Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 11 - 11
1 Jun 2016
O'Neill C Molloy D Patterson C Beverland D
Full Access

Introduction

Radiological Inclination (RI) is defined as the angle formed between the acetabular axis and the longitudinal axis when projected onto the coronal plane. Higher RI angles are associated with adverse outcomes.

Methods

Primary aim: to investigate the effect of adjusting patient pelvic position in the transverse plane by using a ‘head-down’ (HD) operating table position. This was to determine, when aiming for 35° Apparent Operative Inclination (AOI), which operating table position most accurately achieved a target post-operative RI of 42°.

N=270. Patients were randomised to one of three possible operating table positions:

0°HD (Horizontal),

7°HD, or

Y°HD (Patient Specific Table Position)

Operating table position was controlled using a digital inclinometer. RI was measured using EBRA software.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 32 - 32
1 Nov 2015
O'Neill C Molloy D Patterson C Beverland D
Full Access

Introduction

Operative inclination (OI) is defined as the angle between the acetabular axis and the sagittal plane. With the patient in the true lateral decubitus position, this corresponds to the angle formed between the handle of the acetabular component inserter and the theatre floor intra-operatively.

Patients/Materials & Methods

The primary study aim was to determine which method of acetabular component insertion most accurately allows the surgeon to obtain a target OI of 35o.

270 consecutive patients undergoing cementless THA were randomised to one of three possible methods for acetabular component implantation:

1. Freehand,

2. 35o mechanical alignment guide (MAG), or

3. Digital inclinometer assisted

Two surgeons participated. Target OI was 35o in all cases. OI was measured using a digital inclinometer. For the freehand and MAG cases, the surgeon was blinded to inclinometer readings intra-operatively.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 164 - 172
1 Feb 2015
Grammatopoulos G Thomas GER Pandit H Beard DJ Gill HS Murray DW

We assessed the orientation of the acetabular component in 1070 primary total hip arthroplasties with hard-on-soft, small diameter bearings, aiming to determine the size and site of the target zone that optimises outcome. Outcome measures included complications, dislocations, revisions and ΔOHS (the difference between the Oxford Hip Scores pre-operatively and five years post-operatively). A wide scatter of orientation was observed (2sd 15°). Placing the component within Lewinnek’s zone was not associated withimproved outcome. Of the different zone sizes tested (± 5°, ± 10° and ± 15°), only ± 15° was associated with a decreased rate of dislocation. The dislocation rate with acetabular components inside an inclination/anteversion zone of 40°/15° ± 15° was four times lower than those outside. The only zone size associated with statistically significant and clinically important improvement in OHS was ± 5°. The best outcomes (ΔOHS > 26) were achieved with a 45°/25° ± 5° zone.

This study demonstrated that with traditional technology surgeons can only reliably achieve a target zone of ±15°. As the optimal zone to diminish the risk of dislocation is also ±15°, surgeons should be able to achieve this. This is the first study to demonstrate that optimal orientation of the acetabular component improves the functional outcome. However, the target zone is small (± 5°) and cannot, with current technology, be consistently achieved.

Cite this article: Bone Joint J 2015;97-B:164–72.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 12 - 12
1 Jun 2016
Meermans G Peeters W Van Doorn W Kats J
Full Access

Introduction

In total hip arthroplasty (THA), a high radiographic inclination angle (RI) of the acetabular component has been linked to an increased dislocation rate, liner fracture, and increased wear. In contrast to version, we have more proven boundaries when it comes to a safe zone for angles of RI. Although intuitively it seems easier to achieve a target RI, most studies demonstrate a lack of accuracy and the trend towards a high RI with all surgical approaches when using a freehand technique or a mechanical guide. This is due to pelvic motion during surgery, which can be highly variable.

The current study had two primary aims, each with a different primary outcome. The first aim was to determine how accurate a surgeon could obtain the target operative inclination (OI) during THA when using a cementless cup using a digital protractor. The second aim was to determine how accurate a surgeon can estimate the target OI to obtain a RI of 40° based on the patient's hip circumference as demonstrated in a previous study.

