Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_12 | Pages 18 - 18
10 Jun 2024
Haston S Langton D Townshend D Bhalekar R Joyce T
Full Access

Despite advancements, revision rates following total ankle replacement (TAR) are high in comparison to other total joint replacements. This explant analysis study aimed to investigate whether there was appreciable metal particulate debris release from various contemporary TARs by describing patterns of material loss. Twenty-eight explanted TARs (9 designs: 3 fixed and 6 mobile bearing), revised for any reason, were studied. The articulating surfaces of the metal tibial and talar components as well as the polyethylene insert were assessed for damage features using light microscopy. Based on the results of the microscopic analysis, scanning electron microscopy with energy dispersive X-ray spectroscopy was performed to determine the composition of embedded debris identified, as well as non-contacting 3D profilometry. Pitting, indicative of material loss, was identified on the articulating surfaces of 54% of tibial components and 96% of talar components. Bearing constraint was not found to be a factor, with similar proportions of fixed and mobile bearing metal components showing pitting. More cobalt-chromium than titanium alloy tibial components exhibited pitting (63% versus 20%). Significantly higher average surface roughness (Sa) values were measured for pitted areas in comparison to unpitted areas of these metal components (p<0.05). Additionally, metallic embedded debris (cobalt-chromium likely due to pitting of the tibial and talar components or titanium likely from loss of their porous coatings) was identified in 18% of polyethylene inserts. The presence of hard 3rd body particles was also indicated by macroscopically visible sliding plane scratching, identified on 79% of talar components. This explant analysis study demonstrates that metal debris is released from the articulating surfaces and the coatings of various contemporary TARs, both fixed and mobile bearing. These findings suggest that metal debris release in TARs may be an under-recognised issue that should be considered in the study of painful or failed TAR moving forwards.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 31 - 31
1 Apr 2022
Langton D Bhalekar R Joyce T Shyam N Nargol M Pabbruwe M Su E Nargol A
Full Access

Cobalt chrome alloy is commonly used in joint replacement surgery. However, it is recognised that some patients develop lymphocyte mediated delayed type hypersensitivity (DTH) responses to this material, which may result in extensive bone and soft tissue destruction.

Phase 1. United Kingdom: From an existing database, we identified extreme phenotype patient groups following metal on metal (MoM) hip resurfacing or THR: ALVAL with low wearing prostheses; ALVAL with high wear; no ALVAL with high wear; and asymptomatic patients with implants in situ for longer than ten years. Class I and II HLA genotype frequency distributions were compared between these patients’ groups, and in silico peptide binding studies were carried out using validated methodology.

Phase 2. United Kingdom: We expanded the study to include more patients, including those with intermediary phenotypes to test whether an algorithm could be developed incorporating “risk genotypes”, patient age, sex and metal exposure. This model was trained in phase 3.

Phase 3. United Kingdom, Australia, United States. Patients from other centres were invited to give DNA samples. The data set was split in two. 70% was used to develop machine learning models to predict failure secondary to DTH. The predictions were tested using the remaining blinded 30% of data, using time-dependent AUROCs, and integrated calibration index performance statistics.

A total of 606 DNA samples, from 397 males and 209 female patients, were typed. This included 176 from patients with failed prostheses, and 430 from asymptomatic patients at a mean of >10 years follow up. C-index and ROC(t) scores suggested a high degree of discrimination, whilst the IBS indicated good calibration and further backed up the indication of high discriminatory ability. At ten years, the weighted mean survival probability error was < 4%.

At present, there are no tests in widespread clinical use which use a patient's genetic profile to guide implant selection or inform post-operative management. The algorithm described herein may address this issue and we suggest that the application may not be restricted to the field of MoM hip arthroplasty.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 77 - 77
1 Feb 2020
Ramirez-Martinez I Smith S Trail I Joyce T
Full Access

Introduction

Despite the positive outcomes in shoulder joint replacements in the last two decades, polyethylene wear debris in metal-on-polyethylene artificial shoulder joints is well-known as a limitation in the long-term survival of shoulder arthroplasties systems. Consequently, there is an interest in the use of novel materials as an alternative to hard bearing surfaces such as pyrolytic carbon layer (PyroCarbon).

Materials and Methods

In the present study, the unique Newcastle Shoulder Wear Simulator was used (Smith et al., 2015; Smith et al., 2016) to evaluate the wear behavior of four commercially available PyroCarbon humeral heads 43 mm diameter, articulating against conventional ultra-high molecular weight polyethylene (UHMWPE) glenoid inserts with a radius of curvature of 17.5 mm to form an anatomic total shoulder arthroplasty.

A physiological combined cycled “Repeat-motion-load” (RML) (Ramirez-Martinez et al., 2019) obtained from the typical activities of daily life of patients with shoulder implants was applied as a simulator input. A fifth sample of the same size and design was used as a soak control and subjected to dynamic loading without motion during the wear test. The mean volumetric wear rate of PyroCarbon-on-polyethylene was evaluated over 5 million cycles gravimetrically and calculated on the basis of linear regression, as well as the change in surface roughness (Sa) of the components using a non-contacting white light profilometer throughout the test.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 149 - 149
1 Feb 2020
Kandemir G Smith S Joyce T
Full Access

Introduction

Total hip replacement with metal-on-polymer (MoP) hip prostheses is a successful treatment for late-stage osteoarthritis. However, the wear debris generated from the polymer acetabular liners remains a problem as it can be associated with osteolysis and aseptic loosening of the implant. This has led to the investigation of more wear resistant polymers in orthopaedics. Cross-linked polyethylene (XLPE) is now the gold-standard acetabular liner material. However, we asked if carbon fibre reinforced polyether ether ketone (CFR-PEEK) might be a lower wear material. In addition, we sought to understand the influence of contact stress on the wear of both XLPE and CFR-PEEK as this has not previously been reported.

Materials and Methods

A 50-station circularly translating pin-on-disc (SuperCTPOD) machine was used to wear test both XLPE and CFR-PEEK pins against cobalt chromium (CoCr) discs to investigate the influence of contact stress on their wear rates. Fifty XLPE and 50 CFR-PEEK pins were articulated against CoCr discs. The pins, 9 mm in outer diameter and 12 mm in height, were drilled with different diameter holes to generate different sized annuli and thus, different contact areas. The pins were tested at 1.10, 1.38, 1.61, 2.00 and 5.30 MPa, which are typical contact stresses observed in the natural hip joint. An additional pin for every test group was used as a control to track the lubricant uptake. The discs were polished to 0.015 μm Sa prior to testing. The test stations contained 16 ml of diluted newborn calf serum (protein concentration: 22 g/L). Wear was measured gravimetrically with a balance (resolution: 10 μm) every 500,000 cycles. A standardised cleaning and weighing protocol was followed.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 25 - 25
1 May 2019
Langton D Sidaginamale R Wells S Wainwright B Holland J Deehan D Joyce T Jafri A Nargol A Natu S
Full Access

Introduction

We aimed to identify genes associated with the development of ALVAL at relatively low levels of wear.

Methods

At our unit all patients undergoing revision of a MoM hip prosthesis have periprosthetic tissue samples graded for ALVAL. Explants undergo volumetric wear testing of the bearing and taper surfaces. We identified patients with moderate/severe ALVAL who had been exposed to lower than the median wear rate of all recorded patients who had developed ALVAL (<3mm3/year). This was termed the “ALVAL” group. We then identified all patients whose tissues had shown no signs of ALVAL. The patients in the two groups were sent buccal DNA collection kits. DNA was examined using next generation sequencing. Alleleic frequencies in the two groups were compared using Fisher's test and compared to a background UK population group (n=8514). We then conducted binary logistic regression with patient age, sex, primary source of debris (taper/bearing) and HLA genotype as the predictors. With the hypothesis that a cobalt/albumin metalloprotein acts as the epitope, we used validated binding prediction software to determine the relative affinities of the binding grooves created by different DQA1/DQB1 genetic combinations for albumin derived peptides. Given the protection that male sex and younger age appears to confer against ALVAL, we hypothesized that testosterone peptides may compete for these binding sites.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 45 - 45
1 Apr 2019
Joyce T Giddins G
Full Access

Objective

We explanted NeuFlex metacarpophalangeal (MP) joint prostheses to identify common features, such as position of fracture, and thus better understand the reasons for implant failure.

