Advertisement for orthosearch.org.uk
Results 1 - 20 of 91
Results per page:
Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 69 - 69
1 Feb 2020
Kebbach M Geier A Darowski M Krueger S Schilling C Grupp T Bader R
Full Access

Introduction. Persistent patellofemoral (PF) pain is a common postoperative complication after total knee arthroplasty (TKA). In the USA, patella resurfacing is conducted in more than 80% of primary TKAs [1], and is, therefore, an important factor during surgery. Studies have revealed that the position of the patellar component is still controversially discussed [2–4]. However, only a limited number of studies address the biomechanical impact of patellar component malalignment on PF dynamics [2]. Hence, the purpose of our present study was to analyze the effect of patellar component positioning on PF dynamics by means of musculoskeletal multibody simulation in which a detailed knee joint model resembled the loading of an unconstrained cruciate-retaining (CR) total knee replacement (TKR) with dome patella button. Material and Methods. Our musculoskeletal multibody model simulation of a dynamic squat motion bases on the SimTK data set (male, 88 years, 66.7 kg) [5] and was implemented in the multibody dynamics software SIMPACK (V9.7, Dassault Systèmes Deutschland GmbH, Gilching, Germany). The model served as a reference for our parameter analyses on the impact on the patellar surfacing, as it resembles an unconstrained CR-TKR (P.F.C. Sigma, DePuy Synthes, Warsaw, IN) while offering the opportunity for experimental validation on the basis of instrumented implant components [5]. Relevant ligaments and muscle structures were considered within the model. Muscle forces were calculated using a variant of the computed muscle control algorithm. PF and tibiofemoral (TF) joints were modeled with six degrees of freedom by implementing a polygon-contact model, enabling roll-glide kinematics. Relative to the reference model, we analyzed six patellar component alignments: superior-inferior position, mediolateral position, patella spin, patella tilt, flexion-extension and thickness. The effect of each configuration was evaluated by taking the root-mean-square error (RMSE) of the PF contact force, patellar shift and patellar tilt with respect to the reference model along knee flexion angle. Results. The analysis showed that the PF contact force was mostly affected by patellar component thickness (RMSE=440 N) as well as superior-inferior (RMSE=199 N), and mediolateral (RMSE=98 N) positioning.. PF kinematics was mostly affected by mediolateral positioning, patellar component thickness, and superior-inferior positioning. Medialization of the patellar component reduced the peak PF contact force and caused a lateral patellar shift. Discussion. Based on our findings, we conclude that malalignment in mediolateral and superior-inferior direction, tilt and thickness of patellar resurfacing are the most important intraoperative parameters to affect PF dynamics. It could be shown that the translational positioning is more critical than rotational positioning regarding PF contact force. Reported findings are in good agreement with previous experimental and clinical studies [2–4]. Our data reveal that patellar component positioning has to be aligned precisely during total knee arthroplasty to prevent postoperative complications. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 355 - 355
1 Dec 2013
Ishimaru M Shiraishi Y Hino K Onishi Y Miura H
Full Access

Introduction:. The widespread use of TKA promoted studies on kinematics after TKA, particularly of the femorotibial joint. Knee joint kinematics after TKA, including the range of motion (ROM) and the physical performance, are also influenced by the biomechanical properties of the patella. Surgeons sometimes report complications after TKA involvinganterior knee pain, patellofemoral impingement and instability. However, only few studies have focused specially on the patella. Because the patella bone is small and overlapped with the femoral component on scan images. In addition, the patellar component in TKA is made of x-ray–permeable ultra-high molecular weight polyethylene. It is impossible to radiographically determine the external contour of the patellar component precisely. No methods have been established to date to track the dynamic in vivo trajectory of the patella component. In this study, we analyzed the in vivo three-dimensional kinematics of the patellar component in TKA by applying our image matching method with image correlations. Methods:. A computed tomography (CT) and an x-ray flat panel detector system (FPD) were used. FPD-derived post-TKA x-ray images of the residual patellar bone were matched by computer simulation with the virtual simulation images created using pre-TKA CT data. For the anatomic location of the patellar component, the positions of the holes drilled for the patellar component pegs were used. This study included three patients with a mean age of 68 years (three females with right knee replacement) who had undergone TKA with the Quest Knee System and achieved a mean passive ROM of 0 to ≥ 130° after 6 or more month post-TKA. We investigated three-dimensional movements of the patellar component in six degrees of freedom (6 DOF) during squatting and kneeling. Furthermore, we simulated the three-dimensional movement of the patellar component, and we estimated and visualized the contact points between the patellar and femoral components on a three-dimensional model. Results:. Average root mean square errors of this technique with the patellar bone of a fresh-frozen pig complete knee joint have been confirmed as 0.2 mm for the translations and 0.2 degrees for the rotation. The 6 DOF analysis results showed that patellar dynamics were similar for all subjects on squatting and kneeling. For the patellar rotation during squatting, only 1 to 2 additional degrees were noted for all subjects. During kneeling, the patellar rotation noted adduction for all subjects. The patellar contact point on the femoral component gradually showed superior shift, increasing the distance with knee flexion during squatting and kneeling (Fig, 1. 2). Discussions and Conclusions:. In this study, no patellar shifts were detected in rotation or tilt during squatting, suggesting that the patellar component remained in the positions designed for early stages of flexion. And the patellar component shifted towards the lateral side during squatting. This finding suggests the idea that the patellar movement reflected the design of the Quest Knee system. This study demonstrated that the analytical method is useful for evaluating the pathologies and post-surgical conditions of the knee and other joints


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_III | Pages 278 - 278
1 Mar 2004
Weiss R Heisel C Breusch S
Full Access