Methods

In this prospective study, we included 200 consecutive patients undergoing uncemented primary THA in the lateral decubitus position using a posterior approach. Preoperatively, the surgeon determined the target OI based on the patient's hip circumference (22.5°, 25°, 27.5° or 30°). Intraoperatively, the effective OI was measured with the aid of a digital inclinometer after seating of the acetabular component. Six weeks postoperatively anteroposterior pelvic radiographs were made and two evaluators, blinded to the effective OI, measured the RI of the acetabular component. The safe zone for inclination was defined as 30°-45° of inclination.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1662 - 1668
1 Nov 2021
Bhanushali A Chimutengwende-Gordon M Beck M Callary SA Costi K Howie DW Solomon LB

Aims. The aims of this study were to compare clinically relevant measurements of hip dysplasia on radiographs taken in the supine and standing position, and to compare Hip2Norm software and Picture Archiving and Communication System (PACS)-derived digital radiological measurements. Methods. Preoperative supine and standing radiographs of 36 consecutive patients (43 hips) who underwent periacetabular osteotomy surgery were retrospectively analyzed from a single-centre, two-surgeon cohort. Anterior coverage (AC), posterior coverage (PC), lateral centre-edge angle (LCEA), acetabular inclination (AI), sharp angle (SA), pelvic tilt (PT), retroversion index (RI), femoroepiphyseal acetabular roof (FEAR) index, femoroepiphyseal horizontal angle (FEHA), leg length discrepancy (LLD), and pelvic obliquity (PO) were analyzed using both Hip2Norm software and PACS-derived measurements where applicable. Results. Analysis of supine and standing radiographs resulted in significant variation for measurements of PT (p < 0.001) and AC (p = 0.005). The variation in PT correlated with the variation in AC in a limited number of patients (R. 2. = 0.378; p = 0.012). Conclusion. The significant variation in PT and AC between supine and standing radiographs suggests that it may benefit surgeons to have both radiographs when planning surgical correction of hip dysplasia. We also recommend using PACS-derived measurements of AI and SA due to the poor interobserver error on Hip2Norm. Cite this article: Bone Joint J 2021;103-B(11):1662–1668


Bone & Joint Research
Vol. 10, Issue 10 | Pages 639 - 649
19 Oct 2021
Bergiers S Hothi H Henckel J Di Laura A Belzunce M Skinner J Hart A

Aims. Acetabular edge-loading was a cause of increased wear rates in metal-on-metal hip arthroplasties, ultimately contributing to their failure. Although such wear patterns have been regularly reported in retrieval analyses, this study aimed to determine their in vivo location and investigate their relationship with acetabular component positioning. Methods. 3D CT imaging was combined with a recently validated method of mapping bearing surface wear in retrieved hip implants. The asymmetrical stabilizing fins of Birmingham hip replacements (BHRs) allowed the co-registration of their acetabular wear maps and their computational models, segmented from CT scans. The in vivo location of edge-wear was measured within a standardized coordinate system, defined using the anterior pelvic plane. Results. Edge-wear was found predominantly along the superior acetabular edge in all cases, while its median location was 8° (interquartile range (IQR) -59° to 25°) within the anterosuperior quadrant. The deepest point of these scars had a median location of 16° (IQR -58° to 26°), which was statistically comparable to their centres (p = 0.496). Edge-wear was in closer proximity to the superior apex of the cups with greater angles of acetabular inclination, while a greater degree of anteversion influenced a more anteriorly centred scar. Conclusion. The anterosuperior location of edge-wear was comparable to the degradation patterns observed in acetabular cartilage, supporting previous findings that hip joint forces are directed anteriorly during a greater portion of walking gait. The further application of this novel method could improve the current definition of optimal and safe acetabular component positioning. Cite this article: Bone Joint Res 2021;10(10):639–649


