Advertisement for orthosearch.org.uk
Results 1 - 20 of 161
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 450 - 450
1 Dec 2013
Zeng H Karwowski J Lawrynowicz D Nevelos J
Full Access

Introduction:. The purpose of this study was to evaluate the basic corrosion properties of conventional CoCr (ASTM F 1537, UNS R31537), Gas Atomized Dispersion Strengthened (GADS) CoCr (ASTM 1537, UNS R31539), Ti6Al4V ELI (ASTM F 136, UNS R56401) and Ti-12Mo-6Zr-2Fe (TMZF) (ASTM F 1813) alloys that are used in joint replacement applications. Methods:. Test coupon samples made from alloys' wrought bar stocks were polished then tested. Each material had 4 groups (n. 3. 3) that were tested in 4 aqueous solutions: 1) pH 7.4, 2) pH 5, 3) pH 2, and 4) pH 2 with 1 M of H. 2. O. 2. Potential dynamic polarization was conducted using a Princeton Applied Research VMC Potentiostat, with an Ag/AgCl electrode as the reference electrode, and it resembled method in ASTM F 2129-08. The samples were evaluated with Scanning Electron Microscope (SEM). Results:. 1) Both pH and solution chemistry changes affected the corrosion behaviors and caused the polarization curves to shift; Figure 1 & 2 show the polarization curves of CoCr and Ti alloys. 2) The polarization curves for conventional CoCr and GADS alloys were identical with minor differences; Figure 1 shows the polarization curves for CoCr and GADS alloys at pH 7.4 and pH 2 conditions. 3) The polarization curves for Ti6Al4V ELI and TMZF alloys were also similar at most conditions with some minor differences; Figure 2 shows the polarization curves of Ti6Al4V ELI and TMZF alloys at pH 2. It should also be noted that at pH 2 plus H. 2. O. 2. condition, Ti6Al4V ELI showed increased current at voltages near 1V as shown in Figure 2. This might suggest an onset of accelerated corrosion or breakdown that is not normally observed with Ti alloys. TMZF showed constant current in the same region, suggesting no breakdown or accelerated corrosion at this condition. SEM revealed corrosion of TI6Al4V ELI alloy similar to etching as shown in Figure 2. Discussion:. The fact that GADS and conventional CoCr alloys behaved identically suggests similar corrosion properties and resistance in normal conditions; the same applies to TI6Al4V ELI and TMZF alloys. It was unexpected to see the seemingly “breakdown” or accelerated corrosion for Ti6Al4V ELI in pH 2 solution with added H. 2. O. 2. at voltages near 1 V. Note that the test condition was very aggressive and the proper amount or concentration of H. 2. O. 2. in a test solution has not been well documented in the literature. However, it should be kept in mind that alloys' corrosion properties under extreme test conditions should not be overlooked as similar conditions could happen clinically in crevice environments for example. Regardless, it is interesting to observe the difference between Ti6Al4V ELI and TMZF at this particular condition, and it warrants further investigation


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 420 - 420
1 Dec 2013
McPherson E Burgett M Halim T Donaldson T Clarke I
Full Access

Controversy has existed for decades over the role of fretting-corrosion in modular CoCr heads used with stems of CoCr vs Ti6Al4V. Since retrieval data on taper performance remains scant, we report here an18-year survivorship of a Ti6Al4V: CoCr combination (APR design; Intermedics Inc). Unique to this study were the threaded profiles present on both stem and head tapers (Fig. 1). This female patient was revised for pain, osteolysis and recurrent hip dislocation at 17 years, 10 months. A prior MPE hip replacement performed for her severely dysplastic right hip had lasted 11 years. At this 2nd revision, the 28 mm CoCr head was found dislocated posteriorly and superiorly. Metallosis was evident in the tissues. The polyethylene liner showed extensive rim damage on both anterior and posterior aspects. The neck of her APR Revision stem (Intermedics Inc) had worn through the polyethylene rim and impinged on the metal cage. The cage was found loose, the liner had disassociated, and the peri-trochanteric areas were compromised by massive osteolysis. The femoral stem and head were removed together without disassembly. The femoral stem and acetabular construct were replaced by an ARCOS revision system using 36 mm head with a Freedom cup (cemented to Max-Ti cage; Biomet Inc.). The complete femoral neck and head were bi-valved assembled in horizontal plane for direct imaging by interferometry and SEM (Fig. 1a). After sectioning the head separated from the stem. Quantitative imaging used 1 to 5 regions with 6-replicate measurements per region and differentiation into contact and non-contact zones (Fig. 1b). Visual corrosion mapping (3) was recorded digitally in 4 anatomical views (Figs 1b–f). The thread profile on contact zone inside the head (Fig. 2a) had a pitch of approximately 40 μm and a peak-to-valley depth of 4 μm overall (Fig. 2b profile section of thread: PV = 2 μm). The thread profile on stem trunnion (Fig. 3a) had a pitch of approximately 125 μm and a peak-to-valley depth of 3.5 μm overall (Fig. 2b profile section of thread: PV = 1 μm). Thus the stem trunnion thread was much coarser than the head. Overall corrosion grading was judged very mild. Overall we were satisfied that this Ti6Al4V: CoCr combination taper junction with threaded interfaces had performed very well for 18 years. Nevertheless, our visual grading was subject to opinion and thus unrewarding. The continuing project will quantify the contacting and non-contacting regions of head and stem (Fig. 1b)


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 111 - 111
1 Jan 2016
Walsh W Bertollo N Hamze A Christou C Gao B Angibaud L
Full Access

Introduction. Biological fixation through bone ingrowth and ongrowth to implants can be achieved with a variety of surface treatments and technologies. This study evaluated the effect of two different three dimensional surface coatings for CoCr where porosity was controlled through the use of different geometry of CoCr beads in the sintering process. Methods. Test specimens in Group A were coated with conventional spherical porous-bead technology. The porous coating technology used on Group B was a variation of the conventional porous-bead technology. Instead of spherical beads, cobalt-chromium particles in irregular shapes were sieved for a particular size range, and were sintered onto the specimen substrate using similar process as Group A. The geometry and the size variation of the particles resulted in a unique 3D porous structure with widely interconnected pores. Three implants were placed bicortically in the tibia. Two implants were placed in the cancellous bone of the medial distal femur and proximal tibia bilaterally with 4 implantation conditions (2 mm gap, 1 mm gap line-to-line, and press fit). Animals were euthanized at 4 or 12 weeks for standard mechanical, histological and histomorphometric endpoints. Results. Shear strength increased with time for both groups (P<0.001). While no difference was detected between groups at the 4 week time point, the difference was statistically significant at 12 weeks with the irregular shaped beads using in the coating in group B providing a shear strength that outperformed the standard spherical beads. Histomorphometry revealed new bone ingrowth into the porous domains of both implant groups improved with time (P<0.001). Significantly greater (P<0.05) new bone integration was observed with the irregular shaped beads in the cortical as well as cancellous sites at 4 and 12 weeks (Figure 1). Discussion. Significant improvements can be made in the fixation strength of three dimensional CoCr coatings. This holds true in cortical implantation as well as different cancellous implantation scenarios. Material chemical composition of both coating and substrate conforms to ASTM F75 standard. The conventional sintered porous-bead technology in Group A provided a multi-layer porous structure at the bone implant interface has been well-established for the clinical use on TKA implants for over 15 years. This type of coating usually produces an average porosity of 30% to 40%, and an average pore size of 150 µm to 250 µm. The porous coating technology used on Group B was a variation of the conventional porous-bead technology. Instead of spherical beads, cobalt-chromium particles in irregular shapes were sieved for a particular size range, and were sintered onto the specimen substrate using similar process as Group A. Due to the geometry and the size variation of the particles, a true 3D porous structure with widely interconnected pores can be formed. Microstructure analysis on femoral implants showed that this coating technology is able to provide an average porosity of 50% to 70%, and an average pore size of 200 µm to 450 µm. This technology also produces a rougher coating surface appearance which could also play a potential role in the overall performance