Aims: The aim of the study was to determine the efþcacy of jet lavage in comparison to syringe lavage with respect to cement penetration and stability of the poly-ethylene patellar component after patellar resurfacing in total knee arthroplasty. Methods: In a cadaver study, we prepared 37 fresh frozen human patella pairs. The retropatellar bed was randomly cleaned with either jet lavage or a bladder syringe. The polyethylene component was cemented using Palacos R. For 12 patella pairs, sagittal sections were obtained at predeþned levels using a diamond saw. Mikroradiograms were digitised and analysed with respect to cement penetration. For the remaining 25 patella pairs, pullout tests were performed on patellar components using a traction-compression device. Results: Cement penetration was signiþcantly greater (P< 0.0001) in the jet lavage specimens compared to the syringe lavage specimens. The maximum force required to cause mechanical failure was signiþcantly greater (P< 0,0001) in jet lavage specimens compared to syringe lavage specimens. Conclusions: Our results support the routine use of jet lavage for cleansing the patellar bed prior to cement application in cemented patellar resurfacing


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_II | Pages 321 - 321
1 May 2010
Huten D Boyer P Bassaine M
Full Access

Purpose: Patellar complications are among the most frequent after total knee arthroplasty. Encasing the patellar piece is one way of resisting the shear forces leading to loosening. Material and Methods: We studied at more than five years the results obtained with a total knee prosthesis implanted with preservation of the posterior cruciate ligament (PCL). This prosthesis has an asymmetric encased patellar insert with a cemented central pivot. The instrumentation ensures patellar thickness. We reviewed 104 implants at more than five years. Six had been lost to follow-up. Ninety-eight implants were still in place. Results: The following complications were observed: four fractures of the upper rim with little displacement (these fractures healed and pain regressed but the insert had moved); three vertical patellar fractures with little displacement (these fractures healed; two were symptomatic temporarily); one transverse fracture of the upper pole with displacement causing a defect in active extension; eight moderate asymptomatic impactions which were visible on the lateral x-ray (modified orientation of the insert with cement fracture). There was no significant difference for functional results (pain 40.9; movement 21.9; knee score 84.3) between patients with or without a patellar complication. Discussion: Insertion of an asymmetric prosthesis increased the risk of an orientation error (two cases early in our experience). Encasing the patellar insert limits medialisation yet the centering was satisfactory (centred patella 95.2%, shift 3.6%, subluxation 1.2%). Encasing provides a peripheral wall protecting against transverse sheer forces. The lateral wall did not fracture, demonstrating its efficacy. The upper wall can fracture under the force of flexion without functional consequences. The other fractures, favoured by section of the lateral patellar wing (p< 0.05), were not treated. Moderate but certain impaction was noted in eight cases at a mean 3.5 years (1–6 years). It was due to failure of bony support under the effect of the compression forces applied on a small surface. The diameter of the encased patellar inserts was rarely more than 25 mm. Once the prosthesis is in place, the periphery of the patella is the only component articulating with the trochlea and its impaction does not cause further aggravation. This contact did not lead to pain in any patients. Conclusions: Complications observed with encased patellar components differ from the better known apposed prostheses


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_I | Pages 55 - 55
1 Jan 2004
Huten D Boyer P Bassaine M
Full Access

Purpose: Patellar complications are among the most frequent after total knee arthroplasty. Encasing the patellar piece is one way of resisting the shear forces leading to loosening. Material and methods: We studied at more than five years the results obtained with a total knee prosthesis implanted with preservation of the posterior cruciate ligament (PCL). This prosthesis has an asymmetric encased patellar insert with a cemented central pivot. The instrumentation ensures patellar thickness. We reviewed 104 implants at more than five years. Six had been lost to follow-up. Ninety-eight implants were still in place. Results: The following complications were observed: four fractures of the upper rim with little displacement (these fractures healed and pain regressed but the insert had moved); three vertical patellar fractures with little displacement (these fractures healed; two were symptomatic temporarily); one transverse fracture of the upper pole with displacement causing a defect in active extension; eight moderate asymptomatic impactions which were visible on the lateral x-ray (modified orientation of the insert with cement fracture). There was no significant difference for functional results (pain 40.9; movement 21.9; knee score 84.3) between patients with or without a patellar complication. Discussion: Insertion of an asymmetric prosthesis increased the risk of an orientation error (two cases early in our experience). Encasing the patellar insert limits medialisation yet the centering was satisfactory (centred patella 95.2%, shift 3.6%, subluxation 1.2%). Encasing provides a peripheral wall protecting against transverse sheer forces. The lateral wall did not fracture, demonstrating its efficacy. The upper wall can fracture under the force of flexion without functional consequences. The other fractures, favoured by section of the lateral patellar wing (p< 0.05), were not treated. Moderate but certain impaction was noted in eight cases at a mean 3.5 years (1–6 years). It was due to failure of bony support under the effect of the compression forces applied on a small surface. The diameter of the encased patellar inserts was rarely more than 25 mm. Once the prosthesis is in place, the periphery of the patella is the only component articulating with the trochlea and its impaction does not cause further aggravation. This contact did not lead to pain in any patients. Conclusions: Complications observed with encased patellar components differ from the better known apposed prostheses


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 83 - 83
1 Jul 2022
Dandridge O Garner A Amis A Cobb J Arkel RV
Full Access

