Advocates of Debridement-Antibiotics-and-Implant-Retention (DAIR) in hip peri-prosthetic joint infection (PJI) argue that a procedure not disturbing a sound prosthesis-bone interface is likely to lead to better survival and functional outcome compared to revision. However, no evidence supports this. This case-control study's aims were to compare outcome of DAIRs for infected 1° total hip arthroplasty (THA) with outcomes following 1° THA and 2-stage revisions of infected 1° THAs. We retrospectively reviewed all DAIRs, performed for confirmed infected 1° THR (DAIR-Group, n=80), in our unit between 1997–2013. Data recorded included patient demographics, medical history, type of surgery and organism identified. Outcome measures included complications, mortality, implant survivorship and functional outcome using the Oxford Hip Score (OHS). Outcome was compared with 2 control groups matched for gender and age; a cohort of 1° THA (1°-THA-Group, n=120) and a cohort of 2-stage revisions for infection (2-Stage-Revision-Group, n=66).Aim
Method
Native hip joint infection can result in hip arthrosis as a complication requiring the need for subsequent arthroplasty. There is little evidence to support single or staged THA. We present the results of patients who have undergone total hip arthroplasty (THA) following hip joint sepsis in our institution. Patients receiving a THA following previous hip joint sepsis between 2003 and 2015 were identified from the operative records database. Further clinical information was collected from the medical records, including timing of infection and surgery, as well as culture results. Functional outcome (Oxford Hip Score, OHS) and patient satisfaction scores (PSS) were recorded.Introduction
Methods
The value of Debridement-Antibiotics-and-Implant-Retention (DAIR) in prosthetic-joint-infection (PJI) is still a matter of debate as most studies to-date are underpowered with variable end-points. In our, tertiary referral, bone infection unit we consider DAIR to be a suitable option in all PJIs with soundly fixed prostheses, despite chronicity. The aims of this study were to define the long-term outcome following DAIR in hip PJI and identify factors that influence it. This is a retrospective consecutive case series of DAIRs performed between 1997 and 2013. Only infected cases confirmed by established criteria were included. Data recorded included patient demographics, medical history (ASA grade, Charlson and KLICC scores), type of surgery performed (DAIR or DAIR + exchange of modular components) and organism grown. Outcome measures included complications, implant survivorship and functional outcome (Oxford Hip Score, OHS).Introduction
Methods
Debridement, antibiotics and implant retention (DAIR) is a surgical option in the treatment of prosthetic joint infection (PJI). It is thought to be most appropriate in the treatment of early (≤6 weeks post-op) PJI. Most studies to-date reporting on DAIRs in hip PJI have been underpowered by reporting on small cohorts (n= <45), or report on registry data with associated biases and limitations. In our, tertiary referral, bone infection unit we consider DAIR to be a suitable option in all cases of PJI with a soundly fixed prosthesis, with early or late presentation, especially in patients who are too elderly or infirm to undergo major surgery. Aim: To define the 10-year outcome following DAIR in hip PJI and identify factors that influence it. We retrospectively reviewed all DAIRs performed in our unit between 1997 and 2013 for hip PJI. Only infected cases confirmed by histological and microbiological criteria were included. Data recorded included patient demographics and medical history, type of surgery performed (DAIR or DAIR + exchange of modular components), organism identified and type/duration of antibiotic treatment. Outcome measures included complications, mortality rate, implant survivorship and functional outcome. 121 DAIRs were identified with mean age of 71 years (range: 33–97). 67% followed an index procedure of 1° arthroplasty. 53% included exchange of modular components. 60% of DAIRs were for early onset PJI. Isolated staphylococcus was present in 50% of cases and 25% had polymicrobial infection. At follow-up (mean:7 years, range: 0.3 – 18), 83 patients were alive; 5- and 10- year mortality rates were 15% and 35% respectively. 45% had a complication (persistence of infection: 27%, dislocation: 10%) and 40% required further surgery. Twenty hips have been revised to-date (17%). Performing a DAIR and not exchanging the modular components was associated with an almost 3× risk (risk ratio: 2.9) of subsequent implant failure (p=0.04). 10-yr implant survivorship was 80% (95%CI: 70 – 90%). Improved 10-year implant survivorship was associated with DAIR performed for early PJI (85% Vs 68%, p=0.04). Functional outcome will be discussed. DAIR is a particularly valuable option in the treatment of hip PJI, especially in the early post-operative period. Whenever possible, exchange of modular implants should be undertaken, however DAIRs are associated with increased morbidity even in early PJI. Factors that predict success of DAIR in late PJI need to be identified.
