Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 469 - 470
1 Apr 2004
Harker R Beaver R
Full Access

Introduction Maquet views are a well established method of determining the axial alignment of the lower limb in the coronal plane and their use in the assessment of total knee replacement is widespread. It is an awkward investigation for both patient and radiographer and we compared them to the information that can be obtained using the new generation of Helical CT scanners.

Results We prospectively studied a cohort of 60 patents undergoing TKR. As part of their routine post-operative follow-up they had a standard series of AP and Lateral radiographs (performed under fluoroscopic control) in addition to the Maquet views and a CT scan of their lower limbs. All plain films were performed at the same hospital by the same group of radiographers, while all CTs were performed on the same scanner using a predetermined protocol, and post processing performed by the same individual. Maquet views gave a good assessment of coronal alignment but were also shown to have a much high interobserver error than CT. Maquet views took on average three times longer for the radiographers to perform than CT, were often repeated as technically demanding, and patients (especially the elderly and infirm) often found it difficult and uncomfortable to comply with the required positioning. CT views on each patient (which incorporated standard slices and a scanogram) took a few minutes of time in the scanner allowing a quick throughput when patients arrived in clinic. Patients were supine, a position they all found easy to adopt, and radiographers reported that they found them less difficult to perform. The femoral and tibial axes were easily determined, and rotation easily assessed from the femoral epicondyles, negating the projection errors due to malrotation that may compromise the accuracy of Maquet views. Radiation dosage for the CT is higher than a single Maquet view, but these are often repeated due to poor exposure, increasing the dosage above that of CT.

Conclusions We have used a spiral CT protocol with much success and feel that its greater accuracy, coupled with the information gained on sagittal alignment and component rotation means that the older Maquet view has now been superceded. We also feel that the amount of information recorded by a single investigation may prove invaluable in subsequent investigation of pain or suspected loosening, and very helpful in planning any required revision.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 341 - 341
1 Jul 2008
Masood U Williams D Norton M
Full Access

Scarf osteotomy improves hallux valgus and can be used for deformities with large intermetatarsal angles. It is designed to minimise shortening of the first ray. The aim of this radiographic analysis was to assess the outcome of patients undergoing Scarf osteotomy at the Royal Cornwall Hospital.

The initial 18 consecutive cases performed by the senior author were analysed using the guidelines recommended by the American Foot and Ankle Society. Standardised anterior-posterior radiographs of the foot were compared pre-operatively and at 6 weeks postoperatively. Measurements of the intermetatarsal angle (IMA), hallux valgus angle (HVA), joint congruency angle (JCA), distal metatarsal articular angle (DMAA), sesamoid position and metatarsal length were used to assess any improvement.

The results showed a significant median reduction of the IMA of 70, HVA of 180, JCA of 50, and the DMAA of 30 (all p values < 0.001). The medial sesamoid position in relation to the first metatarsal also improved from a mean value of 2.28 to 1 using the American Foot and Ankle Society grading system. There was no shortening of metatarsal length as measured using the Hardy and Clapham method.

This study shows that the radiographic outcome of Scarf osteotomy at the Royal Cornwall Hospital compares favourably with that found in the literature. It provides effective correction of moderate to large intermetatarsal angles.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 5 | Pages 822 - 824
1 Sep 1993
Grelsamer R Bazos A Proctor C

We describe the radiographic measurement of the angle of tilt of the patella and relate it to malalignment of the extensor mechanism. The tilt angle is defined as the angle subtended by a line joining the medial and lateral edges of the patella and the horizontal. The radiograph (Merchant type) is taken with the foot pointing up, the lower edge of the film parallel to the ground, and the knee at 30 degrees flexion. The mean tilt angle of a group of patients with signs and symptoms suggesting patellofemoral malalignment was 12 degrees (+/- 6 degrees); in a similar group of control subjects it was 2 degrees (+/- 2 degrees) (p < 0.01). Tilting of 5 degrees was taken to be the limit of normal. For the detection of patellar malalignment, the tilt angle was almost as specific as the congruence angle (92% v 99%) but more sensitive (85% v 25%) and more accurate (89% v 62%).


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 19 - 19
1 Feb 2017
Hori K Nakane K Terada S Suguro T Niwa S
Full Access

INTRODUCTION

Femur is one of the bones in humans that exhibit ethnic, racial, and gender difference. Several basic and clinical studies were conducted to explore these variations. Clinical anthropological studies have dealt with the compatibility of femoral prostheses and osteosythesis and materials with the femur. If there is a misalignment between the Total Knee Arthroplasy (TKA) femoral comportment installation position, Range of Motion (ROM) failure and several problems may arise. The aim of this study was to evaluate anterior bowing of the Japanese femur and to assess the adequacy of TKA femoral comportment installation position.

METHODS

We analyzed 76 normal Japanese and 97 TKA patients. (June 2014-June 2015) The average age of the normal subjects was 62.0±20.90 (24–88) years old and the average of TKA subjects was 73.6±7.9 (53–89) years old. First we defined and measured the anterior curvature and the posterior condylar offset (PCO) in normal japanese femurs. Then in TKA patients we set the implant as same angle of the component. Third, we measured the post operative anterior curvature and PCO. Then calculated the anterior curvature difference and PCO differences and preformed statistical analysis with ROM.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 51 - 51
1 Feb 2017
Kato M Warashina H
Full Access

Background

We occasionally come across cortical atrophy of the femur with cemented collarless polished triple-taper stem, a short time after the operation. This study aimed to estimate the radiographs of cemented collarless polished triple-taper stem taken at three, six, twelve, and twenty-four months after the initial operation.

Methods

Between May 2009 and April 2011, 97 consecutive patients underwent primary total hip arthroplasty and hemiarthroplasty using a SC-stem or C-stem implant. During the 24 month follow-up, radiographic examination was performed on a total of 95 patients (98 hips). Out of those 95 patients, 52 hips had total hip arthroplasty, 45 had osteoarthritis, 5 had idiopathic osteonecrosis, there were two 2 other cases and 46 hips had hemiarthroplasty for femoral neck fractures. The cementing grade was estimated on the postoperative radiographs. The 24 month postoperative radiographs were analyzed for changes in stem subsidence, cortical atrophy and cortical hypertrophy. According to the Gruen zone, cortical atrophy and cortical hypertrophy were classified on the femoral side. We defined no cortical atrophy as grade 0, cortical atrophy less than 1 mm as grade 1, more than 1 mm and less than 2 mm as grade 2, and more than 2 mm as grade 3. We defined Grade 1 as 1 point, Grade 2 as 2 points, and Grade 3 as 3 points. The points in every zone were calculated, and the average per zone was determined.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_28 | Pages 15 - 15
1 Aug 2013
Ferretti A Conteduca F Mazza D Maestri B Bolle G Argento G Redler A Iorio R
Full Access

Introduction

In total knee arthroplasty extramedullary tibial guides could not to be as accurate as requested in obtaining proper alignment perpendicular to the mechanical axis. The aim of this study was to determine the accuracy of an accelerometer-based system (KneeAlign 2; OrthAlign Inc, Aliso Viejo, California) as evaluated by post-op X-rays analysis.

Methods

Between March 2012 and May 2012 thirty consecutive patients with primary gonarthrosis were selected for unilateral total knee arthroplasty (TKA) using a handheld surgical navigation system to perform the tibial resection.

Navigation procedure: The entire system is provisionally secured to the tibia using a spring placed around the leg and is fixed to the proximal aspect of the tibia using 2-headed pins. Before fixing the system proximally, an aiming arm is used to align the top of the device with the anterior cruciate ligament footprint and the medial one third of the tibial tubercle. Distally, a footplate connected to the tibial jig is used to keep the EM jig a set distance off of the tibial surface. A gyrometer within the navigation unit is then able to calculate the posterior slope of the tibial jig. Subsequent anatomical landmarkings of both the lateral and medial malleoli are identified using the distal aspect of the EM jig to establish the tibia's mechanical axis. Similarly, the gyrometer within the navigation unit is able to calculate the varus or valgus alignment of the tibial jig relative to the tibia's established mechanical axis. Once anatomical registration has been performed, the tibial cutting block is placed at the proximal aspect of the device, and real-time feedback is provided by the navigation unit to the surgeon, who is then able to set the cutting block's varus/valgus and posterior slope alignment before performing the tibial resection.

Postoperatively, standing anteroposterior hip-to-ankle radiographs and lateral knee-to-ankle radiographs were performed to determine the varus/valgus alignment and the posterior slope of the tibial components relative to the mechanical axis in both the coronal and sagittal planes. The difference between the intraoperative reading of the tibial varus/valgus alignment and posterior slope provided by the system was compared to the radiographic measurements obtained postoperatively for each respective case. Differences were analysed via standard t test. The critical level of significance was set at P <0.05.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_7 | Pages 18 - 18
1 May 2016
Anderson J Campbell P Nelson S
Full Access

Avascular necrosis of the femoral head (AVN) is associated with collapse of the femoral head and arthritic degeneration of the joint. The combination of an implant inserted into the femoral head that provides mechanical support and bone grafting to promote bone formation may offer a possible joint-preserving solution1. Seventeen such procedures were performed between November 2012 and March 2014 during an IRB approved clinical trial. Thirteen out of 18 patients remained unrevised at a minimum of 12 months; the results of radiographic and histological analysis of four revisions are presented.

The investigational device (Figure 1) was developed as a joint preserving treatment for AVN with a clinical grade of IIC or less according to the ARCO grading system2.

The device consisted of a braided spherical Nitinol cage with a Titanium / Nitinol orientation feature. It was implanted using fluoroscopic navigation into a spherical cavity cut into the femoral head via an 11mm diameter access tunnel. Once deployed, the implant was filled with a lightly impacted mixture of autologous bone graft and bone marrow soaked Conduit TCP (DePuy CMW, Blackpool, UK). The implant's purpose was to provide mechanical support to the weakened subchondral surface while the bone graft mixture re-integrated with the host bone.

The retrieved femoral heads were trimmed to leave approximately 3mm of bone around the implant, dehydrated, embedded in methacrylate resin, sectioned and thinned into 50–70µm coronal slices for histological analysis. The following observations were made (Figure 2):

Case 1 (Female, age 70, ARCO IIB, revised after 2 days): The patient was revised for spontaneous sub-trochanteric fracture secondary to osteoporosis. Contact between the native bone and bone graft was observed. Marrow elements and repair tissue were visible within the pores in the graft (Figure 2a).

Case 2 (Male, age 67, ARCO IIIC, revised after 82 days): Two wires were broken but retained within the braided structure. A radiolucent gap caused by the presence of fibrous tissue between the graft mixture and native bone was evident suggesting that the implant was unable to prevent progression in this case.

Case 3 (Female, age 70, ARCO IIC, revised after 482 days): The cavity penetrated the subchondral surface; at revision the implant was found to have breached the articular cartilage. There was partial separation of the proximal osteonecrotic fragment and no evidence of graft revascularisation or remodelling within the implant.

Case 4 (Male, age 42, ARCO IIC, revised after 469 days): There was no indication of bone graft re-integration. Collapse of the necrotic bone and deformation of the implant was diagnosed from 1 year follow-up x-rays.