Abstract. Patellofemoral Arthroplasty (PFA) is an alternative to TKA for patellofemoral osteoarthritis that preserves tibiofemoral compartments. It is unknown how implant positioning affects biomechanics, especially regarding the patella. This study analysed biomechanical effects of femoral and patellar component position, hypothesising femoral positioning is more important. Nine cadaveric knees were studied using a repeated-measures protocol. Knees were tested intact, then after PFA implanted in various positions: neutral (as-planned), patellar over/understuffing (±2mm), patellar tilt, patellar flexion, femoral rotation, and femoral tilt (all ±6°). Arthroplasties were implemented with CT-designed patient-specific instrumentation. Anterior femoral cuts referenced Whiteside's line and all femoral positions ensured smooth condyle-to-component transition. Knee extension moments, medial patellofemoral ligament (MPFL) length-change, and tibiofemoral and patellofemoral kinematics were measured under physiological muscle loading. Data were analysed with one-dimensional statistical parametric mapping (Bonferroni-Holm corrected). PFA changed knee function, altering extension moments (p<0.001) and patellofemoral kinematics (p<0.05), but not tibiofemoral kinematics. Patellar component positioning affected patellofemoral kinematics: over/understuffing influenced patellar anterior translation and the patellar tendon moment arm (p<0.001). Medially tilted patellar cuts produced lateral patellar tilt (p<0.001) and vice versa. A similar inverse effect occurred with extended/flexed patellar cuts, causing patellar flexion and extension (p<0.001), respectively. Of all variants, only extending the patellar cut produced near-native extension moments throughout. Conversely, the only femoral effect was MPFL length change between medially/laterally rotated components. PFA can restore native knee biomechanics. Provided anterior femoral cuts are controlled and smooth condyle-to-component transition assured, patellar position affects biomechanics more than femoral, contradicting the hypothesis


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_2 | Pages 83 - 83
1 Mar 2021
McAleese T Quinn M Graves S Clark G
Full Access

Abstract. Objectives. Patella resurfacing in primary total knee arthroplasty (TKA) remains a contentious issue. Australian rates of patellar resurfacing are 66.6%, significantly higher compared to UK rates of 8–15% and Swedish rates of 2%. Resurfacing has gained popularity in Australia since registry data has shown decreased revision rates with no increase in patellar component related complications. We present for discussion an analysis of 113,694 total knee arthroplasties using commonly implanted prostheses in the UK. Methods. We included all TKA's since the Australian register's conception on 01/09/1999 for a primary diagnosis of osteoarthritis involving the use of either the Triathlon or Duracon implant with and without patellar components. The primary outcome of the study was time to revision for Triathlon's resurfaced and non-resurfaced prosthesis compared to the Duracon's equivalent data. We also analysed the reasons for revision between the 4 groups, type of revision and complication rates. We then compared minimally stabilised and posterior stabilised prostheses. Results. The cumulative revision rate for Triathlon prostheses with resurfacing after 12 years was 3.2% (95% CI, 2.9% to 3.6%) compared to 5.6% (95% CI, 5.0% to 6.2%) without resurfacing. Duracon's equivalent data was 6.3% with resurfacing and 5.9% without resurfacing. Triathlon prosthesis with resurfacing have much lower rates of revision due to loosening, patellofemoral pain, patellar erosion compared to unresurfaced Triathlon prostheses. Conclusion. Triathlon with re-surfacing has lower revision rates regardless of age or BMI. Previous concerns regarding patellofemoral loosening, tibial wear, maltracking relate to Duracon only, indicating the importance of implant specific studies. Declaration of Interest. (b) declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research reported:I declare that there is no conflict of interest that could be perceived as prejudicing the impartiality of the research project


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_4 | Pages 127 - 127
1 Apr 2019
Yamada K Hoshino K Tawada K Inoue J
Full Access

Introduction. We have been re-evaluating patellofemoral alignment after total knee arthroplasty (TKA) by using a weight- bearing axial radiographic view after detecting patellar maltracking (lateral tilt > 5° or lateral subluxation > 5 mm) on standard non-weight-bearing axial radiographs. However, it is unclear whether the patellar component shape affects this evaluation method. Therefore, we compared 2 differently shaped components on weight-bearing axial radiographs. Methods. From 2004 to 2013, 408 TKAs were performed with the same type of posterior-stabilized total knee implant at our hospital. All patellae were resurfaced with an all-polyethylene, three-pegged component to restore original thickness. Regarding patellar component type, an 8-mm domed component was used when the patella was so thin that a 10-mm bone cut could not be performed. Otherwise, a 10-mm medialized patellar component was selected. Twenty-five knees of 25 patients, in whom patellar maltracking was noted on standard axial radiographs at the latest follow-up, were included in this study. Knees were divided into 2 groups: 15 knees received a medialized patella (group M) while 10 received a domed patella (group D). Weight-bearing axial radiographs with patients in the semi-squatting position were recorded with the method of Baldini et al. Patellar alignment (tilt and subluxation) was measured according to the method described by Gomes et al. using both standard and weight-bearing axial views. Results. Patients’ demographic data, such as age at surgery, sex, and disease were similar for both groups. The average follow-up period was significantly longer in group D than group M (5.4 years vs. 2.5 years, respectively; p = 0.0045, Mann- Whitney U-test). The lateral tilt angle decreased significantly (p < 0.0001, paired t-test) from 6.5° ± 2.8° to 1.0° ± 1.2° with weight bearing in group M. However, this parameter in group D changed from 6.7° ± 2.7° to 4.7° ± 3.0° with weight bearing; the difference was not significant. Lateral subluxation also decreased significantly (p < 0.0001, paired t-test) from 5.1 mm ± 2.4 mm to 2.5 mm ± 1.4 mm with weight bearing in group M. However, that in group D changed from 2.8 mm ± 2.7 mm to 2.4 mm ± 2.8 mm with weight bearing, and the difference was not significant. On weight-bearing views, patellar maltracking was noted in 4 knees in group D but no knees in group M. The difference was significant (p = 0.017, Fisher's exact test). One of the 21 patients with adequate patellar tracking (4.8%) and 1 of 4 patients with maltracking (25%) complained of mild anterior knee pain. Discussion. Patellar tracking on axial radiographic views improved better in group M than in group D with weight bearing. The patellofemoral contact area was maintained with a domed patella despite tilting, but not with a medialized patella. Our results indicate that the shape difference affected the degree of radiographic improvement. Thus, the weight-bearing axial radiographic view devised by Baldini et al. is useful for evaluating patellofemoral alignment after TKA, but the shape of the patellar component should be considered for result interpretation