Methods

Explanted NeuFlex MP joint prostheses were retrieved as part of an-ongoing implant retrieval programme. Following revision MP joint surgery the implants were cleaned and sent for assessment. Ethical advice was sought but not required. The explants were photographed. The position of fracture, if any, was noted. Patient demographics were recorded.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 15 - 15
1 Apr 2019
Ramirez-Martinez I Smith S Joyce T
Full Access

INTRODUCTION

Shoulder joint prostheses have become the most commonly replaced after knee and hip artificial implants. Reverse shoulder arthroplasty (RSA) is the treatment option for patients with severe osteoarthritis, rotator cuff arthropathy or a massive rotator cuff tear with pseudoparalysis. Though successful, the long-term survival of such implants are limited by wear of the materials in contact [1, 2]. The aim of this study was to investigate RSA wear in vitro using a clinically relevant activities of daily living (ADLs).

MATERIALS AND METHODS

Four new JRI Orthopaedics Reverse Shoulder 42 mm diameter VAIOS with cobalt-chromium (CoCr) glenospheres and ultra-high molecular weight polyethylene (UHMWPE) humeral components were tested. A five million cycles wear test was undertaken using the unique Newcastle Shoulder Wear Simulator with dilute bovine serum as a lubricant. “Mug to mouth” was performed as the ADL to the test prostheses in intervals of 100 cycles, following by 5 seconds of high load (450N) with no motion simulating an ADL such as “lifting an object”. This combined load cycle was then repeated. A fifth reverse shoulder prosthesis was subject to dynamic loading only in a soak control station. Wear was assessed gravimetrically and roughness (Sa) of the articulating surfaces was measured with a non-contacting profilometer.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 17 - 17
1 Apr 2019
Bhalekar R Smith S Joyce T
Full Access

Introduction

Metal-on-polyethylene (MoP) is the most commonly used bearing couple in total hip replacements (THRs). Retrieval studies (Cooper et al, 2012, JBJS, Lindgren et al, 2011, JBJS) report adverse reactions to metal debris (ARMD) due to debris produced from the taper-trunnion junction of the modular MoP THRs. A recent retrospective observational study (Matharu et al, 2016, BMC Musc Dis) showed that the risk of ARMD revision surgery is increasing in MoP THRs. To the authors' best knowledge, no hip simulator tests have investigated material loss from the taper-trunnion junction of contemporary MoP THRs.

Methods

A 6-station anatomical hip joint simulator was used to investigate material loss at the articulating and taper-trunnion surfaces of 32mm diameter metal-on-cross-linked polyethylene (MoXLPE) joints for 5 million cycles (Mc) with a sixth joint serving as a dynamically loaded soak control. Commercially available cobalt-chromium-molybdenum (CoCrMo) femoral heads articulating against XLPE acetabular liners (7.5Mrad) were used with a diluted new-born-calf-serum lubricant. Each CoCrMo femoral head was mounted on a 12/14 titanium alloy trunnion. The test was stopped every 0.5Mc, components were cleaned and gravimetric measurements performed following ISO 14242-2 and the lubricant was changed. Weight loss (mg) obtained from gravimetric measurements was converted into volume loss (mm3) and wear rates were calculated from the slopes of the linear regression lines in the volumetric loss versus number of cycles plot for heads, liners and trunnions. Additionally, volumetric measurements of the head tapers were obtained using a coordinate measuring machine (CMM) post-test. The surface roughness (Sa) of all heads and liners was measured pre and post-test. At the end of the test, the femoral heads were cut and the roughness of the worn and unworn area was measured. Statistical analysis was performed using a paired-t-test (for roughness measurements) and an independent sample t-test (for wear rates).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 16 - 16
1 Apr 2019
Bhalekar R Smith S Joyce T
Full Access

Introduction

The bearing surfaces of ceramic-on-ceramic (CoC) total hip replacements (THR) show a substantially lower wear rate than metal-on-polyethylene (MoP) THR in-vitro. However, revision rates for CoC THR are comparable with MoP. Our hypothesis that an explanation could be adverse reaction to metal debris (ARMD) from the trunnion led us to investigate the wear at both the bearing surfaces and the taper-trunnion interface of a contemporary CoC THR in an in-vitro study.

Methods

Three 36mm CoC hips were tested in a hip simulator for 5 million cycles (Mc). BIOLOX®delta ceramic femoral heads were mounted on 12/14 titanium (Ti6Al4V) trunnions. Wear of femoral heads, acetabular liners and trunnions was determined gravimetrically using the analytical balance. Roughness measurements (Sa) were taken on the articulating surfaces (pre and post-test) and on the trunnion surfaces (worn and unworn). Furthermore, Energy Dispersive X-ray Spectroscopy (EDX) was used to identify and quantify the wear debris present in the lubricant using scanning electron microscope (SEM).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 7 - 7
1 Jun 2016
Sidaginamale R O'Hare J Natu S Joyce T Lord J Nargol A Langton D
Full Access

Background

Some reports have suggested that debris generated from the head neck taper junction is more destructive than equivalent doses from metal bearing surfaces.

Methods

Part 1. We examined the relationship between the source (taper/bearing) and volume of metal debris on Cr and Co concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using regression analysis of prospectively collected data at a single revision unit. Part 2. We investigated variables most strongly associated with macroscopic soft tissue injury as documented at revision surgery using ordinal logistic regression. Independent variables included source and volume of CoCr exposure, Cr and Co joint fluid concentrations, joint fluid grade, ALVAL (Aseptic Lymphocytic Vasculitis Associated Lesion) grade, presence of vascular hyalinisation, agglomerated particle size, implant type, patient sex and age.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 4 - 4
1 Jun 2016
O'Hare J Langton D Nargol A Joyce T Brewster N Cooke N Jafri A Lord J Duffy P Holland J
Full Access

Introduction

Historical studies have reported incidences of taper corrosion in retrieved MoP hips of 20–50%. These studies relied on visual assessments, rather than using modern analytical techniques.

Patients/Materials and Methods

The Northern Retrieval Registry was initiated to routinely analyse all retrieved hips. The volumetric wear rates of retrieved Exeter head tapers were measured using a coordinate measuring machine using validated methods and compared to the available results obtained from an ongoing study of failed MoM prostheses. Power analysis suggested that we would need at least fifty Exeter head tapers to provide significant results. Non-parametric tests were used to assess differences. We have previously identified head diameter/offset/taper angle and taper surface roughness as variables associated with taper wear. The effect of bearing material combination remains unquantified. These design variables were entered into a multiple regression model following log normalisation of taper wear.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 87 - 87
1 May 2016
Clarke I Burgett-Moreno M Bone M Scholes S Joyce T Donaldson T
Full Access

Metal-on-metal retrieval studies indicated that MOM wear-rates could rise as high as 60–70mm3/year in short-term failures (Morlock, 2008). In contrast, some MOM and ceramic-on-ceramic (COC) devices of 1970's era performed admirably over 2–3 decades (Schmalzreid, 1996; Shishido, 2003). While technology has aided analysis of short-term MOM and COC failures (Morlock 2008; Lord 2011), information on successful THA remains scant. Lack of long-term data creates difficulties in setting benchmarks for simulator studies and establishing guidelines for use in standards. In this study we compared clinical and wear histories for a 30-year MOM and a 32-year COC to establish such long-term, wear-rates.

The McKeeTM retrieval was cemented and made 100% of CoCr alloy (Fig. 1a). This patient had a right femoral fracture at 47 years of age, treated by internal-fixation, which failed. Her revision with a Judet implant also failed, leaving her right hip as a Girdlestone. At the age of 68, she had a McKee THA implanted in left hip, and used it until almost 98 years of age (Campbell, 2003). The COC case was a press-fit AutophorTM THA, head and cup made of alumina ceramic, with the only metal being the CoCr stem (Fig. 1c). This was implanted in a female patient 17-years of age active in sports (water-skiing). This modular THA was revised 32-years later due to hip pain from cup migration. Wear on these implants was identified by stereomicroscopy and stained red for photography (Fig. 1). Cup-to-neck impingement was denoted by circumferential neck notching, roughness was assessed by interferometry, and wear determined by CMM (Lord, 2011).

