The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims

We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty.

Materials and Methods

A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 58 - 58
1 Oct 2020
Mihalko WM Heise G Morrow B Arnholt C Kurtz S
Full Access

Introduction. Pitting damage on implants has been reported and attributed to use of electrocautery. This study aimed to distinguish how different TKA bearing surfaces are susceptible to this type of damage and whether surgeons were aware that this damage can occur. Methods. A survey was sent to Hip and Knee Society members to determine what percentage of adult reconstructive surgeons use electrocautery after implantation of components. Three bearing surfaces for primary TKA were selected: CoCr, Oxinium, and zirconium nitride (ZrN) to be damaged by electrocautery with a monopolar (MP, Bovie) and bipolar (BP, Aquamantys) electrocautery with three different energy settings. A comparison of surface damage and backscatter elemental analysis using SEM was performed. Average roughness (Ra), maximal peak-to-valley height (Rz,), kurtosis (Rk), and skewness (Rsk) measurements were collected using a profilometer (DektakXT, Bruker, Tucson, AZ) with a 2.5µm radius stylus to assess an area of 3.8mm by 3.2mm in the central portion of the area for each MP and BP energy setting used to impart damage to the bearing surfaces. A similar undamaged area for each bearing surface was also measured for roughness parameter comparison. Results. Median Rz and Ra measurements were larger for BP damaged areas compared to MP for all bearing surfaces.(Table 1) The Oxinium surface had the greatest increase in roughness parameters of all three bearing surfaces tested. Survey results indicate that a significant percentage of adult reconstructive surgeons use the electrocautery after implants are in place and are not aware of this type of damage. Backscatter SEM analysis found significant changes for BP damage compared to MP. Conclusion. Our in vitro study determined surface damage caused by electrocautery can have significant effects on the bearing surfaces of implants. Our survey determined many arthroplasty experts are unaware that this damage can occur. For any figures or tables, please contact the authors directly


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 916 - 923
1 Sep 2024
Fricka KB Wilson EJ Strait AV Ho H Hopper, Jr RH Hamilton WG Sershon RA

Aims. The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Methods. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6). Results. Using revision for any reason as an endpoint, five-year survival for FB UKAs was 97.2% (95% CI 96.4 to 98.1) compared to 96.0% for MB (95% CI 94.1 to 97.9; p = 0.008). The FB group experienced fewer component revisions (14/2,315, 0.6% vs 12/684, 1.8%; p < 0.001) and conversion arthroplasties (38/2315, 1.6% vs 24/684, 3.5%; p < 0.001). A greater number of MB UKAs underwent revision due to osteoarthritis progression (FB = 21/2,315, 0.9% vs MB = 16/684, 2.3%; p = 0.003). In the MB group, 12 (1.8%) subjects experienced bearing dislocations which required revision surgery. There were 15 early periprosthetic tibia fractures (0.6%) in the FB group compared to 0 for MB (p = 0.035). Conclusion. In similar patient populations, FB UKAs demonstrated slightly higher survival than a commonly used MB design. Adverse event profiles differed by bearing type, with an increased risk of bearing dislocation and OA progression with MB designs, and early periprosthetic tibia fractures for FB designs. Cite this article: Bone Joint J 2024;106-B(9):916–923


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 49 - 49
1 Oct 2019
Noble PC
Full Access