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 23 - 23
1 May 2016
Arnholt C MacDonald D Kocagoz S Chen A Cates H Klein G Rimnac C Kurtz S
Full Access

Introduction. Previous studies of long-term CoCr alloy femoral components for TKA have identified 3rd body abrasive wear and inflammatory cell induced corrosion (ICIC). The extent of femoral condyle surface damage in contemporary CoCr femoral components is currently unclear. The purpose of this study was to investigate the prevalence and morphology of damage (3rd body scratches and ICIC) at the bearing surface in retrieved TKA femoral components from contemporary designs. Methods. 308 CoCr femoral TKA components were collected as part of an ongoing, multi-institutional orthopedic implant retrieval program. The collection included contemporary designs from Stryker (Triathlon n=48, NRG n=10, Scorpio n=31), Depuy Synthes (PFC n=27) and Zimmer (NexGen n=140, Persona n=1) and Biomet (Vanguard n=51). Hinged knee designs and unicondylar knee designs were excluded. Components were split into groups based on implantation time: short-term (1–3y, n=134), intermediate-term (3–5y, n=73) and long-term (6–15y, n=101). Each grouping was mainly revised for instability, infection and loosening. Third-body abrasive wear of CoCr was evaluated using a semi-quantitative scoring method similar to the Hood method (Figure 1). A score of 1 had minimal damage and a score of 4 corresponded to damage covering more than 50% of the evaluated area. ICIC damage was reported as location of affected area. A white light interferometer (Zygo New View 5000) was also used to analyze the topography of severe damage of the bearing surface. For this analysis, three representative components from each cohort were selected and analyzed in three locations on the apex of the bearing surface. We analyzed the following roughness parameters: Ra, Rsk, and Rku. Results. On the CoCr bearing surface, the primary damage mechanisms were large scratches, small random scratches, and ICIC damage (Figure 2). Mild to severe damage (Damage Score ≥ 2) was observed in 96% of the short-term, 98% intermediate-term and 94% of long-term components. Severe damage (Damage Score = 4) was observed in 43% of the short-term, 50% intermediate-term and 56% of long-term components. ICIC damage observed on a portion of the bearing surface was detected in 43% of the short-term components, 30% of the intermediate-term components and 26% of the long term components. Apparent ICIC damage on the bearing and/or a non-bearing region of the component was observed in 85% of the short-term components, 75% of the intermediate-term components and 80of long-term TKA components. The Ra, Rsk, and Rku were similar between cohorts (Table 1). Discussion. Abrasive wear of the femoral components was frequently observed in retrieved contemporary femoral components for TKA, regardless of their implantation time, and can most likely be attributed to third body damage caused by bone or bone cement debris. The prevalence of severe CoCr damage scores was highest in the long-term cohort, while the appearance of ICIC damage was lowest in the long-term cohort. Surface roughness parameters were similar in all three cohorts suggesting that the mechanism for this damage is comparable throughout the first 15 years of service. Future work is necessary to quantify the in vivo release of CoCr from abrasive wear and corrosion mechanisms, and the effects of increased surface roughness on wear of the polyethylene counter face


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 393 - 393
1 Dec 2013
Murphy S Le D
Full Access

Introduction. Adverse Local Tissue Reactions (ALTR) have been reported in association with both wear and corrosion. Tissue reactions have been reported in association with corrosion at CoCr head-CoCr neck, CoCr head-TiAl6V4 neck, and CoCr modular neck on beta-titanium (TMZF) stem junctions. The current abstract reports on 3 cases of ALTR in association with CoCr modular necks on convention titanium (TiAl6V4) stem junctions. Case 1. A 67 year old male (87 kg, 1.73 m, BMI 29.1) presented with new onset hip irritation 11 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 95, CRP = 5, Cr level = 1.0, Co level = 4.1, leukocyte transformation testing = highly reactive to nickel. Hip aspiration was culture negative with 11,250 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the greater trochanter. Case 2. A 52 year old male (89 kg, 1.83 m, BMI 26.5) presented with new onset hip irritation 30 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 7, CRP = 5.4, Cr level = 2.1, Co level = 4.8, leukocyte transformation testing = reactive to nickel. Hip aspiration was culture negative with 3995 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the iliopsoas. Case 3. A 52 year old male (104 kg, 1.85 m, BMI 30.1) presented with new onset hip irritation 26 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 33, CRP = 34.9, Cr level = 1.0, Co level = 3.7, leukocyte transformation testing = no reactivity to any of the biomaterials. Hip aspiration was culture negative with 3,780 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the iliopsoas. Discussion. All three of these patients are scheduled for revision surgery. All three had ceramic-ceramic bearings. We have experience with 1029 ceramic-ceramic THA with fixed neck conventional titanium and modular titanium neck implants with minimum 2 yr f/u and have never diagnosed an adverse reaction in any of these patients. It is possible that corrosion at the CoCr neck on TiAl6V4 stem junction is the root cause of these reactions. Although the incidence of diagnosed reactions is roughly 1%, it appears that the use of CoCr at any junction under significant mechanical stress can result in adverse local tissue reaction and therefore should either be avoided or used with great caution and compelling indications


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 451 - 451
1 Dec 2013
Nguyen D Burgett M Clarke I Halim T Donaldson T
Full Access