Joint preserving procedures have gained popularity in an attempt to delay arthroplasty in young, dysplastic hips. Excellent results can be achieved with peri-acetabular osteotomy (PAO) in congruent non-arthritic hips. The role of salvage procedures such as the Shelf acetabuloplasty remains undefined. This study aims to determine the long-term survival and functional outcome following Shelf acetabuloplasty and to identify factors that influence outcome. This is a retrospective, consecutive, multi-surgeon, case series from a UK referral centre. 125 Shelf procedures were performed between 1987–2013 on 117 patients for symptomatic hip dysplasia. Mean age was 33 years (15–53). Mean follow-up was 10 years (1–27). Radiographic parameters measured included pre-operative arthritis, acetabular-index, centre-edge-angle, joint congruency, subluxation and femoral sphericity. Oxford Hip and UCLA scores were collected at follow-up. Failure was defined as conversion to arthroplasty or OHS<20.Introduction
Patients/Materials & Methods
Since the introduction of 3rd generation Metal-on-Metal-Hip-Resurfacing-Arthroplasty (MoMHRA), thousands of such prostheses have been implanted worldwide in younger patients with end-stage hip osteoarthritis. However, no independent centre has reported their medium-to-long term outcome. The aim of this study is to report the ten year survival and outcome of the Birmingham Hip Resurfacing (BHR), the most commonly used MoMHRA worldwide. Since 1999, 648 BHRs were implanted in 555 patients, the majority of which were male (326). The mean age at surgery was 52.1years (range: 17–82), with primary OA as most common indication (85%). Mean follow up was 7.1years (range: 1–11). The Oxford Hip Score (OHS) and UCLA questionnaires were sent to all patients. Implant survival was established, with revision as the end point. Sub-analysis was performed by gender, femoral component size (small: <45mm, standard: 46–52, large: >53mm) and age at surgery (young:<50yrs, old:>50yrs).Introduction
Methods
Femoral neck narrowing (NN) following Metal-on-Metal Hip Resurfacing Arthroplasty (MoMHRA) is a well-recognised clinical phenomenon. The incidence of resurfaced hips with NN > 10% is reported to be up to 27%. Its pathogenesis is thought to be multi-factorial secondary to stress shielding, impingement, osteolysis secondary to wear/ion particles and as a result of reduced vascularity and pressure effect on cancellous bone secondary to the presence of a soft-tissue mass around the resurfaced hip. Recognised risk factors for its development include: female gender and the presence of a pseudotumour. Serum Chromium (Cr) and Cobalt (Co) are recognised surrogate markers of in-vivo wear of MoMHRA. The aims of this study were to establish whether NN is associated with increased wear. A cohort of 214 patients with unilateral MoMHRA (139M: 75F) was included in this study. Primary osteoarthritis was the diagnosis leading to surgery for the majority of patients (208). The average age at surgery was 54.1 years old (13–73). Six different implants were used; BHR (116), Conserve plus (92), Recap (2), ASR (2), Adept (1) and Cormet (1). The average femoral component size was 49.2mm (range: 38–59). The average follow up was 4.3 years (range: 2–10). Patients were subdivided into 3 groups as per implant size. Small size component group had implants <45mm, average size group had components 45–51mm and large component size group had components >51mm. All patients had Prosthesis-Junction-Ratio (PJR) measured from postoperative (PJRpost) and at latest follow up (PJRfollow) radiographs. Measurements were made using the method described by Lilikakis1. Metal ion levels (Cr/Co) were measured at last follow-up for all patients. Cr level >5.1g/ml and Co levels >4.4 g/ml were considered high2 and patients with such levels formed the high ion group.INTRODUCTION
METHODS
Studies have suggested that there is a reduction in head-neck-ratio (HNR) associated with MoMHRA. A reduction in HNR at operation would decrease range of movement and increase impingement risk. Impingement could lead to 20 edge loading, increasing wear. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Although acetabular component orientation has been shown to contribute to wear and PT development, the role of a decrease in HNR has only been highlighted in PT development. This study aimed to measure changes in HNR that occur at resurfacing and determine any gender- and component size-specific differences. In addition it aimed to determine whether changes in HNR could be associated with increased wear. 84 patients (56M: 28F) with unilateral MoMHRA were included. The mean age at surgery was 57 years. The mean femoral component was 49mm. Components were considered small if <45mm, average if between 45–50mm and large if >50mm. Three designs were implanted; BHR, C+ and Recap. The average follow up was 4 years. All patients had Cr/Co levels measured at follow up. Patients were considered to have high ions if Cr and Co levels were 5.1ppb and 4.4ppb respectively. Pre-operative HNR (HNRpre) and the post-operative HNR (HNRpost) were made from the respective pelvic radiographs. Assuming a 2mm thick cartilage layer, the HNR based on the diameter of the articular cartilage pre-operatively (HNRart) was calculated. The immediate changes in HNR as a result of the operation were expressed relative to articular HNR pre-op: HNRartpost=HNRpost–HNRartINTRODUCTION
METHODS
Although simulation studies have shown superior wear properties of metal-on-metal articulations, increased concern exists regarding the excess in-vivo wear of a small number of Metal-on-Metal-Hip-Resurfacing (MoMHRA) implants. Serum ion levels of Chromium (Cr) and Cobalt (Co) are surrogate markers of wear. Risk factors associated with increased wear include female gender, small components, dysplasia, cup orientation outside safe zone and femoral head downsize during surgery with an associated decrease in Head-Neck-Ratio (HNR). However, these factors are interlinked. This study aims to identify the factors that are most important for subsequent wear of MoMHRA, by performing a multivariate analysis. 206 patients (124M: 82F) with unilateral MoMHRA were included in this study. The average follow up was 3.3 years. All patients had Cr/Co levels measured at follow up. Inclination and anteversion of each cup were measured using EBRA. Cups were analysed as being within or outside the previously defined optimum-zone. HNR measurements were made from pre-operative (HNRpre) and post-operative (HNRpost) radiographs. The immediate changes in HNR (downsize/upsize of femoral head) as a result of the operation were expressed as: HNRprepost=HNRpost–HNRpre Multivariate linear regression modelling was used to explore the association between measures of ions with the following predictor variables (gender, age, diagnosis, femoral component size, orientation of the acetabular component, head/neck ratio and position of femoral stem). Analyses were carried out separately for each outcome (Cr and Co). Classification and Regression Tree (CART) models were fitted as a complimentary approach to regression modelling.INTRODUCTION
METHODS
This study aims to investigate femoral blood flow during Metal-on-Metal Hip Resurfacing (MMHR) by monitoring oxygen concentration during the operative procedure. Patients undergoing MMHR using the posterior approach were evaluated. Following division of fascia lata, a calibrated gas-measuring electrode was inserted into the femoral neck, aiming for the supero-lateral quadrant of the head. Baseline oxygen concentration levels were detected after electrode insertion 2-3cm below the femoral head surface and all intra-operative measures were referenced against these. Oxygen levels were continuously monitored throughout the operation. Data from ten patients are presented. Oxygen concentration dropped most noticeably during the surgical approach and was reduced by 62% (Std.dev +/-26%) following dislocation and capsulectomy. Insertion of implants resulted in a further oxygenation decrease by 18% (Std.dev +/-28%). The last obtained measure before wound closure detected 22% (Std.dev +/-31%) of initial baseline oxygen levels. Variation between subjects was observed and three patients demonstrated a limited recovery of oxygen levels during implant insertion and hip relocation. Intra-operative measurement of oxygen concentration in blood perfusing the femoral head is feasible. Results in ten patients undergoing MMHR showed a dramatic effect on the oxygenation in the femoral head during surgical approach and implant fixation. This may increase the risk of avascular necrosis and subsequent femoral neck fracture. Future experiments will determine if less invasive procedures or specific positioning of the limb can protect the blood supply to femoral neck and head.