Introduction. The association between CoCr joint replacements and adverse tissue reactions has led to increased interest in alternative materials that are both biocompatible and wear-resistant. One approach is to manufacture components from titanium alloys with a hardened articulating surface to increase resistance to scratching and surface damage caused by third-body particles. In this study we investigate methods for characterizing the performance of retrieved TiAlV components with nitrogen-hardened bearing surfaces. Methods. Surface-hardened titanium knee implants (TiNidium) were retrieved from 18 patients (7.7 ±6.8 years) at revision surgery. After processing, the bearing surface of each component was characterized by stereomicroscopy, SEM, optical profilometry, and incremental nano-indentation hardness testing. A case-matched set of 18 CoCr components (6.7 ±5.6 years) were characterized for comparison. Results. Each bearing surface was graded for microscopic damage classified as pitting, abrasion, scratching, and burnishing using an Injury Severity Score. The components were divided into slight, average, and severe damage groups based on the resulting ISS scores. Representative specimens from each group were then subjected to SEM imaging, 12 roughness measurements, and 3 incremental nano-indentation hardness tests. Conclusion. There was no difference between the severity of surface damage of the hardened and CoCr components ((p=0.67); Table 1). The rate of surface damage was greatest in the first 2 years then decreased exponentially (Figure 1). Surface roughness (Ra) values were similar for both groups (TiAlV: 0.771um; CoCr: 0.884um) but decreased with the severity of visual damage in the TiNidium implants due to secondary burnishing of scratches. The hardness of the TiNidium implants varied with depth below the bearing surface in contrast to the CoCr controls in which hardness did not vary with depth (Figure 2). Our findings show that multiple complementary methods of are needed to adequately characterize the performance of surface hardened implants. For figures, tables, or references, please contact authors directly


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 14 - 14
7 Aug 2023
Langton D Bhalekar R Wells S Nargol M Haston S Natu S Nargol A
Full Access

Abstract. Introduction. Several studies have reported significant cobalt(Co) and chromium(Cr) elevations in the blood of patients with total-knee-replacements (TKRs), and histological signs of metal sensitivity have been reported in up to 44% of patients undergoing revision of their TKRs. We carried out this investigation to determine the source and quantity of metal release in TKRs. Methodology. We identified all TKRs with polished CoCr trays (N=59) [Vanguard=29, Attune=4 and PFC=26]. These were analysed using peer-reviewed [coordinate-measuring-machine (CMM)] methodology to measure the volumetric wear of the polyethylene (PE) bearing surfaces and trays. The trays were analysed using 2D-profilometry (surface roughness-Ra) and 4D-microscopy. Histological and blood metal ion concentration analyses were performed. Results. The median(IQR) PE wear rate was 10(6to20) mm3/year. Microscopic analysis identified pitting on superior surface of 36(49%) trays. Ra [median (IQR)] of superior surface of pitted trays [0.076 (0.060–0.084) µm] showed a statistically significant increase (p<0.001) compared with unpitted trays [0.057(0.049–0.066) µm]. 4D-microscopy and CMM analysis estimated wear volumes of up to 2mm3 secondary to pitting. The median (range) Co and Cr concentrations were 2.5µg/l (0.2–69.4) and 1.7µg/l (0.5–12.5) respectively in 40 patients. Of the tissue samples examined in 30 patients,6 had at-least “mild”-ALVAL infiltrate. All corresponding “ALVAL” explants were found to be pitted and/or show evidence of loosening of the tray. Conclusion. This study provides further evidence that CoCr release in TKR appears to be an under-recognised cause of adverse clinical outcomes. Gross metal ion elevations occurred in association with micromotion/loosening of the tray


Bone & Joint Open
Vol. 4, Issue 10 | Pages 776 - 781
16 Oct 2023
Matar HE Bloch BV James PJ

Aims

The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


Bone & Joint Research
Vol. 13, Issue 4 | Pages 149 - 156
4 Apr 2024
Rajamäki A Lehtovirta L Niemeläinen M Reito A Parkkinen J Peräniemi S Vepsäläinen J Eskelinen A

Aims

Metal particles detached from metal-on-metal hip prostheses (MoM-THA) have been shown to cause inflammation and destruction of tissues. To further explore this, we investigated the histopathology (aseptic lymphocyte-dominated vasculitis-associated lesions (ALVAL) score) and metal concentrations of the periprosthetic tissues obtained from patients who underwent revision knee arthroplasty. We also aimed to investigate whether accumulated metal debris was associated with ALVAL-type reactions in the synovium.