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 156 - 156
1 Sep 2012
Fitzpatrick CK Baldwin MA Clary CW Wright A Laz PJ Rullkoetter PJ
Full Access

Complications of the patellofemoral (PF) joint remain a common cause for revision of total knee replacements. PF complications, such as patellar maltracking, subluxation, dislocation and implant failure, have been linked to femoral and patellar component alignment. Computational analyses represent an efficient method for investigating the effects of patellar and femoral component alignment and loading on output measures related to long term clinical success (i.e. kinematics, contact mechanics) and can be utilized to make direct comparisons between common patellar component design types. Prior PF alignment studies have generally involved perturbing a single alignment parameter independently, without accounting for interaction effects between multiple parameters. The objective of the current study was to determine critical alignment parameters, and combinations of parameters, in three patellar component designs, and assess whether the critical parameters were design specific. A dynamic finite element (FE) model of an implanted PF joint was applied in conjunction with a 100-trial Monte Carlo probabilistic simulation to establish relationships between alignment and loading parameters and PF kinematics, contact mechanics and internal stresses (Figure 1). Seven parameters, including femoral internal-external (I-E) alignment, patellar I-E, flexion-extension (F∗∗∗∗∗E) and adduction-abduction (A-A) rotational alignment, and patellar medial-lateral (M-L) and superior-inferior (S-I) translational alignment, as well as percentage of the quadriceps load on the vastus medialis obliquus (VMO) tendon, were perturbed in the probabilistic analysis. Ten output parameters, including 6-DOF PF kinematics, peak PF contact pressure, contact area, peak von Mises stress and M-L force due to contact, were evaluated at 80 intervals during a simulated deep knee bend. Three types of patellar component designs were assessed; a dome-compatible patellar component (dome), a medialized dome-compatible patellar component (modified dome), and an anatomic component (anatomic). Model-predicted bounds at 5 and 95% confidence levels were determined for each output parameter throughout the range of femoral flexion (Figure 2). Traditional sensitivity analysis, in addition to a previously described coupled probabilistic and principal component analysis (probabilistic-PCA) approach, were applied to determine the relative importance of alignment and loading parameters to knee mechanics in each of the three designs. The dome component demonstrated the least amount of variation in contact mechanics and internal stresses, particularly in the 30–100° flexion range, with respect to alignment and loading variability. The modified dome had substantially reduced M-L contact force when compared with the dome. The anatomic design, while wide bounds of variability were predicted, had consistently greater contact area and lowered contact pressure than the dome and modified dome designs. The anatomic design also reproduced more natural sagittal plane patellar tilt than the other components. All three designs were most sensitivity to femoral I-E alignment. Thereafter, sensitivity to component alignment was design specific; for the anatomic component, the main alignment parameter was F-E, while for the domed components it was a combination of F-E and translation (M-L and S-I) (Figure 3). Understanding the relationships and design-specific dependencies between alignment parameters can add value to surgical pre-operative planning, and may help focus instrumentation design on those alignment parameters of primary concern


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 92 - 92
1 Sep 2012
Korduba L Klein R Essner A Kester M
Full Access

INTRODUCTION. Wear and fracture of patellar components has been frequently reported as a failure mode for cemented and press-fit patellar components. Malalignment of the patellar components may cause higher contact stresses, which may lead to excessive wear, delamination, and/or component fracture. In vitro testing of the patella in a clinically relevant malaligned condition is necessary to demonstrate adequate performance of the patellar component and assess the endurance of its fixation features under severe loading conditions. The purpose of this study was to test in vitro the patellar components under malaligned conditions using a knee joint simulator. MATERIALS AND METHODS. A 6 station MTS (Eden Prairie, MN) knee joint wear simulator and Alpha Calf Fraction serum (Hyclone Labs, Logan, UT) diluted to 50% with a pH-balanced 20-mMole solution of deionized water and EDTA was used (protein level = 20 g/l) for testing. Asymmetric, all-polyethylene, patellar components with an overall construct thickness of 11 mm (Duracon®, Stryker Orthopaedics, Mahwah, NJ) were used. Appropriately sized cobalt-chrome femoral components articulated against the patellae. The patellae were cemented (Simplex, Stryker Orthopaedics, Mahwah, NJ) to delrin fixtures, which placed the patella in 10° of lateral tilt (Figure 1). This angle was chosen based off the work of Huang et al, which was one of the larger average tilt angles reported in vivo. Replicating this scenario in vitro allows for observation of the potential scenario that may occur as the femoral component maintains contact strictly on the thinner lateral edge of the patella, concentrating both the axial and shear loads on a small area of polyethylene. The loading and kinematic profiles used for testing were published previously (maximum axial load: 2450N and maximum patellofemoral angle: 54°. Variations of the loading profile were studied by evaluating the effects of heavier patients, which increased the maximum axial load to 3100N(250lb patient) and 3750N(300lb patient) (Figure 2). Lateral offset was tested to evaluate the effect of malalignment. Increments of 1mm were analyzed starting from the neutral position, eventually reaching a maximum lateral offset of 5mm. A 6-dof load cell was placed beneath the patella fixturing to capture dynamic loads (ATI, Apex, NC). The axial and medial/lateral shear loads where used to calculate the resultant medial/lateral shear force being applied to the patellar pegs. RESULTS. The results of using a heavier loading profile and increasing lateral offset are shown in Figure 3. At neutral alignment, the effect of increasing the axial load caused an increase of 10% in resultant shear force. At 5 mm of lateral offset, the increase in loading caused the shear force to increase by 16%. With each loading profile, increasing the lateral offset from 0 to 5 mm caused the resultant shear force to increase two-fold. DISCUSSION. This test model allows for an aggressive method of testing patellar implants and it includes variables to adjust for severity (lateral offset and joint reaction force). Although increasing the amount of lateral patellar offset increases the resultant shear forces, the patellar wear rates remained minimal and constant. Hence, a femoral component that has a forgiving patellar tracking may demonstrate minimal wear, even when evaluated in extremely aggressive test conditions. Note: These results are specific to the device used since the results will be dependant on the function and design of the patellar implant and patella/femur track