McKee head wear covered 1092mm2 area (Figs. 1a, 2: hemi-area ratio 58%). There was no stripe wear and head roughness was 36nm (Ra). Cup wear covered an area of 1790mm2 (hemi-area 63%). Circumferential damage was noted on the supero-posterior femoral neck with scuff marks also on posterior collar (Fig. 2c). Head and cup wear amounted to 37.7 and 25.2mm3, respectively. Total MOM wear was 62.9mm3, indicating a wear-rate of 2.1mm3/year.

Ceramic head wear consisted of two circular patterns (Fig. 1c), the major one of area 1790mm2 (hemi-area 79%). No wear stripes were identified. Non-worn and extensively worn surfaces had roughness (Ra) 17nm and 123nm, respectively. The cup showed 360o circumferential arc of rim wear with a small, non-wear zone inferiorly (Fig. 1c). Gray metallic transfer was evident, EDS revealing Co and Cr (Fig. 3a). Head and cup wear volumes were 77.2 and 54mm3, respectively. Total COC wear amounted to 131.2mm3 indicating a wear-rate of 4.1mm3/year.

These two THA functioned successfully over 3 decades. The McKee retrieval had minor signs of impingement but no adverse “stripe wear”. This MOM performed satisfactorily due to good positioning and patient's advanced age (68 to 98Yrs of age). The COC patient was 17 years of age at index surgery and active. The ceramic cup showed 360o of edge wear, CoCr transfer and a COC wear-rate double that of the MOM retrieval. Thus the high ceramic wear-resistance protected this youthful patient.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_12 | Pages 5 - 5
1 Nov 2015
Langton D Killen M Sidaginamale R Bowsher J Savisaar C Nargol A Joyce T Lord J
Full Access

Introduction

In this study, we aimed to investigate the effect of the topography of the female taper surface on taper wear.

Patients/Materials & Methods

We measured volumetric material loss at the female taper surface of the first consecutive 36mm MoM hips received at our centre. The MoM hips were from a single manufacturer and possessed matching: head size (36mm); taper geometry (5.65 degree 12 14 tapers); and metallurgy (cobalt chrome female taper mated with a titanium stem). Volume loss was measured with a coordinate measuring machine using validated methods. Surface roughness was measured using a Surftest SJ400. There were 111 head tapers available for analysis. The majority had been revised secondary to ARMD, and a minority for loosening. The mean duration of implantation was 56 months. Volumetric wear values were log normalised in order to construct a multiple regression model to investigate the interaction of head offset, the head taper angle and the Rp value of the female taper surface. Rp is a roughness parameter, which is a measure of the peaks above the plateau of a surface.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 75 - 75
1 Mar 2013
Bone M Langton D Lord J Patil S Partington P Joyce T
Full Access

Introduction

There is much current debate concerning wear and corrosion at the taper junctions of large head total hip replacements, particularly metal-on-metal hips. Is such damage a modern concern or has it always occurred in total hip replacement but not previously noted. To investigate this five explanted V40 Exeter femoral stems (Stryker Howmedica) were obtained following revision surgery at a single centre. In all cases, the 24–26 mm femoral heads were still attached.

Hypothesis

In conventional ‘small head’ modular hip prostheses such as the Exeter, negligible wear and corrosion is seen at the taper junction of explanted devices.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 15 - 15
1 Mar 2013
Petheram T Bone M Joyce T Partington P
Full Access

Recent NICE guidance recommends use of a well proven cemented femoral stem for hip hemiarthroplasty in management of fractured neck of femur. The Exeter Trauma Stem (ETS) has been designed based on the well proven Exeter hip stem. It has a double taper polished stem design, proclaimed to share geometry and surface finish with the Exeter hip. This study investigated the surface finish of the two stems in order to investigate the hypothesis that they were different. Two ETS and two Exeter stems were examined using a profilometer with a sensitivity of one nanometer. Macroscopic visual inspection showed that the two Exeter stems had significantly smoother surface finish than the ETS stems. The roughness average (RA) values on the ETS stems were approximately an order of magnitude higher than those of the Exeter stems, mean of 0.235μm compared with 0.025μm (p<0.0001). This difference in surface finish has implications for the biomechanical functioning of the stem. Previous change of the Exeter stem to a matt surface-finish in 1976 resulted in a significant increase in stem failure rates and an understanding of the importance of the polished surface-finish in order to function within a taper-slip philosophy. By changing the surface finish in the ETS stem, longevity of the implant may similarly be affected. Clinical results have yet to be published demonstrating this. We recommend the manufacturer reconsiders the surface finish of the ETS stem to ensure it functions as well as the Exeter primary stem with which it shares a design philosophy.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 51 - 51
1 Mar 2013
Smith S Li L Johnson G Joyce T
Full Access

Wear of polymeric glenoid components has been identified as a cause of loosening and failure of shoulder implants1,2 in vivo. A small number of shoulder joint simulators have been built for in vitro wear testing, however none have been capable of testing with physiological motion patterns in three axes and with physiological loading. The Newcastle Shoulder Wear Simulator was designed with three axes of motion, which are programmable so that different activities of daily living might be replicated. The simulator uses three pneumatic cylinders with integral position encoders to move five shoulder prostheses simultaneously in the flexion-extension, abduction-adduction, and internal-external rotation axes. Axial loading is applied with pneumatic cylinders supplied from a pneumatic proportional valve via a manifold, which also supplies a sixth static control station. In order to establish if that the machine can actually perform as intended, commissioning trials were conducted replicating lifting a 0.5 Kg weight to head height as a daily living activity. During the commissioning trials JRI Orthopaedics Reverse VAIOS shoulder prostheses were tested in 50% bovine serum at ambient temperature. The results show that the shoulder joint wear simulator can satisfactorily reproduce a daily living activity deliberately selected for having a large range of motion and loading. Other daily activities, such as drinking from a mug, are less demanding in the ranges of motion and loading and represent no difficulty in being reproduced on the simulator. Now successfully commissioned, this new multi-station shoulder wear simulator can wear test current and new designs of shoulder prosthesis in vitro


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 52 - 52
1 Mar 2013
Bone M Giddins G Joyce T
Full Access

Introduction

Ten explanted pyrolytic carbon components of a number of finger prostheses were obtained at revision surgery for wear analysis. Implants were removed for either dislocation or failure of fixation. Hypothesis Failure of the components was due to wear from the articulating surfaces, as occurs in many hip and knee prostheses.

Methods

The articulating surfaces were examined using a ZYGO NewView 5000 non-contact profilometer with a resolution of 1nm, to determine the roughness average (RA) of the surface. A total of 86 RA measurements were taken. Detailed images of the surface displayed as a 3D map of were acquired. The RA values for each component were averaged and compared against the British standard for orthopaedic implants, which states that the articulating surfaces of devices made of metal or ceramic should have RA values lower than 0.050 µm.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 28 - 28
1 Jan 2013
Langton D Sidaginamale R Lord J Joyce T Natu S Nargol A
Full Access

Background

Previous studies have suggested that the modular junction of metal on metal (MoM) total hip replacements (THR) is an important source of metallic debris.

Methods

We carried out a prospective study using custom techniques to analyse one of the largest collections of failed contemporary MoM devices in the world. All explants from patients who had suffered adverse reactions to metal debris (ARMD) were included in this study. These explants included: 82 36mm THRs, and 147 resurfacing head THRs and 140 resurfacing arthroplasties from several manufactures. Volumetric wear analysis of the bearing surfaces and taper junctions was carried out using a coordinate measuring machine. The relationships between total metallic loss and metal ion concentrations and the macroscopic and histological tissue appearance of THR patients were compared to those in resurfacing patients. Mann Whitney test for non-parametric data was used to assess significant differences between groups.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 29 - 29
1 Jan 2013
Sidaginamale R Langton D Lord J Joyce T Nargol A
Full Access

Background

We have previously described the relationship between wear rates of MOM components and soft tissue necrosis. In this study we investigated the link between wear rates, metal ion concentrations and osteolysis.

