The February 2024 Spine Roundup360 looks at: Surgeon assessment of bone – any good?; Robotics reduces radiation exposure in some spinal surgery; Interbody fusion cage versus anterior lumbar interbody fusion with posterior instrumentation; Is robotic-assisted pedicle screw placement an answer to the learning curve?; Acute non-traumatic spinal subarachnoid haematomas: a report of five cases and a systematic review of the literature; Is L4-L5 lateral interbody fusion safe and effective?
Aims. To determine whether half-threaded screw holes in a new titanium locking plate design can substantially decrease the notch effects of the threads and increase the plate fatigue life. Methods. Three types (I to III) of titanium locking plates were fabricated to simulate plates used in the femur, tibia, and forearm. Two copies of each were fabricated using full- and half-threaded screw holes (called A and B, respectively). The mechanical strengths of the plates were evaluated according to the American Society for Testing and Materials (ASTM) F382-14, and the screw stability was assessed by measuring the screw removal torque and bending strength. Results. The B plates had fatigue lives 11- to 16-times higher than those of the A plates. Before cyclic loading, the screw removal torques were all higher than the
This study is a prospective, non-randomized trial for the treatment of fractures of the medial malleolus using lean, bioabsorbable, rare-earth element (REE)-free, magnesium (Mg)-based biodegradable screws in the adult skeleton. A total of 20 patients with isolated, bimalleolar, or trimalleolar ankle fractures were recruited between July 2018 and October 2019. Fracture reduction was achieved through bioabsorbable Mg-based screws composed of pure Mg alloyed with zinc (Zn) and calcium (Ca) ( Mg-Zn0.45-Ca0.45, in wt.%; ZX00). Visual analogue scale (VAS) and the presence of complications (adverse events) during follow-up (12 weeks) were used to evaluate the clinical outcomes. The functional outcomes were analyzed through the range of motion (ROM) of the ankle joint and the American Orthopaedic Foot and Ankle Society (AOFAS) score. Fracture reduction and gas formation were assessed using several plane radiographs.Aims
Methods
Aims. To devise a method to quantify and optimize tightness when inserting cortical screws, based on bone characterization and screw geometry. Methods. Cortical human cadaveric diaphyseal tibiae screw holes (n = 20) underwent destructive testing to firstly establish the relationship between cortical thickness and experimental stripping torque (T. str. ), and secondly to calibrate an equation to predict T. str. Using the equation’s predictions, 3.5 mm screws were inserted (n = 66) to targeted torques representing 40% to 100% of T. str. , with recording of compression generated during tightening. Once the target torque had been achieved, immediate pullout testing was performed. Results. Cortical thickness predicted T. str. (R. 2. = 0.862; p < 0.001) as did an equation based on tensile yield stress, bone-screw friction coefficient, and screw geometries (R. 2. = 0.894; p < 0.001). Compression increased with screw tightness up to 80% of the maximum (R. 2. = 0.495; p < 0.001). Beyond 80%, further tightening generated no increase in compression. Pullout force did not change with variations in submaximal tightness beyond 40% of T. str. (R. 2. = 0.014; p = 0.175). Conclusion. Screw tightening between 70% and 80% of the predicted maximum generated optimum compression and pullout forces. Further tightening did not considerably increase compression, made no difference to pullout, and increased the risk of the screw holes being stripped. While further work is needed for development of intraoperative methods for accurate and reliable prediction of the maximum tightness for a screw, this work justifies
Nearly a quarter of screws cause damage during insertion by stripping the bone, reducing pullout strength by over 80%. Studies assessing surgically achieved tightness have predominately shown that variations between individual surgeons can lead to underpowered investigations. Further to the variables that have been previously explored, several basic aspects related to tightening screws have not been evaluated with regards to how they affect screw insertion. This study aims to identify the achieved tightness for several variables, firstly to better understand factors related to achieving optimal intraoperative screw purchase and secondly to establish improved methodologies for future studies. Two torque screwdrivers were used consecutively by two orthopaedic surgeons to insert 60 cortical, non-locking, stainless-steel screws of 3.5 mm diameter through a 3.5 mm plate, into custom-made 4 mm thick 20 PCF sheets of Sawbone, mounted on a custom-made jig. Screws were inserted to optimal tightness subjectively chosen by each surgeon. The jig was attached to a bench for vertical screw insertion, before a further 60 screws were inserted using the first torque screwdriver with the jig mounted vertically, enabling horizontal screw insertion. Following the decision to use the first screwdriver to insert the remaining screws in the vertical position for the other variables, the following test parameters were assessed with 60 screws inserted per surgeon: without gloves, double surgical gloves, single surgical gloves, non-sterile nitrile gloves and, with and then without augmented feedback (using digitally displayed real-time achieved torque). For all tests, except when augmented feedback was used, the surgeon was blinded to the
