cement. Each one of six matched pairs of fresh frozen human femora had either a primary or a revision prosthesis cemented in place. For the ‘revision’ experiments, all cancellous bone was removed from the proximal femur which was then over-reamed to create a smooth-walled cortical shell. An MTS servohydraulic test frame was used to apply axial and torsional loads to each specimen through the prosthetic femoral heads with the femur submerged in isotonic saline solution at 37°C. The mean subsidence was 0.27 ± 0.17 mm for the primary and 0.52 ± 0.30 mm for the revision groups. The difference was statistically significant (p < 0.025), but the mean subsidence was < 1 mm in both groups. The mean maximum torque before failure was 42.9 ± 26.9 N-m for the primary and 34.8± 20.7 N-m for the revision groups. This difference was not statistically significant (p > 0.015). Based on our results we suggest that revision of the femoral component using morsellised cancellous graft followed by cementing with a collarless prosthesis with a polished tapered stem restores the integrity of the proximal femur and provides immediate stability of the implant


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1298 - 1302
1 Sep 2005
Iesaka K Jaffe WL Jones CM Kummer FJ

We have investigated the role of the penetration of saline on the shear strength of the cement-stem interface for stems inserted at room temperature and those preheated to 37°C using a variety of commercial bone cements. Immersion in saline for two weeks at 37°C reduced interfacial strength by 56% to 88% after insertion at room temperature and by 28% to 49% after preheating of the stem. The reduction in porosity as a result of preheating ranged from 71% to 100%. Increased porosity correlated with a reduction in shear strength after immersion in saline (r = 0.839, p < 0.01) indicating that interfacial porosity may act as a fluid conduit.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 454 - 460
1 Mar 2010
Baleani M Bialoblocka-Juszczyk E Engels GE Viceconti M

We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement.

Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue tests, and were examined for porosity.

Pre-heating the stem alone significantly increased the mean temperature values measured within the mantle (+14.2°C) (p < 0.001) and reduced the mean curing time (−1.5 min) (p < 0.001). The addition of vacuum mixing modulated the mean rise in the temperature of polymerisation to 11°C and reduced the mean duration of the process by one minute and 50 seconds (p = 0.01 and p < 0.001, respectively). In all cases, the maximum temperature values measured in the mould simulating the femur were < 50°C. The mixing technique and pre-heating the stem slightly increased the static mechanical strength of bone cement. However, the fatigue life of the cement was improved by both vacuum mixing and pre-heating the stem, but was most marked (+ 280°C) when these methods were combined.

Pre-heating the stem appears to be an effective way of improving the quality of the cement mantle, which might enhance the long-term performance of bone cement, especially when combined with vacuum mixing.