Highly cross-linked polyethylene (HXLPE) has decreased wear and revision rates in total hip replacement (THR) at a long-term. However, the effect of HXLPE manufacturing characteristics on femoral head penetration has not been clearly defined yet. We report this single-institution study to investigate the clinical and radiological results of different HXLPE liners in THR. In this retrospective cohort analysis of our prospective database, we identified 904 THRs performed between 2000 and 2013. Seven different HXLPE liner types were assessed: remelted (3), annealed (2), sequentially annealed (1) and vitamin E-infused (1). The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Roman Software v1.70 package. Thirty hips were revised for the following reasons: aseptic loosening (11), dislocation (12), periprosthetic femoral fracture (6), and infection (1). No hip was revised for wear or osteolysis. 741 THRs were evaluated for a mean follow-up of 15 years (range, 10 to 20). The mean total penetration 15 years after THR was 0.17 mm in the sequentially annealed, and 0.16 mm in the vitamin E-infused groups, whereas it was 0.26 mm in the melted 95 kGy, 0.27 mm in the melted 5 Mrad and 0.25 mm in the melted 100 kGy groups (p=0.001). From one to 15 years after surgery, the mean wear was 0.206 mm in cups with an acetabular inclination greater than 50º and 0.105 mm in those placed between 35º and 50º (p<0.001). Although HXLPE characteristics can result in a different wear performance in patients undergoing THR at a mean follow-up of 15 years, the clinical results are excellent. The position of the acetabular component can influence on the femoral head penetration of modern HXLPEs. Nevertheless, these patients should continue to be monitored to detect future problems


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 69 - 69
19 Aug 2024
Harris MD Thapa S Lieberman EG Pascual-Garrido C Abu-Amer W Nepple JJ Clohisy JC
Full Access

Developmental dysplasia of the hip can cause pain and premature osteoarthritis. However, the risk factors and timing for disease progression in young adults are not fully defined. This study identified the incidence and risk factors for contralateral hip pain and surgery after periacetabular osteotomy (PAO) on an index dysplastic hip. Patients followed for 2+ years after unilateral PAO were grouped by eventual contralateral pain or no-pain, based on modified Harris Hip Score, and surgery or no-surgery. Univariate analysis tested group differences in demographics, radiographic measures, and range-of-motion. Kaplan-Meier survival analysis assessed pain development and contralateral hip surgery over time. Multivariate regression identified pain and surgery risk factors. Pain and surgery predictors were further analyzed in Dysplastic, Borderline, and Non-dysplastic subcategories, and in five-degree increments of lateral center edge angle (LCEA) and acetabular inclination (AI). 184 patients were followed for 4.6±1.6 years, during which 51% (93/184) reported hip pain and 33% (60/184) underwent contralateral surgery. Kaplan-Meier analysis predicted 5-year survivorship of 49% for pain development and 66% for contralateral surgery. Painful hips exhibited more severe dysplasia than no-pain hips (LCEA 16.5º vs 20.3º, p<0.001; AI 13.2º vs 10.0º p<0.001). AI was the sole predictor of pain, with every 1° AI increase raising the risk by 11%. Surgical hips also had more severe dysplasia (LCEA 14.9º vs 20.0º, p<0.001; AI 14.7º vs 10.2º p<0.001) and were younger (21.6 vs 24.1 years, p=0.022). AI and a maximum alpha angle ≥55° predicted contralateral surgery. 5 years after index hip PAO, 51% of contralateral hips experience pain and 34% percent are expected to need surgery. More severe dysplasia, based on LCEA and AI, increases the risk of contralateral hip pain and surgery, with AI being a predictor of both outcomes. Knowing these risks can inform patient counseling and treatment planning


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 43 - 43
1 Aug 2018
Nepple J Graesser E Wells J Clohisy J
Full Access

The purpose of this study was to examine a cohort of patients with minor acetabular dysplasia features in order to identify the preoperative clinical characteristics and imaging findings that differentiate patients with hip instability from patients with impingement. A retrospective cohort study of patients with borderline acetabular dysplasia was performed. All patients were identified by prospective radiographic evaluation with an LCEA between 20° and 25°. Multivariate statistical analyses were used to identify independent predictors of disease type. Of the 143 hips in the cohort, 39.2% (n=56) had the diagnosis of instability, while 60.8% (n=87) had the diagnosis of impingement. The cohort included 109 females (76.2%) and 34 males (23.8%). Hips with instability had a lower LCEA (21.8° vs. 22.8°; p<0.001), lower ACEA (23.3° vs. 26.6°; p=0.002), a higher AI (11.8° vs. 8.5°; p<0.001), and a lower maximum alpha angle (54.4° vs. 61.1°; p=0.001). The odds of instability increased 1.7 times for each one-degree decrease in LCEA, 1.4 times for each one-degree decrease in ACEA, and 1.1 times for each one-degree increase in acetabular inclination (all p0.003). Female sex was strongly associated with instability. The instability subgroup had greater range of motion (IRF, 22.7° vs. 12.4°, p<0.001) and total arc of motion (IRF+ERF, 61.2° vs. 47.4°, p<0.001). We identified predictors of diagnosis including: acetabular inclination (1.49, p<0.001), ACEA (0.89, p=0.007), crossover sign (0.27, p=0.014), preoperative mHHS (0.96, p=0.014), IRF (1.10, p=0.001), and age (0.88, p=0.001). Patients with symptomatic instability tend to have increased acetabular inclination, decreased ACEA, greater functional limitations, younger, greater IRF, while hips with impingement demonstrate the opposite trends