Methods

All unilateral patients who underwent revision of hip resurfacings at our centre were included. Retrieved components were analysed using a coordinate measuring machine to determine total volumetric material loss and rates of wear. Given the accuracy of the wear calculations (which we have previously published), wear rates were considered “abnormal” if ≥3mm3/yr. ROC curves were constructed to determine a Co concentration which would be clinically useful to detect abnormal wear. During revision, the presence/absence of osteolysis was documented.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 82 - 82
1 Sep 2012
Joyce T Lord J Nargol A Langton D
Full Access

Introduction

Total hip prostheses which use a ceramic head within a metal liner are a relatively recent innovation. As such, survivorship rates from independent centres alongside explant analysis are rare. The early clinical experience with this novel ceramic-on-metal (CoM) bearing couple is reported alongside explant analysis of failed devices.

Methods and materials

All CoM hips implanted between 2008 and 2009 at a single hospital by a single surgeon were reviewed. Radiographs were analysed using EBRA software to determine acetabular cup inclination and anteversion angles. Blood metal ion concentrations were measured using inductively coupled plasma mass spectroscopy (ICPMS). Explants were measured for bearing surface and taper wear using a high precision co-ordinate measuring machine (Mitutoyo Legex 322, manufacturer's claimed accuracy 0.8µm). The roughness of the articulating surfaces of heads and liners was measured with a non-contact profilometer (ZYGO NewView 5000, 1nm resolution).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 7 - 7
1 Sep 2012
Gandhi J Sidaginamale R Mereddy P Langton D Joyce T Lord J Natu S Nargol A
Full Access

Background

The failure and subsequent withdrawal of the ASR device in both its resurfacing and THR form has been well documented. The National Joint Registry report of 2010 quoted figures of 12–13% failure at five years. Adverse reaction to metal debris (ARMD) is a poorly understood condition and patients developing severe metal reactions may go unrecognised for sometime.

Patients and Methods

In 2004 a single surgeons prospective study of the ASR bearing surface was undertaken. We present the ARMD failure rates of the ASR resurfacing and ASR THR systems. The diagnosis of ARMD was made by the senior author and was based on clinical history, examination, ultrasound findings, metal ion analysis of blood and joint fluid, operative findings and histopathological analysis of tissues retrieved at revision. Mean follow up was 52 months (24–81) and 70 patients were beyond 6 years of the procedure at the time of writing. Kaplan Meier survival analysis was carried out firstly with joints designated “failure” if the patient had undergone revision surgery or if the patient had been listed. A second survival analysis was carried out with a failure defined as a serum cobalt > 7µg/L. Full explant analysis was carried out for retrieved prostheses.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 17 - 17
1 Sep 2012
Bone M Lord J Patil S Partington P Joyce T
Full Access

Background

Hemiarthroplasty of the hip involves the replacement of the femoral side of the joint with a metal prosthesis, resulting in metal-on-cartilage articulation. The two most common types of hemiarthroplasty used are the Austin Moore and the Thomson, both of which are available in either Titanium (Ti) or cobalt chromium (CoCr). Hemiarthroplasty may be more cost effective in elderly patients who have lower life expectancy and are less active.

Materials and Methods

Three Ti and two CoCr hemiarthroplasty components were obtained following revision surgery. Four had an articulating diameter of 44mm and the other was 46mm diameter. These five hemiarthroplasties were analysed using a Mitutoyo LEGEX322 co-ordinate measuring machine (CMM) (manufacturer's claimed scanning accuracy of 0.8μm). In each case a wear map was generated and the wear volume from the articulating surface was calculated using a bespoke MATLAB program.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 104 - 104
1 Sep 2012
Joyce T Lord J Nargol A Meek D Langton D
Full Access

Metal-on-metal hip resurfacing prostheses are a relatively recent intervention for relieving the symptoms of common musculoskeletal diseases such as osteoarthritis. While some short term clinical studies have offered positive results, in a minority of cases there is a recognised issue of femoral fracture, which commonly occurs in the first few months following the operation. This problem has been explained by a surgeon's learning curve and notching of the femur but, to date, studies of explanted early fracture components have been limited.

Tribological analysis was carried out on fourteen retrieved femoral components of which twelve were revised after femoral fracture and two for avascular necrosis (AVN). Eight samples were Durom (Zimmer, Indiana, USA) devices and six were Articular Surface Replacements (ASR, DePuy, Leeds, United Kingdom). One AVN retrieval was a Durom, the other an ASR. The mean time to fracture was 3.4 months. The AVNs were retrieved after 16 months (Durom) and 38 months (ASR).

Volumetric wear rates were determined using a Mitutoyo Legex 322 co-ordinate measuring machine (scanning accuracy within 1 micron) and a bespoke computer program. The method was validated against gravimetric calculations for volumetric wear using a sample femoral head that was artificially worn in vitro. At 5mm3, 10mm3, and 15mm3 of material removal, the method was accurate to within 0.5mm3. Surface roughness data was collected using a Zygo NewView500 interferometer (resolution 1nm).

Mean wear rates of 17.74mm3/year were measured from the fracture components. Wear rates for the AVN retrievals were 0.43mm3/year and 3.45mm3/year. Mean roughness values of the fracture retrievals (PV = 0.754, RMS = 0.027) were similar to the AVNs (PV = 0.621, RMS = 0.030), though the AVNs had been in vivo for significantly longer.

Theoretical lubrication calculations were carried out which found that in both AVN retrievals and in seven of the twelve cases of femoral fracture the roughening was sufficient to change the lubrication regime from fluid film to mixed. Three of these surfaces were bordering on the boundary lubrication regime. The results show that even before the femoral fracture, wear rates and roughness values were high and the implants were performing poorly.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 3 - 3
1 Sep 2012
Langton D Ghandi J Sidaginamale R Mereddy P Joyce T Lord J Natu S Nargol A
Full Access

Introduction

We conducted independent wear analysis of retrieved metal on metal (MoM) hip components from around the world. All patients with resurfaced hips who developed adverse reactions to metal debris (ARMD) were found to have increased wear of the bearing surfaces. This was untrue in patients with large diameter (?36mm) MoM total hip replacements. This led us to search for other factors leading to ARMD.

Methods

MoM THR explants retrieved from 78 patients suffering ARMD underwent full volumetric wear analysis of bearing surface and taper-junctions using coordinate measuring machine. Scanning electron microscopy (SEM) used to characterise material composition of specific areas.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 91 - 91
1 Aug 2012
Joyce T Lord J Langton D Nargol A
Full Access

Introduction

Total hip prostheses which use a ceramic head within a metal liner are a relatively recent introduction. As such, survivorship rates from independent centres alongside explant analysis are rare. The early experience with this novel ceramic-on-metal (CoM) bearing couple is reported.

Methods and Materials

All CoM hips implanted between 2008 and 2009 at a single hospital by a single surgeon were reviewed. Radiographs were analysed using EBRA software to determine acetabular cup inclination and anteversion angles. Blood metal ion concentrations were measured using inductively coupled plasma mass spectroscopy (ICPMS). Explants were measured for bearing surface and taper wear using a high precision co-ordinate measuring machine. The roughness of the articulating surfaces was measured with a non-contact profilometer.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 92 - 92
1 Aug 2012
Joyce T Langton D Lord J Nargol A
Full Access

Background

The worldwide withdrawal of the DePuy Articular Surface Replacement (ASR) device in both its resurfacing and total hip replacement (THR) form on 26 August 2010, after 93,000 were implanted worldwide, has had major implications. The 2010 National Joint Registry for England and Wales quoted figures of 12-13% failure at five years; however these figures may be an underestimate.

Patients and methods

In 2004 a single surgeon prospective study of the ASR bearing surface was undertaken. Presented are the Adverse Reaction to Metal Debris (ARMD) failure rates of the ASR resurfacing and ASR THR systems. The diagnosis of ARMD was made by the senior author and was based on clinical history, examination, ultrasound findings, metal ion analysis of blood and joint fluid, operative findings and histopathological analysis of tissues retrieved at revision. Acetabular cup position in vivo was determined using EBRA software. Mean follow up was 52 months (24-81) and 70 patients were beyond 6 years of the procedure at the time of writing. Kaplan Meier survival analysis was carried out firstly with joints designated ‘failure’ if the patient had undergone revision surgery or if the patient had been listed for revision. A second survival analysis was carried out with a failure defined as a serum cobalt concentration > 7microgrammes/L (MHRA guideline from MDA-2010-069). Full explant analysis was carried out for retrieved prostheses.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 59 - 59
1 Aug 2012
Bone M Cunningham J Field J Joyce T
Full Access

Finger arthroplasty lacks the success seen with hip and knee joint replacements. The Van Straten Leuwen Poeschmann Metal (LPM) prosthesis was intended for the proximal interphalangeal (PIP) joints. However revision rates of 30% after 19 months were reported alongside massive osteolysis. Three failed LPM titanium niobium (TiNb) coated cobalt chrome (CoCr) components were obtained- two distal and one proximal.