Introduction. Loosening of the baseplate is one of the most common causes of failure in Reverse Shoulder Arthroplasty. To allow osteo-integration to occur and thus provide long-term stability, initial screws fixation plays a pivotal role. In particular, tightening torque and force of nonlocking screws are two parameters that are considered to have a clear impact on implant stability, yet the relation is not fully understood. For this reason, this study aims to define an experimental set-up, to measure force and torque in artificial bone samples of different quality, in order to estimate ranges of optimal surgical values and give guidelines to maximize screw fixation and therefore initial implant stability. Methods. A custom-made torque sensor (Figure 1a) was built and calibrated using a lever deadweight system. To measure the compression force generated by the screw head, three thin FlexiForce sensors (Tekscan, South Boston, US) were enclosed between two 3D printed plates with a central hole to allow screw insertion (Figure 1b). The tightening force, represented by the sum of the three sensors, was calibrated using a uniaxial testing machine (Zwick/Roell, Ulm, Germany). Multiple screw lengths (26mm, 32mm and 47mm) were selected in the protocol. Synthetic bone blocks (Sawbones; Malmö, Sweden) of 20 and 30 PCF were used to account for bone quality variation. To evaluate the effect of a cortical bone layer, for each density three blocks were considered with 0 mm (no layer), 1.5 mm and 3 mm of laminate foam of 50 PCF. The holes for the screws were pre-drilled in the same way as in the operation room. For each combination of screw dimensions and bone quality, ten measurements were performed by acquiring the signal of the
The medial malleolus, once believed to be the primary stabilizer of the ankle, has been the topic of conflicting clinical and biomechanical data for many decades. Despite the relevant surgical anatomy being understood for almost 40 years, the optimal treatment of medial malleolar fractures remains unclear, whether the injury occurs in isolation or as part of an unstable bi- or trimalleolar fracture configuration. Traditional teaching recommends open reduction and fixation of medial malleolar fractures that are part of an unstable injury. However, there is recent evidence to suggest that nonoperative management of well-reduced fractures may result in equivalent outcomes, but without the morbidity associated with surgery. This review gives an update on the relevant anatomy and classification systems for medial malleolar fractures and an overview of the current literature regarding their management, including surgical approaches and the choice of implants. Cite this article: Abstract
Screw plugs have been reported to increase the fatigue strength of stainless steel locking plates. The objective of this study was to examine and compare this effect between stainless steel and titanium locking plates. Custom-designed locking plates with identical structures were fabricated from stainless steel and a titanium alloy. Three types of plates were compared: type I unplugged plates; type II plugged plates with a 4 Nm torque; and type III plugged plates with a 12 Nm torque. The stiffness, yield strength, and fatigue strength of the plates were investigated through a four-point bending test. Failure analyses were performed subsequently.Objectives
Methods
Introduction. Self tapping bone screw has been widely used in the fixation of Arthroplasty implants and bone graft. But the unwanted screw or driver breakage can be a direct result of excessive driving torque due to the thread cutting resistance. Previous studies showed that bone drill bit cutting rake angle was a critical factor and was inversely related to the bone cutting efficiency. 1, 2, 3, 4. (Figure 1) However to date there was no data for how the rake angle could influence the performance of self tapping bone screw. The purpose of this study was to investigate the torque generated by the self tapping cortical screw in simulated bone insertion as a function of the screw tip cutting flute rake angle. Methods. Two 5 mm thick BM5166 polyurethane block were stacked together and drilled through with 2.5mm diameter holes. Five 30mm long 3.5 mm diameter Ti6AL4V alloy self tapping cortical screws with 0°rake angle cutting flutes (Figure 2) were inserted in the holes and driven by the spanner attached to the test machine (Z5.0TN/TC-A-10) with a displacement control of 3 revolutions/min and 30N constant axial loading. The screws were driven into the stacked polyurethane block for 8mm depth. The maximum driving torque was recorded. Procedure was repeated for five same screws but with 7° rake angle cutting flutes. (Figure 2) The driving torqueses were compared. Student t test was performed with confidence level of 95% was assumed. Results. The average
Good lag screw holding power in trabecular bone of the femoral head is a requisite to achieve stability in the management of proximal femoral fractures. It has been demonstrated that
Objectives. Cement augmentation of pedicle screws could be used to improve screw stability, especially in osteoporotic vertebrae. However, little is known concerning the influence of different screw types and amount of cement applied. Therefore, the aim of this biomechanical in vitro study was to evaluate the effect of cement augmentation on the screw pull-out force in osteoporotic vertebrae, comparing different pedicle screws (solid and fenestrated) and cement volumes (0 mL, 1 mL or 3 mL). Materials and Methods. A total of 54 osteoporotic human cadaver thoracic and lumbar vertebrae were instrumented with pedicle screws (uncemented, solid cemented or fenestrated cemented) and augmented with high-viscosity PMMA cement (0 mL, 1 mL or 3 mL). The
Although atlantoaxial rotatory fixation (AARF) is a common cause
of torticollis in children, the diagnosis may be delayed. The condition
is characterised by a lack of rotation at the atlantoaxial joint
which becomes fixed in a rotated and subluxed position. The management of
children with a delayed presentation of this condition is controversial.