Methods

Periprosthetic metal concentrations in the synovia and histopathological samples were analyzed from 230 patients from our institution from October 2016 to December 2019. An ordinal regression model was calculated to investigate the effect of the accumulated metals on the histopathological reaction of the synovia.


Bone & Joint Open
Vol. 5, Issue 2 | Pages 101 - 108
6 Feb 2024
Jang SJ Kunze KN Casey JC Steele JR Mayman DJ Jerabek SA Sculco PK Vigdorchik JM

Aims

Distal femoral resection in conventional total knee arthroplasty (TKA) utilizes an intramedullary guide to determine coronal alignment, commonly planned for 5° of valgus. However, a standard 5° resection angle may contribute to malalignment in patients with variability in the femoral anatomical and mechanical axis angle. The purpose of the study was to leverage deep learning (DL) to measure the femoral mechanical-anatomical axis angle (FMAA) in a heterogeneous cohort.

Methods

Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A DL workflow was created to measure the FMAA and validated against human measurements. To reflect potential intramedullary guide placement during manual TKA, two different FMAAs were calculated either using a line approximating the entire diaphyseal shaft, and a line connecting the apex of the femoral intercondylar sulcus to the centre of the diaphysis. The proportion of FMAAs outside a range of 5.0° (SD 2.0°) was calculated for both definitions, and FMAA was compared using univariate analyses across sex, BMI, knee alignment, and femur length.


Bone & Joint Open
Vol. 4, Issue 6 | Pages 432 - 441
5 Jun 2023
Kahlenberg CA Berube EE Xiang W Manzi JE Jahandar H Chalmers BP Cross MB Mayman DJ Wright TM Westrich GH Imhauser CW Sculco PK

Aims

Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model.

Methods

We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 42 - 42
1 Oct 2018
McCalden RW Clout A Naudie DD MacDonald SJ Somerville LE
Full Access

Purpose. The success of total knee replacement (TKR) surgery can be attributed to improvements in TKR design, instrumentation, and surgical technique. Over a decade ago oxidized zirconium (OxZr) femoral components were introduced as an alternative bearing surface to cobalt-chromium (CoCr), based on strong in-vitro evidence, to improve the longevity of TKR implants. Early reports have demonstrated the clinical success of this material however no long-term comparative studies have demonstrated the superiority of OxZr implants compared to a more traditional CoCr implant. This study aims to compare long-term survivorship and outcomes in OxZr and CoCr femoral components in a single total knee design. Methods. We reviewed our institutional database to identify all patients whom underwent a TKA with a posterior stabilized OxZr femoral component with a minimum of 10 years of follow-up. These were then matched to patients whom underwent a TKA with the identical design posterior stabilized CoCr femoral component during the same time period by gender, age and BMI. All patients had their patella resurfaced. All patients were prospectively evaluated preoperatively and postoperatively at 6 weeks, 3 months, 12 months, 2 years and every 1 to 2 years thereafter. Prospectively collected clinical outcome measures included, Western Ontario and McMaster Universities osteoarthritis index (WOMAC), Short-Form 12 (SF-12) and Knee Society clinical rating scores (KSCRS). Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years allowing comparison between the two cohorts. Paired analysis was performed to determine if differences existed in patient reported outcomes. Results. There were 194 OxZr TKAs identified and matched to 194 CoCr TKAs. There was no difference in average age (OxZr, 54.2 years; CoCr, 54.4 years), Gender (OxZr, 61.9% female; CoCr, 61.9% female) and average body mass index (OxZr, 35.9 kg/m2; CoCr, 36.4 kg/m2) between the two cohorts. There were no significant differences preoperatively in any of the outcome scores between the two groups (WOMAC (p=0.361), SF-12 (p=0.771) and KSCRS (p=0.161)). Both the SF12 (p=0.787) and WOMAC (p=0.454) were similar between the two groups, however the OxZr TKA cohort had superior KSCRS compared to the CoCr TKA cohort at a minimum of 10 years (173.5 vs. 159.1, p=0.002). With revision for any reason as the end point, there was no significant difference in 10-year survivorship between the two groups (OxZr and CoCr, 96.4%, p=0.898). Similarly, aseptic revisions demonstrated comparable survivorship rates at 10-year between the OxZr (98.9%) and CoCr groups (97.9%) (p=0.404). Conclusion. In this matched cohort study, both groups demonstrated similar improvements in patient reported outcomes, although the OxZr cohort had greater KSCRS scores compared to the CoCr cohort. The reason for this difference is not clear but may represent selection bias, where OxZr implants were perhaps used in more active patients. Implant survivorship, based on revision rates for all causes and/or aseptic reasons, was excellent and similar for both the OxZr and CoCr femoral components at 10 years. Therefore, with respect to implant longevity at the end of the first decade, there appears to be no clear advantage of OxZr compared to CoCr for patients with similar demographics with this specific posterior stabilized TKA design. Further follow-up into the second and third decade may be required to demonstrate if a difference does exist