Conclusion

This treatment has preserved the joints of fourteen patients. Of the four revised, two patients had clinical grades or bone quality contra-indicated for the device and three had lesions occupying more than 30% of the femoral head: Improved criteria for patient selection may be required. The device is only partially load-bearing and incapable of stabilising fractures: The radiolucent band associated with fibrous tissue formation may be an early indication of failure.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_30 | Pages 25 - 25
1 Aug 2013
Di Rollo D Rana B
Full Access

The thickness of the cement mantle surrounding total hip replacements has been used to predict the incidence of aseptic loosening. However little work has been done regarding the significance of the cement mantle distal to the tip of the femoral prosthesis. Results are reported of a radiographic audit study analyzing the thickness of the distal cement column in primary total hip replacement.

In this study the thickness of cement distal to the tip of the femoral prosthesis was measured from 80 post-op AP radiographs taken over a period of 5 months using both the PACS system and plain film x-rays. The mean thickness was 2cm (max 8cm) (min −1.8cm) with a std dev of 1.7cm.

This study demonstrated the wide variation in the thickness of the distal cement column achieved by surgeons. It also highlighted the fact that while 2–5mm is the generally accepted optimal cement mantle surrounding the femoral prosthesis, there is little guidance from the literature with regards the optimal thickness of the cement mantle distal to the tip of the femoral prosthesis. Further work requires to be undertaken to determine the optimal thickness of cement in this area, as well as changes in local surgical practice to ensure the optimal thickness of cement distally is consistently achieved. A re-audit of this work after alterations of surgical technique is desirable.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 213 - 213
1 Mar 2013
Kato M Shimizu T Yasura K Aoto T
Full Access

Background

We occasionally came across cortical atrophy of femurs with cemented collarless polished triple-taper stem in a short term period. This study aimed to estimate radiographs of cemented collarless polished triple-taper stem taken 6 months after the initial operation.

Methods

Between May 2009 and April 2011, 97 consecutive patients underwent primary total hip arthroplasty and hemiarthroplasty using SC-stem or C-stem implants. At the 6 month follow-up, a radiographic examination was performed on 70 patients (71 hips). 44 hips had Total Hip Arthoplasty, 35 had osteoarthritis, 5 had idiopathic osteonecrosis, 2 had other diseases and 27 hips had hemiarthroplasty for femoral neck fractures. The postoperative radiographs were used to estimate the cementing grade. Then the 6 month postoperative radiographs were analyzed for changes in stem subsidence, cortical atrophy and cortical hypertrophy. According to the system of Gruen- cortical atrophy and cortical hypertrophy were classified on the femoral side. We defined no cortical atrophy as grade 0, cortical atrophy less than 1 mm as grade 1, more than 1 mm and less than 2 mm as grade 2, more than 2 mm as grade 3.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 309 - 310
1 Jul 2008
Simpson P Smit A Dall G Breusch S
Full Access

Introduction: An intra-medullary cement restrictor is an integral part of modern cementing technique in total hip arthroplasty. Failure of the restrictor to contain cement, flawed surgical technique or dislocation of the restrictor during pressurisation can all result in a deficient cement mantle. A radiographic analysis of hip replacements using a biodegradable restrictor was undertaken to determine the incidence of restrictor failure, the influence of femoral canal morphology on restrictor failure and to describe the cement mantle quality in successful and failed distal cement restriction.

Methods: x-rays from 299 consecutive hip replacements using the amberflex restrictor were analysed. The cortical index, canal-calcar ratio and femoral type, according to Dorr, were recorded. 3 modes of restrictor failure were identified:

Cement leakage –cement was seen to have escaped past the cement restrictor

Restrictor dislocation – the restrictor was 4 or more centimetres distal to the stem tip

Restrictor penetration –the tip of the femoral stem was resting on the restrictor All cement mantles were given a barrack grading.

Results: 84 cases of restrictor failure were observed – 44 dislocations, 24 leakages and 16 penetrations. The mode of failure was not correlated with femoral type, cortical index or canal-calcar ratio. A strong association was found between restrictor failure and grades c and d cement mantles using the chi squared test. A correlation between cortical index, canal-calcar ratio and femoral type was not observed.

Discussion: The ability of a cement restrictor to occlude the femoral canal and resist pressurisation is very important if a good quality cement mantle is to be achieved. Technical error was likely to be an important factor in many of the observed cases of restrictor failure, especially penetrative failure. Surgical technique is more important than femoral morphology in determining the successful use of this restrictor.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 219 - 219
1 Dec 2013
Kurdziel M Ackerman J Salisbury M Baker E Verner JJ
Full Access

Purpose:

Acetabular bone loss during revision total hip arthroplasty (THA) poses a challenge for reconstruction as segmental and extensive cavitary defects require structural support to achieve prosthesis stability. Trabecular metal (TM) acetabular augments structurally support hemispherical cups. Positive short-term results have been encouraging, but mid- to long-term results are largely unknown. The purpose of this study was to determine the continued efficacy of TM augments in THA revisions with significant pelvic bone loss.

Methods:

Radiographs and medical records of 51 patients who had undergone THA revision with the use of a TM augment were retrospectively reviewed. Acetabular defects were graded according to the Paprosky classification of acetabular deficiencies based on preoperative radiographs and operative findings. Loosening was defined radiographically as a gross change in cup position, change in the abduction angle (>5°), or change in the vertical position of the acetabular component (>8 mm) between initial postoperative and most recent follow-up radiographs (Figure 1).


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_IV | Pages 503 - 503
1 Oct 2010
Dargel J Koebke J Mader K Pennig D Schmidt-Wiethoff R Schneider T
Full Access

Introduction: Drilling of the femoral bone tunnel in anterior cruciate ligament reconstruction may be performed in a transtibial drilling technique or via the anteromedial portal.

Purpose: To determine the accuracy of the radiographic bone tunnel position using either a transtibial or anteromedial drilling technique.

Materials & methods: The postoperative lateral radiographs of 100 patients after anterior cruciate ligament reconstruction were reviewed. In each patient, the femoral bone tunnel was created either through the tibial tunnel or via the anteromedial standard arthroscopy portal. The resulting position of the femoral tunnel was evaluated according to reference values reported by Aglietti (65 % of the cortical femoral A-P distance along Blumenstaat’s line), Amis (60 % of the A-P diameter of the posterior lateral femoral condyle parallel to Blumensaat’s line), and Harner (80 % of the A-P length of Blumensaat’s line). The mean deviation of the radiographic tunnel position from the referenced values was statistically evaluated.

Results: Radiographic bone tunnel positions with transtibial drilling were 62.42 ± 8.36, %, 54.53 ± 8.43 %, and 75.84 ± 9.56 % according to Aglietti, Amis, and Harner, respectively. Bone tunnel positions with anteromedial drilling were 65.46 ± 5.29 %, 59.59 ± 4.18 %, and 79.93 ± 4.24 %, respectively. The mean deviation from the reference values was significantly higher when comparing transtibial to anteromedial drilling. Transtibial drilling resulted in a significantly more anterior bone tunnel position.

Conclusion: Precise bone tunnel placement is a prerequisite for proper postoperative knee function and stability. The results of this study indicate that the accuracy of femoral bone tunnel placement through the anteromedial arthroscopy portal was superior to transtibial drilling. It may therefrore be concluded that drilling the femoral tunnel through the anteromedial portal is recommended when using fixation techniques not depending upon placement of a transtibial guide.


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 177 - 177
1 Mar 2010
Lutz M Myerson M
Full Access

We analyzed the radiographic results of patients treated surgically for flatfoot deformity and who underwent medial cuneiform opening wedge osteotomy as part of the operative procedure. The aim of this study was to confirm the utility of the cuneiform osteotomy as part of the correction of hindfoot and ankle deformity. All patients requiring operative management of flatfoot deformity between January 2002 and December 2007 were prospectively entered in a database. We selected all patients who underwent medial cuneiform opening wedge osteotomy. We measured standardized and validated radiographic parameters on pre and post-operative weight bearing radiographs of the foot. All radiographs were assessed using the digital imaging software package (Siemens). The following measurements were used: lateral talus-1st metatarsal angle; medial cuneiform to floor distance (mm), talar declination angle, calcaneal-talar angle, calcaneal pitch angle, 1st metatarsal declination angle, talonavicular coverage angle, and anteroposterior talus-1st metatarsal angle. Other variables including concomitant surgical procedures, healing of the osteotomy, malunion, and adjacent joint arthritis were also noted. There were 86 patients with a mean age of 36 years (range 9–80). 15 patients had bilateral surgery. The aetiology of the deformity was flexible flat-foot in 48, rupture of the posterior tibial tendon in 41, rigid flatfoot deformity with a fixed forefoot supination deformity in 7, and fixed forefoot varus with metatarsus elevatus in 5. In addition to an opening wedge medial cuneiform osteotomy, a lateral column lengthening calcaneus osteotomy was performed in 80, a gastrocnemius recession in 76, a supramalleolar osteotomy in 2, a triple arthrodesis in 4, a subtalar arthroerisis in 13, excision of an accessory navicular in 6, a tendon transfer in 15 and medial-slide calcaneal osteotomy in 8 patients. The mean lateral talus-1st metatarsal angle improved from 23° to 1°; the mean medial cuneiform to floor distance improved from 20mm to 34mm; the mean talar declination angle improved from 39° to 27°; the mean calcaneal-talar angle improved from 64° to 55°; the calcaneal pitch angle improved from 14° to 23°; the mean 1st metatarsal declination angle improved from 17° to 26°; the mean talonavicular coverage angle improved from 45° to 18°; and the mean anteroposterior talus-1st metatarsal angle improved from 19° to 0° Radiographical analysis confirms that the medial cuneiform opening wedge osteotomy is a reliable and valuable surgical tool in the correction of the forefoot which is associated with flatfoot deformity and that arthrodesis of the 1st metatarsocuneiform joint may not be required to obtain correction of the elevated 1st metatarsal.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_I | Pages 183 - 183
1 Mar 2008
Rossi R Castoldi F La Russa M Germano M Rossi P
Full Access

The ABG stem (Anatomique Benoist Girard, How-medica) was realized of titanium alloy (TA6V) and it was designed to provide anchoring only at the femour’s metaphyseal portion. The long-term stability is achived by osteointegration of the implant at the metaphyseal region, which is coated with hydroxyapatite (HA) crystals. The plasma-sprayed HA coating has a tickness of 60 μm for the stem.

The current study presents radiographic outcomes of more than one hundred primary uncemented HA stems with a long-term follow-up. One hundred and eleven ABG hip arthroplasties were evaluated with a mean follow-up of 9.8 years. The radiographical findings have been classified observing Engh’s stability criteria according to Gruen’s subdivision of the periprosthetic femoral zones.

Dividing the hips into two different groups (in the first one the prosthesis implanted 11, 12 or 13 years ago and in the second one the prosthesis with a maximum age of 10 years) it’s possible to see as in the first group the radiolucent lines are quite reduced. Among the older prosthesis the percentage of radiolucent lines is 3.1% in zones 3 and 5 and 18.8% in zone 4. For what concerns the younger ones the percentage are 10.1% (zone 3), 21.5% (zone 4) and 15.2% (zone 5).

We believe that there is a load transfer from the metaphyseal to the metadiaphyseal portion of the femur without a worsening of the clinical outcomes. The absence of reactive lines and lucencies around the proximal HA-coated portion of the stem supports an excellent circumferential bony ingrowth in the metaphyseal area of the proximal femur. According to these percentages we can say that there’s less presence of radiolucency in the oldest prosthesis and it could be possible to argue that radiolucent lines tend to reduce along the time. This is probably due to the increase of the strength of the stem anchorage.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 89 - 89
1 Sep 2012
Karim A Leffers K Kreuzer S
Full Access

Introduction

The advantages of the direct anterior approach (DAA) for total hip arthroplasty include the preservation of external rotators and hip abductors thus leading to quicker recovery times. To our knowledge, there is no objective method in the literature to predict the level of difficulty for femoral exposure through the DAA. It would be beneficial to the surgeon learning the DAA to assess difficulty pre-operatively to avoid prolonged operative times. The purpose of this study was to develop a predictive model of femoral exposure difficulty in the DAA using a combination of demographic data and radiographic measurements.