Damage to metal-on-metal bearings (MOM) has been varyingly described as “edge wear,” third-body abrasive wear and “rim-damage” (1–4). However, no distinction has been made between any of these proposed wear mechanisms. The goal of this study was to discover what features might differentiate between surface damage created by either 2-body or 3-body wear mechanisms in MOM bearings. The hypotheses were that surface damage created by impingement of the cup rim (2-body wear) would be i) linear on the micro-scale, ii) reveal transverse striations (in direction of the sliding rim), iii) have either no raised lip or have a single lip along one side of the track, and iv) have an asymmetrical surface profile across the track width. Five cases with 28 mm MOM, five of 34–38 mm MOM, and five of 50–56 mm diameter were studied (N = 15). The main wear zone (MWZ) was measured in each MOM head and the number of 2-body wear tracks recorded in the non-wear (NWZ) and main wear zone (MWZ). Bearing damage was examined using a white-light interferometer (Zygo Newview 600; 5x lens) and a scanning electron microscope (Zeiss MA15). The depths and slopes were assessed across the width of the damage tracks. Thirteen of the 15 MOM bearings showed wear tracks that exhibited all four of the hypothesized 2-body wear characteristics. These wear tracks will be referred to as “micro-segments”. While micro-segments visually appeared linear, microscopically they revealed a semi-lunar edge coupled with transverse striations leading to a linear edge. This indicated that during impingement episodes, the cup rim ploughed material from the CoCr surface at the semi-lunar edge (Fig. 1), thereby creating the abruptly raised lip on the linear edge of the track. This “snow plough effect” and its distinct edge effect can account for the asymmetrical surface profile. A different type of 2-body wear was identified and referred to as “furrows”. Furrows also visually appeared linear visually, but microscopically revealed longitudinal striations and a symmetrical surface profile (Fig. 2). Furrows had lips raised on both sides of the track, but not circumscribing the terminal ends of the track. Instead, the ends of the furrows are tapered smooth transitions to the articular surface. Thus, 2-body tracks were found to be distinguishable from 3-body tracks (micro-grooves) and were classified as either micro-segments or furrows. Micro-segements supported hypotheses 1–3 and provided a clearer definition for hypothesis-4, while furrows only supported hypothesis 1. The divergence in features between micro-segments and furrows allude to different interactions between the bearing and cup rim that led to each type of track. While these data represent a small set of cases (n = 15) this evidence shows for the first time what was previously only suspected (2), that the CoCr rim can routinely create 2-body wear damage mechanisms in MOM femoral heads


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 153 - 153
1 Jun 2012
Parikh A Hill P Pawar V Morrison M
Full Access

Introduction. Large diameter femoral heads offer increased range of motion and reduced risk of dislocation. However, their use in total hip arthroplasty has historically been limited by their correlation with increased polyethylene wear. The improved wear resistance of highly crosslinked UHWMPE has led a number of clinicians to transition from implanting traditionally popular sizes (28mm and 32 mm) to implanting 36 mm heads. Desire to further increase stability and range of motion has spurred interest in even larger sizes (> 36 mm). While the long-term clinical ramifications are unknown, in-vivo measurements of highly crosslinked UHMWPE liners indicate increases in head diameter are associated with increased volumetric wear [1]. The goal of this study was to determine if this increase in wear could be negated by using femoral heads with a ceramic surface, such as oxidized Zr-2.5Nb (OxZr), rather than CoCrMo (CoCr). Specifically, wear of 10 Mrad crosslinked UHMWPE (XLPE) against 36 mm CoCr and 44 mm OxZr heads was compared. Materials and Methods. Ram-extruded GUR 1050 UHMWPE was crosslinked by gamma irradiation to 10 Mrad, remelted, and machined into acetabular liners. Liners were sterilized using vaporized hydrogen peroxide and tested against either 36 mm CoCr or 44 mm OxZr (OXINIUM(tm)) heads (n=3). All implants were manufactured by Smith & Nephew (Memphis, TN). Testing was conducted on a hip simulator (AMTI, Watertown, MA) as previously described [2]. The 4000N peak load (4 time body weight for a 102 kg/225 lb patient) and 1.15 Hz frequency used are based upon data obtained from an instrumented implant during fast walking/jogging and have previously been shown to generate measurable XLPE wear [2,3]. Lubricant was a serum (Alpha Calf Fraction, HyClone Laboratories, Logan, UT) solution that was replaced once per week [2]. Liners were weighed at least once every million cycles (Mcycle) over the duration of testing (∼ 5 Mcycle). Loaded soak controls were used to correct for fluid absorption. Single factor ANOVA was used to compare groups (a = 0.05). Results. The predominant wear feature displayed on the articular surface of liners was burnishing. There were no signs of fatigue wear or of delamination. Mean wear rates (± std dev) of liners articulated against 36mm CoCr and 44 mm OxZr heads were 3.7 ± 0.4 mm. 3. /Mcycle and 2.7 ± 0.4 mm. 3. /Mcycle, respectively (Figure 1). This difference was statistically significant (p = 0.04). Discussion. Although large diameter heads offer biomechanical advantages, their use in total hip arthroplasty has historically been limited due to correlation with increased polyethylene wear. While highly crosslinked liners exhibit significantly improved wear resistance over conventional UHWMPE, their wear has also been shown to increase with head size [1]. Results presented here indicate that this increase in wear can be negated by using OxZr, rather than CoCr. Specifically, wear of XLPE liners was lower against 44 mm OxZr heads than against 36 mm CoCr heads


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 387 - 387
1 Dec 2013
Kurtz S MacDonald D Higgs G Gilbert J Klein GR Mont M Parvizi J Kraay M Rimnac C
Full Access

Introduction:. Degradation of modular head-neck tapers was raised as a concern in the 1990s (Gilbert 1993). The incidence of fretting and corrosion among modern, metal-on-polyethylene and ceramic-on-polyethylene THA systems with 36+ mm femoral heads remains poorly understood. Additionally, it is unknown whether metal debris from modular tapers could increase wear rates of highly crosslinked PE (HXLPE) liners. The purpose of this study was to characterize the severity of fretting and corrosion at head-neck modular interfaces in retrieved conventional and HXLPE THA systems and its effect on penetration rates. Patients & Methods:. 386 CoCr alloy heads from 5 manufacturers were analyzed along with 166 stems (38 with ceramic femoral heads). Metal and ceramic components were cleaned and examined at the head taper and stem taper by two investigators. Scores ranging from 1 (mild) to 4 (severe) were assigned in accordance with the semi-quantitative method adapted from a previously published technique. Linear penetration of liners was measured using a calibrated digital micrometer (accuracy: 0.001 mm). Devices implanted less than 1 year were excluded from this analysis because in the short-term, creep dominates penetration of the head into the liner. Results:. The majority of the components were revised for instability, infection, and loosening. Mild to severe taper damage (score ≥2) was found in 77% of head tapers and 52% of stem tapers. The extent of damage was correlated to implantation time at the head taper (p = 0.0004) and at the stem taper (p = 0.0004). Damage scores were statistically elevated on CoCr heads than the matched stems (mean score difference = 0.5; p < 0.0001; Figure 2) and the two metrics were positively correlated with each other (ρ = 0.41). No difference was observed between stem taper damage and head material (CoCr, ceramic) (p = 0.56), nor was a correlation found between taper damage and head size (p = 0.85; Figure 3). The penetration rate across different formulations of HXLPE was not found to be significantly different (p = 0.07), and therefore grouped together for further analysis. Within this cohort, penetration rate was not found to be associated with head size (p = 0.08) though a negative correlation with implantation time was noted (ρ = −0.35). When analyzed with taper damage scores, a correlation was not observed between head damage scores and HXLPE penetration rates (p = 0.51). Discussion:. The results of this study do not support the hypothesis that 36+ mm ceramic or CoCr femoral heads articulating on HXLPE liners are associated with increased risk of corrosion among HXLPE liners when compared with smaller diameter heads. A limitation of this study is the semi-quantitative scoring technique, heterogeneity of the retrieval collection and short implantation time of the larger diameter heads. Because corrosion may increase over time in vivo, longer-term follow-up, coupled with quantitative taper wear measurement, will better assess the natural progression of taper degradation in modern THA bearings