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


Bone & Joint Research
Vol. 11, Issue 2 | Pages 91 - 101
1 Feb 2022
Munford MJ Stoddart JC Liddle AD Cobb JP Jeffers JRT

Aims

Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia.

Methods

In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 235 - 241
1 Feb 2022
Stone B Nugent M Young SW Frampton C Hooper GJ

Aims

The success of total knee arthroplasty (TKA) is usually measured using functional outcome scores and revision-free survivorship. However, reporting the lifetime risk of revision may be more meaningful to patients when gauging risks, especially in younger patients. We aimed to assess the lifetime risk of revision for patients in different age categories at the time of undergoing primary TKA.

Methods

The New Zealand Joint Registry database was used to obtain revision rates, mortality, and the indications for revision for all primary TKAs performed during an 18-year period between January 1999 and December 2016. Patients were stratified into age groups at the time of the initial TKA, and the lifetime risk of revision was calculated according to age, sex, and the American Society of Anesthesiologists (ASA) grade. The most common indications for revision were also analyzed for each age group.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims

The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components.

Methods

Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 470 - 475
1 Apr 2011
Kendrick BJL Simpson DJ Kaptein BL Valstar ER Gill HS Murray DW Price AJ

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 3 | Pages 393 - 398
1 Apr 2003
Siddique MS Rao MC Deehan DJ Pinder IM

We carried out 60 revision procedures for failed porous coated anatomic total knee replacements in 54 patients, which were divided into two groups. The 14 knees in group I had a well-fixed femoral component at surgery which was retained, and in the 46 knees in group II both tibial and femoral components were loose and were revised using a variety of implants. Our review comprised clinical and radiological assessment. A total of 13 knees required a second revision. Six (42%) in group I failed very early (mean 2.1 years) when compared with seven (15%) in group II (mean 6.8 years). Failure was due to wear of the polyethylene insert by the abraded, retained femoral component (crude odds ratio 4.07; 95% CI 1.07 to 15.5). We recommend a complete change of primary bearing surfaces at the time of revision of an uncemented total knee replacement in order to prevent early wear of polyethylene


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 145 - 149
1 Jun 2021
Crawford DA Passias BJ Adams JB Berend KR Lombardi AV

Aims

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1261 - 1269
1 Jul 2021
Burger JA Zuiderbaan HA Sierevelt IN van Steenbergen L Nolte PA Pearle AD Kerkhoffs GMMJ

Aims

Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs.

Methods

A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 59 - 66
1 Jun 2021
Abhari S Hsing TM Malkani MM Smith AF Smith LS Mont MA Malkani AL

Aims

Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment.

Methods

A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 113 - 122
1 Jan 2021
Kayani B Tahmassebi J Ayuob A Konan S Oussedik S Haddad FS

Aims

The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups.

Methods

This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.