Abstract
Aims: The aim of the study was to determine the efþcacy of jet lavage in comparison to syringe lavage with respect to cement penetration and stability of the poly-ethylene patellar component after patellar resurfacing in total knee arthroplasty. Methods: In a cadaver study, we prepared 37 fresh frozen human patella pairs. The retropatellar bed was randomly cleaned with either jet lavage or a bladder syringe. The polyethylene component was cemented using Palacos R. For 12 patella pairs, sagittal sections were obtained at predeþned levels using a diamond saw. Mikroradiograms were digitised and analysed with respect to cement penetration. For the remaining 25 patella pairs, pullout tests were performed on patellar components using a traction-compression device. Results: Cement penetration was signiþcantly greater (P< 0.0001) in the jet lavage specimens compared to the syringe lavage specimens. The maximum force required to cause mechanical failure was signiþcantly greater (P< 0,0001) in jet lavage specimens compared to syringe lavage specimens. Conclusions: Our results support the routine use of jet lavage for cleansing the patellar bed prior to cement application in cemented patellar resurfacing.
Theses abstracts were prepared by Professor Dr. Frantz Langlais. Correspondence should be addressed to him at EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.