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 65 - 65
1 May 2016
Campbell P Kung M Ebramzadeh E Van Der Straeten C DeSmet K
Full Access

Bone ingrowth fixation of large diameter, beaded cobalt chromium cups is generally considered to be reliable but this is typically judged radiographically. To date, implant retrieval data of attached bone has been limited. This study evaluated correlations between the pre-revision radiographic appearance and the measured amount of bone attachment on one design of porous coated cup. Methods. Twenty-six monoblock, CoCr Birmingham Hip Resurfacing (BHR, Smith and Nephew, TN, USA) cups with macroscopic beads and hydroxyapatite coating were studied. Seventeen were revised for acetabular malposition with the remainder revised for femoral loosening (4), pain (1), infection (1), dislocation (1) or lysis (2). Median time to revision was 35 months (10 – 70 months). Ten patients were female; the median age of all patients was 54 years. The pre-revision radiographs were visually ranked for cup-bone integration as follows: 0 = none, 1 = < 50%, 2 = 50 – 75%, 3 = 76 – 95%, 4 = > 96% integration. Rankings were made for the superior and inferior aspects, without knowledge of the appearance of bone on the retrievals. The revised cups were photographed at an angle so the dome and the cup periphery were visualized. The area of bone in four equal segments in each of the superior and inferior aspects was measured with image analysis software. A probe was used to differentiate bone from soft tissue. Only bone that covered the beads was counted. Correlation coefficients were calculated for the radiographic and image analysis data. Results. Radiographically, most cups were assessed as having more than 50% of bone attachment and 7 cups were ranked as having almost total integration with bone. Only 2 cups were assessed radiographically as fully loose. Measured total bone attachment ranged from none to 55%. Superior and inferior percent ingrowth were highly correlated (corr=0.68, p<0.001) but there was no correlation between percent bone and x-ray rank (inferior corr=0.01, p=0.96; superior corr=0.23, p=0.26). There was no correlation between cup malpositioning as a reason for revision and x-ray integration ranking (superior p=0.34; inferior p=0.80). Discussion. Despite the radiographic appearance of good fixation, there was little or no correlation between percent area of actual bone attachment and x-ray appearance. One study limitation is the assumption that attached bone was indeed integrated with the beads as destructive sectioning was not done to verify this. Published autopsy retrieval studies have shown that even a small amount of actual ingrowth can provide clinically successful fixation. Another possible limitation was the variable quality of the radiographs. Never-the-less these results raise questions about the accuracy of radiographic analysis of bone fixation. The possibility that inadequate fixation is a cause for pain leading to revision should be considered even when the radiographic appearance indicates otherwise


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_7 | Pages 78 - 78
1 Jul 2020
Somerville L Clout A MacDonald S Naudie D McCalden RW Lanting B
Full Access

While Oxidized Zirconium (OxZr) femoral heads matched with highly cross-linked polyethylene (XLPE) have demonstrated the lowest rate of revision compared to other bearing couples in the Australian National Joint Registry, it has been postulated that these results may, in part, be due to the fact that a single company offers this bearing option with a limited combination of femoral and acetabular prostheses. The purpose of this study was to assess clinical and radiographic outcomes in a matched cohort of total hip replacements (THR) utilizing an identical cementless femoral stem and acetabular component with either an Oxidized Zirconium (OxZr) or Cobalt-Chrome (CoCr) femoral heads at a minimum of 10 years follow-up. We reviewed our institutional database to identify all patients whom underwent a THR with a single cementless femoral stem, acetabular component, XLPE liner and OxZr femoral head with a minimum of 10 years of follow-up. These were then matched to patients who underwent a THR with identical prosthesis combinations with CoCr femoral head by gender, age and BMI. All patients were prospectively evaluated with WOMAC, SF-12 and Harris Hip Score (HHS) preoperatively and postoperatively at 6 weeks, 3 months, 1 and 2 years and every 2 years thereafter. Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years for both cohorts. Paired analysis was performed to determine if differences exist in patient reported outcomes. There were 208 OxZr THRs identified which were matched with 208 CoCr THRs. There was no difference in average age (OxZr, 54.58 years, CoCr, 54.75 years), gender (OxZr 47.6% female, CoCr 47.6% female), and average body max index (OxZr, 31.36 kg/m2, CoCr, 31.12 kg/m2) between the two cohorts. There were no significant differences preoperatively in any of the outcome scores between the two groups (WOMAC (p=0.449), SF-12 (p=0.379), HHS(p=0.3718)). Both the SF12 (p=0.446) and the WOMAC (p=0.278) were similar between the two groups, however the OxZr THR cohort had slightly better HHS compared to the CoCr THR cohort (92.6 vs. 89.7, p=0.039). With revision for any reason as the end point, there was no significant difference in 10 years survivorship between groups (OxZr 98.5%, CoCr 96.6%, p=0.08). Similarly, aseptic revisions demonstrated comparable survivorship rates at 10 year between the OxZr (99.5%) and CoCr groups (97.6%)(p=0.15). Both THR cohorts demonstrated outstanding survivorship and improvement in patient reported outcomes. The only difference was a slightly better HHS score for the OxZr cohort which may represent selection bias, where OxZr implants were perhaps implanted in more active patients. Implant survivorship was excellent and not dissimilar for both the OxZr and CoCr groups at 10 years. Therefore, with respect to implant longevity at the end of the first decade, there appears to be no clear advantage of OxZr heads compared to CoCr heads when paired with XLPE for patients with similar demographics. Further follow-up into the second and third decade may be required to demonstrate if a difference does exist


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 25 - 25
1 Feb 2020
De Villiers D Collins S Taylor A Dickinson A
Full Access