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_1 | Pages 132 - 132
1 Jan 2016
Fitzpatrick CK Nakamura T Niki Y Rullkoetter P
Full Access

Introduction. A large number of total knee arthroplasty (TKA) patients, particularly in Japan, India and the Middle East, exhibit anatomy with substantial proximal tibial torsion. Alignment of the tibial components with the standard anterior-posterior (A-P) axis of the tibia can result in excessive external rotation of the tibial components with respect to femoral component alignment. This in turn influences patellofemoral (PF) mechanics and forces required by the extensor mechanism. The purpose of the current study was to determine if a rotating-platform (RP) TKA design with an anatomic patellar component reduced compromise to the patellar tendon, quadriceps muscles and PF mechanics when compared to a fixed-bearing (FB) design with a standard dome-shaped patellar component. Methods. A dynamic three-dimensional finite element model of the knee joint was developed and used to simulate a deep knee bend in a patient with excessive external tibial torsion (Figure 1). Detailed description of the model has been previously published [1]. The model included femur, tibia and patellar bones, TKA components, patellar ligament, quadriceps muscles, PF ligaments, and nine primary ligaments spanning the TF joint. The model was virtually implanted with two contemporary TKA designs; a FB design with domed patella, and a RP design with anatomic patella. The FB design was implanted in two different alignment conditions; alignment to the tibial A-P axis, and optimal alignment for bone coverage. Four different loading conditions (varying internal-external (I-E) torque and A-P force) were applied to the model to simulate physiological loads during a deep knee bend. Quadriceps muscle force, patellar tendon force, and PF and TF joint forces were compared between designs. Results. The RP design demonstrated consistently lower medial-lateral (M-L) force at the PF joint than the FB design, with greater differences between designs in later flexion once the patella was engaged in the sulcus groove; root-mean-square (RMS) differences in M-L force averaged 50 N less in the RP design throughout the flexion cycle, and 70 N less after 45° flexion (Figure 2). The FB design aligned for optimal bone coverage demonstrated 15% higher M-L forces than the FB design aligned with the tibial A-P axis. RMS load required by the quadriceps muscle was 60 N lower with the RP design than the FB design throughout the cycle (Figure 2). Discussion. Comparing a RP design with an anatomic patellar component and a FB design with a domed patellar component, the RP design demonstrated lower M-L PF joint and soft-tissue extensor mechanism forces. Differences were more pronounced under conditions of high I-E torque where the RP design accommodated large relative TF rotation. Differences in FB alignment resulted in substantially different PF M-L forces; when the FB component was mal-aligned with respect to the tibial A-P axis (and the line-of-action of the patellar tendon) the resulting M-L PF force was increased. The RP design reduced the demands on the extensor mechanism and loads on the PF joint and facilitated better coverage of the resected tibial bone surface


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 73 - 73
1 Apr 2019
Fukunaga M Kawagoe Y Kajiwara T Nagamine R
Full Access

Many recent knee prostheses are designed aiming to the physiological knee kinematics on tibiofemoral joint, which means the femoral rollback and medial pivot motion. However, there have been few studies how to design a patellar component. Since patella and tibia are connected by a patellar tendon, tibiofemoral and patellofemoral motion or contact forces might affect each other. In this study, we aimed to discuss the optimal design of patellar component and simulated the knee flexion using four types of patellar shape during deep knee flexion. Our simulation model calculates the position/orientation, contact points and contact forces by inputting knee flexion angle, muscle forces and external forces. It can be separated into patellofemoral and tibiofemoral joints. On each joint, calculations are performed using the condition of point contact and force/moment equilibrium. First, patellofemoral was calculated and output patellar tendon force, and tibiofemoral was calculated with patellar tendon force as external force. Then patellofemoral was calculated again, and the calculation was repeated until the position/orientation of tibia converged. We tried four types of patellar shape, circular dome, cylinder, plate and anatomical. Femoral and tibial surfaces are created from Scorpio NRG PS (Stryker Co.). Condition of knee flexion was passive, with constant muscle forces and varying external force acting on tibia. Knee flexion angle was from 80 to 150 degrees. As a result, the internal rotation of tibia varied much by using anatomical or plate patella than dome or cylinder shape. Although patellar contact force did not change much, tibial contact balances were better on dome and cylinder patella and the medial contact forces were larger than lateral on anatomical and plate patella. Thus, the results could be divided into two types, dome/cylinder and plate/anatomical. It might be caused by the variations of patellar rotation angle were large on anatomical and plate patella, though patellar tilt angles were similar in all the cases. We have already reported that the anatomical shape of patella would contact in good medial-lateral balance when tibia moved physiologically, therefore we have predicted the anatomical patella might facilitate the physiological tibiofemoral motion. However, the results were not as we predicted. Actually our previous and this study are not in the same condition; we used a posterior-stabilized type of prosthesis, and the post and cam mechanism could not make the femur roll back during deep knee flexion. It might be better to choose dome or cylinder patella to obtain the stability of tibiofemoral joint, and to choose anatomical or plate to the mobility