Metal on metal hip resurfacing (MMHR) is a popular procedure for the treatment of osteoarthritis in young patients. Several centres have observed masses, arising from around these devices, we call these inflammatory pseudotumours. They are locally invasive and may cause massive soft tissue destruction. The aim of this study was to determine the incidence and risk factors for pseudotumours that are serious enough to require revision surgery. In out unit, 1,419 MMHRs were performed between June 1999 and November 2008. All revisions were identified, including all cases revised for pseudotumour. Pseudotumour diagnosis was made by histological examination of samples from revision. A Kaplan-Meier survival analysis was performed, Cox regression analysis was used to estimate the independent effects of different factors. The revision rate for pseudotumour increased with time and was 4% (95% CI: 2.2% to 5.8%) at eight years. Female gender was a strong risk factor: at eight years the revision rate for pseudotumours in men was 0.5% (95% CI 0% to 1.1%), in women over 40 it was 6% (95% CI 2.3% to 10.1%) and in women under 40 it was 25% (95% CI 7.3% to 42.9%) (p<
0.001). Other factors associated with an increase in revision rate were, small components (p=0.003) and dysplasia (p=0.019), whereas implant type was not (p=0.156). We recommend that resurfacings are undertaken with caution in women, especially those younger than 40 years of age, but they remain a good option in men. Further work is required to understand the patho-aetiology of pseudotumours so that this severe complication can be avoided.
8 MoMHRA implants revised due to pseudotumour; 22 MoMHRA implants revised due to other reasons of failure (femoral neck fracture and infection). The linear wear of retrieved implants was measured using a Taylor-Hobson Roundness machine. The average linear wear rate was defined as the maximum linear wear depth divided by the duration of the implant in vivo.
significantly higher median linear wear rate of the femoral component: 8.1um/year (range 2.75–25.4um/year) vs. 1.79um/year (range 0.82–4.15um/year), p=0.002; and significantly higher median linear wear rate of the acetabular component: 7.36um/year (range1.61–24.9um/year) vs. 1.28um/year (range 0.18–3.33um/year), p=0.001. Similarly, differences were also measured in absolute wear values. The median absolute linear wear was significantly higher in the pseudotumour implant group:
21.05um (range 2.74–164.80um) vs. 4.44um (range 1.50–8.80um) for the femoral component, p=0.005; and 14.87um (range 1.93–161.68um) vs. 2.51um (range 0.23–6.04um) for the acetabular component, p=0.008. Wear on the acetabular cup components in the pseudotumour group always involved the edge, indicating edge-loading of the bearing. In contrast, edge-loading was observed in only one acetabular component in the non-pseudotumour group of implants. The deepest wear was observed well within the bearing surface for the rest of the non-pseudotumour group. The difference in the incidence of edge-loading between the two groups was statistically significant (Fisher’s exact test, p=0.03).
For the whole cohort, there was no significant difference in ion levels (Cr: p=0.092. Co=0.075) between cups positioned within Z3 (n=58) versus those outside (n=46 mean). Male patients with cups within Z3 (n=27) had lower ion levels in comparison to those outside Z3, which were significantly lower for Co (p=0.049) but not Cr (p=0.084). Female patients had similar levels within and out of Z3 for both ions (Cr: p=0.83, Co: p=0.84). However, patients with cups within Z1 (n=13) had significantly lower Co (p=0.005) and Cr (p=0.001) than those outside Z1 (n=95). Interestingly, Co levels were significantly lower in Z1 (n=13) in comparison to Z2 (n=33) (p=0.048) but Cr levels were not different (p=0.06).