INTRODUCTION. Hip resurfacing offers a more bone conserving solution than total hip replacement (THR) but currently has limited clinical indications related to some poor design concepts and metal ion related issues. Other materials are currently being investigated based on their successful clinical history in THR such as Zirconia Toughened Alumina (ZTA, Biolox Delta, CeramTec, Germany) which has shown low wear rates and good biocompatibility but has previously only been used as a bearing surface in THR. A newly developed direct cementless fixation all-ceramic (ZTA) resurfacing cup offers a new solution for resurfacing however ZTA has a Young's modulus approximately 1.6 times greater than CoCr - such may affect the acetabular bone remodelling. This modelling study investigates whether increased stress shielding may occur when compared to a CoCr resurfacing implant with successful known clinical survivorship. METHODS. A finite element model of a hemipelvis constructed from CT scans was used and virtually reamed to a diameter of 58mm. Simulations were conducted and comparisons made of the ‘intact’ acetabulum and ‘as implanted’ with monobloc cups made from CoCr (Adept®, MatOrtho Ltd, UK) and ZTA (ReCerf ™, MatOrtho Ltd. UK) orientated at 35° inclination and 20° anteversion. The cups were loaded with 3.97kN representing a walking load of 280% for an upper bound height patient with a BMI of 35. The cup-bone interface was assigned a coulomb slip-stick function with a coefficient of friction of 0.5. The percentage change in strain energy density between the intact and implanted states was used to indicate hypertrophy (increase in density) or stress shielding (decrease in density). RESULTS. Implanting both cups changed the strain distribution observed in the hemipelvis, Figure 1. The change in strain distribution was similar between materials and indicated a similar response from the bone, Figure 2. In both implanted cases, the inferior peri-acetabular bone around the implant indicated a reduction in bone strain. The bone remodelling distribution charts show that regardless of threshold remodelling stimulus level (75% in elderly, 50% in younger patients) the CoCr and ZTA cups were expected to produce the same bone response with only a small percentage of the bone in the hemipelvis indicating stress shielding or hypertrophy, Figure 3. DISCUSSION. Currently only metal cups are used for cementless fixation but improvements in design and technology have made it possible to engineer a thin-walled, direct fixation, all-ceramic cup. Both CoCr and ZTA are an order of magnitude greater than the Young's modulus of cortical bone altering the bone strain but changing the material from CoCr to a stiffer ZTA did not change the expected bone remodelling response. Given the clinical history of metal cups without loosening due to bone remodelling, the study indicates that a ZTA cup should not lead to increased stress shielding and is potentially suitable for as a cementless cup for both resurfacing and THR. SIGNIFICANCE. An all-ceramic cup is unlikely to lead to increased stress shielding around the acetabulum due to the change in material. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 52 - 52
1 Feb 2020
Sadhwani S Picache D Janssen D de Ruiter L Rankin K Briscoe A Verdonschot N Shah A
Full Access

Introduction. Polyetheretherketone (PEEK) has been proposed as an implant material for femoral total knee arthroplasty (TKA) components. Potential clinical advantages of PEEK over standard cobalt chrome alloys include modulus of elasticity and subsequently reduced stress shielding potentially eliminating osteolysis, thermal conduction properties allowing for a more natural soft tissue environment, and reduced weight enabling quicker quadriceps recovery. Manufacturing advantages include reduced manufacturing and sterilization time, lower cost, and improved quality control. Currently, no PEEK TKA implants exist on the market. Therefore, evaluation of mechanical properties in a pre-clinical phase is required to minimize patient risk. The objectives of this study include evaluation of implant fixation and determination of the potential for reduced stress shielding using the PEEK femoral TKA component. Methods and Materials. Experimental and computational analysis was performed to evaluate the biomechanical response of the femoral component (Freedom Knee, Maxx Orthopedics Inc., Plymouth Meeting, PA; Figure 1). Fixation strength of CoCr and PEEK components was evaluated in pull-off tests of cemented femoral components on cellular polyurethane foam blocks (Sawbones, Vashon Island, WA). Subsequent testing investigated the cemented fixation using cadaveric distal femurs. The reconstructions were subjected to 500,000 cycles of the peak load occurring during a standardized gait cycle (ISO 14243-1). The change from CoCr to PEEK on implant fixation was studied through computational analysis of stress distributions in the cement, implant, and the cement-implant interface. Reconstructions were analyzed when subjected to standardized gait and demanding squat loads. To investigate potentially reduced stress shielding when using a PEEK component, paired cadaveric femurs were used to measure local bone strains using digital image correlation (DIC). First, standardized gait load was applied, then the left and right femurs were implanted with CoCr and PEEK components, respectively, and subjected to the same load. To verify the validity of the computational methodology, the intact and reconstructed femurs were replicated in FEA models, based on CT scans. Results. The cyclic load phase of the pull-off experiments revealed minimal migration for both CoCr and PEEK components, although after construct sectioning, debonding at the implant-cement interface was observed for the PEEK implants. During pull-off from Sawbones the ultimate failure load of the PEEK and CoCr components averaged 2552N and 3814N respectively. FEA simulations indicated that under more physiological loading, such as walking or squatting, the PEEK component had no increased risk of loss of fixation when compared to the CoCr component. Finally, the DIC experiments and FEA simulations confirmed closer resemblance of pre-operative strain distribution using the PEEK component. Discussion. The biomechanical consequences of changing implant material from CoCr to PEEK on implant fixation was studied using experimental and computational testing of cemented reconstructions. The results indicate that, although changes occur in implant fixation, the PEEK component had a fixation strength comparable to CoCr. The advantage of long term bone preservation, as the more compliant PEEK implant is able to better replicate the physiological loads occurring in the intact femur, may reduce stress shielding around the distal femur, a common clinical cause of TKA failure. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 90 - 90
1 Apr 2019
Cowie RM Pallem N Briscoe A Fisher J Jennings LM
Full Access

Introduction. PEEK-OPTIMA™ has been considered as an alternative to cobalt chrome in the femoral component of total knee replacements. Whole joint wear simulation studies of both the tibiofemoral and patellofemoral joints carried out to date have shown an equivalent wear rate of UHMWPE tibial and patella components against PEEK and cobalt chrome (CoCr) femoral components. In this study, the influence of third body wear on UHMWPE-on-PEEK was investigated, tests on UHMWPE-on-CoCr were carried out in parallel to compare PEEK to a conventional femoral component material. Methods. Wear simulation was carried out in simple geometry using a 6-station multi-directional pin-on-plate simulator. 5 scratches were created on each PEEK and CoCr plate perpendicular to the direction of the wear test using a diamond stylus to produce scratches with a geometry similar to that observed in retrieved CoCr femoral components. To investigate the influence of scratch lip height on wear, scratches of approximately 1, 2 and 4µm lip height were created. Wear simulation of GUR 1020 UHMWPE pins (conventional, non-sterile) against the plates was carried out for 1 million cycles (MC) using 17g/l bovine serum as a lubricant using kinematic conditions to replicate the average contact pressure and cross-shear in a total knee replacement. Wear of UHMWPE pins was measured gravimetrically and the surface topography of the plates assessed using a contacting Form Talysurf. Wear factors of the pins against the scratched plates were compared to unscratched controls (0µm lip height). Minimum n=3 for each condition and statistical analysis carried out using ANOVA with significance taken at p<0.05. Results. For the control tests (0µm lip height), the wear factor of UHMWPE pins was similar (p=0.64) against PEEK and CoCr plates. Against CoCr, with an increasing lip height, an exponential increase in wear factor of UHMWPE pins was observed; for PEEK, with increasing lip height, the wear factor did not show an exponential increase. When articulated against the largest scratches, 4µm, the wear factor of UHMWPE was significantly higher against CoCr than PEEK (p=0.01). At the conclusion of the study, on the PEEK plates, a polishing effect of the pin against the plates was observed and in the area of the wear test, the lip height of the scratches was lower than pre-test values; for the CoCr plates, no change in lip height was measured after 1MC wear simulation. Conclusion. The exponential relationship between scratch lip height in CoCr and wear of UHMWPE has previously been described. However, the trend in the wear of UHMWPE was different when articulating against scratched PEEK compared to CoCr, with a significantly higher wear factor of UHMWPE against CoCr than PEEK at a scratch lip height of 4µm. This study suggests that the third body wear behaviour of this all-polymer knee replacement will be different to that of conventional implant materials


