SMALL INSERTION TORQUE VALUES ARE NOT AN ACCURATE PREDICTOR OF LAG SCREW PULLOUT STRENGTH

European Orthopaedic Research Society (EORS) 24th Annual Meeting, 14–16 September 2016. Part 2.



Abstract

Good lag screw holding power in trabecular bone of the femoral head is a requisite to achieve stability in the management of proximal femoral fractures. It has been demonstrated that insertion torque and pullout strength of lag screw are linearly correlated. Therefore, insertion torque measurement could be a method to estimate the achieved screw purchase. Manual perception is not reliable [1], but the use of an instrumented screwdriver would make the procedure feasible.

The aim of this study was to assess the accuracy achievable using the insertion torque as predictor of lag screw purchase.

Four different screw designs (two cannulated and two solid-core screws) were investigated in this study. Each screw was inserted into a block of trabecular bone tissue following a standardised procedure designed to maximise the experimental repeatability. The blocks of trabecular tissue were extracted from human as well as bovine femora to increase the range of bone mineral density.

The prediction accuracy was evaluated by plotting pullout strength versus insertion torque, performing a linear regression analysis and calculating the difference (as percentage) between predicted and measured values.

Insertion torque showed a strong linear correlation (coefficient of determination R2: 0.95–0.99) with the pullout strength of lag screw. However the prediction error in pullout strength estimation was greater than 40% for small values of insertion torque, decreasing down to 15% when the lag screw was driven into good quality bone tissue.

Measuring insertion torque can supply quantitative information about the achieved lag screw purchase. Since screw design and insertion procedure have been shown to affect both the insertion torque and the pullout strength [2], the prediction model must be screw-specific and determined, closely simulating the clinical procedure defined by the screw manufacturer. However, the surgeon must be aware that, even under highly repeatable experimental conditions, the prediction error was found to be high when small insertion torque was measured, i.e. when the screw was driven in low quality bone tissue. Therefore, insertion torque is not reliable in evaluating lag screw purchase in the management of proximal femur fracture of osteoporotic patients.