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 55 - 55
1 Nov 2021
Nepple J Freiman S Pashos G Thornton T Schoenecker P Clohisy J
Full Access

Residual Legg-Calve-Perthes (LCP) deformities represent one of the most challenging disorders in hip reconstructive surgery. We assessed mid-term PRO. M. s, radiographic correction, complications and survivorship of combined surgical dislocation (SD) and periacetabular osteotomy (PAO) for the treatment of complex LCP deformities. A longitudinal cohort study was performed on 31 hips with complex LCP deformities undergoing combined SD/PAO. Treatment included femoral head reshaping, trochanteric advancement and relative neck lengthening, management of intra-articular lesions and PAO. Twenty-seven (87.1%) had minimum 5-year follow up. Average age was 19.8 years with 56% female and 44% having previous surgery. At a mean 8.4 years, 85% (23/27) of the hips remain preserved (no conversion to THA). The survivorship estimates at 5 and 10 years were 93% and 85%, respectively. The median and interquartile range for mHHS increased from 64 [55–67] to 92 [70–97] (p<0.001), the WOMAC-pain improved from 60 [45–75] to 86 [75–100] (p= 0.001). An additional 19% (n=5) reported symptoms (mHHS <70) at final follow-up. UCLA activity score increased from a median of 8 [6–10] to 9 [7–10] (p=0.207). Structural correction included average improvements of acetabular inclination 15.3. o. ± 7.6, LCEA 20.7° ± 10.8, ACEA 23.4° ± 16.3, and trochanteric height 18 mm ± 10 mm (all, p<0.001). Complications occurred in two (7%) patients including one deep and one superficial wound infection. At intermediate follow-up of combined SD/PAO for complex LCP deformities, 85% of hips are preserved. This procedure provides reliable deformity correction, major pain relief, improved function and acceptable complication/failure rates


Bone & Joint Open
Vol. 3, Issue 6 | Pages 475 - 484
13 Jun 2022
Jang SJ Vigdorchik JM Windsor EW Schwarzkopf R Mayman DJ Sculco PK

Aims. Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error. Methods. A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (. Δ. sacral slope(SS). stand-sit. > 30°), or stiff (. ∆. SS. stand-sit. < 10°) spinopelvic mobility contributed to increased error rates. Results. The paired absolute difference between CAS and postoperative imaging measurements was 2.3° (standard deviation (SD) 2.6°) for inclination and 3.1° (SD 4.2°) for anteversion. Using a target zone of 40° (± 10°) (inclination) and 20° (± 10°) (anteversion), postoperative standing radiographs measured 96% of acetabular components within the target zone for both inclination and anteversion. Multiple logistic regression analysis controlling for BMI and sex revealed that hypermobile spinopelvic mobility significantly increased error rates for anteversion (odds ratio (OR) 2.48, p = 0.009) and inclination (OR 2.44, p = 0.016), whereas stiff spinopelvic mobility increased error rates for anteversion (OR 1.97, p = 0.028). There were no dislocations at a minimum three-year follow-up. Conclusion. Despite high reliability in acetabular positioning for inclination in a large patient cohort using an optical CAS system, hypermobile and stiff spinopelvic mobility significantly increased the risk of clinically relevant errors. In patients with abnormal spinopelvic mobility, CAS systems should be adjusted for use to avoid acetabular component misalignment and subsequent risk for long-term dislocation. Cite this article: Bone Jt Open 2022;3(6):475–484


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 16 - 16
1 Aug 2021
Gupta V Thomas C Parsons H Metcalfe A Foguet P King R
Full Access