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 149 - 149
1 Feb 2020
Kandemir G Smith S Joyce T
Full Access

Introduction. Total hip replacement with metal-on-polymer (MoP) hip prostheses is a successful treatment for late-stage osteoarthritis. However, the wear debris generated from the polymer acetabular liners remains a problem as it can be associated with osteolysis and aseptic loosening of the implant. This has led to the investigation of more wear resistant polymers in orthopaedics. Cross-linked polyethylene (XLPE) is now the gold-standard acetabular liner material. However, we asked if carbon fibre reinforced polyether ether ketone (CFR-PEEK) might be a lower wear material. In addition, we sought to understand the influence of contact stress on the wear of both XLPE and CFR-PEEK as this has not previously been reported. Materials and Methods. A 50-station circularly translating pin-on-disc (SuperCTPOD) machine was used to wear test both XLPE and CFR-PEEK pins against cobalt chromium (CoCr) discs to investigate the influence of contact stress on their wear rates. Fifty XLPE and 50 CFR-PEEK pins were articulated against CoCr discs. The pins, 9 mm in outer diameter and 12 mm in height, were drilled with different diameter holes to generate different sized annuli and thus, different contact areas. The pins were tested at 1.10, 1.38, 1.61, 2.00 and 5.30 MPa, which are typical contact stresses observed in the natural hip joint. An additional pin for every test group was used as a control to track the lubricant uptake. The discs were polished to 0.015 μm Sa prior to testing. The test stations contained 16 ml of diluted newborn calf serum (protein concentration: 22 g/L). Wear was measured gravimetrically with a balance (resolution: 10 μm) every 500,000 cycles. A standardised cleaning and weighing protocol was followed. Results and Discussion. The wear rates for the XLPE pins were calculated as 1.05, 0.90, 0.77, 0.48 and 0.28 mg/million cycles for the different pin stress groups respectively. The wear rates decreased with increasing contact stress, which was similar to what was observed for ultra-high molecular weight polyethylene (UHMWPE). The change in weight of the discs was insignificant (p-value:0.85). For the CFR-PEEK pin groups, the wear rates were calculated as 0.56, 0.65, 0.61, 0.58 and 0.65 mg/million cycles respectively. The difference between the wear rates was insignificant (p-value: 0.92). However, the weight of the discs decreased significantly (p-value: 0.00). At 1.11 MPa and taking data for UHMWPE tested in the same way, comparison of the three polymers showed that CFR-PEEK produced the lowest wear against CoCr. Although the wear rates for CFR-PEEK were found to be the lowest, the decrease in weight of the CoCr discs articulated against CFR-PEEK was indicative of metallic wear. Conclusion. CFR-PEEK should not be used against orthopaedic metals. XLPE articulating against CoCr was found to be the optimum combination, producing low wear without causing weight change from the counterface, under varying contact stresses


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_8 | Pages 13 - 13
1 Aug 2020
Atrey A Wu J Waddell JP Schemitsch EH Khoshbin A Ward S Bogoch ER
Full Access

The purpose of this investigation is to assess the rate of wear the effect once the “bedding in period”/ poly creep had been eliminated. Creep is the visco-elastic deformation that polyethylene exhibits in the first 6–12 weeks. We also assessed the wear pattern of four different bearing couples in total hip arthroplasty (THA): cobalt-chrome (CoCr) versus oxidized zirconium (OxZir) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) versus highly-crosslinked polyethylene (XLPE) acetabular liners. This was a randomized control study involving 92 patients undergoing THA. They were randomized to one of four bearing couples: (1) CoCr/UHMWPE (n= 23), (2) OxZir/UHMWPE (n=21), (3) CoCr/XLPE (n=24), (4) OxZir/XLPE (n=24). Patients underwent a posterior approach from one of three surgeons involved in the study. All patients received a porous-coated cementless acetabular shell and a cylindrical proximally coated stem with 28 mm femoral heads. Each patient was reviewed clinically and radiographically at six weeks, three and 12 months, two, five and 10 years after surgery. Standardized anteroposterior and lateral radiographs were taken. All polyethylene wear was measured by an independent blinded reviewer. Linear and volumetric wear rates were measured on radiographs using a validated computer software (Polyware Rev. 5). Creep was defined as the wear at 6 or 12 weeks, depending on if there was a more than 10% difference between both measurements. If a greater than 10% difference occurred than the later period's wear would be defined as creep. 72 hips were included in analysis after exclusion of seven revisions, three deaths and 10 losses to follow-up. The annual linear wear rates (in mm/y) at 10 years were (1) 0.249, (2) 0.250, (3) 0.074 and (4) 0.050. After adjusting for creep these rates become were (1) 0.181, (2) 0.142, (3) 0.040 and (4) 0.023. There is statistical differences between raw and adjusted linear wear rates for all bearing couples. The percentage of the radiographically measured wear at 10 years due to creep is (1) 30% (2) 44%, (3) 58.5% and (4) 51.5% with significant differences in couples with XLPE versus those with UHMWPE. There was no significant correlation between age, gender, cup size, tilt, planar anteversion and the linear or volumetric wear rates. The linear wear rate of both UHMWPE and XLPE are even lower thxdsxzan previously described when creep is factored out. XLPE has again demonstrated far superior linear wear rates at 10 years than UHMWPE. There were no significant differences in wear rate at 10 years between CoCr and OxZir, this may be due to an underpowered study. XLPE exhibits proportionally more creep than UHMWPE within the first 6–12 weeks and accounts for more of the total wear at 10 years as measured radiographically at the end period


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 56 - 56
1 Apr 2018
Clarke I Shon W Lu Z Donaldson T
Full Access