Methods

305 post-operative radiographs of consecutive THA's in patients (184 female, 120 male) with primary or secondary osteoarthritis, mean age 64.6 (range 26–91, SD=11.43) performed through the DAA by one of the co-investigators from 12/2005 to 12/2009 were retrospectively reviewed by two separate observers. The observers were blinded to the difficulty level of femoral exposure. Standard post-operative AP pelvis films were assessed with TraumaCad software (TraumaCad 2.2, Voyant Health, Columbia, MD) to make radiographic measurements as shown in Figure 1–2. Each radiograph was calibrated using the size of the femoral head implant. Exclusion criteria included films that had inadequate coverage of the entire pelvis, mal-rotation, or poor exposure. Statistical analysis was performed using STAT 9.1 (StatCorp; College Station, Texas, USA). A two-sided Kruskal–Wallis test was utilized for non-parametric data. Chi-squared tests and Fisher's Exact Test were used to compare proportions. Statistically significant associations were then added to a multivariate model predicting an outcome of difficult exposure.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_II | Pages 146 - 146
1 Feb 2004
Tamaoki S Atsumi T Hiranuma Y Kajiwara T Asakura Y Suzuki J
Full Access

Introduction: The progression of the collapse is influenced by the extent and location of the necrotic focus of the necrotic femoral head. The authors studied the extent of the necrotic focus on the joint surface by conventional antero-posterior radiographs in neutral position, and antero-posterior radiographs at the 45 degrees flexion position. The differences of the lesions in these two types of radiographs were analyzed.

Materials and Methods: The conventional antero-posterior radiographs of the neutral position, and the antero-posterior radiographs at the 45 degrees flexion positions in 115 hip joints of 86 patients with non-traumatic osteonecrosis of the femoral head were obtained for this study. These 115 hips showed extensive lesions and could be divided into two groups: Type C-1 or C-2 by classification of the Specific Disease Investigation Committee under the auspices of the Japanese Ministry of Health Labor and Welfare. Type C1 lesions occupy more than the medial 2/3 of the femoral head and C2 lesions extend laterally to the acetabular edge.

Results: At the neutral position, Type C-1 lesions were noted in 42 joints (36%). In these 42 hips, 13 showed Type C-1 (30%) at the 45 degrees flexion position. In contrast, 29 joints (70%) were Type C-2. Type C-2 lesions in 73 joints (64%) were shown in the neutral position. In these 73 hips, Type C-2 lesions were disclosed on 69 joints (95%) in the 45 degrees flexion position and Type C-1 was noted on 4 hips (5%).

Conclusions: Based on these results, the authors propose that location of the lesion on the joint surface varies with different hip positions.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 225 - 225
1 Jun 2012
Strachan R
Full Access

Degenerate chondral surfaces can be assessed in many ways, but arthroscopy is often performed without proper categorisation, mapping, zoning or sizing of lesions. Progression of disease in un-resurfaced compartments is well-recognised to occur, but is only one of several failure mechanism in partial knee replacement.

A validated ‘Functional Zone’ mapping method was used to document articular surface damage in 250 sequential cases of knee arthroscopy in patients over the age of 40. Size, shape, location and severity of each chondral lesion were noted using the Outerbridge classification. Analysis determined rates of involvement of particular compartments and assessed potential for partial replacement or local treatment and also to consider the risk of future progression. Radiographs including antero-posterior standing, postero-anterior flexion views (Rosenberg), lateral and skyline views were graded (Kellgren and Lawrence) and compared with the arthroscopic findings.

Our results showed that out of the 210 knees with Grade 3 or greater damage 13.3% of knees showed ‘isolated’ medial disease of Outerbridge Grade 3 or worse. Isolated lateral disease was noted in 1.4%, patello-femoral disease in 24.3%, bi-compartmental (Medial/PFJ) disease in 30.9% with tibio-femoral and tri-compartmental disease seen in 15.2%. The combination of lateral and patello-femoral disease was seen in 14.8%. Provided that Grade 1 and 2 changes (which were found in other compartments in high percentages) were ignored and ACL status considered, this information seemed to indicate that at the time these procedures were performed, 13.3% of cases were suitable for a medial uni-compartmental device, with sub-analysis of lesion sizes indicating that 17 out of 28 cases (60.7%) were suitable for a localised resurfacing. Lateral uni-compartmental replacement seemed suitable for only 1.4%, patello-femoral replacement in 24.3%, bi-compartmental in 30.9% and total knee replacement in 30%. The mean age for partial resurfacing was 53years and 59 years for total joint replacements.

Radiological analysis found that the antero-posterior standing views had only 66% sensitivity and 73% specificity for the presence of Grade 3 changes or worse in the medial compartment in comparison with Rosenberg views having a sensitivity of 73% and a specificity of 83%. Skyline views had a sensitivity of 56% and 100% specificity.

This study indicates that a large proportion of cases may be suited to local and limited resurfacing. Cases suitable for Patello-femoral and Bi-compartmental replacements were very common, but with the patella-femoral joint's tendency to be more forgiving in terms of symptoms, meaning that indications for uni-compartmental replacement might well be much higher than the arthroscopic findings suggested. On the other hand, the presence of high levels of Grade 1 and 2 changes in other compartments seems to indicate a need for caution particularly in younger patients. This study also indicates a need for better methods of assessing local cartilage health such as enhanced MRI scanning or spectroscopy.


Orthopaedic Proceedings
Vol. 87-B, Issue SUPP_II | Pages 99 - 100
1 Apr 2005
Laudrin P Babinet A Anract P Tomeno B
Full Access

Purpose: Hinged knee prostheses are mainly used for reconstruction after major tumour resection. Aseptic loosening is the main problem with these implants. One of the solutions proposed to reduce the rate of loosening is to add a hydroxyapatite collar on the shaft stems. This work was conducted to study bone ingrowth with a new hinged implant with a hydroxyapatite collar at the junction between the zone of resection and the shaft.

Material and methods: Twenty-nine massive prostheses with a hydroxyapatite collar were implanted between 1998 and 2001. Nine patients were excluded from the analysis because follow-up was less than two years. This retrospective analysis thus compared twenty massive prostheses with twenty matched hinged GUEPAR prostheses without a collar. Bony ingrowth was measured on plain x-rays (two orthogonal views) at 6, 12, 24, and 36 months. Filling of the gap between the bone and the implant was also assessed. Signs of loosening were noted.

Results: Mean bony ingrowth in implants with a hydroxyapatite collar was 6.58 mm at 6 months 9.84 mm at 12 months, 12.3 mm at 24 months and 13.25 mm at 36 months. Mean bony ingrowth in the implants without a hydroxyapatite collar was 1.65 mm at 6 months, 3.31 mm at 12 months, 4.8 mm at 24 months and 4.35 mm at 36 months. In the implants with a collar, gap filling was partial in five cases and total in 15. In implants without a collar, there was no gap filling in eight cases, partial filling in two cases and total filling in fifteen cases.

Discussion: Prostheses with a hydroxyapatite collar enable better radiological bony ingrowth than observed in implants without a hydroxyapatite collar. Gap filling is better for prostheses with a collar. There was no case of loosening at last follow-up for implants with a hydroxyapatite collar.

Conclusion: In light of these results, shaft anchorage appears to be better with implants with a hydrosyapatite collar. Confirmation of improvement in clinical outcome and lower rate of aseptic loosening will require longer follow-up.


Orthopaedic Proceedings
Vol. 87-B, Issue SUPP_II | Pages 134 - 134
1 Apr 2005
Lazennec J Gorin M Roger B Saillant G
Full Access

Purpose: Uncertain position of the acetabular implant has been the cause of dysfunction in certain cases of total hip arthroplasty (THA). Classical computed tomographic analysis of anteversion has certain limitations. Integrated reconstruction of positions at risk allows a better diagnostic approach.

Material and methods: We studied 46 THA because of posterior malposition (n=17, anterior subluxation in the standing position in twelve, and true dislocation in five) and anterior malposition (n=29, posterior subluxation in sixteen and true dislocation in thirteen). Two groups of 70 naïve hips and a group of 56 THA with no functional problem served as controls. The position of the acetabulum was studied on optimised computed tomography slices reconstructing the planes of analysis for the standing, sitting and reclining positions. The reference planes for the slices was given by the sacral tilt angle measured on the lateral views of the patient in the corresponding positions. The optimised computed tomographic measurements of anteversion were compared with the classical measures. None of the patients had abnormal femoral anteversion and/or an oblique pelvis and/or leg length discrepancy greater than 10 mm. The frontal inclination of the acetabular implants was 40°–50°.

Results: In the naïve hips, acetabular anteversion varied: 19.2 with the conventional method, 15.7 in the standing position and 31 in the sitting position. In the THA controls, anteversion measurements differed: 21.3 with the conventional method, 21.4 in the standing position and 35.8 in the sitting position. In the THA with a posterior malposition, 18/29 could not be explained by the conventional measurement, but the optimised measurement enabled an understanding in 17 hips (defective anteversion in the sitting position).

Discussion: Changes in pelvis orientation between the sitting and standing positions modifies real anteversion of the cup. In particular, subjects with THA tend to have a spontaneous posterior tilt of the pelvis related to trunk ageing. This element should be taken into account for the analysis of both major and minor THA dysfunction.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_II | Pages 309 - 309
1 May 2006
Tamaoki S Atsumi T Hiranuma Y Yamano K Kajiwara T Nakamura K Asakura Y Kato E Watanabe M
Full Access

Introduction: In cases of small and middle size osteonecrosis on conventional antero-posterior (AP) radiographs, we studied the extent of the lesion on AP radiographs at the 45 degrees flexion position for lesions of the anterior area of the femoral head.

Materials and Methods: Classification of Japanese organizing committee was applied for the extent of the lesion on joint surface. Type A lesions occupy the medial one-third or less; Type B, the medial two-thirds or less; Type C-1 occupies more than the medial two-thirds; and Type C-2 extends laterally to the acetabular edge at the neutral position on conventional AP radiographs. Thirty-three hips (25 patients) with small or middle size lesions (Type A;11, Type B;22) were studied. On AP radiographs taken at the 45 degrees flexion position, the extent of the lesion was studied in 33 hips.

Results: Seven of 11 hips of Type A on neutral position showed Type A lesions. For the remaining 4 hips, 2 were Type B, 2 were on Type C-1 at 45 degrees flexion position. For 22 hips with Type B on neutral position, 6 were Type B, 12 were Type C-1, 4 were Type C-2 found at the 45 degrees flexion position.

Discussion: AP radiographs at the 45 degrees flexion position revealed more extensive necrotic area in cases of small and middle size lesions comparing with the neutral position. This finding may be related to progression of the disease.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_I | Pages 39 - 40
1 Jan 2004
Gacon G Philippe M Ray. A
Full Access

Purpose: The purpose of this work was to study the radiological outcome more than seven years after implantation of 89 anatomic non-cemented femoral stems with hydroxyapatite coating around the metaphyseal circumference.