All three components were analysed using an environmental scanning electron microscope (ESEM). This gave the chemical composition of the surface to determine if the TiNb surface coating was still intact. The distal components were analysed using a ZYGO non-contact profilometer (1nm resolution) with the proximal component unable to be analysed due to its shape. ZYGO analysis gave the roughness average (Ra) of the surface and determined the presence of scratches, pitting and other damage.

Images obtained from both the ZYGO and the ESEM indicated that the surfaces of all components were heavily worn. On the articulating surfaces of both distal components unidirectional scratching was dominant, while the non-articulating surface showed multidirectional scratching. The presence of unidirectional scratching suggested two-body wear, whilst the multidirectional scratching on the non-articulating surface of the distal component suggested that trapped debris may have caused three-body wear.

The ESEM chemical analysis showed that in some regions on the distal component the TiNb coating had been removed completely and in other areas it had been scratched or penetrated. On the proximal component the TiNb coating had been almost completely removed from the articulating surfaces and was only present in small amounts on the non-articulating surfaces. There was little evidence of bone attachment to the titanium coating which was intended to help provide fixation.

ESEM images showed the coating had been removed in some sections where there was minimal scratching, suggesting this scratching did not impact significantly in the coating removal. Therefore here the main cause of coating removal may have been corrosion, although scratching may have also have played a part.

The osteolysis reported clinically may have been linked to the wear debris from the failed coating.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 77 - 77
1 Aug 2012
Lord J Langton D Nargol A Meek R Joyce T
Full Access

Metal-on-metal hip resurfacing prostheses are a relatively recent intervention for relieving the symptoms of common musculoskeletal diseases such as osteoarthritis. While some short term clinical studies have offered positive results, in a minority of cases there is a recognised issue of femoral fracture, which commonly occurs in the first few months following the operation. This problem has been explained by a surgeon's learning curve and notching of the femur but, to date, studies of explanted early fracture components have been limited.

Tribological analysis was carried out on fourteen retrieved femoral components of which twelve were revised after femoral fracture and two for avascular necrosis (AVN). Eight samples were Durom (Zimmer, Indiana, USA) devices and six were Articular Surface Replacements (ASR, DePuy, Leeds, United Kingdom). One AVN retrieval was a Durom, the other an ASR. The mean time to fracture was 3.4 months. The AVNs were retrieved after 16 months (Durom) and 38 months (ASR).

Volumetric wear rates were determined using a Mitutoyo Legex 322 co-ordinate measuring machine (scanning accuracy within 1 micron) and a bespoke computer program. The method was validated against gravimetric calculations for volumetric wear using a sample femoral head that was artificially worn in vitro. At 5mm3, 10mm3, and 15mm3 of material removal, the method was accurate to within 0.5mm3. Surface roughness data was collected using a Zygo NewView500 interferometer (resolution 1nm).

Mean wear rates of 17.74mm3/year were measured from the fracture components. Wear rates for the AVN retrievals were 0.43mm3/year and 3.45mm3/year. Mean roughness values of the fracture retrievals (PV = 0.754nm, RMS = 0.027nm) were similar to the AVNs (PV = 0.621nm, RMS = 0.030nm), though the AVNs had been in vivo for significantly longer.

Theoretical lubrication calculations were carried out which found that in both AVN retrievals and in seven of the twelve cases of femoral fracture the roughening was sufficient to change the lubrication regime from fluid film to mixed. Three of these surfaces were bordering on the boundary lubrication regime. The results show that even before the femoral fracture, wear rates and roughness values were high and the implants were performing poorly.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 33 - 33
1 Aug 2012
Lord J Langton D Nargol A Joyce T
Full Access

Wear debris induced osteolysis is a recognized complication in conventional metal-on-polyethylene hip arthroplasty. One method of achieving wear reduction is through the use of metal-on-metal articulations. One of the latest manifestations of this biomaterial combination is in designs of hip resurfacing which are aimed at younger, more active patients. But, do these metal-on-metal hip resurfacings show low wear when implanted into patients?

Using a Mitutoyo Legex 322 co-ordinate measuring machine (scanning accuracy less than 1 micron) and a bespoke computer program, volumetric wear measurements for retrieved Articular Surface Replacements (ASR, DePuy) metal-on-metal hip resurfacings were undertaken. Measurements were validated against gravimetric calculations for volumetric wear using a sample femoral head that was artificially worn in vitro. At 5mm3, 10mm3, and 15mm3 of material removal, the method was shown to be accurate to within 0.5mm3.

Thirty-two femoral heads and twenty-two acetabular cups were measured. Acetabular cups exhibited mean volumetric wear of 29.00mm3 (range 1.35 - 109.72mm3) and a wear rate of 11.02mm3/year (range 0.30 - 63.59mm3/year). Femoral heads exhibited mean wear of 22.41mm3 (range 0.72 - 134.22mm3) and a wear rate of 8.72mm3/year (range 0.21 - 31.91mm3/year). In the 22 cases where both head and cup from the same prosthesis were available, mean total wear rates of 21.66mm3/year (range 0.51 - 95.50mm3/year) were observed.

Revision was necessitated by one of five effects; early femoral neck fracture (4 heads), avascular necrosis (AVN) (2 heads, 1 cup), infection (1 head, 1 cup), adverse reaction to metal debris (ARMD) (19 heads, 18 cups) or ARMD fracture (6 heads, 2 cups). Mean paired wear rates for the AVN and infection retrievals were 0.51mm3/year and 3.98mm3/year respectively. In vitro tests typically offer wear rates for metal-on-metal devices in the region of 2-4mm3.

Mean paired wear rates for ARMD and ARMD fracture were 17.64mm3/year and 68.5mm3/year respectively, significantly greater than those expected from in vitro tests. In the 4 cases of early fracture, only the heads were revised so a combined wear rate calculation was not possible. The heads exhibited mean wear rate of 8.26mm3/year. These high wear rates are of concern.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 51 - 51
1 Aug 2012
Scholes S Joyce T
Full Access

Although bovine serum is the lubricant recommended by several international standards for the wear testing of orthopedic biomaterials there are issues over its use. The inherent batch variation in protein content means that two bovine serum lubricants can give different wear rates. Due to degradation, the lubricant needs to be changed regularly, so that any third body wear particles are removed, thus potentially influencing wear regimes. There are also cost and safety issues with the use of bovine serum. For these reasons, alternative lubricants were investigated.

A 50-station wear test rig was used, which applied multi-directional motion to each ultra-high molecular weight polyethylene (UHMWPE) test pin. Each pin articulated against a cobalt chrome plate polished to better than 0.05 microns Ra. The following lubricants were used: 50% dilute bovine serum; soy protein; olive oil; wheatgerm oil; soya oil; albumin and globulin (AG) mix; albumin, globulin and chondroitin sulphate (AGC) mix; whole milk; Channel Island milk; 11 mg/ml protein egg white; 20 mg/ml egg white; and 40 mg/ml egg white. A minimum of 6 UHMWPE pins per lubricant were wear tested and the tests ran to 2.5 million cycles. Gravimetric measurements were taken throughout the test to determine the volume of wear and at the end of the test the samples were examined using a SEM.

The lubricants giving the closest results to bovine serum were 20 and 40 mg/ml egg white, with mean UHMWPE total wear volumes of 17.4 mm3 and 17.8 mm3 compared to bovine serum which gave 20.7 mm3. Surface topographies showed similar features too. The 11 mg/ml egg white lubricant and the AG and AGC lubricants were next closest in terms of wear. An UV absorbance assay found that all the protein based lubricants suffered from a high degradation rate, and the rate increased with increasing protein content.

Egg white may offer a less expensive alternative to dilute bovine serum as a test lubricant although it is likely that it too would need to be changed as regularly as bovine serum.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVI | Pages 13 - 13
1 Apr 2012
Mangat N Langton D Joyce T Jameson S De Smet K Nargol A
Full Access

Bearing diameter and acetabular component orientation have been shown to be important variables effecting blood metal ion levels following hip resurfacing arthroplasty. So far no studies on bilateral hip resurfacings have taken into account these variables.