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 27 - 27
1 Oct 2018
Sporer S MacLean L Burger A
Full Access

Introduction. Uncemented highly porous titanium implants have been shown to promote osseointegration, and may result in a durable construct for total knee arthroplasty (TKA). Given the mixed results of uncemented TKA, it is important to evaluate the early stability for this product. The objective of the following study was to use radiostereometric analysis (RSA) to assess early fixation of a highly porous tibial baseplate and metal backed patella. Methods. Twenty-seven patients (mean age 64 years, 30% female) undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous tibial baseplate, a metal backed patella and tantalum RSA bone markers. Implant migration was assessed using model-based RSA at 1.5, 3, 6, 12 and 24 months post-operative. Patient reported outcome measures were captured using the same follow-up schedule, and compared to pre-operative measures. Results. There were no adverse events affecting implant fixation, and no revisions. Patient function significantly improved by 3 months post-operation (p < 0.001). The highest rate of tibia and patellar component migration occurred over the first six post-operative weeks, with minimal migration thereafter. Mean maximum total point motion (MTPM) at 24 months was 0.72 (SD 0.34) mm for the tibia, and 0.44 (SD 0.25) mm for the patella. Three tibia baseplates migrated more than 1 standard deviation greater than the mean at 24 months, and also had continuous migration (> 0.2mm of MTPM) in the second post-operative year. One patellar component showed a rapid rate of migration between 6 and 24 months, whereas all other patellar components appeared to stabilize. Conclusions. Osseointegration appears to occur on the highly porous implant surface of the tibia baseplate and metal backed patella, as evidenced by implant stability. Further follow-up is required to determine if clinical loosening will manifest in the continuously migrating implants


Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 470 - 470
1 Apr 2004
Markel D
Full Access

Introduction A biomechanical model was developed to measure wear of all-polyethylene patellar components as it relates to femoral component mal-rotation. The model, based on high load and flexion activities such as stair climbing, was used to differentiate the effects of femoral mal-rotation and differing materials on a single patellar design. Methods The patellar components (Scorpio®, Stryker-Howmedica-Osteonics) were cemented onto metal fixtures and articulated against “aligned” and “mal-aligned” (six degrees internally rotated) femoral components. The patellar components were subjected to a constant force and articulated against femoral components flexing from 600 to 1200. Patellae of identical geometry, made of conventional and highly cross-linked ultra-high molecular weight polyethylene, were tested to 1x106 cycles. Following testing, patellar wear was determined by gravimetric measurement relative to soaked control specimens. Results All conventional polyethylene patellae demonstrated damage in the form of burnishing and scratching of the articular surface. The mal-aligned conventional ultra-high molecular weight patellae demonstrated increased weight loss or wear relative to the aligned components (p=.048). All rotationally mal-aligned highly cross-linked polyethylene components sustained polyethylene fracture or catastrophic failure of the cement-polyethylene construct. Conclusions Rotational mal-alignment of the femoral component will result in increased wear of polyethylene patellar components. The newer highly cross-linked materials failed to resolve this wear problem and sustained catastrophic failure when mal-aligned. Attention needs to be given to the patella-femoral articulation when implanting knee components and when developing new polyethylene as the forces in this articulation may result in polyethylene behaviour that varies dramatically from the femoraltibial articulation. In relation to the conduct of this study, one or more of the authors has received, or is likely to receive direct material benefits


Orthopaedic Proceedings
Vol. 92-B, Issue SUPP_I | Pages 124 - 124
1 Mar 2010
Rosenstein A Postak PD Greenwald S
Full Access

INTRODUCTION: Following Total Knee Arthroplasties, patellofemoral complications have shown to be responsible for approximately 50% of re-operations. Contemporary patellar designs employ both “onlay” and “inset” configurations. The latter promotes ease of placement, reduced bone removal and a heralded theoretic advantage of increased strength at the fixation interface. However, to date, no reports have compared the disassociation strengths of these two patellar component modes of fixation. The purpose of this study is to quantify the shear disassociation strength for both onlay and inset patellar fixation techniques. METHODS: Two sets of synthetic solid foam patellae were prepared using standard milling techniques for symmetrical, three-peg onlay and inset polyethylene cylinders of identical dimension. The use of synthetic bones in mechanical testing was validated in the past. The cylinders were cemented to the synthetic patellae, using standard cementing techniques. The fixation resistance of both groups was measured using an Instron Testing Machine. A compressive joint force simulating chair rise was applied perpendicular to the anterior surface of the patellar component model. A shearing displacement was then applied to the composite until patellar component disassociation. RESULTS: The mean shear strength of the onlay group was 2540 N SD 236 N, (n=7) and 3180 N SD 186 N, (n=6) for the inset group. The inset patellae was 25% (640 N) stronger than the onlay patellae, (p=0.0002, two-tailed student t-test). DISCUSSION/CONCLUSION: The results of the study demonstrated a significantly higher resistance of inset patellar fixation to shear stress compared to onlay patellar fixation. Although further in-vivo studies are indicated, the data suggests that the use of inset patella in total knee replacements may offer stronger fixation and consequently decreasing morbidity associated with patella implant loosening


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVIII | Pages 45 - 45
1 Sep 2012
Amiri S Wilson DR Masri BA Sharma G Anglin C
Full Access