Expectations for ceramic-on-metal (COM) bearings included (i) optimal lubrication due to smoother ceramic heads (ii), reduction of metal ions due to elimination of CoCr heads, and (iii) ‘differential hardness’ reducing adhesive wear and squeaking (Firkins 2001, Williams 2007). Additional benefits included (iv) use of heads larger than for ceramic-on-ceramic (COC), (v) reduction in taper corrosion and (vi) simulator studies clearly demonstrated metal ions and wear both reduced compared to MOM (Firkins 2001, Williams 2007, Ishida 2007). However, contemporary ‘3rd body wear’ paradigms focused only on metal debris size range 0.025–0.035um (Firkins 2001). Thus, neglected was the effect of hip impingement, provoking release of large metal particles sized 20–200um (Clarke 2013). In this study, we compared COM retrievals using hypotheses that adverse COM cases would demonstrate a combination of (a) steeply inclined cups, (b) liner “edge-loading”, (c) Ti6Al4V contamination on ceramic, and (d) evidence of 3rd-body CoCr wear by large particles. As a case example, this 51-year old female had her metal-polyethylene (MPE) bearing revised to COM in June 2011. She reported no symptoms 1-year post-op, but scans revealed a palpable mass in the inguinal region of left hip. By March 2013 the patient reported mild pain in her hip, which progressed to severe by April 2014. Scans showed a solid and cystic iliopsoas bursitis while cup position had changed from 43o to 73o inclination. Revision was performed in June 2014, her joint tissues were found extensively stained due to metal contamination, and histology described formation of a large pseudotumor. Analysis of retrieved components was by interferometry, SEM and EDS. Detailed maps were made of wear areas in heads and cups and volumetric wear was determined by CMM techniques. This adverse COM example revealed large diametral mismatch (595um) compared to COM controls (75–115um). The ceramic head had a broad polar stripe of CoCr contamination, roughness 0.1–0.3um high. Equatorial ceramic areas showed arrays of thin metal smears that demonstrated elemental Ti and Al. The CoCr liner revealed wear area into cup rim, as “edge loading”, and also featured a focal rim-defect over 18o circumferential arc. Liner scratches were 20um wide and larger, and wear-rate of CoCr liner averaged approximately 50mm3 per year. In contrast, ceramic head had minimal wear. Our study highlights the underappreciated risk of impingement by metallic prosthetic components. Prior studies of ceramic heads showed black metallic smears. With COM we can anticipate that the broad polar smear will be CoCr alloy (wear of liner on head). However, Ti6Al4V smearing on ceramic heads is a notable signpost indicating impingement by the Ti6Al4V acetabular shell. The femoral neck (Ti6Al4V: CoCr), may also be damaged. Release of large metal particles, 1500-times larger than prior predictions, provoke a particularly adverse ‘3rd body wear’ (Halim, 2015). Such cases confirm our four hypotheses, that COM bearings will then fail in a way similar to MOM. In contrast, COC bearings are immune to such impingement and 3rd-body metal damage


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 126 - 126
1 Apr 2019
Lal S Hall R Tipper J
Full Access

Currently, different techniques to evaluate the biocompatibility of orthopaedic materials, including two-dimensional (2D) cell culture for metal/ceramic wear debris and floating 2D surfaces or three-dimensional (3D) agarose gels for UHMWPE wear debris, are used. Moreover, cell culture systems evaluate the biological responses of cells to a biomaterial as the combined effect of both particles and ions. We have developed a novel cell culture system suitable for testing the all three type of particles and ions, separately. The method was tested by evaluating the biological responses of human peripheral blood mononuclear cells (PBMNCs) to UHMWPE, cobalt-chromium alloy (CoCr), and Ti64 alloy wear particles. Methods. Clinically relevant sterile UHMWPE, CoCr, and Ti64 wear particles were generated in a pin-on-plate wear simulator. Whole peripheral blood was collected from healthy human donors (ethics approval BIOSCI 10–108, University of Leeds). The PBMNCs were isolated using Lymphoprep (Stemcell, UK) and seeded into the wells of 96-well and 384-well cell culture plates. The plates were then incubated for 24 h in 5% (v/v) CO. 2. at 37°C to allow the attachment of mononuclear phagocytes. Adherent phagocytes were incubated with UHMWPE and CoCr wear debris at volumetric concentrations of 0.5 to 100 µm. 3. particles per cell for 24 h in 5% (v/v) CO. 2. at 37°C. During the incubation of cells with particles, for each assay, two identical plates were set up in two configurations (one upright and one inverted). After incubation, cell viability was measured using the ATPlite assay (Perkin Elmer, UK). Intracellular oxidative stress was measured using the DCFDA-based reactive oxygen species detection assay (Abcam, UK). TNF-α cytokine was measured using sandwich ELISA. DNA damage was measured by alkaline comet assay. The results were expressed as mean ± 95% confidence limits and the data was analysed using one-way ANOVA and Tukey-Kramer post-hoc analysis. Results and Discussion. Cellular uptake of UHMWPE, CoCr and Ti64 particles was confirmed by optical microscopy. PBMNCs incubated with UHMWPE particles did not show any adverse responses except the release of significant levels of TNF-α cytokine at 100 µm. 3. particles per cell, when in contact with particles. PBMNCs incubated with CoCr wear particles showed adverse responses at high particle doses (100 µm. 3. particles per cell) for all the assays. Moreover, cytotoxicity was observed to be a combined effect of both particles and ions, whereas oxidative stress and DNA damage were mostly caused by ions. Ti64 wear particles did not show any adverse responses except cytotoxicity at high particle doses (100 µm. 3. particles per cell). Moreover, this cytotoxicity was mostly found to be a particle effect. In conclusion, the novel cell culture system is suitable for evaluating the biological impact of orthopaedic wear particles and ions, separately


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 48 - 48
1 Feb 2020
Jones H Foley E Garrett K Noble P
Full Access

