Abstract
Metal ion concentrations following metal on metal hip resurfacing arthroplasty remain a concern. Variables associated with increased metal ion concentrations need to be established. This study provides metal ion data from a consecutive cohort of the first 76 patients implanted with a fourth generation hip resurfacing prosthesis.
All patients agreed to post-operative blood metal ion sampling at a minimum of one year. Post-operative radiographic measurements of cup inclination and anteversion were obtained using the EBRA software. Mean whole blood chromium (Cr) and cobalt (Co) concentrations in patients receiving the smallest femoral implants (Ł51mm) were greater than in the patients implanted with the largest prostheses (ł53mm) by a factor of 3 and 9 respectively. Ion concentrations in the small femoral group were significantly related to acetabular inclination (R=0.439, P< 0.001 for Cr, R=0.372, P=0.004 for Co) and anteversion (R=0.330, P=0.010 for Cr, R=0.338, P=0.008 for Co). This relationship was not significant in the large implant group. Mean Cr and Co concentrations in patients with accurately orientated cups (inclination < 45°, anteversion < 20°) were 3.7μg/l and 1.8 μg/l respectively, compared to 9.1μg/l and 17.5μg/l in malaligned cups.
A reduced surface contact area caused by cup malalignment may increase contact stresses, resulting in a high wear rate if fluid film lubrication is inadequate. Improved fluid film lubrication has previously been found in larger heads in vitro. Accurate acetabular component positioning is essential in order to reduce metal ion concentrations following hip resurfacing.
Correspondence should be addressed to Editorial Secretary Mr ML Costa or Assistant Editorial Secretary Mr B.J. Ollivere at BOA, 35–43 Lincoln’s Inn Fields, London WC2A 3PE, England; Email: mattcosta@hotmail.com or ben@ollivere.co.uk