Roentgen Stereophotogrammetric Analysis (RSA) can predict long-term outcome of prostheses by measuring migration over time. The Exeter femoral stem is a double-tapered highly polished implant and has been shown to subside within the cement mantle in 2 year RSA studies. It has a proven track record in terms of long-term survivorship and low revision rates. Several studies have demonstrated excellent clinical outcomes following its implantation but this is the first study to assess stem migration at 10 years, using RSA. This is a single-centre study involving 20 patients (mean age: 63 years, SD=7) undergoing primary total hip replacement for degenerative osteoarthritis using the lateral (Hardinge) approach. RSA radiographs were taken with the patient bearing full weight post-operatively, at 3, 6, 12 months and at 2, 5 and 10 years follow-up. The three-dimensional migration of the Exeter femoral stem was determined. The mean Oxford Hip Score at 10 years was 43.4 (SD=4.6) and there were no revisions. The stems subsided and rotated internally during a 10-year period. The mean migrations of the head and tip of the femoral stem in all three anatomic directions (antero-posterior, medio-lateral &
supero-distal) were 0.69 mm posterior, 0.04 mm lateral and 1.67 mm distal for the head and 0.20 mm anterior, 0.02 mm lateral and 1.23 mm distal for the tip. The total migration at 10 years was 1.81 mm for the head and 1.25 mm for the tip. The Exeter femoral stem exhibits migration which is a complex combination of translation and rotation in three dimensions. Comparing our 10 year with our previous 2 year migration results, the Exeter stems show continued, but slow distal migration and internal rotation. The subsidence continues to compress the cement and bone-cement interface which maintains secure fixation in the long term.
to determine the prevalence of asymptomatic pseudotumours after MoMHRA; and to measure Co and Cr ion levels as well as lymphocyte proliferation responses to Ni, Co and Cr (the principal elements in the CoCr alloy used in MoMHRA) in MoMHRA patients with and without asymptomatic pseudotumours.
Metal Ion Levels – The presence of pseudotumour was associated with significantly higher median serum cobalt levels (9.2mg/L vs. 1.9mg/L, p<
0.001), chromium levels (12.0mg/L vs. 2.1mg/L, p<
0.001), hip aspirate cobalt levels (1182 mg/L vs. 86.2mg/L, p=0.003), and aspirate chromium levels (883mg/L vs. 114.8mg/ L, p=0.006), as well as with inferior functional scores (OHS 41 vs. 47 p<
0.001). There was no significant difference in acetabular cup inclination angle (p=0.51). Lymphocyte Reactivity: A higher incidence and level of enhanced lymphocyte reactivity to Ni (p=0.001), but not to Co or Cr (the principal elements in the CoCr alloy used in metal-on-metal hip resurfacing implants), was found in patients with MoMHRA compared to the patients without MoM implants. However, lymphocyte reactivity to Co, Cr and Ni did not significantly differ in patients with pseudotumours compared to those patients without pseudotumours. Conclusion: The prevalence of asymptomatic pseudotumours in females was high, especially in females with bilateral MoMHRA implants (30%). The patients with ‘asymptomatic’ pseudotumours were in fact mildly symptomatic. Lymphocyte reactivity to Co, Cr and Ni did not differ in patients with pseudotumour compared to those patients without pseudotumours, suggesting that systemic hypersensitivity type IV reactions, mediated by lymphocyte reactivity to these metals, is not the dominant mechanism in pathogenesis of the soft tissue pseudotumours. Furthermore, pseudotumours were not detected in those patients who had normal levels of cobalt and chromium ions. This suggests that pseudotumours do not occur if MoM articulations are well functioning. Therefore, pseudotumours are likely to be a biological consequence of the large amount of metal debris generated in vivo due to excessive wear.
Recently, a series of locally destructive soft tissue pseudotumour has been reported in patients following metal-on-metal hip resurfacing arthroplasty (MoMHRA), requiring revision surgery in a high percentage of patients. Based on the histological evidence of lymphocytic infiltration, a delayed hypersensitivity reaction to nickel (Ni), chromium (Cr) or cobalt (Co) has been suggested to play a role in its aetiology. The aim of this study was to investigate the incidence and level of hypersensitivity reaction to metals in patients with pseudotumour.
