header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Hip

RANDOMIZED CONTROLLED TRIAL OF FUNCTIONAL CT-BASED PLANNING WITH PATIENT-SPECIFIC INSTRUMENTS VERSUS STANDARD CARE IN TOTAL HIP ARTHROPLASTY

British Hip Society (BHS) meeting, held online, 9–11 June 2021.



Abstract

Total hip arthroplasty (THA) is one of the most successful surgical procedures of modern times, however debate continues as to the optimal orientation of the acetabular component and how to reliably achieve this. We hypothesised that functional CT-based planning with patient specific instruments using the Corin Optimised Positioning System (OPS) would provide more accurate component alignment than the conventional freehand technique using 2D templating.

A pragmatic single-centre, patient-assessor blinded, randomised control trial of patients undergoing THA was performed. 54 patients (age 18–70) were recruited to either OPS THA or conventional THA. All patients received a cementless acetabular component. Patients in both arms underwent pre- and post-operative CT scans, and four functional x-rays (standing and seated). Patients in the OPS group had a 3D surgical plan and bespoke guides made. Patients in the conventional group had a surgical plan based on 2D templating x-rays, and the pre-operative target acetabular orientation was recorded by the surgeon. The primary outcome measure was the difference between planned and achieved acetabular anteversion and was determined by post-operative CT scan performed at 6 weeks. Secondary outcome measures included Hip disability and Osteoarthritis Outcome Score (HOOS), Oxford Hip Score (OHS), EQ-5D and adverse events.

In the OPS group, the achieved acetabular anteversion was within 10° of the plan in 96% of cases, compared with only 76% of cases in the conventional group. For acetabular inclination, the achieved position in the OPS group was within 10° of the plan in 96% of cases, compared with in only 84% of cases in the conventional group. These differences were not statistically significant. The clinical outcomes were comparable between the two groups.

Large errors in acetabular orientation appear to be reduced when functional CT-based planning and patient-specific instruments are used compared to the freehand technique, but no statistically significant differences were seen in the difference between planned and achieved angles. Larger studies are needed to analyse this in more detail and to determine whether the reduced numbers of outliers lead to improved clinical outcomes.


Email: