We have reviewed 70 patients with bilateral simultaneous total hip arthroplasties to determine the rate of failure and to compare polyethylene wear and osteolysis between an implant with a cobalt-chrome head and Hylamer liner with that of a zirconia head and Hylamer liner. The mean thickness of the polyethylene liner was 11.0 mm (8.8 to 12.2) in the hip with a zirconia head and 10.7 mm (8.8 to 12.2) in that with a cobalt-chrome head. At follow-up at 6.4 years no acetabular or femoral component had been revised for aseptic loosening and no acetabular or femoral component was loose according to radiological criteria in both the cemented and cementless groups. The mean rate of linear wear and annual wear rate were highest in the 22 mm zirconia femoral head (1.25 mm (SD 1.05) and 0.21 mm (SD 0.18), respectively) and lowest in the 22 mm
CXCR4 gene and protein expression is regulated in a dose and time-dependent manner by metallic wear debris but not polyethylene wear debris in vitro and in vivo. Progressive osteolysis leading to aseptic loosening among metal-on-metal (MoM) total hip arthroplasties (THA's), and adverse reactions to metallic debris (ARMD) are increasing causes for concern among existing patients who have been implanted with MoM hip replacements. Close surveillance of these patients is necessary and difficulties lie in early detection as well as differentiating low-grade infection from ARMD in the early stages. Several inflammatory markers have been investigated in this context, but to date, none is specific with regards to the offending material. In earlier studies, it has been shown that osteoblastic phenotypes and differentiation are regulated by different types of wear particles.Summary Statement
Introduction
Dual mobility (DM) total hip replacements (THRs) were introduced to reduce dislocation risk, which is the most common cause of early revision. Although DM THRs have shown good overall survivorship and low dislocation rates, the mechanisms which describe how these bearings function in-vivo are not fully understood. Therefore, the study aim was to comprehensively assess retrieved DM polyethylene liners for signs of damage using visual inspection and semi-quantitative geometric assessment methods. Retrieved DM liners (n=18) were visually inspected for the presence of surface damage, whereby the internal and external surfaces were independently assigned a score of one (present) or zero (not present) for seven damage modes. The severity of damage was not assessed. The material composition of embedded debris was characterised using energy-dispersive x-ray analysis (EDX). Additionally, each liner was geometrically assessed for signs of wear/deformation [1]. Scratching and pitting were the most common damage modes on either surface. Additionally, burnishing was observed on 50% of the internal surfaces and embedded debris was identified on 67% of the external surfaces. EDX analysis of the debris identified several materials including titanium,
The purpose of the study was to compare the mechanical properties, oxidation and wear resistance of a vitamin E blended and moderately crosslinked polyethylene for total knee arthroplasty (MXE) in comparison with clinically established polyethylene materials. The following polyethylene materials were tested: CPE (30 kGy e-beam sterilized), XLPE (75 kGy gamma crosslinked @ 100°C), ViXLPE (0.1 % vitamin E blended, 80 kGy e-beam crosslinked @ 100°C), and MXE (0.1 % vitamin E blended polyethylene, 30 kGy gamma sterilized). For the different tests, the polyethylene materials were either unaged or artificially aged for two or six weeks according to ASTM F2003-02. The oxidation index was measured based on ASTM F2102 at a 1 mm depth. Small punch testing was performed based on ASTM F2977. Mechanical properties were measured on unaged materials according to ASTM D638. Wear simulation was performed on a load controlled 3 + 1 station knee wear simulator (EndoLab GmbH, Thansau, Germany) capable of reproducing loads and movement of highly demanding activities (HDA) as well as ISO 14243-1 load profiles. The load profiles were applied for 5 million cycles (mc) or delamination of the polyethylene components. Medium size AS e.motion. ®. PS Pro (Aesculap AG, Tuttlingen, Germany) femoral and tibial components with a ZrN-multilayer surface, as well as Columbus. ®. CR
Reports of improved functional outcome of Metal on Metal Hip Resurfacing Arthroplasty (mHRA) to Total Hip Replacement needs to be balanced with concerns of metal ion release. By removing
Abstract. Objectives. Total hip replacement (THR) is one of the most successful and cost-effective interventions in orthopaedic surgery. Dislocation is a debilitating complication of THR and managing an unstable THR constitutes a significant clinical challenge. Stability in THR is multifactorial and is influenced by surgical, patient and implant related factors. It is established that larger diameter femoral heads have a wider impingement-free range of movement and an increase in jump distance, both of which are relevant in reducing the risk of dislocation. However, they can generate higher frictional torque which has led to concerns related to increased wear and loosening. Furthermore, the potential for taper corrosion or trunnionosis is also a potential concern with larger femoral heads, particularly those made from
Abstract. Objectives. Current use of hard biomaterials such as
We have undertaken a series of clinical trials over the last 20 years to look at different bearing surface combinations in young adults. We continue to follow these patients well beyond the planned duration of the trials and new information is constantly becoming available. The first trial compared ceramic-on-ceramic with ceramic-on-standard-polyethylene. These patients have now been followed for 20 years with significant wear in the polyethylene group but virtually identical revision rates. The second trial ceramic-on-ceramic, cobalt-chrome-on-standard-polyethylene and cobalt-chrome-on-cross-linked-polyethylene. In this group the ceramic-on-ceramic patients have the lowest revision rate; the ceramic-on-polyethylene group demonstrates a lower wear rate than cobalt-chrome-on-polyethylene. The third trial looks at