Material: These 81 patients underwent primary arthroplasty (89 hips) between 1991 and 1994 for joint degeneration or necrosis: 48 men and 33 women, mean age 59 years (range 41–78). The inclusion criteria for this study were physical examination and complete radiographic work-up in the second half of 2001. Mean follow-up was nine years (range 7–10).

Methods: All x-rays were analysed by four independent surgeons who used the Engh and Massin criteria. The evaluators, who had not participated in patient care, made their assessment on the basis of the last follow-up clinical report and x-rays. They search for radiographic evidence of stem stability and bony integration as well as signs of osteolysis using the Gruen criteria.

Results: At last follow-up only one femoral stem was painful, but stable. This stem was revised at seven years. Six cups had been changed due to polyethylene wear with iliac osteolysis but without femoral participation. There were no other reoperations. Polyethylene wear was observed in about one-half the hips (44 hips) and was considered severe (1–2 mm) in nine cases. There were no lucent lines nor reactive lines in the metaphyseal area (zones 1 and 7) but 16% of the hips presented reactive lines along the lower, smooth, part of the stem. Endosteal ossification was observed in zones 2 and 6 in 72% of the hips, and less often (13%) in zones 9 and 13. Five hips (5.6%) exhibited bony growth at the tip of the stem producing a thickening in zone 5; these were the only cases with cortical thickening excepting the revised stem (thickening in zones 2 and 6). There was no case of cortical narrowing. Calcar atrophy was observed in 42 hips (47%) with five showing a “drop” aspect. Four hips exhibited osteolysis of the proximal femur in zone 1A, but there were no images of distal osteolysis of the femur.

Discussion: This independent analysis of 89 x-ray files demonstrated that endosteal growth is frequent in the isthmic region. The radiological tolerance was good for this stem (no cortical changes) but there were modifications of the calcar which were difficult; it could be speculated that certain of these modifications might correspond to localised osteolysis at the lowest part of the joint, migration point of polyethyene debris. The role of hydroxyapatite in the observed absence of distal osteolysis is noteworthy.

Conclusion: At nine years follow-up, the absence of osteolysis of the distal femur despite usual polyethylene wear allows the conclusion that hydoxyapatite coating of the metaphyseal circumference creates an effective barrier against wear debris.


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_II | Pages 124 - 124
1 Jul 2002
Koòs Z Kránicz J Bálint L
Full Access

Conservative management of talipes equinovarus has a good effect on adductus deformity of the forefoot, whereas equinus deformity cannot usually be treated well conservatively. However, adductus is the most common recurrent deformity after operations. The aim of the study was to use radiological analysis to explore the reasons that lead to recurrent adductus.

In 86.7% of the cases, either a correction was evaluated as radiologically inadequate but seemed to be good physically, or compensation for an operative over-correction resulted in recurrent adductus some years later. In spite of adequate correction from both a physical and radiological view, recurrent adductus developed in 13.3% of the cases. In our opinion, these recurrences were due to persistent muscle imbalance.

In our department, 458 children were operated on for clubfoot from 1982 to 1997. The patients involved in this study were those managed by medial and posterior soft tissue release after an ineffective six to nine month period of conservative treatment that was started when they were one to two weeks old. Children treated previously in another hospital were excluded from the study. We controlled 228 feet and 42 cases of recurrent adductus were found 2 to 16 years (mean 6.8) after the operations. The radiographs were examined at the end of ineffective conservative treatment, during the early postoperative days, and finally at the follow-up. The anteroposterior talocalcaneal (ATC) angle, the talometatarsal (TM) angle and the naviculometatarsal (NM) angle were measured in all of the radiographs. Based on the measured angles, three main groups of patients were formed.

Recurrent adductus in 24 feet (Group A) was caused by inadequate operative corrections, including inappropriate correction of either the hind foot (reduced ATC angle) or the forefoot (reduced NM angle), or both. Although the talocalcaneal and talometatarsal positions were normal in early postoperative radiographs, adductus developed again two to five years later in seven cases (Group B). In these cases, we think that persistent muscle imbalance was responsible for the recurrent deformity.

In 11 feet the ATC angles were in normal range or increased (Group C). These adductus deformities were caused by either an overcorrected talocalcaneal position resulting in compensatory metatarsal varus or medial subluxation of the talonavicular joint, which had been only partially compensated by the lateral deviation of the 1st ray.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 87 - 97
1 Jan 2021
Burssens A De Roos D Barg A Welck MJ Krähenbühl N Saltzman CL Victor J

Aims

Patients with a deformity of the hindfoot present a particular challenge when performing total knee arthroplasty (TKA). The literature contains little information about the relationship between TKA and hindfoot alignment. This systematic review aimed to determine from both clinical and radiological studies whether TKA would alter a preoperative hindfoot deformity and whether the outcome of TKA is affected by the presence of a postoperative hindfoot deformity.

Methods

A systematic literature search was performed in the databases PubMed, EMBASE, Cochrane Library, and Web of Science. Search terms consisted of “total knee arthroplasty/replacement” combined with “hindfoot/ankle alignment”. Inclusion criteria were all English language studies analyzing the association between TKA and the alignment of the hindfoot, including the clinical or radiological outcomes. Exclusion criteria consisted of TKA performed with a concomitant extra-articular osteotomy and case reports or expert opinions. An assessment of quality was conducted using the modified Methodological Index for Non-Randomized Studies (MINORS). The review was performed according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines and registered in the PROSPERO database (CRD42019106980).


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_II | Pages 210 - 210
1 May 2006
Rydholm U Li Q Kesteris U
Full Access

Different resurfacing implants offer different kinds of positioning instruments. As it is of outmost importance to position the components within rather narrow limits to diminish the risk of femoral notching or impingement we decided to measure the position achieved in 72 hips resurfaced with the Durom® resurfacing hip and instruments.

There were 38 males and 27 females with 72 hips (7 bilateral). The indication was OA in 51 cases, RA in 12 and ON in 2. We compared 2 groups, 26 hips operated with an antero-lateral approach (A) and 46 with a postero-lateral approach (B).

The acetabular cup anteversion angle was 22±11° in group A and 15±9° in group B. The abduction angle was 38±9 ° in group A and 44±7° in group B. The acetabular gap was 2±1 mm, resp. 2±2 mm. The stem-shaft angle was 140±5° resp. 141±6°. Retroverted cups averaged 7±4°.

The difference between pre- and postoperative acetabular size was 3 mm in group A (mostly RA patients) and 5 mm in group B (mostly OA patients).

Conclusions: We have obtained a fairly good implant position. The only significant differences between the two groups were decreased acetabular cup abduction angle compared to the preoperative angle in the antero-lateral group, but increased angle in the postero-lateral group, and that less acetabular bone was removed in the antero-lateral group (patients with RA included) compared to the postero-lateral group.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 601 - 606
1 May 2017
Narkbunnam R Amanatullah DF Electricwala AJ Huddleston III JI Maloney WJ Goodman SB

Aims

The stability of cementless acetabular components is an important factor for surgical planning in the treatment of patients with pelvic osteolysis after total hip arthroplasty (THA). However, the methods for determining the stability of the acetabular component from pre-operative radiographs remain controversial. Our aim was to develop a scoring system to help in the assessment of the stability of the acetabular component under these circumstances.

Patients and Methods

The new scoring system is based on the mechanism of failure of these components and the location of the osteolytic lesion, according to the DeLee and Charnley classification. Each zone is evaluated and scored separately. The sum of the individual scores from the three zones is reported as a total score with a maximum of 10 points. The study involved 96 revision procedures which were undertaken for wear or osteolysis in 91 patients between July 2002 and December 2012. Pre-operative anteroposterior pelvic radiographs and Judet views were reviewed. The stability of the acetabular component was confirmed intra-operatively.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 36 - 42
1 Jan 2014
Liebs T Nasser L Herzberg W Rüther W Hassenpflug J

Several factors have been implicated in unsatisfactory results after total hip replacement (THR). We examined whether femoral offset, as measured on digitised post-operative radiographs, was associated with pain after THR. The routine post-operative radiographs of 362 patients (230 women and 132 men, mean age 70.0 years (35.2 to 90.5)) who received primary unilateral THRs of varying designs were measured after calibration. The femoral offset was calculated using the known dimensions of the implants to control for femoral rotation. Femoral offset was categorised into three groups: normal offset (within 5 mm of the height-adjusted femoral offset), low offset and high offset. We determined the associations to the absolute final score and the improvement in the mean Western Ontario and McMaster Universities osteoarthritis index (WOMAC) pain subscale scores at three, six, 12 and 24 months, adjusting for confounding variables.

The amount of femoral offset was associated with the mean WOMAC pain subscale score at all points of follow-up, with the low-offset group reporting less WOMAC pain than the normal or high-offset groups (six months: 7.01 (sd 11.69) vs 12.26 (sd 15.10) vs 13.10 (sd 16.20), p = 0.006; 12 months: 6.55 (sd 11.09) vs 9.73 (sd 13.76) vs 13.46 (sd 18.39), p = 0.010; 24 months: 5.84 (sd 10.23) vs 9.60 (sd 14.43) vs 13.12 (sd 17.43), p = 0.004). When adjusting for confounding variables, including age and gender, the greatest improvement was seen in the low-offset group, with the normal-offset group demonstrating more improvement than the high-offset group.

Cite this article: Bone Joint J 2014;96-B:36–42.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 25 - 25
1 Jan 2016
Stevens A Wilson C Shunmugam M Ranawat V Krishnan J
Full Access

Inter- and intra-observer variation has been noted in the analysis of radiographic examinations with regard to experience of surgeons, and the monitors used for conducting the evaluations. The aim of this study is to evaluate inter/intra observer variation in the measurement of mechanical alignment from long-leg radiographs.

40 patients from the elective waiting list for TKA underwent long leg radiographs pre-operatively and 6 months post-operatively (total of 80 radiographs). The x-rays were analysed by 5 observers ranging in experience from medical student to head orthopaedic surgeon. Two observers re-analysed their results 6 months later to determine intraobserver correlation, and one observer re-measured the alignment on a different monitor. These measurements were all conducted blindly and none of the observers had access to the others’ results.

80 radiographs were analysed in total, 40 pre-op and 40 post-op. The mechanical alignment was analysed using Pearson's correlation (r = 0 no agreement, r = 1 perfect agreement) and revealed that experience as an orthopaedic surgeon has little effect on the measurement of mechanical alignment from long leg radiograph. The results for the different monitor analysis were also analysed using Pearson's correlation of long leg alignment. Monitor quality does seem to affect the correlation between alignment measurements when reviewing both intra and inter observer correlation on different computer monitors.

Surgical experience has little impact on the measurement of alignment on long leg radiographs. Of greater concern is that monitors of different resolution can affect measurement of mechanical alignment. As there might be a range of monitors in use in different institutions, and also in outpatient clinics to surgical theatres, close attention should be paid to the implications of these results.


Orthopaedic Proceedings
Vol. 84-B, Issue SUPP_II | Pages 195 - 196
1 Jul 2002
Thomas S Sforza G Levy O Copeland A
Full Access

The aim of this studdy was to examine the effect of cementless surface replacement shoulder arthroplasty (CSRA) on proximal humeral anatomy in eroded shoulder joints.

AP radiographs of 39 shoulders in 37 patients that underwent CSRA for arthritis were examined for geometry of the glenohumeral joint with correction for the magnification of our apparatus. Thirty-two were hemiarthroplasties and seven were total shoulder replacements (TSR). Average age was 70 years (range: 29–88 years). Mean clinical and radiological follow-up was 38 month (range: 24–72 months) and 16 months (range: 10–65 months) respectively. We measured reliable values on the proximal humerus and the lateral glenohumeral offset (LGHO) relative to the coracoid base. Preoperative and last follow-up Constant scores were recorded.

Based on anatomical data with respect to humeral head radius there was a mean 6mm preoperative loss in LGHO (95% CI 3.6–8.8, p< 0.01). The mean value of 53mm increased postoperatively to 59mm and was 57mm at last follow-up. The mean changes were an increase of 6mm (95% CI 4.4–8.5, p< 0.01) and then a decrease of 2mm (95% CI 0.1–5.4, p=0.04) respectively. The lever arm, measured from the greater tuberosity to the centre of instant rotation, increased a mean 5mm post-operatively (95% CI 3.8–6.4, p< 0.01) with no significant fall at last follow-up. Humeral head size and medial offset relative to shaft width increased by 13% and 30% respectively. The humeral head centre moved superiorly relative to the glenoid a mean 2mm after operation (95% CI −0.2–3.5, p=0.08) and a further 1mm at last follow-up (95% CI −0.1–3.0, p=0.07). Forward flexion and abduction improved from 66′ and 58′ preoperatively to 124′ and 112′ postoperatively, with age/ sex-adjusted Constant scores increasing by 53 (95% CI 43.0–64.4, p< 0.01) from a mean 25 preoperatively to 79 at last follow-up.

For hemiarthroplasty the LGHO increased by 9% and for TSR by 24%, with greater increases in flexion and abduction in the latter group.

The Copeland CSRA is centred on the native humeral neck for head version and offset. This preserves maximal bone stock and avoids the need for modularity which some modern stemmed prostheses use to reconcile differences between proximal shaft and humeral head anatomy. The inherent limitation is the requirement for preservation of sufficient humeral head to permit resurfacing. In this group with fairly marked degrees of joint erosion the CSRA, using autogenous bone graft and prostheses of variable width, achieved statistically and clinically significant increases in the lever arm. The improved biomechanics and soft tissue tension correlated to a good clinical outcome with no evidence of significant early subsidence.


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 460 - 460
1 Apr 2004
McAfee P Cunningham B Holtsapple G Bussard K Guyer R Blumenthal S Dmitriev A Maxwell J Isaza J Regan J
Full Access

Introduction: A prospective randomized study of artifical disc replacement vs. lumbar fusion for one-level disc pathology with 2 year minimum follow-up was completed in compliance with a U.S. FDA protocol.

Methods: A total of 15 investigational sites enrolled 375 subjects with a randomization in a 2:1 ratio. Of the 375, 205 were randomized to receive the Charité artificial disc, and 99 were randomized to receive anterior lumbar interbody fusion with BAK cages. An additional group of 71 patients received the Charité disc as “training cases” prior to beginning randomization. Clinical outcome measures included VAS, Oswestry Disability Index, and SF-36 Healthy Surveys. A total of 6,900 radiographs were digitized throughout the 24 month treatment interval. The 276 disc replacement patients were allocated into one of three groups based on radiographic technical paramenters-- Group I – Ideal, defined as Charité disc placement within 3 mm of ideal in both planes. Coronal plane = AP radiograph = midline or within 3 mm of midline. Mid-Sagittal plane = Lateral radiograph = 2mm posterior to middle of vertebral body or within 3mm of this axis. Group II —Suboptimal (not ideal) and Group III – Poor.

Results: The Charité prosthesis was significantly more effective than BAK in restoring the height of the collapsed disk space (p < 0.001). In Charité cases, the mean initial disc space height at the L5-S1 operative level was 5.2 mm +/− 1.44 (Std Dev) and increased to a mean of 13.5 mm +/− 1.18 (Std Dev). For BAK, the initial disc space height was 5.9.mm +/− 1.74 and increased to an immediate post-operative disk space height of 11.9 mm +/− 2.07. There was less subsidence with the Charite disk replacement than the BAK control at 2 years (p < 0.001). Of the 276 subject radiographs analysed with Charité disc replacement, 83% were classified as Group I, 11% as Group II, and 6% as Group III.

The mean Oswestry Disability Index scores at 2 years correlated with technical accuracy in placement of the prosthesis: Group I – 24.1; Group II – 30.3; and Group III – 36.3 (p < .05). The Mean VAS scores at 2 years correlated with technical accuracy in placement of the prosthesis: Group I – 28.3; Group II – 35.4; and Group III – 48.4 (p = 0.016). The mean flexion/extension range of motion and prosthesis function also correlated with device placement: Group I – 7.12 +/− 4.06 degrees; Group II – 7.47 +/− 4.41 degrees; and Group III – 3.15 +/− 3.51 degrees (p = 0.003).

Discussion: The surgical technical accuracy of Charité artificial disc placement correlated with clinical outcome, range of motion, and device functionality at 2 years. The Charité Lumbar Disk replacement proved to be a successful alternative to traditional lumbar fusion in every parameter. The results from this U.S. Investigational study confirm that proper placement of the Charité artificial disc improves clinical and radiological outcomes.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 30 - 30
1 Dec 2022
McGoldrick N Cochran M Biniam B Bhullar R Beaulé P Kim P Gofton W Grammatopoulos G
Full Access

Short cementless femoral stems are increasingly popular as they allow for less dissection for insertion. Use of such stems with the anterior approach (AA) may be associated with considerable per-operative fracture risk. This study's primary aim was to evaluate whether patient-specific femoral- and pelvic- morphology and surgical technique, influence per-operative fracture risk. In doing so, we aimed to describe important anatomical thresholds alerting surgeons. This is a single-center, multi-surgeon retrospective, case-control matched study. Of 1145 primary THAs with a short, cementless stem inserted via the AA, 39 periprosthetic fractures (3.4%) were identified. These were matched for factors known to increase fracture risk (age, gender, BMI, side, Dorr classification, stem offset and indication for surgery) with 78 THAs that did not sustain a fracture. Radiographic analysis was performed using validated software to measure femoral- (canal flare index [CFI], morphological cortical index [MCI], calcar-calcar ratio [CCR]) and pelvic- (Ilium-ischial ratio [IIR], ilium overhang, and ASIS to greater trochanter distance) morphologies and surgical technique (% canal fill). Multivariate and Receiver-Operator Curve (ROC) analysis was performed to identify predictors of fracture. Femoral factors that differed included CFI (3.7±0.6 vs 2.9±0.4, p3.17 and II ratio>3 (OR:29.2 95%CI: 9.5–89.9, p<0.001). Patient-specific anatomical parameters are important predictors of fracture-risk. When considering the use of short stems via the AA, careful radiographic analysis would help identify those at risk in order to consider alternative stem options


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 29 - 29
19 Aug 2024
Kayani B Konan S Tahmassebi J Giebaly D Haddad FS
Full Access

The direct superior approach (DSA) is a modification of the posterior approach (PA) that preserves the iliotibial band and short external rotators except for the piriformis or conjoined tendon during total hip arthroplasty (THA). The objective of this study was to compare postoperative pain, early functional rehabilitation, functional outcomes, implant positioning, implant migration, and complications in patients undergoing the DSA versus PA for THA. This study included 80 patients with symptomatic hip arthritis undergoing primary THA. Patients were prospectively randomised to receive either the DSA or PA for THA, surgery was undertaken using identical implant designs in both groups, and all patients received a standardized postoperative rehabilitation programme. Predefined study outcomes were recorded by blinded observers at regular intervals for two-years after THA. Radiosteriometric analysis (RSA) was used to assess implant migration. There were no statistical differences between the DSA and PA in postoperative pain scores (p=0.312), opiate analgesia consumption (p=0.067), and time to hospital discharge (p=0.416). At two years follow-up, both groups had comparable Oxford hip scores (p=0.476); Harris hip scores (p=0.293); Hip disability and osteoarthritis outcome scores (p=0.543); University of California at Los Angeles scores (p=0.609); Western Ontario and McMaster Universities Arthritis Index (p=0.833); and European Quality of Life questionnaire with 5 dimensions scores (p=0.418). Radiographic analysis revealed no difference between the two treatment groups for overall accuracy of acetabular cup positioning (p=0.687) and femoral stem alignment (p=0.564). RSA revealed no difference in femoral component migration (p=0.145) between the groups at two years follow-up. There were no differences between patients undergoing the DSA versus PA for THA with respect to postoperative pain scores, functional rehabilitation, patient-reported outcome measurements, accuracy of implant positioning, and implant migration at two years follow-up. Both treatment groups had excellent outcomes that remained comparable at all follow-up intervals


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_16 | Pages 20 - 20
17 Nov 2023
van Duren B France J Berber R Matar H James P Bloch B
Full Access

Abstract. Objective. Up to 20% of patients can remain dissatisfied following TKR. A proportion of TKRs will need early revision with aseptic loosening the most common. The ATTUNE TKR was introduced in 2011 as successor to its predicate design The PFC Sigma (DePuy Synthes, Warsaw, In). However, following reports of early failures of the tibial component there have been ongoing concerns of increased loosening rates with the ATTUNE TKR. In 2017 a redesigned tibial baseplate (S+) was introduced, which included cement pockets and an increased surface roughness to improve cement bonding. Given the concerns of early tibial loosening with the ATTUNE knee system, this study aimed to compare revision rates and those specific to aseptic loosening of the ATTUNE implant in comparison to an established predicate as well as other implant designs used in a high-volume arthroplasty centre. Methods. The Attune TKR was introduced to our unit in December 2011. Prior to this we routinely used a predicate design with an excellent long-term track record (PFC Sigma) which remains in use. In addition, other designs were available and used as per surgeon preference. Using a prospectively maintained database, we identified 10,202 patients who underwent primary cemented TKR at our institution between 01/04/2003–31/03/2022 with a minimum of 1 year follow-up (Mean 8.4years, range 1–20years): 1) 2406 with ATTUNE TKR (of which 557 were S+) 2) 4652 with PFC TKR 3) 3154 with other cemented designs. All implants were cemented using high viscosity cement. The primary outcome measures were all-cause revision, revision for aseptic loosening, and revision for tibial loosening. Kaplan-Meier survival analysis and Cox regression models were used to compare the primary outcomes between groups. Matched cohorts were selected from the ATTUNE subsets (original and S+) and PFC groups using the nearest neighbor method for radiographic analysis. Radiographs were assessed to compare the presence of radiolucent lines in the Attune S+, standard Attune, and PFC implants. Results. At a mean of 8.4 years follow-up, 308 implants underwent revision equating to 3.58 revisions per 1000 implant-years. The lowest risk of revision was noted in the ATTUNE cohort with 2.98 per 1000-implant-years where the PFC and All Other Implant groups were 3.15 and 4.4 respectively. Aseptic loosing was the most common cause for revision across all cemented implants with 76% (65/88) of involving loosening of the tibia. Survival analysis comparing the ATTUNE cohort to the PFC and All Other Cemented Implant cohorts showed no significant differences for: all-cause revision, aseptic loosening, or tibial loosening (p=0.15,0.77,0.47). Radiolucent lines were detected in 4.6%, 5.8%, and 5.0% of the ATTUNE S+, standard ATTUNE, and PFC groups respectively. These differences were not significant. Conclusion. This study represents the largest non-registry review of the original and S+ ATTUNE TKR in comparison to its predicate design as well as other cemented implants. There appears to be no significant increased revision rate for all-cause revision or aseptic loosening. Radiographic analysis also showed no significant difference in peri-implant radiolucency. It appears that concerns of early loosening may be unfounded. Declaration of Interest. (a) fully declare any financial or other potential conflict of interest


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 11 - 11
1 Apr 2022
McGoldrick NP Cochran M Biniam B Bhullar R Beaulé PE Kim PR Gofton W Grammatopoulos G
Full Access

Short cementless femoral stems are increasingly popular as they allow for less dissection for insertion. Use of such stems with the anterior approach (AA) may be associated with considerable per-operative fracture risk. This study's primary aim was to evaluate whether patient-specific femoral- and pelvic- morphology and surgical technique, influence per-operative fracture risk. In doing so, we aimed to describe important anatomical thresholds alerting surgeons. This is a single-centre, multi-surgeon retrospective, case-control matched study. Of 1145 primary THAs with a short, cementless stem inserted via the AA, 39 periprosthetic fractures (3.4%) were identified. These were matched for factors known to increase fracture risk (age, gender, BMI, side, Dorr classification, stem offset and indication for surgery) with 78 THAs that did not sustain a fracture. Radiographic analysis was performed using validated software to measure femoral- (canal flare index [CFI], morphological cortical index [MCI], calcar-calcar ratio [CCR]) and pelvic- (Ilium-ischial ratio [IIR], ilium overhang, and ASIS to greater trochanter distance) morphologies and surgical technique (% canal fill). Multivariate and Receiver-Operator Curve (ROC) analysis was performed to identify predictors of fracture. Femoral factors that differed included CFI (3.7±0.6 vs 2.9±0.4, p<0.001) and CCR (0.5±0.1 vs 0.4±0.1, p=0.006). The mean IIR was higher in fracture cases (3.3±0.6 vs 3.0±0.5, p<0.001). % Canal fill was reduced in fracture cases (82.8±7.6 vs 86.7±6.8, p=0.007). Multivariate analysis and ROC analyses revealed a threshold CFI of 3.17 was predictive of fracture (sensitivity:84.6% / specificity:75.6%). Fracture risk was 29 times higher when patients had CFI>3.17 and II ratio>3 (OR:29.2 95%CI: 9.5–89.9, p<0.001). Patient-specific anatomical parameters are important predictors of fracture-risk. When considering the use of short stems via the AA, careful radiographic analysis would help identify those at risk in order to consider alternative stem options


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_8 | Pages 64 - 64
1 May 2019
Rodriguez J
Full Access

Modifiable factors contributing to stiffness include alignment, implant size, implant position and rotation, and soft tissue tightness or laxity. Less modifiable factors include genetics as in predisposition to inflammation and fibrosis, aberrations in perception and experience of emotional pain, and preoperative range of motion. We reviewed 559 knees undergoing revision between 2007 and 2014, selecting out patients with a diagnosis of stiffness and greater than one-year follow-up. Stiffness was defined as greater than 15 degrees of flexion contracture or less than 75 degrees of flexion or less than 90 degrees of active motion and a chief complaint of limited motion and pain. Radiographic analysis used a set of matched controls with greater than 90 degrees and full extension prior to surgery and were further matched by age, gender, BMI. Flexion contracture changed from an average of 9.7 to an average of 2.3 degrees, flexion changed from an average of 81 to an average of 94 degrees, active motion changed from an average of 72 to an average of 92 degrees, pain scores improved from 44 to 72 points, and Knee Society function scores improved from an average of 49 to an average of 70 points. There were four failures for stiffness, two knees underwent additional manipulation, gaining an average of 10 degrees; and two knees were revised. Radiographic analysis demonstrated stiffness to be strongly correlated to anterior condylar offset ratio and to patellar displacement by multivariant regression analysis, suggesting that overstuffing the patellofemoral joint by anteriorization of the femoral component is associated with stiffness. Using modern revision techniques, revision for stiffness creates reliable improvements in pain, Knee Society clinical and functional scores, and motion


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 6 - 6
1 Aug 2021
Kennedy I Hrycaiczuk A Ng N Sheerins O Patil S Jones B Stark A Meek D
Full Access

Periprosthetic fractures (PPF) of the femur following total hip arthroplasty represent a significant complication with a rising incidence. The commonest subtype is Vancouver B2 type, for which revision to a long uncemented tapered fluted stem is a widely accepted management. In this study we compare this procedure to the less commonly performed cement-in-cement revision. All patients undergoing surgical intervention for a Vancouver B2 femoral PPF in a cemented stem from 2008 – 2018 were identified. We collated patient age, gender, ASA score, BMI, operative time, blood transfusion requirement, change in haemoglobin (Hb) level, length of hospital stay and last Oxford Hip Score (OHS). Radiographic analysis was performed to assess time to fracture union and leg length discrepancy. Complications and survivorship of implant and patients were recorded. 43 uncemented and 29 cement-in-cement revisions were identified. There was no difference in patient demographics between groups. A significantly shorter operative time was found in the cement-in-cement group, but there was no difference in transfusion requirement, Hb change, or length of hospital stay. OHS was comparable between groups. A non-significant increase in overall complication rates was found in the revision uncemented group, with a significantly higher dislocation rate. Time of union was comparable and there were no non-unions in the cement-in-cement group. A greater degree of stem subsidence was found in the uncemented group. There was no difference in any revision surgery required in either group. Three patients in the uncemented group died in the perioperative period, compared to none in the cement-in-cement group. With appropriate patient selection, both cement-in-cement and long uncemented tapered stem revision represent appropriate treatment options for Vancouver B2 fractures


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 39 - 39
1 Aug 2021
Rajan A Leong J Singhal R Siney P Shah N Board T
Full Access

Trabecular metal (TM) augments are designed to support an uncemented socket in revision surgery when adequate rim fit is not possible. We have used TM augments in an alternative arrangement, to contain segmental defects to facilitate impaction bone grafting (IBG) and cementation of a cemented socket. However, there is a paucity of literature supporting the use of this technique. We present one of the largest studies to date, reporting early outcomes of patients from a tertiary centre. A single-centre retrospective analytical study of prospectively collected data was performed on patients who had undergone complex acetabular reconstruction using TM augments, IBG and a cemented cup. All patients operated between 2015 and 2019 were included. We identified 105 patients with a mean age of 74yrs. The mean follow-up was 2.3 years(1–5.5yrs). Our primary outcome measure was all-cause revision of the construct. The secondary outcome measures were, Oxford hip score (OHS), radiographic evidence of cup migration/loosening and post-op complications. Eighty-four out of 105 patients belonged to Paprosky grade IIb, IIc or IIIa. Kaplan-Meier survivorship for all-cause revision was 96.36% (CI, 90.58–100.00) at 2 years with 3 failures. Two were due to early infection which required two-stage re-revision. The third was due to post-operative acetabular fracture which was then re-revised with TM augment, bone graft and large uncemented cup. Pre-op and post-op matched OHS scores were available for 60 hips(57%) with a mean improvement of 13 points. Radiographic analysis showed graft incorporation in all cases with no evidence of cup loosening. The mean vertical cup migration was 0.5mm (Range −5 to 7mm). No other complications were recorded. This study shows that reconstruction of large acetabular defects during revision THA using a combination of TM augments to contain the acetabulum along with IBG to preserve the bone stock and a cemented socket is a reliable and safe technique with low revision rates and satisfactory clinical and radiographic results. Long term studies are needed to assess the possibility of preservation and regeneration of bone stock


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 48 - 48
1 May 2017
Cooper J McKinnon J Walsh W Oliver R Rawlinson J Cristou C
Full Access

Background. Calcium sulfate and phosphate have a long clinical history of use as bone-void fillers (BVF) with established biocompatibility and resorption profiles. It has been widely reported that the addition of ‘impurity’ elements such as Silicon, Strontium and Zinc to calcium phosphate is advantageous, resulting in an improved bone healing response. Methods. This study examined the in vivo response of two formulations of calcium sulfate, as 3mm diameter hemispherical beads, in critical sized defects created in cancellous bone of distal femur and proximal tibia (10mm diameter × 13mm depth) in adult sheep; beads prepared from recrystallised pharmaceutical grade calcium sulfate (RPCS, Stimulan, Biocomposites Ltd, UK) and a lower purity medical grade material containing 1% strontium (SrCS). The animals were sacrificed at 3, 6 and 12 weeks post implantation and the surgical sites analysed using microCT and decalcified histology. Results. Radiographic analysis showed a slower resorption for SrCS compared to RPCS. Radiographic analysis for both materials confirmed little residual beads at three weeks post implantation. Radiographs at sacrifice confirmed no adverse reactions at any sites at 3, 6 and 12 weeks. Radiographic data alone was not adequate to determine the status of the bone formation and the implant resorption at the implant site. Histological analysis confirmed little or no adverse tissue reactions to either material. However, RPCS outperformed the modified material in terms of new bone formation at all time points post implantation. At 3 weeks histology for RPCS confirmed that residual beads were still visible with active new bone growth appearing to penetrate centripetally into the defect with some resorption of the implant material. By 6 weeks significant new bone was present throughout the defect. In comparison, absorption of the modified material was slower, and penetration of new bone into the defect was less progressed. Conclusions. The rapid bone regenerative ability of the recrystallised pharmaceutical grade calcium sulfate was demonstrated. The presence of 1% Strontium impurity acted to delay implant absorption and bone healing in this model


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 70 - 70
1 Oct 2020
Staats K Sosa BR Kuyl E Niu Y Suhardi VJ Turajane K Windhager R Greenblatt MB Ivashkiv L Bostrom MP Yang X
Full Access

Introduction. Initial post-operative implant instability leads to impaired osseointegration, one of the most common reasons for aseptic loosening and revision surgery. In this study, we developed a novel murine model of implant instability and demonstrated the anabolic effect of immediate and delayed intermittent Parathyroid Hormone (iPTH) treatment in the setting of instability-induced osseointegration failure. Methods. 3D-printed titanium implants were inserted in an oversized drill-hole in the tibia of C57Bl/6 mice (n=54). After implantation, the mice were randomly divided in 3 treatment groups (control: PBS-vehicle; iPTH; delayed iPTH). Radiographic analysis was performed to confirm signs of implant loosening. Peri-implant tissue formation was assessed through histology. Osseointegration was assessed through µCT and biomechanical pullout testing. Results. Immediate iPTH treatment reduced radiolucencies and induced a distinct pedestal sign distal to the implant stem (white arrow Fig 1A). The PBS treated mice had fibrous tissue implant encapsulation, whereas new mineralized tissue and no fibrous tissue was observed with immediate iPTH treatment (Fig 1E). Delayed iPTH treatment was also able to exhibit significant peri-implant bone mineralization, osteoblasts, angiogenesis, and a reduction of fibrous tissue (Fig 2A-B). iPTH treatment increased the force required to pull out the implant significantly from 8.41 ± 8.15N in the PBS group to 21.49 ± 10.45N and 23.68 ± 8.99N, in the immediate and delayed iPTH treatment groups, respectively (Fig 2D). PBS vehicle resulted in a bone volume/trabecular volume (BV/TV) of 0.23 ± 0.03, while immediate and delayed iPTH treatment increased BV/TV significantly to 0.46 ± 0.07 and 0.34 ± 0.10, respectively (Fig 2E). Conclusion. Immediate iPTH treatment prevents peri-implant fibrous tissue formation and enhances peri-implant bone formation in our murine model of mechanical instability. Delayed iPTH treatment was able to resolve the peri-implant fibrous tissue and stimulate bone formation. This study potentially addresses a leading cause of aseptic loosening by demonstrating that initial implant instability can be rescued by iPTH even with delayed start of treatment. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 53 - 53
1 Oct 2020
Larson D Rosenberg J Lawlor M Garvin KL Hartman C Lyden E Konigsberg B
Full Access

Introduction. Revision total knee arthroplasty (TKA) is becoming increasingly common in the United States as the population ages and larger numbers of primary TKA are performed in younger individuals. Cemented or uncemented tibial stems are frequently used in revision cases. Decreased clinical outcomes and patient satisfaction have been described for revision TKA. This study aims to determine if the presence of overall pain and tibial pain at the end of the stem differs between cemented and uncemented tibial stems in revision TKA. Methods. This was a retrospective cohort study comparing patients who underwent revision TKA utilizing cemented or uncemented tibial stems in a 15-year period at a single institution with at least two-year follow-up. Exclusion criteria included age under 18, isolated revisions of the femoral component or polyethylene exchanges, lack of preoperative or postoperative imaging, insufficient operative or implant records available for electronic chart review, revision procedures performed at outside facilities, patients who were deceased at the time of survey administration, refusal to participate in the study, and failure to return the mailed survey or respond to a telephone follow-up questionnaire. Radiographic analysis included calculation of the percentage of the tibial canal filled with the implant, as well as measurement of the diameter of the tibial stem. Radiographs were also reviewed for evidence of cavitary defects, pedestal formation, radiolucent lines, and periprosthetic fractures. Mailed surveys addressing overall pain, tibial pain, and satisfaction were analyzed using Fisher's exact test and the independent sample t-test. Logistic regression was used to adjust for age, gender, and preoperative bone loss. Results. A total of 110 patients were included (63 cemented and 47 uncemented stems). No statistically significant differences in stem length, operative side, or indications for revision were found. The uncemented group had a significantly higher percent canal fill (p < 0.0001). Tibial pain at the end of the stem was present in 25.3% of cemented stems and 25.5% of uncemented stems (p = 1.00). There was a trend towards more overall pain in the uncemented cohort, but this did not reach statistical significance. Only 74.6% of cemented patients and 78.7% of uncemented patients were satisfied following revision TKA (p = 0.66). Conclusion. The data supports our hypothesis that there are no differences in end-of-stem pain or overall pain between cemented and uncemented tibial stems in revision TKA. High rates of dissatisfaction were noted in both cohorts postoperatively, consistent with previous literature. Patient factors likely play a large role in the presence of postoperative pain. These factors should be further evaluated in future studies in an effort to reduce pain and improve patient satisfaction


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 55 - 55
1 Oct 2020
Mahan C Blackburn B Anderson LA Peters CL Pelt CE Gililland JM
Full Access

Introduction. Porous metaphyseal cones are increasingly used for fixation in revision total knee arthroplasty (RTKA). Both cemented shorter length stems and longer diaphyseal engaging stems are currently utilized with metaphyseal cones with no clear evidence of superiority. The purpose of this study was to evaluate our experience with 3D printed titanium metaphyseal cones with both short cemented and longer cementless stems from a clinical and radiographic perspective. Methods. In total 136 3D printed titanium metaphyseal cones were implanted. The mean patient age was 63 and 48% were female. The mean BMI was 33 and the mean ASA class was 2.5. There were 42 femoral cones in which 28 cemented and 14 cementless stems were utilized. There were 94 tibial cones in which 67 cemented and 27 cementless stems were utilized. The choice for stem fixation was surgeon dependent and in general cones were utilized for AORI type 2 and 3 bone defects on the femur and tibia. The most common fixation scenario was short cemented stems on both the femur and tibia followed by cemented stem fixation on the tibia and cementless fixation on the femur. Clinical data such as revision, complication, and PRO was collected at last follow-up (minimum follow-up 1 year). Radiographic analysis included cone bony ingrowth and coronal and sagittal alignment on long-standing radiographs. Descriptive statistics were used to compare demographics between patients who had malalignment (HKA beyond +/− 3 degrees and flexion/extension beyond +/− 3 degrees). Adjusted logistic regression models were run to assess malalignment risk by stem type. Results. Patient reported outcomes demonstrated modest improvements with Pre-op KOOS improving from 44 pre-op to 59 post -op and PF-CAT improving from 33 to 37 post-op. PROMIS pain scores decreased significantly from 54 to 44 post-op. 36% of patients had malalignment in either the coronal or sagittal plane. Patients with malalignment were more likely to be female (66.7% vs 40.4%, p-value=0.02). After adjusting for age, sex and BMI, there was a significantly increased risk for coronal plane malalignment when both the femur and tibia had cementless compared to cemented stems (odds ratio=5.54, 95%CI=1.15, 26.80). There was no significantly increased risk when comparing patients with mixed stems to patients with cemented stems. Sagittal plane malalignment was more common with short cemented stems although both coronal plane and sagittal plane malalignment with either stem type was not associated with inferior clinical outcome. Overall cone survivorship was excellent with only two cones removed for infection. Conclusion. Metaphyseal titanium cones provide reliable fixation in revision TKA. However, PROs in this complex patient population show only modest improvement consistent with other variables such as co-morbidities and poor baseline physical function. Small cone inner diameter may adversely influence cementless stem position leading to coronal plane malalignment. Short cemented stems are subject to greater sagittal plane malalignment with no apparent influence on clinical outcome


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 142 - 142
1 Feb 2020
Nizam I Batra A Gogos S
Full Access

INTRODUCTION. The Woodpecker pneumatic broaching system facilitates femoral preparation to achieve optimal primary fixation of the stem in direct anterior hip replacement using a standard operating table. The high-frequency axial impulses of the device reduce excess bone tension, intraoperative femoral fractures and overall operating time. The Woodpecker device provides uniformity and enhanced control while broaching, optimizing cortical contact between the femur and implant and thereby maximizing prosthetic axial stability and longevity. This study aims to describe a single surgeon's experience using the Woodpecker pneumatic broaching system in 649 cases of direct anterior approach (DAA) total hip arthroplasties to determine the device's safety and efficacy. METHODOLOGY. All consecutive patients undergoing elective anterior bikini total hip arthroplasties (THA) performed by a single surgeon between July 2013 and June 2018 were included. Patients undergoing a THA with the use of the Woodpecker device through a different surgical approach, revision THA or arthroplasties for a fractured neck of femur were excluded (n=219). The pneumatic device was used for broaching the femoral canal in all cases. Pre-operative and post-operative Harris Hip Scores (HHS) and post-operative radiographs were analyzed to identify femoral fractures and femoral component positioning at 6 weeks, 6 months and 12 months post-operative. Any intra-operative or post-operative surgical complications and component survivorship until most recent follow up were recorded in the clinical notes. RESULTS. A total of 649 patients (L THA=317, R THA=328 and bilateral=2) with a mean age of 69 (range 46–91yrs) and mean BMI of 28.3 (range = 18.4–44.0) underwent a DAA THA using a Woodpecker device were included in the study. Of these patients, 521 (80%) underwent uncemented and 128 (20%) underwent cemented femoral components. The time taken to broach the femur using Woodpecker broaching this system averaged 2.8 minutes (1.4 to 7.5 minutes) in both cemented and uncemented cases. In 91% of cases the templated broach size was achieved with the remaining 9% within +/− 1 size of the planned template. Radiographic analysis revealed 67.3% of the stems placed in 0–1.82 degrees of varus and 32.7% placed in 0–1.4 degrees of valgus. Average HHS were 24.4 pre-operatively, with drastic improvements shown at 6 weeks (80.95), 6 months (91.91) and 12 months (94.18) after surgery. Intraoperative femoral fractures occurred in three patients (0.4%) during trial reduction, a further three patients had periprosthetic post-operative fractures (0.4%) from falls, two patients had stem subsidence (0.3%) and a further two patients had wound infections (0.3%). At the most recent follow up, the survivorship of the acetabular component was 99.7% and the femoral component was 99.1%, with mean follow up of 2.9 years (0.5 to 5 years). No intraoperative or post-operative complications could be directly attributed to the Woodpecker broaching system. CONCLUSION. The pneumatic Woodpecker device is a safe and effective alternative tool in minimally invasive direct anterior hip replacement surgery for femoral broaching performed on a standard operating table. The skill and experience of the surgeon must be taken into consideration when utilizing new surgical devices


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims. Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. Materials and Methods. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum. Results. There were 151 patients who met the classification of high-risk according to the inclusion criteria and received DM THA during the study period. Mean age was 82 years old (73 to 95) and 114 patients (77.5%) were female. Mean follow-up was 3.6 years (1.9 to 6.1), with five patients lost to follow-up and one patient who died (for a reason unrelated to the index procedure). One patient (0.66%) sustained an intraprosthetic dislocation; there were no other dislocations. Conclusion. At mid-term follow-up, the use of a DM bearing for primary THA in patients at high risk of dislocation provided a stable reconstruction option with excellent radiographic results. Longer follow-up is needed to confirm the durability of these reconstructions


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1292 - 1297
1 Sep 2005
Lietman SA Inoue N Rafiee B Deitz LW Chao EYS

We used a canine intercalary bone defect model to determine the effects of recombinant human osteogenic protein 1 (rhOP-1) on allograft incorporation. The allograft was treated with an implant made up of rhOP-1 and type I collagen or with type I collagen alone. Radiographic analysis showed an increased volume of periosteal callus in both test groups compared with the control group at weeks 4, 6, 8 and 10. Mechanical testing after 12 weeks revealed increased maximal torque and stiffness in the rhOP-1 treated groups compared with the control group. These results indicate a benefit from the use of an rhOP-1 implant in the healing of bone allografts. The effect was independent of the position of the implant. There may be a beneficial clinical application for this treatment


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 9 - 9
1 Apr 2018
Kweon S
Full Access

Purpose. To evaluate the radiographic long-term result of femoral revision hip arthroplasty using impacted cancellous allograft combined with cemented, collarless, polished and tapered stem. Materials and methods. Among 28 patients with impacted cancellous allograft with a cemented stem, 28 hips from 26 consecutive patients were analyzed retrospectively. The average patient age was 59 years. The follow-up period ranged 9 years 6 months to 14 years 5 months (mean, 12, 5 years). Radiographic parameters analyzed in this study included subsidence of the stem in the cement, subsidence of the cement mantle in the femur, bone remodeling of the femur, radiolucent line, and osteolysis. Results. Radiographic analysis showed very stable stem initially. 27 stems showed minimal subsidence (less than 5 mm) and 1 stem showed moderate subsidence (about 10 mm) in the cement. But there was no mechanical failure and subsidence at the composit-femur interface. Evidence of cortical and trabecular remodeling were observed in all cases. No radiolucent line or osteolysis were found in the follow-up period. There were 4 proximal femoral cracks and 1 distal femoral splitting during operation. Conclusion. The result of cemented stem revision with the use of impacted cancellous allograft was good long-terand femoral bone stock deficiency may be reconstructed successfully


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 407 - 407
1 Sep 2009
Ollivere B Duckett S August A Porteous M
Full Access

Introduction: The Birmingham Hip resurfacing was commercially introduced in 1997 and early originating centre series show good functional outcomes. Concerns have been raised over the longevity, functional outcomes and metal ion release from the implants. There are no series of medium term results from an independent district general hospital reported in the literature. We present a prospective series of 100 patients with mean 5 year follow up from a district general hospital. Each patient underwent yearly clinical, hip scoring, and regular radiographic evaluation. Radiographic analysis was undertaken using Harris’, Hodgkinson’s and Amstutz’s criteria, evaluation of component position, neck narrowing and migration using diagnostic PACS workstations with standardised scaled images. Results: Between June 2001 and Feb 2004 100 Birmingham Hip replacements were performed by two consultant surgeons (MP, AA). Mean follow up is 61.2 months (range 38–76 months). Harris hip scores (fig 1) improved from 46 pre-operatively to 90 post-operatively and no significant change over the next five years. There were no revisions in this period. Obese patients (BMI> 30) had a significantly (p< 0.03) lower post operative functional score as compared normal patients. No other factors were significant for outcome. Component position was satisfactory in 93% of cases. Radiographic analysis showed no cups, or stems were definitely loose. Radiolucent lines were present in 8/100 acetabular and 3/100 femoral components, osteolytic lesions were seen in three acetabular components. Mean neck narrowing was 9mm. No patients show any radiographic evidence of avascular necrosis. Conclusion This independent series shows the results of the Birmingham hip resurfacing are reproducible and comparable to those reported in the originating centre. The Birmingham hip resurfacing gives excellent clinical results, and there is no early evidence of radiographic failure. The high rate of neck narrowing gives us cause for concern and we would recommend regular radiographic follow up


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 86 - 86
1 Apr 2019
Al-Zibari M Everett SJ Afzal I Overschelde PV Skinner J Scott G Kader DF Field RE
Full Access

Background. In the late 1980's Michael Freeman conceived the idea that knee replacement would most closely replicate the natural knee joint, if the medial Tibio-Femoral articulation was configured as a “ball-in-socket”. Over the last three decades, medial rotation and medial pivot designs have proved successful in clinical use. Freeman's final iteration of the medial ball-in-socket concept was the Medial Sphere knee. We report the three-year survivorship, clinical outcomes, patient reported outcome measures (PROMs) and radiographic analysis of this implant in a multi-centre, multi-surgeon, prospective observational study. Methods. Patients awaiting total knee replacement were recruited by four centres. They had no medical contraindication to surgery, were able to provide informed consent and were available for follow-up. Primary outcome was implant survival at six months, one, two, three and five years. Secondary outcomes were patient reported outcome measures: Oxford Knee Score (OKS), Euroqol (EQ-5D), International Knee Society Score (IKSS), IKSS Functional score and Health State score, complications and radiographic outcomes. Radiographic analysis was undertaken using the TraumaCad software and data analysis was undertaken using SPSS. Results. To date, 328 female and 202 male patients with a mean age 66.9 years and mean body mass index 30.0 were recruited. Three year Kaplan-Meier survivorship analysis of cumulative failure showed an implant survival of 99.46% (95% confidence interval 100 – 96.74), when deaths and withdrawals were treated as censored data. Twelve patients withdrew (2.26%), seven died (1.32%) and two knees were revised (0.38%). The mean EQ5D, Health State Scores, OKS, IKSS & IKSS Function scores at three years improved significantly from pre- operative scores (Health State Score: 9.91 (65.59 pre-op to 75.50); OKS: 18.82 (19.90 pre-op to 38.72); IKSS: 39.87 (44.39 pre-op to 92.09); IKSS Function Score: 35.03 (49.42 pre-op to 84.45). The mean improvement of EQ5D at three years was: 0.34 (0.48 pre-op to 0.82). Discussion. Survival of the GMK Sphere to three years in this study was over 99%. Risk of revision compares favourably with UK National Joint Registry (NJR) data. The improvements that are seen in patient reported outcome measures reflect an enhancement in patient function and quality of life. Conclusion. At three years follow-up, the implant demonstrates satisfactory survival and outcomes. Patient matching and evaluation of more cases, at more time points will allow outcome comparison with other implant options


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 22 - 22
1 Apr 2019
Massari L Bistolfi A Grillo PP Causero A
Full Access

Introduction. Trabecular Titanium is a biomaterial characterized by a regular three-dimensional hexagonal cell structure imitating trabecular bone morphology. Components are built via Electron Beam Melting technology in aone- step additive manufacturing process. This biomaterial combines the proven mechanical properties of Titanium with the elastic modulus provided by its cellular solid structure (Regis 2015 MRS Bulletin). Several in vitro studies reported promising outcomes on its osteoinductive and osteoconductive properties: Trabecular Titanium showed to significantly affect osteoblast attachment and proliferation while inhibiting osteoclastogenesis (Gastaldi 2010 J Biomed Mater Res A, Sollazzo 2011 ISRN Mater Sci); human adipose stem cells were able to adhere, proliferate and differentiate into an osteoblast-like phenotype in absence of osteogenic factors (Benazzo 2014 J Biomed Mater Res A). Furthermore, in vivo histological and histomorphometric analysis in a sheep model indicated that it provided bone in-growth in cancellous (+68%) and cortical bone (+87%) (Devine 2012 JBJS). A multicentre prospective study was performed to assess mid-term outcomes of acetabular cups in Trabecular Titanium after Total Hip Arthroplasty (THA). Methods. 89 patients (91 hips) underwent primary cementless THA. There were 46 (52%) men and 43 (48%) women, with a median (IQR) age and BMI of 67 (57–70) years and 26 (24–29) kg/m2, respectively. Diagnosis was mostly primary osteoarthritis in 80 (88%) cases. Radiographic and clinical evaluations (Harris Hip Score [HHS], SF-36) were performed preoperatively and at 7 days, 3, 6, 12, 24 and 60 months. Bone Mineral Density (BMD) was determined by dual-emission X-ray absorptiometry (DEXA) according to DeLee &Charnley 3 Regions of Interest (ROI) postoperatively at the same time-points using as baseline the measureat 1 week. Statistical analysis was carried out using Wilcoxon test. Results. Median (IQR) HHS and SF-36 improved significantly from 48 (39–61) and 49 (37–62) preoperatively to 99 (96–100) and 76 (60–85) at 60 mo. (p≤0.0001). Radiographic analysis showed evident signs of bone remodelling and biological fixation, with presence of superolateral and inferomedial bone buttress, and radial trabeculae in ROI I/II. All cups resulted radiographically stable without any radiolucent lines. The macro-porous structure of this biomaterial generates a high coefficient of friction (Marin 2012 Hip Int), promoting a firm mechanical interlocking at the implant-bone interface which could be already observed in the operating room. BMD initially declined from baseline at 7 days to 6 months. Then, BMD slightly increased or stabilized in all ROIs up to 24 months, while showing evidence of partial decline over time with increasing patient' age at 60 months, although without any clinical significance in terms of patients health status or implant stability. Statistical significant correlations in terms of bone remodeling were observed between groups of patients on the basis of gender and age (p≤0.05). No revision or implant failure was reported. Conclusions. All patients reported significant improvements in quality of life, pain relief and functional recovery. Radiographic evaluation confirmed good implant stability at 60 months. These outcomes corroborate the evidence reported on these cups by orthopaedic registries and literature (Perticarini 2015 BMC Musculoskelet Disord; Bistolfi 2014 Min Ortop)


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 89 - 89
1 Feb 2017
Levy J Kurowicki J
Full Access

Background. Peri-prosthetic humerus fractures are relatively uncommon occurrences that can be difficult to manage non-operatively. Locking plate technology has enhanced the surgical management of these fractures. We describe an osteosynthesis technique utilizing a locking plate with eccentrically placed screw holes to place “skive screws” in the proximal end of the plate to achieve fixation around the stem of the implant. Methods. A retrospective review of prospectively collected data was performed for a consecutive series of patients treated with this skive screw technique from May 2011 to September 2014. Seven patients presented with postoperative type B peri-prosthetic humerus fractures. Average follow-up was 24 months. Radiographic analysis was performed on most recent postoperative imaging. Clinical outcomes were assessed using VAS pain, ASES total score, ASES functional score, SST, SANE, range of motion and strength. Results. At an average of follow-up of 24 months, all patients demonstrated fracture healing. Functional outcomes were limited with only two patients achieving forward elevation above 90 degrees and average ASES Function score was 27.5. Pain relief was nearly uniform with an average VAS Pain score of 0.5 (Figure 1). Conclusions. Peri-prosthetic humeral shaft fractures can be successfully treated with hybrid fixation technique using a locking plate with eccentric holes that facilitate placement of proximal “skive screws”. Using this technique, a 100% union rate was observed with excellent pain relief


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_17 | Pages 21 - 21
1 Nov 2017
Surendran S Patinharayil G Raveendran M
Full Access

It is a well-known fact that total knee arthroplasty is a soft tissue operation. Soft tissue balancing is the key to success in total knee arthroplasty. It is paramount importance to preserve the maximal amount of bone on both the femur and tibial side. In Indian scenario, majority of the patients present relatively late with varus or valgus deformity. Adding to this problem is poor bone quality due to osteoporosis. Our technique of Posterior cruciate ligament (PCL) retaining TKA with tibial end plate resection facilitates soft tissue balancing, preserves PCL and maximizes bone preservation on both tibial and femoral side achieving good results in minimum seven year follow up. We retrospectively analyzed seven year outcomes of 120 knees (110 patients), mean age was 65 years (range 55 to 75 years), who received contemporary cruciate-retaining prostheses with tibial end plate resection technique. The diagnosis was osteoarthritis in 96%, Rheumatoid arthritis in 2% and posttraumatic arthritis in 2% cases. There were more number of flexible varus knees as compared to flexible valgus knees. All the patients were followed up for minimum of 84 months with average follow up of 96 months. They were followed up at 3mths, 6mths, 1,3,5,7,9 and 10 years. The functional assessment was done using knee society knee and function scores. Radiographic analysis was done to rule out subsidence and aseptic loosening. The statistical significance was assessed using chi square test. Survival analysis was done using the Kaplan Meier analysis with revision taken as the endpoint. The average ROM was 100 degrees preoperatively and 120 degrees at last follow-up. The average knee society knee score improved from 45 points preoperatively to 90 points at last follow-up. The average knee society functional score improved from 48 points preoperatively to 84 points at last follow-up (p<0.05). Radiolucency was observed in 20 knees but all except four were non-progressive lesions smaller than 2 mm. None of the implants were revised for subsidence or aseptic loosening of tibial component. The technique of PCL retaining total knee arthroplasty with tibial end plate resection in arthritic knees with flexible varus or valgus deformity yields good functional outcome in medium term follow up with relatively low incidence of subsidence of the tibial implant. This technique appears promising for total knee arthroplasty in osteoporotic bones where retaining the strong subchondral bone increases the longevity of the implant


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 131 - 131
1 May 2016
Kweon S Jeong K
Full Access

Purpose. This studyevaluated the results of the acetabular medial wall osteotomy to reconstruct the acetabulum in dysplastic hip during total hip athroplasty. Materials and Methods. A total of 30 hips of 30 patients who underwent THA between March 1999 and October 2002 were clinically and radiogically evaluated. The average age at the time of operation was 46.5 years (range: 17 to 73 years), and the mean follow-up period was 5 years (range: 5.3 to 8.7 years). 26 cases, a cementless hemispherical acetabular cup and 4 cases, reinforced ring were inserted in the true acetabulum. Only 2 hips needed structural bone graft. Results. The average Harris hip score improved from 56.3 points preoperatively to 93.2 points at the last follow up. Radiographic analysis revealed no aseptic loosening or radiolucent line, and showed stable bony fixation at the true acetabulum. The mean thickness of the medial acetabular wall postoperative was 20.5 mm. Bone union of the medial wall observed at a mean of four months post-operatively. Conclusion. The acetabular medial wall osteotomy can provide the integrity of acetabular medial wall while achieving enhanced acetabular coverage and more normal hip biomechanics