Total hip arthroplasty (THA) is one of the most successful surgical procedures of modern times, however debate continues as to the optimal orientation of the acetabular component and how to reliably achieve this. We hypothesised that functional CT-based planning with patient specific instruments using the Corin Optimised Positioning System (OPS) would provide more accurate component alignment than the conventional freehand technique using 2D templating. A pragmatic single-centre, patient-assessor blinded, randomised control trial of patients undergoing THA was performed. 54 patients (age 18–70) were recruited to either OPS THA or conventional THA. All patients received a cementless acetabular component. Patients in both arms underwent pre- and post-operative CT scans, and four functional x-rays (standing and seated). Patients in the OPS group had a 3D surgical plan and bespoke guides made. Patients in the conventional group had a surgical plan based on 2D templating x-rays, and the pre-operative target acetabular orientation was recorded by the surgeon. The primary outcome measure was the difference between planned and achieved acetabular anteversion and was determined by post-operative CT scan performed at 6 weeks. Secondary outcome measures included Hip disability and Osteoarthritis Outcome Score (HOOS), Oxford Hip Score (OHS), EQ-5D and adverse events. In the OPS group, the achieved acetabular anteversion was within 10° of the plan in 96% of cases, compared with only 76% of cases in the conventional group. For acetabular inclination, the achieved position in the OPS group was within 10° of the plan in 96% of cases, compared with in only 84% of cases in the conventional group. These differences were not statistically significant. The clinical outcomes were comparable between the two groups. Large errors in acetabular orientation appear to be reduced when functional CT-based planning and patient-specific instruments are used compared to the freehand technique, but no statistically significant differences were seen in the difference between planned and achieved angles. Larger studies are needed to analyse this in more detail and to determine whether the reduced numbers of outliers lead to improved clinical outcomes


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 36 - 43
1 Jan 2018
Hambright D Hellman M Barrack R

Aims. The aims of this study were to examine the rate at which the positioning of the acetabular component, leg length discrepancy and femoral offset are outside an acceptable range in total hip arthroplasties (THAs) which either do or do not involve the use of intra-operative digital imaging. Patients and Methods. A retrospective case-control study was undertaken with 50 patients before and 50 patients after the integration of an intra-operative digital imaging system in THA. The demographics of the two groups were comparable for body mass index, age, laterality and the indication for surgery. The digital imaging group had more men than the group without. Surgical data and radiographic parameters, including the inclination and anteversion of the acetabular component, leg length discrepancy, and the difference in femoral offset compared with the contralateral hip were collected and compared, as well as the incidence of altering the position of a component based on the intra-operative image. Results. Digital imaging took a mean of five minutes (2.3 to 14.6) to perform. Intra-operative changes with the use of digital imaging were made for 43 patients (86%), most commonly to adjust leg length and femoral offset. There was a decrease in the incidence of outliers when using intra-operative imaging compared with not using it in regard to leg length discrepancy (20% versus 52%, p = 0.001) and femoral offset inequality (18% versus 44%, p = 0.004). There was also a difference in the incidence of outliers in acetabular inclination (0% versus 7%, p = 0.023) and version (0% versus 4%, p = 0.114) compared with historical results of a high-volume surgeon at the same centre. Conclusion. The use of intra-operative digital imaging in THA improves the accuracy of the positioning of the components at THA without adding a substantial amount of time to the operation. Cite this article: Bone Joint J 2018;100B(1 Supple A):36–43


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 760 - 767
1 Jul 2019
Galea VP Rojanasopondist P Laursen M Muratoglu OK Malchau H Bragdon C

Aims. Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. Patients and Methods. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level, acetabular inclination, anteversion, and head size. Plain radiographs were assessed for radiolucency and patient-reported outcome measures (PROMs) were administered at each follow-up. Results. In total, 73 patients (82%) completed the seven-year visit. Mean seven-year linear proximal penetration was -0.07 mm (. sd. 0.16) and 0.00 mm (. sd. 0.22) for the VEPE and ModXLPE cohorts, respectively (p = 0.116). PROMs (p = 0.310 to 0.807) and radiolucency incidence (p = 0.330) were not different between the polyethylene cohorts. The mean proximal shell migration rate was 0.04 mm per year (. sd. 0.09). At seven years, patients with radiolucency (34%) demonstrated greater migration (mean difference: 0.6 mm (. sd. 0.2); p < 0.001). PROMs were lower for patients with radiolucency and greater proximal migration (p = 0.009 to p = 0.045). No implants were revised for aseptic loosening. Conclusion. This is the first randomized controlled trial to report seven-year RSA results for VEPE. All wear rates were below the previously reported osteolysis threshold (0.1 mm per year). PTC shells demonstrated acceptable primary stability through seven years, as indicated by low migration and lack of aseptic loosening. However, patients with acetabular radiolucency were associated with higher shell migration and lower PROM scores. Cite this article: Bone Joint J 2019;101-B:760–767