Introduction. Corrosion products from modular taper junctions are a potent source of adverse tissue reactions after THR. In an attempt to increase the area of contact and resistance to interface motion in the face of taper mismatches, neck trunnions are often fabricated with threaded surfaces designed to deform upon assembly. However, this may lead to incomplete contact and misalignment of the head on the trunnion, depending upon the geometry and composition of the mating components. In this study we characterized the effect of different femoral head materials on the strength and area of contact of modular taper constructs formed with TiAlV trunnions. Materials and Methods. Three groups of 36mm femoral heads (CoCr, Biolox ceramic; Oxinium) and matching Ti-6Al-4V rods with 12/14 trunnions were selected for use in this study. The surface of each trunnion was coated with a 20nm layer of gold applied by sputter-coating in vacuo. Each head/trunnion pair was placed in an alignment jig and assembled with a peak axial impaction force of 2000N using a drop tower apparatus. After assembly, each taper was disassembled in a custom apparatus mounted in a mechanical testing machine (Bionix. MTS. After separation of the components, the surface of each trunnion was examined with backscattered electron microscopy to reveal the area of disruption of the original gold-coated surface. Images encompassing the entire surface of the trunnion were collected and quantified by image processing. Results. The force required to disassemble the Oxinium and Biolox heads from their mating tapers were 2153±104N and 2200±145N, respectively (p-=0.5359). In contrast, the average disassembly force of the CoCr-TiAlV couples was 47% less (1165±156N, p<0.0001). Direct contact between the trunnion and the femoral head was only present over 3.7±0.3% of the nominal surface area of the modular junctions and was limited to the crests of the threads. Contact area did not vary as a function of head composition (p>0.4). However, there were noticeable differences in terms of the distribution of contact between the head and the trunnion. CoCr heads typically had large spans of noncontact immediately below the apex of the taper and opposite each other at the trunnion base. Biolox heads tended to have complete contact at the apex but only extended down 30% of the taper and intermittently at the base. Oxinium heads had comparable complete contact areas to Biolox at the apex but unlike Biolox and CoCr, a uniform band of contact existed at the base. Conclusions. CoCr heads provided only half the resistance to disassembly of Biolox and Oxinium heads. The total area of direct head-trunnion contact is minimal and is not affected by head composition. The heads studied had characteristic patterns of interface contact. This may be due to variations in the geometry of the bores within each head combined with cocking of the femoral head during seating as the thread peaks are being deformed. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 91 - 91
1 Jan 2016
Derasari A Gold J Alexander J Kim SW Patel R Parekh J Incavo S Noble P
Full Access

Introduction. Mechanically-assisted corrosion of the head-neck junction present a dilemma to surgeons at revision THR whenever the femoral component is rigidly fixed to the femur. Many remove the damaged femoral head, clean the femoral taper and fix a new head in place to spare the patient the risks associated with extraction and replacement of the well-functioning femoral stem. This study was performed to answer these research questions:. Will new metal heads restore the mechanical integrity of the original modular junction after impaction on corroded tapers?. Which variables affect the stability of the new interface created at revision THR?. Materials and Methods. Twenty-two tapers (CoCr, n=12; TiAlV, n=10) were obtained for use in this study. Ten stems were in pristine condition, while 12 stems had been retrieved at revision THR and with corrosion damage to the trunnion (Goldberg scale 4). Twenty-two new metal heads were obtained for use in the study, each matching the taper and manufacturer of the original component. The following test states were performed using a MTS Machine: 1. Assembly, 2. Disassembly, 3. Assembly, 4. Toggling and 5. Disassembly. All head assemblies were performed wet using 50% calf serum in accordance to ISO 7206-10. During toggling, each specimen's loading axis was aligned 25° to the trunnion axis in the frontal plane and 10° in the sagittal plane (Figure 1). Toggling was performed at 1Hz for 2,000 cycles with a sinusoidal loading function (230N–4300N). During loading, 3D motion of the head-trunnion junction was measured using a custom jig rigidly attached to the head and the neck of each prosthesis. Relative displacement of the head with respect to the neck was continuously monitored using 6 high resolution displacement transducers with an accuracy of ±0.6µm. Displacement data was independently validated using FEA models of selected constructs. Results. The average micromotion of the head vs trunnion interface was greatest at the start of loading and stabilized after approximately 50 loading cycles at an average of 30.6±3.2µm (Figure 2). For CoCr couples, interface motion dropped by 17% when a pristine head was mounted on a corroded stem compared to a new stem (25.7±2.7µm (pristine stem), vs. 30.1±4.6µm (corroded stem), p= 0.4023) (Figure 3). However, addition of a new CoCr head with a corroded titanium stem led to an 73% increase in interface motion after assembly with a new CoCr head (Corroded: 43.4±9.8µm, Pristine: 25.2±7.0µm, p=0.1661). The resistance to head-neck disruption was 15% higher in TIALV/CoCr couples compared to CoCr/CoCr (TiAlV: 2558 ±63N, CoCr: 2226±99N, p=0.0111) and was not affected by the presence of corrosion of the trunnion (1% loss of strength in each case). Discussion. Corrosion at the trunnion does not disrupt the mechanical integrity of the junction when a CoCr head is replaced on a CoCr taper. We are less sure about the mechanical integrity of a TiAlV taper demonstrated by a trend towards increased micromotion at this junction. Further work is required to better elucidate the role of dissimilar metals in the mechanical integrity of the head-neck junction


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 150 - 150
1 May 2016
Lerf R Reimelt I Dallmann F Delfosse D
Full Access

Background. When reversing the hard-soft articulation in inverse shoulder replacement, i.e. hard inlay and soft glenosphere (cf. Figure 1), the tribological behaviour of such a pairing has to be tested thoroughly. Therefore, two hard materials for the inlay, CoCr alloy and alumina toughened zirconia ceramic (ceramys®) articulating on two soft materials, conventional UHMWPE and vitamin E stabilised, highly cross-linked PE (vitamys®) were tested in a joint simulator. Methods. The simulator tests were performed at Endolab GmbH, Rosenheim, Germany, analogue to standardised gravimetric wear tests for hip prosthesis (ISO 14242-1) with load and motion curves adapted to the shoulder. The test parameters differing from the standard were the maximum force (1.0 kN) and the range of motion. A servo-hydraulic six station joint simulator (EndoLab) was used to run the tests up to 5*106 cycles with diluted calf serum at 37° C as lubricant. Visual inspection and mass measurements were done at 0.1, 0.5, 1, 2, 3, 4 and 5 million cycles using a high precision scale and a stereo microscope, respectivly. Results. The wear rates measured in the simulator are summarised in the table below and illustrated in Figure 2. The simulator wear rate of the standard articulation CoCr – UHMWPE is similar to that found in the corresponding pairing for hip endoprosthesis, although the articulation diameter of the glenospheres tested is larger (42 mm compared to 28 – 32 mm in hip joints). Replacing UHMWPE by the cross-linked vitamys®, the wear rate is reduced to about 1/3 for both hard counterparts, CoCr and ceramys®, respectively. Replacing the CoCr inlay by a part made from ceramys® lowers wear by about 37 % in articulation against UHMWPE. This difference is significant (p = 0.002, significance level 5 %). And comparing CoCr and ceramys® against vitamys®, yields a reduction of about 44 %. Which is significant again (p = 0.015, significance level 5 %). The lowest wear rate, with a reduction of about 80 % compared to the standard CoCr – UHMWPE, exhibits the pairing of both advanced materials, ceramys® – vitamys®. Conclusions. Long-term clinical follow-up will confirm if this in-vitro wear reduction leads to longer in-vivo survival of reverse total shoulder arthroplasty. Such a study is under ethic approval, currently. However, the ceramys® inlay offers the benefits of a nickel free inverse shoulder replacement with less x-ray opacity, compared to CoCr. To view tables/figures, please contact authors directly