We examined the serum ion results of patients under the care of two experienced hip resurfacing surgeons who carry out ion analysis as part of routine post operative care. Surgeon 1: Patients were implanted exclusively with a “third generation” resurfacing device. Surgeon 2: Patients were implanted with the same “third generation” device and also a low clearance “fourth generation” resurfacing device. Only ion results from patients who were 12 months post surgery were included. Bilateral patients were matched to unilateral patients according to the surgeon performing the operation; the resurfacing system implanted and cup inclination and anteversion angles. The ion data from each bilateral group was tested against the corresponding unilateral groups using the Mann Whitney U test for non parametric data. Significance was drawn at p<0.05.

Surgeon 1: There were 310 patients with unilateral joints and 50 with bilateral joints. There were no significant differences with regard to time to follow up, activity levels, joint sizes or cup orientations. Serum chromium (Cr) and cobalt (Co) concentrations were significantly greater in the bilateral group (p<0.001). Median ion levels were greater in the bilateral group by a factor of >2 in the smallest joint sizes and <2 in the largest joints. Surgeon 2: There were 11 patients with bilateral third generation resurfacing joints and 50 with unilateral joints of the same design. The same relationship as described above was identified. There was a notable difference in the fourth generation implant group (n=13 bilateral, n=100 unilateral). Median ion levels for patients receiving bilateral joints of sizes <47mm were ten times greater than in the corresponding unilateral group.

Bearing diameter and component design are critical factors in determining metal ion levels following bilateral hip resurfacings. Surgeons must consider the potential implications of gross increases in metal ion levels prior to performing bilateral hip resurfacings in smaller patients.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Joyce T Langton D Lord J Grigg H Jameson S Cooke N Tulloch C Logishetty R Meek D Nargol A
Full Access

Background: There is a paucity of published data with regard to the wear of failed metal on metal (MoM) resurfacing devices.

Materials and Methods: MoM components retrieved from patients from two independent centres experiencing failure secondary to ARMD were analysed using a Mitutoyo Legex 322 coordinate measuring machine (CMM) which has an accuracy of 0.8 microns. Between 4000 – 6000 points were taken on each explant, dependent on the size of the bearing surface. Maximum wear depths and total volumetric wear were calculated. These values were compared to those from control samples retrieved following uncomplicated fractures/femoral collapse secondary to avascular necrosis (after calculating equivalent yearly wear rates).

Results: 58 ARMD components were analysed. This included 22 36mm MoM THRs (DePuy Pinnacle), 28 DePuy ASRs and 8 Zimmer Duroms. There were 30 resurfacing fracture/avascular necrosis controls. Volumetric wear rates and maximum wear depths of ARMD resurfacing components were significantly greater than the resurfacing control group for both the ASR and Duroms (p< 0.05) however 2 ARMD components exhibited similar amounts of wear compared to controls. Wear rates of the ARMD THR group were significantly lower than the ARMD resurfacing group (p< 0.05).

Conclusions: Increased articular wear is associated with an increased incidence of local adverse effects including tissue necrosis, joint effusions and fractures. However, there are a minority of patients who can develop tissue necrosis in the absence of accelerated wear, implying a spectrum of sensitivity This is reflected in the incidence of ARMD in the patient groups at the main study centre: > 5% in the ASR group and approx 1% in the THR group. We believe this indicates a failure of adequate lubrication and the resultant negative effects in larger bearing devices.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 549 - 549
1 Nov 2011
Langton D Jameson S Joyce T Nargol A
Full Access

Background: Blood metal ion levels have been shown to correlate with wear of retrieved components. Increased articular wear is associated with early failure. It is not clear what the management of patients with raised metal ions who remain asymptomatic should be.

Methods: A prospective study of the DePuy ASR resurfacing device was commenced in 2004 at our independent centre. Blood and serum metal ion analysis is carried out routinely. We assessed the clinical outcome of all patients with cobalt (Co) concentrations > 5μg/L (n=25). We consider this value to be a clear indicator of a poorly performing bearing surface. Patients who remained asymptomatic were brought back for repeat ions testing and clinical assessment. A parallel study was carried out to examine the relationship between the wear depth of retrieved explants (assessed using a coordinate measuring) and blood ion levels

Results: Increases in chromium(Cr) and Co levels from the second blood sample correlated with the first sample (p< 0.05). The four patients with the highest Cr Co levels went on to suffer spontaneous femoral neck fractures between 3 and 4 years post op (pseudotumours in two of these cases). Of the remaining 25, 5 were revised for worsening pain (large effusions and areas of tissue necrosis were seen at revision) and 10 were under investigation for new onset of pain. Wear depths correlated well with blood ion concentrations (p< 0.05) substantiating the results of another centre.

Conclusion: 60% of patients with raised metal ion levels went on to develop complications within two years of the blood test. This suggests patients may only have temporary tolerances to excessive concentrations of metal wear debris. One novel association appears to be spontaneous delayed femoral fracture. Results suggest increased ion concentrations associated with small joint size/sub optimal cup orientation will continue to rise and these patients should be followed up closely.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_III | Pages 298 - 298
1 Jul 2011
Langton D Jameson S Joyce T Ramasetty N Natu S Antoni N
Full Access

In our independent centre, from 2002 to 2009, 155 BHRs (mean F/U 60 months) have been implanted as well as 420 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems (mean F/U 35)

During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 17 failures of this nature in patients with ASR implants (12 females) and 0 in the BHR group. This amounts to a failure of 3.5% in the ASR group.

Tissue specimens from revision surgery showed varying degrees of “ALVAL” as well as consistently high numbers of histiocytes. Particulate metal debris was also a common finding.

The mean femoral size and acetabular anteversion and inclination angles of the ARMeD group/all asymptomatic patients was 45/49mm (p< 0.001), 27/20°(p< 0.001) and 53/48°(p< 0.08). Median blood chromium(Cr) and cobalt(Co) was 29 and 69 μg/L respectively in the ARMeD group versus 3.9 and 2.7 μg/L in the asymptomatic patients (n=160 with ion levels). Explant analysis confirmed greater rates of wear than expected.

Lymphocyte proliferation studies involving ARMeD patients showed no hyper reactivity to Cr and Co in vitro implying that these adverse clinical developments are mediated by a toxic reaction or a localised immune response.

Our overall results suggest that the reduced arc of cover of the fourth generation ASR cup has led to an increased failure rate secondary to the increased generation of metal debris. This failure rate is 7% in ASR devices with femoral components _47mm.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 38 - 38
1 Jan 2011
Langton D Jameson S Joyce T Webb J Nargol A
Full Access

This report documents the clinical and histological findings, the operative management and the explant analyses of patients with early aseptic failure of large metal-on-metal (MonM) bearing joints.

Three hundred and fifty patients have been implanted with the ASR bearing surface (resurfacing or a modular THR) by a single surgeon at an independent centre since 2004. Six patients (all female) have been revised secondary to aseptic failure. All complained of severe groin pain exacerbated by straight leg raise and routine investigations were unable to establish a cause. Large amounts of sterile, highly viscous green fluid were aspirated from the hip joints in each case. Gross swelling of the pseudocapsule and a similar green fluid surrounding the implants were found at revision surgery. Histological examination of periprosthetic tissue samples showed changes consistent with ALVAL/metallosis, and analysis of the fluid revealed dense numbers of inflammatory cells. Symptoms in patients revised to ceramic-on-ceramic bearings improved post operatively. This was not the case with those reimplanted with MoM joints. Data from a subset of 76 patients (all unilateral resurfacings) showed that malaligned cups (anteverted > 20° +/− inclination angle > 45°) were associated with significantly higher whole blood metal ion levels than cups positioned within this range. All the patients with early aseptic failure had malaligned cups. Independent explant analysis revealed significant increases in the surface roughness values of the articular surfaces.

Our results suggest that some patients develop a significant inflammatory reaction to metal implants. This may be due to high levels of metal debris around the implant as a result of accelerated wear. Poor component alignment was found in all our patients with early aseptic failure.


Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_I | Pages 38 - 38
1 Jan 2011
Jameson S Langton D Joyce T Webb J Nargol A
Full Access

Metal ion concentrations following metal on metal hip resurfacing arthroplasty remain a concern. Variables associated with increased metal ion concentrations need to be established. This study provides metal ion data from a consecutive cohort of the first 76 patients implanted with a fourth generation hip resurfacing prosthesis.

All patients agreed to post-operative blood metal ion sampling at a minimum of one year. Post-operative radiographic measurements of cup inclination and anteversion were obtained using the EBRA software. Mean whole blood chromium (Cr) and cobalt (Co) concentrations in patients receiving the smallest femoral implants (Ł51mm) were greater than in the patients implanted with the largest prostheses (ł53mm) by a factor of 3 and 9 respectively. Ion concentrations in the small femoral group were significantly related to acetabular inclination (R=0.439, P< 0.001 for Cr, R=0.372, P=0.004 for Co) and anteversion (R=0.330, P=0.010 for Cr, R=0.338, P=0.008 for Co). This relationship was not significant in the large implant group. Mean Cr and Co concentrations in patients with accurately orientated cups (inclination < 45°, anteversion < 20°) were 3.7μg/l and 1.8 μg/l respectively, compared to 9.1μg/l and 17.5μg/l in malaligned cups.

A reduced surface contact area caused by cup malalignment may increase contact stresses, resulting in a high wear rate if fluid film lubrication is inadequate. Improved fluid film lubrication has previously been found in larger heads in vitro. Accurate acetabular component positioning is essential in order to reduce metal ion concentrations following hip resurfacing.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 390
1 Jul 2010
Langton D Sprowson A Jameson S Joyce T Reed M Partington P Carluke I Nargol A
Full Access

Background: There are no large comparative metal ion studies of commercially available hip resurfacing devices which have taken into account the effects of femoral size and cup inclination and anteversion.

Patients and methods: Metal ion analysis is carried out routinely at our independent centre. We present the metal ion results of 95 unilateral ASR patients and 70 unilateral BHR patients. For all patients, acetabular cup orientation was assessed using EBRA software. Patients with other metallic implants and those within 12 months of surgery were excluded.

Results: Whole blood/serum chromium (Cr) and cobalt (Co) concentrations were inversely related to femoral component size in both the ASR and BHR group (p< 0.05). Cr and Co levels were only seen to increase in the BHR group when the cup was implanted with an inclination greater than 55°. A significant relationship was identifed between the anteversion of the BHR cup and Cr and Co (p< 0.05 for Co, Spearman Rank correlation), with an increase in ions observed at anteversion angles > 17°. Cr and Co were more strongly influenced by cup position in the case of the ASR, with an increase in metal ions observed at inclinations greater than 45° and anteversion angles of < 10° and > 20°.

Discussion: The increased tolerance of the BHR cup to inclinations between 45–55° is likely due to the larger BHR cup providing greater protection against edge loading. When the cohort was divided by gender, the median Cr concentrations of the male ASR patients were significantly lower than those of the BHR males (p< 0.001). This suggests that in larger components positioned at more satisfactory angles of inclination and anteversion, the lower clearance of the ASR proves more significant than the extra coverage provided by the BHR cup. The BHR appears to be more sensitive to changes in anteversion than inclination.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 397 - 398
1 Jul 2010
Langton D Joyce T Jameson S Nargol A
Full Access

Background: There is a paucity of published literature with regard to metal ion levels following bilateral hip resurfacings.

Method: Recent research has identified femoral component size and acetabular cup anteversion and inclination as important variables involved in metal ion release into the blood. We compared 13 patients with bilateral joints resurfaced using the ASR and 11 patients with bilateral BHRs to those with unilateral joints of similar size and cup inclinations/anteversions. Patients were excluded if the last procedure was within 12 months of blood sampling and if they had other metallic implants. Analysis of standing radiographs to determine cup orientation was carried out using EBRA software. Blood samples were analysed using ICPMS.

Results: Median whole blood Cr and Co values were higher in the BHR bilateral group when compared to the unilateral BHR group by a factor of 1.12 and 1.24 respectively. (5.17 vs 4.59 for Cr, 2.40 vs 1.93 for Co). The difference was significant for Co (p=0.030) but not for Cr (p=0.136). For the large ASR joints (53mm), median Cr and Co values were higher in the bilateral group by a factor of 1.5 and 1.85 respectively. (4.55 vs 2.97 (Cr) 2.83 vs 1.53 (Co)). The difference was significant for both Cr (p=0.001) and Co (p=0.022). For the small ASR joints (< 53mm), median Cr Co values were higher in the bilateral group by a factor of 1.95 and 2.30 respectively. (8.29 vs 4.25 (Cr), 6.78 vs 2.94 (Co)) (p=0.019 Co) (p=0.007 Co).

Conclusion: Metal ion concentrations are significantly greater in patients with bilateral resurfacings. The Cr Co concentrations observed in our patients with bilateral small ASR joints are double those in the published literature on bilateral 28mm metal on metal(MoM) joints implying that the lubrication achieved by small ASRs is sub optimal.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_III | Pages 390 - 391
1 Jul 2010
Langton D Jameson S Joyce T Natu S Logishetty R Tulloch C Nargol A
Full Access

In our independent centre, in the period from January 2003 to august 2008, over 1100 36mm MoM THRs have been implanted as well as 155 Birmingham Hip Resurfacing procedures, 402 ASR resurfacings and 75 THRs using ASR XL heads on SROM stems.

During this period we have experienced a number of failures with patients complaining of worsening groin pain at varying lengths of time post operatively. Aspiration of the hip joints yielded a large sterile effusion on each occasion. At revision, there were copious amounts of green grey fluid with varying degrees of necrosis. There were 11 failures of this nature in patients with ASR implants (10 females) and 2 in the 36 MoM THR group (one male one female).

Tissue specimens from revision surgery showed varying degrees of ‘ALVAL’ as well as consistently high numbers of histiocytes. Metal debris was also a common finding.

A fuller examination of our ASR cohort as a whole has shown that smaller components placed with inclinations > 45° and anteversions < 10 or > 20° are associated with increased metal ion levels. The 11 ASR failed joints were all sub optimally positioned (by the above definition), small components.

Explant analysis using a coordinate measuring machine and out of roundness device confirmed greater than expected wear of each component. The lower number of failures in the 36mm MoM group, as well as the equal sex incidence, suggests that the majority of these failures are due to the instigation of an immune reaction by large amounts of wear debris rather than adverse reactions to well functioning joints. It is likely that small malpositioned ASRs function in mixed to boundary lubrication, and this, combined with the larger radius of these joints compared to the 36mm MoM joints, results in more rapid wear.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 57 - 57
1 Mar 2010
Joyce* T Langton D Jameson S Nargol A
Full Access

Resurfacing metal-on-metal hip arthroplasty is currently showing promising clinical results. However there are concerns related to such implants, including the elevated levels of metal ions typically seen in patients. Valuable data can be obtained from explanted prostheses but due to their recent introduction few retrieval studies on resurfacing hip prostheses have been published.

Five ASR hip resurfacing prostheses were revised due to pain. From two patients, head and cup were available for independent explant analysis. In the other three cases only femoral components were available. All were removed from female patients and all were revised to ceramic-on-ceramic hip prostheses. Post-operative radiographic measurements of cup inclination and ante-version were obtained using the EBRA software. The surface roughness values of the articulating surfaces of the explants were measured using a non-contacting profilometer. A co-ordinate measuring machine was used to measure the diameter of the head and the cup and thus the diametral clearance. The same measurements were then taken from a new unused ASR prosthesis and compared. Using elastohydrodynamic theory the minimum effective film thickness of the implant was calculated. In turn this allowed the lubrication regime to be determined.

The average roughness values of the head and the cup of one implant were found to be 0.135microns and 0.058microns respectively, with a diametral clearance of 110microns. These results indicated that, at the time of removal, the prosthesis would have operated in the boundary lubrication regime. Other explants showed evidence of localised contact between the head and the rim of the acetabular cup, and these showed articulating surfaces with typical roughness values of between 0.025microns and 0.050microns. The new ASR had head and cup surface roughness values of 0.010microns and 0.012microns respectively and a diametral clearance of 87microns, implying that a new implant would operate under fluid film lubrication. All cups five were implanted with inclination angles over 45 degrees and anteversion over 25 degrees.