Purpose. Measurements of patellar kinematics are essential to investigate the link between anterior knee pain following knee arthroplasty and patellar maltracking. A major challenge in studying the patellofemoral (PF) joint postoperatively is that the patellar component is only partially visible in the sagittal and close-to-sagittal radiographs. The narrow angular distance between these radiographs makes the application of conventional bi-planar fluoroscopy impossible. In this study a methodology has been introduced and validated for accurate estimation of the 3D kinematics of the PF joint post-arthroplasty using a novel multi-planar fluoroscopy approach. Method. An optoelectronic camera (Optotrak Certus) was used to track the motion of an ISO-C fluoroscopy C-arm (Siemens Siremobil) using two sets of markers attached to the X-ray source and detector housings. The C-arm was used in the Digital Radiography (DR) mode, which resembles an ordinary X-ray fluoroscopy image. A previously-developed technique (Cho et al., 2005; Daly et al., 2008) was adapted to find the geometric parameters of the imaging system. Thirty-eight DRs of the calibration phantom were obtained for the 190 of rotation of the C-arm at 5 rotational increments while data from motion markers were recorded continuously at a frequency of 100 Hz. A total knee replacement prosthesis was implanted on an artificial bone model of the knee, and the implant components and bones were rigidly fixed in place using a urethane rigid foam. For the purpose of validation, positions of the implant components were determined using a coordinate measuring machine (CMM). Sagittal and obliquely sagittal radiographs of the model were taken where the patellar component was most visible. For each DR the geometric parameters of the system were interpolated based on the location of the motion markers. The exact location of the projection was then determined in 3D space. JointTrack Bi-plane software (Dr. Scott Banks, University of Florida, Gainesville) was used to conduct 2D-3D registration between the radiographs and the reverse-engineered models of the implant components. Results of the registration were directly compared to the ground-truth obtained from the CMM to calculate the accuracies. Results. The accuracies for the PF were found to be 0.48 mm and 1.32 for position and orientation of the components. For the tibiofemoral joint these values were found to be 0.89 mm and 1.43, respectively. Conclusion. The multi-planar method can be used to assess the sequential kinematics of the patellofemoral and tibiofemoral joints including the mediolateral translation and tilt of the patellar component, which are obscured in standard 2D sagittal measurements and are not possible using the traditional bi-planar setup. A limitation is that it can only be used for static imaging of the joint. It has the advantage of a relatively low radiation dose. This methodology can be used to investigate the relationship between maltracking of the patella and anterior knee pain as well as other postoperative complications


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 95 - 95
1 Feb 2020
Harris A Christen B Malcorps J O'Grady C Sensiba P Vandenneucker H Huang B Cates H Hur J Marra D Kopjar B
Full Access

Introduction/Aim. Outcomes for guided motion primary total knee arthroplasty (TKA) in obese patients are unknown. Materials and Methods. 1,684 consecutive patients underwent 2,059 primary TKAs with a second-generation guided motion implant (Journey II Bi-Cruciate Stabilized Knee System, Smith & Nephew, Inc., Memphis) between 2011–2017 at three European and seven US sites. Results. Of 2,003 (97.3%) TKAs in 1,644 patients with BMI data: average age 64.5 years; 58.4% females; average BMI 32.5 kg/m. 2. ;13.4% had BMI ≥ 40 kg/m. 2. Subjects with BMI ≥ 40 kg/m. 2. had longest length of hospital stay (LOS) at European sites; LOS similar at US sites. Subjects with BMI ≥ 40 kg/m. 2. (P=0.0349) had longest surgery duration. BMI ≥ 40 kg/m. 2. had more re-hospitalizations or post-TKA reoperations than BMI < 40 kg/m. 2. (12.7% and 9.2% at five-year post-TKA, P<.0495). 62 TKAs were revised (3.39/100 TKA at five years) with no differences in revision risk between groups. 17 (27.4%) revisions involved femoral or tibial component removal; 45 (72.6%) involved tibial insert or patellar component removal only with revision risk similar between groups. Discussion. Our results corroborate literature-reported revision outcomes of standard TKAs in obese patients. Conclusion. Surgery duration and long-term complication rates are higher in patients with BMI ≥ 40 kg/m. 2. , but device revision risk is not elevated


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 534 - 534
1 Sep 2012
Heinert G Preiss S Klauser W Kendoff D Sussmann P
Full Access

Introduction. Patellar tracking in total knee replacements has been extensively studied, but little is known about patellar tracking in isolated patellofemoral replacements. We compared patellar tracking and the position of the patellar groove in the natural knee, followed by implantation of the femoral component of a PFR (patella unresurfaced) and after implantation of the femoral & patellar component of the PFR. Methods. Computer navigation was used to track the patella in eight whole lower extremities of four cadavers in the natural knee, in the same knee with the femoral component of the PFR (PFR-P) and with the femoral and patellar component of the PFR (PFR+P, patella resurfaced) (Depuy Sigma PFR). The form and position of the trochlea in the natural knee and the patellar groove of the femoral component was also analysed. Values are means+/−SD, two tailed Student's t-test for paired samples. Results. With a PFR-P the patella had a slightly more lateral tilt (0.8+/−0.8° to 2.8+/−2.5° at 40–100° of flexion, p<0.05 vs. Nat), this was more pronounced with the PFR+P (2.0+/−0.7° to 4.9+/−1.8° at 20–90° flexion, p<0.05 vs. Nat., p<0.05 vs. PFR-P at 20–80° flexion). No differences in patella rotation were seen between the three groups. In the PFR-P group the patella tracked a little more medially compared to the natural knee (0.6+/− 0.7mm to 1.3+/−2.6mm, p<0.05 at 20°,80°,90° flexion). The difference was more pronounced after patella resurfacing (PFR+P) (2.1+/−2.0mm to 3.0+/−2.2mm, p<0.05 vs. Nat. at 10°–100°, p<0.05 vs. PFR-P from 10–100°). When analysed relative to the patellar groove of the trochlea/femoral component the patella in the natural knee tracked slightly lateral to the groove (2.0+/−1.7mm to 2.9+/−2.0mm at 50–100° p<0.05), so did the patella of PFR-P (2.0+/−2.3mm to 2.3+/−2.3mm at 60–90° flexion, p<0.05), whilst the PFR+P tracked right on the groove (0.6+/−3.7mm medially to 0.6+/−2.9mm laterally, p<0.05 vs Nat at 10–30° & 70–100°, p<0.05 vs. PFR-P at 10–100°). Distance from the patellatot the epicondylar axis was slightly larger in the PFR-P group (0.6+/− 0.7mm to 1.3+/−1.4mm, p<0.05 vs. Natu at 20,80 & 90°. This was more pronounced with patellar resurfacing (2.1+/−2.0 to 3.0+/−2.2mm, p<0.05 vs. Nat at 10–100°, p<0.05 vs. PFR-P at 20–100°) The patella groove on the natural knee and the implanted femoral component of the implanted PFR had the same radius, inclination relative to the femoral mechanical axis, antero-posterior position and medio-lateral orientation. As intended by the designers the groove of the patellar component extended about 13mm further superiorly and 0.5mm more inferiorly. Discussion. The patella groove on the femoral component of the PFR reproduces the natural trochlear anatomy well. Patella tracking in the PFR-P shows only minor differences compared to the natural knee. Resurfacing of the patella in the PFR+P group causes the patella to tilt a little more laterally and track a little more medially, the distance to the epicondylar axis is slightly larger but this allows the patella to engage better in the patellar groove of the femoral component


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_I | Pages 95 - 95
1 Mar 2006
Harman M Markovich G Banks S Hodge W
Full Access

Introduction: Patellar complications after total knee arthroplasty (TKA) remain a common reason for failure. Fully congruent patellar components, with larger contact areas and a polyethylene articular surface that is free to rotate in the frontal plane (LCS design), were designed to accommodate patellar mechanics and decrease wear. However, it remains to be determined whether the perceived advantages of a mobile-bearing, fully congruent patella design are realized in-vivo. The purpose of this study is correlate wear patterns on congruent mobile-bearing patellar components with patellar mechanics that existed after TKA. Methods: Uncemented metal-backed patellar components were retrieved at revision surgery from 26 knees with meniscal bearing (23 knees) and rotating platform (3 knees) LCS mobile bearing prostheses (DePuy Orthopaedics). Mean patient age and time in-situ was 75+11 years and 11+4 years, respectively. Revision reasons included bearing wear (11), patella wear (7), instability (2), pain (3), loosening (1), osteolysis (1), and unknown (1). Polyethylene damage was assessed using optical microscopy. Articular wear area was measured using digital image analysis and the % area, location and incidence of each damage mode was calculated. Results: The average damage area on the retrieved patella occupied 69%+15% of the surface. Burnishing, delamination and scratching modes occupied the largest areas. Delamination was noted on 58% of the retrieved patellae, predominantly located in the superior-medial quadrant. Nine (35%) patellae were fractured, with the fracture plane typically oriented in the medial-lateral direction or along the lateral edge. Twently one (81%) patellae had subsurface cracks oriented along the superior-inferior axis on the extreme lateral edge and along the medial-lateral axis. None of the patellae had embedded third body debris, but the embedded superior metal pin was exposed due to extreme damage in 4 patellas. The original femoral and tibial components were left in-situ in all knees at the time of revision, such that only the polyethylene tibial and patellar articulations were exchanged. Discussion: Despite severe wear of the components, there was only a 5% incidence of osteolysis noted intra-operatively. Cyclic compressive and tensile forces during knee flexion likely caused initiation and propagation of cracks resulting in patellar bearing fracture. The delami-nation patterns on the retrieved patellae are consistent with bearing rotation into an incongruent bearing position during knee flexion, with presumably high contact stresses occurring in the delaminated superior-medial quadrant. Fully congruent mobile-bearing patella components must maintain mobility between the articular surface and metal back so that areas of incongruent contact, and associated high contact stresses and delamination, do not occur during in-vivo function


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_8 | Pages 28 - 28
1 May 2019
Thornhill T
Full Access

There is no question that at some point many TKAs will be cementless-the question is when. The advantages of cementless TKA include a shorter operative time, no need for a tourniquet, more suitability for MIS, no concern for cement extrusion, and the history of THA. The concerns for cementless TKA include the history to date with cementless TKA (tibia and metal-backed patella), variable bony substrate, surgical cut precision, cost, revision concerns, and the patella (for patella component resurfacers). Cemented total knee arthroplasty remains the gold standard and has proven to provide durable results in most patients. The early experience with cementless tibial fixation was problematic due to tibial micromotion leading to pain and loosening. Screw fixed tibial components had additional problems as portals for polyethylene debris leading to tibial osteolysis. Moreover, metal-backed patellar components were associated with a high failure rate and most surgeons began to cement all three components. Renewed interest in cementless tibial fixation is driven in part by newer materials felt to be more suitable for ingrowth and by the perceived benefit of minimally invasive surgery. One of the concerns in limited exposure total knee arthroplasty is the difficulty in preventing the extravasation of cement posteriorly. If there is evidence-based data that quad sparing non-patella everting and limited incision length facilitates rehabilitation and does not jeopardise outcome, cementless tibial fixation will be a more attractive option in some patients. An additional concern is that the tibial surface is frequently quite variable in terms of the strength of the cancellous bone. Bone cement stabilises those differences and provides a homogeneous platform for load bearing through the tibial component