Group 1: MoMHRA patients with pseudotumours, detected on the ultrasound and confirmed with MRI (n=6, 5 F:1 M, mean age 53 years); Group 2: MoMHRA patients without pseudotumours (n=13, 7 F:6 M, mean age 55 years); and Group 3: age-matched control subjects without metal implants (n=6, 4 F:2 M, mean age 54 years). Lymphocyte transformation tests (LTT) were used to measure lymphocyte proliferation responses to metals. Peripheral blood mononuclear cells were isolated from heparinized blood samples using standard Ficoll–Hypaque® (Pharmacia). The PBMC were cultured at a cell density of 106 cells/mL. Culture was set up in the presence of either:
medium alone; nickel chloride (Sigma; 10-4M-10-6M); cobalt chloride (10-4M-10-6M); and chromium chloride (10-4M-10-6M). After 5 days of culture, cells were pulsed with [3H]-thymidine and proliferation was assessed by scintillation counting. The stimulation index (SI) was calculated by the ratio of mean counts per minute of stimulated to unstimulated cultures. A SI value of greater than 2.0 was interpreted as a positive result.
Tribological studies of hip arthroplasty suggest that larger diameter metal-on-metal (MOM) articulations would produce less wear than smaller diameter articulations. Other advantages using these large femoral heads implants include better stability with lower dislocation rates and improved range of motion. The aim of the present study was to compare chromium (Cr), cobalt (Co) and titanium (Ti) ion concentrations up to 1-year after implantation of different large diameter MOM total hip arthroplasty (THA).
Statistical group comparison revealed significant difference for Cr (p=0.006), Co (p=0.047) and Ti (p=<
0.001). With Biomet implants presenting the best results for Cr and Co and Zimmer the highest Ti level.
Metal on Metal Hip Resurfacing Arthroplasty (MoMHRA) has gained popularity due to its perceived advantages of bone conservation and relative ease of revision to a conventional THR if it fails. Known MoMHRA-associated complications include femoral neck fracture, avascular necrosis/collapse of the femoral head/neck, aseptic loosening and soft tissue responses such as ALVAL and pseudotumours. This study’s aim was to assess the functional outcome of failed MoMHRA revised to THR and compare it with a matched cohort of primary THRs.
We report on a group of 20 metal-on-metal resurfaced hips (17 patients) presenting with a soft tissue mass associated with various symptoms. We describe these masses as pseudotumours. All patients underwent plain radiography and fuller investigation with CT, MRI and ultrasound. Where samples were available, histology was performed. All patients in this series were female. Presentation was variable; the most common symptom was pain or discomfort in the hip region. Other symptoms included spontaneous dislocation, nerve palsy, an enlarging mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. Fourteen of the 20 cases (70%) have so far required revision to a conventional hip replacement and their symptoms have either settled completely or improved substantially since the revision surgery. Two of the three bilateral cases have asymptomatic pseudotumours on the opposite side. We estimate that about 1% of patients develop a pseudotumour in the first five postoperative years after a hip resurfacing. The cause of these pseudotumours is unknown and is probably multi-factorial, further work is required to define this; they may be manifestations of a metal sensitivity response. We are concerned that with time the incidence of these pseudotumours will increase.
Significant proximal femoral bone deficiency in revision hip surgery poses a considerable challenge, especially as revision hip surgery is increasing in prevalence. Many femoral prosthetic devices, which compensate for this bone loss, are available to surgeons but ideally one wants to restore viable proximal bone stock. The Oxford Universal Hip (OUH) has been designed and modified with these challenges in mind. This is atrimodular, non-locking device that allows for some rotation and subsidence in the cement. Load transfer occurs at the proximal wedge, thus avoiding proximal stress shielding. The OUH has been used extensively in both primary and revision hip surgery and is particularly useful when used in conjunction with proximal impaction bone grafting. The main aim of this study was to determine the clinical outcome of the patients. The viability and survivorship of the impacted bone graft was assessed in order to give some indication of adequate perfusion of the incorporated bone grafting. Between 1999 and 2002 the OUH, in combination with proximal impaction bone grafting, was implanted in 72 hip revisions in 69 patients. All patients were clinically evaluated using the Oxford Hip score. Histological samples of impacted bone grafting were analysed by a pathologist 2 years after implantation to assess viability and perfusion of the impacted bone graft. The Oxford Hip Score improved from a pre-operative to a post-operative score of 45 (26–58) to 24.3 (12–43) respectively (p<
0.001). The mean follow up time was 32.7 months and in that time there were no re-revisions for aseptic loosening. Histology demonstrated viable bone stock in the proximal femur where bone grafting had been impacted 2 years previously. The OUH is a versatile hip prosthesis for use in revision hip arthroplasty and is especially well suited to proximal impaction bone grafting where it is possible to restore viable bone stock
Bone stock loss secondary to debris and mechanic alin-stability presents a challenge in revision hip arthroplasty. The aim of our study is to evaluate the clinical outcome of revision hip arthroplasty using the Oxford hip prosthesis combined with impaction allografting. Between 1999 and 2002, we revised 72 hips in 69 patients using this technique (mean age 65years). Indications were aseptic loosening (56), infection (8), peri-prosthetic fracture (7) and a broken stem (1). The mean time to revision was 8.5 years (1–21years). Patients were assessed clinically and with the Oxford Hip Score (OHS) pre- and post-operatively. Fifty-seven patients also had acetabular revision. Four patients required femoral osteotomy to remove the old prosthesis. We used a mean of 1.8 (1–4) femoral heads per operation. Patients were mobilised partially weight bearing (8weeks) followed by a gradual return to full loading. Complications included peri-operative femoral fracture (6), infection (6), dislocation (10), DVT (1)and PE (2). The average blood transfusion was 1.8 units (0–9). The OHS improved from 45 (26–58) pre-operatively to 24.3 post-operatively (12–43). No hip has been re-revised for aseptic loosening at a mean follow-up of 32.7months (16–51). The Oxford hip is a trimodular prosthesis with a polished tapered metaphyseal section that is free to slide and rotate on the stem. The stem is first inserted uncemented into the diaphysis. Bone graft is impacted proximally, with mesh if necessary, and then the proximal wedge is cemented in. The wedge allows for some subsidence in the cement and creates optimal radial force transmission, which is essential for bone-remodelling stimulation and preventing proximal stress shielding. Although this is a short-term experience, we believe that the use of the tri-modular Oxford stem combined with minimal proximal impaction allografting is a reliable method of dealing with difficult revision femoral surgery. The results were comparable with a primary arthroplasty in terms of pain relief and functional results.
Oxford Unicompartmental knee arthroplasty (UKA) is now performed using a minimally invasive surgical (MIS) technique. Although early results are encouraging, the studies assessing outcome could be criticised for the restricted number of patients and limited follow-up. Aim of this study was to assess clinical outcome and prosthetic survival rate inpatients with minimally invasive Oxford medial UKA. This prospective study assessed 500 consecutive patients, who underwent cemented Oxford UKA for medial OA using MIS technique. Patients were assessed using objective and functional Knee Society Score (KSS). This study has confirmed preliminary findings that Oxford UKA using a minimally invasive approach is safe, reliable and effective.
The Charnley Elite femoral component was first introduced in 1992 as a new design variant of the original Charnley femoral component (De Puy, Leeds, UK) with modified neck and stem geometry. The original component had undergone few changes in nearly forty years and has excellent long-term results. Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA)1. Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is indicated as stems of this type are still being implanted. One hundred Charnley Elite stems, implanted in our centre between 1994 and 1997 were included in a prospective, cross-sectional follow-up study. Outcome measures include validated clinical scores (Charnley hip score, Harris hip score and Oxford hip score) and radiological scores (Gruen classification) as well as revision rates over the past 10 years. The clinical follow-up supports the RSA predictions of early failure of the Charnley Elite femoral stem.
Early migration of the new stem design was determined by Roentgen Stereophotogrammetric Analysis (RSA). Rapid early migration of a component relative to the bone, measured by RSA, is predictive of subsequent aseptic loosening for a number of femoral stems. As there was rapid early migration and rotation of the Charnley Elite stem, we predicted that the long-term results would be poor. An outcome assessment is required as stems of this type are still being implanted.
Preliminary clinical scores in the patients who had not undergone any subsequent surgery were adequate (Oxford Hip Score mean average of 23.9). Thirteen percent of radiographs analysed had evidence of loosening, giving an overall loosening rate of 14% at 8 years.
This study presents the 2 year migration results of the BHR femoral component using Roentgen Stereophoto-grammteric Analysis (RSA).
Between 1999 and 2002, we revised 72 hips in 69 patients using this technique (mean age 65years, 28 to 88). Fifty-six cases had aseptic loosening, 8 had infection (2 stages), 7 had peri-prosthetic fractures and 1 had a broken stem. The mean time to revision was 8.5years (1 to 21). Patients were assessed clinically and with the Oxford Hip Score (OHS) pre- and post-operatively. Fifty-seven patients also had acetabular revision. Four patients required femoral osteotomy to remove the old prosthesis. We used a mean of 1.8 (1 to 4) femoral heads per operation. Complications included 6 peri-operative femoral fractures diagnosed at operation and fixed successfully, 6 infections, 10 dislocations (2 were recurrent), one deep vein thrombosis, 2 pulmonary embolism and one gastrointestinal haemorrhage. The average blood transfusion was 1.8 units (0 to 9). The OHS improved from 45 (26 to 58) pre-operatively to 24.3 post-operatively (12 to 43). No hip has been re-revised for aseptic loosening at a mean follow-up of 32.7 months (16 to 51).
There is an increasing interest amongst surgeons and demand from patients for hip resurfacing. One concern regarding resurfacing is the incidence of femoral neck fracture post operatively. McMinn and Treacy report an incidence of 0.4% in their series, our finding was of an incidence of over four times as high (1.9%). We looked at our database of hip resurfacings and tried to identify the risk factors for fracture. We identified 11 fractures and compared these with 22 controls selected by choosing the cases performed by the surgeon immediately before and after the fracture case. We analysed their medical notes and x-rays. Statistical analysis was performed using a package in ™Excel. The implants were either Birmingham Hip (Midland Medical Technologies) or Cormet (Corin) resurfacings. No statistically significant correlation was found for sex, age or body mass index. We found that fracture was twice as likely in the presence of possible or probable osteopenia. We did not find that fracture was more likely to occur in patients with a previous diagnosis of Perthes, DDH, SUFE and avascular necrosis (AVN). We found patients with a superior overhang of the femoral component on the neck did not risk fracture, however we could not demonstrate that notching in itself increased the risk of fracture. There was no correlation with neck-shaft and stem-shaft angle or neck lengthening and offset and subsequent neck fracture. In 13 bilateral cases there was fracture in 3 (incidence 23%). Apart from one fracture that occurred at 18 weeks post-operatively all the others occurred before eight weeks. Five fractures occurred in patients who subsequently on histological analysis were found to have avascular necrosis. We conclude that bilateral surgery is probably unwise. That a superior overhang seems to protect against fracture as long as this is not at the expense of creating an inferior notch. Finally, we find AVN in a number of retrieved heads, what is the true incidence of AVN and does the approach adopted cause the avascular process and if so why do we see so few fractures?
Despite impressive overall results at one year, lower knee scores were associated with a surgeons ‘learning curve’. After this ‘learning curve’, increased surgical experience led to further improvement with 90% achieving an excellent result, 8% a good, 2% a fair and 0% a poor result.
Polyethylene wear remains an important cause of failure in knee replacements. Retrieval studies, simulators and simple X-ray methods produce wear data that may be inaccurate or unrealistic. We have developed an accurate RSA system for measuring wear in-vivo. Using this system we have found wear rate in a fixed bearing TKR to be about 0.1mm/yr[ Four Oxford uni-compartmental knee replacements, with excellent clinical results were studied ten years after implantation. RSA X-rays were taken in double leg stance with the knee in full extension and 15 and 30 degrees flexion. Following RSA calibration, silhouettes of the components on the stereo X-rays were extracted using a Canny edge detector and were matched to silhouettes projected using CAD models to determine the 3D component position. The average minimum thickness of the bearing was determined and was compared with the measured minimum thickness of 14 unused bearings to calculate linear penetration. The average linear penetration after average 10 years implantation (range 8.5 to 10.25 years) was 0.16 mm (SD 0.13 mm). The average penetration rate was 0.017 mm/year (SD 0.011 mm/year). The maximum linear penetration rate was 0.027 mm/year. The penetration rate is similar to that obtained in a retrieval study [
Increasing the width of the proximal section of a polished tapered stem enhances its rotational stability.