The long-term biological effects of wear debris are unknown. We have investigated whether there is any evidence of cumulative mutagenic damage in peripheral blood lymphocytes of patients undergoing revision arthroplasty of predominantly metal-on-plastic total hip replacements compared with those at primary arthroplasty. There was a threefold increase in aneuploidy and a twofold increase in chromosomal translocations which could not be explained by the confounding variables of smoking, gender, age and diagnostic radiographs. In the patients with TiVaAl prostheses there was a fivefold increase in aneuploidy but no increase in chromosomal translocations. By contrast, in patients with
We examined stainless-steel,
ABSTRACT. The friction and lubrication behaviour of four Biomet ReCap components with a nominal diameter of 52 mm and diametral clearance ranging from 167-178 μm were investigated using a friction hip simulator. Friction testing was carried out using pure bovine serum and aqueous solutions of bovine serum (BS), with and without carboxymethyl cellulose (CMC), adjusted to a range of viscosities (0.001-0.236 Pas). The Stribeck analyses suggested mixed lubrication as the dominant mode with the lowest friction factor of 0.07 at a viscosity of 0.04 Pas. INTRODUCTION. The femoral resurfacing systems provide an alternative to hemi and total hip arthroplasty and offer several unique advantages including large resurfacing heads (>35–60 mm diameter) allowing increased range of motion (and stability) over the traditional 28 mm artificial hip joints, with excellent tolerances and surface finish leading to a reduction in wear, as well as preserving primary bone with the femoral canal remaining untouched. This work has investigated the friction and lubrication behaviour of four 52 mm metal-on-metal Biomet ReCap components with a clearance of 167-178 μm using serum-based lubricants. MATERIALS AND METHODS. Four as-cast, high carbon,
ABSTRACT. The viscosity and shear stress versus shear rate relationship for pure bovine serum (BS) and its aqueous solutions with and without carboxymethyl cellulose (CMC) were investigated. BS and diluted BS without CMC showed pseudoplastic flow curves up to a critical shear rate of ∼100 s-1 above which a Newtonian flow with significant rise in shear stress was observed. The viscosity flow curve for the diluted BS+5g CMC showed only shear thinning up to a shear rate of 3000 s-1 whereas diluted BS+1g or +2g CMC showed similar flow curves to pure BS. The shear rate application modified the flow behaviour of BS from a pseudoplastic to a Newtonian flow depending on its purity and CMC content. Friction factor was dependent on viscosity and clearance with mixed lubrication as the dominant mode within the viscosity range 0.001-0.044 Pas. INTRODUCTION. Pure BS and diluted BS are used as in vitro lubricants for tribological studies. Boundary, mixed and fluid film lubrication are the mechanisms involved in the lubrication of both natural and artificial joints. Clearance and lubricant viscosity will influence the nature of contact between the articulating surfaces. The objectives of this work were to study the flow properties of serum-based lubricants with different viscosities and the correlation between the rheological properties and frictional (and lubrication) behaviour of large diameter Biomet ReCaps with various clearances. MATERIALS AND METHODS. Rheological analysis was performed using a cone-on-plate rheometer (RHEOPLUS/32 V3.40) with a gap of 0.049mm at a constant temperature of 25°C on pure BS as base, 25BS+75 distilled water (DW), and 25BS+75DW+1g, +2g, or +5g CMC. The viscosity and shear stress were measured within shear rates of 0.3-3000 s-1. Frictional measurements of all the joints were carried out at the University of Bradford using a Prosim Friction Simulator on four, as-cast, high carbon,
Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues. A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.Objectives
Methods
We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants. This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.
Wear of polyethylene is associated with aseptic loosening of orthopaedic implants and has been observed in hip and knee prostheses and anatomical implants for the shoulder. The reversed shoulder prostheses have not been assessed as yet. We investigated the volumetric polyethylene wear of the reversed and anatomical Aequalis shoulder prostheses using a mathematical musculoskeletal model. Movement and joint stability were achieved by EMG-controlled activation of the muscles. A non-constant wear factor was considered. Simulated activities of daily living were estimated from After one year of use, the volumetric wear was 8.4 mm3 for the anatomical prosthesis, but 44.6 mm3 for the reversed version. For the anatomical prosthesis the predictions for contact pressure and wear were consistent with biomechanical and clinical data. The abrasive wear of the polyethylene in reversed prostheses should not be underestimated, and further analysis, both experimental and clinical, is required.
An experimental sheep model was used for impaction allografting of 12 hemiarthroplasty femoral components placed into two equal-sized groups. In group 1, a 50:50 mixture of ApaPore hydroxyapatite bone-graft substitute and allograft was used. In group 2, ApaPore and allograft were mixed in a 90:10 ratio. Both groups were killed at six months. Ground reaction force results demonstrated no significant differences (p >
0.05) between the two groups at 8, 16 and 24 weeks post-operatively, and all animals remained active. The mean bone turnover rates were significantly greater in group 1, at 0.00206 mm/day, compared to group 2 at 0.0013 mm/day (p <
0.05). The results for the area of new bone formation demonstrated no significant differences (p >
0.05) between the two groups. No significant differences were found between the two groups in thickness of the cement mantle (p >
0.05) and percentage ApaPore-bone contact (p >
0.05). The results of this animal study demonstrated that a mixture of ApaPore allograft in a 90:10 ratio was comparable to using a 50:50 mixture.
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented