There remains concern with the use of constrained liners (CL) implanted at the time of acetabular cup revision in revision total hip replacement (rTHA). The aim of this study was to determine the implant survival in rTHA when a CL was implanted at the same time as acetabular cup revision. We reviewed our institutional database to identify all consecutive rTHAs where a CL was implanted simultaneously at the time acetabular cup revision from 2001 to 2021. One-hundred and seventy-four revisions (173 patients) were included in the study. Mean follow-up of 8.7 years (range two – 21.7). The most common indications for rTHA were instability (35%), second-stage periprosthetic joint infection (26.4%), and aseptic loosening (17.2%). Kaplan Meier Analysis was used to determine survival with all-cause re-revision and revision for cup aseptic loosening (fixation failure) as the endpoints. A total of 32 (18.3%) patients underwent re-revision at a mean time of 2.9 years (range 0.1 – 14.1). The most common reasons for re-revision were instability (14), periprosthetic joint infection (seven), and loosening of the femoral component (four). Three (1.7%) required re-revision due to aseptic loosening of the acetabular component (fixation failure) at a mean of two years (0.1 – 5.1). Acetabular component survival free from re-revision due to aseptic loosening was 98.9% (95% CI 97.3 – 100) at five-years and 98.1% (95% CI 95.8 – 100) at 10-years. There were no acetabular component fixation failures in modern highly porous shells. CLs implanted at the time acetabular cup revision in rTHA have a 98.1% 10-year survival free from acetabular cup aseptic loosening (fixation failure). There were no cup fixation failures in modern highly porous shells. Thus, when necessary, implanting a CL during revision of an acetabular component with stable screw fixation is safe with an extremely low risk of cup fixation failure.
There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification.Aims
Methods
Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size.Aims
Methods
Simultaneous bilateral total knee arthroplasty (TKA) has been used due to its financial advantages, overall resource usage, and convenience for the patient. The training model where a trainee performs the first TKA, followed by the trainer surgeon performing the second TKA, is a unique model to our institution. This study aims to analyze the functional and clinical outcomes of bilateral simultaneous TKA when performed by a trainee or a supervising surgeon, and also to assess these outcomes based on which side was done by the trainee or by the surgeon. This was a retrospective cohort study of all simultaneous bilateral TKAs performed by a single surgeon in an academic institution between May 2003 and November 2017. Exclusion criteria were the use of partial knee arthroplasty procedures, staged bilateral procedures, and procedures not performed by the senior author on one side and the trainee on another. Primary clinical outcomes of interest included revision and re-revision. Primary functional outcomes included the Oxford Knee Score (OKS) and patient satisfaction scores.Aims
Methods
Anterior cruciate ligament (ACL) rupture commonly leads to post-traumatic osteoarthritis, regardless of surgical reconstruction. This study uses standing MRI to investigate changes in contact area, contact centroid location, and tibiofemoral alignment between ACL-injured knees and healthy controls, to examine the effect of ACL reconstruction on these parameters. An upright, open MRI was used to directly measure tibiofemoral contact area, centroid location, and alignment in 18 individuals with unilateral ACL rupture within the last five years. Eight participants had been treated nonoperatively and ten had ACL reconstruction performed within one year of injury. All participants were high-functioning and had returned to sport or recreational activities. Healthy contralateral knees served as controls. Participants were imaged in a standing posture with knees fully extended.Aims
Methods
Our objective is to describe early and midterm results with the use of a new knee prosthesis as an articulating spacer in planned two-stage management for infected total knee arthroplasty. As a second objective, we compared outcomes between the group with a retained first stage and those with a completed 2-stage revision. Forty-seven patients (48 knees) from January 2012 and November 2017 underwent a 2-stage exchange with an articulating spacer with new implants was used for a chronic knee periprosthetic joint infection with a mean follow-up of 3.7 years (2–6.5 years). The most frequently identified infecting organism was MSSA (31%), MRSA (21%) or MRSE (20%). At the first stage, a new PS femoral component and a new all-polyethylene posterior stabilized (PS) tibial component or a standard PS tibial liner were cemented with antibiotic-cement, typically 3.6 gm tobramycin and vancomycin 1.5 gm. IV antibiotics for six weeks were administered. The planned reimplantation was at 3 months, but ninetteen spacers (14 all poly tibias and 5 tibial liner) were retained for over 12 months. Postoperative assessment included knee range of motion (ROM), quality of life (QOL) scores (SF-12, WOMAC, KOOS, Oxford, and UCLA scores), and a satisfaction scale from 0–100%.Introduction
Methods
Fully constrained liners are used to treat recurrent dislocations or patients at high risk after total hip replacements. However, they can cause significant morbidities including recurrent dislocations, infections, aseptic loosening and fractures. We examine long term results of 111 patients with tripolar constrained components to assess their redislocation and failure rate. The purpose of this study was to assess survivorship, complications and functional outcomes at a minimum 10 years after the constrained tripolar liners used in our institute. We retrospectively identified 111 patients who had 113 revision tripolar constrained liners between 1998 and 2008. Eighty-nine were revised due to recurrent dislocations, 11 for pseudotumor with dysfunctional abductors, and 13 for periprosthetic infection with loss of soft tissue stabilizers. All patients had revision hip arthroplasty before the constrained liner was used: 13 after the first revision, 17 after the second, 38 after the third, and 45 had more than 3 revisions. We extracted demographics, implant data, rate of dislocations and incidence of other complications. Kaplan Meier curves were used to assess dislocation and failure for any reason. WOMAC was used to assess quality of life. At 10 years, the survival free of dislocation was 95.6% (95%CI 90- 98), and at 20 years to 90.6% (95% CI 81- 95.5). Eight patients (7.1%) had dislocations of their constrained liners: 1 patient had simultaneous periprosthetic infection identified at the time of open reduction, and 1 patient sustained stem fracture 3 months prior to the liner dislocation. At 10 years, the survival to any further surgery was 89.4% (95% CI 82–93.8), and at 20 years, this was 82.5 (95% CI 71.9–89.3). Five patients (4.4%) had deep infection: 4 of these had excision arthroplasty due to failure to control infection, while 1 patient was treated successfully with debridement, exchange of mobile components and intravenous antibiotics. Two patients (1.8%) had dissociated rings that required change of liner, ring and head. Two patients (1.8%) had periprosthetic femoral fractures that were treated by revision stems and exchange of constrained liners. The mean WOMAC functional and pain scores were 66.2 and 75.9 of 100, respectively. Constrained tripolar liners in our institute provided favourable results in the long term for recurrent dislocation hip arthroplasty with dysfunctional hip stabilizers. Infection in these patients can prove to be difficult to treat due to their poor soft tissue conditions from repeated surgeries. Comparing long terms results from other types of constrained liners is essential to evaluate these salvage liners.
A large proportion of wait times for primary total knee (TKA) and hip (THA) arthroplasty is the time from primary care referral to surgical consultation. To our knowledge, no study has investigated whether a referral Oxford Knee or Hip Score (OKHS) could be used to triage non-surgical referrals appropriately. The primary purpose of the current study was to determine if a referral OKHS has the predictive ability to discriminate when a knee or hip referral will be deemed conservative as compared to surgical by the surgeon during their first consultation, and to identify an OKHS cut-off point that accurately predicts when a primary TKA or THA referral will be deemed conservative. We retrospectively reviewed all consecutive primary TKA and THA consultations from a single surgeon's tertiary, high volume practice over a three-year period. Patients with a pre-consultation OKHS, BMI < 4 1, and no absolute contraindication to TJR were included. Consultation knees/hips were categorized into two groups based on surgeon's decision, those that were offered TJR during their first consolation (surgical) versus those that were not (conservative). Baseline demographic data and OKHS were abstracted. Variables of interest were compared between cohorts using the exact chi-square test and Wilcoxon rank-sum test. Spearman's rank correlation coefficients were used to measure association between pre-consult OKHS and the surgeon's decision. A receiver operator characteristic (ROC) curve analysis was used to calculate the area under the curve (AUC) and to identify a cut-off point for the pre-operative OKHS that identified whether or not a referral was deemed conservative. TKA and THA referrals were analyzed separately. The study included 1,436 knees (1,016 patients) with a median OKS of 25 (IQR 19–32) and 478 hips (388 patients) with a median OHS of 22 (IQR 16–29). Median pre-consultation OKHS demonstrated clinically and statistically significant differences between the surgical versus conservative cohorts (p 32 (sensitivity=0.997, NPV=0.992) and for hips is OHS >34 (sensitivity=0.997 NPV=0.978). ROC analysis identified severable potential lower OKHS thresholds, depending on weight of prioritization of sensitivity, specificity, and NPV. Referral OKS and OHS demonstrate good ability to discriminate when a primary TKA or THA referral will be deemed non-surgical versus surgical at their first consultation in a single surgeon's practice. Multiple potential effective OKHS thresholds can be applied as a tool to decrease wait times for primary joint arthroplasty. However, a cost analysis would aid in identifying the optimal cut-off score, and these findings need to be externally validated before they can be broadly applied.
High complication rates and poor outcomes have been widely reported in patients undergoing revision of large head metal-on-metal arthroplasty. A previous study from our centre showed high rates of dislocation, nerve injury, early cup loosening and pseudotumor recurrence. After noting these issues, we implemented the following changes in surgical protocol in all large head MOM revisions: One: Use of highly porous shells in all cases. Two: Use of largest femoral head possible. Three: Low threshold for use of dual mobility and constrained liners when abductors affected or absent posterior capsule. Four: Use of ceramic head with titanium sleeve in all cases. Five: Partial resection of pseudotumor adjacent to sciatic and femoral nerves. The purpose of the present study is to compare the new surgical protocol above to our previously reported early complications in this group of patients We specifically looked at (1) complications including reoperations, (2) radiologic outcomes, and (3) functional outcomes. Complication rates after (Group 1), and before (Group 2) modified surgical protocol were compared using Chi-square test, assuming statistical significance p < 0 .05. Major complications occurred in 4 (8.3%) of 48 patients who had modified surgical technique, compared to 12 (38%) of 32 revisions prior to modification (p < 0 .05). Two hips of 48 (4.17%) endured dislocations in Group 1, compared to 9 of 32 (28%) in Group 2 (p < 0 .05). Four patients of 48 had repeat revision in Group 1: 2 for recurrence of pseudotumor, 1 for dislocation, and 1 for infection, compared to 6 patients who had 7 repeat revisions of 32 patients in Group 2: 3 for acetabular loosening, 3 for dislocation, and 1 for recurrence of pseudotumor (p=0.1). None of 48 revisions in Group 1 had acetabular loosening, compared to 4 of 32 in Group 2 (p=0.02). Two patients had nerve injury in Group 2, compared to none in Group 1 (p=0.16). The mean WOMAC pain score was 87.1 of 100 and the function score was 88.4 of 100 in Group 1, compared to a mean WOMAC pain score of 78 of 100 (p=0.6) and a function score of 83 of 100 in Group 2 (p=0.8). Modification of the surgical techniques described in the introduction has resulted in a significant decrease in complications in revision of large head MOM total hips. We continue to use this protocol and recommend it for these difficult cases.
Our objective is to describe our early and mid-term results with the use of a new simple primary knee prosthesis as an articulating spacer in planned two-stage management for infected knee arthroplasty. As a second objective, we compared outcomes between the group with a retained first stage and those with a complete two-stage revision. We included 47 patients (48 knees) with positive criteria for infection, with a minimum two-year follow-up, in which a two-stage approach with an articulating spacer with new implants was used. Patients with infection control, and a stable and functional knee were allowed to retain the initial first-stage components. Outcomes recorded included: infection control rate, reoperations, final range of motion (ROM), and quality of life assessment (QoL) including Western Ontario and McMaster Universities osteoarthritis index, Knee Injury and Osteoarthritis Outcome Score, Oxford Knee Score, 12-Item Short-Form Health Survey questionnaire, and University of California Los Angeles (UCLA) activity score and satisfaction score. These outcomes were evaluated and compared to additional cohorts of patients with retained first-stage interventions and those with a complete two-stage revision. Mean follow-up was 3.7 years (2.0 to 6.5).Aims
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic surgery, aiming to provide efficient, high-quality care. However, there are concerns about the safety of this practice. The purpose of this study was to examine the safety and efficacy of a model of partially overlapping surgery that we termed ‘swing room’ in the practice of primary total hip (THA) and knee arthroplasty (TKA). A retrospective review of prospectively collected data was carried out on patients who underwent primary THA and TKA between 2006 and 2017 in two academic centres. Cases were stratified as partially overlapping (swing room), in which the surgeon is in one operating room (OR) while the next patient is being prepared in another, or nonoverlapping surgery. The demographic details of the patients which were collected included operating time, length of stay (LOS), postoperative complications within six weeks of the procedure, unplanned hospital readmissions, and unplanned reoperations. Fisher's exact, Wilcoxon rank-sum tests, chi-squared tests, and logistic regression analysis were used for statistical analysis.Aims
Methods
The number of medial unicompartmental knee replacements (UKR) performed for arthritis has increased and as such, revisions to total knee replacement (TKR) is increasing. Previous studies have investigated survivorship of UKR to TKR revision and functional outcomes compared to TKR to TKR revision, but have failed to detail the surgical considerations involved in these revisions. Our objectives are to investigate the detailed surgical considerations involved in UKR to TKR revisions. This study is a retrospective comparative analysis of a prospectively collected database. From 2005 to 2017, 61 revisions of UKR to TKR were completed at a single center. Our inclusion criteria included: revision of UKR to TKR or TKR to TKR with minimum 1 year follow-up. Our exclusion criteria include: single component and liner revisions and revision for infection. The 61 UKR to TKR revisions were matched 2:1 with respect to age, ASA and BMI to a group of 122 TKR to TKR revisions. The following data was collected: indication for and time to revision, operative skin to skin surgical time, the use of specialized equipment (augment size/location, stem use), intraoperative and postoperative complications, re-operations and outcome scores (WOMAC, Oxford 12, SF 12, satisfaction score).Introduction
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic care, aiming to provide efficient, high-quality care. However, there have been concerns about the safety of this practice. The purpose of this study is to examine safety and efficacy of a model of partially overlapping surgery that we termed “the swing room” in practice in primary hip and knee arthroplasty. A retrospective review of prospectively collected data using an administrative database was carried out on patients who underwent primary unilateral total hip and total knee arthroplasty from 2006 to 2017 at two sites of one academic center staffed by four arthroplasty surgeons. All revisions and bilateral primary procedures were excluded. Cases were stratified as overlapping or non-overlapping. Overlapping was defined when a surgeon had access to two operating rooms with two teams, and non-overlapping was defined as when a surgeon only had access to a single operating room on a particular day. Patient demographic characteristics, operating room time, procedure time, length of stay, Postoperative complications within 30 days of the procedure, unplanned hospital readmissions, unplanned reoperations, and emergency department visits were collected. The Fisher's exact Wilcoxon rank-sum test and logistic regression analysis were used for statistical analysis.Introduction
Methods
The aim of this study was to determine if the Oxford Knee and Hip Score (OKHS) can accurately predict when a primary knee or hip referral is deemed nonsurgical We retrospectively reviewed pre-consultation OKHS for all consecutive primary total knee arthroplasty (TKA) and total hip arthroplasty (THA) consultations of a single surgeon over three years. The 1436 knees (1016 patients) and 478 hips (388 patients) included were categorized based on the surgeon’s decision into those offered surgery during the first consultation Aims
Patients and Methods
Fully constrained liners are used to treat recurrent dislocations or patients at high risk after total hip replacements. However, they can cause significant morbidities including recurrent dislocations, infections, aseptic loosening and fractures. We examine long term results of 111 patients with tripolar constrained components to assess their redislocation and failure rate. The purpose of this study was to assess survivorship, complications and functional outcomes at a minimum 10 years after the constrained tripolar liners used in our institute.Background
Questions/purposes
High complication rates and poor outcomes have been widely reported in patients undergoing revision of large head metal-on-metal arthroplasty. A previous study from our center showed high rates of dislocation, nerve injury, early cup loosening and pseudotumor recurrence. After noting these issues, we implemented the following changes in surgical protocol in all large head MOM revisions: 1. Use of highly porous shells in all cases 2. Use of largest femoral head possible 3. Low threshold for use of dual mobility and constrained liners when abductors affected or absent posterior capsule 4. Use of ceramic head with titanium sleeve in all cases 5. Partial resection of pseudotumor adjacent to sciatic and femoral nerves. The purpose of the present study is to compare the new surgical protocol above to our previously reported early complications in this group of patients We specifically looked at (1) complications including reoperations; (2) radiologic outcomes; and (3) functional outcomes. Complication rates after (Group 1), and before (Group 2) modified surgical protocol were compared using Chi-square test, assuming statistical significance p<0.05.Background
Questions/purposes
Pseudotumor or high grade ALVAL (aseptic lymphocyte-dominated vasculitis-associated lesion) development around total hip replacements secondary to local metal debris generation – especially in the setting of metal-on-metal bearings – is a well-recognised histopathologic phenomenon. Recent work has suggested a strong genetic correlation with an individual's risk of such lesion development following metal exposure. Emerging data have highlighted a similar potential concern around total knee replacements (TKRs), particularly with increasing construct modularity. To date, the body-of-knowledge pertaining to TKR-associated ALVALs has largely been limited to individual case reports or small retrospective case series’, with no large-scale investigation looking at this potential complication. This study sought to establish the preliminary prevalence of pseudotumor or high grade ALVAL formation seen at the revision of primary TKRs and to establish the correlation between histologic ALVAL grade and patient-reported functional outcomes. Findings of 1263 consecutive patients undergoing revision knee surgery, at a high-volume referral center, were reviewed. 161 cases of active infection were excluded. Complete histopathology reports were subsequently available for 321 (29.1%) non-infective cases. Each case was independently histologically-classified using a previously validated scoring system reflecting ALVAL grade. Post-operative patient-reported outcomes measures (PROMs) were available for 134 patients (41.7%), allowing direct correlation between functional performance and the established histopathology results.Background
Methods
Delay in access to primary total hip (THA) arthroplasty continues to pose a substantial burden to patients and society in publicly funded healthcare systems. The majority of strategies to decrease wait times have focused on the time from surgical consult to surgery, however a large proportion of total wait time for these patients is the time from primary care referral to surgical consultation. Prioritization scoring tools and patient reported outcome measures are being used in an attempt to ration limited resources in the face of increasing demand. However, to our knowledge, no study has investigated whether a referral Oxford Hip Score (OHS) could be used to triage non-surgical referrals appropriately, in an effort to increase timely access to specialists for patients that are candidates for total joint replacement (TJR). 1) To determine if a referral OHS has the predictive ability to discriminate when a hip patient will be deemed surgical versus conservative by the surgeon during their first consultation 2) To identify an OHS cut-off point that can be used to accurately predict when a primary THA referral will be deemed conservative by the consultant surgeon during the first consultation.Background
Purpose
Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients.Aims
Patients and Methods
Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
Non-modular tapered fluted, titanium stems are
available for use in femoral revision. The combination of taper
and flutes on the stem provides axial and rotational stability,
respectively. The material and surface properties of the stem promotes
bone on-growth. If the surgeon is confident and reasonably experienced
in the surgical use of this sort of design and the case is relatively
straightforward, a non-modular design is effective. It also potentially reduces
implant inventory, and circumvents the potential problems of taper
junction corrosion and fatigue fracture. There are reports of excellent
survival, good clinical and functional results and evidence of subsequent
increase in proximal bone stock. Cite this article: Bo
Tapered, fluted, modular, titanium stems are
increasingly popular in the operative management of Vancouver B2 and
selected B3 peri-prosthetic femoral fractures. We have reviewed
the results at our institution looking at stem survival and clinical
outcomes and compared this with reported outcomes in the literature.
Stem survival at a mean of 54 months was 96% in our series and 97%
for combined published cases. Review of radiology showed maintenance
or improvement of bone stock in 89% of cases with high rates of
femoral union. Favourable clinical outcome scores have reported
by several authors. No difference in survival or clinical scores
was observed between B2 and B3 fractures. Tapered stems are a useful
option in revision for femoral fracture across the spectrum of femoral
bone deficiency. Cite this article:
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in tapered stems despite
worse pre-operative femoral deficiency. Tapered stems have proved
to be a useful alternative in revision total hip arthroplasty across
the spectrum of femoral bone deficiency.
Antibiotic impregnated articulating spacers are
used in two-stage revision total knee arthroplasty to deliver local antibiotic
therapy while preserving function. We have observed infection control
in greater than 95% of cases with functional outcomes approaching
those seen in revision for aseptic loosening. Higher failure has
been observed with methicillin resistant organisms.
There is a postulated association between increased serum metal ions and pseudotumour formation in patients with metal-on-metal hip replacements. The primary aim of this study was to assess the prevalence of pseudotumour in 31 asymptomatic patients with a large femoral head (LFH) metal-on-metal hip implant. This was compared to the prevalence of pseudotumour in 20 matched asymptomatic patients with a hip resurfacing (HRA) and 24 matched asymptomatic patients with a standard metal-on-polyethylene (MOP) total hip. A secondary objective was to assess possible correlation between increased serum metal ions and pseudotumour formation Ultrasound examination of the three groups was performed at a minimum follow up of two years. Serum metal ions were measured in the metal-on-metal LFH and HRA groups at a minimum of two years.Purpose
Method
Dislocation after revision total hip is a common complication. The purpose of this study was to assess whether a large femoral head (36/40mm) would result in a decreased dislocation rate compared to a standard head (32mm). A randomized clinical trial was undertaken to assess the effect of large femoral heads on dislocation after revision total hip. Patients undergoing revision hip arthroplasty at seven centers were randomized to 32mm head or 36/40mm head. Patients were stratified according to surgeon. Primary endpoint was dislocation. Rates were compared with Fishers exact test. Secondary outcome measures were quality of life: WOMAC, SF-36 and satisfaction. One hundred eighty four patients were randomized: 92 in the 32mm head group and 92 in the large head group. Baseline demographics were similar in the two groups. Patients were followed from two to five years postoperativelyPurpose
Method
Measurements of patellar kinematics are essential to investigate the link between anterior knee pain following knee arthroplasty and patellar maltracking. A major challenge in studying the patellofemoral (PF) joint postoperatively is that the patellar component is only partially visible in the sagittal and close-to-sagittal radiographs. The narrow angular distance between these radiographs makes the application of conventional bi-planar fluoroscopy impossible. In this study a methodology has been introduced and validated for accurate estimation of the 3D kinematics of the PF joint post-arthroplasty using a novel multi-planar fluoroscopy approach. An optoelectronic camera (Optotrak Certus) was used to track the motion of an ISO-C fluoroscopy C-arm (Siemens Siremobil) using two sets of markers attached to the X-ray source and detector housings. The C-arm was used in the Digital Radiography (DR) mode, which resembles an ordinary X-ray fluoroscopy image. A previously-developed technique (Cho et al., 2005; Daly et al., 2008) was adapted to find the geometric parameters of the imaging system. Thirty-eight DRs of the calibration phantom were obtained for the 190 of rotation of the C-arm at 5 rotational increments while data from motion markers were recorded continuously at a frequency of 100 Hz. A total knee replacement prosthesis was implanted on an artificial bone model of the knee, and the implant components and bones were rigidly fixed in place using a urethane rigid foam. For the purpose of validation, positions of the implant components were determined using a coordinate measuring machine (CMM). Sagittal and obliquely sagittal radiographs of the model were taken where the patellar component was most visible. For each DR the geometric parameters of the system were interpolated based on the location of the motion markers. The exact location of the projection was then determined in 3D space. JointTrack Bi-plane software (Dr. Scott Banks, University of Florida, Gainesville) was used to conduct 2D-3D registration between the radiographs and the reverse-engineered models of the implant components. Results of the registration were directly compared to the ground-truth obtained from the CMM to calculate the accuracies.Purpose
Method
The outcome following isolated liner exchange for revision knee arthroplasty, while an attractive option for its simplicity, has a mixed outcome reported in the literature. We report our experience in patients who had a minimum of two years follow-up. From our database we identified 44 cases in 41 patients who had an isolated liner exchange for a failed primary knee replacement. Twenty were female and the mean body mass index was 33 (range 20 to 49). The mean time to revision from their index procedure was 76 months (range 8 to 152). The mean age at revision was 69 years (range 45 to 90). Patients were assessed by use of validated quality of life questionnaires: Oxford-12, UCLA Activity Level, WOMAC and SF-12. Patients current scores were compared to pre-operative scores in 19 cases. Radiographs were assessed for polyethylene wear, osteolysis and alignment with respect to the mechanical axis. Intra-operative findings were compared to radiographic findings.Purpose
Method
The purpose of this study was to compare and evaluate the cost-effectiveness of the MIS Anterolateral Approach to that of the MIS Posterolateral and MIS Direct Lateral Approach. A prospective randomized control trial was designed and conducted to compare the MIS Anterolateral Approach to that of the MIS Posterolateral and MIS Direct Lateral Approach. Contemporary methods for economic evaluation were used to ascertain direct and indirect costs (in Canadian dollars) along with clinical effectiveness outcomes (SF6D and Pat5D utility measures). University and hospital ethics was obtained and patients were recruited and consented to participate in the RCT resulting in the assignment of 130 patients MIS hip arthroplasty procedures. Baseline patient demographics, comorbidity, quality of life, and utility were obtained for all patients. In-hospital costing data was obtained including operating room and patient room costs as well as medication, rehab and complications. Post-discharge costs were calculated from direct and indirect costs of medication, rehab, medical costs and complications until one year post-operatively. Clinical effectiveness measures were administered at intervals until one year post-operatively.Purpose
Method
computed tomography absorptiometry (CT-OAM) which uses maximum intensity projections to assesses peak density values within subchondral bone, and our novel computed tomography topographic mapping of subchondral density (CT-TOMASD) technique, which uses surface projections to assess both cortical and trabecular bone density at specific depths from the subchondral surface. Average BMD at normalized depths of 0–2.5mm, 2.5–5.0mm, and 5.0–10mm from the surface were assessed using CT-TomasD. Regional analyses were performed consisting of:
medial/lateral (M/L) BMD ratio, and BMD of a 10mm diameter core identified as having the maximum regional BMD. Each bone was assessed for OA using a modified-KL scoring system: Normal (mKL=0); Early-OA (1–2); and Late-OA (3–4).
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
Revision of a failed acetabular reconstruction in total hip arthroplasty (THA) can be challenging when associated with significant bone loss. In cementless revision THA, achieving initial implant stability and maximising host bone contact is key to the success of reconstruction. Porous tantalum acetabular shells may represent an improvement from conventional porous coated uncemented cups in revision acetabular reconstruction associated with severe acetabular bone defects.
determine predictors of pain, function and activity level 1–2 years after revision hip arthroplasty and define quality of life outcomes after revision total hip replacement.
When considering WOMAC pain as an outcome variable, factors predictive of improving category outcome included baseline WOMAC function (p= 0.001), age between 60–70 (p<
0.004), male gender (p= 0.005), lower Charnley class (p<
0.001) and no previous revisions (p <
0.023). Baseline WOMAC pain did not predict final pain outcome. Baseline WOMAC function (p=0.001), the indication for the operation (p=0.007), and the operating surgeon were significant predictors of UCLA activity at follow up. Peri or post-operative complications were not an adverse predictor of physical function, pain or activity.
A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score. There was a significant improvement ( Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.
We studied various aspects of graft impaction and penetration of cement in an experimental model. Cancellous bone was removed proximally and local diaphyseal lytic defects were simulated in six human cadaver femora. After impaction grafting the specimens were sectioned and prepared for histomorphometric analysis. The porosity of the graft was lowest in Gruen zone 4 (52%) and highest in Gruen zone 1 (76%). At the levels of Gruen zones 6 and 2 the entire cross-section was almost filled with cement. Cement sometimes reached the endosteal surface in other Gruen zones. The mean peak impaction forces exerted with the impactors were negatively correlated with the porosity of the graft.
Revision of a total knee arthroplasty may require an extensile approach to permit a satisfactory exposure without compromising the attachment of the patellar tendon. It has been assumed that a rectus snip is a relatively benign form of release, but the effect of using this approach on function, pain and patient satisfaction is not known. From January 1997 to December 1999, 107 patients who underwent revision of total knee arthroplasty were followed up at a minimum of two years (mean 40.5 months) and assessed by the Oxford Hip Score, the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), the Short-Form (SF)-12 and patient satisfaction. Co-morbidity, surgical exposure, the Hospital for Special Surgery (HSS) knee scores and the range of movement were also used. A standard medial parapatellar approach was used in 57 patients and the rectus snip in 50. The two groups were equivalent for age, sex and co-morbidity scores. The WOMAC function, pain, stiffness and satisfaction scores demonstrated no statistical difference. The use of a rectus snip as an extensile procedure has no effect on outcome.
Removal of well-fixed, cementless, acetabular components during revision arthroplasty remains a challenging problem. Further damage to host bone may limit options for reconstruction and compromise the long-term result of the revision operation. We report the results of 31 hips with well-fixed, cementless sockets which were removed using a new cup extraction system. In all hips the socket was removed without difficulty and with minimal further bone loss.
There are few medium- and long-term data on the outcome of the use of proximal femoral structural allografts in revision hip arthroplasty. This is a study of a consecutive series of 40 proximal femoral allografts performed for failed total hip replacements using the same technique with a minimum follow-up of five years (mean 8.8 years; range 5 to 11.5 years). In all cases the stem was cemented into both the allograft and the host femur. The proximal femur of the host was resected in 37 cases. There were four early revisions (10%), two for infection, one for nonunion of the allograft-host junction, and one for allograft resorption noted at the time of revision of a failed acetabular reconstruction. Junctional nonunion was seen in three patients (8%), two of whom were managed successfully by bone grafting, and bone grafting and plating respectively. Instability was observed in four (10%). Trochanteric nonunion was seen in 18 patients (46%) and trochanteric escape in ten of these (27%). The mean Harris hip score improved from 39 to 79. Severe resorption involving the full thickness of the allograft was seen in seven patients (17.5%). This progressed rapidly and silently, but has yet to cause failure of any of the reconstructions. Profound resorption of the allograft may be related to a combination of factors, including a slow form of immune rejection, stress shielding and resorption due to mechanical disuse with solid cemented distal fixation, and the absence of any masking or protective effect which may be provided by the retention of the bivalved host bone as a vascularised onlay autograft. Although continued surveillance is warranted, the very good medium-term clinical results justify the continued use of structural allografts for failed total hip replacements with severe loss of proximal femoral bone.
The PROSTALAC functional spacer is made of antibiotic-loaded acrylic cement but has a small metal-on-polythene articular surface. We have used it as an interim spacer in two-stage exchange arthroplasty for infected total knee replacement. PROSTALAC allows continuous rehabilitation between stages as it maintains good alignment and stability of the knee and a reasonable range of movement. It also helps to maintain the soft-tissue planes, which facilitates the second-stage procedure. We reviewed 45 consecutive patients, treated over a period of nine years. The mean follow-up was for 48 months (20 to 112). At final review, there was no evidence of infection in 41 patients (91%); only one had a recurrent infection with the same organism. There was improvement in the Hospital for Special Surgery knee score between stages and at final review. The range of movement was maintained between stages. Complications were primarily related to the extensor mechanism and stability of the knee between stages. Both of these problems decreased with refinement of the design of the implant. The rate of cure of the infection in our patients was similar to that using other methods. Movement of the knee does not appear to hinder control of infection.
We describe a U-shaped approach to the distal femur which, having divided the extensor mechanism and elevated the entire quadriceps muscle, gives excellent exposure and allows a number of reconstructive options. It was used in 14 patients, 13 of whom were followed up for a mean of 3.5 years (1 to 11). There was no case of flap necrosis, and complications related to the reconstruction were acceptable.
An analysis of the cement mantle obtained with the Exeter impaction allografting system at one centre showed that it was either deficient or absent in almost 47% of Gruen zones. We therefore examined the mantle obtained using this system at another hospital and compared the results with those from the CPT and Harris Precoat Systems at other centres. The surgical indications for the procedure and the patient details were broadly similar in all four hospitals. There was some variation in the frequency of use of cortical strut allografts, cerclage wires and wire mesh to supplement the impaction allograft. Analysis of the cement mantles showed that when uncertain Gruen zones were excluded, the incidence of zones with areas of absence or deficiency of the cement was 47% and 50%, respectively, for the two centres using the Exeter system, 21% for the CPT system and 18% for the Harris Precoat system. We measured the difference in size between the proximal allograft impactors and the definitive prosthesis for each system. The Exeter system impactors are shorter than the definitive prosthesis and taper sharply so that the cavity created is inadequate, especially distally. The CPT proximal impactors are considerably longer than the definitive prosthesis and are designed to give a mantle of approximately 2 mm medially and laterally and 1.5 mm anteriorly and posteriorly. The Harris Precoat proximal impactors allow for a mantle with a circumference of 0.75 mm in the smaller sizes and 1 mm in the larger. Many reports link the longevity of a cemented implant to the adequacy of the cement mantle. For this reason, femoral impaction systems require careful design to achieve a cement mantle which is uninterrupted in its length and adequate in its thickness. Our results suggest that some current systems require modification.