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 65 - 65
19 Aug 2024
Walter W Lin D Weinrauch P de Smet K Beaule P Young D Xu J Manktelow A
Full Access

Hip resurfacing arthroplasty (HRA) is a bone conserving alternative to total hip arthroplasty. We present the early 2-year clinical and radiographic follow-up of a novel ceramic-on-ceramic (CoC) HRA in an international multi-centric cohort. Patients undergoing HRA between September 2018 and January 2021 were prospectively included. Patient-reported outcome measures (PROMS) in the form of the Forgotten Joint Score (FJS), HOOS Jr, WOMAC, Oxford Hip Score (OHS) and UCLA Activity Score were collected preoperatively and at 1- and 2-years post-operation. Serial radiographs were assessed for migration, component alignment, evidence of osteolysis/loosening and heterotopic ossification formation. 200 patients were identified to have reached 2-year follow-up. Of these, 185 completed PROMS follow-up at 2 years. There was significant improvement in HOOS (p< 0.001) and OHS (p< 0.001) and FJS (p< 0.001) between the pre-operative and 2-year outcomes. Patients reported improved pain (p<0.001), function (p<0.001) and reduced stiffness (p<0.001) as measured by the WOMAC score. Patients had improved activity scores on the UCLA Active Score (P<0.001) with 53% reporting return to impact activity at 2 years. There was no osteolysis and the mean acetabular cup inclination angle was 41deg and the femoral component shaft angle was 137deg. No fractures were reported over but there was one sciatic nerve palsy with partial recovery. Two patients were revised; one at 3 months for pain due to a misdiagnosed back problem and another at 33 months for loosening of the acetabular component with delamination of the titanium ingrowth surface. CoC resurfacing at 2-years post-operation demonstrate promising results with satisfactory PROMS


Bone & Joint Open
Vol. 2, Issue 6 | Pages 365 - 370
1 Jun 2021
Kolodychuk N Su E Alexiades MM Ren R Ojard C Waddell BS

Aims. Traditionally, acetabular component insertion during total hip arthroplasty (THA) is visually assisted in the posterior approach and fluoroscopically assisted in the anterior approach. The present study examined the accuracy of a new surgeon during anterior (NSA) and posterior (NSP) THA using robotic arm-assisted technology compared to two experienced surgeons using traditional methods. Methods. Prospectively collected data was reviewed for 120 patients at two institutions. Data were collected on the first 30 anterior approach and the first 30 posterior approach surgeries performed by a newly graduated arthroplasty surgeon (all using robotic arm-assisted technology) and was compared to standard THA by an experienced anterior (SSA) and posterior surgeon (SSP). Acetabular component inclination, version, and leg length were calculated postoperatively and differences calculated based on postoperative film measurement. Results. Demographic data were similar between groups with the exception of BMI being lower in the NSA group (27.98 vs 25.2; p = 0.005). Operating time and total time in operating room (TTOR) was lower in the SSA (p < 0.001) and TTOR was higher in the NSP group (p = 0.014). Planned versus postoperative leg length discrepancy were similar among both anterior and posterior surgeries (p > 0.104). Planned versus postoperative abduction and anteversion were similar among the NSA and SSA (p > 0.425), whereas planned versus postoperative abduction and anteversion were lower in the NSP (p < 0.001). Outliers > 10 mm from planned leg length were present in one case of the SSP and NSP, with none in the anterior groups. There were no outliers > 10° in anterior or posterior for abduction in all surgeons. The SSP had six outliers > 10° in anteversion while the NSP had none (p = 0.004); the SSA had no outliers for anteversion while the NSA had one (p = 0.500). Conclusion. Robotic arm-assisted technology allowed a newly trained surgeon to produce similarly accurate results and outcomes as experienced surgeons in anterior and posterior hip arthroplasty. Cite this article: Bone Jt Open 2021;2(6):365–370


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 321 - 328
1 Feb 2021
Vandeputte F Vanbiervliet J Sarac C Driesen R Corten K