This is a retrospective study of a group of such children. Children who were admitted to two institutions between 1988 and
2014 with a diagnosis of AARF were included. We identified 12 children
(four boys, eight girls), with a mean age of 7.3 years (1.5 to 13.4),
in whom the duration of symptoms on presentation was at least four weeks
(four to 39). All were treated with halo traction followed by a
period of cervical immobilisation in a halo vest or a Minerva jacket.
We describe a simple modification to the halo traction that allows
the child to move their head whilst maintaining traction. The mean follow-up
was 59.6 weeks (24 to 156).Aims
Patients and Methods
Summary. A rotational limit for screw insertion may improve screw purchase and plate compression by reducing stripping, as compared to a torque based limit. Introduction. Over-tightening screws results in inadvertent stripping of 20% of cortical bone screws. The current method of “two-fingers tight” to insert screws relies on the surgeon receiving torque feedback. Torque, however, can be affected by screw pitch, bone density and bone-thread friction. An alternative method of tightening screws is the “turn-of-the-nut” model, commonly used in engineering applications. In the “turn-of-the-nut” method, nuts used to fasten a joint are rotated a specific amount in order to achieve a pre-specified bolt tension. When applied to orthopaedics, bone assumes the role of the nut and the screw is the bolt. The screw is turned a set angular rotation that is independent of torque feedback. Potentially the “turn-of-the-nut” method provides an easier way of screw insertion that might lessen inadvertent screw stripping. The purpose of the current study was to use the “turn-of-the-nut” method to determine the angular rotation that results in peak plate compression and peak screw pullout force. Methods. Three pairs of human humeri in each of three groups (osteopenic, osteoporotic, and normal) underwent plate compression and pullout protocols. For plate compression, 3.5-mm screws were tightened into strain gauge instrumented plate until screw stripping occurred.
Introduction. Femoral neck fractures are a large clinical and economical problem. One of the most common fixation options for femoral neck fractures are multiple cancellous screws. A previous clinical study has shown the lack of washers to be the single largest predictor of fixation failure in the treatment of femoral neck fractures with cancellous screws. This finding was somewhat surprising as washers do not prevent the screws from backing out and do not provide any increase resistance to varus collapse. Therefore a follow-up biomechanical study was designed to test this observation. The purpose was to evaluate the maximal
Introduction. Conventional screws achieve sufficient
Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles ( In conclusion, filling the empty combination locking/compression
holes in peri-articular distal femur locking plates at the level
of supracondylar comminution does not increase the fatigue life
of the fixation in a comminuted supracondylar femoral fracture model
(AO 33-A3) with a 6 cm gap.
In patients with osteoporosis there is always
a strong possibility that pedicle screws will loosen. This makes
it difficult to select the appropriate osteoporotic patient for
a spinal fusion. The purpose of this study was to determine the
correlation between bone mineral density (BMD) and the magnitude
of torque required to insert a pedicle screw. To accomplish this,
181 patients with degenerative disease of the lumbar spine were
studied prospectively. Each underwent dual-energy x-ray absorptiometry
(DEXA) and intra-operative measurement of the torque required to
insert each pedicle screw. The levels of torque generated in patients
with osteoporosis and osteopenia were significantly lower than those
achieved in normal patients. Positive correlations were observed between
BMD and T-value at the instrumented lumbar vertebrae, mean BMD and
mean T-value of the lumbar vertebrae, and mean BMD and mean T-value
of the proximal femur. The predictive torque (Nm) generated during pedicle
screw insertion was [-0.127 + 1.62 × (BMD at the corresponding lumbar
vertebrae)], as measured by linear regression analysis. The positive
correlation between BMD and the maximum torque required to insert
a pedicle screw suggests that pre-operative assessment of BMD may
be useful in determining the ultimate strength of fixation of a
device, as well as the number of levels that need to be fixed with
pedicle screws in patients who are suspected of having osteoporosis.
Introduction: With non-fusion stabilization surgery technique, the demands on the pedicle screw system have increased. The screw implants require a high press fit for optimum bone integration and a high pullout strength to guarantee primary stability. We compared the cylindrical screw design from a pedicle screw system with the conical design in regard to the