These results suggest that components with high inclination and anteversion angles display greater than expected wear and may operate in boundary rather than fluid film lubrication which may eventually lead to early failure.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 61 - 61
1 Mar 2010
Joyce* T Huang Y
Full Access

Injections of hyaluronic acid solutions, often known as visco-supplements, into the joints of patients suffering from osteoarthritis are an accepted therapy. In most replacement joints, wear of the biomaterials used in them is a critical concern. For in vitro wear tests of such materials the recommended lubricant is one based on bovine serum. However, unlike synovial fluid, bovine serum does not contain hyaluronic acid. The aim of the work reported here was to take a clinically used hyaluronic acid solution, Ostenil, and to investigate its influence on the wear of two orthopaedic biopolymers

Ultra high molecular weight polyethylene (UHMWPE) and poly tetra fluoro ethylene (PTFE) were tested in turn using a four-station, multi-directional, pin-on-plate wear test rig which had previously been shown to reproduce clinical wear factors for UHMWPE, PTFE and polyacetal. For each biopolymer three lubricants were employed: 33% bovine serum (2 stations); 33% bovine serum + Ostenil (1 station); and distilled water + Ostenil (1 station). Polymeric test pins were subject to a load of 40N and articulated against polished stainless steel plates. Wear factors were determined by dividing the volume lost by the product of the load and the sliding distance (units × 10-6mm3/Nm).

The UHMWPE wear tests ran to 66.3km sliding distance. The addition of Ostenil to dilute bovine serum resulted in a wear factor of 1.4 × 10-6mm3/Nm for UHMWPE. The wear factor was 1.6 × 10-6mm3/Nm when dilute bovine serum alone was used as the lubricant. This shows good agreement with a wear factor of 2.1 × 10-6mm3/Nm reported for failed UHMWPE acetabular cups. PTFE provides an accelerated wear test with clinical validity. In the presence of 33% bovine serum a mean wear factor for PTFE of 40 × 10-6mm3/Nm was measured. The wear factor was 59 × 10-6mm3/Nm for dilute bovine serum plus Ostenil. For explanted PTFE acetabular cups a wear factor of 37 × 10-6mm3/Nm has been calculated. For both polymers wear was least when the lubricant was distilled water plus Ostenil. However a transfer film was found and such films are not clinically valid.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1537 - 1540
1 Nov 2009
Khan WS Dunne NJ Huntley JS Joyce T Reichert ILH Snelling S Scammell BE

This paper outlines the recent development of an exchange Travelling Fellowship scheme between the British and American Orthopaedic Research Societies.


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 164 - 164
1 Mar 2009
Green S Lee S Joyce T Unsworth A
Full Access

The first metatarsophalangeal (MTP) joint is the key joint of the foot in terms of function during gait. Various replacement toe joint prostheses are commercially available but unlike other replacement joints such as the hip or knee, few simulator based studies have been conducted to evaluate the performance and reliability of these prostheses. Presented are results obtained using a newly developed and validated multi-station MTP joint test-rig that is able to simulate the natural biomechanics of the toe joint. The developed simulator is a multi-station computer controlled electro-pneumatic metataso-phalangeal joint simulator that applies dynamic loading and motions commensurate with the walking gait cycle. This involves the combination of plantar-dorsi flexion range of 32 degrees, 5 degrees of inversion/eversion and toe-off dynamic loading peaking at up to 820 N. Presented are the validation and in vitro test results of MTP joint simulations carried out on silastic and articulating metal and polymer designs of MTP prostheses. Despite being subjected to a reduced loading regime of 300 N peak force, the silastic prostheses were found to perform poorly in the simulator, ultimately failing due to a combination of fatigue crack growth and joint collapse. This behaviour and failure mode was consistent with that of ex vivo origin silastic MTP prostheses examined and provides confidence in the validity of the simulator.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_III | Pages 377 - 377
1 Oct 2006
Joyce T Monk D Thompson P Chiu P Unsworth A Green S
Full Access

Introduction: The wear of orthopaedic biopolymers is recognised as a major factor in the failure of total joint replacements. Clinical wear data exists for acetabular cups manufactured from three biopolymers: ultra high molecular weight polyethylene (UHMWPE); poly tetra fluoro ethylene (PTFE); and polyacetal. The aim of this paper was to wear test these biopolymers and compare the results with clinical data.

Method and Materials: The biopolymers were tested using a modified, four-station, pin-on-plate wear rig [1]. In the tests, two of the four stations applied reciprocating motion and two applied multi-directional motion. Biopolymer pins articulated against stainless steel plates under a load of 40N. The lubricant consisted of 25% bovine serum and 75% distilled water. A standardised cleaning and weighing protocol was followed, and the biopolymer wear factors were calculated by dividing the volume lost by the product of the load and the sliding distance.

Discussion and Conclusions: Failed and retrieved UHMWPE acetabular cups have been reported as having a clinical wear factor of 2.1 x 10−6mm3/Nm [2]. However, UHMWPE cups which have been functioning well until removal at post-mortem have been said to show 45 to 69% less wear than revised UHMWPE cups [3]. Combining these values suggests clinical wear factors for functional UHMWPE in the range of 0.95 to 1.45 x 10−6mm3/Nm. This range fits well with the value of 1.1 x 10−6mm3/Nm shown in table 1 for UHMWPE under multi-directional motion. A clinical wear factor of 37 x10−6mm3/Nm has been calculated for PTFE acetabular cups [4]. When compared with the mean wear factor for PTFE pins under multi-directional motion obtained from the pin-on-plate rig, the match is remarkable. For polyacetal cups a mean volumetric wear of 136mm3/ year has been reported [5] and it has been calculated that explanted hip prostheses averaged 1.54 million cycles/year [2]. In polyacetal acetabular cups of 37mm diameter, an average sliding distance of 25mm/cycle can be calculated [6] and it has been said that an equivalent static load of 1000N applies [7]. Taking these four values permits a clinical wear factor for polyacetal cups of 3.5 x 10−6mm3/Nm to be calculated. This number compares well with the value of 3.8 x 10−6mm3/Nm seen for the polyacetal test pins under multi-directional motion. In summary, all three biopolymers subject to multi-directional motion exhibited clinically relevant values of wear.


Orthopaedic Proceedings
Vol. 85-B, Issue SUPP_I | Pages 14 - 14
1 Jan 2003
Joyce T Milner R Unsworth A
Full Access

Metacarpophalangeal (MCP) arthroplasty usually involves the fitting of a silicone spacer, commonly Swanson prosthesis, but more recently the Sutter prosthesis has been introduced.

Four Sutter MCP prostheses, two each sized 30 and 40, were removed from the right hand of a female patient. The patient aged 61 years ate revision, had longstanding rheumatoid arthritis. Using a single station finger stimulator1 two Sutter size 50 MCP prostheses were tested. This stimulator ran at a speed of 100 cycles per minute. During each of these cycles, which flexed the test prosthesis through a 90° arc of motion, the load across the test prosthesis varied between 10N and 15N after 3000 cycles, the stimulator applied a static ‘pinch’ load and the whole combined load cycle began again. Ringer solution heated at 37°C was used as a lubricant. Clinically, the prostheses had been implanted for 53 months. All four had fractured at the junction of the hinge and distal stem. In the simulator tests the Sutter size 50 prosthesis managed just over 10 million cycles of flexion-extension, including over 3300 ‘pinch’ loads before fracture occurred, at the junction of the distal stem and hinge. The second prosthesis fractured in the same manner after 5.3 million cycles of flexion-extension.

These are the first reported in vitro results of fracture of Sutter prosthesis as well as the first paper to state the site of ex vivo fractures of Sutter prostheses. A computer model described in a recent paper 2 indicated that failure of the Sutter prosthesis should occur at the central hinge region. Clearly the in vitro results and the ex vivo experience disagree with the computer model. McArthur and Milner 3 have shown clinically that the Swanson joint appears to be superior to the Sutter implant, a result confirmed elsewhere4. The finger stimulator has previously caused fracture of Swanson pros-thesis in a time and a manner comparable with surgical experience1. Therefore another correlation with ex vivo results, but testing the Sutter prostheses has further validated the finger simulator.