Aims. Optimal exposure through the direct anterior approach (DAA) for total hip arthroplasty (THA) conducted on a regular operating theatre table is achieved with a standardized capsular releasing sequence in which the anterior capsule can be preserved or resected. We hypothesized that clinical outcomes and implant positioning would not be different in case a capsular sparing (CS) technique would be compared to capsular resection (CR). Methods. In this prospective trial, 219 hips in 190 patients were randomized to either the CS (n = 104) or CR (n = 115) cohort. In the CS cohort, a medial based anterior flap was created and sutured back in place at the end of the procedure. The anterior capsule was resected in the CR cohort. Primary outcome was defined as the difference in patient-reported outcome measures (PROMs) after one year. PROMs (Harris Hip Score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), and Short Form 36 Item Health Survey (SF-36)) were collected preoperatively and one year postoperatively. Radiological parameters were analyzed to assess implant positioning and implant ingrowth. Adverse events were monitored. Results. At one year, there was no difference in HSS (p = 0.728), HOOS (Activity Daily Life, p = 0.347; Pain, p = 0.982; Quality of Life, p = 0.653; Sport, p = 0.994; Symptom, p = 0.459), or SF-36 (p = 0.338). Acetabular component inclination (p = 0.276) and anteversion (p = 0.392) as well as femoral component alignment (p = 0.351) were similar in both groups. There were no dislocations, readmissions, or reoperations in either group. The incidence of psoas tendinitis was six cases in the CS cohort (6%) and six cases in the CR cohort (5%) (p = 0.631). Conclusion. No clinical differences were found between resection or preservation of the anterior capsule when performing a primary THA through the DAA on a regular theatre table. In case of limited visibility during the learning curve, it might be advisable to resect a part of the anterior capsule. Cite this article: Bone Joint J 2021;103-B(2):321–328


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1428 - 1434
1 Nov 2008
Dandachli W Kannan V Richards R Shah Z Hall-Craggs M Witt J

We present a new CT-based method which measures cover of the femoral head in both normal and dysplastic hips and allows assessment of acetabular inclination and anteversion. A clear topographical image of the head with its covered area is generated. We studied 36 normal and 39 dysplastic hips. In the normal hips the mean cover was 73% (66% to 81%), whereas in the dysplastic group it was 51% (38% to 64%). The significant advantage of this technique is that it allows the measurements to be standardised with reference to a specific anatomical plane. When this is applied to assessing cover in surgery for dysplasia of the hip it gives a clearer understanding of where the corrected hip stands in relation to normal and allows accurate assessment of inclination and anteversion


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 315 - 320
1 Mar 2011
Hart AJ Ilo K Underwood R Cann P Henckel J Lewis A Cobb J Skinner J

We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine. We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of > 5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups. We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 9 - 9
1 Nov 2021
Farey J Chai Y Xu J Sadegpour A Jones DM Baker N Vigdorchik J Walter W
Full Access

Imageless computer navigation systems in total hip arthroplasty (THA) improve acetabular cup position, thereby reducing the risk of revision surgery for all causes as well as dislocation. We aimed to evaluate the registration accuracy of 3 alternate registration planes. A prospective, observational study was conducted with 45 THA in the supine position using two imageless navigation systems and 3 registration planes. Patient position was registered sequentially using an optical system (Stryker OrthoMap) and an inertial sensor-based system (Navbit Sprint) with 3 planes of reference: (Plane 1) an anatomical plane using the anterior superior iliac spines (ASISs) and the pubic symphysis; (Plane 2) a functional plane parallel to the line between the ASISs and the table plane; and, (Plane 3) a functional plane that was perpendicular to the gravity vector and aligned with the longitudinal axis of the patient. The 3 measurements of acetabular cup inclination and anteversion were compared with the measurements from postoperative computed tomography (CT) scans. For inclination, the mean absolute error was significantly lower for Plane 3 (1.80°) than for Plane 2 (2.74°), p = .038 and was lower for both functional planes than for the anatomical plane (3.75°), p < .001. For anteversion, the mean absolute error was significantly lower for Plane 3 (2.00°) than for Plane 2 (3.69°), p = .004 and was lower for both functional planes than for the anatomical plane (8.58°), p < .001. Patient registration using functional planes more accurately measured the acetabular cup position than registration using anatomic planes